Breakage of a volar locking plate after delayed union of a distal radius fracture

Similar documents
CASE REPORT. Distal radius nonunion after volar locking plate fixation of a distal radius fracture: a case report

COMPARATIVE STUDY OF FUNCTIONAL OUTCOME OF EXTERNAL AND INTERNAL FIXATION IN TREATMENT OF COMMINUTED DISTAL RADIUS FRACTURES

ORIGINAL ARTICLE TREATMENT DISTAL RADIUS FRACTURE WITH VOLAR BUTTRESS TECHNIQUE- A CLINICAL STUDY

Case Report Volar Locking Plate Breakage after Nonunion of a Distal Radius Osteotomy

MINIMALLY INVASIVE PLATE OSTEOSYNTHESIS FOR DISTAL RADIUS FRACTURES: SURGICAL TECHNIQUE M. TOBE 1, K. MIZUTANI 1, Y. TSUBUKU 1, Y.

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Factors Predicting Late Collapse of Distal Radius Fractures

Locked Plating: Clinical Indications

Therapeutic Effects of Volar Anatomical Plates Versus Locking Plates for Volar Barton s Fractures

University of Groningen. Fracture of the distal radius Oskam, Jacob

Index. Note: Page numbers of article titles are in boldface type. Hand Clin 21 (2005)

ROLE OF LOCKING COMPRESSION PLATES IN THE TREATMENT OF FRACTURES DISTAL END OF RADIUS

Clinical Study Rate of Improvement following Volar Plate Open Reduction and Internal Fixation of Distal Radius Fractures

Orthopedics in Motion Tristan Hartzell, MD January 27, 2016

Kousuke Iba 1*, Yasuhiro Ozasa 1, Takuro Wada 1, Tomoaki Kamiya 1, Toshihiko Yamashita 1, Mitsuhiro Aoki 2. Abstract

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

Complications of Distal Radius Fractures. How to Treat a Distal Radius Fx 11/13/2017. Michael S. Bednar, M.D. Loyola University Chicago

Distal Femur Fractures in The Elderly The Ideal Construct

University of Groningen. Fracture of the distal radius Oskam, Jacob

Malunited extra-articular distal radius fractures: corrective osteotomies using volar locking plate

Treatment Approach To Cases Of Nonunion Intercondylar Fracture Humerus

Locked Plate Fixation of Osteoporotic Humeral Shaft Fractures: Are Two Locking Screws Per Segment Enough?

Surgical Management of Distal end Radius Fractures by Various Methods: A Prospective Study

A Patient s Guide to Adult Distal Radius (Wrist) Fractures

journal ORIGINAL RESEARCH

Distal radius fractures raises considerable interest and

ORIGINAL ARTICLE. possible. Accurate assessment of standard radiographs is essential for appropriate 3. management. And includes true posterior- 4

Differentiated approach to repair of displaced distal radial metaepiphyseal fractures. R.P. Matveev, S.V. Bragina, A.M. Shneiveis

Distal radius fractures

E ORIGINAL ARTICLE Low extra-articular (transcondylar) fractures of the distal humerus

Fractures of the distal end of the radius should be

Ultrasound-guided reduction of distal radius fractures

Common Limb Fractures. Mr Sheraz Malik MB BS MRCS Instructor Mr Paul Ofori-Atta Mb ChB FRCS President Motc Life UK April 2009

Screw Placement and Types Matter

Carpal rows injuries!

INTERNAL FIXATION OF THE METACARPALS AND PHALANGES P. BURGE

Functional Outcome in Distal Radius Fractures Treated with Locking Compression Plate

Treatment of non-union of forearm bones with a free vascularised cortico - periosteal flap from the medial femoral condyle

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

The Kienböck disease and scaphoid fractures. Mariusz Bonczar

Disclosure. 85% Dorsal. 15% Volar. Distal Radius Fractures- Volar Plating for All. Skeletal Dynamics- Consultant

Posteromedial approach to the distal humerus for fracture fixation

Treatment of unstable distal radius fractures with the volar locking plate

Conservative treatment of the distal radius fracture using thermoplastic splint: pilot study results

Surgical correction for supination loss following malunited radial fractures

Locked Plating: Biomechanics and Biology

Distal radius fracture Golden standard and breaking news

Results of tibia nailing with Angular Stable Locking Screws (ASLS); A retrospective study of 107 patients with distal tibia fracture.

A Patient s Guide to Adult Olecranon (Elbow) Fractures

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

St Mary Orthopaedic Conference. Steven A. Caruso, MD Trenton Orthopaedic Group Trauma and Complex Fracture Surgeon October 25, 2014

Scaphoid Fractures. Mohammed Alasmari. Orthopaedic Surgery Demonstrator Majmaah University

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

International Journal of Health Sciences and Research ISSN:

Angle Stable Distal Radial Plate System WINSTA-R

University of Groningen. Fracture of the distal radius Oskam, Jacob

Comparison between Distractor Application on Both Radial & Ulnar Side and Radial Side Only for Fracture Distal Radius with Ulnar Styloid Fracture

Fractures of the tibia shaft treated with locked intramedullary nail Retrospective clinical and radiographic assesment

Complications and functional outcome following fixation of complex, intra-articular fractures of the distal radius with the AO Pi-Plate

Distal Radius Plate 2.4/2.7 dorsal and volar

Comparative Study between Bridging External Fixation vs Volar Plating (Ellis-T Plate) for Comminuted Fracture of the Distal End Radius

Volar fixed-angle plating for distal radius fractures 刀圭会協立病院 津村敬

VOLAR LOCKING COMPRESSION SYSTEM DISTAL AND RADIUS FRACTURE DISSERTATION

A Patient s Guide to Adult Forearm Fractures

Olecranon fracture. Lonnie Froberg, MD, Ph.D Rigshospitalet, Copenhagen University Hospital

Prospective Randomised Study of Intra-Articular Fractures of the Distal Radius: Comparison Between External Fixation and Plate Fixation

Forearm Fracture Solutions. Product Overview

2016 SIGN Conference: Treatment of Difficult Fractures Around The World

Mark VanDer Kaag 1, Ajmal Ikram 2. Hand Unit, Tygerberg Hospital University of Stellenbosch

A Patient s Guide to Adult Metacarpal Fractures of the Hand

Wrist Arthritis & Partial Wrist Fusion

JMSCR Vol 05 Issue 06 Page June 2017

Research Article How Early Can We Mobilise 4 th And 5 th Metacarpal Shaft Fractures? A Retrospective Study

Technique Guide. Rotation Correction Plates 1.5 and 2.0. Reposition plates for fractures and osteotomies at the metacarpals and phalanges.

Hand and wrist emergencies

Redisplacement of Distal Radius Fracture after Initial Closed Reduction: Analysis of Prognostic Factors

The value of manipulation of displaced distal radius fractures in the emergency department

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

Clinical Study Dorsally Comminuted Fractures of the Distal End of the Radius: Osteosynthesis with Volar Fixed Angle Locking Plates

Recurrent subluxation or dislocation after surgical

Bridging external fixation and supplementary Kirschner-wire fixation versus volar locked plating for unstable fractures of the distal radius

JMSCR Vol. 03 Issue 08 Page August 2015

4/28/2010. Fractures. Normal Bone and Normal Ossification Bone Terms. Epiphysis Epiphyseal Plate (physis) Metaphysis

University of Groningen. Fracture of the distal radius Oskam, Jacob

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Evaluation of volar locking plate fixation for management of intraarticular fractures of distal end of radius

Ankle Fractures in the Elderly: How to Deal with Poor Bone Quality

IS CASTING AN ACCEPTABLE TREATMENT APPROACH FOR A DISTAL RADIUS FRACTURE THAT HAS UNDERGONE A SATISFACTORY CLOSED REDUCTION?

The NBX Non-Bridging External Fixator A Non-Bridging External Fixator/Locking Plate capturing a series of.062mm K-wires and 3mm half-pins that are

FOOSH It sounded like a fun thing at the time!

Raju Vaishya, Amit Kumar Agarwal, Nishint Gupta & Vipul Vijay

ComparisonofUlnar-ShorteningOsteotomyWitha NewTrimedDynamicCompressionSystemVersusthe SynthesDynamicCompressionSystem:ClinicalStudy

Functional Outcome Of Intraarticular Distal Radius Fractures Managed By Volar Locking Compression Plate A Prospective Study

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System.

FUNCTIONAL AND RADIOLOGICAL OUTCOMES AFTER VOLAR LOCKING PLATE FIXATION IN TYPE C DISTAL RADIUS FRACTURES A RETROSPECTIVE STUDY

D-RAD SMART PACK Plating System. A series of case studies

NE Nebraska Trauma Conference Tristan Hartzell, MD November 8, 2017

Transcription:

cta Orthop. elg., 2007, 73, 785-790 CSE REPORT reakage of a volar locking plate after delayed union of a distal radius fracture Tom DE ERE, Frédéric LECOUVET, Olivier RIER From Cliniques Universitaires St-Luc, ruxelles, elgium This is a case report of a 58-year-old obese, diabetic female who sustained a dorsally angulated distal radius fracture, treated with a volar locking plate. Union was delayed and the plate broke. The fracture healed after repeated surgery involving replacement of the plate and grafting with demineralised bone matrix. Delayed unions of distal radius fractures are rare. This case shows that fracture treatment with locking compression plates has some specific biomechanical consequences which have to be taken into account, and special attention will be given to technical peculiarities and possible drawbacks. Keywords : distal radius ; locking plate ; hardware failure ; delayed union. CSE REPORT 58-year-old right handed woman presented at our department after a fall. She complained of pain at the right wrist. The radiographs showed a dorsally displaced fracture of the distal radius with intra-articular extension (fig 1). ccording to the classification proposed by Jupiter and Fernandez this fracture was classified as type I (8). Her medical history revealed type II diabetes treated with oral antidiabetics, hypertension and significant obesity. She was not a smoker. Measurement of bone mineral density 7 months after this initial trauma was within the normal range (T-score : - 0.34 at the lumbar spine, i.e. 96% of the normal value and T-score : - 0.66 at the left hip, i.e. 91% of INTRODUCTION Fractures of the distal radius are among the most common musculoskeletal injuries (1). Non-union of these fractures however is uncommon (5,7,10). The functional outcome after distal radius fractures correlates closely with radiocarpal alignment and restoration of length (13,14). Recently fixed angle internal fixation devices were developed to combine the facilities of conventional plate osteosynthesis with those of an internal fixator system (12). We present a case of non-union followed by plate breakage after volar plating of a distal radius fracture without bone grafting. Tom De aere, MD, Orthopaedic Surgeon. Department of Orthopaedics, Clinique St-Jean and St- Etienne, russels. Olivier arbier, MD, Orthopaedic Surgeon. Department of Orthopaedic surgery and Traumatology, Cliniques Universitaires St-Luc, russels, elgium. Frédéric Lecouvet, MD, PhD, Radiologist. Department of Radiology, Cliniques Universitaires St-Luc, russels, elgium. Correspondence : Olivier arbier, Cliniques Universitaires St-Luc, Service d Orthopédie et de Traumatologie de l appareil locomoteur, venue Hippocrate 10, 1200 ruxelles, elgium. E-mail : Olivier.arbier@orto.ucl.ac.be 2007, cta Orthopædica elgica. No benefits or funds were received in support of this study

786 T. DE ERE, F. LECOUVET, O. RIER Fig. 1. Preoperative radiographs : Distal radius fracture with dorsal displacement. : nteroposterior view ; : Lateral view the normal value) for sex and age (Hologic, dual X- ray absorptiometry, measurements at the level of left hip and lumbar spine). Open reduction and internal fixation was scheduled three days after the accident. The fracture was approached through a volar incision and was fixed after reduction with a 3.5 T-type Locking Compression Plate (LCP) (Mathys Medical Ltd, ettlach, Switzerland). ngle-stable locking screws were used proximally as well as distally. Postoperatively, the forearm was immobilised in a plaster cast. The postoperative radiographs showed radial shortening, 10 palmar inclination and discontinuity in the anterior cortical alignment at the fracture site. The plate was not in contact with the anterior cortex (fig 2). fter two weeks the plaster cast was replaced by a removable splint allowing gentle mobilisation. Six weeks after surgery, the splint was removed and a rehabilitation program under supervision of a physiotherapist was started. t this stage only mobilisation of the wrist and hand was allowed without strengthening exercises. One month later functional improvement was good but the patient reported some pain during and after effort. The radiographs showed a slight posterior collapse of the fracture site (fig 3), due to slight bending of the plate without screw loosening. Rehabilitation was continued. Fourteen weeks after the operation the pain had significantly increased. The patient reported a painful event during a rehabilitation session, a few days earlier. Thereafter, when leaning on the wrist, she heard a cracking. Clinical examination revealed shortening and radial displacement at the fracture site. Radiographs showed a broken plate at the original fracture site. This breakage occurred through an empty screw hole and the distal radius was displaced dorsally and radially (fig 4). Five days later the patient was re-operated and the plate and screws were removed. The non-union was fixed with a similar plate in conjunction with a posterior bone graft. s iliac crest harvesting would have been difficult due to the morbid obesity of the patient, the nonunion site was grafted with demineralised bone matrix (llomatrix, Wright Medical Corporation) (fig 5). Locking screws were used distally but this time classical bicortical nonlocking screws were used for proximal fixation of the plate, giving compression at the non-union site.

REKGE OF VOLR LOCKING PLTE 787 Fig. 2. Postoperative radiographs : View after fracture fixation with an anterior LCP-plate. : nteroposterior view ; : Lateral view. fter the surgery the forearm was again immobilised in a plaster cast and the patient left the hospital three days later. lthough the peroperative fluoroscopic images taken with the C-arm were reassuring, the radiograph a few days after the intervention showed an intra-articular positioning of one of the two distal screws. It was decided to leave this screw in place and remove it after fracture healing. This time the plate was in close contact with the anterior cortex (fig 6). The plaster cast was removed at 7 weeks and replaced by a removable splint and physiotherapy was begun. Complete fracture consolidation was obtained 8 months after the initial trauma : the fracture line disappeared and there was no pain on palpation of the fracture or mobilisation of the wrist (fig 6). The Disability of rm, Shoulder and Hand (DSH) score at the time of healing was 56.2 (normal value : 50 ; standard deviation : 10) (6). DISCUSSION Non-union of the distal radius is very rare but cases have been reported in association with diabetes mellitus and obesity (11). These non-unions are usually symptomatic (5,7) and can be treated by rigid internal fixation and iliac crest bone grafting (2). Routine bone grafting of unstable fractures of the distal radius has even been suggested by several authors to promote fracture healing (3,12). ngular stability of the LCP-plate is reported to maintain fragments in place in distal radius fractures that might have been at risk of tilting, without opening the fracture site for grafting, although osteoporosis is known to be a risk factor for failure of the LCP-system (12). It has been demonstrated that the transmission of forces to the proximal radius is greater through the volar LCP-system than through a conventional volar or dorsal compression plate (9). In such an

788 T. DE ERE, F. LECOUVET, O. RIER Fig. 3. Radiographs at 10 weeks : Dorsal impaction of the fracture site with plate bending. : nteroposterior view ; : Lateral view. Fig. 4. Radiographs at 14 weeks : reakage of the plate with secondary displacement of the fracture. : nteroposterior view ; : Lateral view.

REKGE OF VOLR LOCKING PLTE 789 Fig. 5. CT-scan 3 months after the second operation, showing the demineralised bone matrix graft in the metaphyseal bone area. osteosynthesis, the plate is exposed to high bending forces, particularly because there is no bony support from the opposite cortex. Non-union in our case was probably caused by several factors. The reduced capillary vascularity associated with diabetes and the surgical periosteal devascularisation created a biological situation which was unfavourable for fracture healing. Mechanical weaknesses were the lack of continuity of the anterior cortex as well as the poor contact of the plate with the anterior cortex. In combination with the dorsal fracture comminution this gave an unstable construct that went on to failure. This is evidenced by the change in the angulation between the plate and the distal screws on early postoperative radiographs (fig 2) and radiographs at 10 weeks (fig 3). s shown by the lateral radiographs (figs 2, 3), all the forces were transmitted through the distal part of the plate, acting as an internal fixator without cortical contact. Initial bone grafting to support the posterior cortex might have been indicated in this situation. Radiographs 2.5 months after initial surgery showed a slight posterior collapse of the fracture site (fig 3). The plate was submitted to prolonged high bending forces and ultimately broke. Fig. 6. Radiographs 8 months after the second operation, showing bony healing. : nteroposterior view ; : Lateral view.

790 T. DE ERE, F. LECOUVET, O. RIER The stability of the locking screw system is demonstrated by the fact that none of the screws had loosened or pulled out when the plate broke (fig 4). Fracture union was obtained after posterior bone grafting and revision of the osteosynthesis with a plate that was in close contact with the anterior cortex. Moreover in the proximal part of the plate, bicortical screws without angular stability were used to allow a progressive transmission of load between the plate and the bone (fig 6). These bicortical screws have a longer working length, which enables them to withstand axial and bending forces over a longer time (4). We found one similar report of failure of the LCP-system in the distal radius : it was a case of cut out of the epiphyseal screws in the distal fragment because of insufficient fixation in osteoporotic bone, combined with anterior instability through fracture comminution and lack of contact between the anterior cortex and the plate (12). This failure however went on to fracture healing without plate breakage ; besides, the patient in our case showed no osteoporosis as confirmed by a bone mineral density measurement 7 months after the initial trauma. CONCLUSION The great variety of possibilities offered by the LCP-system requires a good biomechanical analysis of the lesion and careful surgical technique. In our case restoration of length and reduction of the anterior cortex of the distal radius was insufficient and the construct was too weak to withstand the forces transmitted through it and that is why the osteosynthesis went on to failure. Initial bone grafting would probably have been better as well to avoid non-union in this patient. REFERENCES 1. ngermann P, Lohmann M. Injuries to the hand and wrist : a study of 50,272 injuries. J Hand Surg 1993 ; 18- : 642-644. 2. Fernandez DL, Ring D, Jupiter J. Surgical management of delayed union and non-union of distal radius fractures. J Hand Surg 2001 ; 26- : 201-209. 3. Fernandez D, Wolfe S. Green s Operative Hand Surgery. 5 th Ed, Elsevier Churchill Livingstone, Philadelphia, 2005, pp 645-704. 4. Gautier E, Sommer C. Guidelines for the clinical application of the LCP. Injury 2003 ; 34 Suppl 2 : 43-54. 5. Harper WM, Jones JM. Non-union of Colles fracture : Report of two cases. J Hand Surg 1990 ; 15- : 121-123. 6. Hudak PL, madio PC, ombardier C. Development of an upper extremity outcome measure : the DSH (disabilities of the arm, shoulder and hand). m J Industr Med 1996 ; 29 : 602-608. 7. Jupiter J, Fernandez DL, Toh CL Fellman T, Ring D. Operative treatment of volar intra-articular fractures of the distal end of the radius. J one Joint Surg 1996 ; 78- : 1817-1828. 8. Jupiter J, Fernandez DL. Fractures of the Distal Radius. Springer-Verlag, New York, 1996, pp 49-50. 9. Leung F, Zhu L, Ho H, Lu WW, Chow SP. Palmar plate fixation of O type C2 fracture of distal radius using a locking compression plate : a biomechanical study in a cadaveric model. J Hand Surg 2003 ; 28- : 263-266. 10. Margaliot Z, Haase SC, Kotsis SV, Kim HM, Chung KC. meta-analysis of outcomes of external fixation versus plate osteosynthesis for unstable distal radius fractures. J Hand Surg 2005 ; 30- : 1185-1199. 11. Segalman K, Clark GL. Un-united fractures of the distal radius : a report of 12 cases. J Hand Surg 1998 ; 23- : 914-919. 12. Sommer C, abst R, Muller M, Hanson. Locking compression plate loosening and plate breakage : a report of four cases. J Orthop Trauma 2004 ; 18 : 571-577. 13. Taleisnik J, Watson HK. Midcarpal instability caused by malunited fractures of thedistal radius. J Hand Surg 1984 ; 9- : 350-357. 14. Kim HM, Chung KC et al. Intra-articular fractures of the distal aspect of the radius. Instr Course Lect 1999 ; 48 : 465-480.