Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Similar documents
The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff

QA for Clinical Dosimetry with Emphasis on Clinical Trials

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

RPC s Credentialing Programs for Clinical Trials

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Quality assurance in external radiotherapy

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

INTRODUCTION. Material and Methods

EORTC Member Facility Questionnaire

International Practice Accreditation

Patient and Linac QA - present and the future

3D DOSIMETRY IN END-TO-END DOSIMETRY QA

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

SBRT Credentialing: Understanding the Process from Inquiry to Approval

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

Eric E. Klein, Ph.D. Chair of TG-142

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a

Limits of Precision and Accuracy of Radiation Delivery Systems

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Credentialing for Clinical Trials -IGRT. Evidence Based Radiation Oncology. Levels of Clinical Evidence. Why Credentialing?

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

EVALUATION OF INTENSITY MODULATED RADIATION THERAPY (IMRT) DELIVERY ERROR DUE TO IMRT TREATMENT PLAN COMPLEXITY AND IMPROPERLY MATCHED DOSIMETRY DATA

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output

Quality Assurance of Helical Tomotherapy Machines

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Verification of Advanced Radiotherapy Techniques

Overview of Advanced Techniques in Radiation Therapy

Catharine Clark NPL, Royal Surrey County Hospital and RTTQA

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

Activity report from JCOG physics group

Concept for quantifying the dose from image guided radiotherapy

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

SRS/SBRT Errors and Causes

BELdART-2: Moving towards safer radiotherapy. Steven Lelie 1,2, Wouter Schroeyers 1, Brigitte Reniers 1, Sonja Schreurs 1

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

SunCHECK Patient Comprehensive Patient QA

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017

Electron Beam ET - Reloaded Therapy - Reloaded

DEPT OF RADIATION ONCOLOGY DEPT OF RADIATION ONCOLOGY

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017

The Journey of Cyberknife Commissioning

Independent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA

Leila E. A. Nichol Royal Surrey County Hospital

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Commission, Ganakbari, Savar, Dhaka, Bangladesh

The Radixact System Experience at Miami Cancer Institute. Tino Romaguera, D.Sc. Senior Physicist

Learning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan

ph fax

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Quality Assurance in Radiation Therapy

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Clinical Trial Credentialing:

Effective and efficient radiotherapy dosimetry audit: Where to next?

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

How Should We Verify Complex Radiation Therapy Treatments?

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

Small field diode dosimetry

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

Objective. Evaluation of Linear Accelerator Performance Standards using an Outcome Oriented Approach. Linac performance standards*

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

doi: /j.ijrobp

A rose is a rose by any name

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning

Guideline & Reports 医学物理学会教育委員会資料

PGY-1. Resident Review Session Schedule

Transcription:

Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment planning data (On-site dosimetry reviews) Consistency of treatment records (Chart reviews) Understanding of advanced technology procedures (Questionnaires, phantoms,, etc.) AAPM Philadelphia July 2010

Constancy of Basic Machine Calibration RPC monitors 1,768 institutions, of which ~1,600 are in the US US Machines & Facilities Increase from 1,338 in 2005 (32%) Machines Number of radiation beams has increased more rapidly Year Facilities AAPM Philadelphia July 2010

Annual TLD/OSLD Audits Monitor ~ 14,000 beams/yr Conversion from TLD to OSLD AAPM Philadelphia July 2010

Distribution of TLD results Photons beams within 7% Number of beams: 3051 Avg. RPC/Inst.: 0.999 Stdev.: 1.6% Electrons beams within 7% Number of beams 4310 Avg. RPC/Inst: 0.998 Stdev.: 1.7% AAPM Philadelphia July 2010

Data for other machines AAPM Philadelphia July 2010

Data for other machines AAPM Philadelphia July 2010

Data for other machines AAPM Philadelphia July 2010

TLD measurements in proton beams Proton TLD Frequency Distribution 25 Protons: 109 measurements Photons: > 6,000 measurements 20 15 10 Frequency (%) 5 0 <0.94 0.94 0.95 0.96 0.97 0.98 0.99 1 1.01 1.02 1.03 1.04 1.05 1.06 >1.06 RPC/Institution AAPM Philadelphia July 2010

On-Site Dosimetry Review Visit Parameters Measured Review QA Program Implementation of TG-51 Review Temp/Press Correction Photon Calibration Photon FSD (incl. small field) Photon Depth Dose Off-axis Factors/Beam symmetry Electron Calibration Electron Depth Dose Electron Cone Ratios Wedge Transmission Consistent use of Data

Comprehensive On-Site Audit Improvements and Additions In response to new radiotherapy treatments in trials, new audit techniques have been implemented such as: 1. TomoTherapy 2. CyberKnife 3. Proton therapy 4. Small field dosimetry 5. 2D water scanner and electrometers 6. Electronic transfer of Visit data (2010) 7. Incorporate new AAPM TG-142 QA review (2010) 8. Image guidance (in development)

Comprehensive On-Site Audit Improvements and Additions In response to new radiotherapy treatments in trials, new audit techniques have been implemented such as: 1. TomoTherapy 2. CyberKnife 3. Proton therapy 4. Small field dosimetry 5. 2D water scanner and electrometers 6. Electronic transfer of Visit data (2010) 7. Incorporate new AAPM TG-142 QA review (2010) 8. Image guidance (in development)

Comprehensive On-Site Audit Improvements and Additions In response to new radiotherapy treatments in trials, new audit techniques have been implemented such as: 1. TomoTherapy 2. CyberKnife 3. Proton therapy 4. Small field dosimetry 5. 2D water scanner and electrometers 6. Electronic transfer of Visit data (2010) 7. Incorporate new AAPM TG-142 QA review (2010) 8. Image guidance (in development)

Comprehensive On-Site Audit Improvements and Additions In response to new radiotherapy treatments in trials, new audit techniques have been implemented such as: 1. TomoTherapy 2. CyberKnife 3. Proton therapy 4. Small field dosimetry 5. 2D water scanner and electrometers 6. Electronic transfer of Visit data (2010) 7. Incorporate new AAPM TG-142 QA review (2010) 8. Image guidance (in development)

Comprehensive On-Site Audit Improvements and Additions In response to new radiotherapy treatments in trials, new audit techniques have been implemented such as: 1. TomoTherapy 2. CyberKnife 3. Proton therapy 4. Small field dosimetry 5. 2D water scanner and electrometers 6. Electronic transfer of Visit data (2010) 7. Incorporate new AAPM TG-142 QA review (2010) 8. Image guidance (in development) As radiotherapy treatment techniques change, so do the visit audit techniques

Measurement of small-field output factors 6 MV, d = 10 cm Field Size Varian(64) Siemens(4) Elekta(10) 10 x 10 1.000 1.000 1.000 6 x 6 0.922 (0.012) 0.910 (0.004) 0.924 (0.004) 4 x 4 0.866 (0.019) 0.851 (0.004) 0.869 (0.007) 3 x 3 0.833 (0.025) 0.817 (0.003) 0.837 (0.005) 2 x 2 0.785 (0.013) 0.757 (0.014) 0.793 (0.007) 1 x 1 0.694 (0.042) - 0.659 (0.025)

Virtual Dosimetry Review Visit Use of the RPC s standard data - Compilation of RPC measured avg. data 2700 photon beams and 81 linac model/ energy combinations - Specific to make/model/energy with 5 sets of RPC measured data - Analyses of these data indicate that machines of same make/model/energy have same radiation characteristics. - Successful at predicting specific errors 88% of the time. - Available to Med. Phys. community upon request.

Web-Based Facility Questionnaire

Web-Based Facility Questionnaire

RPC Phantoms Pelvis (10) Spine (3) Thorax (13) H&N (31) TPC Houston: Apr. 20, 2010 SRS Head (4) Liver (2)

Treat phantom as if it were a patient

Primary PTV OAR Good Agreement Secondary PTV Measurement AAPM Philadelphia July 2010 Plan

Phantom Results Comparison between institution s plan and delivered dose. Phantom H&N Prostate Spine Lung Liver Irradiations 752 174 19 174 23 Pass 585 143 13 124 12 Pass % 78% 82% 68% 71% 52% Criteria 7%/4mm 7%/4mm 5%/3mm 5%/5mm 7%/4mm Year introduced 2001 2004 2009 2004 2005 AAPM Philadelphia July 2010

HN results grouped by TPS Treatment planning system Pass Rate (%) Attempts Criteria Failed Dose DTA Dose and DTA Corvus 75 32 7 0 1 Eclipse 85 114 10 4 3 Pinnacle 73 168 33 4 8 TomoTherapy 73 22 5 1 0 XiO 73 59 7 4 5 Other 79 24 3 0 2 Total 419 65 13 19

Explanations for Failures Explanation Minimum # of occurrences incorrect output factors in TPS 1 incorrect PDD in TPS 1 IMRT Technique 3 Software error 1 inadequacies in beam modeling at leaf ends (Cadman, et al; PMB 2002) QA procedures 3 errors in couch indexing with Peacock system equipment performance 2 setup errors 7 14 3 IAEA, July 6, 2010

AAPM Philadelphia July 2010 Supported by NCI grants CA10953 and CA81647