Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)

Similar documents
Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations

Basic dosimetric data for treatment planning system

Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Limits of Precision and Accuracy of Radiation Delivery Systems

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

Small field diode dosimetry

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Eric E. Klein, Ph.D. Chair of TG-142

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

The Journey of Cyberknife Commissioning

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Study of Motion-induced Dose Error Caused by Irregular Tumor Motion in Helical Tomotherapy

PRECISE DOSIMETRY OF SMALL PHOTON BEAMS COLLIMATED BY AN ADD-ON DYNAMIC MICRO-MULTILEAVES COLLIMATOR.

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Unrivaled, End-to-End

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Detector comparison for small field output factor measurements with flattening filter free photon beams

Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

Quality Assurance of Helical Tomotherapy Machines

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Nuclear Associates

IMRT QA: Point Dose Measurements or 2D Array?

Amendment No. 2. Item No. 2 (Rfx/ Event number )

A new geometric and mechanical verification device for medical LINACs

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery

ph fax

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

I. Equipments for external beam radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM

Mitsubishi Heavy Industries Technical Review Vol. 51 No. 1 (March 2014)

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Neutron Measurements for Intensity Modulated Radiation Therapy

Electron Beam ET - Reloaded Therapy - Reloaded

Quality assurance in external radiotherapy

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Radiochromic film dosimetry in water phantoms

Verification of micro-beam irradiation

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Additional Questions for Review 2D & 3D

A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

IMRT Planning Basics AAMD Student Webinar

Debate Motion. Theoretical benefits of VMAT: Real Costs of VMAT. "Arc Based Techniques Will Make Conventional IMRT Obsolete

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors

Comparison of stereotactic plans for brain tumors with two different multileaf collimating systems

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Trajectory Modulated Arc Therapy: Application to Partial Breast Irradiation. Research and development to advance radiotherapy

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

Technology Challenges of Commercial Medical Electron Accelerators

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

Chapters from Clinical Oncology

Case Study. Institution Farrer Park Hospital

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

EPID dosimetry for pretreatment quality assurance with two commercial systems

Elekta Synergy Digital accelerator for advanced IGRT

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

Dosimetry of the Gamma Knife TM

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe

Follow this and additional works at:

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Dosimetric characterization of surface applicators for use with the Xoft ebx system

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Online Care, Online Control, Online Confidence

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Transcription:

Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 4, December, 2015 http://dx.doi.org/10.14316/pmp.2015.26.4.223 Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC) Ae Ran Kim*, Jae-Hyuk Seo*, Hun-Joo Shin*, Hyeong Wook Park, Ki Woong Lee, Jae Choon Lee, Shin-Wook Kim*, Ji Na Kim*, Hyeli Park, Heui-Kwan Lee, Young-Nam Kang *Department of Biomedical Engineering, College of Medicine, The Catholic University of Korea, Seoul, Department of Radiation Oncology, Presbyterian Medical Center of Korea, Jeonju, Department of Radiation Oncology, Bucheon St. Mary's Hospital, The Catholic University of Korea, Bucheon, Department of Radiation Oncology, Incheon St. Mary's Hospital, The Catholic University of Korea, Incheon, Department of Medical Physics, Kyonggi University of Korea, Suwon, Department of Radiation Oncology, Seoul St. Mary's Hospital, The Catholic University of Korea, Seoul, Korea Double-focused micro-multileaf Collimator (μmlc) is able to create radiation fields having sharper dose gradients at the field edges than common MLC. Therefore, μmlc has been used for the stereotactic radiosurgery (SRS) and Stereotactic Radiotherapy (SRT). We evaluated the dosimetric characteristics of a doublefocused Dynamic-μMLC (DMLC) attached to the Elekta Synergy linear accelerator. For this study, the dosimetric parameters including, Percent Depth Dose (PDD), Leaf leakage and penumbra, have been measured by using of the radiochromic films (GafChromic EBT2), EDGE diode detector and three-dimensional water phantom. All datas were measured on 6 MV x-ray. As a result, The DMLC shows transmission below to 1% and because of double-focused construction of the DMLC, the penumbras of fields with DMLC are independent from the field sizes. In this paper, the resulting dosimetric evaluations proved the applicability of the DMLC attached to the Elekta Synergy linear accelerator. Key Words: Dynamic micro-multileaf Collimator, PDD, Leaf transmission, Penumbra Introduction A Multileaf collimator (MLC) has been clinically used for both conventional and irregular radiation fields shaping. Due to individually movable leaves fields of high complexity can be generated. The MLC was integrated in almost every modern linac. The commonly used leaf width of MLC is 1 1.25 cm. Stereotactic Radio-Surgery (SRS) and Stereotactic Radiotherapy This work was supported by the fusion research program of Korea research council for industrial science and technology (B551179-12- 08-00, Development of Convergent Radio Therapy Equipment with O-arm CT) funded by the Ministry of Trade, Industry & Energy (MOTIE, Korea). Received 1 December 2015, Revised 28 December 2015, Accepted 28 December 2015 Correspondence: Young-Nam Kang (ynkang33@gmail.com) Tel: 82-2-2258-6849, Fax: 82-2-2258-1532 cc This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. (SRT) are concerned to treat small lesions with irregular field shape. The smaller MLC dimensions, the more useful for treatment with small radiation fields. A variation of this device is the so-called micro-mlc (μmlc). Nowadays, the leaf width of the μmlc is less than 5 mm. 1,2) The μmlc designs can be classified into two groups, single-focused and double focused. The leaf ends as well as the leaf sides match the beam divergence, making the configuration double-focused. The advantage of such a precise collimation is the independence of the penumbra from the location of the field. 3) This focusing with respect to dose distributions have been presented by Meeks et al. 4) Therefore, the double-focused μmlc is a very useful collimation system for SRS and SRT. 5) In this study, we evaluated the dosimetric parameters including, Percentage Depth Dose (PDD), Interleaf leakage, Intra-leaf transmission, leaf-end transmission and penumbra of a double-focused Dynamic-μMLC (DMLC, 3D Line, USA Inc.). It was attached to the Elekta Synergy Linear accelerator. All data were measured on 6 MV x-ray. - 223 -

Ae Ran Kim, et al:evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC) Materials and Methods 1. DMLC system The Dynamic Micro-multileaf collimator (DMLC) provided by 3D Line is part of the DynART (Dynamic Arc Radiation Therapy) system. The DMLC is double-focusing and has 24 pairs of tungsten leaves that each project 2.9 mm width at the isocenter. Therefore, the maximum field size is 68.3 69.7 mm 2. The DMLC was attached to the Elekta Synergy linear accelerator (Fig. 1). The DMLC is coupled to the linac (linear accelerator) at 624 mm from the source (Fig. 2(a)). 6) Leaves describe a convex path, so the photon beam will be focused both in the motion direction and in the transverse direction. In order to reduce transmission through the leaves, DMLC are slightly divergent that leads rather to the so-called triangular tongue and groove effect (Fig. 2(b)). 6,7) When the gantry was positioned at 180 o, the DMLC can be manually attached to the accessory mount of the linac by the help of a specially designed trolley. The electrical connection between the DMLC and gantry of the linac is based on a single cable, which can be connected very easily. The physical characteristics of this DMLC and BrainLAB m3 are described in Table 1. 2. Percent depth dose (PDD) The measurements were obtained using a EDGE Detector TM (Sun Nuclear Corporation, Melbourne, FL) and a Blue Phantom 2 three-dimensional scanning system (IBA Dosimetry GmbH). The central axis PDDs were measured in water for square field sizes, 3.6 3.6 cm 2 and 6.0 6.0 cm 2 and Source to Surface Distance (SSD) at 100 cm. The measured datas were normalized to the maximum depth dose of 6 MV x-ray. 3. Leaf leakage We measured interleaf leakage, intra-leaf and leaf-end transmission. 3) To measure interleaf leakage and intra-leaf transmission, all leaves of the DMLC were closed. Due to the Fig. 1. Dynamic micro-multileaf collimator (DMLC) attached to the Elekta Synergy linear accelerator. The electrical connection between the DMLC and the gantry is established by only on cable. Fig. 2. (a) Scheme of the transverse view of the DMLC. (b) Detail of the leaves of the DMLC. All distance are in mm. - 224 -

PROGRESS in MEDICAL PHYSICS Vol. 26, No. 4, December, 2015 Table 1. Physical characteristics of the 3D Line DMLC and BrainLAB m3. 3D Line DMLC BrainLAB m3 Number of leaves 48 (24 pairs) 52 (26 pairs) Leaf Material Tungsten (93%) Coated tungsten Leaf Height 80 mm - Real leaf width 1.9 mm - Leaf width at isocenter 2.9 mm 14 pairs at 3 cm 6 pairs at 4.5 cm 6 pairs at 5.5 cm Maximum field size at isocenter 68.3 69.7 mm 2 98 98 mm 2 Weight (approx.) 30 kg 30 kg Design Double-Focused - Leaf speed 1.0 cm/s 1.5 cm/s Fig. 3. Percent depth dose of (a) field size 6 6 cm 2, (b) 3.6 3.6 cm 2 for 6 MV. small amount of collimator transmission, the linac jaws were opened as 7 7 cm 2. The radiochromic films (GafChromic EBT2) were used for the measurement. The films were positioned perpendicularly to the beam central axis in the solid water phantom at a depth 5 cm and SAD 100 cm. The films were exposed to 6 MV x-ray for 6,000 monitor units. Also, to confirm the leaf-abutting setup, the five sequential field shapes were used. The shape of each field was closed and the abutment location was translated by 1.5 cm steps over the course of the sequence. 8) Each exposure was for 1,200 MU. We got the vertical profiles for first film and cross-line profiles for second film. The films were scanned with a Epson Perpection V1000 scanner. The software we used to analyze the scanned 48-bit RGB tiff images of the EBT2 films was based on MATLAB R2015a. 4. Penumbras and field sizes The measurements of Penumbras were made at the depth of 1.5 and 5 cm, SSD 100 cm. Because the leaves are designed to match the beam divergence, the penumbra should be independent from the field sizes. To confirm this, penumbra datas were obtained. We measured beam profiles of 3.6 3.6 cm 2, 4.8 4.8 cm 2, 6.0 6.0 cm 2 field sizes. Results and Discussions 1. Percent depth dose (PDD) As shown in Fig. 3, measured PDD datas with or without - 225 -

Ae Ran Kim, et al:evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC) the DMLC were similar. The DMLC does not significantly affect the depth-dose characteristics of the linear accelerator. 2. Leaf leakage Fig. 4 shows the leaf leakage of the DMLC. The maximum interleaf leakage between the leaf sides was determined as 0.94%, where the maximum transmission through the leaves 0.79%. The transmission slightly increased more than 4.7 mm width μmlc. 2) We considered the scatter of interleaf uptake as short as the leaf width. The maximum transmission between leaf-ends (i.e., where the leaves from opposite sites met at the central axis) was determined to be 1.37%. The transmission between leaf-ends is dependent on leaf alignment. Table 2 9) shows the comparison of transmission and leakage between DMLC and commercial μmlc (m3; BrainLAB, Heimstetten, Germany). Currents measurements are in agreement with the AAPM TG 50 recommendation which recommends 2% as a maximum value for transmission. As shown in Fig. 5, we could check the leaf-end leakage peaks at 3.0 cm, 1.5 cm, 0 cm, 1.5 cm and 3.0 cm. These peaks shown the stability of the abutted leaf setup. 3. Penumbras and field sizes As shown in Table 3, The 80 20% beam penumbras average over all field size were 2.66 mm along the leaf end. These values are in good agreement with measurements of BrainLAB m3, which are 2.31 mm. 9) The measured width of penumbras were not significantly different compared to the linac jaw and independent from the field size. Table 2. Transmission & Leakage of 3D Line DMLC and BrainLAB m3. Transmission (%) Leakage (%) m3-mlc 0.93 1.18 3D Line DMLC (This work) 0.79 0.94 Fig. 4. In-plane profile about interleaf leakage and intra-leaf transmission. The field closed at isocenter. Fig. 5. Cross-plane profile about abutted leaf setup field. The position of 5 peaks shown the abutted leaf position. Table 3. Penumbras (mm) for 3.6 3.6, 4.8 4.8, 6.0 6.0 cm 2 field sizes defined using the DMLC and Linac jaws. The datas were measured at SSD 100 cm at depths of 1.5 cm and 5 cm in water using a EDGE Detector. Field Size 80 20% d=1.5 cm 80 20% d=5.0 cm 3.6 3.6 cm 2 4.8 4.8 cm 2 6.0 6.0 cm 2 DMLC Linac jaws DMLC Linac jaws DMLC Linac jaws 2.6 2.8 2.7 2.9 2.7 2.9 3.0 3.0 3.1 3.2 3.1 3.1-226 -

PROGRESS in MEDICAL PHYSICS Vol. 26, No. 4, December, 2015 Conclusions We evaluated the dosimetric characteristics of DMLC attached to the Elekta Synergy linear accelerator. PDDs measured by the DMLC is equal to a depth-dose curves of the original linac. That means that the DMLC affects only the offaxis ratio (OAR). Due to the double-focused construction of DMLC, the penumbras of fields with DMLC are independent from the field sizes. The DMLC shows transmission below to 1%. Because of this minimal transmission, the linac jaws can be fixed while the DMLC is moving. It enhances the sparing of normal tissue by minimizing unwanted dose exposure. In this paper, the dosimetric investigations proved the applicability of the DMLC attached to the Elekta Synergy linear accelerator. References 1. Bortfeld T, Oelfke U, Nill S, What is the optimum leaf width of a multileaf collimator? Med Phys 27(11):2494-502 (2000) 2. Fischer M, Todorovic M, Drud E, and Cremers F, Commissioning of a double-focused micro multileaf collimator (μmlc). J Appl Clin Med Phys 11(2):81 91 (2010) 3. AAPM Report No. 72: Basic Applications of Multileaf Collimators, Report of Task Group 50 (2001) 4. Meeks SL, Bova FJ, Kim S, Tome WA, and Buatti JM, Dosimetric characteristics of a double-focused miniature multileaf collimator. Med Phys 26(5):729 733 (1999) 5. Galal MM, Keogh S, and Khalil S, Dosimetric and mechanical characteristics of a commercial dynamic μmlc used in SRS Med Phys 38(7):4225 (2011) 6. Gonzalez W, Lallena AM, Alfonso R, Monte Carlo simulation of the dynamic micro-multileaf collimator of a LINAC Elekta Precise using PENELOPE. Phys Med Biol 56(11):3417-3431 (2011) 7. Yu CX, Design considerations for the sides of multileaf collimator leaves. Phys Med Biol 43:1335-1342 (1998) 8. Bayouth JE, Morrill SM, MLC dosimetric characteristics for small field and IMRT applications, Med Phys 30(9) :2545-52 (2003) 9. Garcia-Garduno OA, Celis MA, Manuel J, et al., Radiation transmission, leakage and beam penumbra measurements of a micro-multileaf collimator using GafChromic EBT Film. J Appl Clin Med Phys 9(3):2802 (2008) - 227 -

Ae Ran Kim, et al:evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC) 이중으로집중된동적미세다엽콜리메이터의선량학적특성평가 * 가톨릭대학교의과대학의공학교실, 전주예수병원방사선종양학과, 가톨릭대학교부천성모병원방사선종양학과, 가톨릭대학교인천성모병원방사선종양학과, 경기대학교의학물리학과, 가톨릭대학교서울성모병원방사선종양학과 김애란 * ㆍ서재혁 * ㆍ신헌주 * ㆍ박형욱 ㆍ이기웅 ㆍ이재춘 김신욱 * ㆍ김지나 * ㆍ박혜리 ㆍ이희관 ㆍ강영남 이중으로집중된미세다엽콜리메이터 (Double-focused micro Multileaf Collimator: μmlc) 는보통의다엽콜리메이터 (Multileaf Collimator: MLC) 에비하여조사면가장자리부분의선량을급격하게줄여준다. 이러한특성때문에, 미세다엽콜리메이터는정위적방사선수술과치료 (Stereotactic Radio-Surgery/RadioTtherapy, SRS/SRT) 에사용되어왔다. 우리는 Elekta Synergy 선형가속기에이중으로집중된동적미세다엽콜리메이터 (Double-focused Dynamic micro-multileaf Collimator: DMLC) 를부착하여선량학적특성을평가하였다. 본연구에서는, 필름 (GafChromic EBT2 film), EDGE 다이오드검출기, 3차원물팬텀을이용하였다. 깊이선량백분율 (Percent Depth Dose, PDD), 엽투과도 (leaf leakage), 반그림자 (Penumbra) 를측정하였고, 모든데이터들은 6MV 광자선으로측정하였다. 그결과, DMLC가 1% 이내의투과도를갖는것을확인할수있었다. DMLC는이중으로집중되는구조를가졌기때문에반그림자가조사야크기에대하여독립적인것을확인하였다. 본연구에서는 DMLC의선량학적특성을바탕으로, Elekta Synergy에부착된 DMLC의적용가능성을증명하였다. 중심단어 : 동적미세다엽콜리메이터, 깊이선량백분율, 엽투과도, 반그림자 - 228 -