Linking structure and function in glaucoma

Similar documents
The Role of the RNFL in the Diagnosis of Glaucoma

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx

OCT in the Diagnosis and Follow-up of Glaucoma

Structural examina.on: Imaging

Glaucoma: Diagnostic Modalities

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

CLINICAL SCIENCES. Felipe A. Medeiros, MD; Linda M. Zangwill, PhD; Christopher Bowd, PhD; Robert N. Weinreb, MD

C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

The Measure of Confidence

Corporate Medical Policy

3/16/2018. Perimetry

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC)

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography

Advances in OCT Murray Fingeret, OD

OtticaFisiopatologica

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma.

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

Evolving glaucoma management True diagnostic integration for the preservation of vision

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD

Summary HTA HTA-Report Summary Validity and cost-effectiveness of methods for screening of primary open angle glau- coma

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements

PRIMUS 200 from ZEISS The essential OCT

PRIMUS 200 from ZEISS The essential OCT

21st Century Visual Field Testing

Comparative study of new imaging technologies for the diagnosis of glaucoma: Protocol Approved by the Ethics Committee

Il contributo dell'angio-oct: valutazione integrata della componente nervosa e vascolare della malattia glaucomatosa

Cirrus TM HD-OCT. Details defi ne your decisions

Early Detection Of Glaucoma Clinical Clues. Points To Live By. Glaucoma Risk Factors 10/3/2014

Structure WGA. Structure WGA. Structural Assessment in Glaucoma. What s New

Intro to Glaucoma/2006

MOVE IT OR LOSE IT: THE ROLE OF KINETIC VISUAL FIELDS

Cirrus TM HD-OCT. Details define your decisions

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

A Review Of Risk Factors. Early Detection Of Glaucoma Clinical Clues. A risk factor analysis is critical. Points To Live By

Diagnosing open-angle glaucoma may be particularly

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma

Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients

Financial Disclosure. Visual Field Interpretation RELIABILITY VISUAL FIELD INTERPRETATION IN GLAUCOMA METHODS OF DATA PRESENTATION

LARGE DISCS WITH LARGE CUPS A DIAGNOSTIC CHALLENGE IN AFRICAN PATIENTS. Darshana Soma

MEDICAL POLICY. Proprietary Information of YourCare Healthcare

Position of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC

Diagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note.

International Journal of Advance Engineering and Research Development EARLY DETECTION OF GLAUCOMA USING EMPIRICAL WAVELET TRANSFORM

Relationship between GDx VCC and Stratus OCT in juvenile glaucoma

This policy is applicable to Commercial Products only. For BlueCHiP for Medicare, see related policy section.

The Evolution of Fundus Perimetry

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS

New Concepts in Glaucoma Ben Gaddie, OD Moderator Murray Fingeret, OD Louis Pasquale, MD

Ganglion cell analysis by optical coherence tomography (OCT) Jonathan A. Micieli, MD Valérie Biousse, MD

Correspondence should be addressed to Verena Prokosch;

Mark Dunbar: Disclosure

Relationship Between Structure

tracking progression we can better manage our patients. Like any tool, any instrument you ve got to

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography

CHAPTER 13 CLINICAL CASES INTRODUCTION

NIH Public Access Author Manuscript Br J Ophthalmol. Author manuscript; available in PMC 2010 April 29.

Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects

Ganglion cell complex scan in the early prediction of glaucoma

3/23/2016. Regina Smolyak, MD Glaucoma Service Flaum Eye Institute

Ophthalmologic Techniques That Evaluate the Posterior Segment for Glaucoma

International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN

Reliability analyses of the GDx nerve-fiber analyzer

A comparison of HRT II and GDx imaging for glaucoma detection in a primary care eye clinic setting

University Hospital Basel. Optical Coherence Tomography Emerging Role in the Assessment of MS PD Dr. Konstantin Gugleta

SOCT Copernicus REVO. * - Currently import and overlay are avaibale in manual mode only

Correlating Structure With Function in End-Stage Glaucoma

Retinal Nerve Fiber Analysis: the Role It Plays in Assessing Glaucoma Grace Martin, CPOT

The World s fastest OCT. As simple as pressing. the start button

Available online at Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6):

Scanning laser polarimetry (SLP) provides real-time, objective

Early diagnosis of glaucoma

Sequential non-arteritic anterior ischemic optic neuropathy (NAION) Jonathan A. Micieli, MD Valérie Biousse, MD

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection

Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthalmology

Key words: Glaucoma, Imaging, Ophthalmoscopy, Optic neuropathy, Topography

The only measure of RNFL Integrity : GDxPRO

Access to the published version may require journal subscription. Published with permission from: Elsevier

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7

CLINICAL SCIENCES. optic neuropathy characterized

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography

RETINAL NERVE FIBER LAYER

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation

Moving forward with a different perspective

In some patients with glaucoma, standard (achromatic) automated

Differences between Non-arteritic Anterior Ischemic Optic Neuropathy and Open Angle Glaucoma with Altitudinal Visual Field Defect

VISUAL FIELDS. Visual Fields. Getting the Terminology Sorted Out 7/27/2018. Speaker: Michael Patrick Coleman, COT & ABOC

Clinical decision making based on data from GDx: One year observations

Research Article The Pattern of Retinal Nerve Fiber Layer and Macular Ganglion Cell-Inner Plexiform Layer Thickness Changes in Glaucoma

The only measure of RNFL Integrity: GDxPRO

Although measurements of the optic disc and retinal nerve

Glaucoma Diagnosis. Definition of Glaucoma. Diagnosing Glaucoma. Vision Institute Annual Fall Conference

Glaucoma Progression. Murray Fingeret, OD

Visual Fields Shawn L. Cohen, M.D. Part 2 of 4. Definitions / Tables (Part 2 of 2) Static Perimetry (Humphrey, Octopus)

eye as a camera Kandel, Schwartz & Jessel (KSJ), Fig 27-3

Transcription:

CET CONTINUING Sponsored by 1 CET POINT Linking structure and function in glaucoma 50 Dr Samantha McGinnigle PhD, BSc (Hons), MCOptom, AHEA This article will give an overview of the latest imaging technology available to assess the optic disc. This will then be compared with the information provided by visual field assessment including standard automated perimetry (SAP), short-wavelength automated perimetry (SWAP) and frequency doubling technology (FDT). The different mechanisms for each type of perimetry will be discussed. Course code: C-34074 Deadline: November 29, 2013 Learning objectives To understand different approaches when undertaking visual field assessment (Group 3.1.5) To recognise when to refer patients with suspected glaucoma based upon clinical findings (Group 6.1.8) To understand the needs of patients with advanced disease when examining visual fields (Group 7.1.7) Learning objectives To understand the instruments involved in visual field analysis (Group 3.1.5) To understand the implications of glaucoma manifestations (Group 8.1.5) About the author Numerous instruments to capture images of the fundus have been developed in recent years, enabling the practitioner to undertake objective assessment of the optic disc. This article will give an overview of the latest imaging technology available, comparing this with information obtained by visual field assessment.

Sponsored by UNDETECTABLE DISEASE Ganglion cell death/ axon loss Retinal nerve fiber layer change (undetectable) ASYMPTOMATIC DISEASE Retinal nerve fiber layer change (undetectable) Short wavelength automated perimetry VF change Short wavelength automated perimetry VF change VF change (moderate) FUNCTIONAL IMPAIRMENT 51 Acceleration of apoptosis Normal VF change (severe) Blindness Figure 1 Glaucoma progression as a continuum. Reprinted from Weinreb et al. Am J Ophthalmol, 2004:138(3);458 467 Introduction Glaucoma is a progressive optic neuropathy, associated with characteristic visual field defects, caused by structural changes in the optic nerve head and peripapillary retinal nerve fibre layer (see Figure 1). 1 Research in an established glaucoma referral refinement pathway has shown a decline in referral accuracy by optometrists with no special interest in glaucoma as opposed to those with additional training. Specifically, this relates to the detection of disc abnormalities and abnormal intraocular pressure (IOP) following the publication of the National Institute for Health and Clinical Excellence (NICE) glaucoma guidelines. 2 Monitoring optic disc changes, and more specifically, assessing the retinal nerve fibre layer (RNFL), can provide predictive data on which patients with elevated IOP will go on to develop glaucoma. Over the past two decades there have been considerable advances in technology, leading to the development of a vast array of equipment for fundus imaging. Structural changes to the optic disc and retinal nerve fibre layer often precede visual field defects, which can be measured with standard achromatic perimetry in early glaucoma. 3 This could lead to the conclusion that visual field tests, particularly standard automated perimetry (SAP), are not the ideal screening tools for optometrists. However, structural and functional measures in glaucoma contribute different and, arguably, equally important information for assessing the eye, despite the variable correlation between these techniques, in relation to the disease stage. Standard automated perimetry (SAP) SAP is the gold standard measure for visual fields. The most effective detection criteria for glaucomatous visual field loss are localised loss for individual points, hemifield clusters, or asymmetrical sensitivity across the horizontal meridian. 4 SAP is not selective for a particular ganglion cell type, therefore, any of the primary ganglion cell types can respond to the achromatic stimulus presented on a white background. A significant amount of ganglion cell loss can occur before SAP can identify functional deficits. However, progression of visual field defects can be difficult to quantify due to high test-retest variability. This was highlighted in the ocular hypertension study where 85.9% of initial field assessments showing defects were found to be normal on repeat testing. The findings in this study led to the recommendation of three consecutive tests. However, even with this more stringent criteria applied, some eyes still produce normal visual field tests on further follow-up. 5 Approximately one third of the variability in SAP test results can be accounted for by the severity of the defect, it s location and the patient s diagnosis. 6 However, factors such as patient performance and reliability, fixation losses, fatigue, learning effects, changes in pupil size, improper refractive correction, and true physiological variability also play a part. 7 The Swedish Interactive Thresholding Algorithm (SITA) uses an adaptive algorithm based on the patient s responses, which significantly decreases testing time (to four or five minutes for 24-2) without compromising accuracy, as compared to much longer, full- Figure 2 The GDxPRO Scanning Laser Polarimeter. Image courtesy of Carl Zeiss, Meditec For the latest CET visit www.optometry.co.uk/cet

1 CET POINT CET CONTINUING 52 threshold testing. 8 The SITA fast 24-2 reduces testing time further, to just over three minutes. However, this is at the expense of precision as it allows for a greater variability in responses on individual point re-testing. 9 Short wavelength automated perimetry (SWAP) SWAP projects a Goldmann size V blue target against a bright yellow background. The background reduces the sensitivity of the green and red cones, therefore the blue (shortwavelength sensitive) cones and their small, bistratified retinal ganglion cells are isolated (the koniocellular pathway). Longitudinal studies have shown that SWAP defects may occur three to five years before abnormalities are seen on full-threshold SAP, and are predictive of both the onset and location of future SAP defects. 10 Approximately 10% of ganglion cells are of the bistratified variety. The most widely accepted theory suggests that by testing fewer sparsely distributed cells with large retinal fields, the results are less likely to be masked by other pathways. 10 SWAP has traditionally been recommended as a test to evaluate visual function in younger glaucoma patients. This is due to the yellow tone of the nuclear cataract acting as a blue filter and causing significant diffuse depression of sensitivity. The limitations of prolonged testing have largely been overcome with the introduction of the SWAP SITA. However, the yellow background and the blue spot target of SWAP are still more difficult to recognise than the white-on-white test, leading to increased patient fatigue and discomfort during the test. Testing the blue cone system has also been shown to have higher intra- and inter-variability compared to white testing. 11 In advanced disease, the Figure 3 HRT takes data from a 3D stack of tomographic images of the optic nerve and RNFL, aligns the images and computes a 3D topographic map of the retinal surface. Image courtesy of Heidelberg Engineering patient may not recognise even the brightest target. Therefore, due to the narrowing of the dynamic range, the test may not be sensitive enough to monitor progression in these cases. Frequency doubling technology (FDT) When a low spatial frequency sinusoidal grating (less than one cycle per degree) undergoes high temporal frequency counterphase flicker (greater than 15 Hz), the stimulus appears to have twice as many light and dark bars than are physically present a phenomenon known as frequency doubling. The origins of this response are thought to be cortical, although it has been reported that FDT selectively tests the contrast sensitivity of the magnocellular cells. 12 Magnocellular retinal ganglion cells are primarily involved in motion and flicker detection and comprise approximately 10% of the entire retinal ganglion cell population. Evidence of a reduced ability of the visual system to use other subsets of retinal ganglion cells to compensate for damaged retinal ganglion cells of the type being tested supports the utility of the test. 13 A 24-2 version using an algorithm called Zippy Estimation of Sequential Thresholds (ZEST) with similar forecasting principles to SITA is available, and has comparable performance. 14 However, there are also reports that FDT identifies field loss before SAP. 15,16 FDT has shown lower test-retest variability than SAP and appears to have higher sensitivity than SWAP at a comparable level of specificity. 17 Disadvantages include: fading of the testing screen due to the Troxler phenomenon (where stimuli away from the fixation point may seem to disappear) and unreliable results in patients with cataracts, 19 high myopia or small pupils. 20 Imaging technology Scanning laser polarimetry (SLP), confocal scanning laser ophthalmoscopy (CSLO) and optical coherence tomography (OCT) make use of the different properties of light and the varied characteristics of retinal tissue to provide real-time, non-invasive and high-resolution images. There have been numerous reports indicating how the sensitivity and specificity of diagnosis could be significantly increased by the combined use of imaging techniques and psychophysical testing. These will be detailed after each device description. Scanning laser polarimetry SLP is a technique for assessing the thickness of the RNFL, with potential for diagnosis and follow-up of patients with glaucoma. The RNFL exhibits bi-refringence due to the structure of retinal nerve fibre bundles. The instrument measures the retardation or phase shift of the polarised light as it passes through the tissue. The amount of retardance exhibited by the RNFL is proportional to its thickness. 21 Larger amounts of retardation are seen next to the blood vessels, superior and inferior to the optic nerve head in most normal healthy subjects. However, scans from eyes with lightly pigmented fundi, myopes,

Sponsored by and some elderly patients can show an provided, although interestingly, this does atypical retardation pattern. 22 In addition, not rely on manual delineation of the the cornea (and to a lesser extent, the lens) optic disc margin. Rather, a predefined are also bi-refringent. When assessing RNFL model of the optic nerve head structure retardation, anterior segment retardation is used. As the name suggests, confocal must be neutralised. However, both atypical scanning laser ophthalmoscopy provides retardation and residual anterior segment a global and sectorial quantification of bi-refringence can have a detrimental effect on the accuracy of measurements the region. However, there are limitations to the technique and the investigators 53 taken by SLP. Fixed and variable corneal who derived the calculation noted that compensators have been utilised to cancel the early pattern of glaucomatous change bi-refringence arising from the anterior may be different to the most substantially segment, based on population averages altered shape parameters they identified and individual, eye-specific compensation. Enhanced corneal compensation (ECC) uses an algorithm to further improve the signalto-noise ratio of RNFL retardation, reducing the errors associated with atypical scan results. 23,24,25 Images of the fundus are obtained by a scanning beam of near-infrared laser Figure 4 Cirrus HD-OCT with an axial resolution of 5µm. Image courtesy of Carl Zeiss, Meditec generally slow. However, this is significantly reduced following the application of topical therapy, regardless of whether they develop POAG. 30 in the study. 32 An alternative method of analysis the Moorfields regression analysis (MRA) requires manual outlining of the disc boundaries. However, the diagnostic performance of the methods appear to be similar, with accuracy for classification dependent on optic disc size and glaucoma severity. 33 The development of an ethnic- light (780nm), which moves in a raster pattern to include the peripapillary and macular regions of the eye. The latest SLP devices such as the GDx PRO (Carl Zeiss Meditec) can compare retinal nerve fibre thickness at the optic nerve head and indicate significant thinning or thickening of the RNFL (see Figure 2). The nerve fibre index (NFI) is an algorithm that takes this thickness information and combines it with other variables to give a value indicating disease status. The device also provides characterisation of the microstructures within the RNFL as it has been shown that changes in the orientation and density of the microtubules occur before thickness changes become apparent. 26,27 How does SLP compare with perimetry? SLP with variable corneal compensation has been shown to correlate well with perimetry performed by the Humphrey Field Analyser, using the 24-2 full threshold or 24-2 SITA standard programme. 28 The ocular hypertensive treatment study demonstrated that lowering intraocular pressure in eyes with ocular hypertension without glaucomatous optic neuropathy or visual field damage decreased the incidence of primary open angle glaucoma (POAG) by 50%. 29 The average velocity of visual field Confocal scanning laser ophthalmology (CSLO) The Heidelberg Retina Tomograph III (HRT III; Heidelberg Engineering) is the latest commercially available CSLO. The device uses a diode laser (670nm wavelength) to sequentially scan 15 x 15 areas of the retinal surface in horizontal and vertical directions in multiple focal planes. The coronal (longitudinal) planes are effectively stacked up (see Figure 3) and this reassembly enables an assessment of the optic nerve head topography. 31 Three sets of scans are acquired in rapid succession and, after the images are aligned, the average measurement is used for analysis. The operator must then outline the optic nerve head margin to enable the device to define a reference plane. Parameters can then be calculated to quantify the optic nerve head region and produce a reflectance image overlaid with a topographic map. The reference plane is automatically generated by the CSLO software and calculated as 50µm below the mean retinal surface level. All structures within the contour line and above the reference plane are considered to be neuro-retinal rim, and all structures below the reference plane are considered to be the cup. The specific normative database may improve sensitivity and specificity in cases with very large or small discs. 33 Advantages of HRT include good image quality through undilated pupils (though dilation may be necessary at times), and the ability to upgrade existing machines with newer software, allowing the clinician to build upon long-term databases. Most importantly, the sophisticated registration capability of HRT to superimpose baseline and follow-up images allows for change to the ONH to be detected automatically. The use of HRT in the ancillary study to the ocular hypertension treatment study has resulted in a well-characterised data set which is beneficial for future investigations of this technique. How does CSLO compare with perimetry? An interesting study using patternrecognition algorithms known as machine learning classifiers, has been applied to optic disc imaging results and visual field data. Machine learning classifiers are well suited for evaluating these large data as they summarise the information in order to detect complex patterns and trends. This method has shown improved accuracy in glaucoma diagnosis for combined optimised change in untreated ocular hypertension is glaucoma probability score (GPS) is also HRT and SWAP data. 34 A study investigating For the latest CET visit www.optometry.co.uk/cet

1 CET POINT CET CONTINUING structure-function relationships using refining diagnostic parameters as reduced subjective examination and permits the perimetry and scanning laser ophthalmoscopy choroidal thickness appears to be the most measurement of features not otherwise in glaucoma patients showed correlations influential factor on the occurrence of normal possible using other methods. The are strongest in temporal areas where tension glaucoma in high myopes. 38 technology allows for an expert examination 54 glaucomatous damage tends to occur first, particularly for FDT compared with SAP. 35 Optical coherence tomography OCT works in a similar way to ultrasound, using near-infrared light instead of sound waves to produce high resolution, crosssectional images. Time domain OCT has been superseded by the higher quality spectral domain (SD-OCT), which permits direct, realtime visualisation of retinal pathology. SD-OCT also provides quantitative measurements of retinal architecture at higher resolutions than CSLO and SLP (see Figure 4). The increased speed reduces motion artefacts and improves patient compliance, although patients may still require dilation and image quality can be compromised in patients with ocular opacities. Repeatability of nerve fibre layer thickness measurements in patients with glaucoma and healthy volunteers has shown good intra- and inter-observer repeatability in comparison with time domain OCT. 36 SD-OCT has been used to quantify the loss of retinal ganglion cells in experimentally induced anterior ischemic optic neuropathy, and highpenetration OCT using long wavelengths has been used to characterise choroidal thickness in highly myopic eyes with normal tension glaucoma. 37 This may have implications for How does SD-OCT compare with perimetry? The RNFL assessment with SD-OCT performed well in detecting pre-perimetric glaucomatous damage in a cohort of glaucoma suspects and delivered a better performance than CSLO. 39 For diagnostic purposes, machine learning classifiers trained on OCT and SAP data can successfully discriminate between healthy and glaucomatous eyes. When combined, these measurements improved diagnostic accuracy compared with OCT data alone. 40 A study evaluating the relationship between SAP, changes in estimated retinal ganglion cell (RGC) counts and measures of RNFL thickness showed the same amount of RGC loss corresponded to the mean deviation (MD) change depending on the stage of the disease, that is, small changes in MD, but relatively larger changes in RNFL thickness for early disease stages as opposed to larger changes in MD, but only small or no changes in average RNFL thickness in later disease stages. 41 Conclusion Imaging allows an objective, non-contact measure that is more reproducible than a by non-experts, however, this relies on quality control measures being carried out by the instrument software to establish accurate diagnosis and change criteria. The operator s level of skill at the time of image acquisition and careful interpretation are vital. Scans containing artefacts from movements or areas with signal dropout cannot be interpreted correctly as it is impossible for the analysis algorithm to accurately predict measurement details. Certain cases, such as high myopes or tilted discs, are not suitable for comparison with data from a normative database and may be normal, but flagged as abnormal by the statistical analysis failing to recognise that they fit at the low end of the distribution. In summary, imaging findings should never be accepted as the absolute measure and must be considered alongside more low-tech examination findings. Combined method analysis can prove to be more sensitive and specific, however, they are not readily available to non-research optometrists. Therefore, simply assessing correspondence between multiple techniques (even if they are individually variable) is far more likely to give an accurate and more complete overview of the patient s disease status. MORE INFORMATION References Visit www.optometry.co.uk/clinical, click on the article title and then on references to download. Exam questions Under the new enhanced CET rules of the GOC, MCQs for this exam appear online at www.optometry.co.uk/cet/exams. Please complete online by midnight on November 29, 2013. You will be unable to submit exams after this date. Answers will be published on www.optometry.co.uk/cet/exam-archive and CET points will be uploaded to the GOC every two weeks. You will then need to log into your CET portfolio by clicking on MyGOC on the GOC website (www.optical.org) to confirm your points. Reflective learning Having completed this CET exam, consider whether you feel more confident in your clinical skills how will you change the way you practice? How will you use this information to improve your work for patient benefit?