An Ultra-High-Speed Scheimpflug Camera for Evaluation of Corneal Deformation Response and Its Impact on IOP Measurement

Similar documents
Jose M. Martinez-de-la-Casa, Julian Garcia-Feijoo, Ana Fernandez-Vidal, Carmen Mendez-Hernandez, and Julian Garcia-Sanchez METHODS

INTRODUCTION S. HERDENER, D. HAFIZOVIC, M. PACHE, S. LAUTEBACH, J. FUNK. University Eye Hospital, Freiburg - Germany

True IOP No Doubt Facts and Figures Figures. Ziemer Ophthalmic Systems AG a Ziemer Group Company Allmendstrasse 11 CH-2562 Port, Switzerland

Dina H. Erickson, O.D., a Denise Goodwin, O.D., a Michael Rollins, O.D., b Amber Belaustegui, O.D., c and Chad Anderson a

Evaluation of the Effect of Diabetes Mellitus on Corneal Biomechanics

Peretyagin O.A., Cand. Med. Sc., Ass. Prof.; Dmitriev S.K., Dr. Med. Sc., Prof.; Lazar Yu.M., Cand. Med. Sc.; Tatarina Yu.A., Junior Research Fellow

Make a more confident glaucoma risk assessment with Corneal Hysteresis, a superior predictor of glaucoma progression.

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

The innovative aspects are that ORA G3 is currently the only device capable of measuring corneal hysteresis.

Central Corneal Thickness-An important variable for prognostication in Primary Open Angle glaucoma; A Kolkata based study in Eastern India

Role of Central Corneal Thickness in Circadian Intraocular Pressure Fluctuations among Patients with Primary Open Angle Glaucoma

Intraocular pressure (IOP) is of fundamental importance in

Although measurements of the optic disc and retinal nerve

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

NIIOS. Cornea Specialist, Melles Cornea Clinic, NIIOS, Rotterdam, the Netherlands Naval Hospital, Athens, Greece VASILIS S.

Comparison of Intraocular Pressure Measurements by the Ocular Response Analyzer and Goldmann Applanation Tonometer after Penetrating Keratoplasty

Glaucoma: Diagnostic Modalities

A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements

STANDARD AUTOMATED PERIMETRY IS A GENERALLY

LARGE DISCS WITH LARGE CUPS A DIAGNOSTIC CHALLENGE IN AFRICAN PATIENTS. Darshana Soma

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD

The use of an air puff to produce a deformation of the cornea. Contributing Factors to Corneal Deformation in Air Puff Measurements.

Impact of Myopia on Corneal Biomechanics in Glaucoma and Nonglaucoma Patients

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

The Evolution of Fundus Perimetry

CLINICAL SCIENCES. Comparison of Ocular Response Analyzer Parameters in Chinese Subjects With Primary Angle-Closure and Primary Open-Angle Glaucoma

Scanning laser polarimetry (SLP) provides real-time, objective

The Ocular Response Analyzer (ORA; Reichert Ophthalmic

CLINICAL SCIENCES. Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation Tonometry

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC)

INTRODUCTION MIKLOS SCHNEIDER 1, GABOR BORGULYA 2, ANDRAS SERES 1, ZOLTAN Z. NAGY 1, JANOS NEMETH 1

Analysis of differences in intraocular pressure evaluation performed with contact and non-contact devices

Deep anterior lamellar keratoplasty (DALK) is considered

Refractive corneal surgery (RCS) using a laser, such as. Glaucoma Progression in Eyes with a History of Refractive Corneal Surgery.

Policy #: 354 Latest Review Date: July 2011

Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer

Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography

Evolving glaucoma management True diagnostic integration for the preservation of vision

Comparison of IOP measurement by ocular response analyzer, dynamic contour, Goldmann applanation, and noncontact tonometry

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma.

December 2016, Volume: 8, Issue: 12, Pages: , DOI:

INTRODUCTION J. DAWCZYNSKI, E. KOENIGSDOERFFER, R. AUGSTEN, J. STROBEL. Department of Ophthalmology, University Hospital Jena, Jena - Germany

RETINAL NERVE FIBER LAYER

MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND PACHYMETRY. POLICY NUMBER: CATEGORY: Technology Assessment

Elevated intraocular pressure (IOP) is a major risk factor for

Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach

Optical coherence tomography (OCT) is a noninvasive,

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography

Intro to Glaucoma/2006

Correspondence should be addressed to Fusako Fujimura;

Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science

The Role of the RNFL in the Diagnosis of Glaucoma

1/31/2018. Course Objectives. Diagnostic Testing. Optic Nerve Damage ANATOMY AND PHYSIOLOGY OF A GLAUCOMA WORK-UP/TONOMETRY TECHNICIAN: -SDP

Glaucoma is the leading cause of irreversible blindness in

C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma

Self-monitoring of disease has long been of proven value. For

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography

Advances in OCT Murray Fingeret, OD

Fluctuation of Intraocular Pressure and Glaucoma Progression in the Early Manifest Glaucoma Trial

Correlating central corneal thickness and intraocular pressure in ocular hypertension and glaucoma

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS

Role of central corneal thickness measurement in management of open angle glaucoma and glaucoma suspects in Calabar, Nigeria

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

Science & Technologies

Correlation between presence of primary irisand cilliary body cysts and intraocular pressure

CLINICAL SCIENCES. Felipe A. Medeiros, MD; Linda M. Zangwill, PhD; Christopher Bowd, PhD; Robert N. Weinreb, MD

Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation

Effects of laser in situ keratomileusis (LASIK) on corneal biomechanical measurements with the Corvis ST tonometer

Efficacy of transepithelial corneal cross-linking using iontophoresis

Comparison of Goldmann applanation tonometry and dynamic contour tonometry in healthy and glaucomatous eyes

Intraocular pressure changes and relationship. Corneal Biomechanics After SMILE and FS-LASIK

GLAUCOMA SUMMARY BENCHMARKS FOR PREFERRED PRACTICE PATTERN GUIDELINES

CLINICAL SCIENCES. Clinical Significance of Central Corneal Thickness in the Management of Glaucoma

Correlating Structure With Function in End-Stage Glaucoma

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS

Reproducibility of Bruch Membrane Opening-Minimum Rim Width Measurements With Spectral Domain Optical Coherence Tomography

Clinical Study Evaluation of the New Digital Goldmann Applanation Tonometer for Measuring Intraocular Pressure

Corneal hysteresis in patients with glaucoma-like optic discs, ocular hypertension and glaucoma

Early detection and control of disease progression or

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS

1/25/2019 OCT & OCTA RETINAL IMAGING: HOW TO PREVENT RAGING GLAUCOMA! THE ORIGINAL RAGING GLAUCOMA OCT RETINAL IMAGING OPTIC NERVE HEAD EXAMINATION

21st Century Visual Field Testing

Targeting Intraocular Pressure in Glaucoma: a Teaching Case Report 1

Elevated intraocular pressure (IOP) is the most important. The Effect of Trabeculectomy on Ocular Pulse Amplitude. Glaucoma

3/16/2018. Perimetry

Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes

Effects of Intracorneal Ring Segment on Corneal Biomechanics in Keratoconic Eyes. Abstract

The Heritability of Corneal Hysteresis and Ocular Pulse Amplitude: A Twin Study

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography

Kobe University Repository : Kernel

Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT

Comparison between Pascal dynamic contour tonometer and Goldmann applanation tonometer after different types of refractive surgery

Transcription:

Glaucoma An Ultra-High-Speed Scheimpflug Camera for Evaluation of Corneal Deformation Response and Its Impact on IOP Measurement Christopher Kai-Shun Leung, 1 Cong Ye, 1 and Robert N. Weinreb 2 1 Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong, People s Republic of China 2 Hamilton Glaucoma Center and Department of Ophthalmology, University of California, San Diego, California Correspondence: Christopher Kai- Shun Leung, Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong Eye Hospital, 147K Argyle Street, Kowloon, Hong Kong; tlims00@hotmail.com. Submitted: December 26, 2012 Accepted: March 1, 2013 Citation: Leung CK-S, Ye C, Weinreb RN. An ultra-high-speed Scheimpflug camera for evaluation of corneal deformation response and its impact on IOP measurement. Invest Ophthalmol Vis Sci. 2013;54:2885 2892. DOI:10.1167/iovs.12-11563 PURPOSE. To investigate the test retest variability of and factors associated with the corneal deformation response, and evaluate its impact on IOP measurement. METHODS. Forty normal and 140 glaucoma suspect/glaucomatous eyes of 104 individuals were prospectively included for repeated measurements of corneal deformation response by an ultra-high-speed Scheimpflug camera, followed by corneal hysteresis and corneal resistance factor measurements by the ocular response analyzer, and dynamic contour tonometry (DCT) (the reference standard) and Goldmann applanation tonometry (GAT) during the same visit. The test retest variability of corneal deformation response was evaluated. Univariate and multivariate linear mixed models were used to identify factors (age, manifest spherical refractive error, axial length, IOP, central corneal thickness [CCT], corneal curvature, corneal hysteresis, and corneal resistance factor) associated with corneal deformation response and the difference between DCT and GAT measurements. RESULTS. Corneal deformation amplitude (CDA) had an intraclass correlation coefficient of 0.86. There was no difference in CDA between the glaucoma and nonglaucoma groups. A higher IOP, a younger age, and a greater CCT were associated with a smaller CDA (P 0.002) (n ¼ 180). In the univariate analysis (n ¼ 180), the difference between DCT and GAT measurements was associated with spherical refractive error (P ¼ 0.037), CCT (P ¼ 0.004), and CDA (P < 0.001) after adjusting the effect of IOP (DCT). In the multivariate analysis, the only factor associated with the IOP difference was CDA (P ¼ 0.003). CONCLUSIONS. CDA was a reliable indicator to quantify corneal deformation response. CDA, rather than CCT, was the key source of measurement error of GAT. Keywords: ultra-high-speed Scheimpflug camera, intraocular pressure, corneal deformation response Goldmann applanation tonometry (GAT) is the reference standard for measurement of IOP in clinical practice and clinical trials. 1 Although GAT overestimates IOP in healthy eyes with thick corneas and underestimates IOP in healthy eyes with thin corneas, the corneal biomechanical response has been suggested to be more influential than central corneal thickness (CCT) in measuring IOP. 2 In a mathematical simulation model, Liu and Roberts 2 examined the association between IOP and corneal parameters, including CCT, corneal curvature, and corneal elasticity. They showed that normal variation of CCT (443 629 lm) contributed to a difference of only 2.87 mm Hg of predicted GAT IOP measurements. By contrast, variation of corneal elasticity, measured in Young s modulus with an estimated normal range between 0.1 and 0.9 MPa, accounted for a difference of 17.26 mm Hg. This is supported by an experimental study by Dupps and colleagues 3 using sonic wave propagation velocity on the corneal surface as a measure of corneal stiffness. They showed that corneal stiffening after cross-linking in two human globes resulted in a dramatic increase in IOP measured by pneumatonometer (Mentor O&O, Inc., Norwell, MA) and TonoPen (Medtronic Solan, Jacksonville, FL) despite a relatively low intravitreal pressure. These studies indicated that the biomechanical properties of the cornea are more important than CCT in determining GAT measurement. Clinical investigation of the stress and strain response of the cornea has been obfuscated by the lack of instrumentation to capture the dynamic behavior of the cornea. Although the ocular response analyzer (ORA) (Reichert, Inc., Depew, NY) provides a measure of corneal hysteresis and corneal resistance factor, 4 direct measurement of the corneal dynamics, such as the displacement of the cornea (strain) in response to an external pressure (load) has not been attainable in vivo. The recent introduction of an ultra-high-speed Scheimpflug imaging device (Corvis ST; Oculus, Arlington, WA) has allowed visualization and measurement of corneal deformation response to a standardized air-puff pressure. The instrument captures the corneal deformation response at a rate of 4330 image frames per second. With frame-by-frame analysis of the corneal images, parameters, including corneal deformation amplitude, corneal applanation length, and corneal velocity, can be quantified and analyzed, providing insights into the biomechanical properties of the cornea. Using the ultra-highspeed Scheimpflug camera, we evaluated the test retest repeatability of the parameters describing the corneal defor- Copyright 2013 The Association for Research in Vision and Ophthalmology, Inc. www.iovs.org j ISSN: 1552-5783 2885

Corneal Deformation Response and IOP Measurement IOVS j April 2013 j Vol. 54 j No. 4 j 2886 mation response, identified factors determining the corneal deformation response, and investigated the impact of corneal deformation response on GAT IOP measurement. METHODS Subjects One hundred eighty eyes, including 40 normal, 39 glaucoma suspect, and 101 glaucomatous eyes of 104 individuals were prospectively examined at the University Eye Center, the Chinese University of Hong Kong. They underwent a complete ophthalmic assessment, including visual acuity, manifest autorefraction, axial length, corneal curvature, CCT measurement, and fundus examination. All subjects had visual acuity of at least 20/40, manifest spherical error within the range between þ4.00 and 8.0 diopters, and with no clinical evidence of ocular disease other than mild cataract and/or glaucoma. Subjects with corneal refractive surgery were excluded. Glaucoma patients had narrowing of the neuroretinal rim and/or thinning of the retinal nerve fiber layer (RNFL) evident in clinical examination with corresponding visual field defects (described below). Glaucoma suspects had either a history of IOP of 22 mm Hg or higher during three separate visits or evidence of narrowing of the neuroretinal rim and/or thinning of the RNFL but without visual field defects. Healthy subjects had no record of GAT IOP of 22 mm Hg or higher, and no evidence of glaucomatous optic disc or visual field changes. All participants had repeated corneal deformation response measurements by Corvis ST (Oculus), and corneal hysteresis and corneal resistance factor measurements by ORA (Reichert, Inc.). This was followed by dynamic contour tonometry (DCT, Pascal; Swiss Microtechnology AG, Port, Switzerland) and GAT during the same visit. The study was conducted in accordance with the ethical standards stated in the 1964 Declaration of Helsinki with approval obtained from the local ethnics committee. Informed consent was obtained from all participants. Measurement of Corneal Deformation Response The Corvis ST (Oculus) is an ultra-high-speed Scheimpflug camera imaging the corneal deformation response to a symmetrically metered air puff (size: 3.06 mm; pressure: 60 mm Hg) at a rate of 4330 image frames per second covering a horizontal distance of 8 mm. A total of approximately 140 images of the cornea on one two-dimensional cross-section were collected over a constant metered collimated air puff for 30 ms. By tracing the anterior and posterior corneal boundaries at individual image frames, parameters describing the corneal deformation response were automatically measured by the built-in software. These include (1) corneal deformation amplitude; (2) inward and (3) outward corneal applanation lengths; and (4) inward and (5) outward corneal velocities at the time of applanation (the study was conducted at a time when the Corvis ST was not capable of analyzing corneal curvature at the maximum deformation and peak distance ). Although the Corvis ST also measured the IOP and CCT, we used only DCT and GAT for measurement of IOP, and ultrasound pachymetry for measurement of CCT. Corneal deformation amplitude was the maximum displacement of the corneal apex in response to the air puff (Fig. 1C). The inward and outward applanation lengths represented the distance of the cornea flattened by the air puff at the time of inward and outward applanation, respectively (Figs. 1B, 1D). The Corvis ST recorded the movement of the cornea at high temporal resolution and the velocity was calculated by tracking the corneal displacement with time. All participants had two consecutive measurements for each eye during the same visit. Measurement of IOP The systolic IOP was recorded for both GAT and DCT. The systolic diastolic IOP variation can be estimated from the GAT by observing the relative movement of the upper and lower semicircular rings during the pulsation. The systolic IOP was obtained when the force of applanation was adjusted so that the rings did not cross each other and the diastolic pulsation rested on the touching of the inner rings. The working principle of DCT has been described. 5 In brief, the matching of the contour of the tonometer and the contour of the cornea allows the IOP to be measured by a pressure sensor located at the tip of the tonometer. In this study, the recording duration of each measurement was approximately 5 seconds. Measurement with a quality score higher than 3 (scale 1 5) was repeated until two measurements of quality score of 3 or lower (as recommended by the manufacturer) were obtained in each eye. Likewise, two consecutive IOP measurements were obtained with GAT. For both DCT and GAT, if the difference between the first and second readings were greater than 2 mm Hg, a third measurement would be obtained, and the average of two measurements was calculated. Measurement of Corneal Hysteresis and Corneal Resistance Factor Each ORA measurement generates a signal profile with 2 distinct peaks corresponding to the first (P1) and second (P2) applanation pressures. A quality score was provided by the instrument with reference to the signal profile. Only measurements with a quality score (waveform score) greater than 4 were included. Four parameters were derived from each ORA signal profile: (1) corneal compensated IOP, (2) Goldmanncorrelated IOP, (3) corneal hysteresis, and (4) corneal resistance factor. Corneal hysteresis is an indicator of viscous dampening effect of the cornea and corneal resistance factor is an indicator of the overall resistance of the cornea. These parameters are derived from the difference between the 2 applanation pressures using the formula P1 kp2 þ c, where P1 and P2 are the first and second applanation pressures, respectively, and k and c are constants. 6 Only corneal hysteresis and corneal resistance factor were analyzed in the study. Visual Field Testing The diagnostic groups were defined with reference to visual field results. Standard visual field testing was performed using static automated white-on-white threshold perimetry (SITA Standard 24-2, Humphrey Field Analyzer II; Carl Zeiss Meditec, Dublin, CA). A visual field was defined as reliable when fixation losses, and false-positive and false-negative rates were less than 20%. Only reliable tests were included in the analysis. A visual field defect was defined as having three or more significant (P < 0.05) nonedge contiguous points with at least one at the P less than 0.01 on the same side of horizontal meridian in the pattern deviation plot. Further, it had to be classified as outside normal limits in the glaucoma hemifield test and confirmed with at least two consecutive visual field tests. Statistics Statistical analyses were performed using Stata 12.1 (StataCorp, College Station, TX). test retest variability was quantified with

Corneal Deformation Response and IOP Measurement IOVS j April 2013 j Vol. 54 j No. 4 j 2887 FIGURE 1. Serial images of the cornea captured by an ultra-high-speed Scheimpflug camera at 0.000 ms (A), 7.623 ms (B), 18.018 ms (C), 22.407 ms (D), and 32.109 ms (E) after a symmetrically metered air puff (size: 3.06 mm; pressure: 60 mm Hg) was ejected. The inward and outward corneal applanations are indicated in (B) and (D).

Corneal Deformation Response and IOP Measurement IOVS j April 2013 j Vol. 54 j No. 4 j 2888 TABLE 1. Age, Refractive Errors, IOP, Corneal Dimensions, and Corneal Deformation Response Measurements All Subjects, n ¼ 180 Normal group, n ¼ 40 Glaucoma group*, n ¼ 140 Mean 6 SD Range Mean 6 SD Range Mean 6 SD Range P Age, y 56.6 6 14.0 17 to 81 56.5 6 11.0 29 to 68 56.7 6 14.8 17 to 81 0.986 Spherical error, diopters 1.2 6 2.8 7.8 to 3.3 0.5 6 2.4 7.3 to 3.3 1.4 6 2.8 7.8 to 3.3 0.249 Axial length, mm 24.2 6 1.6 20.6 to 27.8 23.9 6 1.5 21.4 to 27.8 24.3 6 1.6 20.6 to 27.4 0.844 DCT IOP, mm Hg 21.8 6 4.6 12.1 to 38.6 19.7 6 2.9 14.3 to 25.4 22.4 6 4.8 12.1 to 38.6 0.004 GAT IOP, mm Hg 16.0 6 3.9 9 to 29 14.1 6 2.7 9 to 21 16.6 6 4.0 9 to 29 0.004 CCT, lm 548 6 38 425 to 637 541 6 35 473 to 603 550 6 39 425 to 637 0.759 Corneal curvature, mm 7.7 6 0.3 7.0 to 8.8 7.7 6 0.2 7.1 to 8.0 7.7 6 0.3 7.0 to 8.8 0.944 Corneal hysteresis, mm Hg 9.9 6 1.6 5.6 to 15.6 10.2 6 1.5 6.3 to 12.7 9.8 6 1.7 5.6 to 15.6 0.481 Corneal resistance factor, mm Hg 10.0 6 1.6 5.8 to 14.4 9.9 6 1.3 7.0 to 12.7 10.1 6 1.7 5.8 to 14.4 0.537 Corneal deformation amplitude, mm 1.09 6 0.12 0.79 to 1.80 1.13 6 0.08 0.95 to 1.33 1.08 6 0.13 0.79 to 1.80 0.072 Inward corneal applanation length, mm 1.75 6 0.19 1.28 to 2.31 1.76 6 0.20 1.30 to 2.21 1.75 6 0.19 1.28 to 2.31 0.607 Outward corneal applanation length, mm 1.88 6 0.30 0.96 to 2.67 1.81 6 0.25 1.19 to 2.34 1.90 6 0.30 0.96 to 2.67 0.061 Inward corneal velocity, m/s 0.12 6 0.02 0.05 to 0.18 0.13 6 0.02 0.08 to 0.16 0.12 6 0.02 0.05 to 0.18 0.174 Outward corneal velocity, m/s 0.33 6 0.08 0.71 to 0.12 0.34 6 0.07 0.58 to 0.2 0.33 6 0.08 0.71 to 0.12 0.546 * The glaucoma group included 39 glaucoma suspect and 101 glaucomatous eyes. repeatability coefficient (RC), coefficient of variation (CV), and intraclass correlation coefficient (ICC). RC, defined as 2.77 3 within subject standard deviation (Sw), is an estimated average of measurement variability in a population. 7 The Sw is the square root of the within-subject mean square of error (the unbiased estimator of the component of variance due to random error) in a one-way random effects model. The sample size required for a repeatability study depends on the estimated confidence interval of Sw (1.96/=[2n(m 1)]), where n is the number of subjects and m is the number of observations. 8 Depending on the stringency for accuracy, the confidence interval is usually set as 10% to 20% either side of Sw. With a sample size of 180 eyes with two observations in each eye, the confidence interval was estimated at 12.7%. Coefficient of variation (CVw) was defined as 100 3 Sw/overall mean. ICC was interpreted as follows: less than 0.75 represents poor to moderate reliability; 0.75 to 0.90 represents good reliability; greater than 0.90 represents excellent reliability for clinical measures. 9 Univariate and multivariate linear mixed models were used to investigate the associations between corneal deformation amplitude (the average of two measurements during the same visit was analyzed) and each of the following: age, spherical error, axial length, IOP (DCT), CCT, corneal curvature, corneal hysteresis, corneal resistance factor, use of topical IOP-lowering medication, use of topical prostaglandin analogue, and diagnosis (normal, glaucoma suspect, glaucoma). Likewise, factors associated with the difference in DCT and GAT measurements were examined with univariate and multivariate linear mixed models. P less than 0.05 was considered statistically significant. RESULTS The mean (SD) and the range of age, spherical error, axial length, DCT IOP, GAT IOP, CCT, corneal curvature, corneal hysteresis, corneal resistance factor, and corneal deformation response measurements are shown in Table 1. Except for IOP measurements (P ¼ 0.004), there were no significant differences between the normal and glaucoma suspect/glaucoma groups (P 0.061, linear mixed modeling with adjustment of correlation between fellow eyes). Test Retest Variability of Corneal Deformation Measurements The test retest variability of corneal deformation amplitude was low and the RC, CV, and ICC were 0.13 mm (95% confidence interval [CI]: 0.12 0.14 mm), 4.29% (95% CI: 3.85% 4.73%), and 0.86 (95% CI: 0.82 0.89), respectively (Table 2). However, the inward and outward corneal applanation lengths and corneal velocities had high test retest variabilities. The values of ICC were less than 0.60 and the CVs were greater than 10%. For this reason, in the analyses of determinants of corneal deformation response and its impact on IOP measurement, we included only corneal deformation amplitude. Determinants of Corneal Deformation Amplitude In the univariate analysis, corneal deformation amplitude was significantly associated with IOP (DCT), CCT, and age (all with P 0.006) (Table 3). Spherical error, axial length, corneal curvature, corneal hysteresis, corneal resistance factor, use of topical IOP-lowering medication(s), and diagnosis (normal, glaucoma suspect, glaucoma) were not associated with corneal deformation amplitude (P 0.116). Multivariate analysis revealed that a higher IOP, a younger age, and a greater CCT were associated with a smaller corneal deformation amplitude (all with P 0.002). Figure 2 illustrates the dynamic corneal response in an eye with a high IOP and an eye with a low IOP. The movie captures are shown in Supplementary Movies S1 and S2. Impact of Corneal Deformation Amplitude on IOP Measurement DCT measures IOP largely independent of CCT and corneal biomechanical properties and closely approximates the true IOP. 10 13 Using DCT IOP as a reference standard, the difference between DCT and GAT measurements increased with increasing IOP (P < 0.001) (Fig. 3). After adjusting the effect of IOP (DCT), the IOP difference (DCT GAT) was significantly associated with spherical error (P ¼ 0.037, R 2 ¼ 0.295), CCT (P ¼ 0.004, R 2 ¼ 0.309), and corneal deformation amplitude (P <

Corneal Deformation Response and IOP Measurement IOVS j April 2013 j Vol. 54 j No. 4 j 2889 TABLE 2. RC, CV, and ICC of Corneal Deformation Response Parameters RC CV (%) ICC Corneal deformation amplitude, mm 0.13 (0.12 0.14) 4.29 (3.85 4.73) 0.86 (0.82 0.89) Inward corneal applanation length, mm 0.74 (0.67 0.82) 15.36 (13.77 16.95) 0.01 ( 0.14 0.15) Outward corneal applanation length, mm 1.21 (1.08 1.33) 23.19 (20.79 25.58) 0.05 ( 0.19 0.10) Inward corneal velocity, m/s 0.08 (0.07 0.09) 22.84 (20.48 25.20) 0.17 (0.03 0.31) Outward corneal velocity, m/s 0.16 (0.14 0.17) 17.11 ( 15.35 to 18.88) 0.59 (0.49 0.68) 0.001, R 2 ¼ 0.327) in the univariate analyses (Table 4). For each 100-lm difference in CCT (ranged between 427 lm and 637 lm in this study, Table 1), DCT and GAT differed by 1.3 mm Hg. By contrast, for each 1-mm difference in corneal deformation amplitude (ranged between 0.790 mm and 1.795 mm in this study, Table 1), the IOP measurements differed by 6.1 mm Hg. Corneal hysteresis and corneal resistance factor had no association with the IOP difference (P 0.095). In the multivariate analysis, the only significant factor associated with the IOP difference was corneal deformation amplitude (P ¼ 0.003). DISCUSSION To our knowledge, this is the first study investigating the corneal deformation response obtained with the ultra-highspeed Scheimpflug camera. With direct visualization of the corneal deformation response, we showed that the corneal deformation amplitude had a low test retest variability. A higher IOP, a greater CCT, and a younger age were independently associated with a smaller corneal deformation amplitude. More important, in support with the simulation model proposed by Liu and Roberts, 2 this study provides clinical data demonstrating that corneal biomechanical response is more influential than CCT in GAT measurement. Measuring corneal deformation amplitude can thus provide insights into the biomechanical property of the cornea and its impact on IOP measurement. The Corvis ST ultra-high-speed Scheimpflug camera is a new technology to evaluate the dynamic corneal biomechanical response. Investigating its measurement variability is of fundamental importance. It is notable that although corneal deformation amplitude, corneal applanation length, and corneal velocity measurements were all automatically derived from the same set of images for each eye by the instrument, their measurement repeatabilities varied considerably. The corneal deformation amplitude had the highest repeatability (ICC ¼ 0.86), which was followed by outward corneal velocity (ICC ¼ 0.59). Inward corneal velocity and inward/outward corneal applanation lengths had poor repeatabilities (ICC ranged between 0.05 and 0.17). Such disparity can be explained by the differences in clarity and definitiveness of the landmarks used in the measurements. The well-defined corneal apex is used for measurement of the corneal deformation amplitude (Fig. 1C). However, determining the ends of the flattest portion of the cornea for measurement of corneal applanation length could be less consistent (Figs. 1B, 1D). Microsaccadic eye movement may add variations to the measurement of corneal velocity analyzed in a relatively short time frame (30 ms). We showed that corneal deformation amplitude was negatively associated with IOP (Table 3). A higher IOP renders the cornea less likely to deform in response to stress (Fig. 2). Of note, the association between corneal biomechanical response and IOP has been previously reported in porcine and human eyes ex vivo. Examining 17 porcine eyes under inflation conditions, Bao and colleagues 14 measured the corneal deformation with a laser displacement sensor, estimated Young s modulus of the cornea, and found a linear relationship between Young s modulus and IOP. Elsheikh and colleagues 15 measured the corneal stiffness (expressed in Young s modulus) in 37 human donor corneas using inflation tests and demonstrated that there was a positive linear relationship between Young s modulus and IOP. These ex vivo studies signify that the cornea is less likely to deform when the IOP is high. The implication is that corneal deformation amplitude should always be interpreted with reference to the level of IOP. After adjusting the effect of IOP in the multivariate model, CCT and age were also associated with corneal deformation amplitude (Table 3). Although it is conceivable that the cornea would deform less in eyes with a greater CCT, the positive association between age and corneal deformation amplitude is unexpected. Previous ex vivo studies suggest that TABLE 3. Factors Associated With Corneal Deformation Amplitude Analyzed With Linear Mixed Models Parameter Coefficient 95% Confidence Interval P Univariate analysis Age, y 0.0020 0.0006 to 0.0035 0.006 Spherical error, diopter 0.0027 0.0046 to 0.0099 0.470 Axial length, mm 0.0068 0.0205 to 0.0070 0.335 IOP (DCT), mm Hg 0.0161 0.0193 to 0.0129 <0.001 CCT, lm 0.0010 0.0015 to 0.0005 <0.001 Corneal curvature, mm 0.0249 0.1063 to 0.0564 0.548 Corneal hysteresis, mm Hg 0.0032 0.0079 to 0.0143 0.570 Corneal resistance factor, mm Hg 0.0084 0.0188 to 0.0021 0.116 Use of topical IOP-lowering medication(s) 0.0090 0.0310 to 0.0489 0.661 Use of topical prostaglandin analogue 0.0238 0.0159 to 0.0635 0.240 Diagnosis* 0.0008 0.0252 to 0.0236 0.949 Multivariate analysis Age, y 0.0018 0.0007 to 0.0028 0.001 IOP (DCT), mm Hg 0.0156 0.0186 to 0.0126 <0.001 CCT, lm 0.0006 0.0010 to 0.0002 0.002 * Diagnosis (0, normal; 1, glaucoma suspect; 2, glaucoma).

Corneal Deformation Response and IOP Measurement IOVS j April 2013 j Vol. 54 j No. 4 j 2890 FIGURE 2. Images captured by an ultra-high-speed Scheimpflug camera at the time when the corneal deformation was maximum. An eye with a low IOP (10.5 mm Hg) (A) had a greater corneal deformation amplitude (1.70 mm vs. 0.99 mm) than an eye with a high IOP (30.0 mm Hg) (B). Despite the fact that a younger age and a greater CCT were associated with a smaller corneal deformation amplitude (Table 3), the IOP was more influential in determining the corneal deformation amplitude, as shown in the example. corneal stiffness increases with age. 16 18 Notably, these studies examined corneoscleral buttons only, and the measurement of pressure deformation response may not reflect the actual corneal biomechanical response in an intact globe. Although there was no significant difference in age between the glaucoma and normal groups (Table 1), glaucomatous eyes may exhibit different corneal biomechanical properties compared with normal eyes, confounding the analysis between corneal deformation amplitude and age. More investigation is needed to validate the present finding. CCT impacts Goldmann applanation tonometry. Although a number of correction formulas have been proposed to adjust FIGURE 3. A scatter plot between IOP (DCT) and the difference between DCT and GAT measurements.

Corneal Deformation Response and IOP Measurement IOVS j April 2013 j Vol. 54 j No. 4 j 2891 TABLE 4. Factors Associated With the Difference Between DCT and GAT Measurements Analyzed With Linear Mixed Models After Adjusting the Effect of IOP (DCT) Univariate analysis (adjusted for IOP) Multivariate analysis (adjusted for IOP) Parameter Coefficient 95% Confidence Interval P Age, y 0.0074 0.0159 to 0.0307 0.534 Spherical error, diopters 0.1224 0.0071 to 0.2377 0.037 Axial length, mm 0.2057 0.4115 to 0.0001 0.050 CCT, lm 0.0125 0.0209 to 0.0040 0.004 Corneal curvature, mm 0.2188 1.0771 to 1.5148 0.741 Corneal hysteresis, mm Hg 0.0549 0.2479 to 0.1381 0.577 Corneal resistance factor, mm Hg 0.1653 0.3591 to 0.0286 0.095 Corneal deformation amplitude, mm 6.0842 2.8807 to 9.2877 <0.001 Diagnosis* 0.1864 0.5806 to 0.2077 0.354 Spherical error, diopters 0.0591 0.1048 to 0.2230 0.480 Axial length, mm 0.1072 0.3995 to 0.1852 0.473 CCT, lm 0.0083 0.0169 to 0.0002 0.055 Corneal deformation amplitude, mm 5.0140 1.7496 to 8.2784 0.003 * Diagnosis (0, normal; 1, glaucoma suspect; 2, glaucoma). GAT IOP readings with reference to CCT, such adjustment is not recommended in clinical practice. 1 In fact, adjusting GAT IOP using CCT-based correction formulae results in poorer agreement with DCT IOP compared with unadjusted GAT IOP. 19 CCT-based correction formulas are not applicable to individual corneas because factors other than CCT can influence IOP measurement. In this study, in additional to CCT, corneal deformation amplitude was a significant factor accounting for the difference between GAT and DCT IOP measurements. In fact, corneal deformation response is more instrumental than CCT in determining the IOP measurement obtained with GAT. This is supported by the finding that although CCT was associated with the difference between DCT and GAT measurements, it became insignificant in the multivariate model when corneal deformation amplitude was also included (Table 4). The relative impact of corneal deformation amplitude and CCT on GAT IOP measurement can also be inferred from the respective univariate analyses (Table 4). With a range of CCT between 425 and 637 lm (Table 1), the estimated variation in IOP is 2.65 mm Hg (0.0125 3 [637 425]). By contrast, with a range of corneal deformation amplitude between 0.790 and 1.795 mm in the same group of individuals, the variation in IOP is greater: 6.11 mm Hg (6.0842 3 [1.795 0.790]). This observation supports the mathematical simulation study by Liu and Roberts 2 demonstrating that corneal biomechanical property is more influential than CCT in GAT. Although clinical measurement of corneal elasticity is not yet feasible, corneal deformation amplitude can serve as a useful indicator of corneal biomechanical response. In fact, it is a direct measurement of strain, in response to a fixed load. For a given level of IOP, a smaller corneal deformation response may signify a stiffer cornea. Measurement of the corneal deformation amplitude with the ultra-high-speed Scheimpflug camera can contribute to a better understanding of the corneal biomechanical properties. Incorporation of individualized corneal deformation amplitude in GAT may provide a more accurate estimation of IOP. Corneal deformation amplitude measured by the ultra-highspeed Scheimpflug camera is different from corneal hysteresis measured by the ORA, although both instruments measure the corneal response to a fixed load of air puff. The former directly measures the strain response (displacement, expressed in mm), whereas the latter (expressed in mm Hg) represents the difference between the two peak applanation pressures. Hysteresis is defined by the area bounded by the load-unload displacement curve. Because the ORA does not measure corneal displacement, the biomechanical implication of corneal hysteresis reported by the ORA remains ambiguous. In this study, corneal hysteresis and corneal resistance factor were not associated with corneal deformation amplitude, nor accounted for the IOP difference between GAT and DCT (Tables 3, 4). There are limitations in the measurement of corneal deformation response. Similar to the ORA, the load-unload (pressure) displacement relationship cannot be determined by the Corvis ST. The Corvis ST measures only corneal displacement, but not the corresponding load during the deformation. In other words, direct measurement of corneal elasticity and corneal hysteresis would not be possible with the Corvis ST. As corneal deformation response was significantly related to IOP (Table 3), we investigated only intravisit repeatability, but not intervisit reproducibility. IOP readings likely varied from visit to visit, rendering determination of the variability of corneal deformation responses impractical. To investigate the impact of corneal deformation amplitude on IOP measurement, we maximized the range of IOP in the analysis by including both healthy subjects and glaucoma patients. Notably, there was no significant difference in corneal deformation amplitude between the normal and glaucoma groups even after adjustment of IOP, age, and CCT (P ¼ 0.425) (data not shown). The role of measurement of corneal deformation response in risk assessment of the development and progression of glaucoma requires further investigation. Last, it is worth noting that high-quality standards were imposed on data collection, including IOP, ORA, and Corvis ST measurements. Our finding may not be generalized to other populations. To summarize, the ultra-high-speed Scheimpflug camera provides a novel approach to quantify the corneal deformation response. The corneal deformation amplitude had a low test retest variability, associated with age, IOP, and CCT, and was more influential than CCT in GAT measurement. This new measure would be pertinent to evaluating GAT IOP measurement and investigating the biomechanical properties of the cornea. Acknowledgments Disclosure: C.K.-S. Leung, Carl Zeiss Meditec (F, R); Y. Cong, None; R.N. Weinreb, Carl Zeiss Meditec (F, C) References 1. Weinreb RN, Brandt JD, Garway-Heath D, eds. Intraocular Pressure. Consensus Series 4. The Hague, The Netherlands: Kugler Publications; 2007:17 58.

Corneal Deformation Response and IOP Measurement IOVS j April 2013 j Vol. 54 j No. 4 j 2892 2. Liu J, Roberts CJ. Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis. J Cataract Refract Surg. 2005;31:146 155. 3. Dupps WJ Jr, Netto MV, Herekar S, et al. Surface wave elastometry of the cornea in porcine and human donor eyes. J Refract Surg. 2007;23:66 75. 4. Terai N, Raiskup F, Haustein M, et al. Identification of biomechanical properties of the cornea: the ocular response analyzer. Curr Eye Res. 2012;37:553 562. 5. Kaufmann C, Bachmann LM, Thiel MA. Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis. Invest Ophthalmol Vis Sci. 2003;44:3790 3794. 6. Kotecha A, Elsheikh A, Roberts CR, et al. Corneal thicknessand age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci. 2006;47:5337 5347. 7. British Standards Institution. Precision of Test Methods 1: Guide for the Determination and Reproducibility for a Standard Test Method (BS 597, Part 1). London: BSI; 1975. 8. Bland JM. How can I decide the sample size for a repeatability study? Available at: http://www-users.york.ac.uk/~mb55/ meas/sizerep.htm. Accessed October 28, 2012. 9. Portney LG, Watkins MP. Foundations of Clinical Research: Applications to Practice. Norwalk, CT: Appleton and Lange, 1993:505 528. 10. Kniestedt C, Nee M, Stamper RL. Dynamic contour tonometry: a comparative study on human cadaver eyes. Arch Ophthalmol. 2004;122:1287 1293. 11. Kniestedt C, Nee M, Stamper RL. Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol. 2005;243:359 366. 12. Kanngiesser HE, Kniestedt C, Robert YC. Dynamic contour tonometry: presentation of a new tonometer. J Glaucoma. 2005;14:344 350. 13. Boehm AG, Weber A, Pillunat LE, et al. Dynamic contour tonometry in comparison to intracameral IOP measurements. Invest Ophthalmol Vis Sci. 2008;49:2472 2477. 14. Bao F, Jiang L, Wang X, et al. Assessment of the ex vivo biomechanical properties of porcine cornea with inflation test for corneal xenotransplantation. J Med Eng Technol. 2012;36: 17 21. 15. Elsheikh A, Alhasso D, Rama P. Biomechanical properties of human and porcine corneas. Exp Eye Res. 2008;86:783 790. 16. Knox Cartwright NE, Tyrer JR, Marshall J. Age-related differences in the elasticity of the human cornea. Invest Ophthalmol Vis Sci. 2011;52:4324 4329. 17. Elsheikh A, Geraghty B, Rama P, et al. Characterization of agerelated variation in corneal biomechanical properties. J R Soc Interface. 2010;7:1475 1485. 18. Elsheikh A, Wang D, Pye D. Determination of the modulus of elasticity of the human cornea. J Refract Surg. 2007;23:808 818. 19. Park SJ, Ang GS, Nicholas S, et al. The effect of thin, thick, and normal corneas on Goldmann intraocular pressure measurements and correction formulae in individual eyes. Ophthalmology. 2012;119:443 449.