BD Simultest CD3/CD16+CD56

Similar documents
BD Simultest CD3/CD4 IVD

BD Simultest CD3/CD4 IVD

BD Simultest CD3/CD8 IVD

BD Simultest CD3/ Anti HLA-DR

BD Simultest IMK Plus

BD Tritest CD3/CD16+CD56/CD45

CD4 (SK3) IVD. For determining percentages of human helper/ inducer lymphocytes in peripheral blood. Catalog No.

BD Multitest IMK Kit

BD Oncomark CD8/CD56/CD3

CD3 (SK7) IVD. For identifying and enumerating T lymphocytes. Catalog No.

BD OneFlow LST Application Guide

BD Tritest CD3/CD8/CD45

_a qêáíéëí»= `apl`ansh`arsl`aqr

BD Simultest CD8/CD38

BD Leucocount Kit. Rx Only IVD

CD8 (SK1) Monoclonal Antibodies Detecting Human Antigens

BD Procount Progenitor Cell Enumeration Kit

BD Multitest CD3/CD8/CD45/CD4

BD FastImmune. BD FACS lysing solution and. using BD FACS Permeabilizing Solution

WHO PQDx PR June/2016, version 2.0. WHO Prequalification of Diagnostics Programme PUBLIC REPORT. Abstract

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

BD OneFlow LST. 20 tests per kit Catalog No

Technical Resources. BD Immunocytometry Systems. FastImmune Intracellular Cytokine Staining Procedures

Application Note. Productivity and Efficiency of 6-Color BD Multitest and BD Trucount Technologies for Enumeration of Lymphocyte Subsets

Flow Cytometric Analysis of T -Lym phocytes Subsets in Adult Thais

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Plasma

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Cell culture

BD Multitest CD8/CD38/CD3/HLA-DR

Human CD4+T Cell Care Manual

ab Exosome Isolation and Analysis Kit - Flow Cytometry, Cell Culture (CD63 / CD81)

Page 1 of 2. Product Information Contents: ezkine Th1 Activation 2 Whole Blood Intracellular Cytokine Kit

BD Multitest CD45RA/CD45RO/CD3/CD4. Catalog No Tests 20 µl/test

MagniSort Mouse CD4 Naive T cell Enrichment Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures.

CD45RA/CD62L/CD3/CD8. Catalog No Tests 20 µl/test

MagniSort Human CD4 Memory T cell Enrichment Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures.

MagniSort Mouse CD4 T cell Enrichment Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures.

MagniSort Human CD4 T cell Enrichment Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures.

Gladstone Institutes, University of California (UCSF), San Francisco, USA

Application Information Bulletin: Human NK Cells Phenotypic characterizing of human Natural Killer (NK) cell populations in peripheral blood

SIV p27 Antigen ELISA Catalog Number:

Rat Hemoglobin A1c (HbA1c) Kit Instructions

Human Umbilical Cord Blood CD34 + /133+ Progenitor Cell Care Manual

NEPHROCHECK Calibration Verification Kit Package Insert

7-AAD/CFSE Cell-Mediated Cytotoxicity Assay Kit

ezkine Th1/Th17 Whole Blood Intracellular Cytokine Kit Catalog Number: RUO: For Research Use Only. Not for use in diagnostic procedures.

Concentration Estimation from Flow Cytometry Exosome Data Protocol

In vitro human regulatory T cell expansion

7-AAD/CFSE Cell-Mediated Cytotoxicity Assay Kit

In vitro human regulatory T cell expansion

BD FastImmune. Catalog No Tests 20 µl/test

IV2-113E Use by. Invitron Glargine ELISA Kit REF LOT IVD. Definitions. English. For in-vitro diagnostic use. Instructions for use.

ELEGANCE Chlamydia pneumoniae IgG ELISA KIT

MALDI Sepsityper Kit

In vitro human regulatory T cell suppression assay

Annexin V APC Assay Kit

BD Simultest. Monoclonal Antibodies Detecting Human Antigens. CD8/Anti-HLA-DR

25(OH) Vitamin D ELISA (BD-220BA), 192 Tests

Technical Bulletin No. 157

Cotinine (Mouse/Rat) ELISA Kit

Morinaga Mouse C-peptide ELISA Kit

See external label 2 C-8 C 96 tests CHEMILUMINESCENCE. CMV IgG. Cat # Step (20-25 C Room temp.) Volume

BD Phosflow T Cell Activation Kit

Evaluation of a Method for Counting Absolute Numbers of Cells with a Flow Cytometer

Thyroid Stimulating Hormone (S-TSH) Thyroid Stimulating

ABIOpure TM Viral (version 2.0)

NEPHROCHECK Liquid Control Kit Package Insert

Direct ex vivo characterization of human antigen-specific CD154 + CD4 + T cells Rapid antigen-reactive T cell enrichment (Rapid ARTE)

BD Cytometric Bead Array (CBA) Mouse Th1/Th2/Th17 Cytokine Kit

CYTOMEGALOVIRUS (CMV) IgM ELISA Kit Protocol

HIV-1 ICx/CRx Kit Reagents for Immune Dissociation and Reactivity Confirmation

CD28 (L293) Monoclonal Antibodies Detecting Human Antigens

Detailed step-by-step operating procedures for NK cell and CTL degranulation assays

COMPONENT NAME COMPONENT # QUANTITY STORAGE SHELF LIFE FORMAT. Store at 2-8 C. Do not freeze. Store at 2-8 C. Do not freeze.

Mouse Leptin ELISA Kit Instructions

Mercodia Glucagon ELISA

TNF-alpha ELISA. For Research Use Only. Not For Use In Diagnostic Procedures.

Mouse C-Peptide ELISA Kit

Triiodothyronine (T3) ELISA

VGKC-Autoantibody Assay RIA

Establishing a Pure Lymphocyte Gate for Subset Analysis by Flow Cytometry

Procine sphingomyelin ELISA Kit

COMPONENT NAME COMPONENT # QUANTITY STORAGE SHELF LIFE FORMAT. Store at 2-8 C. Do not freeze. Store at 2-8 C. Do not freeze.

Cryopreserved Human Hepatocytes (Cat # HP-F)

DIAGNOSTIC AUTOMATION, INC.

Instructions for use. TSH rat ELISA. Please use only the valid version of the Instructions for Use provided with the kit AR E-8600

BD Cytometric Bead Array (CBA) Human Th1/Th2/Th17 Cytokine Kit

For research or further manufacturing use only. Not for injection or diagnostic procedures.

Influenza A IgG ELISA

BD FastImmune. Catalog No Tests 20 µl/test

Human Carbamylated LDL ELISA Kit (CBL-LDL Quantitation)

See external label 96 tests HSV 2 IgA. Cat #

Rapid antigen-specific T cell enrichment (Rapid ARTE)

IVD information *Droppers for the sensitized and control cells. Not for use other than dispensing the sensitized and control cells.

E.Z.N.A. SQ Blood DNA Kit II. Table of Contents

Human HBcAb IgM ELISA kit

Step-by-Step Instructions For OraQuick HCV Rapid Antibody Test

COMPONENT NAME COMPONENT # QUANTITY STORAGE SHELF LIFE FORMAT. Store at 2-8 C. Do not freeze. Store at 2-8 C. Do not freeze.

Bovine Insulin ELISA

Transcription:

2/2014 23-5267-02 IVD BD Simultest CD3/CD16+CD56 For determining human natural killer lymphocytes in peripheral blood 50 Tests Catalog No. 342403 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2014 BD Becton, Dickinson and Company BD Biosciences 2350 Qume Drive San Jose, CA 95131 USA Benex Limited Pottery Road, Dun Laoghaire, Co. Dublin, Ireland Tel +353.1.202.5222 Fax +353.1.202.5388 BD Biosciences European Customer Support Tel +32.2.400.98.95 Fax +32.2.401.70.94 help.biosciences@europe.bd.com Becton Dickinson Pty Ltd, 4 Research Park Drive, Macquarie University Research Park, North Ryde NSW 2113, Australia Becton Dickinson Limited, 8 Pacific Rise, Mt. Wellington, Auckland, New Zealand 1. INTENDED USE BD Simultest CD3/CD16+CD56 is a two-color direct immunofluorescence reagent for enumerating percentages of mature human natural killer (NK) lymphocytes in erythrocyte-lysed whole blood (LWB). 2. SUMMARY AND EXPLANATION Human lymphocytes can be divided into three major populations based on their biologic function and cell-surface antigen expression: T lymphocytes, B lymphocytes, and NK lymphocytes. T lymphocytes participate in antigenspecific cell-mediated immunity and regulate the secretion of immunoglobulin by B lymphocytes. NK lymphocytes mediate cytotoxicity against certain tumors and virus-infected target cells, without requiring the presence of class I or II major histocompatibility complex (MHC) molecules on the target. 1 Clinical Applications NK lymphocytes, identified as CD3 and CD16 + and/or CD56 +, 2 have been shown to mediate cytotoxicity against certain tumors and virus-infected cells. NKmediated cytotoxicity does not require class I or class II MHC molecules to be present on the target cell. 1 3. PRINCIPLES OF THE PROCEDURE When monoclonal antibody reagents are added to human whole blood, the fluorochrome-labeled antibodies bind specifically to antigens on the surface of leucocytes. Monoclonal antibodies can be used to identify lymphocyte subpopulations. bdbiosciences.com ClinicalApplications@bd.com 1

An aliquot of the stained patient sample is introduced into the flow cytometer and passed in a narrow stream through the path of a laser beam. The stained cells fluoresce when excited by the laser beam and the emitted light is collected and processed by the flow cytometer. 4. REAGENT Reagent Provided, Sufficient for 50 Tests The BD Simultest CD3/CD16+CD56 reagent, sufficient for 50 tests, is provided in 1 ml of buffered saline with gelatin and 0.1% sodium azide. It contains FITClabeled CD3, clone SK7, 3-6 for the identification of T lymphocytes and the T- lymphocyte subset that expresses the CD16 and/or CD56 antigens. It also contains PE-labeled CD16, clone B73.1 7-9, and PE-labeled CD56, clone MY31, 10,11 to ensure identification of the NK-lymphocyte population. The fluorescein-to-protein ratio (F:P) for BD IgG monoclonal antibody reagents is 2 to 5. The F:P ratio for CD3 FITC has been optimized for its intended use. The CD3 antibody is composed of mouse IgG 1 heavy chains and kappa light chains. CD3 reacts with the epsilon chain of the CD3/T-cell antigen receptor (TCR) complex. 12 This complex is composed of at least six proteins that range in molecular weight from 20 to 30 kilodaltons (kda). 13 The antigen recognized by CD3 antibodies is noncovalently associated with either α/β or γ/δ TCR (70 to 90 kda). 14 The CD16 antibody is composed of mouse IgG 1 heavy chains and kappa light chains. 8,10 The antigen recognized by CD16 antibodies is a 50 to 65-kDa protein that is the IgG Fc III receptor present on NK lymphocytes and neutrophils. 2 The CD56 15 antibody is composed of mouse IgG 1 heavy chains and kappa light chains. The CD56 antigen is present on NK lymphocytes. 2,16 The antibody recognizes the 140-kDa isoform of the neural cell adhesion molecule (NCAM). 9,17 The 140-kDa core protein is extensively glycosylated to give a mature antigen. 11,18 The CD56 antigen is present on essentially all resting and activated CD16 + NK lymphocytes and approximately 5% of CD3 + peripheral blood lymphocytes. 2 CD56 antigen density increases when NK lymphocytes are activated. 19 CD3 + CD56 + T lymphocytes compose a unique subset of cytotoxic T lymphocytes that mediates non MHC-restricted cytotoxicity. 2 The CD56 antigen is involved in neuronal homotypic cell adhesion and cell differentiation during embryogenesis. 11 Precautions For in vitro diagnostic use. When stored at 2 C 8 C, the antibody reagent is stable until the expiration date shown on the label. Do not use after the expiration date. The antibody reagent should not be frozen or exposed to direct light during storage or during incubation with cells. Keep the reagent vial dry. 2

Alteration in the appearance of the reagent, such as precipitation or discoloration, indicates instability or deterioration. In such cases, the reagent should not be used. The antibody reagent contains sodium azide as a preservative; however, care should be taken to avoid microbial contamination, which can cause erroneous results. WARNING All biological specimens and materials coming into contact with them are considered biohazards. Handle as if capable of transmitting infection 20,21 and dispose of with proper precautions in accordance with federal, state, and local regulations. Never pipette by mouth. Wear suitable protective clothing, eyewear, and gloves. 5. INSTRUMENT BD Simultest CD3/CD16+CD56 reagent is designed for use on a BD FACS brand flow cytometer equipped with appropriate computer hardware, software, and gating electronics. The flow cytometer must be equipped to detect two-color fluorescence, forward scatter (FSC), and side scatter (SSC). We recommend using BD Simulset software, version 2.5 or later, for data acquisition and analysis. All performance characteristics were obtained using a BD FACScan flow cytometer and verified on a BD FACStrak flow cytometer. Other systems can have different characteristics and should be verified by the user. 6. SPECIMEN AND COLLECTION PREPARATION Collect blood aseptically by venipuncture 22,23 into a sterile (lavender top) BD Vacutainer EDTA blood collection tube. A minimum of 1 ml of whole blood is required for this procedure. Blood should be stained within 6 hours of venipuncture for optimal results. Anticoagulated blood can be stored at room temperature (20 C 25 C) for up to 6 hours until ready for staining. Blood samples refrigerated prior to staining can give aberrant results. A white blood cell (WBC) count and a differential white cell count should be obtained from the same sample of whole blood before staining. An acceptable WBC concentration range is from 3.5 x 10 3 to 9.4 x 10 3 WBC/µL. Samples with counts greater than 9.4 x 10 3 WBC/µL must be diluted with 1X phosphatebuffered saline (PBS) containing 0.1% sodium azide. For samples with counts less than 3.5 x 10 3 WBC/µL, more blood might be needed and a separation procedure can be required to concentrate the cells. Interfering Conditions Previously fixed and stored patient specimens should not be used. Whole blood samples refrigerated prior to staining can give aberrant results. For optimal results, blood samples should be stained within 6 hours of venipuncture. Samples obtained from patients taking immunosuppressive drugs can yield poor resolution. 24 The presence of blast cells or unlysed or nucleated red blood cells (RBCs) can interfere with test results. 3

Hemolyzed samples should be rejected. Follow the collection tube manufacturer s guidelines for the minimum volume of blood to be collected. CAUTION Use standard precautions when obtaining, handling, and disposing of all human blood samples and potentially carcinogenic reagents. 7. PROCEDURE Reagent Provided See Reagent Provided, Sufficient for 50 Tests and Precautions in Section 4, Reagent Reagents and Materials Required But Not Provided BD Simultest Leucogate (CD45/ CD14) reagent, 1 ml (Catalog No. 342408). For determining a lymphocyte analysis gate, refer to the BD Simulset Leucogate instructions for use (IFU) and the BD Simulset Software User s Guide. Store at 2 C 8 C. BD Simultest Isotype Control γ 1 /γ 2a (IgG 1 FITC/IgG 2a PE) (both keyhole limpet hemocyanin specific), 1 ml (Catalog No. 342409). Store at 2 C 8 C. BD FACS lysing solution (10X) (Catalog No. 349202). For dilution instructions and warnings, refer to the IFU. BD Calibrite beads (Catalog No. 349502. For detailed information on use, please refer to the BD Calibrite Beads IFU. BD Vacutainer EDTA blood collection tubes or equivalent. Falcon * disposable 12 x 75-mm polystyrene test tubes or equivalent. Vortex mixer. Low-speed centrifuge (minimum speed 200g) with swinging-bucket rotor and 12 x 75-mm tube carriers. Vacuum aspirator with trap. Micropipettor with tips. BD CellWASH (Catalog No. 349524) or a wash buffer of PBS with 0.1% sodium azide. BD CellFIX (Catalog No. 340181) or 1% paraformaldehyde solution in PBS with 0.1% sodium azide. Store at 2 C 8 C in amber glass for up to 1 week. BD FACSFlow sheath fluid (Catalog No. 342003) or equivalent. CAUTION Use only BD FACSFlow sheath fluid diluent to dilute BD Calibrite beads. Reagent-grade (both distilled and deionized) water. Staining and Fixing the Cells Whole blood samples are first stained with BD Simultest Leucogate (Tube A), BD Simultest Isotype Control (Tube B), and BD Simultest CD3/CD16+CD56 (Tube C) reagents. Diluted (1X) BD FACS lysing solution is used to lyse RBCs following staining. Use care to protect the tubes from direct light. Perform the procedure at room temperature (20 C 25 C) using room temperature reagents. Refer to Precautions in Section 4 Reagent. 1. For each patient sample, label three 12 x 75-mm Tubes A, B, and C. Also label each tube with the sample identification number. * Falcon is a registered trademark of Corning Incorporated. 4

2. Place 20 µl of BD Simultest Leucogate reagent into Tube A, 20 µl of BD Simultest Isotype Control reagent into Tube B, and 20 µl of BD Simultest CD3/CD16+CD56 reagent into Tube C. 3. For each sample tube, use a fresh micropipettor tip and carefully add 100 µl of well-mixed, anticoagulated whole blood into the bottom of each of the three labeled tubes. The recommended WBC concentration is 3.5 x 10 3 to 9.4 x 10 3 WBC/µL. Vortex thoroughly at low speed for 3 seconds and incubate for 15 to 30 minutes at room temperature (20 C 25 C). NOTE Protect samples from direct light during this incubation procedure and use care to prevent blood from running down the side of the tube. If whole blood remains on the side of the tube, it will not be stained with the reagent. 4. Add 2 ml of room temperature (20 C 25 C) 1X BD FACS lysing solution to each tube. Immediately vortex thoroughly at low speed for 3 seconds and incubate for 10 to 12 minutes at room temperature (20 C 25 C) in the dark. Do not exceed 12 minutes. NOTE Avoid prolonged exposure of the cells to lytic reagents, which can cause white cell destruction. 5. Immediately after incubation, centrifuge tubes at 300g for 5 minutes at room temperature(20 C 25 C). 6. Aspirate the supernatant leaving approximately 50 µl of residual fluid in the tube to avoid disturbing the pellet. 7. Vortex thoroughly at low speed to resuspend the cell pellet in the residual fluid, and then add 2 ml of BD CellWASH solution or PBS with 0.1% sodium azide to each tube. Vortex thoroughly at low speed for 3 seconds. Centrifuge at 200g for 5 minutes at room temperature (20 C 25 C). 8. Aspirate the supernatant leaving approximately 50 µl of residual fluid in the tube to avoid disturbing the pellet. 9. Vortex thoroughly at low speed to resuspend the cell pellet in the residual fluid, and then add 0.5 ml of BD CellFIX solution or 1% paraformaldehyde to each tube. Vortex thoroughly at low speed for 3 seconds. Make sure the cells are well mixed with the fixing solution. 10. The cells are now ready to be analyzed on the flow cytometer. Cap or cover the prepared tubes and store at 2 C 8 C in the dark until flow cytometric analysis. Analyze the fixed cells within 24 hours after staining. Vortex the cells thoroughly (at low speed) to reduce aggregation before running them through the flow cytometer. 25 Flow Cytometry Follow the BD instructions for two-color flow cytometric analysis and refer to the BD Simulset Software User s Guide. Quality Control For optimal results, we recommend using BD Calibrite beads and BD Autocomp software for setting the photomultiplier tube (PMT) voltages, setting the fluorescence compensation and checking 5

instrument sensitivity prior to use of BD Simultest CD3/CD16+CD56 reagent on a BD FACScan or BD FACStrak flow cytometer. We recommend that a control sample from a normal adult subject be run daily to optimize instrument settings and as a quality control check of the system. Correct results for a hematologically normal patient sample are shown in Figure 1. BD Simultest Isotype Control reagent is run with each patient sample to set fluorescence 1 (FL1) and fluorescence 2 (FL2) markers between negatively and positively stained lymphocyte clusters and to detect the presence of nonspecific staining that would indicate erroneous patient results. Visually inspect the dot plot for BD Simultest Isotype Control reagent Tube B. If the negative cluster is diffuse and smeared over the FL2 intensity range, the marker might not be set correctly and results can be suspect. Figure 1 BD FACScan LWB sample, from a hematologically normal patient, stained with BD Simultest Leucogate (Tube A), BD Simultest Isotype Control (Tube B), and BD Simultest CD3/CD16+CD56 (Tube C) reagents. BD Simultest Leucogate reagent was used to reduce debris, monocytes, and granulocytes in the gate shown under Tube A. Dot plot displays of FL1 (x-axis) vs FL2 (y-axis) are shown for Tubes B and C. Plots A, B, C, display the data for tubes A, B, and C respectively. 2a IgG SSC B FSC IgG 1 CD3 BD Simulset software will automatically inspect the data and alert the operator with a number of possible error messages. Refer to the BD Simulset Software User s Guide for a list of possible messages. Use the following criteria for inspection of the dot plots obtained for each sample to evaluate the quality of the data obtained. The operator should reject the results if one or more of the following error messages is received for the normal control: no separation between cellular populations; too few lymphocytes (less than 500); excessive RBC or nucleated RBC contamination and debris (greater than 10%); or excessive monocyte (greater than 3%) or granulocyte (greater than 6%) contamination of the lymphocyte gate. A CD16+CD56 C 6

If there is no obvious reason for the normal control to fail, a sample from another normal control should be stained and run and the entire staining procedure repeated on all subsequent samples. Samples with nucleated RBCs can contain too much debris because of incomplete lysis of nucleated erythrocytes with BD FACS lysing solution. Too much debris can also occur when assaying blood samples from patients with certain hematologic disorders where red cells are difficult to lyse, as in myelofibrosis and spherocytosis. Nucleated erythrocytes will be counted as debris and, if debris exceeds 10%, the software will flag the sample as too many nonlymphs in gate and the sample results should be rejected. The tube-to-tube difference in gated events must be less than or equal to 500 events for all tubes in the panel except the BD Simultest Leucogate reagent (Tube A). (However, if the BD Simultest Isotype Control reagent tube is problematic, then the panel is suspect.) 8. RESULTS Percent Lymphocyte Conversion When the Percent Lymphocyte Conversion computation is performed, the CD3 and CD16 + and/or CD56 + lymphocyte subset is reported as a percentage of lymphocytes in the lymphocyte analysis gate. If the computation is not performed, results will be reported as a percentage of the gated events. Three-Part Differential For lysed whole blood, it is possible to estimate monocytes, lymphocytes, and granulocytes as a percentage of leucocytes using the BD Simultest Leucogate reagent (Tube A). BD Simulset software automatically calculates a three-part differential. Please refer to the BD Simulset Software User s Guide for representative data printouts. NOTE The differential provided by BD Simulset software should be used only for comparison with an independent differential white cell count for quality control purposes and should not be used in place of an independent laboratory differential white cell count in patient charts or entered into BD Simulset software to obtain absolute counts. Absolute Counts An absolute cell count can be computed if a WBC and the lymphocyte percentage from an independent differential white cell count are obtained using standard laboratory procedures. 9. LIMITATIONS Laboratories must establish their own normal reference ranges for the BD Simultest CD3/CD16+CD56 parameters that can be affected by sex of patient, age of patient, and preparative technique. Race of patient can also have an effect, although sufficient data is not available to establish this. Age, sex, clinical status, and race of subjects should be known when a reference range is determined. 7

Results from BD Simultest CD3/ CD16+CD56 reagent must be used in conjunction with other information available from the clinical evaluation and additional independent diagnostic procedures. BD Simultest CD3/CD16+CD56 reagent is not intended for screening samples for the presence of leukemic cells or for use in phenotyping samples from leukemia patients. The presence of blast cells might not allow BD Simultest Leucogate to set an adequate lymphocyte analysis gate. The software will flag the sample and results will not be printed. Absolute count values of lymphocyte subsets might not be comparable across laboratories because of variations in methods for determining white blood cell counts and/or white cell differential values. 10. EXPECTED VALUES Leucocyte Subsets The natural killer lymphocyte population present in quadrant 1 (Q1) of Tube C is shown in Figure 1 and in the BD Simulset Software User s Guide. The clinically significant results are found in the Physician Report. BD has investigated the normal reference ranges for BD Simultest CD3/ CD16+CD56 reagent parameters in 160 subjects using the BD FACScan flow cytometer at three sites (two US clinical sites and BD in San Jose, California). The expected normal reference ranges of natural killer lymphocytes for LWB are shown in Table 1. 26 Table 1 Representative reference ranges for BD Simultest CD3/CD16+CD56 reagent parameters in hematologically normal adults as percentages of total gated lymphocytes (converted) (data pooled from three clinical sites) Mean 95% Parameter Sex Age n (%) Range NK lymphs Both 18 70 160 14 5.6 31.0 Adult reference ranges should not be used with pediatric blood samples. Race can also be a variable in the establishment of normal reference ranges, 27 although insufficient data was collected by BD to determine this. NOTE Expected normal values can vary depending upon age, sex, or race of patient. Because of these differences, each laboratory should establish its own normal reference range for each parameter. 11. PERFORMANCE CHARACTERISTICS Performance of the BD Simultest CD3/ CD16+CD56 reagent was established by testing at two US clinical sites and BD laboratories in San Jose, California. Within-Sample Reproducibility Table 2 shows average within-sample reproducibility obtained for both normal and abnormal subjects. Between-Instrument Reproducibility Table 3 shows between-instrument reproducibility results. 8

Between-Laboratory Reproducibility Between-laboratory reproducibility is indicated by the ability to pool the normal reference ranges for BD Simultest CD3/ CD16+CD56 reagent parameters (Table 1). BD Simultest CD3/CD16+CD56 vs Comparative Method A summary of the results is presented in Table 4 Table 2 Within-sample reproducibility for BD Simultest CD3/CD16+CD56 reagent (six normal subjects and four abnormal subjects) as percentages of lymphocytes (converted) Subjects Parameter Mean (%) SD ab df c SD bd df c Normal NK lymphs 13.3 1.4 12 1.4 17.8 Abnormal NK lymphs 8.0 0.7 8 0.7 11.5 a. Within stain b. SD = standard deviation c. df = degrees of freedom d. Between stain Table 3 Between-instrument reproducibility for BD Simultest CD3/CD16+CD56 reagent parameters (five normal subjects and two instruments) as percentages of lymphocytes (converted) Parameter Instrument Mean (%) SD a CV b NK lymphs 1 12.9 0.46 3.58 2 13.4 0.41 3.06 a. SD = standard deviation b. CV = coefficient of variation Table 4 BD Simultest CD3/CD16+CD56 reagent vs comparative method a (BD Simultest IMK Plus reagent panel and BD Simultest IMK-Lymphocyte reagent panel) Range of Data (%) Parameter Slope Intercept r n IMK Plus IMK-Lymphocyte NK lymphs 0.96 0.37 0.96 85 2 32 1 33 a. This data was collected on a BD FACScan flow cytometer using BD Simultest IMK-Lymphocyte reagents, which included Leucogate, Control γ 1 / γ 2a, and CD3/CD16+CD56 reagents. Software employed was BD Simultest IMK-Lymphocyte software. 9

Stability of Stained Cell Preparations We recommend analyzing samples within 24 hours of staining. Cross-Reactivity The CD16 antigen is expressed on neutrophils. 2 CD56 antigen is present on approximately 5% of CD3 + peripheral blood lymphocytes. 2 Linearity-Recovery Results are expected to be linear from 3.5x10 3 to 9.4 x 10 3 WBC/µL. WARRANTY Unless otherwise indicated in any applicable BD general conditions of sale for non-us customers, the following warranty applies to the purchase of these products. THE PRODUCTS SOLD HEREUNDER ARE WARRANTED ONLY TO CONFORM TO THE QUANTITY AND CONTENTS STATED ON THE LABEL OR IN THE PRODUCT LABELING AT THE TIME OF DELIVERY TO THE CUSTOMER. BD DISCLAIMS HEREBY ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY AND FITNESS FOR ANY PARTICULAR PURPOSE AND NONINFRINGEMENT. BD S SOLE LIABILITY IS LIMITED TO EITHER REPLACEMENT OF THE PRODUCTS OR REFUND OF THE PURCHASE PRICE. BD IS NOT LIABLE FOR PROPERTY DAMAGE OR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES, INCLUDING PERSONAL INJURY, OR ECONOMIC LOSS, CAUSED BY THE PRODUCT. REFERENCES 1. Fitzgerald-Bocarlsy P, Herberman R, Hercend T, et al. A definition of natural killer cells. In: Ades E, Lopez C, eds. Natural Killer Cells and Host Defense. Basel: Karger; 1989:1. 2. Lanier L, Le A, Civin C, Loken M, Phillips J. The relationship of CD16 (Leu-11) and Leu-19 (NKH-1) antigen expression on human peripheral blood NK cells and cytotoxic T lymphocytes. J Immunol. 1986;136:4480-4486. 3. Haynes B. Summary of T cell studies performed during the second international workshop and conference on human leukocyte differentiation antigens. In: Reinherz E, Haynes B, Nadler L, Bernstein I, eds. Leukocyte Typing II: Human T Lymphocytes. New York: Springer-Verlag; 1986:3-30. 4. Ledbetter J, Evans R, Lipinski M, Cunningham- Rundles C, Good R, Herzenberg L. Evolutionary conservation of surface molecules that distinguish T lymphocyte helper/inducer and T cytotoxic/ suppressor subpopulations in mouse and man. JExp Med. 1981;153:310-323. 5. Kan E, Wang C, Wang L, Evans R. Noncovalently bonded subunits of 22 and 28 kd are rapidly internalized by T cells reacted with Anti-Leu-4 antibody. J Immunol. 1983;131:536-539. 6. Knowles R. Immunochemical analysis of the T cellspecific antigens. In: Reinherz E, Haynes B, Nadler L, Bernstein I, eds. Leukocyte Typing II: Human T Lymphocytes. New York: Springer-Verlag; 1986:259-288. 7. Perussia B, Starr S, Abraham S, Fanning V, Trinchieri G. Human natural killer cells analyzed by B73.1, a monoclonal antibody blocking Fc receptor functions. I. Characterization of the lymphocyte subset reactive with B73.1. J Immunol. 1983;130:2133-2141. 8. Perussia B, Acuto O, Terhorst C, et al. Human natural killer cells analyzed by B73.1, a monoclonal antibody blocking Fc receptor functions. II. Studies of B73.1 antibody-antigen interaction on the lymphocyte membrane. J Immunol. 1983;130:2142-2148. 9. Schmidt R. Non-lineage/natural killer section report: New and previously defined clusters. In: Knapp W, Dörken B, Gilks W, et al, eds. Leucocyte Typing IV: White Cell Differentiation Antigens. New York, NY: Oxford University Press; 1989;517-542. 10. Perussia B, Trinchieri G, Jackson A, et al. The Fc receptor for IgG on human natural killer cells: Phenotypic, functional, and comparative studies with monoclonal antibodies. J Immunol. 1984;133:180-189. 11. Schubert J, Lanier L, Schmidt R. Cluster report: CD56. In: Knapp W, Dörken B, Gilks W, et al, eds. Leucocyte Typing IV: White Cell Differentiation Antigens. New York, NY: Oxford University Press; 1989;699-702. 12. van Dongen J, Krissansen G, Wolvers-Tettero I, et al. Cytoplasmic expression of the CD3 antigen as a diagnostic marker for immature T-cell malignancies. Blood. 1988;71:603-612. 13. Brenner M, Groh V, Porcelli S, et al. Structure and distribution of the human γδ T-cell receptor. In: Knapp W, Dörken B, Gilks W, et al, eds. Leucocyte Typing IV: White Cell Differentiation Antigens. New York, NY: Oxford University Press; 1989:1049-1053. 14. Clevers H. Alarcón B, Wileman T, Terhorst C. The T cell receptor/cd3 complex: a dynamic protein ensemble. Annual Rev Immunol. 1988;6:629-662. 10

15. Edelman G. Cell adhesion molecules. Science. 1983;219:450-457. 16. Knapp W, Reiber P, Dörken B, Schmidt R, Stein H, von dem Borne AEG. Towards a better definition of leucocyte surface molecules. Immunol Today. 1989;10:253-258. 17. Lanier L, Testi R, Bindl J, Phillips J. Identity of the Leu-19 (CD56) leucocyte differentiation antigen and neural cell adhesion molecule. J Exp Med. 1989;169:2233-2238. 18. Lanier L, Chang C, Azuma M, Ruitenberg J, Hemperly J, Phillips J. Molecular and functional analysis of human natural killer cell-associated neural cell adhesion molecule (N-CAM/CD56). J Immunol. 1991;146:4421-4426. 19. Phillips J, Lanier L. Dissection of the lymphokineactivated killer phenomenon: Relative contribution of peripheral blood natural killer cells and T lymphocytes to cytolysis. J Exp Med. 1986;164:814-825. 20. Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline Third Edition. Wayne, PA: Clinical and Laboratory Standards Institute; 2005. CLSI document M29-A3. 21. Centers for Disease Control. Perspectives in disease prevention and health promotion update: universal precautions for prevention of transmission of human immunodeficiency virus, hepatitis B virus, and other bloodborne pathogens in health-care settings. MMWR. 1988;37:377-388. 22. Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard Sixth Edition. Wayne, PA: Clinical and Laboratory Standards Institute; 2007. CLSI document GP41-A6. 23. Enumeration of Immunologically Defined Cell Populations by Flow Cytometry; Approved Guideline Second Edition. Wayne, PA: Clinical and Laboratory Standards Institute; 2007. CLSI document H42-A2. 24. Giorgi J. Lymphocyte subset measurements: significance in clinical medicine. In: Rose N, Friedman H, Fahey J, eds. Manual of Clinical Laboratory Immunology. 3rd ed. Washington DC. American Society for Microbiology; 1986:236-246. 25. Jackson A, Warner N. Preparation, staining, and analysis by flow cytometry of peripheral blood leukocytes. In: Rose N, Friedman H, Fahey J, eds. Manual of Clinical Laboratory Immunology. 3rd ed. Washington DC: American Society for Microbiology; 1986:226-235. 26. Reichert T, DeBruyère M, Deneys V, et al. Lymphocyte subset reference ranges in adult Caucasians. Clin Immunol Immunopathol. 1991;60:190-208. 27. Prince H, Hirji K, Waldbeser L, Plaeger-Marshall S, Kleinman S, Lanier L. Influence of racial background on the distribution of T cell subsets and Leu-11-positive lymphocytes in healthy blood donors. Diag Immunol. 1985;3:33-37. 11