AAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites

Similar documents
U.S. Nuclear Regulatory Commission

Brachytherapy Sources, Dosimetry and Quality Assurance

Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations

Supplement 2 for the 2004 update of the AAPM Task Group No. 43 Report: Joint recommendations by the AAPM and GEC-ESTRO

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

ADVANCED TECHNOLOGY CONSORTIUM (ATC) CREDENTIALING PROCEDURES FOR LUNG BRACHYTHERAPY IMPLANT PROTOCOLS

Jeffrey F. Williamson a) Radiation Oncology Center, Washington University, St. Louis, Missouri 63110

Transperineal Interstitial Permanent Prostate Brachytherapy (TIPPB) Quality Assurance Guidelines

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Larry A. DeWerd Accredited Dosimetry and Calibration Laboratory, University of Wisconsin, Madison, Wisconsin, 53706

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Disclosure. Societal Guidelines in High Energy and Skin Brachytherapy. Content. Acknowledgments. Jose Perez-Calatayud.

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

Implantable MOSFET dosimeter response to 192 Ir HDR radiation

Material-specific Conversion Factors for Different Solid Phantoms Used in the Dosimetry of Different Brachytherapy Sources

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy

Current Status of Electronic Brachytherapy Dosimetry

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors

Dose Calculation for Photon-Emitting Brachytherapy Sources with Average Energy Higher than 50 kev: Full Report of the AAPM and ESTRO

Bell in 1903 was first suggested implanting. First Report on Study of Quality Control Program of High Dose Rate (Hdr) Brachytherapy.

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output

MAMMOSITE BRACHYTHERAPY DOSIMETRY EFFECT OF CONTRAST AND AIR INTERFACE ON SKIN DOSE. A RESEARCH PAPER SUBMITTED TO THE GRADUATE SCHOOL

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

U.S. DEPARTMENT OF ENERGY LABORATORY ACCREDITATION PROGRAM (DOELAP) FOR PERSONNEL DOSIMETRY SYSTEMS

Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

Brachytherapy Planning and Quality Assurance

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

Optimisation of eye plaque dosimetry using Monte Carlo method. J. Green, D. Cutajar, S. Guatelli, & Rosenfeld, A.B.

Dosimetry benchmark for MBDCA

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

Approval of Dosimetry Services in Ireland Guidelines for Applicants

Comparative Study on Homogeneous and Inhomogeneous Photon Dose Distribution from a High Dose Rate 192 Ir Brachytherapy Source using MCNP5

Review of TG-186 recommendations

Applications in brachytherapy

QA for Clinical Dosimetry with Emphasis on Clinical Trials

Brachytherapy Dosimetry in the Clinic Margaret Bidmead Royal Marsden NHS Trust

Method for verifying the air kerma strength of I-125 plaques for the treatment of ocular melanoma

International Practice Accreditation

Monte Carlo and experimental relative dose determination for an Iridium-192 source in water phantom

International Journal of Radiation Research, July 2014

Monte Carlo Simulation Study on Dose Enhancement by Gold Nanoparticles in Brachytherapy

UNCERTAINTIES REPORTING IN MEDICAL PHYSICS. Jing Cui, DSc. Assistant Professor Department of Radiation Oncology UC Davis Comprehensive Cancer Center

A commissioning procedure for breast intracavitary electronic brachytherapy systems

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

The need for standardization of dosimetry in experimental radiation biology

INTERNSHIP DUE PROCESS GUIDELINES

Sterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control

Dosimetric Characteristics of the Brachytherapy Sources Based on Monte Carlo Method

Developing a Verification and Training Phantom for Gynecological Brachytherapy System

LAB5 EDITION 2 NOVEMBER 2012

Radioactive sources in brachytherapy

Metrology Laboratory of Ionizing Radiation

Testing Railway Safety Critical Workers for Drugs and Alcohol

TRAUMA RECOVERY/HAP OPERATING GUIDELINES

Metrology Laboratory of Ionizing Radiation

Gafchromic film dosimetry of a new HDR 192 Ir brachytherapy source

IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Bone Densitometry Foundation Course (e-learning)

Radiochromic film dosimetry in water phantoms

Purpose: Policy: The Fair Hearing Plan is not applicable to mid-level providers. Grounds for a Hearing

Please find enclosed Dosimetry Service Licence No which replaces Dosimetry Service Licence No

COUNCIL OF THE EUROPEAN UNION. Brussels, 7 September 2009 (OR. en) 11261/09 Interinstitutional File: 2008/0002 (COD) DENLEG 51 CODEC 893

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

INTERNATIONAL STANDARD ON ASSURANCE ENGAGEMENTS 3000 ASSURANCE ENGAGEMENTS OTHER THAN AUDITS OR REVIEWS OF HISTORICAL FINANCIAL INFORMATION CONTENTS

EHR Developer Code of Conduct Frequently Asked Questions

Assurance Engagements Other than Audits or Review of Historical Financial Statements

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Episcleral eye plaque dosimetry comparison for the Eye Physics EP917 using Plaque Simulator and Monte Carlo simulation

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

The Role of Monte Carlo and other Advanced Dose Calculation Algorithms for Breast APBI

Inter-society standards for the performance of brachytherapy: a joint report from ABS, ACMP and ACRO

UKAS Guidance on the Application of ISO/IEC Dealing with Expressions of Opinions and Interpretations

Pmasonic Dosimetry System Peformance Testing and Results at Nuclear Accident Dose Levels

Calculation of Dose Distribution Around a Clinical 252. Cf Source for Neutron Therapy Based on AAPM, TG-43 Protocol

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Commissioning and Radiobiology of the INTRABEAM System

The Status of Guidance for Safety, Quality Management and Practice for High Dose-Rate Brachytherapy

SunCHECK Patient Comprehensive Patient QA

A COMPARISON BETWEEN GEANT4 AND MCNPX ON THE DOSIMETRY OF THE 192IR MICROSELECTRON V2 HDR BRACHYTHERAPY

Z E N I T H M E D I C A L P R O V I D E R N E T W O R K P O L I C Y Title: Provider Appeal of Network Exclusion Policy

GRIEVENCE PROCEDURES INFORMAL REVIEWS AND HEARINGS

A resident's salary will continue, during the time they are exercising the Grievance Procedure rights, by requesting and proceeding with a hearing.

Dosimetric parameters of three new solid core I-125 brachytherapy sources

Commissioning of brachytherapy Treatment Planning Systems

AUDIT OUTLINE INFORMATION SUMMARY

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005

SBRT Credentialing: Understanding the Process from Inquiry to Approval

Clinical Trial Credentialing:

Financial Administration and Control of Research and Special Funds

Physics. Copyright 2010 Xoft, Inc.

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Digital film dosimetry in radiotherapy and the development of analytical applications software

Transcription:

AAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites A. Purpose and Rationale Many individual users contact the AAPM Brachytherapy Subcommittee (BTSC) and IROC Houston asking which of the available photon-emitting brachytherapy source models conform to the AAPM s guidelines. The AAPM guidelines are based on reports for low-energy and high-energy photonemitting brachytherapy sources: 1. R. Nath, L. L. Anderson, G. Luxton, K. A. Weaver, J. F. Williamson, and A. S. Meigooni, "Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43," Med. Phys. 22, 209 234 (1995). 2. R. Nath, L. L. Anderson, J. A. Meli, A. J. Olch, J. A. Stitt, and J. F. Williamson, "Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56," Med. Phys. 24, 1557 1598 (1997). 3. H. D. Kubo, G. P. Glasgow, T. D. Pethel, B. R. Thomadsen, and J. F. Williamson, "High dose rate brachytherapy treatment delivery: Report of the AAPM Radiation Therapy Committee Task Group No. 59," Med. Phys. 25, 375 403 (1998). 4. J. F. Williamson, B. M. Coursey, L. A. DeWerd, W. F. Hanson, R. Nath, and G. Ibbott, Guidance to users of Nycomed Amersham and North American Scientific, Inc. I-125 Interstitial Sources: Dosimetry and calibration changes: Recommendation of the American Association of Physicists in Medicine Radiation Therapy Committee Ad Hoc Subcommittee on Low-Energy Seed Dosimetry, Med. Phys. 26, 570 573 (1999). 5. J. F. Williamson, B. M. Coursey, L. A. DeWerd, W. F. Hanson, R. Nath, M. J. Rivard, and G. Ibbott, On the use of apparent activity (A app) for treatment planning of 125 I and 103 Pd interstitial brachytherapy sources: Recommendations of the American Association of Physicists in Medicine Radiation Therapy Committee Subcommittee on Low-Energy Brachytherapy Source Dosimetry, Med. Phys. 26, 2529 2530 (1999). 6. M. J. Rivard, B. M. Coursey, L. A. DeWerd, W. F. Hanson, M. S. Huq, G. S. Ibbott, M. G. Mitch, R. Nath, and J. F. Williamson, "Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations," Med. Phys. 31, 633 674 (2004). 7. L. A. DeWerd, M. S. Huq, I. J. Das, G. S. Ibbott, W. F. Hanson, T. W. Slowey, J. F. Williamson, and B. M. Coursey, "Procedures for establishing and maintaining consistent air-kerma strength standards for low-energy, photon-emitting brachytherapy sources: Recommendations of the Calibration Laboratory Accreditation Subcommittee of the American Association of Physicists in Medicine," Med. Phys. 31, 675 681 (2004). 8. Z. Li, R.K. Das, L. A. DeWerd, G. S. Ibbott, A. S. Meigooni, J. Perez-Calatayud, M. J. Rivard, R. S. Sloboda and J. F. Williamson, "Dosimetric prerequisites for routine clinical use of photon emitting brachytherapy sources with average energy higher than 50 kev," Med. Phys. 34, 37-40 (2007). 9. M. J. Rivard, W. M. Butler, L. A. DeWerd, M. S. Huq, G. S. Ibbott, A. S. Meigooni, C. S. Melhus, M. G. Mitch, R. Nath, and J. F. Williamson, "Supplement to the 2004 update of the AAPM Task Group No. 43 Report," Med. Phys. 34, 2187 2205 (2007). 10. W. M. Butler, W. S. Bice Jr., L. A. DeWerd, J. M. Hevezi, M. S. Huq, G. S. Ibbott, J. R. Palta, M. J. Rivard, J. P. Seuntjens, and B. R. Thomadsen, "Third-party brachytherapy source calibrations and physicist responsibilities: Report of the AAPM Low Energy Brachytherapy Source Calibration Working Group," Med. Phys. 35, 3860 3865 (2008). The Radiation Therapy Oncology Group (RTOG) has asked IROC Houston to identify those brachytherapy sources that meet the AAPM dosimetric prerequisites, since participating institutions will be expected to use sources meeting these prerequisites. This joint AAPM/IROC Houston Source Registry is intended to satisfy the RTOG request and similar requests that have arisen from other cooperative clinical trial study groups. The AAPM and IROC Houston work together to identify and register brachytherapy sources that meet the AAPM criteria. A webpage-based Registry is maintained by IROC Houston, listing sources that meet the AAPM criteria. IROC Houston relies on the AAPM s BTSC, through its Brachytherapy Source Registry Work Group (WGBSR), to establish the criteria for including or excluding source models from this Registry, and to determine which source models meet the criteria. As the recommendations are intended to apply internationally, some of the agencies, organizations and standard laboratories identified within the U.S. could be, at the discretion of the BTSC, interpreted in the context of the arrangements in other countries where applicable. AAPM-IROC Houston Brachytherapy Source Registry Policy v13 Page 1 of 7

B. Compliance Registry Appearance and Contents The Registry is in the form of a webpage reachable via hot links from both the AAPM and IROC Houston websites, and is entitled Joint AAPM/IROC Houston Registry of Brachytherapy Sources Meeting the AAPM Dosimetric Prerequisites. The webpage displays the following information for each source model listed on the Registry: The vendor (manufacturer if different), radionuclide, model number, trade name, date of addition to the Registry, and dosimetry publications taken as evidence of compliance. Links to this Policy, the AAPM prerequisites publications, other publications subsequently determined to be relevant, as well as an application form to be used by source vendors are kept accessible on the webpage. The following text appears verbatim on the webpage: The AAPM, through its Brachytherapy Subcommittee, has determined that the following brachytherapy source models comply with the AAPM s dosimetric prerequisites as set forth in one of two publications: "Dosimetric prerequisites for routine clinical use of new low-energy photon interstitial brachytherapy sources: Recommendations of the American Association of Physicists in Medicine Radiation Therapy Committee". Med. Phys. 25, 2269-2270 (1998), or Dosimetric prerequisites for routine clinical use of photon emitting brachytherapy sources with average energy higher than 50 kev," Med. Phys. 34, 37-40 (2007). The manufacturers must also satisfy criteria established by the AAPM Subcommittee on Calibration Laboratory Accreditation and described in a publication (hereafter referred to as the 2004 CLA Report ): Procedures for establishing and maintaining consistent air-kerma strength standards for low-energy, photon-emitting brachytherapy sources: Recommendations of the Calibration Laboratory Accreditation Subcommittee of the American Association of Physicists in Medicine. Med. Phys., 35 671-685 (2004). The criteria are summarized here. (Note: The final phrase is a link to a new page that lists the following text: The AAPM dosimetric prerequisites for brachytherapy sources may be summarized as follows The vendor provides air-kerma strength calibrations that are directly or indirectly traceable to the air-kerma strength standards (S K,N99 for low energy sources) for photon emitting brachytherapy sources maintained by NIST, or in the case of HDR 192 Ir, the standard developed by the University of Wisconsin and maintained by the ADCLs. A full set of TG-43 dosimetric parameters is available, supporting both calculation of the 2-D dose-rate distribution and, for low-energy seed models, the 1-D isotropic point source approximation. This set of dosimetric parameters must be based upon at least one experimental study and at least one Monte Carlo study of the source model s dosimetric parameters. These studies must be performed by investigators having independence from the manufacturer, and have been accepted for publication in a peer-reviewed journal. The vendor has assured that the calibration from NIST has been transferred to the ADCLs. The vendor has implemented a program that is compliant with the 2004 CLA Report for periodically comparing its air-kerma strength calibrations with the NIST primary standard and the secondary standards maintained by the ADCLs. The complete text of the 2004 CLA report is available here.) AAPM-IROC Houston Brachytherapy Source Registry Policy v13 Page 2 of 7

The AAPM dosimetric prerequisites for high-energy sources may be summarized as follows: The single-source dose distribution used for clinical treatment planning should be based upon two dose-rate determinations, one of which is a theoretical calculation method such as the Monte Carlo method, and the other an experimental measurement. Note that a single dosimetric study is acceptable for certain 137 Cs, 192 Ir, and 60 Co sources under specific circumstances (see the complete text.) A full set of TG-43 dosimetric parameters is available, supporting calculation of the 2-D dose-rate distribution. In addition, a set of along and away tables must be provided. The vendor provides air-kerma strength calibrations that are directly or indirectly traceable to the air-kerma strength standards for photon emitting brachytherapy sources maintained by NIST. The vendor has implemented a program of periodically comparing its air-kerma strength calibrations with the NIST primary standard or the secondary standards maintained by the ADCLs that is compliant with the CLA Report. Note, however, that while the CLA Report specifically addresses low-energy sources, the AAPM has confirmed that manufacturers are expected to comply with its requirements for highenergy sources also. The CLA Report was dedicated to low-energy sources. Until a similar report is available for high-energy sources, vendors must apply these guidelines with appropriate variations. For example, it may be necessary to send a source or a measurement system to NIST or an ADCL. In addition, the AAPM requires biennial comparisons for high-energy sources through Measurement Quality Assurance (MQA) tests. In the case of experimental sources for which no ADCL interpolative or NIST primary standard exists, the end user institution is responsible for calibrating the sources with an ionization chamber of appropriate dimensions that has calibrations directly traceable to an appropriate NIST air-kerma standard. The webpage will contain the following disclaimer (the disclaimer will be displayed in a new page by clicking on a link labeled Disclaimer ): The AAPM and IROC Houston maintain this Registry solely as a service to their members and clients. Neither the AAPM nor IROC Houston endorses or approves specific products. No statement regarding the quality of construction, safety, or clinical effectiveness of specific sources is expressed or implied by inclusion or exclusion of sources from this Registry. The AAPM Brachytherapy Subcommittee bases its decisions to place or exclude source models in this Registry on the content of published papers and the vendor s descriptions of its calibration procedure. The AAPM and IROC Houston neither warrant nor are responsible for (a) accuracy of the published dosimetry studies and applicability to the sources as manufactured; (b) compliance of the vendor with its stated procedures; or (c) the accuracy of any particular brachytherapy source calibration. The AAPM does not monitor on a continuing basis the compliance of the vendor s calibration procedures with the AAPM prerequisites nor the compatibility of its source design and manufacturing practices with the published dosimetry data. AAPM-IROC Houston Brachytherapy Source Registry Policy v13 Page 3 of 7

C. Procedures for Registry Inclusion 1. Request for inclusion of a source model is initiated by the vendor applying to the AAPM BTSC for this purpose. 2. A special case exists for orphaned sources; those no longer commercially available, but still in regular use in hospitals. These must be sources with long half-life and suitable dose rates which consequently comprise only certain models of 137 Cs and 60 Co sources. In the case of these sources, there is no manufacturer available to submit the Registry application forms. For these orphaned sources, the AAPM and IROC Houston have agreed to an alternative procedure for Registry application: a hospital that wishes to participate in a clinical trial that involves brachytherapy sources not currently posted on the Registry may submit the application, listing the dosimetric studies available and the dosimetry parameters to be used for treatment planning. The hospital must also describe their method of source strength traceability for review by IROC Houston to assure the correct calibration of the sources. In the special case of source trains, in which individual sources cannot be removed for calibration with a well chamber, the hospital may describe a method of calibration at a distance in a phantom, in accordance with calibration procedures described in the peer-reviewed literature. 3. Upon receipt, BTSC will deliver the application to WGBSR, which will review the published/accepted dosimetry papers. WGBSR will review the vendor s statements regarding calibration procedures with appropriate NIST Group Leader as well as the ADCL Directors and the AAPM CLA Subcommittee chair. WGBSR may withhold entry into the Registry until any discrepancies between the vendor s statements and calibration lab personnel observations are resolved satisfactorily. Following its review, WGBSR will forward a recommendation for or against inclusion in the Registry to BTSC. 4. BTSC has adopted the following Independence Policy (reproduced from Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. Med. Phys. 31, 633-674 (2004): The first meaning of independent studies is that they are performed, written, and published by investigators who are affiliated with institutions independent of the source vendor and who have no major conflicts-of interest with that vendor. The second meaning of independent studies is that they are scientifically independent of one another, i.e., they represent independent and distinct estimations of the same quantities. In the case of two measurement based studies, this will usually mean that two different investigators have used their own methodologies for measuring Λ and sampling the relative dose distribution, as TLD dosimetry is highly technique and investigator dependent. In the case of an empirical study and a Monte Carlo study, if properly executed, they will yield scientifically independent estimates of the TG-43 parameters. Thus, so long as the two studies are successfully scrutinized by the peer-review process and satisfy the AAPM scientific requirements, the empirical and Monte Carlo investigator author lists can overlap or even be identical. It is permissible to publish the Monte Carlo and measured estimates in the same paper so long as the two datasets are independently tabulated. In this context, Not independent means that the one study is used to modify the outcomes and methods of the other to improve agreement between the two datasets in a manner that is not scientifically justified. 5. BTSC has adopted the following Data Overlap Policy : We require that, for sources other than conventionally-encapsulated 137 Cs, 192 Ir, and 60 Co sources. two dosimetry studies each assess: (a) Λ; (b) g L (r) or g P (r); (c) F(r,θ) and φ an (r) with r and θ sampling resolution at least as good as that recommended in the 2004 AAPM TG-43U1 (low-energy) and Li et al. 2007 (highenergy) reports so that overlap between the two studies may readily reveal significant discrepancies if present; and (d) L or L eff. As provided in reference (8) above, a single dosimetry study is acceptable for conventionally-encapsulated 137 Cs 192 Ir, and 60 Co sources. 6. BTSC expects that vendors will comply with the AAPM prerequisite requiring periodic comparisons of their calibrations with those of an independent laboratory. Vendors can best do this by participating in comparisons with both NIST and the ADCLs offering calibration AAPM-IROC Houston Brachytherapy Source Registry Policy v13 Page 4 of 7

services for that source model. The AAPM has recommended a procedure and a frequency for such comparisons in: Procedures for establishing and maintaining consistent air-kerma strength standards for low-energy, photon-emitting brachytherapy sources: Recommendations of the Calibration Laboratory Accreditation Subcommittee of the American Association of Physicists in Medicine. Med. Phys. 31, 675-681 (2004). The AAPM High Energy Brachytherapy Source Dosimetry working group (HEBD) has set a calibration comparison frequency of two years for 60 Co, 137 Cs, and 192 Ir. Vendors of these high-energy sources should comply with this comparison frequency. 7. Inclusion into the Registry is by majority vote of the BTSC. Written notification of any unfavorable decision will be provided to the vendor by the WGBSR chair along with an explanation. 8. Upon completion of a positive vote for inclusion by the BTSC, IROC Houston will be informed of the decision and instructed to add the source model to the Registry. 9. BTSC reserves the right to withhold entry into the Registry if it believes that the dosimetry articles contain technical errors or omissions; or rely on unproven or inappropriate computational or experimental dosimetry methodologies. 10. Upon receipt of notification of rejection of a source from the Registry, a manufacturer may appeal the decision by submitting a written request for review to the AAPM Therapy Physics Committee (TPC). The TPC will assign a minimum of three physicists to review the materials submitted to the BTSC and the decision made by BTSC. TPC may decide to endorse or reverse the decision of the BTSC. 11. This Policy may be revised as new source models or technologies emerge. AAPM-IROC Houston Brachytherapy Source Registry Policy v13 Page 5 of 7

D. Procedures for Registry Removal Vendors (or hospitals according to section C.2) may request that a source be removed from the Registry on the grounds that it will no longer be available to the medical community. Immediate steps will be taken to remove the source from the Registry upon receipt of a specific written request from the manufacturer, source distributor, or hospital. While manufacturers are encouraged to contact the BTSC should there be any changes in their source manufacturing processes, the BTSC retains the discretion to remove a product from the Joint AAPM/IROC Houston Registry should it have grounds to believe that a product, as currently manufactured, is no longer accurately characterized by the reference-quality dose-distribution data submitted in support of Registry posting. Reasonable grounds include, but are not limited to: change of manufacturing venue or process, changes in source geometry and/or internal design, and discovery of deficiencies in the published dosimetry data. To retain Registry status, the vendor may submit for BTSC review a set of revised dosimetry data that meet the standards specified by the AAPM dosimetric prerequisites. Alternatively, the vendor may provide evidence to BTSC demonstrating that the revised product is dosimetrically equivalent to the original source from which the published and accepted dosimetry data were derived. Such information should include, as applicable, Assurances from the vendor that manufacturing processes are unchanged and the geometric and compositional structure of relevant source components are unchanged. Review of quality assurance data (e.g., source radiograph, optical micrographs, measurements, etc.) and vendor manufacturing processes with original dosimetry investigators. These investigators must make a recommendation regarding equivalence to BTSC and be willing to discuss findings with BTSC in closed session. Review of NIST anisotropy and spectroscopy measurements before and after the product change. A description of the vendor s efforts and checks, along with their outcome, to assure the equivalence of their current product with that at the time of original dosimetric characterization. The above information should be submitted to BTSC with sufficient lead-time that BTSC can make a recommendation prior to the vendor distributing the modified product commercially for routine clinical use. Failure of a vendor to maintain a regular program of comparisons with NIST and an ADCL, in accordance with the recommendations of the AAPM Calibration Laboratory Accreditation (CLA) Subcommittee (Med. Phys. 31, 675-681 (2004) or Med. Phys. 34, 37-40 (2007)) may result in removal. When CLA determines that a vendor has exceeded the recommended interval, the CLA chair will issue a warning. If a response to the chair s communication is not received within 60 calendar days, the CLA chair will notify the BTSC chair that the source model should be removed from the Registry. The decision to remove a source model from the Registry will be made by majority vote of the BTSC. Upon completion of a vote to remove, IROC Houston will be instructed to remove the source from the webpage Registry. The following application form is available for download from the Registry webpage: AAPM-IROC Houston Brachytherapy Source Registry Policy v13 Page 6 of 7

Application for Inclusion in the Joint AAPM/IROC Houston Registry of Photon-Emitting Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites To: Chair, Brachytherapy Subcommittee of the AAPM Therapy Physics Committee From: (Manufacturer) Technical contact and address: The source identified below is believed by the manufacturer to meet the AAPM Prerequisites for dosimetry of brachytherapy sources. The prerequisites for low-energy sources are defined in: "Dosimetric prerequisites for routine clinical use of new low energy photon interstitial brachytherapy sources: Recommendations of the American Association of Physicists in Medicine Radiation Therapy Committee". Med. Phys., 25, 2269-2270 (1998). The prerequisites for high-energy sources are defined in: "Dosimetric prerequisites for routine clinical use of photon emitting brachytherapy sources with average energy higher than 50 kev," Med. Phys. 34, 37-40 (2007). Note that the recommendation regarding the calibration frequency of the vendor s sources for in-house QA (recommendation number 6 in the reference for high-energy sources) has been extended by the AAPM High Energy Brachytherapy Source Dosimetry Working Group to 2 years for 192 Ir, 137 Cs, and 60 Co sources. As the recommendations are intended to apply internationally, some of the agencies, organizations, and standard laboratories identified within the U.S. could be, at the discretion of the BTSC, interpreted in the context of the arrangements in other countries where applicable. The manufacturer hereby requests that the source be included by the AAPM and IROC Houston on its web-based Registry of sources complying with the Prerequisites. Source Name: Model Number: Manufactured by: Marketed by: Accredited Laboratory(s) providing initial reference calibration(s), and date of calibration(s): Dates of comparisons between accredited laboratory and vendor calibration procedures during the last 2 years: Do you provide your clients with air-kerma strength calibrations directly or indirectly traceable to the appropriate NIST standard? Please attach publications describing the dosimetric parameters of the source and provide citations. Please describe your program of measurement quality assurance and calibration constancy. Indicate the laboratory(s) with which frequent comparisons will be made. AAPM-IROC Houston Brachytherapy Source Registry Policy v13 Page 7 of 7