Case Report Multimodality Fusion with MRI, CT, and Ultrasound Contrast for Ablation of Renal Cell Carcinoma

Similar documents
Radiofrequency Ablation of Liver Tumors

Accurate Guidance. Confident Outcomes.

Patient Selection for Ablative Therapies. Adrian D Joyce Leeds UK

Mahrad Paymani, MD Lake Medical Imaging

Aid to percutaneous renal access by virtual projection of the ultrasound puncture tract onto fluoroscopic images

Fusion Ultrasound: Characterization of Abdominal Masses with MR, CT, PET, and Contrast Ultrasound

S th US Contrast

Outline - MRI - CT - US. - Combinations of imaging modalities for treatment planning

FOR IMMEDIATE RELEASE

CT Guided Procedures And Interesting Cases. Stephen Kim, MD Diagnostic and Interventional Radiology

Virtually-tracked US-guided Radiofrequency Ablation(RFA) of Benign Thyroid Nodules: Preliminary Results

Case Report Three-Dimensional Dual-Energy Computed Tomography for Enhancing Stone/Stent Contrasting and Stone Visualization in Urolithiasis

1. INTRODUCTION ABSTRACT

A Navigation System for Minimally Invasive CT-Guided Interventions

Human Systems. Technology - Ultrasounds

Effect of Electromagnetic Navigation on CT-Guided Percutaneous Thermal Ablation or Biopsy of Lung Tumors

General summary GENERAL SUMMARY

Abdominal Ultrasound

Welcome to a better way for

NIH Public Access Author Manuscript J Urol. Author manuscript; available in PMC 2011 October 16.

Expanding therapy options for

Image Fusion, Contouring, and Margins in SRS

Welcome to a better way for YOU and YOUR patient.

Error analysis in neuronavigation

SURGICAL ULTRASOUND (University Diploma)

Certification Review. Module 28. Medical Coding. Radiology

PINPOINTING RADIATION THERAPY WITH THE PRECISION OF MR.

MRI to fit your planning. Philips Panorama HFO Oncology Configuration

NIH Public Access Author Manuscript Diagn Imaging Eur. Author manuscript; available in PMC 2014 November 10.

Tracked Ultrasound in Navigated Spine Interventions

Killing Tumors with Scans Not Scalpels: Kidney Cancer Ablation. Basics. What is Percutaneous Ablation? Where are your kidneys?

Needle Placement Robots

Thermal Damage of the Genitofemoral Nerve Due to Radiofrequency Ablation of Renal Cell Carcinoma: A Potentially Avoidable Complication

Thermal Damage of the Genitofemoral Nerve Due to Radiofrequency Ablation of Renal Cell Carcinoma: A Potentially Avoidable Complication

Percutaneous ablation of renal cell carcinoma. Where do we stand now? Sanja Stojanović, Spasić Aleksandar

POSITRON EMISSION TOMOGRAPHY (PET)

Siemens Healthineers Erlangen, August 27, Siemens Healthineers presents imaging for complex, minimally invasive procedures

NIH Public Access Author Manuscript Proc SPIE. Author manuscript; available in PMC 2013 December 31.

An abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen.

Three-Dimensional Reconstruction of Renovascular-Tumor Anatomy to Facilitate Zero-Ischemia Partial Nephrectomy

HEPATIC METASTASES. We can state 3 types of metastases depending on their treatment options:

Biologic Effects of Diagnostic Imaging Modalities

The bright Future of Diagnostic Imaging. Paul Smit Senior Vice President Medical Systems

Over 35,000 new cases of renal

Role of DE-CT in Oncology

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

The next generation of prostate care

Ultrasound - Prostate

State-of-the-art minimally invasive interventions for liver tumors

FOR CMS (MEDICARE) MEMBERS ONLY NATIONAL COVERAGE DETERMINATION (NCD) FOR MAGNETIC RESONANCE IMAGING:

Designing the smart operating room of the future

7th November, Labex TRAIL

AUA Guidelines Renal Mass and Localized Kidney Cancer

Children's (Pediatric) Ultrasound - Abdomen

8/10/2016. PET/CT Radiomics for Tumor. Anatomic Tumor Response Assessment in CT or MRI. Metabolic Tumor Response Assessment in FDG-PET

cryosurgical_ablation_of_miscellaneous_solid_tumors 1/2007 5/2017 5/2018 5/2017

MEASUREMENT OF EFFECT SOLID TUMOR EXAMPLES

Is renal cryoablation becoming an effective alternative to partial nephrectomy?

How I do it: percutaneous radiofrequency ablation (RFA) Igor Sorokin, MD 1 Murthy Chamarthy, MD, 2 Jeffrey A. Cadeddu, MD 1,2 1

RF Ablation: indication, technique and imaging follow-up

Esophageal Cancer. What is esophageal cancer?

FIELD SAFETY NOTICE / PRODUCT NOTIFICATION

Timothy L. Miao 1, Ania Z. Kielar 2,3, Rebecca M. Hibbert 2, Nicola Schieda 2,3

Integrated Planning, Navigation and Targeting for Tumor Ablation

Introduction of FIREFLY Technology

CT-Guided Liver Biopsy With Electromagnetic Tracking: Results From a Single-Center Prospective Randomized Controlled Trial

SYSTEM OVERVIEW MINIMALLY INVASIVE ROBOTIC LASER THERMOTHERAPY

Section Activity Activity Description Details Reference(s)

Computer-Assisted Navigation on the Arrested Heart during CABG Surgery

Diagnostic Ultrasound. Sutiporn Khampunnip, M.D.

Hepatic Imaging: What Every Practitioner Should Know

Acknowledgements. QA Concerns in MR Brachytherapy. Learning Objectives. Traditional T&O. Traditional Summary. Dose calculation 3/7/2015

Abdominal Imaging Update. Tom Sutherland MBBS MMed FRANZCR

Improving Methods for Breast Cancer Detection and Diagnosis. The National Cancer Institute (NCI) is funding numerous research projects to improve

D. J. Margolis 1, S. Natarajan 2, D. Kumar 3, M. Macairan 4, R. Narayanan 3, and L. Marks 4

Robotic HIFU intervention under US guidance

Research Article Comparison of Colour Duplex Ultrasound with Computed Tomography to Measure the Maximum Abdominal Aortic Aneurysmal Diameter

See it! Trust it! Treat it!

FOR CMS (MEDICARE) MEMBERS ONLY NATIONAL COVERAGE DETERMINATION (NCD) FOR MAGNETIC RESONANCE IMAGING:

IMAGE-GUIDED SPINE SURGERY

Multiparametric MR Imaging of the Prostate after Treatment of Prostate Cancer

China Medical Technologies, Inc.

Framework for 3D TransRectal Ultrasound (TRUS) Image-Based Tracking - Example of Use Evaluation of 2D TRUS Prostate Biopsies Mapping

Disclosure. Acknowledgement. What is the Best Workup for Rectal Cancer Staging: US/MRI/PET? Rectal cancer imaging. None

Questions to ask before you have a kidney removed for kidney cancer

CoATherm. AK Series. APRO KOREA Inc. Radio Frequency Thermal Ablation System

Pioneering the future of general imaging.

Ultrasound imaging is a noninvasive medical test that helps physicians diagnose and treat medical conditions.

MR-US Fusion. Image-guided prostate biopsy. Richard E Fan Department of Urology Stanford University

MRI Applications in Radiation Oncology:

Augmented Reality Needle Guidance Improves Facet Joint Injection Training

Intuitive. Intelligent. Innovative. General Imaging

Prostate MRI Hamidreza Abdi, MD,FEBU Post Doctoral Fellow Vancouver Prostate Centre UBC Department of Urologic Sciences May-20144

SUPERIORITY OF A REAL TIME PLANNING TECHNIQUE OVER IMAGE GUIDED RADIATION THERAPY FOR THE TREATMENT OF PRIMARY PROSTATE CANCERS

RADIOFREQUENCY ABLATION

OPTO-ACOUSTIC BREAST IMAGING

MR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015

CT Scanning Protocol For V2R Guided Surgery Solutions

Breast Cancer Screening

Transcription:

Hindawi Publishing Corporation Case Reports in Urology Volume 2012, Article ID 390912, 5 pages doi:10.1155/2012/390912 Case Report Multimodality Fusion with MRI, CT, and Ultrasound Contrast for Ablation of Renal Cell Carcinoma Hayet Amalou 1 and Bradford J. Wood 1, 2 1 Center for Interventional Oncology, NIH Clinical Center, National Institutes of Health, MSC 1182, Bethesda, MD 20892, USA 2 Urologic Oncology Branch, National Cancer Institute, National Institutes of Health, Bethesda, MD 20892, USA Correspondence should be addressed to Bradford J. Wood, bwood@nih.gov Received 13 August 2012; Accepted 14 October 2012 Academic Editors: M. Gallucci, S. Giulini, A. Goel, A. Marte, and T. Nomura Copyright 2012 H. Amalou and B. J. Wood. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Fusion technology with electromagnetic (EM) tracking enables navigation with multimodality feedback that lets the operator use different modalities during different parts of the image-guided procedure. This may be particularly helpful in patients with renal insufficiency undergoing kidney tumor ablation, in whom there is a desire to minimize or avoid nephrotoxic iodinated contrast exposure. EM tracking software merges and fuses different imaging modalities such as MRI, CT, and ultrasound and can also display the position of needles in real time in relation to preprocedure imaging, which may better define tumor targets than available intraoperative imaging. EM tracking was successfully used to ablate a poorly visualized renal tumor, through the combined use of CT, gadolinium-enhanced MR, and contrast-enhanced US imaging to localize the tumor. 1. Introduction Although surgical resection remains the gold standard for treating renal cell carcinoma, radiofrequency ablation (RFA) or cryoablation of renal lesions is widely becoming accepted as an effective treatment modality for patients that are poor surgical candidates, or in whom poor renal function necessitates a nephron-sparing approach [1]. The most technically challenging portion of an ablation procedure may be the localization of the tumor margins, which may be best defined on a modality (or enhancement phase) not immediately available during the procedure. Renal tumors may be poorly visualized with unenhanced CT or US and may be only visible on contrast-enhanced ultrasound (US), CT, and MRI [1]. However, MR-guided interventions are limited by cost, availability, and special equipment needs. EM tracking allows real-time visualization of needle-tip position and angle of trajectory superimposed upon a preprocedural image. Such systems require an electromagnetic field generator and a special introducer needle or stylet with a sensor coil embedded within an introducer needle or clipped to the needle hub. A small current is induced by the coil, as it moves within the changing electromagnetic field. This changing current reports three dimensional position coordinates and trajectory. Coregistration or matching of modalities is accomplished by placing fiducials or a fiducial patch on the skin near the field of interest. The coordinates are semiautomatically matched between the image and the fiducial, then a virtual image of the needle position is superimposed upon prior CT, MRI, and/or PET [2]. A composite fusion image can be displayed blending (or displaying side by side) two or more modalities along with the needle position. The accuracy and clinical impact of EM tracking systems for biopsy and ablation have been described in early reports [3, 4]. RF ablation in a patient with renal insufficiency was facilitated with EM tracking and multimodality fusion imaging that included real-timeultrasoundwithcontrastaswellaspreprocedural unenhanced CT and MR. Using this approach, iodinated and gadolinium contrasts were not used for the ablation, and thus risks of renal damage (from iodine) and nephrogenic systemic fibrosis (from gadolinium) were minimized.

2 Case Reports in Urology Figure 1: Preprocedure enhanced CT shows a mixed solid and cystic enhancing posterior renal tumor (6 o clock in kidney). 2. Case Report A 53-year-old-female patient with a history of Von Hippel Lindau disease, multiple prior complex surgeries complicated by massive blood loss and transfusions, prior renal ablations, and recurrent renal tumors presented with a growing left, posterior, midpole, exophytic, 3 cm renal tumor, declining renal function, and renal insufficiency, with creatinine = 2.0 mg/dl. The patient gave a written informed consent for a clinical trial for EM tracking and ablation, approved by the institutional investigational review board. Noncontrast CT scan did not adequately visualize the tumor (Figure 4(c)), and contrast was not desirable due to renal insufficiency. MR (Philips 3.0T Achieva scanner, Philips Medical Systems, Best, The Netherlands) with gadolinium- DTPA contrast (Magnevist; Berlex Laboratories, Wayne, NJ, USA) and preprocedure enhanced CT showed the tumor margins well (Figure 1). An EM tracking system and field generator were used (Figure 2) (Northern Digital Inc., Waterloo, ON, USA) interfaced with custom registration and display software (Philips Research, Briarcliff, NY, USA) and a commercially available tracked 22 G stylet inside a standard 19 G outer biopsy guider needle stylet (Philips Healthcare, Toronto, ON, USA). Five fiducial markers were placed near quadrant of interest. Ventilation was interrupted for CT imaging and for navigation. A preprocedural CT scan (MX 8000, Philips Medical Systems, Cleveland, OH, USA) was obtained during the interruption of ventilation, with 3 mm thick sections, 1.5 mm overlap. The skin fiducials were identified manually on the CT. The patient was supine for the planning MR and decubitus during CT and throughout the ablation procedure. Therefore, registration rotated the prior MR to match the preprocedural CT scan, and the target tumor was defined on the fusion images. The skin entry site was estimated using a radio-opaque grid (E-Z-EM Inc., Lake Success, NY, USA). The guider needle was inserted while the tracking software displayed the virtual needle position and trajectory, along with the target point, on the MR-CT composite image. Another CT scan was then taken to confirm the actual location of the needle, which was later compared to the displayed location, using previously described techniques [2 4]. This target to registration error was 1.5 mm. The root mean square registration error of image to actual fiducials was 1.7 mm. To better delineate the tumor margin, as well as the postablation devascularized zone, US contrast was administered (Definity, Bristol Myers Squibb, N. Billerica, MA, USA) in 2 ml boluses (total 4 ml) during needle placement (Figure 2) as well as after RFA to confirm target and to confirm treatment zone, respectively. Similar to the tracked needle, the ultrasound transducer also had sensor coils attached, which allowed the ultrasound image to be tracked and referenced to the needle and the preprocedural CT and MRI images (Figure 4). A single water cooled 17 G needle was placed using tandem technique with a 12-minute ablation, with needle position verified with fusion and ultrasound (Figure 4). Ultrasound contrast then confirmed residual tumor blood flow and viability; so a cluster needle (Cool-tip, Covidien, Boulder, CO, USA) was then introduced by similar technique. After-ablation, US contrast was not seen to perfuse the tumor, suggesting complete treatment (Figure 3). Postablation creatinine actually went down from 2.0 to 1.8 mg/dl. No tumor recurrence was seen at 5-year 1-month postablation MRI (Figure 5). 3. Discussion With the increasing reliance of nephron-sparing techniques in the management of renal cancer, RFA has become a widely accepted form of treatment with even less damage to renal function than surgery [1, 5]. Also, renal tumors may only be seen on specific CT phases of the contrast bolus, or on specific sequences of MRI. This spatial information alone does not guarantee accurate nephron-sparing treatment, unless it is used in combination with fusion. Ideal visualization may improve accuracy for needle-based ablation procedures. The needle was guided with a CT-MR composite image overlaid with a real-time US. When using standard tools alone (and no EM tracking) in cases with limited tumor conspicuity, the physician mentally registers the information from offline preprocedure modalities onto the blind guidance modality using estimates of shared anatomical landmarks. This conventional method is not standardized and highly prone to human error inherent to such estimates. In this case, we electronically superimposed the MR target, well delineated by gadolinium enhancement, upon the procedural CT and US images, allowing the adequate visualization of the tumor without using nephrotoxic CT contrast (Figures 2 4). Limitations and sources of error for fusion imaging with electromagnetic tracking include nearby metal interference, limited working space, rigid registration, organ motion, and respiratory motion [2, 3]. The combination of US with CT and MR can make use of all three complementary modalities: US is an inexpensive and readily available method of obtaining real-time imaging feedback without ionizing radiation. US contrast adds information on tumor perfusion and viability. MR can localize soft tissue tumors often better than US or CT. However,

Case Reports in Urology 3 Figure 2: Ultrasound contrast image and grey scale ultrasound before ablation show the perfusion of solid mass and hypoechoic mixed cystic and solid mass. Figure 3: Ultrasound contrast image and grey scale ultrasound show the lack of the perfusion of solid mass with central needle and now increased echoes in tumor from ablation. US is often obscured by microbubbles released during heating. Multiplanar image reconstruction and real-time position feedback can facilitate needle placement. Although speculative, this could potentially reduce procedure time, decrease radiation exposure, reduce the need for multiple confirmation CTs, and may even avoid multiple needle insertions [6, 7]. Accuracy may be especially important for nephron-sparing procedures. Disclosure NIH has an intellectual property in the field. This work was supported by the Center for Interventional Oncology and the Intramural Research Program of the National Institutes of Health Clinical Center. Philips Healthcare and NIH have a cooperative research and development agreement.

4 Case Reports in Urology (c) (d) Figure 4: Fusion images combine procedural CT with MRI ( and color images) and ultrasound (d) to display tumor target (red sphere and and red dot (d)). The tumor target is displayed in relation to the tracked needle location (cross hairs). Figure 5: T2 weighted MRI 5 years 1 month after ablation shows total disappearance of solid lesion.

Case Reports in Urology 5 Conflict of Interests The authors report no conflict of interests. Acknowledgments Thanks are due to Jochen Kruecker, Ph.D. and Sheng Xu, Ph.D. for assistance with software. Ibrahim Al Bakir, B.S. and Kiran P. Sekhar, M.D. assisted in paper editing. This work was supported in part by the Center for Interventional Oncology and the Intramural Research Program of the National Institutes of Health. NIH and Philips Healthcare have a Cooperative Research and Development Agreement. References [1] M. J. Stone, A. M. Venkatesan, J. Locklin, P. Pinto, M. Linehan, and B. J. Wood, Radiofrequency ablation of renal tumors, Techniques in Vascular and Interventional Radiology, vol. 10, no. 2, pp. 132 139, 2007. [2] B. J. Wood, H. Zhang, A. Durrani et al., Navigation with electromagnetic tracking for interventional radiology procedures: a feasibility study, Journal of Vascular and Interventional Radiology, vol. 16, no. 4, pp. 493 505, 2005. [3] J. Krücker, S. Xu, N. Glossop et al., Electromagnetic tracking for thermal ablation and biopsy guidance: clinical evaluation of spatial accuracy, Journal of Vascular and Interventional Radiology, vol. 18, no. 9, pp. 1141 1150, 2007. [4] J. Krücker, S. Xu, A. Venkatesan et al., Clinical utility of real-time fusion guidance for biopsy and ablation, Journal of Vascular and Interventional Radiology, vol. 22, no. 4, pp. 515 524, 2011. [5] E.Wehrenberg-Klee,T.W.Clark,S.B.Malkowiczetal., Impact on renal function of percutaneous thermal ablation of renal masses in patients with preexisting chronic kidney disease, Journal of Vascular and Interventional Radiology, vol. 23, no. 1, pp. 41 45, 2012. [6] T. Penzkofer, P. Bruners, P. Isfort et al., Free-hand CT-based electromagnetically guided interventions: accuracy, efficiency and dose usage, Minimally Invasive Therapy and Allied Technologies, vol. 20, no. 4, pp. 226 233, 2011. [7] L. Appelbaum, J. Sosna, Y. Nissenbaum, A. Benshtein, and S. N. Goldberg, Electromagnetic navigation system for CT-guided biopsy of small lesions, American Journal of Roentgenology, vol. 196, no. 5, pp. 1194 1200, 2011.