Comparison of Energy Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams using Magnetic Resonance Imaging

Similar documents
Comparison of Percentage Depth Dose (PDD) and Post Time Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams Using MRI Technique

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Optimization of the T2 parametric image map calculation in MRI polymer gel dosimetry

A feasibility study of multislice X-ray CT imaging of gel dosimeters using the zero scan method

RadioTherapy. 80 wt% 90 wt% 3wt% 5wt% 3 wt% 10 wt% 1

Evaluation of the basic properties of the BANGkit gel dosimeter

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

3D MRI gel dosimetry based on image intensity (A new approach)

The Development of Normoxic Polymer Gel Dosimetry using High Resolution MRI

Australian Journal of Basic and Applied Sciences. Evaluation of CT based MAGAT gel dosimetry system for measuring 3D dose distribution

3D Dosimetry with Polymer Gel

X-ray computed tomography imaging of polymer gel dosimeters

NMR spectroscopy of saccharide-doped PAGAT dosimeters

Historical overview of the development of gel dosimetry: a personal perspective

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

How to scan polymer gels with MRI?

Preliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter

Formaldehyde increases MAGIC gel dosimeter melting point and sensitivity

Evaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

The use of the BANG-3 polymer gel to quantify the three-dimensional dose distribution of IMRT

Clinical Applications of Gel Dosimeters

Diode calibration for dose determination in total body irradiation

High resolution gel-dosimetry by optical-ct and MR scanning

Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance


Radiotherapy and Oncology 48 (1998) Received 10 July 1997; revised version received 15 June 1998; accepted 14 August 1998

Assessment of X-ray Computed Tomography Dose in Normoxic Polyacrylamide Gel Dosimetry. Patricia Lynn Baxter. Masters of Science

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy

Fundamentals of Polymer Gel Dosimeters

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

Radiochromic 3D Detectors

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Radiochromic film dosimetry in water phantoms

High Resolution Polymer Gel Dosimetry for Small and Micro Field Dosimetry, and Development of Innovative Polymer Gel Dosimeters

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital


Three Dimensional Dosimetry. Session Program: Therapy Education Session Organizer: L. John Schreiner

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

Y FILMS DOSIMETR Nederland België / Belgique

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

Small field dose delivery evaluations using cone beam optical computed tomography-based polymer gel dosimetry

"Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 : Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY *

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Gel dosimetry: The effect of gold nanoparticles on the dose enhancement in the external radiation therapy

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose

VIP. MRI magnetic resonance imaging. Keywords: PHITS, gel dosimeter, three dimensional dose distribution, LET dependence, dose verification MRI 2)

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

True 3D chemical dosimetry (gels, plastics): Development and clinical role

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD)

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Dosimetry Standards and Dissemination Systems for Radiation Processing in China

Verification of micro-beam irradiation

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

Calibration of two 90 Sr+ 90 Y dermatological applicators

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

A practical three-dimensional dosimetry system for radiation therapy

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Dosimetric Analysis Report

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

The determination of timer error and its role in the administration of specified doses

ROLE OF ANTI-OXIDANTS ON NORMOXIC METHACRYLIC ACID GELATIN (MAG) POLYMER GEL DOSIMETER AT 6-MV PHOTON BEAM USING SINGLE SPIN ECHO MRI

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics

Characterization of the effect of MRI on Gafchromic film dosimetry

Patient dosimetry for total body irradiation using single-use MOSFET detectors

Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

9.7 Dosimetry Study on A p-n Junction Semiconductor Detector

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Cochlear Implants. Medical Procedures. for MED EL Implant Systems. AW33290_2.0 (English US)

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison

MODELING AND DEVELOPMENT OF THREE-DIMENSIONAL GEL DOSIMETERS

Performance of several active personal dosemeters in interventional radiology and cardiology

Transcription:

Research Journal of Applied Sciences, Engineering and Technology (): 6-, ISSN: -767 Maxwell Scientific Organizational, Submitted: November 9, Accepted: November 9, Published: February, Comparison of Energy Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams using Magnetic Resonance Imaging B. Azadbakht and K. Adinehvand Department of Engineering, Borujerd Branch, Islamic Azad University, Borujerd, Iran Abstract: The purpose of this study was to evaluate dependence of PAGAT polymer gel dosimeter /T on different electron and photon energies for a standard clinically used 6 Co therapy unit and an electa linear accelerator.using MRI, the formulation to give the maximum change in the transverse relaxation rate R(/T) was determined to be.% N,N'-methylen-bis-acrylamide(bis),.% acrylamid(aa), % gelatine, mm tetrakis (hydroxymethyl) phosphonium chloride (THPC),. mm hydroquinone (HQ) and 86% HPLC(Water). When the preparation of final polymer gel solution is completed, it is transferred into phantoms and allowed to set by storage in a refrigerator at about C. The optimal post-manufacture irradiation and post imaging times were both determined to be h. The sensitivity of PAGAT polymer gel dosimeter with irradiation of photon and electron beams was represented by the slope of calibration curve in the linear region measured for each modality. The response of PAGAT gel with photon and electron beams is very similar in the lower dose region. The R-dose response was linear up to Gy and the R-dose response of the PAGAT polymer gel dosimeter is linear between to Gy. In electron beams the R-dose response for doses less than Gy is not exact, but in photon beams the R-dose response for doses less than Gy is not exact. Dosimeter energy dependence was studied for electron energies of, and 8MeV and photon energies of.,, 6 and 8 MV. Evaluation of dosimeters were performed on Siemens Symphony, Germany.T Scanner in the head coil. In this study no trend in polymer-gel dosimeter /T dependence was found on mean energy for electron and photon beams. Key words: Electron and photon beams, MRI, PAGAT gel, polymer gels INTRODUCTION In 98, magnetic resonance imaging (MRI) demonstrated great potential in visulazing three dimensional (D) dose distributions of ferric or ferrous sulphate gel dosimeters (Gore et al., 98). Subsequently, studies were undertaken to investigate the feasibility of using Ferric gel as a D dosimetry system in radiation oncology (Olsson et al., 99). The major limitation in Ferric gel dosimetry is that it suffers from blurness of dose with time which is due to the migration of ferrous and ferric ions in gel matrix, known as diffusion (Schulz and Gore, 99). In 99, a polymer gel dosimeter was developed that maintained spatial information following irradiation which could be visualized using MRI (Maryanski et al., 99). Polymer gel dosimetry is a technique that has the ability to map absorbed radiation dose distribution in three dimensions (D) with high spatial resolution. Polymer gel dosimeters offer a number of advantages over traditional dosimeters such as ionization chambers, Thermoluminescent Dosimeter (TLD) and radiographic film. These advantages include independence of radiation direction, radiological soft tissue equivalence, integration of dose for a number of sequential treatment fields, and perhaps most significantly, evaluation of a complete volume at once (De Deene, ) and (Brindha et al., ). In, the first normoxic gels were suggested that could be produced, stored and irradiated in a normal condition. Magnetic Resonance Imaging (MRI) has been most extensively used for the evaluation of absorbed dose distributions in polymer gel dosimeters. In the MRI evaluation of polymer gel dosimeters, changes in T is a result of physical density changes of irradiated polymer gel dosimeters. Many factors such as polymer gel composition, temperature variation during irradiation, type and energy of radiation, dose rate, temperature during MRI evaluation, time between irradiation to MRI evaluation, and strength of magnetic field have been studied by different authors (Maryanski et al., 99); Maryanski et al., 996; Kennan et al., 996). All these factors can potentially affect polymer gel dosimeter response and significantly influence measured results. Consequently, it is important to evaluate and quantify each individual factor. In PAGAT polymer gel dosimeter, the gel itself forms both a multi dimensional phantom and Corresponding Author: B. Azadbakht, Department of Engineering, Borujerd Branch, Islamic Azad University, Borujerd, Iran 6

Res. J. Appl. Sci. Eng. Technol., (): 6-, Table : Different chemicals and percent weight of PAGAT gel Component Percent weight Gelatine ( bloom) % N,N'-methylen-bis- acrylamide (bis).% Acrylamide (AA).% Tetrakis-phosphonium chloride (THPC) m M Hydroquinone (HQ). mm HPL C (water) 86% Table : The protocol of magnetic resonance imaging (MRI) Parameters Field of view (FOV) (mm) 6 Marrix size (MS) Slice thickness (d) (mm) Repetition time (TR) (ms) Echo time (TE) (ms) Number of slices,,, Number of echoes Total measurement time (min) - Resolution (mn). Band with (Hz/Pixel) the detector (Venning et al., ). The gel can be modified to be almost completely soft-tissue equivalent. Considering factors such as accuracy, sensitivity, the time needed for dosimetry, three-dimensional capabilities, energy independence, dose rate independence, and costs, we believe that PAGAT polymer gel dosimeter is the closest to ideal dosimetry method comparing with TLDs, ion chambers, film dosimetry, Fricke gels and anoxic gels. MATERIALS AND METHODS Preparation of PAGAT gel: The PAGAT polymer gel formulation by % mass consisted of.% N,N'-methylenbis-acrylamide (bis),.% acrylamid (AA), % gelatine, mm tetrakis (hydroxymethyl) phosphonium chloride (THPC),. mm hydroquinone (HQ) and 86% HPLC (Water) (6). All components were mixed on the bench top under a fume hood. The gelatine was added to the ultra-pure de-ionized water and left to soak for min, followed by heating to 8ºC using an electrical heating plate controlled by a thermostat. Once the gelatine completely dissolved the heat was turned off and the cross-linking agent, bis was added and stirred untile dissolved. Once the bis was completely dissolved the AA was added and stirred untile dissolved. Using pipettes, various concentration of the polymerization inhibitor HQ and the THPC anti-oxidant were combined with the polymer gel solution. When the preparation of final polymer gel solution is completed, it is transferred into phantoms and allowed to set by storage in a refrigerator at about ºC (Venning et al., ). Table lists the component with different percent weight in normoxic PAGAT polymer gel dosimeter. Irradiation: Irradiation of vials was performed using electron beams by an electa linear accelerator with SSD R-dose (/sec) 6 6 6 Fig. : Sensitivity of PAGAT gel with different range of doses (-6Gy) = 8 cm, field size of cm and the depth was selected at cm. The optimal post-manufacture irradiation was determined to be day. Imaging: Before imaging, all polymer gel dosimeters were transferred to a temperature controlled MRI scanning room to equilibrate to room temperature. The PAGAT polymer gel dosimeters were imaged in a Siemens Symphony. Tesla clinical MRI scanner using a head coil. T weighted imaging was performed using a standard Siemens -echo pulse sequence with TE of ms, TR of ms, slice thickness of mm, FOV of 6 mm. The optimal post imaging times was determined to be day. The images were transferred to a personnel computer where T and R maps were computed using modified radiotherapy gel dosimetry image processing software coded in MATLAB (The Math Works, Inc).The mean T value of each vial was plotted as a function of dose with the quasi-linear section being evaluated for R- dose sensitivity. Table lists the protocol of Magnetic Resonance Imaging (MRI) was used in PAGAT polymer gel dosimeter. RESULTS Verification of R-dose sensitivity of PAGAT polymer gel dosimeter: Verification of R-dose sensitivity of PAGAT polymer gel dosimeter with electron beams: PAGAT gels with optimum value of ingredient was manufactured and irradiated to different doses. Polymer gel dosimeters in Perspex phantoms were homogeneously irradiated with 6MeV electron beam with an electa linear accelerator located in Tehran. Delivered doses were from -6cGy. The calibration curve (transverse relaxation rate (/T) versus applied absorbed dose) was obtained and plotted. Dependence of /T response to the absorbed dose in the range of -6cGy is shown is Fig.. As it can be seen in Fig., PAGAT has a linear response up to Gy. the response of the PAGAT gel is very similar in the lower 7

Res. J. Appl. Sci. Eng. Technol., (): 6-, Table : Sensitivity of PAGAT gel with different range of doses R- Dose sensitivity (SG GyG ) ) Correlation coefficient - -.6.997 -.89.9 -.6.99-6.9.997 Table : Sensitivity of PAGAT with different range of doses R- Dose sensitivity (SG GYG ) Correlation coefficient -.6.67 -..999 -.98.999 -.77.98 R (/sec) 7 6 Fig. : PAGAT polymer gel dosimeter R (/T) calibration response in range of -Gy dose region and The R-dose response for doses less than Gy is not exact. The R-dose response of the PAGAT polymer gel dosimeter is linear between - Gy and - Gy. Figure shows that PAGAT polymer gel has a dynamic range of at least.66 for doses up to Gy. Table lists the Sensitivity of PAGAT polymer gel dosimeter with different range of doses. Verification of R-dose sensitivity of PAGAT polymer gel dosimeter with photon beams: Polymer gel dosimeters in Perspex phantoms were homogeneously irradiated with. MV photon beam with a Co-6 therapy unit located in Tehran. Delivered doses were from - cgy. The calibration curve (transverse relaxation rate (/T) versus applied absorbed dose) was obtained and plotted. Dependence of /T response to the absorbed dose in the range of - cgy is shown is Fig.. As it can be seen in Fig., PAGAT has a linear response up to Gy. the response of the PAGAT gel is very similar in the lower dose region and The R-dose response for doses less than Gy is not exact. The R-dose response of the PAGAT polymer gel dosimeter is linear between - Gy and - Gy. Figure shows that PAGAT polymer gel has a dynamic range of at least.88 SG for doses up to Gy compared with less than. SG for doses up to Gy in the preliminary study of the PAGAT polymer gel dosimeter (De Deene, ). Table lists the Sensitivity of PAGAT polymer gel dosimeter with different range of doses. R -dose (/sec) Slop........8.7.6.. MEV MEV 8MEV Y =.6x+.7 R =.986 (MeV) Y =.6x+.67 R =.987 (MeV) Y =.8x+.696 R =.986 (8MeV) Fig. : Verification response of PAGAT polymer gel dosimeter with different energies (e.g, and 8 MeV) Y = 8E-.x+. R =.976. Fig. : Sensitivity of PAGAT polymer gel dosimeter with different energies (e.g,, and 8 MeV) Verification of R-dose response of PAGAT gel dosimeter with different energies: Verification of R-dose response of PAGAT gel dosimeter with different electron energies: Response of the energy dependence was obtained for, and 8 MeV nominal electron energy and is shown in Fig.. a linear function is fitted to each set of data and their slope and correlation were found. As it can be seen from Fig., dependence of PAGAT polymer gel response to the energy of electron is shown. As the energy increases from to 8 MeV the sensitivity is nearly stable:. 8. 6 % 8. which is % variation energy. Due to the large proportion of water (86%), the PAGAT polymer gel is nearly soft-tissue equivalent and no energy corrections are required for electron beams in radiotherapy. Figure Shows the sensitivity of PAGAT polymer gel dosimeter with different energies. The R-dose sensitivity of PAGAT gel with energies (e.g,, and 8 MeV) are.6,. and.8, respectively. Verification of R-dose response of PAGAT gel dosimeter with different photon energies: Response of 8

Res. J. Appl. Sci. Eng. Technol., (): 6-, Slop R (/sec)..... 6 R for - 6Co R for MV R for 6MV R for 8MV Y =.9x+. R =.976 Y =.x+.96 R =.98 Y =.x+. R =.976 Y =.x+.99 R =.99 Fig. : Verification response of PAGAT polymer gel dosimeter with different energies (e.g.,.,, 6 and 8 MV) Slope Linear (slope) Y =.x+. R =. Fig. 6: Sensitivity of PAGAT polymer gel dosimeter with different energies (e.g.,.,, 6 and 8 MV) the energy dependence was obtained for.,, 6 and 8 MV nominal photon energy and is shown in Fig.. a linear function is fitted to each set of data and their slope and correlation were found. As it can be seen from Fig., dependence of PAGAT polymer gel response to the energy of photon is shown. As the energy increases from. to 8 MV the sensitivity is nearly stable:.. 9 %. which is % variation energy. Due to the large proportion of water (86%), the PAGAT polymer gel is nearly soft-tissue equivalent and no energy corrections are required for photon beams in radiotherapy. Figure 6 shows the sensitivity of PAGAT polymer gel dosimeter with different energies. The R-dose sensitivity of PAGAT gel with energies (e.g,.,, 6 and 8 MV) are.9, o.,. and., respectively. DISCUSSION AND CONCLUSION According to the best knowledge of the authors only a few studies were published to evaluate energy and dose rate dependence for the polymer gel dosimeters. Ibbott et al. (997) which worked on BANG polymer gel, concluded that the polymer gel dose response is independent of both dose rate and beam modality and used the polymer gel calibration curve determined with 6 MV x rays for evaluation of the gel dosimeter irradiated with 6 Co without any limitations or approximations (Ibbott et al., 997). Similarly De Wagter et al. (996) which worked on polymer gel stated in their study that polymer gel responds practically independently of beam quality (De Wagter et al., 996). Another study was also focused on the BANG polymer gel dosimeter response in different physical (Farajollahi et al., 999). They studied the dosimeter response for four different energies: kv x rays,. MeV 6 Co gamma rays, 6MV x rays, and 8MV x rays. They homogeneously irradiated gel dosimeters with the same dose by four mentioned photon beams and compared the dosimeter response to conclude that there is no energy dependence (Farajollahi et al., 999; Farajollahi, 999). Baldock et al. (996) studied PAG (poly acrylamide gelatin) polymer gel dosimeter response for a range of electron and photon energies. They used incident nominal electron energies of, 7, 9,,, 7, and MeV and photon energies of 66 kev from 7 Cs,. MeV from 6 Co, 6MV and 6MV to deliver the dose of Gy in each sample. Based on measured polymer gel dosimeter response they concluded that the energy response of the polymer gel retains suitable characteristics for a range of commonly used electron and photon beams (Baldock et al., 996). Maryanski et al. (996) used central axis percentage depth dose of 6 MV x-ray and MeV electron beams measured in the BANG- gel dosimeter (composed of: % Gelatin, % N,N -methylene-bis-acrylamide(bis), % Acrylic acid, % NaOH & 88% Water) and compared them with diode measurements in water. They concluded that there is no energy dependence of gel dosimeter for electron beams in the range -MeV (Maryanski et al., 996). This is similar to our work however the ingredients in BANG- and PAGAT are totally different. To avoid potential problems with different dosimeter response in different physical conditions one should perform calibration of polymer gel dosimeter and exposure of test phantom under the same or very similar physical conditions. PAGAT polymer gel dosimeter displayed good energy responses, which are important consideration when developing polymer gel dosimeters. REFERENCES Baldock, C., A.G. Greener, N.C. Billingham, R. Burford and S.F. Keevil, 996. Energy response and tissue equivalence of polymer gels for radiation dosimetry by MRI. Proc. Eur. Soc. Magn. Resonance Med. Biol., :. 9

Res. J. Appl. Sci. Eng. Technol., (): 6-, Brindha, S., A.J. Vennin, B. Hill and C. Baldock,. Experimental study of attenuation properties of normoxic polymer gel dosimeters. Phys. Med. Biol., 9: -6. De Deene, Y.,. Essential characteristics of polymer gel dosimeters: Third international conference radiotherapy gel dosimeter. J. Phys. Conf. Ser. : - 7. De Wagter, C., Y. De Deene, B. Mersseman and W. De Neve, 996. Comparison between polymer gel and film dosimetry in anthropomorphic phantoms. Abstract No., th Annual ESTRO Meeting, Vienna, Austria, pp: -6. Farajollahi, A.R., 999. An investigation into the applications of polymer gel dosimetry in radiotherapy. Abstract of Ph.D. Thesis, Med. Phys., 6: 9. Farajollahi, A.R. D.E. Bonnett, A.J. Ratcliffe, R.J. Aukett and J.A. Mills, 999. An investigation into the use of polymer gel dosimetry in low dose rate brachytherapy. Br. J. Radiol., 7: 8-9. Gore, J.C., Y.S. Kang and R.J. Schulz, 98. Measurement of radiation dose distributions by Nuclear Magnetic Resonance (NMR) imaging. Phys. Med. Biol., 9: 89-97. Ibbott, G.S., M.J. Maryanski, P. Eastman, S.D. Holcomb, Y. Zhang, R.G. Avison, M. Sanders and J.C. Gore, 997. Three-dimensional visualization and measurrment of conformal dose distributions using magnetic resonance imaging of BANG polymer gel dosimeters. Int. J. Radiat. Oncol., Biol., Phys., 8: 97-. Kennan, R.P., K.J. Richardson, J. Zhong, M.J. Maryanski and J.C. Gore, 996. The effects of cross-link density and chemical exchange on magnetization transfer in polyacrylamide gels. J. Magn. Reson., Ser. B: 67-77. Maryanski, M.J., J.C. Gore, R. Kennan and R.J. Schulz, 99. NMR relaxation enhancement in gels polymerized and cross linked by ionizing radiation: A new approach to D dosimetry by MRI (Magn. Reson. Imag.,, -8. Maryanski, M.J., R.J. Schulz, G.S. Ibbott, J.C. Gatenby, J. Xie, D. Horton and J.C. Gore, 99. Magnetic resonance imaging of radiation dose distributions using o polymer gel dosimeter. Phys. Med. Biol., 9: 7-. Maryanski, M.J., G.S., Ibbott, P. Estman, R.J. Schulz and J.C. Gore, 996. Radiation therapy dosimetry using magnetic resonance imaging of polymer gels. Med. Phys., : 699-7. Olsson, L.E., A. Fransson, A. Ericsson and S. Mattsson, 99. MR imaging of absorbed dose distributions for radiotherapy using ferrous sulphate gels. Phys. Med. Biol., : 6-6. Schulz, R.J. and J.C. Gore, 99. Reported in imaging of D dose Distributions by NMR National Cancer Institute Grant Application. ROCA66-. Venning, A.J., B. Hill, S. Brindha, B.J. Healy and C. Baldock,. Investigation of the PAGAT polymer gel dosimeter using magnetic resonance imaging. Phys. Med. Biol., : 87-888.