Implementing a Generic Methods Development Strategy for Enantiomer Analysis

Similar documents
CE Columns and Accessories

Impurity Profiling of Carbamazepine by HPLC/UV

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Analysis of L- and D-Amino Acids Using UPLC Yuta Mutaguchi 1 and Toshihisa Ohshima 2*

column If you would like a subscription, or more information, please contact your nearest Waters Office.

[ APPLICATION NOTE ] The Separation of 8 -THC, 9 -THC, and Their Enantiomers by UPC 2 Using Trefoil Chiral Columns INTRODUCTION APPLICATION BENEFITS

Analysis of the Non-Ionic Surfactant Triton-X Using UltraPerformance Convergence Chromatography (UPC 2 ) with MS and UV Detection

Independent Column Temperature Control Using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds

Tenofovir disoproxil fumarate (Tenofoviri disoproxili fumaras)

Heparin Sodium ヘパリンナトリウム

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

Capillary electrophoresis facilitates determination of metal complex stoichiometry by Job s method of continuous variation

Ultra High Performance Liquid Chromatograph. Nexera X2 C196-E079

Chromatography Vacuum Ultraviolet Spectroscopy

A RAPID AND SENSITIVE UPLC/UV/MS METHOD FOR SIMVASTATIN AND LOVA S TAT IN IN SU P P O RT O F C L E A NING VA L I DAT IO N S T U DIES

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

Capillary electrophoretic techniques toward the metabolome analysis*

SUPPLEMENTAL INFORMATION

Determination of Benzodiazepines in Urine by CE-MS/MS

SIMULTANEOUS ESTIMATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN TABLETS BY RP-HPLC METHOD

ABACAVIR SULFATE Proposal for revision of The International Pharmacopoeia (August 2012)

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations

Choosing Preparative Columns

CONTENT. i iv ix. SVKM s NMIMS, School of Pharmacy and Technology Management

Phenyl-Hexyl. UHPLC Columns. lternate, complementary selectivity to C18 and C8 bonded phases

Analysis of Isoflavones with the PerkinElmer Flexar FX-15 UHPLC System Equipped with a PDA Detector

International Journal of Pharma and Bio Sciences

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS)

Rapid Quantitative Analysis of Cannabidiol from Consumer Products Using UltraPerformance Convergence Chromatography

LC/MS Analysis of Various Hydrophilic Compounds Using a Polymer-Based Amino Column - Shodex TM HILICpak TM VG-50 2D

CYCLOSERINI CAPSULAE - CYCLOSERINE CAPSULES (AUGUST 2015)

TENOFOVIR TABLETS: Final text for addition to The International Pharmacopoeia (June 2010)

Comprehensive Two-Dimensional HPLC and Informative Data Processing for Pharmaceuticals and Lipids

Introduction. and aryl-α-amino ketones, 11

Improved CE/MS Sensitivity by Operating the Triple-Tube Coaxial Sheath-Flow Sprayer Without Appling Nebulizing Gas

Maximizing chromatographic peak capacity with the Agilent 1290 Infinity LC system

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Sepax Technologies, Inc.

Simultaneous determination of l- and d-lactic acid in plasma by capillary electrophoresis

CHAPTER 3 Amino Acids, Peptides, Proteins

Poster No. PT [ ANALYSIS OF ANION CONSTITUENTS OF URINE BY INORGANIC CAPILLARY ELECTROPHORESIS

A Validated Chiral Liquid Chromatographic Method for The Enantiomeric Separation of Dapoxetine Hydrochloride

Improved Pirkie Type 1 Amide Type: Asymmetric carbon atoms are bonded directly with CONH group

Metabolomics: quantifying the phenotype

Catalogue. Resins and Columns For High Performance Liquid Chromatography

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Enantioresolution of Amphetamine, Methamphetamine, and Deprenyl (Selegiline) by LC, GC, and CE

176 YMC Chiral Columns

The total protein test is a rough measure of all of the proteins in the plasma. Total protein measurements can reflect:

Chromatography on Immobilized Artificial Membrane

Internal Validation Guide of Y-STR Systems for Forensic Laboratories Printed in USA. 11/12 Part# GE713

BASIC ENZYMOLOGY 1.1

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

Isolation of five carotenoid compounds from tangerine tomatoes

John Haselden 1, Gordon Dear 1, Jennifer H. Granger 2, and Robert S. Plumb 2. 1GlaxoSmithKline, Ware, UK; 2 Waters Corporation, Milford, MA, USA

LC Columns with Liquid Separation Cell Technology

Simultaneous estimation of Metformin HCl and Sitagliptin in drug substance and drug products by RP-HPLC method

Analytical and Preparative SFC Columns

Characterization and application of single-isomer cyclodextrin derivatives

Systematic Evaluation of Solid Phase Extraction (SPE) Chemistries for the Determination of Acidic, Neutral, and Basic Drugs

Capillary electromigration techniques:

Chemical Nature of the Amino Acids. Table of a-amino Acids Found in Proteins

130327SCH4U_biochem April 09, 2013

Rapid Gradient and Elevated Temperature UHPLC of Flavonoids in Citrus Fruit

ZIDOVUDINE, LAMIVUDINE AND ABACAVIR TABLETS Draft proposal for The International Pharmacopoeia (September 2006)

Characterization of Purified MS2 Bacteriophage by the Physical Counting Methodology used in the Integrated Virus Detection System (IVDS)

Development, Estimation and Validation of Lisinopril in Bulk and its Pharmaceutical Formulation by HPLC Method

Institute of Physics and Chemistry, Chinese Academy of Sciences, Beijing, , P. R.

Comparison of Relative Quantification of Monoclonal Antibody N-glycans Using Fluorescence and MS Detection

Electronic Supporting Information

Stability of Vancomycin 31 mg/ml in Extemporaneous Eye Drops Determined with Capillary Electrophoresis

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Journal of Chemical and Pharmaceutical Research, 2018, 10(3): Research Article

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Residue Monograph prepared by the meeting of the Joint FAO/WHO Expert Committee on Food Additives (JECFA), 82 nd meeting 2016.

Molecules in the mirror

New Automated Screening System for the Determination of Basic Compounds in Urine by On-Line Extraction-HPLC- DAD

A CZE Method for the Identification of Therapeutic Antibodies and Quality Control of Infusion Bags at the Hospital

Rebaudioside a From Multiple Gene Donors Expressed in Yarrowia Lipolytica

Studies on Stationary Phase Selectivity for Solid-Core Particles

Micellar Electrokinetic

N-terminal charge-driven de novo sequencing by using ASDF-incorporated Curved Field Reflectron

The HPLC Preparative Scale-Up of Soybean Phospholipids Application

Agilent Anion-Exchange Media for Proteins - Loading vs Resolution - Effect of Flow Rate and Example Protein Separations

*, G. Manetto, Z. Deyl, F. Tagliaro, I. Miksık ˇ

Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page:

SYNAPT G2-S High Definition MS (HDMS) System

Chiral Screening for SFC and UHPLC with the Agilent 1260 Infinity II SFC/UHPLC Hybrid System

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE

Analysis of Common Sweeteners and Additives in Beverages with the PerkinElmer Flexar FX-15 System Equipped with a PDA Detector

Thermo Scientific. GlycanPac AXR-1. Column Product Manual. P/N: April, Part of Thermo Fisher Scientific

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

The effect of temperature and incubation time on the analysis of highly sialylated glycans from bovine fetuin

The Raptor HILIC-Si Column

CORESTA RECOMMENDED METHOD NÄ 9

CESI-MS of Intact Proteins

Transcription:

2 Chapman and Fu-Tai Chen Implementing a Generic Methods Development Strategy for Enantiomer Analysis Jeff Chapman and Fu-Tai A. Chen, Beckman Coulter Inc., Fullerton, California, USA. Introduction Whether one s approach to drug discovery involves natural products screening or the synthesis of innovative new compounds, the challenges presented to the analytical chemist are complicated when a product contains one or more chiral centres. Yet, these optically active compounds tend to make good drug candidates and, as such, require stereochemically selective assay methods in place for their analysis (1). An important step towards separating enantiomers is to first create diastereomers. Diastereomers may be created through chemical derivatization with a chiral reagent or may be formed transiently through interactions with chiral selectors. The latter approach is the most desirable because it is the simplest to employ. Using capillary electrophoresis (CE), a chiral environment may be easily created, by filling the capillary with an electrolyte solution containing a chiral additive. Although many types of chiral additives have been used in CE (2 4) the most comprehensive and effective separation strategies have been achieved with sulfated cyclodextrins (5 8). However, to implement an effective separation strategy, a clear definition and an understanding of the chiral selector used are essential. Although used generically in many publications the term sulfated cyclodextrins actually represents several different types of selectors supplied from different manufacturers. ur experience has been that these selectors all behave differently so it is essential to know which selector you are working with when specifying it in a method. In this article we describe an enantiomer methods development strategy using a family of highly sulfated cyclodextrins (HSCDs) to resolve a diverse array of compounds. Using this approach we successfully resolved 156/160 racemic mixtures tested (9) and propose this as a generic methods strategy to be considered as a first step in developing methods for separating enantiomers. Highly Sulfated Cyclodextrins In our experience, the most successful separations using HSCDs have been achieved at low ph. Under these conditions of suppressed electroosmotic flow (EF), the negatively charged cyclodextrins have strong electrophoretic mobility towards the positive electrode (anode). If the enantiomers interact with these HSCDs they will be swept towards the anode regardless of charge state. Neutral compounds interact hydrophobically with the cavity of the HSCDs, while basic compounds, being strong cations at low ph, interact with both the hydrophobic cavity and with the negatively charged sulfates. Many acidic compounds will be protonated at low ph and behave in a similar manner as neutral species, while zwitterionic analytes will be positively charged eliciting behaviour similar to basic compounds. Figure 1 illustrates the analysis of these four classes of enantiomers using a single analytical condition. The methods development strategy we propose here uses a family of HSCDs developed in-house. These reagents include the BCI, and HSCDs and comprise a controlled distribution of product containing an average of 11, 12 and 13 sulfates/molecules, respectively. Detailed characterizations of these reagents have been described elsewhere (4 7). Although there are trends developing that should allow future prediction of the best starting selector, we recommend implementing a strategy that uses the, and selectors independently to determine which one will provide the best resolution of an analyte. At low ph the EF in an untreated fused-silica capillary is minimal. In this instance the electrophoretic mobility

Chapman and Fu-Tai Chen 3 of an analyte (µ) in the presence of HSCD can be represented by a weighted sum of the mobility of free analyte µ F and that in the complexed state, µ C, as shown in equation 1. µ µ F K 1 (HSCD)µ C /1 K 1 (HSCD) [1] In this equation, the (HSCD) term is the concentration of free HSCD after interaction with the analyte. If this number is small, the mobility of the complex would be affected significantly by a change in analyte concentration. For this reason we recommend a starting concentration of the HSCD at 5%, (25 mm), as this should be one to two orders of magnitude greater (typically 0.1 to 1 mm) than the concentration of the chiral species being (a) (b) (c) (d) 1 6.0 4.0 3.0 9.0 4.0 separated. The excess amount of chiral selector over the chiral species makes the [free HSCD] virtually independent of variations in analyte concentration. As a greater population of analyte will be in the complexed state, a net higher mobility of analyte will result. The mobility of two enantiomers (racemates), 1 and 2, can be compared according to equation 2. µ 1 µ 2 {µ F1 K 1 (HSCD)µ C1 /1 K 1 (HSCD)} {µ F2 + K 2 (HSCD)µ C2 /1 K 2 (HSCD)} [2] Because HSCDs are highly negatively charged with substantially greater molecular weight than that of the analytes being separated, the mobility of the 10 H R 3.0 R 6.1 H R 6.1 R 2.3 Figure 1: Electropherograms of zwitterionic, acidic, basic and neutral chiral compounds using 5% -HSCD in 25 mm phosphate buffer ph 2.5 (a) carbidopa (zwitterionic drug), (b) ketoprofen (acidic drug), (c) ritalin (basic drug) and (d) warfarin (neutral drug). Conditions: untreated fused-silica capillary 50 µm 30 cm (20 cm to detector); capillary temperature: 22 C (recirculating liquid); field strength: 500 V/cm; detection: PDA 200 nm; sample introduction: pressure 1.2 psi*s; current generated: 156 µa. H H H N H H 3 C HN NH 2 complex can be approximated as µ C1 µ C2 and nearly equal to µ HSCD. Neutral enantiomers should satisfy the condition of µ F1 µ F2. Thus, equation 2 can be modified as equation 3: µ 1 µ 2 (µ C1 µ F1 )(K 1 K 2 ) (HSCD)/ [1 K 1 (HSCD)] [1+ K 2 (HSCD)] [3] For a neutral species, µ F1 µ F2 0, and for basic compounds, µ C1 >> µ F1 Thus, as in equation 4 µ 1 µ 2 µ HSCD (K 1 K 2 ) (HSCD)/ [1 K 1 (HSCD)] [1+ K 2 (HSCD)] [4] In principle, the mobility difference µ is intrinsically related to K 1 and K 2 and when [HSCD] is substantially large, a small variation of racemate concentration would not affect the total free (HSCD), thus the µ (resolution term) will remain essentially constant. This is what we have seen in practice, when maintaining a high concentration of chiral selector. In such instances the resolution term and migration time remain relatively constant even with changing analyte concentration. Although racemic mixtures may be baseline resolved at a calculated resolution greater than, detection of low-level enantiomers can be easily masked when the concentration of the eutomer (major enantiomer) is increased to the level necessary to detect the lower concentration enantiomer. However, as resolution values of 5 or greater are commonly obtained using the HSCD approach, low-level enantiomer quantification below 0.1% levels can be readily achieved. Figure 2 illustrates the detection of ( ) phenylpropanolamine enantiomer at 0.1%. The good resolution afforded by this assay makes this formerly difficult task routine. Analyte Detection During the methods development phase, photodiode array (PDA) detection can prove to be a valuable tool in removing the ambiguity of enantiomer assignment. As enantiomers are physicochemically identical, they will have matching spectra, allowing discrimination of enantiomer pairs from contaminating analytes. Although most drug candidates are ultraviolet (UV)-absorbing, there may be instances in which the chiral compound of interest does not have significant UV absorbance. In these examples, one can employ a preseparation derivatization to render either a UV-absorbing or a fluorescing product. An example is shown

4 Chapman and Fu-Tai Chen in Figure 3 with the analysis of homoproline. Shown is the separation of dansylated homoproline enantiomers, both as a racemic mixture and the single enantiomer preparation. With the determination of low-level Absorbance units Absorbance units 10 3 0.014 0.01 0.006 0.002 10.0 5.0 0.0 15.0 10.0 5.0 0.0 Racemates enantiomers, the linearity of detection is also an important consideration. Figure 4 illustrates the detection response linearity for the phenylpropanolamine enantiomers described above. This plot represents the percent impurity (lower level enantiomer) 0.002 0.0 0.2 0.4 0.6 0.8 1.1 1.3 1.5 1.7 1.9 2.1 2.3 2.6 2.8 3.0 3.2 3.4 3.6 3.8 Figure 2: Electropherogram of phenylpropanolamine enantiomers, with the ( ) form readily detected at 0.1%. Conditions: buffer: 5% -HSCD in 25 mm phosphate, ph 2.5; capillary: untreated fused-silica capillary 50 µm 30 cm (20 cm to detector); capillary temperature: 22 C (recirculating liquid); field strength: ( )500 V/cm (anode at outlet); detection: PDA 200 nm; sample introduction: pressure 1.2 psi*s; current generated: 152 µa. (a) (b) 0.1% ( ) R 2.9 Pure enantiomer Reagent peak Reagent peak 99.9% ( ) Figure 3: Electropherogram of dansylated derivatives of homoproline. (a) Racemic mixture of homoproline, (b) L-homoproline enantiomer. Conditions: buffer: 5% -HSCD in 25 mm phosphate, ph 2.5; capillary: untreated fused-silica capillary 50 µm 30 cm (20 cm to detector); capillary temperature: 22 C (recirculating liquid); field strength: ( )500 V/cm (anode at outlet); detection: PDA 200 nm; sample introduction: pressure 1.2 psi* s; current generated: 155 µa. as a function of the ratio of velocity corrected peak area of the impurity/total. A good linear response is generated allowing an accurate determination of enantiomer impurity even at levels as low as 0.1%. Capillary Selection and Conditioning Within this methods development strategy, we recommend using a 50 µm bare fusedsilica capillary with a relatively short effective length (10 or 20 cm to detector), allowing you to operate under high field strengths, while still maximizing mass load. Both the speed and resolution of this approach are highlighted in Figure 5 where dansylated phenylalanine enantiomers are well resolved in less than 1 min. It is notable that a resolution value of 3 could be generated in such a rapid manner. However, this really has become a trademark for this type of analysis. As the capillaries are operated at low ph it is best not to use NaH rinses between runs. In fact, we typically only rinse with run buffer prior to each analysis. To reduce the daily equilibration cycle we recommend rinsing at least once a day with a polyethylene oxide solution (25 mm lithium acetate, 0.4% PE (MW 300 000), 10% ethylene glycol, adjusted to ph 4.75). This polymeric solution acts to coat the capillary surface and in our experience has reduced the need for the traditional daily start-up equilibration runs. In this manner, reproducible migration times are established quickly. However, should a capillary become excessively fouled, treatment with a strong acid, followed by the polymer solution, followed by a run buffer is recommended. Dedicating capillaries for this assay will also be of benefit; although a capillary may be quickly exchanged with different buffers, the surfaces do take time to become uniform, especially when different buffer ions are used. Electrophoresis Parameters For the purposes of comparative analysis we recommend that the electrophoresis conditions used for the initial screening runs are standardized. nce the best cyclodextrin type and concentration are determined the process of optimization can be started. The starting conditions that we have chosen are as follows: Capillary temperature 22 C** Field strength 500 V/cm (generating currents of approximately 150 µa) Detection: PDA UV 200 nm (Scanning 190 320 nm) Sample introduction 1.2 psi*s (4 s at 0.3 psi). **The capillary temperature represents

Chapman and Fu-Tai Chen 5 that of the environment used to thermoregulate the capillary. When transferring methodology between instruments using different thermoregulatory mechanisms it is useful to use the run current as a diagnostic to assess differences in actual capillary temperature. In this way you can fine-tune your capillary temperature by lowering or elevating your system temperature until a similar current is achieved. As these reagents are viscous, simple steps, such as water dips after both sample introduction and the end of the run, can be used to prevent sample carryover/ contamination. Another important variable to pay attention to is the total run time programmed. As the resolution of enantiomer pairs can be quite high, a failed separation attempt can easily be mistaken by simply not allowing enough separation time to resolve the two enantiomers. In Figure 6 we illustrate the analysis of amphetamine enantiomers for which such an oversight could easily be made. Notice the elapsed run time from the first enantiomer to the second. If the run were aborted too soon, the assumption could have been made that no separation was achieved. A good experiment to consider is to run your sample with and without the chiral selector. If complete resolution is achieved, the peak heights will be approximately half that of the control where no selectors were present. Alternatively, by keeping the capillary as short (10 cm) as possible the total run time required is significantly reduced. high, forcing a stronger interaction than desired. In this instance, lowering the concentration is advised. An HSCD dilution series of 1 to 5% can rapidly provide you with a standard curve to estimate the best concentration of selector to choose. If the answer to the above question was no (i.e., that there was no change in peak shape), the HSCD concentration should be increased in a series from 5 to 10%. Corrected Peak Area Ratio ( )/Total 0.1 0.08 0.06 0.04 0.02 0 0 Automation of this optimization process can be readily achieved with the introduction of a buffer array. Results Using only the primary screen of, and HSCD, and the standard method described above, we have been able to resolve (R > ) 156/160 drugs tested, with no optimization process employed. % ( ) PPA Figure 4: Plot of the ratio of the impurity velocity corrected area/total velocity corrected peak area versus the % contaminating enantiomer (impurity) for phenylpropanolamine. A linear response from 0.1% to 5% is readily attained, allowing effective quantification of the lower level enantiomer to levels as low as 0.1%. Methods ptimization nce the primary three runs, evaluating the, and HSCDs, have been performed, an optimization process may be implemented. First, choose the HSCD that gave the best resolution. You may wish to vary the concentration between 2% and 10% to evaluate whether any enhancement of resolution is achieved. Capillary temperature, length and field strength may all be used to fine-tune the resultant method. However, it is preferable that the field strength be optimized last, once you have decided upon the optimum capillary length and temperature. If resolution of enantiomers is not achieved with any of the three cyclodextrins, attention should be focused first on the peak shape. Did any of the runs illustrate a broadening of peak shape? If the answer is yes, there is a good chance that the HSCD concentration is too Absorbance (AU, 10 4 ) 3 1 R 3.2 0 30 60 90 120 Time (s) Figure 5: Electropherogram of dansylated-phenylalanine enantiomers separated using a short capillary (10 cm sample introduction from the outlet side). Conditions: buffer: 5% -HSCD in 25 mm phosphate, ph 2.5; capillary: untreated fused-silica capillary 50 µm 30 cm (10 cm to detector); capillary temperature: 22 C (recirculating liquid); field strength: ( )500 V/cm (cathode at outlet); detection: PDA 200 nm; sample introduction: pressure 1.2 psi*s sample introduced from outlet side; current generated: 157 µa.

6 Chapman and Fu-Tai Chen Absorbance (AU, 10 4 ) R 23.4 5 10 Figure 6: Electropherogram of amphetamine enantiomers demonstrating the high resolution achieved with highly sulfated cyclodextrins. Conditions: buffer: 5% -HSCD in 25 mm phosphate, ph 2.5; capillary: untreated fused-silica capillary 50 µm 30 cm (20 cm to detector); capillary temperature: 22 C (recirculating liquid); field strength: ( ) 500 V/cm (anode at outlet); detection: PDA 200 nm; sample introduction: pressure 1.2 psi*s; current generated: 156 µa. The compounds tested included 56/58 neutral drugs 92/92 basic drugs 8/10 acidic drugs. A library of these drugs, their structures and the separations achieved have been published and can be reviewed on the Internet at www.beckmancoulter.com/chiral38 In summary, we have presented a powerful chiral analysis strategy using a combination of capillary electrophoresis and a family of highly sulfated cyclodextrins. The result is, in our opinion, a sound enantiomer analysis strategy that uses a single generic method approach to resolve a broad array of compounds. An approach we suggest you try first, when developing methodology for enantiomer separations. Coulter Inc., Fullerton, California, USA, and Albert Fu-Tai Chen is a senior scientist, also at Beckman Coulter. References: (1) http://www.fda.gov/cder/guidance/stereo.htm (2) K. Verleysen and P. Sandra, Electrophoresis, 19 (16 17), 2798 2833 (1998). (3) R. Vespalec and P. Bocek, Electrophoresis, 15(6), 755 762 (1994). (4) S. Fanali, J. Chromatogr. A, 875(1 2), 89 122 (2000). (5) F.-T.A. Chen and R.A. Evangelista, J. Chinese Chemical Society, 46, 847 860 (1999). (6) I.S. Lurie et al., Electrophoresis, 19, 2918 2925 (1998). (7) K. Verleysen, T. Van den Bosch and P. Sandra, Electrophoresis, 20, 2650 2655 (1999). (8) F.-T.A. Chen, G. Shen and R.A. Evangelista, Chirality, (Submitted 2000). (9) www.beckmancoulter.com/chiral38 Jeff Chapman is the Strategic Market Manager for CE technology at Beckman