Research Article Validation of an Endoscopic Fibre-Optic Pressure Sensor for Noninvasive Measurement of Variceal Pressure

Similar documents
Michele Bettinelli RN CCRN Lahey Health and Medical Center

Index. Note: Page numbers of article titles are in boldface type.

Evidence-Base Management of Esophageal and Gastric Varices

Information Technology Solutions

Primary Prophylaxis against Variceal Hemorrhage Pharmacotherapy vs Endoscopic Band Ligation

Tranjugular Intrahepatic Portosystemic Shunt

Index. Note: Page numbers of article titles are in boldface type.

NYU School of Medicine Department of Radiology Rotation-Specific House Staff Evaluation

Transjugular intrahepatic portosystemic shunt (TIPS) Information for patients Sheffield Vascular Institute

VARICEAL BLEEDING. Ram Subramanian MD Hepatology & Critical Care Medical Director of Liver Transplant Emory University, Atlanta.

LIVER CIRRHOSIS. The liver extracts nutrients from the blood and processes them for later use.

Transjugular Intrahepatic

PORTAL HYPERTENSION. Tianjin Medical University LIU JIAN

Cirrhosis and Portal Hypertension Gastroenterology Teaching Project American Gastroenterological Association

Thrombocytopenia and Chronic Liver Disease

The Leeds Teaching Hospitals NHS Trust Transjugular Intrahepatic Portosystemic Shunt (TIPS)

Carvedilol or Propranolol in the Management of Portal Hypertension?

COPYRIGHTED MATERIAL. 1 Approach to the patient with gross gastrointestinal bleeding. Grace H. Elta, Mimi Takami

Portogram shows opacification of gastroesophageal varices.

Background: Bedside ultrasound is emerging as a useful tool in the assessment of

Esophageal Varices Beta-Blockers or Band Ligation. Cesar Yaghi MD Hotel-Dieu de France University Hospital Universite Saint Joseph

Obliterative hepatocavopathy ultrasound and cavography findings

Barbara Rus Gadžijev Peter Popovič Klinični inštitut za radiologijo UKC Ljubljana

INTRAHEPATIC PSS: AN INTERVENTIONALIST S PERSPECTIVE

Diagnostic Procedures. Measurement of Hepatic venous pressure in management of cirrhosis. Clinician s opinion

Transjugular Intrahepatic Portosystemic Shunt (TIPS) About your procedure

Transjugular Intrahepatic Portosystemic Shunt Reduction for Management of Recurrent Hepatic Encephalopathy

Evaluation of liver and spleen stiffness using a ultrasound guided method: Accuracy of ARFI(R) measurements in liver disease patients

Anaesthetic considerations for laparoscopic surgery in canines

Sample page. POWER UP YOUR CODING with Optum360, your trusted coding partner for 32 years. Visit optum360coding.com.

CPT 2014 Overview of GI Changes

Left-sided portal hypertension with a patent splenic vein: An impossible or a not-so-uncommon scenario?

Clinical Trials & Endpoints in NASH Cirrhosis

DRAFT. Angiography: Transjugular Intrahepatic Portosystemic Shunt (TIPS) What to expect. What is a transjugular intrahepatic portosystemic shunt?

Evaluation of liver and spleen stiffness using a ultrasound guided method: Accuracy of ARFI(R) measurements in liver disease patients

Denver Shunts vs TIPS for Ascites

Matching study design to research question-interactive learning session

Variceal bleeding. Mainz,

Virtual Mentor American Medical Association Journal of Ethics December 2008, Volume 10, Number 12:

ACTIVITY USING RATS A METHOD FOR THE EVALUATION OF ANALGESIC. subject and a variety of stimuli employed. In the examination of new compounds

Vascular Technology Examination Content Outline

Sandipan Ghose, Md. Azizul Hoque, Md. Khalilur Rahman, Mohd. Harun-or-Rashid

AN UNUSUAL CAUSE OF HEPATIC ENCEPHALOPATHY

Privileges for San Francisco General Hospital

Eun-Ju Kang, M.D., Seong Kuk Yoon, M.D., Sang-Hyeon Kim, M.D., Jin Han Cho, M.D., Myong Jin Kang, M.D., Byeong-Ho Park, M.D.

Journal of American Science 2014;10(10)

RadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.

History and Science of CO 2 Digital Subtraction Angiography

Title: Cholestasis after TIPS placement in a patient with primary sclerosing cholangitis: an uncommon complication

Instruction for Use of 3-stage Dilation Balloon Catheter JS-CE-SM-08-01, A/0

Preliminary study of the permeability and safety of covered stents-grafts in pediatric TIPS

2012 CPT Changes Affecting Radiology REVISIONS

Barrett s Esophagus. lining of the lower esophagus that bears his name (i.e., Barrett's esophagus). We now

Transjugular liver access and biopsy

Ascites Management. Atif Zaman, MD MPH Oregon Health & Science University Professor of Medicine Division of Gastroenterology and Hepatology

Wireless Capsule Endoscopy

Tools of the Gastroenterologist: Introduction to GI Endoscopy

Catheter-directed Thrombolysis

Transcatheter Aortic Valve Implantation Procedure (TAVI)

Stripping, the most common method

September Copyright 2017 Cath Lab Solutions LLC All Rights Reserved.

Risk factors for 5-day bleeding after endoscopic treatments for gastroesophageal varices in liver cirrhosis

Liver failure &portal hypertension

Initial placement 24FR Pull PEG kit REORDER NO:

Management of Ascites and Hepatorenal Syndrome. Florence Wong University of Toronto. June 4, /16/ Gore & Associates: Consultancy

Guidelines, Policies and Statements D5 Statement on Abdominal Scanning

Spontaneous portosystemic venous shunts in liver cirrhosis: Anatomy, pathophysiology, hemodynamic changes and imaging findings

Hepatopulmonary Syndrome: An Update

Variceal wall tension is thought to represent the key

Historical perspective

The Frequency of Gastroesophageal Reflux Disease in Nutcracker Esophagus and the Effect of Acid-Reduction Therapy on the Motor Abnormality

Acute Upper Gastrointestinal Hemorrhage Surgical Perspective. Dr.J.H.Barnard Dept. of Surgery PAH

iu22 Liver Shear Wave ElastPQ

DR. DO NGUYEN TIN CHILDREN HOSPITAL 1

Definition: HPS is a disease process with a triad of: 1- Liver disease. 2- Widespread intrapulmonary vasodilatation. 3- Gas exchange abnormality prese

CPT COD1NG UPDATES Gastroenterology CPT Advisors

Relative Values for Physicians. Relative Value Studies, Inc.

CY2017 Hospital Outpatient: Vascular Procedure APCs and Complexity Adjustments

AN EXPERIMENTAL TUBE PEDICLE LINED WITH SMALL BOWEL. By J. H. GOLDIN, F.R.C.S.(Edin.) Plastic Surgery Unit, St Thomas' Hospital, London

Alpha-1 Antitrypsin Deficiency: Liver Disease

INITIAL CLINICAL PRIVILEGES DELINEATION FORM Department of Medicine Monmouth Medical Center

Initial placement 20FR Guidewire PEG kit REORDER NO:

MANAGEMENT OF LIVER CIRRHOSIS: PRACTICE ESSENTIALS AND PATIENT SELF-MANAGEMENT

Detection of Esophageal Varices in Liver Cirrhosis Using Non-invasive Parameters

The Value of Renal Artery Resistive Indices: Association with

Lecture 2: Clinical anatomy of thoracic cage and cavity II

following the last documented transfusion; thereafter, evaluate the residual impairment(s).

clearing activity is produced and destroyed in the rat. Both the

Background & Indications Probe Selection

Review Article Self-Expandable Metal Stents in the Treatment of Acute Esophageal Variceal Bleeding

Clinical Study Correlation between Endosonographic and Doppler Ultrasound Features of Portal Hypertension in Patients with Cirrhosis

Pressure. 3. Direct Pressure. Pressure in Human Body. Pressures and Ranges. Effects of Kinetic Energy. Effect of Gravitational Force

Over the Wire Technique vs. Modified Seldinger Technique in Insertion of PICC

For Reference Only PEDIATRIC MEDICINE 2013

MODULE 2 THE LABORATORY RAT

הצגת מקרה- מפגש מכונים מאי 2017

Detection of Esophageal Varices Using CT and MRI

UNC HOSPITALS CHAPEL HILL, NORTH CAROLINA REQUEST AND AUTHORIZATION FOR UPPER GASTROINTESTINAL ENDOSCOPY AND BIOPSY MIM#180

Original article Correlation between serum-ascites albumin concentration gradient and endoscopic parameters of portal hypertension

Original Article PLATELET COUNT TO SPLEEN DIAMETER RATIO AS A PREDICTOR OF ESOPHAGEAL VARICES IN PATIENTS OF LIVER CIRRHOSIS DUE TO HEPATITIS C VIRUS

Transcription:

Hindawi Publishing Corporation BioMed Research International Volume 2016, Article ID 1893474, 7 pages http://dx.doi.org/10.1155/2016/1893474 Research Article Validation of an Endoscopic Fibre-Optic Pressure Sensor for Noninvasive Measurement of Variceal Pressure Bin Sun, 1 De-Run Kong, 1 Su-Wen Li, 1 Dong-Feng Yu, 2 Ging-Jing Wang, 1 Fang-Fang Yu, 1 Qiong Wu, 1 and Jian-Ming Xu 1 1 Department of Gastroenterology, The First Affiliated Hospital of Anhui Medical University, Hefei, Anhui 230022, China 2 Department of Surgery, The First Affiliated Hospital of Anhui Medical University, Hefei, Anhui 230022, China Correspondence should be addressed to Jian-Ming Xu; xujm1017@163.com Received 20 December 2015; Revised 15 April 2016; Accepted 27 April 2016 Academic Editor: John J. Gildea Copyright 2016 Bin Sun et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. In this study, the authors have developed endoscopic fibre-optic pressure sensor to detect variceal pressure and presented the validation of in vivo and in vitro studies, because the HVPG requires catheterization of hepatic veins, which is invasive and inconvenient. Compared with HVPG, it is better to measure directly the variceal pressure without puncturing the varices in a noninvasive way. 1. Introduction Varicealpressureisnotonlyamajorpredictivefactorfor variceal bleeding, but also a useful parameter to evaluate the effect of pharmacotherapy of portal hypertension [1]. Therefore, it is important to regularly measure variceal pressure for cirrhotic patients. The conventional method of determining variceal pressure is directly puncturing the varix by using a fine needle [2]. This method needs to be improved because it was invasive and has the high risk of bleeding. Several attempts have been taken to develop a noninvasive or minimally invasive measurement, but none of them has been successful due to their inconvenience and impracticability in use [3 8]. We recently developed an endoscopic fibreoptic pressure sensor (EFPS) to measure variceal pressure [9]. Compared with conventional methods of measuring variceal pressure, EFPS is much easier to place the pressure sensor on the suitable varices through the endoscopic biopsy channel. In this study we tested the accuracy and variability of EFPS in vitro using a varix model and in vivo using canine mesenteric veins that are comparable in size to esophageal varices. 2. Material and Methods Measurement of Variceal Pressure Using the Endoscopic Fibre- Optic Pressure Sensor (EFPS). EFPSwasmanufacturedby Jiuhong Medical Instrument Co., Ltd. (Changzhou, China). Variceal pressure measurement by using EFPS has been described previously [9]. EFPS main unit is connected to a fibre optical that ends in a miniature fibre-optic pressure sensor (FOP-F125, FISO Technologies Inc., Quebec, Canada). When the endoscope is introduced into the distal end of the esophagus, the fibre-optic pressure sensor is compressed by the retrograde insertion through the working channel on the varix with endoscopic control (Figure 1). However, we noticedthatthesensorneedstobeheldbyhandsothat EFPS connects well with the varix during thewhole process. The pressure tracing detected by EFPS was shown as a sharp upwardandstableplateaucurve(figure2).forthein vitro studies, EFPS also showed a plateau phase that is a stable line. However, the major difference between the in vivo study and the in vitro study by EFPS is that the pressure tracing has periodic fluctuations in accord with breathing variation when used in vivo. During the experimental procedure, 2 independent measurers need to be blinded to the pressure in the artificial varix and in the actual mesenteric vein. 2.1. In Vitro Study. Previously,wehavedevelopedanartificial variceal model with a diameter of 4 or 8 mm and a length of 5 cm. It was made from 0.2 mm thick latex and was fixed to the inner wall of an artificial esophagus

2 BioMed Research International P 1 d H a b P 2 c Endoscope Figure 1: Noninvasive measurement of variceal pressure using the endoscopic fibre-optic pressure sensor inserted through the working channel. (mmhg) 30 27 24 21 18 15 12 9 6 3 0 0 5 10 15 20 25 30 35 40 45 50 55 60 (s) Figure 2: Variceal pressure recording by EFPS in vitro study. (diameter: 2.8 cm) [8]. These artificial varices are filled with water and then connected to a liquid column manometer (5 cm in diameter); its height determines the artificial variceal pressure; the zero reference point was established at the level of the middle of artificial varix. The sensor was gently compressed on the artificial varix to measure the artificial variceal pressure (Figure 3). Theheightoftheliquidcolumnrepresentingtheartificial variceal pressure was recorded by one operator, while the other operator that was blinded to the artificial variceal pressure was recording the measuring pressure by the EFPS. In each experiment, 5 measurements were performed in random order for any pressure (increased by 2 cm H 2 Oatone time) varied from 6 to 40 cm H 2 O. 2.2. Animal Study 2.2.1. Ethics Statements. This study was carried out in strict accordance with the recommendations in the Regulations for the Administration of Affairs Concerning Experimental AnimalsapprovedbytheStateCouncilofPeople srepublic Figure 3: Experimental model of an esophageal varix for comparison of the endoscopic fibre-optic pressure sensor with direct measurement of artificial intravariceal pressure: a: esophageal model; b: artificial varices filled with water and connected to a water column; c: fibre-optic pressure sensor that was inserted through the working channel and compressed on the varix gently; d: polygraph. The pressure in the balloon (P 1 ) was recorded by polygraph of endoscopic fibre-optic pressure sensor. The pressure in the artificial varix (P 2 ) was measured using the height of the water column (H). of China. The research protocol was approved by the Ethics Committee of Anhui Medical University. 2.2.2. Animals. Ten mongrel dogs, weighing 13.8 16.7 kg (average 15.2±0.9 kg), were used in this experiment. All dogs were purchased from the Laboratory Animal Center of Anhui Medical University. Animals were housed in appropriate stainless steel cages with the size of 0.80 m 0.80 m 1.00 m. All cages were placed in an air-conditioned room; the recommended temperature and humidity were 20 C 25 C and 40% 70%, respectively. The cages were washed twice daily. In addition, during the cage washing, the dogs were taken outside for exercise to reduce their stress. The light cycle was controlled using an automatic timer (12-hour light, 12-hour dark). All animals were provided with 600 g of solid food per day with free access to water. Dogs were allowedatleast7daystoadapttotheirhomeenvironment before experiments. Dogs were considered to be fine based on medical history, physical examination, and laboratory test. The health of the dogs was monitored every week. During this period, all animals stayed healthy, with no clinical or pathological abnormalities based on the report from the veterinary examinations. Beforetheoperation,thesedogswerestarvedfor12hours. After anaesthetized with pentobarbital sodium (30 mg/kg) initially, these dogs were weighted. With the dogs supine, the limbswerefixedintheoperationtablewiththebandages,and the respiration and heart rates were monitored by a multiparameter monitor. All animals were breathing spontaneously during surgery and maintaining the anesthetization. The skin of abdomen was sterilized with 3% iodine and 75% alcohol. Laparotomy was performed by right epigastric incision to

BioMed Research International 3 (mmhg) 30 26 22 18 14 10 6 2 2 6 10 0 5 10 15 20 25 30 35 40 45 50 55 60 (s) Figure 5: Variceal pressure recording by EFPS in animal study. Figure 4: Portal vein pressure of dog was measured by directly compressing the fibre-optic-sensor on the exposed portal vein. expose and isolate the portal vein, superior mesenteric vein, and the inferior vena cava (IVC). A 7-F indwelling catheter that was filled with saline solution and linked to a manometer was maintained in the proximal branch of the superior mesenteric vein, which was at the same level as the portal vein. The mesenteric venous pressure was measured continuouslybyphilipssuresignsvm8multiparametermonitor.the zero point was set as the ventral surface of the IVC. After the portal vein pressure (PVP) was determined, the sensor pressures measured by EFPS were determined in each dog by direct compressing of the sensor on the exposed portal vein (Figure 4). In practice of animal experiments, in order to make a good contact of the sensor with the varix, the fibre-optic pressure sensor was held by hand and directly compressed on the portal vein. All pressure tracings were obtained using EFPS and recorded on a Yoto polygraph (Yoto Technologies Co., Hefei, China). Two operators performed each experiment. The sensor was gently compressed on the vein by one operator and the other set the zero reference point prior to the application of the sensor on the vein and assessed the quality and stability of the tracing as well as recorded sensor pressure. Each operator was unaware of the placement of the sensor or the sensor pressure recording. The criteria required for a satisfactory measurement included a zero reference tracing and a variceal pressure tracing with stable plateau for at least 10 seconds, with the sensor being maintained on the vein under direct view (Figure 5). The mean value of two satisfactory measurements determined the variceal pressure. At the end of the study the dogs were euthanized with an excess of anesthetic. 2.3. Clinical Study. Between December 2013 and April 2014, EFPS was assessed in eight cirrhotic patients (5 men and 3 women; mean age 53.5 years; etiology of cirrhosis for eight patients was hepatitis B virus) who are scheduled Figure 6: The variceal pressure of patient was measured by EFPS which was gently pressed onto the variceal surface under direct visual control. for transjugular intrahepatic stent-shunt (TIPS). All had a history of episodes of esophageal variceal bleeding and presented large esophageal varices (F2, F3) under endoscopy. Exclusioncriteriaincludedportalveinthrombosis,previous endoscopic treatment of varices (sclerotherapy or endoscopic band ligation), previous surgical portosystemic shunt or an intrahepatic portosystemic stent-shunt, hepatocellular carcinoma or other malignancy, severe clotting defects, and hepatic encephalopathy grade III or IV. Before variceal pressure measurement started, the patients were sedated with intravenous premedication (diazepam 5 mg; butylbromide scopolamine 20 mg), to reduce as much as possible excessive esophageal contractions. First, the endoscope was introduced into the distal esophagus. Then the fibre-optic-probe was protruded to the tip of endoscope through the working channel; the largest varix at the level of the distal portion of the esophageal varices was chosen for measurement (Figure 6). Between peristaltic waves, the probe was gently compressed on the largest varix and lasted for 5 seconds in order to obtain a stable variceal pressure tracing. The criteria required for a satisfactory measurement included a zero reference tracing and a variceal pressure tracing with stable plateau for at least 5 seconds (Figure 7). The mean value of two satisfactory measurements determined the variceal pressure.

4 BioMed Research International (mmhg) 30 27 24 21 18 15 12 9 6 3 0 0 5 10 15 20 25 30 35 40 45 50 55 60 Figure 7: Endoscopic variceal pressure recording showed the satisfactory variceal pressure tracing (note the fluctuations following cardiac and respiratory cycles as well as the stability of the tracing). Portal pressure was measured by insertion of a 7-F curved catheter into the portal vein connected to an electromanometric transducer during TIPS placement, which was performed within 24 hours after variceal pressure measurement. The difference between portal vein pressure and inferior vena caval pressure was recorded as portal pressure gradient (PPG). The midchest was used as the external zero reference. Pressure tracing with stable plateau lasting for at least 10 seconds was considered satisfactory. 2.4. Statistical Methods. Quantitative data were expressed as mean ± standard deviation (SD) and were compared using Student s t-test if the data were normally distributed. The correlation between variables was analyzed using the Pearson linear correlation test. Statistical analysis was done using the SPSS 10.0 software package. P < 0.05 was considered to be statistically significant. 3. Results 3.1. In Vitro Study. A total of 180 fibre-optic-sensor pressure measurements were obtained successfully by two operators after a brief training. Linear regression analysis showed a good correlation between fibre-optic-sensor pressure and actual intraluminal pressure in the variceal model for different diameters (r 0.998, P < 0.001; Figure 8). For the varix models with the diameter of 5 cm, the correlation coefficient between the fibre-optic-sensor pressure and actual intraluminal pressure for Dr. Sun was r = 0.999. The correlation coefficient between the fibre-optic-sensor pressure and actual intraluminal pressure for Dr. Li was r = 0.999. For the varix models with the diameter of 8 cm, the correlation coefficient between the fibre-optic-sensor pressure and actual intraluminal pressure for Dr. Sun was r = 0.999. The correlation coefficient between the fibre-optic-sensor pressure and actual intraluminal pressure for Dr. Li was r = 0.998. The correlation coefficient betweentheoperators(dr.sunanddr.li)fortheartificial intravariceal pressures was r = 0.998 (Figure 9). 3.2. Animal Study. Determination of fibre-optic-sensor pressure was technically successful in all of the session. The (s) tracing obtained by fibre-optic-sensor conformed to the tracing criteria stated above for acceptance. We noticed that the selection of satisfactory fibre-optic-sensor pressure tracings was done without the result of actual mesenteric venous pressure. There was a linear correlation between fibreoptic-sensor pressure and PVP values (r = 0.757, P < 0.001, Figure 10). 3.3. Clinical Study. Variceal pressure measurements in eight patients were technically successful. There was no difficulty in measuring the variceal pressure because of easy positioning of the fibre-optic-probe on the varix. The pressure tracing detected by EFPS conformed to the criteria stated above. No adverse effect related to the variceal pressure measurement was observed, especially no variceal bleeding during the measuring procedure. After the patients were given intravenous premedication with diazepam and butylbromide scopolamine, the esophageal peristalsis or breathing that would cause the excessive esophageal peristalsis and then affect the variceal pressure measurement was significantly decreased. We also found that the periods between two peristaltic waves were suitable for obtaining the satisfactory pressure tracing, because the esophagus was at rest during the session. In eight patients, the mean PPG value was 34.7 ± 5.9 mmhg, and mean variceal pressure value was 21.8 ± 4.4 mmhg. There was a linear correlation between variceal pressure and PPG values (r = 0.973, P < 0.001, Figure11). ItwasobviousthatPPGvalueswerehigherthanthevariceal pressure. 4. Discussion The variceal pressure can be measured either via directly puncturing the varix by using a fine needle or with noninvasive methods. In recent years, our team described a new noninvasive technique for measuring the variceal pressure. Compared with previous noninvasive technology of measuring variceal pressure, it was much easier to place the fibre-optic-sensor (diameter 2 mm) on the varices through the endoscopic biopsy channel and set up a good contact between the sensor and the vessel [9]. In order to be used clinically, variceal measurements obtained with this device must be accurate and reproducible. Therefore, we testified the reliability and reproducibility of this method in vitro and in vivo. In this study, we showed that the variceal pressure measurement using fibre-optic-sensor did not require any specialized training in addition to operating EFPS and standard endoscopy. Like the operation of the esophageal mucosal biopsy, the fibre-optic-sensor is easy to operate for measuring the variceal pressure with an endoscopic control by an operator as described in our previous study [9]. The pressuretracingobtainedwithefpswascharacterizedbya sharp upstroke and stable plateau phase during both in vitro and in vivo studies. By in vitro study, linear regression analysis demonstrated a good correlation between the fibre-opticsensor and the direct intraluminal pressure measurements for

BioMed Research International 5 Actual pressure (mmhg) r = 0.999 Actual pressure (mmhg) 5.00 r = 0.999 5.00 0.00 0.00 Measured pressure (mmhg) 0.00 Measured pressure (mmhg) (a) (b) Figure 8: Correlation between measured pressure and the actual intraluminal pressure for model varices of different diameters: (a) 8 mm and (b) 5 mm. Operator 2 (mmhg) Operator 2 (mmhg) 5.00 r = 0.999 5.00 r = 0.998 0.00 0.00 5.00 Operator 1 (mmhg) 0.00 5.00 Operator 1 (mmhg) (a) (b) Figure 9: The correlation coefficient between the operators for the artificial intravariceal pressures: (a) 8 mm and (b) 5 mm. different diameters of artificial varices. More importantly, the coefficient of variance showed that the results of fibre-opticsensor manometry were also reliable and reproducible. Intheanimalstudy,wealsoobservedapositivecorrelation between the mesenteric venous pressure and PVP in experimental dogs. We attributed these results to the fact that the diameter of the measuring surface of the fibreoptic-sensor was less than the diameter of these vessels. This may lead to a good contact between the fibre-optic-sensor and the vessel, thereby enabling pressure measurement to obtain a stable plateau phase. In clinical study, the varices of F2 to F3 in patients may be related to the high rate of success. We found that variceal pressure measured by EFPS correlated significantly with the PPG, which indicated the results of EFPS were reliable. As the previous studies showed, the portal vein pressure was higher than the variceal pressure, because of the prevariceal resistance in collateral vessels [10]. However, the clinical study that enrolled a small group of cirrhotic patients probably indicated the preliminary result. In order to further verify the accuracy of EFPS, larger studies with more cases of cirrhosis are underway. To our knowledge, this fibre-optic-sensor is the smallest gauge to measure variceal pressure, which can be retrogradely inserted through the working channel of an endoscope. In prior literature, a pressure-sensitive gauge attached to the tip of the endoscope was used to measure the variceal pressure. This noninvasive method was first developed by Rigau et al. andcontinuouslyimprovedbyothers.however,thereliability of the endoscopic gauge in vivo remained to be tested because ideal positioning of the gauge with diameter of 5.5 mm on the varix may be difficult and may result in variable readings, especially for the measurement in small varix [3, 11, 12].

6 BioMed Research International Measured pressure (mmhg) 12.00 9.00 6.00 r = 0.757 the clinical applicability of this device. We believe that this new device may help provide a simple and accurate method for determining variceal pressure, an important parameter related to the risk of initial variceal bleeding and the effectiveness of pharmacologic therapy. Competing Interests The authors declare that they have no competing interests. Authors Contributions 3.00 4.00 5.00 6.00 7.00 8.00 Actual pressure (mmhg) Figure 10: Correlation between measured pressure by EFPS and the actual portal vein pressure in animal study. 45.00 BinSunandDe-RunKongcontributeequallytothiswork and share first authorship. Acknowledgments The authors would like to thank Professor Qiyi Tang, from Howard University College of Medicine, for language assistance. This work was supported by Grant 81271736 from the National Natural Science Foundation of China. PPG (mmhg) 40.00 35.00 17.50 22.50 27.50 VP (mmhg) r = 0.973 Figure 11: Correlation between the variceal pressure (VP) and the portal pressure gradient (PPG) in cirrhotic patients. In addition to the diameter of gauge, the measurement of variceal pressure by using the conventional pressure-sensitive gaugewaslikelytobeaffectedbyperistalsisintheesophagus, which would make pressure measurement more difficult [11 13]. With the endoscope insertion and air insufflations, the esophageal contractions would increase more and more frequently even though the patients received intravenous premedication such as diazepam and butylbromide scopolamine to reduce possible excessive esophageal movement [9, 13]. Owing to the thermal stability problems that occurred in fibre-optic-sensor, the measured variceal pressures are susceptible to temperature. Before each measurement, the zero reference of fibre-optic-sensor must be set in order to acclimatize to the temperature at the time. In conclusion, our results indicate that EFPS has the potential to measure variceal pressure in vivo, which is independent of variceal size. Further studies in patients using this fibre-optic-sensor are now undertaken to determine References [1] R. K. Tandon and N. Saikia, Measuring intravariceal pressure, Gastrointestinal Endoscopy,vol.70,no.3,pp.414 416,2009. [2] E. D. Palmer, On correlations between portal venous pressure and the size and extent of esophageal varices in portal cirrhosis, Annals of Surgery,vol.138,no.5,pp.741 744,1953. [3] J.Rigau,J.Bosch,J.M.Bordasetal., Endoscopicmeasurement of variceal pressure in cirrhosis: correlation with portal pressure and variceal hemorrhage, Gastroenterology, vol. 96, no. 3, pp. 873 880, 1989. [4] P. H. Gertsch and J.-J. Meister, Pressure measurement in oesophageal varices: preliminary report on a new non-invasive method, Gut, vol. 28,no. 9, pp. 1162 1165,1987. [5]E.S.Miller,J.K.Kim,J.Gandehoketal., Anewdevice for measuring esophageal variceal pressure, Gastrointestinal Endoscopy,vol.56,no.2,pp.284 291,2002. [6] L. S. Miller, Q. Dai, A. Thomas et al., A new ultrasound-guided esophageal variceal pressure-measuring device, American Journal of Gastroenterology,vol.99,no.7,pp.1267 1273,2004. [7] A. K. Vegesna, C. Y. Chung, A. Bajaj et al., Minimally invasive measurement of esophageal variceal pressure and wall tension (with video), Gastrointestinal Endoscopy,vol.70,no.3,pp.407 413, 2009. [8] D.-R. Kong, J.-M. Xu, L. Zhang et al., Computerized endoscopic balloon manometry to detect esophageal variceal pressure, Endoscopy,vol.41,no.5,pp. 415 420, 2009. [9] D.-R. Kong, B.-B. He, A.-J. Wu, J.-G. Wang, F.-F. Yu, and J.-M. Xu, Fiberoptic sensor for noninvasive measurement of variceal pressure, Endoscopy, vol. 45, no. 2, pp. E55 E56, 2013. [10] J. Bosch, J. M. Bordas, J. Rigau et al., Noninvasive measurement of the pressure of esophageal varices using an endoscopic gauge: comparison with measurements by variceal puncture in patients undergoing endoscopic sclerotherapy, Hepatology, vol. 6, no. 4, pp. 667 672, 1986. [11] J. Polio, J. Hanson, E. Sikuler et al., Critical evaluation of a pressure-sensitive capsule for measurement of esophageal varix

BioMed Research International 7 pressure. Studies in vitro and in canine mesenteric vessels, Gastroenterology, vol. 92, no. 5, pp. 1109 1115, 1987. [12] J. L. Puckett, J. Liu, V. Bhalla et al., Ultrasound system to measure esophageal varix pressure: an in vitro validation study, American Journal of Physiology Gastrointestinal and Liver Physiology, vol. 288, no. 5, pp. G914 G919, 2005. [13] P.Gertsch,G.Fischer,G.Kleber,A.M.Wheatley,G.Geigenberger, and T. Sauerbruch, Manometry of esophageal varices: comparison of an endoscopic balloon technique with needle puncture, Gastroenterology, vol. 105, no. 4, pp. 1159 1166, 1993.