The HPLC Preparative Scale-Up of Soybean Phospholipids Application

Similar documents
Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

High-Resolution Analysis of Intact Triglycerides by Reversed Phase HPLC Using the Agilent 1290 Infinity LC UHPLC System

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

Authors. Abstract. Introduction. Environmental

Application Note. Abstract. Authors. Pharmaceutical

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Enrichment of Phospholipids from Biological Matrices with Zirconium Oxide-Modified Silica Sorbents

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

Lipid Analysis. Andréina Laffargue, IRD CRYMCEPT Montpellier workshop, October 17th Introduction to lipid structures

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

HPLC Analysis with Fluorescence Detection of Chlorophyll Degradation Products Pheophytins and Pyropheophytin in Virgin Olive Oil

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Fast Separation of Triacylglycerols in Oils using UltraPerformance Convergence Chromatography (UPC 2 )

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts

Separation of 15 Underivatized Saccharide and Sialic Acid USP Standards

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

Analytical Method Development for USP Related Compounds in Paclitaxel Using an Agilent Poroshell 120 PFP

Application Note. Author. Abstract. Introduction. Food Safety

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Application Note. Authors. Abstract. Petrochemical

application Natural Food Colorants Analysis of Natural Food Colorants by Electrospray and Atmospheric Pressure Chemical Ionization LC/MS

Author. Introduction. Abstract

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

Fat-Soluble Vitamins Analysis on an Agilent ZORBAX Eclipse PAH Polymeric C18 Bonded Column

TENOFOVIR TABLETS: Final text for addition to The International Pharmacopoeia (June 2010)

Comprehensive Two-Dimensional HPLC and Informative Data Processing for Pharmaceuticals and Lipids

Glycerolipid Analysis. LC/MS/MS Analytical Services

Using Hydrophilic Interaction Chromatography (HILIC) for the Retention of Highly Polar Analytes

CYCLOSERINI CAPSULAE - CYCLOSERINE CAPSULES (AUGUST 2015)

Converting a CHP Method for Insulin to Agilent Poroshell 120 Columns

ZIDOVUDINE, LAMIVUDINE AND ABACAVIR TABLETS Draft proposal for The International Pharmacopoeia (September 2006)

using the Agilent 7696A Sample Prep

Time-of-Flight LC/MS Identification and Confirmation of a Kairomone in Daphnia magna Cultured Medium. Application. Authors. Abstract.

HPLC SEPARATION OF PHOSPHOLIPIDS

SeQuant ZIC -HILIC For all who expect more...

Author. Introduction. Small Molecule Pharmaceuticals & Generics

RITONAVIRI COMPRESSI RITONAVIR TABLETS. Final text for addition to The International Pharmacopoeia (July 2012)

A Novel HILIC Column for High Speed N-linked Glycan Analysis

The Raptor HILIC-Si Column

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Chromatography Vacuum Ultraviolet Spectroscopy

Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

Determination of Aflatoxins in Food by LC/MS/MS. Application. Authors. Abstract. Experimental. Introduction. Food Safety

Phospholipid characterization by a TQ-MS data based identification scheme

SeQuant ZIC -HILIC For all who expect more...

Automated Purification and Analytical Reinjection of a Small Molecule Drug, Probenecid, on a Gilson LC/MS Dual Function System

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON LAMIVUDINE TABLETS

Effective use of Pharmacopeia guidelines to reduce cost of chromatographic analysis for Fluticasone propionate

DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA EFAVIRENZ, EMTRICITABINE AND TENOFOVIR TABLETS

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

Lipids Analysis. Lipids

Independent Column Temperature Control Using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Lipid Analysis by Evaporative Light Scattering Detection

Lipid Class Separation Using UPC 2 /MS

Agilent 1260 Infinity Analytical SFC System

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column

LC/MS Analysis of Various Hydrophilic Compounds Using a Polymer-Based Amino Column - Shodex TM HILICpak TM VG-50 2D

Thermo Scientific. GlycanPac AXR-1. Column Product Manual. P/N: April, Part of Thermo Fisher Scientific

EMTRICITABINE AND TENOFOVIR TABLETS

CHAPTER I. Introduction. Lipids play a variety of roles in biological systems. They are molecules that are

Systematic Evaluation of Solid Phase Extraction (SPE) Chemistries for the Determination of Acidic, Neutral, and Basic Drugs

Extended Application Note

Fast determination of residual glycerol and glycerides in biodiesel by SFC/MS using the Agilent 1260 Infinity Analytical SFC System

DRAFT PROPOSAL FOR THE INTERNATIONAL PHARMACOPOEIA: CARBAMAZEPINI COMPRESSI - CARBAMAZEPINE TABLETS

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column

Tenofovir disoproxil fumarate (Tenofoviri disoproxili fumaras)

Analysis of HMF by HPLC

Draft monograph for inclusion in. The International Pharmacopoeia. Dextromethorphani solutionum peroralum - Dextromethorphan oral solution

Application Note. Authors. Abstract. Food

Isolation of five carotenoid compounds from tangerine tomatoes

PHOTOCATALYTIC DECONTAMINATION OF CHLORANTRANILIPROLE RESIDUES IN WATER USING ZnO NANOPARTICLES. DR. A. RAMESH, Ph.D, D.Sc.,

SUPPLEMENTARY DATA. Materials and Methods

Title Revision n date

Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS

GlycanPac AXH-1 Columns

Analytical and Preparative SFC Columns

EXPERIMENT 13: Isolation and Characterization of Erythrocyte

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

ARTESUNATE TABLETS: Final text for revision of The International Pharmacopoeia (December 2009) ARTESUNATI COMPRESSI ARTESUNATE TABLETS

Direct Analysis of Folic Acid in Digestive Juices by LC/TOF-MS Application

PROPOSAL FOR REVISION OF MONOGRAPH PUBLISHED IN The International Pharmacopoeia: REVISION OF ph test ABACAVIR ORAL SOLUTION (JULY 2012)

2D-LC as an Automated Desalting Tool for MSD Analysis

Quantitative Analysis of Carbohydrates and Artificial Sweeteners in Food Samples Using GFC- MS with APCI Interface and Post-column Reagent Addition

Analysis of several common. organic acids in tobacco leaf, snus, and moist snuff

Separation of Macrocyclic Lactones (Avermectins) on FLARE C18 MM & FLARE C18+ Columns

Sample Preparation is Key

Simultaneous Analysis of Active Pharmaceutical Ingredients and Their Counter-Ions Using a Mixed-Mode Column

Thermo Scientific. Acclaim C30 Columns. Product Manual. P/N: February Part of Thermo Fisher Scientific

Comparison of a UPLC Method across Multiple UHPLC Systems

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

CLINDAMYCIN PALMITATE

The phosphate group replaces the fatty acid on C number 3 of a triacylglycerol molecule O O CH 2 O C R CH 2 O P O X OH.

Determination of Tetracyclines in Chicken by Solid-Phase Extraction and High-Performance Liquid Chromatography

Transcription:

The HPLC Preparative Scale-Up of Soybean Phospholipids Application Food and Flavors Authors Cliff Woodward and Ron Majors Agilent Technologies, Inc. 285 Centerville Road Wilmington, DE 1988-161 USA Abstract Classes of major soybean phospholipids can be separated from a crude mixture, under both analytical and preparative conditions, using a matched set of columns, normal phase conditions, and a small amount of volatile salt added to the mobile phase. At higher concentrations of phosphatidylcholine, however, two distinct peaks, whose areas changed with the mass injected, were observed. These two peaks were determined by mass spectrometry to be the same or, at least, very similar. Introduction Phospholipids have an important role in animal and plant systems since they are commonly found in cellular and intracellular membranes where they fulfill the role of primary structural components. Phospholipids are also widely used in the food, cosmetic and industrial manufacturing industries and account for 8% of worldwide plant oil production. The structure of a typical phospholipid is depicted in Figure 1. Phospholipids are the major components of soybean lipids and are often desired in a pure form as standards for chromatography and other studies. Preparative amounts are therefore needed and HPLC is often the first choice for this purpose. This application note describes both the development of an analytical normal phase HPLC separation of the most common phospholipids found in crude soy lecithin extracts and the possibility of performing a linear scale-up to preparative columns filled with the same packing. Properties of Phospholipids From a chromatographic viewpoint, phospholipids are in a class of polar lipids and are amphipathic. They contain an ionic portion due to the phosphate moiety and they may contain additional hydrophilic or ionic portions due to the presence of choline, serine, ethanolamine, myo-inositol and/ or glycerol functionality. All phospholipids contain a hydrophobic portion due to the presence of the fatty acid moieties and their complexity results from the two fatty acids on the 1,2-diacylglycerol portion. These fatty acid moieties may be the same or different. The main phospholipids present in soybeans along with their abbreviations are shown in Table 1. In a typical soybean extract, about half of the extractables are phospholipids with neutral lipids, mainly triglycerides, and carbohydrates making up the remainder. Since phospholipids are generally very complex and difficult to prepare in a pure form, most commercially prepared mixtures contain varying amounts of these extractables. Phospholipids generally have low water solubility and can be dissolved in nonpolar solvents. Since their ultraviolet (UV) absorbance arises only from the double bond in the fatty acid portion, they provide low response with UV detectors. Refractive index-, evaporative light scattering- or mass spectrometric-detectors are, therefore, more frequently used.

Glycerol Phosphatidylcholine (PC) Fatty acid C O CH 2 O C O CH O O CH 2 O P O CH 2 O _ Phosphate Choline CH 3 CH 2 N + CH 3 CH 3 Hydrophobic tail Polar head group Figure 1. Structure of typical phospholipid. Table 1. Important Phospholipids in Commercial Plant Oil Extracts Phospholipid Abbreviation Phosphatidylethanolamine PE Phosphatidylserine PS Phosphatidylinositol PI Phosphatidylcholine PC Chromatographic Separation of Phospholipids Because of their chemical properties, phospholipids may be separated by both normal phase and reversed-phase chromatography. For preparative scale-up from analytical dimensions, the use of silica normal phase chromatography has advantages when compared to reverse phase chromatography (RPC): 1) The mobile phases are more volatile; thus, collected fractions can easily be recovered. 2) The viscosity of the common normal phase solvents is low, allowing higher flow rates to be employed, thereby improving throughput. 3) Using the adsorption mechanism, class separations of major phospholipids are easier to perform compared to RPC. 4) In normal phase work, since phospholipids absorb poorly in the UV, detection is better accomplished using the evaporative light scattering detector (ELSD) that allows gradient systems to be employed and provides excellent sensitivity in organic mobile phases. Row and Kang demonstrated successful separation of phospholipids using normal phase chromatography but they operated at 28- or 21-nm in the UV [1]. In their work, they used a ternary gradient system. However, since our preparative chromatograph employed a binary pumping system, we had to adapt our gradient conditions for both the analytical and preparative chromatography accordingly. Evaporative Light Scattering as a Detection Principle The ELSD detects compounds based somewhat on a compound's mass, and not on its spectroscopic characteristics. It is nearly universal and it responds to all compounds as long as they are less volatile than the mobile phase. As seen in Figure 2, the chromatographic flow stream enters the heated nebulizer and is nebulized using a stream of dry nitrogen; the resultant mist containing the analyte is sequentially directed through the heated evaporator where is loses the solvent and becomes a particle, which is then directed through the polychromatic light beam. The particles cause photon scattering as they traverse the path of the light beam. The signal generated is proportional to the analyte particle size present. The solvent is evaporated and does not interfere with the analyte measurement. The presence or absence of chromophores, fluorophores, or electroactive groups has no impact on the signal. The ELSD can be optimized for analyte sensitivity and baseline stability by adjustment and optimization of evaporation and nebulization temperatures. For preparative applications, the upper flow rate capability of the ELSD is 4 ml/min, and thus requires post-column splitting of the chromatographic effluent. 2

Detection chamber Light source Heated nebulizer Heated evaporating chamber Syphon Figure 2. Schematic of ESA Chromachem ELSD. Scaling Up from Analytical to Preparative Chromatography In scaling up to a preparative separation, it is recommended that initial method development use an analytical-sized column. Since flow rates and sample injection sizes are lower than for prep, solvent costs are lowered and less sample is used, which is important when sample mass is limited or for samples that are relatively expensive. Once the separation is optimized using an analytical (or scalar) column, linear scale-up to a preparative column of the same materials is easily achieved. Initially, we adjusted our gradient to approximate the one used by Row and Kang [1]. Using 2.5% methanol (MeOH), 2.5% isopropanol (IPA), and 95% hexane as solvent A, and using 4% IPA and 6% MeOH as solvent B, we achieved a successful separation of the four standard phospholipids but found that our retention times often varied from run to run. In earlier normal phase work [2], Christie found that the addition of a small amount of ionic material to the most polar solvent, in his case, isopropanol-water, gave better resolution and extended the column life. However, several of his suggested ionic additives were nonvolatile. Since the ELSD requires the use of a volatile mobile phase, we chose a low concentration of ammonium acetate (1-mM) but added it to both mobile phase constituents. Higher concentrations could be used to give improved retention behavior, but at >1-mM levels there is danger of salt precipitation, and should therefore be avoided. Using this hexane-methanol-isopropanol-ammonium acetate (NH 4 Ac) gradient system allowed us to fine tune the separation, especially relative to isocratic methods of the past. The final set of chromatographic conditions are depicted in Table 2. Figure 3 shows the optimized separation on a silica-gel scalar column as a function of the injected amount of phospholipid standards: PE, PS, PI, and PC. The small amount of NH 4 Ac, a volatile salt, gave much better retention behavior for PC, a zwitterion with a quaternary amine functionality (see Figure 1). The salt addition also assured minimal retention variation with load at low injection mass. Surprisingly, at higher injection masses, we observed two peaks for PC. 3

Table 2. Chromatography Analytical runs: Column: Mobile phase: Final Set of Experimental Conditions Agilent Prep SIL Scalar Column, 4.6 15 mm, 1 µm A = 95:2.5:2.5% hexane-isopropanolmethanol B = 4:6% isopropanol-methanol Both solvents contain 1-mM ammonium acetate 1.5 ml/min Flow rate: Gradient: Time, min %B 2 18.7 2.2 1 25 1 25.1 Temperature: Ambient Preparative runs: Column: Agilent Prep SIL Column, 21.2-15 mm, 1 µm Mobile phase: Same as above Flow rate: 31.9 ml/min Gradient: Same as above Temperature: Ambient Detection Detector: Chromachem ELSD, (ESA, Cambridge, MA) Temp. nebulizer: 45 C Temp. evaporator: 65 C Filter: High Gas pressure: 23 psi Attenuation: 1 Sample Dissolved in: CHCl 3:Hexane 2:1 Analytical: 5 mg/ml Prep: 25 mg/ml mv 2 PE PS PI 175 15 PC 125 Z2 1 75 Z1 5 µg 5 25 5 µg 1 µg 2 µg 25 µg 5 1 15 2 25 3 35 min Figure 3. Agilent prep 4.6 15 mm, 1 µm Mixture of soy phospholipid standards. - Loadability. Study of PC Elution Behavior As stated above, at higher injection masses, in Figure 3, we observed two peaks for PC, one eluting at a 16.1 minute (PC-Z1) and the second 21.8 minutes (PC-Z2). As sample mass increased (above 75 µg on the 4.6 15-mm column), the latter peak increased while the former decreased but the total area increased as expected. As can be seen in Figure 4, with increasing sample mass, the peak areas for all of the phospholipids, including PC, increased. Typical of the ELSD, response was nonlinear, especially at the lowest concentrations (less than 5 µg); however, since the double peak phenomenon occurred with PC only, we suspected that it had something to do with the zwitterion functionality. We, therefore, investigated this aspect more thoroughly. 4

8 Area counts 7 6 5 4 3 PE PI PS 2 1 PC (Total) PC-Z2 PC-Z1-1 1 2 3 4 5 6 Micrograms injected Figure 4. Area counts for ELSD versus micrograms injected (analytical column, 4.6 15 mm). Using liquid chromatography/mass spectrometry (LC/MS), we attempted to determine the structure of the two peaks observed with PC. Since in normal phase chromatography ionization could be a problem, we first investigated MS ionization approaches. MS sensitivity is related to the formation of stable ions in the source. Using electrospray ionization, a minimal signal was observed; so, we resorted to atmospheric pressure photometric ionization (APPI), a more useful ionization technique for neutral organic compounds. Since PC is zwitterionic, its natural ionic character allowed it to be detected in either positive or negative APPI. To enhance sensitivity, acetone and toluene dopants were added post-column as a 5:5 mixture. From other studies [3] it is known that soybean PC consists of multiple phospholipid entities with varying fatty acid content. We were able to assign certain m/z values to individual PCs on the basis of the fatty acid portion of the 1,2- diacylglycerol functionality. Figure 5 shows the raw, unsmoothed mass spectral data of the two PC peaks (PC-Z1 and PC-Z2). 6 5 4 C18:2, C18:2 PC (dilinoleoyl) MW 781 C16:, C18:2 PC (palmitoyl/lineoleoyl) MW 757 C18:1, C18:2 PC (oleoyl/linoleoyl) MW 783 C18:2, C18:3 PC (lineoleoyl/linolenoyl) MW 779 3 PC-Z1 2 PC-Z2 1 16 18 2 22 24 min Figure 5. Overlay of the most common phosphatidylcholine (lecithin) (PC) species present in crude soy lecithin Extracted ions from LC/MS (unsmoothed data). 5

Clearly, one can determine that there was partial resolution by the type of fatty acid moiety present. Note that the same species are present in both peaks PC-Z1 and PC-Z2. Furthermore, as can be seen in Table 3, if one explores the peak height ratios of the individual PCs within the PC-Z1 and PC-Z2 umbrellas, they are similar enough to deduce that the peaks are the same. In addition, based on fatty acid ratio studies of the individual PCs in purified PC as determined by gas chromatography [3], for a first approximation, the fatty acid composition of the PCs depicted in Figure 5 are similar. This observation also led us to believe that these two PC peaks were the same compound (Table 3). Our speculation is that the two peaks are a result of the competitive silanol binding between the positively charged ammonium ion in the mobile phase modifier and the positively charged portion of PC. At low PC concentrations, the ammonium binds more to the silica silanol groups and the concentration of unbound PC-Z1 is insufficient to displace it. As the concentration of PC increases, however, some of the PC adsorbs to the silanol sites in competition with the ammonium ion. This strongly held PC (identified as PC-Z2) elutes only at very high concentrations of IPA- MeOH. As larger amounts of crude soy lecithin are injected, the available PC increases so that more binds to the surface, resulting in an increase in its elution peak, PC-Z2. If one estimates the total amount of available silanol groups (assuming 8-µmoles/m 2 for a totally hydroxylated surface) on the surface of the packing inside of the column, the number of µmoles of PC potentially bound to the surface at saturation (around 6 µmoles) is consistent with the calculated level of binding sites available. Preparative Studies of Crude Soy Lecithin Figure 6 depicts the preparative scale-up of the same phospholipid standard to a 21.2-mm 15-mm column packed with the same material as the scalar column. The flow rate was increased to 31.88 ml/min to achieve the same linear velocity and separation time as under the analytical conditions. The same relative amounts injected onto the preparative column allowed as much as 2 mg of total phospholipid before the ELSD detector showed signs of overload as noted in the upper chromatogram. Note that the same PC-Z1 and PC-Z2 peaks can be observed in the chromatogram of Figure 6. Table 3. Table of Relative Extracted Ion Currents Fatty acid Weight,% Relative Relative Relative composition typical soybean* weight.% height PC-Z1 height PC-Z2 C18:2/ C18:2 PC 47.5 5 39 45 (dilinoleoyl) C16:/ C18:2 PC 3. 31 31 24 (palmitoyl/ lineoleoyl) C18:1/ C18:2 PC 1.5 13 28 21 (oleoyl/ linoleoyl) C18:2/ C18:3 PC 6.1 6 3 9 (lineoleoyl/ linolenoyl) *B. Herslof, U. Olsson, and P. Tingvall in I. Hanin and G. Pepeu,Eds. (199). Phospholipids: Biochemical, Pharmaceutical, and Analytical Considerations, Plenum Press, New York, pp. 295 298. 6

mv 2 PE PS PI PC 175 15 Z1 Z2 125 1 75 5 25 2 mg 4 mg 1 mg 2 mg 1 mg 5 1 15 2 25 3 35 min Figure 6. Agilent prep 21.2 15 mm, 1-µm Mixture of soy phospholipid standards. - Loadability. Finally, in Figure 7, we investigated the scale-up of a 2-mg injection of a crude soy lecithin sample which, by overlaying the standards run under the same conditions, revealed the presence of three phospholipids, including PE, PI, and PC. No PS is seen in this crude extract. Norm. 12 Standard run conditions 1 Crude soy PC 8 Pure soy PC 6 Pure soy PI 4 Pure soy PS 2 Pure soy PE 5 1 15 2 25 min Figure 7. Agilent prep SIL, 21.2 15 mm,1 µm Crude soy lecithin (PC) and pure individual standards. 7

Conclusion www.agilent.com/chem These experiments show that the classes of major soybean phospholipids can be successfully separated from a crude mixture, under both analytical and preparative conditions, using a matching set of columns, normal phase conditions, and a small amount of volatile salt added to the mobile phase to cut down on possible electrostatic interactions. At higher concentrations of PC, however, two distinct peaks, whose areas change with mass injected, are observed. By mass spectrometric measurements of the eluted peaks, these two peaks were determined to be the same or, at least, very similar. It is suspected that PC is strongly binding to the silica silanol group in competition with the ammonium acetate, a mobile phase modifier, and is eluting at high gradient solvent polarity. Acknowledgement We would like to thank ESA for the loan of the Chromachem ELSD detector for the duration of this study. References 1. Kyung Ho Row and Duk Hui Kang, (23) Amer. Biotech. Lab., 21(9) 4 43. 2. W. M., Christie, (1986) J. Chromatogr. 361, 396 399. 3. B. Herslof, U. Olsson, and P. Tingvall in I. Hanin and G. Pepeu, Eds. (199) Phospholipids: Biochemical, Pharmaceutical, and Analytical Considerations, Plenum Press, New York, pp. 295 298. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 25 Printed in the USA April 28, 25 5989-2848EN