Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

Similar documents
Supplemental Material

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand:

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

RN-BC, MS, CCRN, FAHA

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART

Pediatric pacemakers & ICDs:

Cardiac Electrical Therapies. By Omar AL-Rawajfah, PhD, RN

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

Figure 2. Normal ECG tracing. Table 1.

Automatic External Defibrillators

Biphasic Clinical Summaries

SUPPLEMENTAL MATERIAL

Review of Pacemakers and ICD Therapy: Overview and Patient Management

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

Supplementary Online Content

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

ICD: Basics, Programming and Trouble-shooting

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

The Management of Heart Failure after Biventricular Pacing

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012

Rhythm Control: Is There a Role for the PCP? Blake Norris, MD, FACC BHHI Primary Care Symposium February 28, 2014

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm

Practice Questions.

Title: Automatic External Defibrillators Division: Medical Management Department: Utilization Management

Advanced ICD Concepts

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

Call Medtronic at 1 (800) to verify the patient s current implanted system

MEDICAL POLICY Cardioverter Defibrillators

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network

Intraoperative and Postoperative Arrhythmias: Diagnosis and Treatment

Cardiology. Objectives. Chapter

AF :RHYTHM CONTROL BY DR-MOHAMMED SALAH ASSISSTANT LECTURER CARDIOLOGY DEPARTMENT

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

CPR Guidelines - Update

INTERPRETING THE ECG IN PATIENTS WITH PACEMAKERS

Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition

NORTH OF ENGLAND CARDIOVASCULAR NETWORK (NECVN)

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker

The ARREST Trial: Amiodarone for Resuscitation After Out-of-Hospital Cardiac Arrest Due to Ventricular Fibrillation

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

University of Wisconsin - Madison Cardiovascular Medicine Fellowship Program UW CCU Rotation Goals and Objectives Goals

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

AF in the ER: Common Scenarios CASE 1. Fast facts. Diagnosis. Management

Ventricular tachycardia Ventricular fibrillation and ICD

Dual-Chamber Implantable Cardioverter-Defibrillator

Chain of Survival. Highlights of 2010 American Heart Guidelines CPR

Different indications for pacemaker implantation are the following:

Atrial Fibrillation 10/2/2018. Depolarization & ECG. Atrial Fibrillation. Hemodynamic Consequences

Transvenous Pacemaker Procedures

Preoperative Evaluation: Patients with Cardiac Disease

Defibrillation threshold testing should no longer be performed: contra

Peri-operative management of pacemakers and implantable cardiac defibrillators

PEDIATRIC SVT MANAGEMENT

Defibrillator. BiomedGuy

Management of RT patients with implanted cardiac devices: From recommendation to implementation

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

Model 5392 EPG Temporary Pacer

Pacing Lead Implant Testing. Document Identifier

Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report

ACLS Prep. Preparation is key to a successful ACLS experience. Please complete the ACLS Pretest and Please complete this ACLS Prep.

Chapter 80 Implantable Cardiac Devices

Wide QRS Tachycardia in a Dual Chamber Pacemaker Patient: What is the Mechanism?

Biventricular Pacing - Hemodynamic Benefit for Patients with Congestive Heart Failure

BHRS Prep course Pub style Quiz NOT A

CATHETER ABLATION FOR TACHYCARDIAS

Implantation of a Permanent Tined Endocardial Electrode into Right Atrium during Open-heart Surgery: Report of 3 Cases

Early Defibrillation. Dr. M. Ravishankar

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Cardiac Imaging in abnormal rhythm Role of MDCT

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Chapter 9. Learning Objectives. Learning Objectives 9/11/2012. Cardiac Arrhythmias. Define electrical therapy

A Rare Type of Pacemaker Mediated Tachycardia

Diploma in Electrocardiography

L: Cardiovascular. Saskatchewan Association of Licensed Practical Nurses, Competency Profile for LPNs, 3rd Ed. 107

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

Visualization of Defibrillation Simulation Using Multiple Transthoracic Electrodes

UNDERSTANDING YOUR ECG: A REVIEW

Need to Know: Implantable Devices. Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia

National Coverage Determination (NCD) for Cardiac Pacemakers (20.8)

Clinical Policy: Holter Monitors Reference Number: CP.MP.113

Reversion of ventricular tachycardia by pacemaker stimulation

Leadless Pacing. Osama Diab Assistant Prof. of Cardiology Ain Shams University Egypt

EXAMPLE PROTOCOL FOR PATIENT SCANNING PROCESS

The New England Journal of Medicine

11/21/18. EKG Pop Quiz. Michael Giocondo, MD Cardiac Electrophysiology Saint Luke s Cardiovascular Consultants

Clinical Cardiac Electrophysiology

MEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring

The pill-in-the-pocket strategy for paroxysmal atrial fibrillation

Transcription:

Case Report Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Taku Nishida MD, Tamio Nakajima MD, PhD, Yutaka Goryo MD, Ken-ichi Ishigami MD, PhD, Hiroyuki Kawata MD, PhD, Manabu Horii MD, PhD, Shiro Uemura MD, PhD, Yoshihiko Saito MD, PhD The First Department of Internal Medicine, Nara Medical University In May 2005 a 68-year-old woman received a VDD pacemaker implantation in the right pectoral region at our hospital for the treatment of complete atrioventricular block. In July 2008, she was diagnosed with dilated cardiomyopathy based on histological testing. In November 2008, she developed syncope due to ventricular tachycardia while at another hospital. She underwent external electrical cardioversion with an anterior-lateral paddle position using a single shock of 100 J. This shock led to severe bradycardia resulting in a transfer to our hospital. The physician who provided the shock could not have been aware that the patient had an implanted pacemaker. The skin above the pulse generator was burned. The electrocardiogram showed no pacing spikes or ventricular escape rhythm. Investigation of the pacemaker 3 hours after cardioversion revealed reprogramming of the device and a marked rise in the lead impedance (>3;000 ohm). Removal of the generator and implantation of a biventricular cardioverter defibrillator were required. The emergency situation, the small size of the generator, the small incision made using the buried suture method, and the patient s obesity all probably contributed to the physician s not noticing the implanted pacemaker. It is important to increase awareness of the severe consequences that may follow if the physician administering external defibrillation does not know about the patient s implanted pacemaker. (J Arrhythmia 2009; 25: 209 213) Key words: External cardioversion, Pacemaker malfunction, Ventricular tachycardia Introduction Implantable cardiac rhythm management devices have become more widely used in patients with not only bradycardia, but also life threatening tachycardia and severe heart failure. Therefore it is common to encounter a patient with implanted devices needing external electrical cardioversion to terminate life-threatening tachycardia. A recent study examined the safety and efficacy of external electrical cardioversion in patients with state-of-the-art rhythm-pacing devices. 1) Here, we report lead failure and pacemaker malfunction following external defibrillation in a patient with an implanted single-lead VDD pacemaker, in which the sternal paddle was placed directly over the pulse generator. Received 10, July, 2009: accepted 8, September, 2009. Address for correspondence: Taku Nishida MD, The First Department of Internal Medicine, Nara Medical University, 840 Shijocho, Kashihara, Nara, 634-0006 Japan. Phone: 0744-24-3051 Fax: 0744-22-9726 209

J Arrhythmia Vol 25 No 4 2009 Case report In May 2005, a 68-year-old obese woman received a single-lead VDD pacemaker at our hospital for the treatment of complete atrioventricular block. The pulse generator (ELA Medical Symphony DR model 2550) was implanted in the right pectoral region using a buried suture method to close the incision. The ventricular pacing electrode configuration was bipolar. In July 2008, the patient was admitted to our hospital with congestive heart failure. Cardiac catheterization showed no coronary stenosis and markedly decreased left ventricular ejection fraction of 25.7%. The patient was diagnosed with dilated cardiomyopathy based on histology of biopsy samples from the left ventricular myocardium. Optimal medication including a beta blocker and angiotensin receptor blocker improved the heart failure. In November 2008, she developed syncope while at another hospital. The electrocardiogram revealed that this symptom was caused by ventricular tachycardia. Intravenous lidocaine could not terminate the tachycardia, but resulted in hemodynamic collapse. External electrical cardioversion was undergone as an emergency procedure with an anterior-lateral paddle position using a single shock of 100 J. The ventricular tachycardia was terminated; however, severe bradycardia followed which was associated with sustained cardiogenic shock. Therefore, the patient was transferred to our hospital. Inspection of the patient s chest revealed a skin burn overlying the pulse generator. A blood test revealed mild anemia but no evidence of myocardial damage or any electrolyte abnormalities. One month before the third admission, testing of the pacemaker had shown normal sensing and pacing thresholds, and pacing impedance in the normal range. However, investigation of the pacemaker 3 hours after cardioversion revealed that the device was in standby mode, with settings of: VVI mode, basic rate of 70 beats per minute, pacing amplitude of 5.0 V with pulse width of 0.5 ms, ventricular sensitivity of 2.0 mv, and unipolar pacing configuration. After re-initialization of the pacemaker, measurement of the pacemaker output revealed a marked rise in the lead impedance, and no paced response could be evoked despite increase to the maximum pacing amplitude (Figure 1, 2a). A chest X-ray revealed cardiomegaly and mild pulmonary congestion but did not show the lead fracture (Figure 2b). Percutaneous intracardiac pacing inserted through right internal jugular vein and continuous injection of low-dose dopamine dramatically stabilized the hemodynamic state. After the patient had recovered from the serious hazard without any neurological abnormalities, the pulse generator was removed and a biventricular implantable cardioverter defibrillator was implanted successfully in the left pectoral region. (Figure 3) Discussion Safety of external defibrillation on implantable devices Since the late 1960s, a protective mechanism, the Zener diode, has been incorporated into pacemakers and defibrillators to insulate the electronic circuit from damage due to electrical shocks. 2) Nevertheless, there have been numerous case reports of pacemaker and/or lead dysfunctions caused by external electrical cardioversion. 3 6) Gould et al. reported the case of an 81-year-old female with sick sinus syndrome who developed ventricular fibrillation and was successfully defibrillated. In that case, as in the case reported here, the defibrillation paddle was placed on the pulse generator, which led to a complete loss of function of the pulse generator. 7) To the best of our knowledge, only two studies have analyzed the safety of external electrical cardioversion in patients with implanted cardiac devices. 1,4) The more recent report included 44 patients with pacemakers, implantable cardioverterdefibrillators, or cardiac resynchronization systems who underwent external electrical defibrillation for atrial fibrillation. Fifteen devices were implanted in the right pectoral region, and 8 leads were unipolar. There was no incidence of device or lead dysfunction following electrical cardioversion. In this study, however, a strict anterior-posterior paddle position was used to protect the device and lead system. Moreover, no case involved emergency electrical defibrillation. The older of the two studies included 36 patients with unipolar pacemakers implanted in the right pectoral region who received external electrical cardioversion with an anterior-lateral paddle position. This study reported an incidence of transient (up to 30 min) loss of capture in 50% of patients, sensing failure in 41% of patients, and three cases of pacemaker malfunction of which two required generator replacement. While most of the indications for cardioversion were atrial fibrillation or flutter, this study included one case of emergency direct current (DC) shock for ventricular fibrillation. In the patient with ventricular fibrillation, several 360 J shocks were required for VF recurrence, which resulted only in persistent elevation of the pacing threshold. In this study, however, the paddle had 210

Figure 1 Measurement of the pacemaker after re-initialization. Patient s data registered at the time of implantation could not be read, and the lead impedance was elevated to more than 3,000 ohm (box). been positioned as far as possible away from the pacemaker before shock delivery. Pitfalls in the clinical setting In our case, an emergency DC shock was delivered with the sternal paddle positioned directly over the pulse generator. The physician who provided the primary care, including the DC shock, could not have noticed that the patient had an implanted pacemaker. It seems difficult for emergency care providers to recognize implanted devices in patients with syncope due to ventricular tachycardia, or to obtain the patient s entire medical history. In addition, in this case, the small pulse generator, the small incision done by the buried suture method, and the patient s obesity (BMI of approximately 25) probably prevented the physician from noticing the implanted pacemaker. Possible causes of the pacemaker system malfunction Interestingly, after the external electrical cardioversion in this case, the lead impedance was elevated in such a measure as to indicate complete lead disconnection, whereas the chest radiograph did not provide evidence of lead fracture. In some reports, transient or persistent elevation in the pacing threshold has been observed, and this elevation was suggested to be caused by current-induced tissue damage at the electrode-endomyocardial interface. 3 5) However, the degree of elevation in lead impedance observed in this case has never been reported, nor has there been any report of lead fracture associated with cardiopulmonary resuscitation. The manufacturer s instructions do not specify the upper limit of defibrillation energy which the pulse generator could resist, but they do indicate that 211

J Arrhythmia Vol 25 No 4 2009 a) b) Figure 2 a) The 12-lead electrocardiogram on admission showed a wide QRS rhythm of 48 beats/min without any pacing spikes despite the programmed ventricular pacing rate. b) The chest radiogram after insertion of the pacing lead. Careful observation could not identify the lead fracture. a) b) QRS 167 ms Figure 3 Chest radiogram a) and 12-lead electrocardiogram b) after biventricular implantable cardioverter defibrillator implantation. a) The active fixation coil lead was placed at the right ventricular septum. b) A regular atrioventricular sequential pacing rhythm of 50 beats/min. The QRS width was 167 ms. shock paddles should be positioned as far as possible away from the generator. The single-lead VDD pacing system and the short distance between the paddle position and the entire pacemaker system could have magnified the tissue damage. The remaining possibility was that the defibrillation energy directly damaged the Zener diode or the electrical circuit of the pulse generator. 8) We did not perform lead measurement and analysis of the pulse generator. Therefore we could not identify the precise cause of the marked rise in lead impedance. We selected a left-sided approach when we performed the CRT-D implantation. Even though the previously implanted VDD leads were intact, we favored the left side since the defibrillation threshold would be higher in right-sided implantation than in left-sided implantation, 9) and there was a possibility of a need for atrial pacing when uptitrating beta blocker dose. 212

Paddle position The current (2006) ACC/AHA/ESC guidelines for the management of patients with atrial fibrillation recommend that the paddle for external electrical cardioversion be positioned as remotely as possible, preferably in the anterior-posterior configuration. 10) Unlike atrial fibrillation, the feasibility of external cardioversion with an anterior-posterior paddle position for ventricular fibrillation/tachycardia remains unknown. 11) The AHA 2005 guidelines for cardiopulmonary resuscitation do not mention the anteriorposterior paddle position, but recommend lateral (biaxillary) position and posterior-apical position in addition to conventional anterior-lateral position. 12) Although the lateral (biaxillary) position and the posterior-apical position seem to be somewhat safer for the implanted devices than the anterior-lateral position, these positions have not been widely applied, and the posterior-apical position is apparently not practical without the use of an adhesive pad, since interruptions in chest compressions should be minimized. It should be a required design consideration that the pacemaker and pacing lead be made completely resistant to defibrillation. Reducing the problem by better education of primary care providers may be necessary but impractical. Nevertheless, we must increase the awareness that severe hazards are possible when external electrical cardioversion is required in patients with implanted cardiac devices. References 1) Johannes C Manegold, Carsten W Israel, Joachim R Ehrlich, et al: External cardioversion of atrial fibrillation in patients with implanted pacemaker or cardioverterdefibrillator systems: a randomized comparison of monophasic and biphasic shock energy application. Eur Heart J 2007; 28: 1731 1738 2) Lau FYK, Bilitch M, Wintrob HJ: Protection of implanted pacemakers from excessive electrical energy of DC shock. Am J Cardiol 1969; 23: 244 249 3) Waller C, Callies F, Langenfeld H: Adverse effects of direct current cardioversion on cardiac pacemakers and electrodes: is external cardioversion contraindicated in patients with permanent pacing systems? Europace 2004; 6: 165 168 4) Altamura G, Bianconi L, Lo Bianco F, et al: Transthoracic DC shock may represent a serious hazard in pacemaker dependent patients. Pacing Clin Electrophysiol 1995; 18: 194 198 5) Gould L, Patel S, Gomes GI, et al: Pacemaker failure following external defibrillation. Pacing Clin Electrophysiol 1981; 4: 575 577 6) Giedwoyn JO: Pacemaker failure following external defibrillation. Circulation 1971; 44: 293 7) Das G, Eaton J: Pacemaker malfunction following transthoracic countershock. Pacing Clin Electrophysiol 1981; 4: 487 490 8) Aylward P, Blood R, Tonkin A: Complications of defibrillation with permanent pacemaker in situ. Pacing Clin Electrophysiol 1979; 2: 462 9) Gold MR, Shih HT, Herre J, et al: Comparison of defibrillation efficacy and survival associated with right versus left pectoral placement for implantable defibrillators. Am J Cardiol 2007; 100: 243 246 10) ACC/AHA/ESC 2006 Guidelines for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines. (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation) Circulation 2006; 114: e257 e354 11) Kirchhof P, Eckhardt L, Loh P, et al: Anterior posterior versus anterior lateral electrode positions for external cardioversion of atrial fibrillation: a randomised trial. Lancet 2002; 360: 1275 1279 12) 2005 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care, Part 5: Electrical Therapies Automated External Defibrillators, Defibrillation, Cardioversion, and Pacing. Circulation 2005; 112: IV-35 IV-46 213