Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Similar documents
DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Diode calibration for dose determination in total body irradiation

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

Quality assurance in external radiotherapy

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

INTRODUCTION. Material and Methods

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Detector comparison for small field output factor measurements with flattening filter free photon beams

QA for Clinical Dosimetry with Emphasis on Clinical Trials

Small field diode dosimetry

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment

Protection of the contralateral breast during radiation therapy for breast cancer

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn

SRS/SBRT Errors and Causes

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Brachytherapy Planning and Quality Assurance

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

A thesis presented to the: DEPARTMENT OF MEDICAL PHYSICS SCHOOL OF NUCLEAR AND ALLIED SCIENCES UNIVERSITY OF GHANA JOSEPH EDMOND OSEI

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

The determination of timer error and its role in the administration of specified doses

Expectations of Physics Knowledge for Certification

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

A new geometric and mechanical verification device for medical LINACs

IMPLEMENTATION OF IN VIVO DOSIMETRY FOR EXTERNAL PHOTON BEAM RADIOTHERAPY USING GAFCHROMIC EBT 2 FILM A THESIS PRESENTED TO THE:

The Journey of Cyberknife Commissioning

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Eric E. Klein, Ph.D. Chair of TG-142

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Dosisverifikation mit Delta 4 Discover während der Behandlung

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS

Basic dosimetric data for treatment planning system

Patient-specific quality assurance for intracranial cases in robotic radiosurgery system

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Limits of Precision and Accuracy of Radiation Delivery Systems

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

Characteristics Analysis of High Energy External Radiotherapy Beams in Water

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

IMRT QA: Point Dose Measurements or 2D Array?

IORT with mobile linacs: the Italian experience

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Influence of the exposure time in the area monitors at radiodiagnostic

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.

Twelfth Annual Warren K. Sinclair Keynote Address

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Major accidents in radiotherapy

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

Application(s) of Alanine

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

The clinical use of OSLD

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

ABSTRACT I. INTRODUCTION

IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Computed tomography Acceptance testing and dose measurements

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Absolute Dosimetry. Versatile solid and water phantoms. Introduction Introduction Introduction Introtro Intro Intro

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Transcription:

2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe 123, Michael Nyamadi 3, Francis Doughan 3, Evans Sasu 3, Anthony Ashun 3 1 Department of Radiography, School of Biomedical and Allied Health Sciences, College of Health Sciences, University of Ghana, Legon, Accra, Ghana 2 Department of Physics, University of Cape Coast, Cape Coast, Ghana 3 National Centre for Radiotherapy and Nuclear Medicine, Korle Bu Teaching Hospital, Korle Bu, Accra, Ghana ABSTRACT For good treatment outcome in external beam radiotherapy (EBRT), it is imperative to know with great accuracy the amount of radiation dose that will be deposited at any point within the irradiated region of a patient. And because absorbed dose distributions cannot be measured directly in a patient, there is therefore the need to calculate or estimate the dose distributions.the dose calculations are based on dosimetric functions determined or measured in full scatter water phantom, which are used to try to link the doses measured in water to what would be pertain in the patient. Two of these dosimetric functions: percentage depth dose and relative dose factor, had been assessed for an Equinox 100 telecobalt machine and compared with those of "" beam provided by the manufacturer of the teletherapy machine. The measured percentage depth doses were also compared with those of a published beam. The said dosimetric functions were measured to facilitate the commissioning of a treatment planning system for treatment simulation in external beam radiotherapy. The irradiation geometries used for the measurements were based on recommendations of the vendor of the treatment planning system, which were in tandem with what are generally recommended for the measurements of the dosimetric functions. The measured percentage depth doses compared favourably well with those of the published beam than those of the "" beam, and were within ± 6.33% (mean of 1.93 ± 1.67%) and ±17.09% (mean of 4.98 ± 4.12%) when compared with percentage depth doses : the published beam and the "" beam respectively. The differences in the percentage depth doses the published beam relative to those measured for field sizes ranging : 4 x 4 cm 2 to 30 x 30 cm 2 for depths in water up to 20 cm were within ± 2% (recommended tolerance for central axis dosimetry parameter constancy). The differences in the relative dose factor values the "" beam relative to those measured were within ± 1% (mean of 0.49 ± 0.34%). This reiterates the fact that one needs to be circumspective in the use of published and teletherapy machine manufacturer provided beam for clinical applications, though these beam may be used to assess one's measured beam. Keywords: "Silver" Beam Data, Published Beam Data, Percentage Depth Dose and Relative Dose Factor I. INTRODUCTION For favourable treatment outcome in external beam radiotherapy (EBRT), it is imperative to know with great accuracy the amount of radiation dose that will be deposited at any point within the irradiated region of a patient. Placing radiation dosimeters within the patient is not practicable. Dosimetric functions determined or measured in a full scatter water phantoms are used to try to link the doses measured in the water to what will pertaining within the patient [1]. Two most crucial dosimetric functions in use are percentage depth dose (PDD) and relative dose factor (output factor). PDD gives the relative variation of dose with depth within the patient for a particular field size, source-to-surface distance (SSD) and beam energy [1]. It is one of the dosimetric functions which is used extensively for all beam energies applicable to radiotherapy [1]. PDD is also used as beam quality specifier for megavoltage x- ray beams. Other dosimetric functions such as tissuephantom ratios (TPRs ) are usually derived measured PDD [1-3]. TPRs are independent of SSD and are used for the same purpose as PDDs for isocentric treatment technique. PDD is define at the IJSRST1733180 18 April 2017 Accepted: 29 April 2017 March-April-2017 [(2)3: 592-596] 592

percentage ratio of the dose at a required depth in water for a specific field size to that at the depth of maximum dose for the same field size for a particular SSD and beam energy [1,3]. Since it is time consuming to obtained teletherapy machine outputs for all field sizes that may be used clinically, machine output is usually determined for a reference field size and relative dose factors (RDFs) are used to obtain machine output for other field sizes that of the reference field size [3]. RDF gives an indication how machine output for a particular field size relates to that of the reference field size ( usually, 10 cm x 10 cm ) which is used for beam output calibration. RDF is defined as the ratio of the dose at a reference depth (usually the depth of maximum dose) in water for required field size to that of the reference field measured at the same depth and source to detector distance [3]. These beam are obtained during the initial commissioning of a teletherapy machine and are treated as the standard for clinical use and need to be verified periodically by a qualified medical physicist to ensure that the machine parameters have not changed during normal operation [3]. These are input into the treatment planning system (TPS) based on the manufacturer of the TPS requirements to facility simulation of treatment with the TPS prior to treatment delivery. As manufacturers perfect the manufacturing processes for teletherapy machines, there has been an attempt by vendors to standardize machines to have identical beam characteristics [3]. In some cases, "golden" or "" beam sets are provided which contain most or all of the commissioning beam required by the TPS. These beam are acquired by the manufacturer of the teletherapy machine same or similar model of the teletherapy machine at the manufacturer's facility. The one performing the commissioning has the choice of measuring all the required, or verifying a carefully selected subset of the at time of beam commissioning. The preferred option will depend on a number of factors, such as the make and model of the teletherapy machine and TPS, as well as the accuracy required for clinical use [3]. Published beam may be used to assist in arriving at a decision. One needs to be circumspective with the use of any golden or beam within a clinic, as variations in beam parameters have been noted between beams with the same nominal energies [3]. Also, conditions prevailing during the time of the commissioning will not be the same as at the manufacturer's facility. Measuring all the required beam also provide a thorough check of the teletherapy machine, which may unearth problems that may not otherwise be revealed with a mere spot check. However, a golden and or published beam set may be an excellent source of quality assurance for verifying one s commissioning results [3]. Beam measured to determine beam characteristics a teletherapy machine are usually measured with square field sizes, and through the equivalent square field concept beam for other field sizes are obtained [3]. Sometimes owing to inadequacies in shielding requirement for a room housing a teletherapy machine (specifically telecobalt machines), the teletherapy machine is fitted with a beam stopper at the opposite end of the treatment head to attenuate the beam as it transmits through the patient. Some of the beam stoppers are not retractable and their orientation hinder the positioning of a motorized water tanks, which are used to measure and acquire most of the beam required during commissioning process. With reference to this, clinics therefore seek to use alternative means in obtaining the required beam. Also, lack of appropriate dosimetry equipment to facilitate acquisition of PDD compel some radiotherapy centre to rely solely on published PDD for clinical use. This is often the case for telecobalt machines where the beam quality does not change. This is not the case for the relative dose factor (RDF) which is very dependent on collimator design. This paper seeks to assess measured PDD and RDF for an Equinox 100 cobalt 60 teletherapy machine and compare them with those of beam provided by the manufacturer of the teletherapy machine. The PDDs are also compared to published PDD. II. METHODS AND MATERIAL PDD and RDF were measured with Blue Phantom2 motorized water tank with OmniPro-Accept software (IBA dosimetry GmbH, Germany). 0.125 cc cylindrical ionization chamber (TW 31002-1505; PTW Freiburg, Germany) was used in the beam acquisition. During the measurements of the RDFs, the ionization chamber 593

was connected to a UNIDOS electrometer (T10002-20427; PTW Freiburg, Germany), and the electrometer readings corrected for influencing factors (temperature and pressure). The beam under consideration formed part of beam acquired for an Equinox 100 cobalt 60 teletherapy machine (Best Theratronics, Canada) to facilitate the commissioning of a Prowess Panther treatment planning system (Prowess Inc., USA) for treatment planning. The depth dose measurements were done at the beam central axis for depths ranging : 0 to 30 cm for SSD of 100 cm. The depth doses were normalized to that of the depth of maximum dose, Dmax (at 100 %). The PDDs were measured for field sizes ranging : 4 x 4 cm2 to 43 x 43 cm2. The RDFs were determined by measuring outputs at the beam central axis for field sizes ranging : 3 x 3 to 43 x 43 cm2 at depth of Dmax for SSD of 100 cm, and then normalizing to that of the reference field size of 10 x 10 cm2. For all the measurements it was ensured that there is at least 10 cm of water below the ionization chamber to provide the needed backscattered radiation. The irradiation geometries used for the beam acquisition were based on recommendations provided by the vendor of the treatment planning system being commissioned, which were in tandem with what are generally recommended for the measurements of the dosimetric functions. III. RESULTS AND DISCUSSION The measured PDD are listed in table 1 for some selected depths within the phantom and field sizes. Corresponding PDDs obtained supplementary 25 [2] and that "" beam provided by the manufacturer teletherapy machine are also listed in table 1. The differences between the measured PDDs and their corresponding PDDs obtained the published beam and the "" beam respectively, are expressed as percentage of the measured PDDs, and are also listed in table 1. Comparing PDDs the "" beam to their measured counterparts resulted in discrepancies ranging : 0.00% to 17.08% (mean of 4.98%; standard deviation of 4.12%). The PDDs the "" beam were generally lower in value than their measured counterparts. On the other hand, comparing PDDs the published beam to their measured counterparts resulted in discrepancies ranging : -0.22% to 6.32% (mean of 1.93%; standard deviation of 1.67%). The PDDs the published beam were also lower in value than their measured counterparts, with the exception of those for depths ranging 1 to 5 cm for field size of 4 x 4 cm 2. The PDDs the published beam compared favourable well to the measured PDDs than those of the "" beam provided by the manufacturer of the teletherapy machine. Majority of the discrepancies between the measured PDDs and those of the published beam for field sizes ranging : 4 x 4 cm 2 to 30 x 30 cm 2 and for depths ranging 1 to 20 cm were within ± 2% tolerance stipulated for central axis dosimetry parameter constancy [4]. The same could not be said about deviations of the PDDs the "" beam, which are more dispersed. This makes it non-productive task by trying to verify a selected subset of the PDD one wish to use. The large differences in PDD values of the "" beam relative to those measured may be attributed to detector types used in the acquisition of the beam [4]. The "" beam PDDs were measured with a diode (Si P) type detector (PFD3G, Scanditronix/Wellhofer) with thickness of active volume of 0.06 mm and diameter of active area 2 mm [5]. Also differences in source design specifications may contribute to the discrepancies in the PDD values. Although collimator design differences may also contribute to the differences in the PDD values relative to those measured, the significance should have been felt in the discrepancies recorded for the RDF values when compared to those measured, which were within the ± 2% recommended tolerance level set for output constancy[4]. Field size (cm 2 ) 4 x 4 Table 1. Comparison of Percentage depth dose Depth (cm) 1 97.96 98.03 97.50 0.07-0.47 3 87.00 88.65 86.40 1.90-0.69 594

8 x 8 10 x 10 5 75.57 75.40 75.40-0.22-0.22 8 60.01 60.70 60.40 1.15 0.65 10 51.47 52.09 51.70 1.20 0.45 13 40.68 41.21 41.10 1.30 1.03 15 34.66 35.21 35.20 1.59 1.56 18 27.47 27.88 27.90 1.49 1.57 20 23.34 23.84 23.90 2.14 2.40 24 16.97 17.69 17.60 4.24 3.71 26 14.47 14.92 15.00 3.11 3.66 28 12.30 12.80 12.90 4.07 4.88 30 10.44 11.16 11.10 6.90 6.32 1 98.29 98.68 98.30 0.40 0.01 3 88.99 90.62 89.00 1.83 0.01 5 79.23 79.96 79.30 0.92 0.09 8 65.18 66.09 65.20 1.40 0.03 10 56.85 57.63 57.00 1.37 0.26 13 45.95 46.77 46.00 1.78 0.11 15 39.60 40.54 39.70 2.37 0.25 18 31.89 32.64 32.10 2.35 0.66 20 27.37 28.20 27.60 3.03 0.84 24 20.32 21.11 20.70 3.89 1.87 26 17.53 18.26 18.00 4.16 2.68 28 15.10 15.81 15.70 4.70 3.97 30 12.94 13.62 13.50 5.26 4.33 1 98.27 98.65 98.50 0.39 0.23 3 89.67 90.81 89.70 1.27 0.03 5 79.99 80.67 80.40 0.85 0.51 8 66.66 67.63 66.80 1.46 0.21 10 58.36 59.39 58.70 1.76 0.58 13 47.53 48.55 47.90 2.15 0.78 15 41.47 42.51 41.60 2.51 0.31 18 33.47 34.32 33.80 2.54 0.99 20 29.02 29.97 29.30 3.27 0.96 24 21.63 22.38 22.20 3.47 2.64 26 18.71 19.73 19.30 5.45 3.15 28 16.23 17.03 16.80 4.93 3.51 30 13.90 14.79 14.60 6.40 5.04 Table 1. Comparison of Percentage depth dose (continuation ). 15 x 15 30 x 30 43 x 43 1 97.94 98.74 98.6 0.82 0.67 3 90.01 92.12 90.5 2.34 0.54 5 81.26 82.9 82 2.02 0.91 8 68.63 70.73 69.3 3.06 0.98 10 61.03 62.9 61.6 3.06 0.93 13 50.68 52.9 51.00 4.38 0.63 15 44.39 46.38 44.9 4.48 1.15 18 36.40 38.27 37.00 5.14 1.65 20 31.90 33.76 32.40 5.83 1.57 24 24.33 26.03 24.90 6.99 2.34 26 21.18 22.71 21.90 7.22 3.40 28 18.47 19.99 19.30 8.23 4.49 30 15.99 17.46 16.90 9.19 5.69 1 98.38 98.83 98.70 0.46 0.33 3 91.28 93.49 91.60 2.42 0.35 5 83.32 85.57 83.90 2.70 0.70 8 72.13 75.08 72.50 4.09 0.51 10 64.92 68.30 65.20 5.21 0.43 13 55.00 58.57 55.60 6.49 1.09 15 49.09 52.65 49.70 7.25 1.24 18 41.45 44.83 41.90 8.15 1.09 20 36.77 40.12 37.30 9.11 1.44 24 28.92 32.07 29.70 10.89 2.70 26 25.61 28.54 26.40 11.44 3.08 28 22.72 25.51 23.60 12.28 3.87 30 20.07 22.52 20.90 12.21 4.14 1 96.95 98.88 99.02 1.99 2.14 3 90.36 94.40 91.98 4.47 1.79 5 82.49 87.13 84.64 5.62 2.61 8 71.79 76.73 73.60 6.88 2.52 10 65.17 70.13 66.60 7.61 2.19 13 55.43 60.90 57.13 9.87 3.07 15 49.83 54.98 51.40 10.34 3.15 18 42.32 47.53 43.67 12.31 3.19 20 38.05 42.74 39.14 12.33 2.86 24 30.23 34.53 31.44 14.22 4.00 26 26.76 30.76 28.08 14.95 4.93 28 23.78 27.62 25.08 16.15 5.47 30 21.14 24.75 22.25 17.08 5.25 Field size (cm 2 ) Depth (cm) Graphical representation of the PDDs as a function of depth in water is depicted in figure 1 for field sizes of 4 x 4 cm 2 and 43 x 43 cm 2, representing the field sizes 595

PDD PDD (cm) where the maximum deviations the measured PDDs were recorded for comparisons with the published beam and the "" beam respectively. The same axis was used for all the plots. The PDDs for a particular field size followed similar trend. For the specified field sizes, the maximum deviations occurred at a depth of 30 cm. 4 x 4 cm^2 IV.CONCLUSION This shows that it is not prudent for one to solely rely on published and teletherapy machine manufacturer provided beam for clinical applications, though these beam may be used to assess one's measured beam. V. REFERENCES depth in water (cm) 43 x 43 cm^2 PDD PDD depth in water (cm) [1]. E.B. Podgorsak. Radiation oncology physics: A handbook for teachers and students. International Atomic Energy Agency, Vienna. 2005, Pp. 161-216. [2]. British Journal Of Radiology, Central Axis Depth Dose Data for Use in Radiotherapy, Suppl. 25 (1996). Pp. 58-59, 153-157. [3]. Das I J, Cheng C, Watts R J, Ahnesjö A. et al. Accelerator beam commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM. Med. Phys. 2008; 35 (9): 4186-4211. [4]. Kutcher G J, Coia L, Gillin M, Hanson W F.et al. Comprehensive QA for radiation oncology: Report of AAPM radiation therapy committee task group 40 Med. Phys. 1994;21 (4): 581-618. [5]. MDS Nordion Inc. Theratron equinox eternal beam therapy system: Beam summary sheet. MDS Nordion Inc., Canada. 2006, Pp. 1-4. Figure 1. Variation of percentage depth dose as a function of depth in water for field sizes of 4 x 4 cm 2 and 43 x 43 cm 2. The measured RDFs are listed in table 2 together with their counterpart the "" beam for the various field sizes. Also shown in table 2 are the discrepancies between the measured RDFs and those of the "" beam, which are expressed as percentage of the measured RDFs for the respective field sizes. The percentage deviations between the measured RDFs and those obtained the "" beam were within ± 1% (mean of 0.49% and standard deviation of 0.34%); 596