Midfoot Fusion Bolt 6.5 mm. Intramedullary fixation of the medial column of the foot.

Similar documents
Technique Guide. 6.5 mm Midfoot Fusion Bolt. For intramedullary fixation of the medial column of the foot.

6.5 mm midfoot fusion bolt

LCP Metaphyseal Plates. For extra-articular fractures.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

The Calcaneal Plate. The Synthes non-locking solution for the Calcaneus.

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

LCP DISTAL TIBIA PLATE

Distal Radius Plate 2.4/2.7 dorsal and volar

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

3.0/3.5/4.0/4.5/6.5/7.0/7.3. Cannulated Screws. Surgical Technique

Button Plate. Reinforcement for transosseous fixations.

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Wrist Fusion Instrument and Implant Set.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

VA LOCKING CALCANEAL PLATES 2.7

LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

VA LCP MEDIAL COLUMN FUSION PLATES 3.5

Surgical Technique. This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

3.5 mm LCP Extra-articular Distal Humerus Plate

LCP Periarticular Proximal Humerus Plate 3.5. The anatomic fixation system with anterolateral shaft placement.

VA-LCP Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

VA-LCP Olecranon Plates 2.7/3.5. The fracture-specific low-profile fixation system with variable angle locking technology.

ANGLED BLADE PLATES FOR ADULTS

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Superior Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

Low Bend Distal Tibia Plates

3.5 mm LCP Olecranon Plates

LCP Percutaneous Aiming System 3.5 for PHILOS. For less invasive surgery at the proximal humerus.

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

LCP Proximal Tibial Plate 4.5/5.0 with Periarticular Aiming Arm Instruments

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Femoral Neck System. Surgical Technique

Technique Guide. PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Wrist Fusion Set. Anatomic plates for total wrist fusion.

LOW PROFILE NEURO. This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

VECTRA-T SURGICAL TECHNIQUE. The Translational Anterior Cervical Palate System. This publication is not intended for distribution in the USA.

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

VA-LCP Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Mini External Fixator.

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

VA-LCP Ankle Trauma System 2.7/3.5. Our most comprehensive ankle plating system.

LCP Ulna Osteotomy System 2.7. Low profile angular stable fixation for ulna shortening osteotomies.

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Humerus Block. Discontinued December 2016 DSEM/TRM/0115/0296(1) Surgical Technique. This publication is not intended for distribution in the USA.

CSLP-Cervical Spine Locking Plate

Part of the DePuy Synthes Locking Compression Plate (LCP ) System. 3.5 mm LCP Medial Proximal Tibia Plates

VA-LCP Distal Humerus Plates 2.7/3.5. The low-profile fixation system with variable angle locking technology.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Technique Guide. LCP Posterior Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment LCP system.

2.7 mm/3.5 mm LCP Distal Fibula Plate

VA-LCP Ankle Trauma System 2.7/3.5. Our most comprehensive ankle plating system.

TSLP Thoracolumbar Spine Locking Plate

PROXIMAL FEMORAL NAIL REMOVAL SET

Olecranon Osteotomy Nail. For simple fractures and osteotomies of the olecranon.

The Locking Calcaneal Plate Instrument and Implant Sets

Periarticular Aiming Arm Instruments for LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large).

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures.

3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments

3.5 mm LCP Hook Plate

Technique Guide. VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Orthodontic Bone Anchor (OBA) System

3.5 mm LCP Clavicle Hook Plates

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

IMF Screw Set. For temporary, peri opera tive stabilisation of the occlusion in adults.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

3.5 mm LCP Distal Humerus Plates

External Distal Radius Fixator. Supplement to the 8 mm rod fixator system

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Collinear Reduction Clamp

Expert HAN. Expert Hindfoot Arthrodesis Nail.

4.5 mm LCP Medial Proximal Tibia Plates

Variable Angle LCP Volar Rim Distal Radius Plate 2.4. For fragment-specific fracture fixation with variable angle locking technology.

3.5 mm Clavicle Hook Plates

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

2.4 mm Variable Angle Locking Intercarpal Fusion System

LCP Wrist Fusion Set. Anatomic plates for total wrist fusion.

Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area.

ECD EXPANDABLE CORPECTOMY DEVICE Continuously Expandable Vertebral Body Replacement for Tumour Cases

Cannulated Angled Blade Plate 3.5 and 4.5, 90.

Technique Guide. Compact 2.0 LOCK Mandible. The locking system for the mandible.

2.4 mm Variable Angle LCP Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

The Versatile Polyaxial Solution for the Universal Spine Systems. USS II Polyaxial. Surgical Technique

Surgical Technique. Customer Service:

3.5 mm LCP Anterolateral Distal Tibia Plates

DOUBLE/TRIPLE PELVIC OSTEOTOMY PLATES For Treating Coxofemoral Joint Instability and Subluxation in Immature Dogs

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Surgical Technique 4.5/8.5MM BEAMING SYSTEM. Customer Service:

Transcription:

Midfoot Fusion Bolt 6.5 mm. Intramedullary fixation of the medial column of the foot. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

Table of Contents Introduction Midfoot Fusion Bolt B 6.5 mm 2 AO Principles 4 Indications and Contraindications 5 Surgical Technique Preoperative Planning 6 Preparation 7 Implantation 10 Supplemental Fixation 20 Implant Removal (optional) 21 Product Information Implants 22 Instruments 23 Vario Cases 25 Set Lists 26 Bibliography 29 MRI Information 30 Image intensifier control Warning This description alone does not provide sufficient background for direct use of DePuy Synthes products. Instruction by a surgeon experienced in handling these products is highly recommended. Processing, Reprocessing, Care and Maintenance For general guidelines, function control and dismantling of multi-part instruments, as well as processing guidelines for implants, please contact your local sales representative or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance For general information about reprocessing, care and maintenance of Synthes reusable devices, instrument trays and cases, as well as processing of Synthes non-sterile implants, please consult the Important Information leaflet (SE_023827) or refer to: http://emea.depuysynthes.com/hcp/reprocessing-care-maintenance Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 1

Midfoot Fusion Bolt B 6.5 mm. Intramedullary fixation of the medial column of the foot. Overview The Midfoot Fusion Bolt is a solid 6.5 mm intramedullary implant that can be used to fuse the medial metatarsocuneiform, naviculocuneiform and talonavicular joints. Alternative uses of the midfoot fusion bolt may also be the possibility of fusing the lateral column, calcaneocuboid and 4th metatarsocuboid joint. The midfoot fusion bolt aims to achieve permanent fusion of these joints in patients suffering from gross instability such as Charcot neuroarthropathy with or without collapse of the midfoot. The midfoot fusion bolt provides stabilization and alignment, enhancing the possibility of a functional foot and increasing the likelihood of limb salvage. Features and Benefits Identical pitch of head and tip threads (2.7 mm) Single cutting head thread Allows easy insertion and countersinking in cortical bone. Solid 6.5 mm diameter shaft For added strength in bending and shear 2 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

Functional principle of threads in Midfoot Fusion Bolt The 6.5 mm Midfoot Fusion Bolt is: Solid bolt Available in Titanium and Stainless Steel Available in lengths 50 160 mm Highly resistant to bending and shear Headless design Complete countersinking Able to achieve compression 1 MFB insertion Insertion of the MFB with the compression T-handle Blunt tip Prevents damage to soft tissues if over-inserted. 2 Closure of gap and compression As the end of the compression T-handle comes into contact with the bone it acts like the head of a screw and allows for closure of the gap by compression Leading threads and shaft have the same diameter Easy insertion with reduced risk of threads cutting out of the bone 3 Countersinking Once the desired degree of compression is reached the head is countersunk by holding the T-handle and turning the bolt into the bone with the screwdriver Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 3

AO Principles In 1958, the AO formulated four basic principles, which have become the guidelines for internal fixation. 1, 2 Anatomic reduction Fracture reduction and fixation to restore anatomical relationships. 1 2 Stable fixation Fracture fixation providing absolute or relative stability, as required by the patient, the injury, and the personality of the fracture. Early, active mobilization Early and safe mobilization and rehabilitation of the injured part and the patient as a whole. 4 3 Preservation of blood supply Preservation of the blood supply to soft tissues and bone by gentle reduction techniques and careful handling. Copyright 2007 by AO Foundation 1 Mü ller ME, M Allgöwer, R Schneider, H Willenegger. Manual of Internal Fixation. 3rd ed. Berlin Heidelberg New York: Springer. 1991. 2 Rü edi TP, RE Buckley, CG Moran. AO Principles of Fracture Management. 2nd ed. Stuttgart, New York: Thieme. 2007. 4 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

Indications and Contraindications Indications Treatment of gross collapse of the mid/hindfoot associated with: Charcot foot Neuropathic disease Charcot foot can develop in several areas of the foot. The most common is the Lisfranc area, secondarily the Chopart area, thirdly the ankle area and fourthly the calcaneus subtalar area. It often develops in the forefoot. The idea of the Midfoot Fusion Bolt is to treat the Lisfranc and Chopart area. Contraindications Peripheral vascular disease Serious and life-threatening co-morbidities Warning: The Midfoot Fusion Bolt is not to be used as a stand alone implant and needs to be used with supplemental fixation, such as additional screws and plates across the arthrodesed joints. Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 5

Preoperative Planning Evaluation of X rays For good preoperative planning it is important to have the following X rays: Weight bearing in 3 projections: Lateral, AP; 45º oblique AP AP of ankle Comparative X rays of both sides Preoperative examination and evaluation Estimate the balance of the foot and the function of the vital tendons, especially the length of the tendo Achilles and gastrocnemius. A gastrocnemius slide or percutaneous tendo Achilles lengthening is needed as equinus contracture causes abnormally high stresses across the midfoot. 6 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

Preparation 1 Patient Positioning Place the patient in the supine position, with a roll under the affected hip to position the foot in neutral (toes straight up in a resting position) 2 Gastrocnemius slide or percutaneous tendo Achilles lengthening In almost all cases, a gastrocnemius slide or percutaneous tendo Achilles lengthening is needed to relieve stresses across the midfoot. Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 7

Preparation 3 Surgical Approach Two or three incisions are usually needed to expose and prepare the joints to be fused. Make a medial utility incision to expose and denude the talonavicular, navicularcuneiform and tarsometatarsal joints. Take care to avoid the tibialis anterior tendon Make a dorsal straight incision over the 1st metatarsal phalangeal joint (MPJ) to expose the joint. This allows for access of the drill bit and the midfoot fusion bolt through the 1st metatarsal articular surface. Note: At this stage of the procedure a subtalar fusion may be performed. 8 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

4 Prepare joints Instrument 292.200 K-Wire B 2.0 mm with trocar tip, length 150 mm Expose and prepare all the joints for fusion. Correct deformities with resections, where necessary. These corrections should be performed to the estimated final shape of the foot. Place temporary K-wires to hold the joints in place, taking care not to place them in the path of the final implant. Note: If desired add graft before provisionally fixing the joints with K-wires. 5 Prepare metatarsal Access the 1st metatarsal head through the dorsal incision. Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 9

Implantation 1a Insert guide wire Instruments 292.680 Guide Wire B 2.8 mm 312.080 Drill Sleeve 8.5/2.8 312.050 Protection Sleeve 12.0/8.5 Insert the 2.8 mm guide wire under image intensifier control, through the upper center of the first metatarsal head close to the dorsal cortex and across the tarsometatarsal, the navicularcuneiform and the talonavicular joints into the talus. Place the guide wire to the required depth. Note: Measurement is to the tip of the guide wire. 11 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

1b Insert guide wire for alternative/reverse technique Instruments 292.680 Guide Wire B 2.8 mm 312.080 Drill Sleeve 8.5/2.8 312.050 Protection Sleeve 12.0/8.5 For the alternative technique insert the 2.8 mm guide wire under image intensifier control, through the upper center of the first metatarsal head close to the dorsal cortex and straight across the tarsometatarsal, the navicularcuneiform and the talonavicular joints into the talus. Continue inserting the guide wire through the posterior of the talus making sure it exits posterior avoiding the subtalar joint. From the back of the foot re-adjust the guide wire position in the metatarsal head to allow measurement of the desired length of the midfoot fusion bolt. Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 11

Implantation 2a Measure for bolt length Instrument 03.227.030 Direct Measuring Device for HCS-Headless Compression Screw B 6.5 mm Place the measuring device over the guide wire to the surface of the metatarsal head. Definite bolt length should consider joint spaces, depth of countersink and desired position of the leading tip of the bolt. 2b Measure for bolt length for alternative/reverse technique Instruments 03.227.030 Direct Measuring Device for HCS-Headless Compression Screw B 6.5 mm Place the measuring device over the guide wire to the bone. Definite bolt length should consider joint spaces, depth of countersink and desired position of the leading tip of the bolt. 11 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

3a Drill Instruments 03.111.002 Drill Bit B 5.0 mm, calibrated, cannulated, length 300/250 mm, 3-flute, for Quick Coupling 357.047 Drill Bit B 6.5 mm, cannulated, length 330 mm, for Quick Coupling When the reduction is complete insert the 5.0 mm drill bit over the guide wire. Under image intensification drill to the final desired position of the bolt. Note: The length of the bolt can also be checked with the calibrated measurements on the 5.0 mm drill bit. Verify under image intensification that the drill bit follows the correct path and does not penetrate the end of the talus. It is recommended to over-drill especially in hard bone. Do not drill the entire length, as this will prevent purchase of the leading bolt threads. 6.5 mm drill bit 5.0 mm drill bit Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 11

Implantation 3b Drill for Alternative/reverse technique Instruments 03.111.002 Drill Bit B 5.0 mm, calibrated, cannulated, length 300/250 mm, 3-flute, for Quick Coupling 357.047 Drill Bit B 6.5 mm, cannulated, length 330 mm, for Quick Coupling When the reduction is complete insert the 5.0 mm drill bit over the guide wire. Under image intensification drill to the final desired position of the bolt. Note: The length of the bolt can also be checked with the calibrated measurements on the 5 mm drill bit. Verify under image intensification that the drill bit follows the correct path. It is recommended to over-drill especially in hard bone. Do not drill the entire length, as this will prevent purchase of the leading bolt threads. 6.5 mm drill bit 5.0 mm drill bit 11 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

4 Remove guide wire Remove the guide wire to allow for the insertion of the solid midfoot fusion bolt. If stability is not maintained, add supplemental K-wires. Pay attention that the wires are not in the path of the bolt. 5 Select the midfoot fusion bolt Select the midfoot fusion bolt. Definite bolt length should consider, joint spaces, depth of countersink and desired position of the leading tip of the bolt (step 6). It is important that the midfoot fusion bolt is well inserted into the body of the talus. Note: If there is excessive joint space, this should be considered when choosing the bolt. 6 Attach the midfoot fusion bolt to T-handle Instrument 03.111.001 T-Handle for Midfoot fusion bolt Connect the correct length midfoot fusion bolt to the T-handle and finger tighten. Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 11

Implantation 7a Insert the midfoot fusion bolt Insert the midfoot fusion bolt into metatarsal head. Use image intensification to visualize and control bolt insertion. Place allograft or autograft in the prepared joints, if required. Turn the T-handle until it reaches the cartilage of the joint. 11 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

7b Insert the midfoot fusion bolt for alternative/reverse technique Insert the midfoot fusion bolt into the talus. Use image intensification to visualize and control bolt insertion. Place allograft or autograft in the prepared joints, if required. Turn the T-handle until it reaches the posterior aspect of the talus. 8a Compress joints Under image intensification turn the T-handle until the desired compression is achieved. Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 11

Implantation 8b Compress joints for alternative/reverse technique Under image intensification turn the T-handle until the desired compression is achieved. 9a Countersink the midfoot fusion bolt Instrument 314.164 Screwdriver Stardrive 4.5/5.0, T25 Place the Stardrive screwdriver through the cannulation in the T-handle and engage the screwdriver into the head of the bolt. Hold the T-handle stationary and use the screwdriver to countersink the threaded head of the midfoot fusion bolt into the metatarsal. 11 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

Note: When the groove in the screwdriver is in line with the top end of the T-handle, this indicates that the top end of the thread is even with the bone surface. Check the length of the MFB, making sure that it is well inserted into the body of the talus. Precaution: If the bolt is short, there is a risk of cut-out along the plantar aspect of the distal talus when full weight bearing starts. 9b Countersink the midfoot fusion bolt for alternative/ reverse technique Instrument 314.164 Screwdriver Stardrive 4.5/5.0, T25 Place the Stardrive screwdriver through the cannulation in the T-handle and engage the screwdriver into the head of the bolt. Hold the T-handle stationary and use the screwdriver to countersink the threaded head of the midfoot fusion bolt into the posterior of the talus. Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 11

Supplemental Fixation The Midfoot Fusion Bolt is to be used with supplemental fixation, such as additional screws and plates across the arthrodesed joints. Synthes offers a wide selection of plates and screws which cover a correspondingly extensive variety of indications. For this reason this technical guide does not cover specific indications or the selection of a plate type for specific clinical situations. For treatment of these subjects, please refer to AO Principles of Fracture management, courses offered by the AO (www.aofoundation.org), and the corresponding professional literature. Example Example 22 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

Implant Removal (optional) Clear the Stardrive recess at the end of the bolt from any tissue ingrowth. Screw out the bolt until the head is no longer engaging the bone. Attach T-handle and place the Stardrive screwdriver through the cannulation in the T-handle and engage the screwdriver into the head of the bolt. Holding the T-handle turn the screwdriver anticlockwise to partially jam the bolt in the T-handle. Remove the screwdriver. While gently pulling, turn the T-handle anticlockwise. Postoperative treatment guidelines Immobilization in a well padded splint for the first 2 weeks. During this time a dependent position should be avoided. The dressing is removed at 2 weeks, the wound is evaluated and gentle range of motion of the ankle and 1st metatasophalangeal joints is initiated. Full weight-bearing between 10 12 weeks based on radiographic evidence of healing. Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 22

Implants Midfoot Fusion bolt B 6.5 mm Length (mm) Stainless steel TAN 50 02.111.150 04.111.150 55 02.111.155 04.111.155 60 02.111.160 04.111.160 65 02.111.165 04.111.165 70 02.111.170 04.111.170 75 02.111.175 04.111.175 80 02.111.180 04.111.180 85 02.111.185 04.111.185 90 02.111.190 04.111.190 95 02.111.195 04.111.195 100 02.111.200 04.111.200 105 02.111.205 04.111.205 110 02.111.210 04.111.210 115 02.111.215 04.111.215 120 02.111.220 04.111.220 125 02.111.225 04.111.225 130 02.111.230 04.111.230 135 02.111.235 04.111.235 140 02.111.240 04.111.240 145 02.111.245 04.111.245 150 02.111.250 04.111.250 155 02.111.255 04.111.255 160 02.111.260 04.111.260 All Midfoot Fusion Bolts are available nonsterile and sterile packed. For sterile implants add suffix S to the article number. 22 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

Instruments 03.111.001 T-Handle for Midfoot Fusion Bolt 314.164 Screwdriver Stardrive 4.5/5.0, T25, with Groove, length 240 mm 03.111.002 Drill Bit 5.0 mm, calibrated, cannulated, length 300/250 mm, 3-flute, for Quick Coupling 357.047 Drill Bit 6.5 mm, cannulated, length 330 mm, for Quick Coupling 292.680 Guide Wire 2.8 mm with threaded tip, length 300 mm 03.227.030 Direct Measuring Device for HCS- Headless Compression Screw 6.5 mm 292.200 Kirschner Wire 2.0 mm with trocar tip, length 150 mm Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 23

Optional Instruments 312.080 Drill Sleeve 8.5/2.8 312.050 Protection Sleeve 12.0/8.5 02.227.001 Guide Wire B 2.8 mm with trocar tip, length 300 mm, Stainless Steel 338.490 Quick Coupling for Small Air Drill or 511.761 Quick Coupling for DHS/DCS Triple Reamers 22 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

Vario Cases 68.111.041 Insert, size 1/1, for Instrument Set Midfoot Fusion Bolt B 6.5 mm 68.000.101 Lid for Instrument Tray, size 1/1 68.111.047 Module with lid for 6.5 mm Midfoot Fusion Bolts (TAN or Stainless Steel) Optional insert The implant module fits into the insert 68.111.042 68.111.042 Insert, size 1/1, for Midfoot Fusion Bolt B 6.5 mm 68.000.101 Lid for Instrument Tray, size 1/1 Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 22

Set Lists Instrument and implant set for Midfoot Fusion Bolt in Vario insert and module 01.111.042 Midfoot Fusion Bolt set (Stainless Steel) 01.111.044 Midfoot Fusion Bolt set (TAN) Case 68.111.041 Insert, size 1/1, for Instrument Set Midfoot Fusion Bolt B 6.5 mm 68.000.101 Lid for Instrument Tray, size 1/1 68.111.047 Module with lid for 6.5 mm Midfoot Fusion Bolts (TAN or Stainless Steel) The Vario Case insert offers space for standard and optional instruments 22 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

Instruments Article No: Description No: 03.111.001 T-Handle for Midfoot Fusion Bolt 1 314.164 Screwdriver Stardrive 4.5/5.0, T25, 1 with Groove, length 240 mm 03.111.002 Drill Bit B 5.0 mm, calibrated, 1 cannulated, length 300/250 mm, 3-flute, for Quick Coupling 292.680 Guide Wire B 2.8 mm with 8 threaded tip, length 300 mm 03.227.030 Direct Measuring Device for HCS- 1 Headless Compression Screw B 6.5 mm 357.047 Drill Bit B 6.5 mm, cannulated, 1 length 330 mm, for Quick Coupling Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 22

Set Lists Implants 0X.111.150 length 50 mm 0X.111.220 length 120 mm 0X.111.155 length 55 mm 0X.111.225 length 125 mm 0X.111.160 length 60 mm 0X.111.230 length 130 mm 0X.111.165 length 65 mm 0X.111.235 length 135 mm 0X.111.170 length 70 mm 0X.111.240 length 140 mm 0X.111.175 length 75 mm 0X.111.245 length 145 mm 0X.111.180 length 80 mm 0X.111.250 length 150 mm 0X.111.185 length 85 mm 0X.111.255 length 155 mm 0X.111.190 length 90 mm 0X.111.260 length 160 mm 0X.111.195 length 95 mm 0X.111.200 length 100 mm 0X.111.205 length 105 mm 0X.111.210 Midfoot Fusion Bolt, dia. 6.5mm, length 110 mm 0X.111.215 length 115 mm X = 2: stainless steel X = 4: TAN 22 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

Bibliography Sigvard T Hansen, JR Functional reconstruction of the foot and ankle: 243 256 Trepman Elly, Nihal Aneel, Pinzur Michael Charcot Neuroarthropathy of the foot and ankle. Foot & ankle 2005 Rajbhandari S, Jenkins R, Davies C, Tesfaye S (2002): Charcot neuroarthropathy in diabetes mellitus. Diabetologia 45: 1085 1096 Midfoot Fusion Bolt 6.5 mm Surgical Technique DePuy Synthes 22

MRI Information Torque, Displacement and Image Artifacts according to ASTM F 2213-06, ASTM F 2052-06e1 and ASTM F2119-07 Non-clinical testing of worst case scenario in a 3 T MRI system did not reveal any relevant torque or displacement of the construct for an experimentally measured local spatial gradient of the magnetic field of 3.69 T/m. The largest image artifact extended approximately 169 mm from the construct when scanned using the Gradient Echo (GE). Testing was conducted on a 3 T MRI system. Radio-Frequency-(RF-)induced heating according to ASTM F2182-11a Non-clinical electromagnetic and thermal testing of worst case scenario lead to peak temperature rise of 9.5 C with an average temperature rise of 6.6 C (1.5 T) and a peak temperature rise of 5.9 C (3 T) under MRI Conditions using RF Coils [whole body averaged specific absorption rate (SAR) of 2 W/kg for 6 minutes (1.5 T) and for 15 minutes (3 T)]. Precautions: The above mentioned test relies on non-clinical testing. The actual temperature rise in the patient will depend on a variety of factors beyond the SAR and time of RF application. Thus, it is recommended to pay particular attention to the following points: It is recommended to thoroughly monitor patients undergoing MR scanning for perceived temperature and/or pain sensations. Patients with impaired thermo regulation or temperature sensation should be excluded from MR scanning procedures. Generally it is recommended to use a MR system with low field strength in the presence of conductive implants. The employed specific absorption rate (SAR) should be reduced as far as possible. Using the ventilation system may further contribute to reduce temperature increase in the body. 33 DePuy Synthes Midfoot Fusion Bolt 6.5 mm Surgical Technique

DSEM/TRM/0815/0451 09/15 Synthes GmbH Eimattstrasse 3 4436 Oberdorf Switzerland Tel: +41 61 965 61 11 Fax: +41 61 965 66 00 www.depuysynthes.com This publication is not intended for distribution in the USA. All surgical techniques are available as PDF files at www.depuysynthes.com/ifu 0123 DePuy Synthes Trauma, a division of Synthes GmbH. 2015. All rights reserved. 036.000.994