Methods Forty patients representing all consecutive patients referred to. our laboratory on randomly selected days were chosen for

Similar documents
Doppler Basic & Hemodynamic Calculations

found that some patients without stenotic lesions had blood velocity or pressure measurement across the

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics

PART II ECHOCARDIOGRAPHY LABORATORY OPERATIONS ADULT TRANSTHORACIC ECHOCARDIOGRAPHY TESTING

Uncommon Doppler Echocardiographic Findings of Severe Pulmonic Insufficiency

Appendix II: ECHOCARDIOGRAPHY ANALYSIS

Giovanni Di Salvo MD, PhD, FESC Second University of Naples Monaldi Hospital

Diagnostic approach to heart disease

Cardiac ultrasound protocols

Pulsed Doppler echocardiographic determination of stroke volume and cardiac output: clinical validation of two new methods using the apical window

METHODS WILLIAM J. STEWART, MD,* KATHLEEN A. GALVIN, LINDA D. GILLAM, MD, FACC, DAVID E. GUYER, MD, FACC, ARTHUR E.

Adult Echocardiography Examination Content Outline

Right Ventricle Steven J. Lester MD, FACC, FRCP(C), FASE Mayo Clinic, Arizona

P = 4V 2. IVC Dimensions 10/20/2014. Comprehensive Hemodynamic Evaluation by Doppler Echocardiography. The Simplified Bernoulli Equation

Imaging Guide Echocardiography

HEMODYNAMIC ASSESSMENT

Department of Pediatrics and Child Health, Kurume University School of Medicine, Kurume, 830 Japan. Received for publication October 26, 1992

British Society of Echocardiography

left heart catheterization

ECHOCARDIOGRAPHY DATA REPORT FORM

Echo Doppler Assessment of Right and Left Ventricular Hemodynamics.

Comprehensive Hemodynamics By Doppler Echocardiography. The Echocardiographic Swan-Ganz Catheter.

Certificate in Clinician Performed Ultrasound (CCPU) Syllabus. Rapid Cardiac Echo (RCE)

Doppler echocardiographic evaluation of the normal

Noninvasive estimation of right ventricular systolic

ASCeXAM / ReASCE. Practice Board Exam Questions Monday Morning

DOPPLER HEMODYNAMICS (1) QUANTIFICATION OF PRESSURE GRADIENTS and INTRACARDIAC PRESSURES

The Doppler Examination. Katie Twomley, MD Wake Forest Baptist Health - Lexington

cardiac imaging planes planning basic cardiac & aortic views for MR

Pulmonary valve echo motion in pulmonary

Doppler Color Flow Imaging #4

Pediatric Echocardiography Examination Content Outline

Pulsed Wave Doppler and Color Flow Doppler Evaluation in Healthy Dogs and Dogs with Cardiac Disease

5 Working With Measurements

B-Mode measurements protocols:

Aortic Stenosis: Spectrum of Disease, Low Flow/Low Gradient and Variants

Assessment of Aortic Blood Flow Velocities With Continuous Wave Doppler Ultrasound in the Neonate and Young Child

The background of the Cardiac Sonographer Network News masthead is a diagnostic image:

Color Doppler Echocardiographic Assessment of Valvular Regurgitation in Normal Infants

Adult Congenital Heart Disease: What All Echocardiographers Should Know Sharon L. Roble, MD, FACC Echo Hawaii 2016

Doppler-echocardiographic findings in a patient with persisting right ventricular sinusoids

Adel Hasanin Ahmed 1

Echocardiography in Adult Congenital Heart Disease

the aorta and its branch vessels were isolated and

Feasibility and variability of six methods for the echocardiographic and Doppler determination of cardiac output

Noninvasive pulsed Doppler echocardiographic. with atrial septal defect: usefulness of the right parasternal approach

Two-dimensional echocardiography with a wide angle (600) sector scanner'

Recommendations for Standards in Transthoracic Two-Dimensional Echocardiography in the Dog and Cat

Accuracy and Pitfalls of Doppler Evaluation of the Pressure Gradient in Aortic Coarctation

DOPPLER ECHOCARDIOGRAPHY DIAGNOSTIC METHODS

Pulsed and Continuous Wave Doppler Echocardiographic Assessment of Valvular Regurgitation in Normal Subjects

regurgitant fraction by two-dimensional Doppler echocardiography

Assessment of LV systolic function

M-Mode Echocardiography Is it still Alive? Itzhak Kronzon, MD,FASE. Sampling Rate M-Mode: 1800 / sec 2D: 30 / sec

Non-invasive estimation of the mean pressure

Reverse Doming of the Anterior Mitral Leaflet With Severe Aortic Regurgitation

The Newborn Transitional Circulation: A Two-Dimensional Doppler Echocardiographic Study

Introduction to Fetal Doppler Echocardiography

Adopted by Council March 2006, Revised March 2012, September 2015

HISTORY. Question: What category of heart disease is suggested by this history? CHIEF COMPLAINT: Heart murmur present since early infancy.

Quantitative Assessment of Pulmonary Hypertension in Patients With Tricuspid Regurgitation Using Continuous Wave Doppler Ultrasound

Normal pulsed wave Doppler echocardiographic parameters of Turkmen horses of Iran

Case # 1. Page: 8. DUKE: Adams

RIGHT VENTRICULAR SIZE AND FUNCTION

volume allows characterization of areas of normal and abnormal blood flow within the cardiac chambers and vessels. If an abnormal highvelocity

The impacts of pericardial effusion on the heart function of infants and young children with respiratory syncytial virus infection

The Doppler echocardiographic assessment of valvular regurgitation in normal children

Iwo-dimensional Echocardiographic Doppler Studies

HISTORY. Question: What category of heart disease is suggested by the fact that a murmur was heard at birth?

Echocardiography in Congenital Heart Disease

ICE: Echo Core Lab-CRF

Pulmonary Hypertension. Echocardiography: Pearls & Pitfalls

Echocardiographic Evaluation of the Jarvik 2000 Axial-Flow LVAD

Atrial Septal Defects

Cath Lab Essentials: Basic Hemodynamics for the Cath Lab and ICU

Journal of the American College of Cardiology Vol. 33, No. 6, by the American College of Cardiology ISSN /99/$20.

Reliability of non-invasive estimates of pulmonary hypertension by pulsed Doppler echocardiography

Right Heart Evaluation ASE Guidelines Review. Chris Mann RDCS, RCS, FASE Faculty, Echocardiography Pitt Community College Greenville, NC

Tricuspid and Pulmonic Valve Disease

American Society of Echocardiography

TEE Outside of the Cardiac OR

Doppler Echocardiographic Studies of Diastolic Function in the Human Fetal Heart: Changes During Gestation

HISTORY. Question: What type of heart disease is suggested by this history? CHIEF COMPLAINT: Decreasing exercise tolerance.

Back to Basics: Common Errors In Quantitation In Everyday Practice

Noninvasive Quantification of Stenotic Semilunar Valve Areas by Doppler Echocardiography

Little is known about the degree and time course of

Intro to Bedside Ultrasound. Cardiac Ultrasound

PRACTICAL GUIDE TO FETAL ECHOCARDIOGRAPHY IC Huggon and LD Allan

Aortic stenosis (AS) is common with the aging population.

ADULT CONGENITAL HEART DISEASE. Stuart Lilley

Vevo 2100 System Cardio Measurements. Dieter Fuchs, PhD FUJIFILM VisualSonics, Inc.

Policy #: 222 Latest Review Date: March 2009

Quantitation of right ventricular dimensions and function

좌심실수축기능평가 Cardiac Function

Right Heart Catheterization. Franz R. Eberli MD Chief of Cardiology Stadtspital Triemli, Zurich

ViosWorks: A Paradigm Shift in Cardiac MR Imaging

The Normal Echocardiogram

Non-invasive assessment of aortic stenosis by

Cardiology Fellowship Manual. Goals & Objectives -Cardiac Imaging- 1 P a g e

TAVR TTE INTERROGATION BY ALAN MATTHEWS

Transcription:

DIAGNOSTIC METHODS DOPPLER ECHOCARDIOGRAPHY Comparison of multiple views for the evaluation of pulmonary arterial blood flow by Doppler echocardiography GEORGE W. LIGHTY, JR., M.D., PH.D., ANTHONY GARGIULO, R.D.M.S., ITZHAK KRONZON, M. D., F. A. C. C., AND FRANK POLITZER, M. D. ABSTRACT Forty adult patients underwent Doppler and two-dimensional echocardiographic examination of the pulmonary artery from multiple views to determine the variability in the magnitude of Doppler-determined flow velocity and pulmonary arterial diameter from various echocardiographic windows. Flows were recorded from two or more views in 32 patients (80%). Twelve of these patients (38%) had flow velocities recorded from two or more views that were within 6% of each other. Twenty of these patients (62%) had view-dependent differences in measured flow velocity ranging from 7% to 48%. The commonly used parasternal short-axis view yielded the highest pulmonary arterial flow velocity in only 35% of the patients studied. Determinations of pulmonary arterial blood flow can vary markedly when measured from different sites, and this is presumably due to varying ability to approximate a zero-degree Doppler angle from different views. Measurement of pulmonary arterial flow velocity should be attempted from multiple views, and the highest flow velocity should be selected as that obtained with the best zero-degree Doppler angle approximation. Circulation 74, No. 5, 1002-1006, 1986. THE MEASUREMENT of pulmonary arterial blood flow velocity by Doppler echocardiography can be useful in the assessment of pulmonic valvular disease,, 2pulmonary hypertension,3 cardiac output,4 and intracardiac shunt magnitude.5 For quantitative measurements of blood flow velocity, the Doppler equation requires the precise determination of the Doppler angle or orientation of the ultrasound beam parallel to blood flow. It has been shown that multiple ultrasound views are often needed to obtain a reliable measure of aortic flow velocity.6 Since similar anatomic constraints are present for measurement of pulmonary arterial blood flow velocity, we evaluated and compared images from three two-dimensional echocardiographic views that potentially permit visualization of the pulmonary artery while simultaneously allowing nearparallel orientation of the ultrasound beam to the direction of blood flow, thus minimizing Doppler angledependent errors in measurement of flow velocity. From the Department of Medicine, New York University Medical Center, New York. Address for correspondence: Itzhak Kronzon, M.D., New York University Medical Center, 560 First Ave., Suite 2E, New York, NY 10016. Received April 22, 1986; revision accepted Aug. 7, 1986. 1002 Methods Forty patients representing all consecutive patients referred to our laboratory on randomly selected days were chosen for study. They ranged in age from 16 to 84 years (mean 59 + 16 years); sixteen (40%) were female patients. Two-dimensional images and Doppler flow signals from the parasternal short-axis and the parastemal right ventricular outflow tract views with patients in the left lateral decubitus position, and from the subcostal short-axis view with patients in the supine position, were recorded with an ATL Mark 600 Duplex ultrasound instrument (Advanced Technology Laboratories, Inc., Bellevue, WA) (figures 1 and 2). In the commonly used parastemal short-axis view, the coaptation point of the pulmonic valve leaflets and the greater curvature of the pulmonary artery are rarely visualized. The parasternal right ventricular outflow tract view provides better visualization of the pulmonic valve and the greater curvature of the pulmonary artery; it is obtained by angulating the transducer anteriorly and superiorly toward the right shoulder in a plane approximately midway between the parastemal long- and shortaxis views. In the subcostal short-axis view the pulmonic valve plane is seen as a horizontal structure with the vessel walls clearly visualized. The parastemal right ventricular outflow tract and subcostal short-axis views appear to provide more optimal imaging of the pulmonary artery for measurement of its diameter at a known distance beyond the plane of the pulmonic valve. By use of the cross-sectional area (CSA) of the pulmonary artery calculated from the diameter measurement and the flow velocity integral (FVI) obtained from the Doppler recording of flow velocity, stroke volume (SV) may be calculated with the equation SV - CSA x FVI. A 9 mm Doppler sample volume was positioned at approxi-

DIAGNOSTIC METHODS-DOPPLER ECHOCARDIOGRAPHY mately the midline of the pulmonary artery, 1 cm distal to the coaptation point of the pulmonic valve leaflets in each view. Scan planes and Doppler sample volume position were adjusted in each view until the highest velocity flow signal was obtained. Peak Doppler flow velocities and pulmonary arterial diameters were measured with the Cardiologic Analysis System (SONY Medical Systems, Park Ridge, New Jersey). Pulmonary arterial diameters were measured as inside-to-inside distance at peak systole, 1 cm distal to the coaptation point of the pulmonic valve leaflets. Results Doppler recordings of good quality were obtained in 31 patients from the parasternal short-axis view, in 33 patients from the parasternal right ventricular outflow tract view, and in 21 patients from the subcostal shortaxis view (table 1). Doppler recordings could not be W"0;:: i: -~ ~ ~ ~....,......E..E :::: :D.. B. FIGURE 1. Two-dimensional echocardiograms demonstrating visualization and pulsed wave Doppler interrogation of the pulmonary artery from the parastemal short-axis (top), parastemal right ventricular outflow tract (middle), and subcostal short-axis (bottom) views. Ao aorta; LA left atium; LV left ventricle; PA pulmonary artery; PV pulmonic valve; PW - left ventricular posterior wall; RA right atrium; RV right ventricle; S interventricular septum; TV tricuspid valve. Vol. 74, No. 5, November 1986.c FIGURE 2. Pulsed Doppler waveforms of pulmonary arterial blood flow velocity from a single patient recorded from the parasternal shortaxis (A), parasternal right ventricular outflow tract (B), and subcostal short-axis (C) views. Horizontal scale: I sec between lines. Vertical scale: 0.2 m/sec between lines; 0 m/sec line is in boldface. 1003

LIGHTY et al. TABLE 1 Doppler measurements of pulmonary arterial flow in 40 patients from three transducer positions Highest PA Flows flow velocity Transducer position recorded recorded None 1 NA PSAX only 1 4 PRVOT only 2 12 SCSAX only 4 9 PSAX and PRVOT only 15 8 PSAX and SCSAX only 1 1 PRVOT and SCSAX only 2 3 PSAX, PRVOT, and SCSAX 14 1 NA not applicable; PA pulmonary arterial; PSAX parasternal short axis; PRVOT parastemal right ventricular outflow tract; SCSAX subcostal short axis. obtained from more than one view in seven patients; recordings from a single view only were obtained in one patient from the parasternal short-axis view, in two patients from the parasternal right ventricular outflow tract view, and in four patients from the subcostal short-axis view. No single view consistently provided the highest peak flow velocity. The highest flow velocity was recorded from the parasternal short-axis view in 14 of 31 patients in whom flow recordings from this view could be obtained, from the parasternal right ventricular outflow tract view in 24 of 33 patients, and from the subcostal short-axis view in 14 of 21 patients. Based on the cosine function in the Doppler equation, an error of up to ± 20 degrees in the assumed zero-degree Doppler angle results in only a 6% error in measured flow velocity. Due to the nonlinearity of the cosine function, an error in the assumed zero-degree Doppler angle of greater than 20 degrees results in proportionately much greater error in measured flow velocity. We therefore considered flow velocities within 6% of each other measured from different views in the same patient to represent equivalent flow velocities and those with a greater than 6% difference to represent view-dependent-and therefore Doppler angle-dependent-differences in measured flow velocity. Thirteen patients had flow velocities recorded from two or more views that were within 6% of each other; this was achieved from two views in 12 patients and from all three views in one patient (table 1). Peak flow velocity ranged from 49 to 159 cm/sec, with a mean peak flow velocity of 94 + 26 cm/sec. Twenty patients had view-dependent differences in measured flow velocity ranging from 7% to 48%. It is thus likely that in these patients, the actual Doppler angle was 1004 TABLE 2 Pulmonary arterial imaging in 40 patients from three transducer positions Solitary view View Adequate Adequate for PA (n 40) PA image PA diameter diameter PSAX 29 0 PRVOT 32 30 15 SCSAX 22 20 5 PA pulmonary arterial; PSAX parasternal short axis; PRVOT parastemal right ventricular outflow tract; SCSAX subcostal short axis. greater than 20 degrees from the assumed zero-degree angle in the view providing the lower flow velocity. Satisfactory Doppler recordings could not be obtained in one patient. Of the 40 patients studied, clinically satisfactory pulmonary arterial images were obtained in 29 patients from the parasternal short-axis view, in 32 patients from the parasternal right ventricular outflow tract view, and in 22 patients from the subcostal short-axis view (table 2). Of these images, pulmonary arterial diameters were measurable on none of those from the parasternal short-axis view, on 30 of those from the parasternal right ventricular outflow tract view, and on 20 of those from the subcostal short-axis view. Pulmonary arterial diameters were measurable from both the parasternal right ventricular outflow tract and subcostal short-axis views in 15 patients; for these 15 patients, excellent correlation was found between diameter measurements from these two views, with a correlation coefficient of.92 (figure 3). In 15 patients, pulmonary arterial diameters could be measured only from the parasternal right ventricular outflow tract view and in five patients they could be measured only from the subcostal short-axis view. The pulmonary arterial diameter could not be measured in five patients from any view. Pulmonary arterial diameters in the study group ranged from 1.8 to 4.5 cm, with a mean diameter of 3.1 ± 0.4 cm. Both pulmonary arterial diameter and satisfactory Doppler recordings of flow velocity were obtained for 35 of the 40 patients (88%), thus permitting the calculation of volume blood flow. Discussion Combined two-dimensional imaging and pulsed Doppler echocardiography allows rapid positioning of the Doppler sample volume in the proximal pulmonary artery from multiple views; this is particularly helpful in the measurement of pulmonic flow velocity from the subcostal short-axis approach. The commonly used

DIAGNOSTIC METHODS-DoPPLER ECHOCARDIOGRAPHY 4 0 i 3 0 "/ 0 > 2 0. / Y m 0.895 X +0.3 R - 0.92 FIGURE 3. Correlation of pulmonary arterial diameters measured from the parasternal right ventricular outflow tract view and from the subcostal shortaxis view in the 15 patients in whom both views yielded a measurable diameter. PRVOT parasternal right ventricular outflow tract; SCSAX subcostal short axis. 1 2 SCSAX (CM) parasternal short-axis view yielded the highest pulmonic flow velocity in only 35% of our patient population. The parasternal right ventricular outflow tract view provided a higher recorded pulmonic flow velocity than the parasternal short-axis view in 38% of the patients studied. In an additional nine patients (23%), the highest flow velocities could be obtained only by the subcostal approach. Our findings are very similar to those reported by Vijayaraghavan et al.6 on the use of multiple views to obtain the highest Doppler aortic flow velocity. Thus the need to measure Doppler flow velocity from multiple views to obtain an adequate quantitative assessment of flow cannot be overemphasized. Although the highest obtained flow velocity in each view was attributed to the closest approximation of a zero-degree Doppler angle, it is acknowledged that some variation in measured blood flow velocity within the pulmonary artery may be due to a nonlinear velocity profile. Since our Doppler measurements of pulmonic flow velocity were not compared with measurements of pulmonic flow by alternative methods, the accuracy of the data may be questioned. However, the maximum flow velocities we obtained compare favorably with those reported by other investigators.7'8 Other investigators have suggested that continuous-wave Doppler measurements of aortic blood flow may be superior to Vol. 74, No. 5, November 1986 those obtained by pulsed Doppler techniques.9 These arguments are largely based on the ability to either better approximate a zero Doppler angle with the use of the Doppler flow signal for guidance of transducer positioning (such as the right parasternal border view for aortic stenosis), or the elimination of flow-signal aliasing that occurs when high-velocity blood flow is measured with the pulsed Doppler recording technique. In adults, pulmonic stenosis is a rarity, and even in patients with significant pulmonic insufficiency, it is unusual to obtain flow velocities of sufficient magnitude that they cannot be resolved with use of either standard or high pulse-repetition frequency pulsed Doppler echocardiography. However, smaller nonimaging Doppler transducers using either continuous or pulsed mode might well permit alternative transthoracic ultrasound windows with more optimum Doppler flow signals than can be obtained with an imaging scanhead in some patients. This aspect deserves further investigation. We were able to measure pulmonary arterial diameters in 35 of the 40 patients studied using either the parastemal right ventricular outflow tract or the subcostal short-axis views. Thus the diameter measurement was the major limiting factor in our ability to estimate volume blood flow in the pulmonary artery, reducing our theoretical volume flow measurement group from 39 to 35 patients (39 patients had satisfac- 1005

LIGHTY et al. tory Doppler recordings). In both views, the vessel walls are relatively parallel to the ultrasound beam, thus reducing measurement resolution, which becomes dependent upon geometry of the lateral and azimuthal beams. In our experience, the subcostal two-dimensional view of the pulmonary artery, when obtainable, provided the best images of the vessel. However, there was a high correlation between blinded measurements of vessel diameter from the two views. While not used in the present study, additional images of the pulmonary artery for diameter measurements might be obtained from the suprasternal approach. However, it is frequently difficult to obtain high-quality images from this view in adult patients. The data from this study lead us to conclude that determinations of pulmonary arterial blood flow velocity can vary markedly when it is measured from different sites, due presumably to varying ability to approximate a zero-degree Doppler angle from multiple views. Therefore, measurements of pulmonary arterial flow velocity should be attempted from all three (parasternal short-axis, parasternal right ventricular outflow tract, and subcostal short-axis) views, and the highest flow velocity should be selected. References 1. Oliveira Lima C, Sahn DJ, Valdes-Cruz LM, Goldberg SJ, Vargas Barron J, Allen HD, Grenadier E: Noninvasive prediction of transvalvular pressure gradient in patients with pulmonary stenosis by quantitative two-dimensional echo Doppler studies. Circulation 67: 866, 1983 2. Miyatake K, Okamoto M, Kinoshita N, Matsuhisa M, Nagata S, Beppu S, Park Y, Sakakibara H, Nimura Y: Pulmonary regurgitation studied with the ultrasonic pulsed Doppler technique. Circulation 65: 969, 1982 3. Kitabatake A, Inoue M, Asao M, Masuyama T, Tanouchi J, Morita T, Mishima M, Uematsu M, Shimazu T, Hori M, Abe H: Noninvasive estimation of pulmonary hypertension by a pulsed Doppler technique. Circulation 68: 302, 1983 4. Goldberg SJ, Sahn DJ, Allen HD, Valdes-Cruz LM, Hoenecke H, Carnahan Y: Evaluation of pulmonary and systemic blood flow by 2- dimensional Doppler echocardiography using fast Fourier transform spectral analysis. Am J Cardiol 50: 1394, 1982 5. Sanders SP, Yeager S, Williams RG: Measurements of systemic and pulmonary blood flow and QP/QS ratio using Doppler and twodimensional echocardiography. Am J Cardiol 51: 952, 1983 6. Vijayaraghavan G, Singham KT, Tei C, Wong AL, Wong M, Shah PM: Aortic flow velocity in older subjects using continuous-wave Doppler echocardiography. Am Heart J 107: 1275, 1984 7. Hatle L, Angelsen B: Doppler ultrasound in cardiology. Philadelphia, 1985, Lea and Febiger, p 93 8. Risoe C, Wille SO: Blood velocity in human arteries measured by bidirectional ultrasonic Doppler flowmeter. Acta Physiol Scand 103: 370, 1978 9. Sahn D, Valdes-Cruz L, Scaguelli S, Young K, Larson D, Garcia D: Comparison of continuous wave and high PRF (multigate) 2D echo Doppler for pressure gradient estimation in animal models and human patients. Circulation 68(suppl III): 111-228, 1983 (abst) 1006