Elbow Hinge Fixator. Guided Flexion/Extension for Unstable Elbow Fractures.

Similar documents
Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area.

Monolateral External Fixation System for Trauma and Orthopaedics

External Distal Radius Fixator. Supplement to the 8 mm rod fixator system

Mini External Fixator.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Technique Guide. The Distraction Osteogenesis Ring System. Nonarticular tibia frame.

LCP Distal Humerus Plates

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

Alignment Rod. For intraoperatively confirming correction of the mechanical leg axis.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90

Small External Fixator

Variable Angle LCP Volar Rim Distal Radius Plate 2.4. For fragment-specific fracture fixation with variable angle locking technology.

Technique Guide. PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

Medium External Fixator Humeral Shaft Frame. Modular frame for upper extremity use.

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Humerus Block. Discontinued December 2016 DSEM/TRM/0115/0296(1) Surgical Technique. This publication is not intended for distribution in the USA.

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

3.5 mm LCP Extra-articular Distal Humerus Plate

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

VA-LCP Anterior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

LCP Extra-articular Distal Humerus Plate.

USS Variable Axis Screw

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

Small External Fixator Wrist Spanning Frame. For the treatment of wrist fractures.

3.5 mm LCP Olecranon Plates

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Small External Fixator

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

The Distraction Osteogenesis Ring System

Small External Fixator Nonspanning Wrist Frame. For the treatment of wrist fractures.

Technique Guide. Compact 2.0 LOCK Mandible. The locking system for the mandible.

LCP Percutaneous Aiming System 3.5 for PHILOS. For less invasive surgery at the proximal humerus.

Technique Guide. LCP Posterior Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment LCP system.

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Technique Guide. Adjustable Distal Radius Fixator. Part of the Synthes External Fixation Systems.

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

MEFiSTO. Monolateral External Fixation System for Trauma and Orthopaedics.

Olecranon Osteotomy Nail. For simple fractures and osteotomies of the olecranon.

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures.

Surgical technique. IMF Screw Set. For temporary, peri opera tive stabilisation of the occlusion in adults.

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System.

EXTERNAL FIXATION SYSTEM

Locking Proximal Humerus Plate. For complex and unstable fractures.

The Titanium Tibial Nail System

3.5 mm LCP Hook Plate

Knee Surgical Technique

TSLP Thoracolumbar Spine Locking Plate

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM

Handling Technique. Collinear Reduction Clamp. For minimally invasive fracture reduction.

Surgical Technique. This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

Large External Fixator Modular Knee Bridge. Using multi-pin clamps.

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments

2.4 mm Variable Angle LCP Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

3.5 mm LCP Distal Humerus Plates

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

Small External Fixator, Radiolucent, Sterile. Specific design for distal radius fractures.

DOUBLE/TRIPLE PELVIC OSTEOTOMY PLATES For Treating Coxofemoral Joint Instability and Subluxation in Immature Dogs

For Small-Statured Adults and Pediatric Patients. Medium External Fixator Basic Modular Frame

Periarticular Aiming Arm Instruments for LCP Proximal Tibial Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large).

LCP Periarticular Proximal Humerus Plate 3.5. The anatomic fixation system with anterolateral shaft placement.

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

Large External Fixator Modular Knee Bridge

Hoffmann II Micro External Fixation System

3. Insert Tocar Sleeves Insert the NCB tissue protection sleeve assembly 1.6 to 10mm through a skin incision (Fig. 38).

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Low Bend Distal Tibia Plates

Large External Fixator Delta Frame Ankle Bridge. For staged fixation of the distal tibia.

The CerviFix System. Including the StarLock Components. CerviFix Clamp and Screw. StarLock Clamp and Screw

Technique Guide. TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee.

Technique Guide. VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Collinear Reduction Clamp

Large External Fixator Delta Frame Ankle Bridge. Using pin clamps with outrigger posts.

ARCH Laminoplasty System. Dedicated System for Open-door Laminoplasty.

Mandible External Fixator II

LCP Wrist Fusion Set. Anatomic plates for total wrist fusion.

Technique Guide. DHS Blade. For osteoporotic bone.

Cannulated Angled Blade Plate 3.5 and 4.5, 90.

QUICK REFERENCE GUIDE. The PreFix Fixator (92000 Series) ALWAYS INNOVATING

Technique Guide. Rotation Correction Plates 1.5 and 2.0. Reposition plates for fractures and osteotomies at the metacarpals and phalanges.

Mini External Fixator

Distal Radius Plate 2.4/2.7 dorsal and volar

LCP Proximal Tibial Plate 4.5/5.0 with Periarticular Aiming Arm Instruments

LCP Metaphyseal Plates. For extra-articular fractures.

VA-LCP Condylar Plate 4.5/5.0. Part of the Synthes Variable Angle Periarticular Plating System.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

Medium External Fixator Delta Frame Ankle Bridge

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Transcription:

Elbow Hinge Fixator. Guided Flexion/Extension for Unstable Elbow Fractures. Surgical Technique MR Safe Radiolucent

Table of Contents System Description 3 Indications and Contraindications 4 Fixation Components 5 Implants 6 Instruments 7 Surgical Technique 8 Bibliography 12 Image intensifier Warning This description is not sufficient for immediate application of the instrumentation. Instruction by a surgeon experienced in handling these instruments is highly recommended. Synthes Elbow Hinge Fixator. Surgical Technique 1

Elbow Hinge Fixator. Guided Flexion/Extension for Unstable Elbow Fractures. Free frame design Patient-specific assembly of the fixation frame Free pin placement in the humerus and ulna Elbow Hinge Fixator is compatible with 8.0 and 11.0 mm carbon fibre rods Can supplement internal fixation Monitored axis determination Axis determined with 2.5 mm guide wire Metal sleeve enables the coaxiality to be determined at all times with the image intensifier Radiolucent materials allow easier visualization of the affected anatomy Early mobilization Active and passive post-operative mobilization of the elbow joint Spares the joint surface thanks to the guided hinge movement The joint can be blocked by bridging with an additional rod 2 Synthes Elbow Hinge Fixator. Surgical Technique

AO Principles In 1958, the AO/ASIF (Association for the Study of Internal Fixation) established four basic principles 1 that became guidelines for osteosynthesis. Anatomical repositioning To ensure successful treatment using mobile, joint-bridging external fixators, the joint must be reduced. Stabile fixation The proven modular rod construction technique ensures that the bridged joint is sufficiently stabilized. The integrated hinged joint permits specific, guided movement, thus relieving the stress on the ligamentous apparatus. Maintenance of the blood supply The use of Schanz screws is minimally harmful to tissue and enables the bone fragments to be fixed distant from the fracture. The approach minimizes soft tissue trauma during surgery and retains the necessary blood supply for bone healing. Early mobilization The Elbow Hinge Fixator is designed to absorb force during flexion and extension, transmit it to the center of movement of the fixation element, and relieve the stress on the internal osteosynthesis. Depending on the patient s situation, the affected joint can be directly mobilized after surgery during rehabilitation. This reduces the risk of ankylosis and helps accelerate the restoration of the original mobility and function of the patient. 1 M.E. Müller, M. Allgöwer, R. Schneider, and R. Willenegger (1991) AO Manual of Internal Fixation, 3 Edition. Berlin: Springer. 3

System Description As a component of the AO family of external fixators, the joint bridging mobile fixator enables a high degree of flexibility thanks to its modular frame construction. The free pin placement allows greater adaptation to the particular circumstances of each injury. The Elbow Hinge as the central element is compatible with fixation components of large and medium-size external fixators. This increases the flexibility in selecting a patient-specific, optimum form of therapy. Furthermore, little effort is required to switch from rigid to movable external fixation. The radiolucency of the Elbow Hinge Fixator helps you find the anatomical joint axis and makes it easier to position the mechanical axis of the Elbow Hinge. The fixation components are all comprised of MR-Safe materials. Patients with external fixators can be reliably examined at later times in magnetic resonance tomograms. 4 Synthes Elbow Hinge Fixator. Surgical Technique

Indications and Contraindications Indications The guided, joint-bridging external fixator assembly is suitable for supplementary treatment of complex, unstable elbow injuries when early functional stress is impossible due to persistent ligamental instability. The most important indications for guided joint bridging with external fixators are: Delayed treatment of dislocated and rigid elbows Chronic, persistent joint instability Acute joint instability after complex ligamentary injuries Unstable elbow fractures For adults, Elbow Hinge Fixator is preferably configured with the components of the large external fixator (rod diameter: 11 mm), and with components of the medium-size external fixator (rod diameter: 8 mm) for children and small adults. Contraindications Patients who for social and physical reasons are not suitable for an external fixator. Patients in whom no screws can be inserted due to a bone or soft tissue disease. 5

Fixation components 394.055 Hinged Joint Rod for External Fixator Fits the large external fixator 390.008 Clamp, clip-on, self-holding, MR-safe 390.005 Combination Clamp, clip-on, self-holding, MR-safe 394.800 Carbon Fibre Rod 11.0 mm 394.840 Fits the medium external fixator 390.035 Clamp, clip-on, self-holding, MR-safe 390.031 Combination Clamp, clip-on, self-holding, MR-safe 390.037 Combination Clamp, 8.0/11.0, clip-on, selfholding, MR-safe 395.779 Carbon Fibre Rod 8.0 mm 395.786 6 Synthes Elbow Hinge Fixator. Surgical Technique

Implants Self-Drilling Schanz Screws Reinforced bone anchoring due to radial preloading and symmetrical thread profile Reduced heat generation and insertion torque due to optimized drill tip geometry Titanium Steel Diameter Length (mm) (mm) 494.771 294.771 4.0/3.0 80 494.772 294.772 4.0/3.0 100 494.774 779 294.774 779 4.0 60 175 Self-Tapping Schanz Screws Titanium Steel Diameter Length (mm) (mm) 494.300 294.300 4.0/3.0 80 494.430 460 294.430 460 4.0 60 125 Guide Wire The guide wire is used to determine the anatomical joint axis during surgery and position the Elbow Hinge Fixator in reference to this axis. 292.260 Kirschner Wire 2.5 mm with trocar tip, length 280 mm, Stainless Steel 7

Instruments Adapter 393.101 Adapter for Seldrill Schanz Screw 4.0 mm 393.103 Adapter for Seldrill Schanz Screw 5.0 mm Combination wrenches 321.160 Combination Wrench 11.0 mm 321.158 Combination Wrench 8.0 mm Drill Guide System 395.911 Handle for Drill Sleeve Drill sleeves 5.0 mm 395.921 Drill Sleeve 6.0/5.0, short, with thread 395.912 Drill Sleeve 5.0/3.5, short 394.181 Trocar 3.5 mm, short Drill sleeves 4.0 mm 395.922 Drill Sleeve 4.0, with thread 392.955 Drill Sleeve 4.0/2.5 394.183 Trocar 2.5 mm 8 Synthes Elbow Hinge Fixator. Surgical Technique

Surgical Technique The following steps are explained with reference to the large external fixator and self-drilling, self-tapping (Seldrill) Schanz screws. Optionally, conventional self-tapping Schanz screws can also be used. For a detailed description of the handling of the screws, please refer to the surgical technique for large and medium-size external fixators (036.000.237). The joint bridging mobile fixator for the elbow joint is assembled in modules and is oriented around the pivot of the condyles. Therefore, for correct application, it is necessary to accurately align the upper arm segment prior to connecting the frame to the Schanz screws in the ulna. 1 Insert the Schanz screws in the humerus Required components Seldrill Schanz Screw 5.0 mm X94.774 779 Handle for Drill Sleeve 395.911 Drill Sleeve 4.0, short with thread 395.922 Drill Sleeve 4.0/2.5 392.955 Trocar 2.5 mm 394.183 Adapter for Seldrill Schanz Screw 4.0 mm 393.101 Drill with attachment for AO/ASIF quick coupling Type-dependent Insert the first two Schanz screws in the humerus. Select their position as appropriate for the soft tissue and anatomical situation. When setting the Schanz screws, pay special attention to the radial nerve. It is therefore recommended to spread the tissue with a blunt clamp to the bone, and to use a protection sleeve. The greater the distance between the Schanz screws, the greater the stability of the frame. 9

2 Insert the Schanz screws into the ulna Required components Seldrill Schanz Screw 5.0 mm X94.774 779 Handle for Drill Sleeve 395.911 Drill Sleeve 4.0, short with thread 395.922 Drill Sleeve 4.0/2.5 392.955 Trocar 2.5 mm 394.183 Adapter for Seldrill Schanz Screw 4.0 mm 393.101 Drill with attachment for AO/ASIF quick coupling Type-dependent Insert two other Schanz screws into the ulna from a dorsal direction. To optimize lower arm rotation, the two screws in the ulna should lie far in a dorsal direction at the edge of the ulna. 3 Connect the Schanz screws with carbon fibre rods Required components Clamp, clip-on, self-holding, MR-safe 390.008 Carbon Fibre Rod 11 mm 394.800 394.840 Combination Wrench 11 mm 321.160 Connect the Schanz screws in the humerus with two clip-on, self holding clamps and a carbon fibre rod. Repeat this process with the Schanz screws in the ulna. Depending on the assembly, make sure that the two rods project somewhat on the joint side to provide sufficient room for the addition frame. Tighten all the clamp nuts. Note: If the reduction needs to be secured, it is recommended to hold it during surgery with an additional rod. 10 Synthes Elbow Hinge Fixator. Surgical Technique

4 Determine the anatomical joint axis Required components Kirschner Wire 2.5 mm with trocar tip, length 280 mm, Stainless Steel 292.260 Drill Type-dependent Determine the anatomical axis using the image intensifier, and laterally insert a 2.5 mm guide wire for reference. To prevent damaging the ulnar nerve, do not drill through the medial cortex. When it is in the correct position, the guide wire represents a point in the center of the rotational axis. Note: A skilled practitioner can also determine the joint axis directly by the projection of the two condyles. The hole for the joint rod must represent a concentric circle to the superimposed condyles (see step 7). 11

5 Position the joint rod and connect to the humerus with a carbon fibre rod Required components Carbon Fibre Rod 11 mm 394.800 394.840 Hinged Joint Rod for External Fixator 394.055 Combination Clamp, clip-on, self-holding, MR-safe 390.005 Combination Wrench 11 mm 321.160 Guide the Elbow Hinge Fixator over the reference wire, and connect it to the partial frame of the humerus using an additional rod and a combination clamps. When connecting Elbow Hinge Fixator to the humoral partial frame, it is important to not bend the reference wire since the anatomical axis would then no longer correspond to the mechanical axis of the joint. Note: Depending on the position of the partial frame, the joint rod can also be connected directly by means of a rod-to-rod clamp. 6 Remove the reference wire Remove the reference wire after all the clamps of the proximal part of the fixator have been tightened. 12 Synthes Elbow Hinge Fixator. Surgical Technique

7 Check the axes The Elbow Hinge position is checked by visualizing the elbow joint in lateral position. When positioned correctly, the center of the fixator and the radial and ulnar condyles project exactly over each other. The center of the Elbow Hinge Fixator will appear as a small circle in the center of the superimposed images of the radial and ulnar condyles. 8 Connect the joint rod to the carbon fibre rod in the ulna Required components Combination Clamp, clip-on, self-holding, MR-safe 390.005 Carbon Fibre Rod 11 mm 394.800 394.840 Combination Wrench 11 mm 321.160 As a final assembly step after reducing the joint, the end of the Elbow Hinge that is distal from the central guide hole is connected to the ulnar partial frame. Note: Depending on the position of the partial frame, the joint rod can also be connected directly by means of a rod-to-rod clamp. 13

9 Check the Mobility of the Joint After finishing with assembly, check the mobility of the joint and its position. When the position is correct, the central, cannulated metal sleeve of the joint rod projects exactly in the center of the concentric condyles in the form of a circle in a lateral visualization. In the anterior-posterior plane, the central metal sleeve must be parallel to the joint surface since otherwise radial or ulnar deviations of the ulna are generated during movement. 14 Synthes Elbow Hinge Fixator. Surgical Technique

10 Blocking the Joint Required components Combination Clamp, clip-on, self-holding, MR-safe 390.005 Carbon Fibre Rod 11 mm 394.800 394.840 Combination Wrench 11 mm 321.160 Depending on the situation, the elbow joint can be temporarily blocked directly after surgery with an additional rod that connects the two partial frames proximal and distal from the joint. The duration of the blocking depends on the patient s situation. 15

Bibliography Jupiter J.B, and Ring D, (2002) Treatment of unreduced elbow dislocations with hinged external fixation, J Bone Joint Surg Am 84-A, 1630-1635 Schmickal T, Heppert V, Wentzensen A (2002) Einsatzmöglichkeiten des Fixateur externe bei Oberarm- und Ellbogenverletzungen. Trauma Berufskrankheiten 4: 337 382 Schmickal T, Wentzensen A (2000) Die Behandlung komplexer Ellbogenverletzungen mit einem gelenksüberbrückenden Fixateur mit Bewegungsachse. Unfallchirurg, 103: 191 196 Tan V, Daluiski A, Capo J and Hotchkiss R, (2005) Hinged Elbow External Fixators: Indications and Uses. J Am Acad of Orthop Surg 13, 503-514 16 Synthes Elbow Hinge Fixator. Surgical Technique

Synthes GmbH Eimattstrasse 3, CH-4436 Oberdorf www.synthes.com Presented by: 0123 036.000.663 SE_046486 AA 30050105 Synthes 2006 Printed in Switzerland Subject to modifications.