Case Report Exactech Opteon Femoral Component Fracture 12 Years after Arthroplasty

Similar documents
Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs: Report of Two Cases

Case Report Bilateral Distal Femoral Nailing in a Rare Symmetrical Periprosthetic Knee Fracture

CAUTION: Ceramic liners are not approved for use in the United States.

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

ESC. Enhanced Stability Liners. Design Rationale & Surgical Technique

Case Report Osteolysis of the Greater Trochanter Caused by a Foreign Body Granuloma Associated with the Ethibond Suture after Total Hip Arthroplasty

Aesculap Trilliance Triple Tapered Polished Hip Stem

28 Surgical Technique

Optimum implant geometry

Optimum implant geometry

operative technique Kent Hip

The proven, simple solution. CPT 12/14 Hip System

Optimum implant geometry

Case Report Pediatric Transepiphyseal Seperation and Dislocation of the Femoral Head

Case Report Patellofemoral Joint Replacement and Nickel Allergy: An Unusual Presentation

VerSys LD/Fx Cemented and Press-Fit Hip Prostheses. Surgical Technique IMAGE TO COME. Versatile solutions for total and partial hip replacement

Bone Bangalore

TaperFit. Cemented Total Hip Replacement Surgical technique

Enhanced Stability Constrained Liners. Design Rationale Surgical Technique

Hip Resurfacing System

Tri-Lock Bone Preservation Stem

Furlong H-A.C. THR System

ACETABULAR CUP SURGICAL TECHNIQUE

Duraloc CONSTRAINED LINER

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Featuring. Technology. Product Rationale

Efficacy Innovation ABG II. Brochure. Surgical Protocol. Cemented Stem

Bone Preservation Stem

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

Segmental Fracture Of A Cemented Femoral Stem - A Case Report And Review Of Litrature

Templating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5.

This publication is not intended for distribution in the USA.

Minimally Invasive System for Total Hip Arthroplasty. Surgical Technique

DIRECT ANTERIOR APPROACH. Guide for use with the Furlong Evolution Femoral Stem & CSF Plus Acetabular Cup

EXTENDED TROCHANTERIC OSTEOTOMY SURGICAL TECHNIQUE FPO EXTENSIVELY COATED FIXATION

Case Report Multiple Giant Cell Tumors of Tendon Sheath Found within a Single Digit of a 9-Year-Old

Progeny Hip Stem. Surgical Protocol and Product Specifications

Biomet Large Cannulated Screw System

CPCS. Training Manual. Collarless Polished Cemented Stem

This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

*smith&nephew CONTOUR

PRODUCT SCHEDULE FOR CERTIFICATE CE JRI ORTHOPAEDICS LIMITED

Constrained Acetabular Insert

Fracture through a femoral stem prosthesis after trauma: A case report and review of the literature

Ceramic Femoral Heads and Acetabular Cup Liners

Optimizing function Maximizing survivorship Accelerating recovery

1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe

Surgical procedure. METS Modular Proximal Femur

The Leader in Orthopaedic Innovation

21st Century Fracture Management ETS. Surgical Protocol

Case Report An Undescribed Monteggia Type 3 Equivalent Lesion: Lateral Dislocation of Radial Head with Both-Bone Forearm Fracture

Continuing the Tradition. VerSys Heritage Hip System

Rx90 Total Hip System Acetabular Series

Case Report Intra-Articular Entrapment of the Medial Epicondyle following a Traumatic Fracture Dislocation of the Elbow in an Adult

SURGICAL TECHNIQUE. Protrusio Cage A COMPREHENSIVE ACETABULAR REVISION SYSTEM

Total Hip Replacement. Find out why the Anterior Approach may be right for you.

ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY

DURALOC. Acetabular Cup System. Surgical Technique

PLR. Proximal Loading Revision Hip System

CASE REPORT. Bone transport utilizing the PRECICE Intramedullary Nail for an infected nonunion in the distal femur

ANTERIOR TOTAL HIP ARTHOPLASTY

JRI Thompson Hemiarthroplasty

DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574)

LPS SYSTEM POCKET GUIDE

Baris Beytullah Koc, 1 Martijn Schotanus, 1 Bob Jong, 2 and Pieter Tilman Introduction. 2. Case Presentation

Arcos Modular Femoral Revision System

Avenir Femoral Hip System

Clinical Evaluation Surgical Technique

Following a tradition of success. VerSys Heritage Primary Hip Prosthesis Surgical Technique

CAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician.

9800 Metric Blvd. Austin, Texas

Bi-Polar 22.2mm & 28mm System - Operative technique

Tradition Hip Primary Surgical Technique

Case Report Combined Effect of a Locking Plate and Teriparatide for Incomplete Atypical Femoral Fracture: Two Case Reports of Curved Femurs

AVANTEON. Operative Technique & Catalogue Information AVANTEON

HOSPITAL "A" (2007) ORTHOPEDIC JOINT IMPLANT COMPONENT MATRIX

Total Knee Original System Primary Surgical Technique

OPTIONAL/ADDITIONAL INSTRUMENTS

Trilogy Acetabular System

Nuclear medicine and Prosthetic Joint Infections

Hip Replacement - Anterior

US Patent No. 6,309,396B1. G. David Ritland, M.D.

DIRECT ANTERIOR APPROACH

Trilogy Acetabular System

Design Rationale. ECHELON Primary Hip System

Clinical Study Rate of Improvement following Volar Plate Open Reduction and Internal Fixation of Distal Radius Fractures

PediNail Pediatric Femoral Nail

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

Arcos Interlocking Distal Stem. Surgical Technique Addendum to the Arcos Modular Femoral Revision System

Restoration Modular Revision Hip System. Choice Matters

Case Report Inflammatory Pseudotumor Complicated by Recurrent Dislocations after Revision Total Hip Arthroplasty

Case Report Successful Closed Reduction of a Lateral Elbow Dislocation

Surgical Technique r5.indd 1 12/8/10 10:36 AM

Peggers Super Summaries: THR and Revision Concepts

CORAIL HIP SYSTEM SURGICAL TECHNIQUE

AUTOBLOQUANTE AUTOBLOQUANTE. Product Rationale Surgical Technique

EC Certification. FULL QUALITY ASSURANCE SYSTEM Directive 93/42/EEC for Medical Devices, Annex II excluding (4)

VERSYS HERITAGE REVISION HIP PROSTHESIS. Surgical Technique for Revision Hip Arthroplasty

Transcription:

Case Reports in Orthopedics Volume 2016, Article ID 1789197, 4 pages http://dx.doi.org/10.1155/2016/1789197 Case Report Exactech Opteon Femoral Component Fracture 12 Years after Arthroplasty Shaun P. Patel, Valentin Antoci Jr., John J. Kadzielski, and Mark S. Vrahas DepartmentofOrthopaedicSurgery,MassachusettsGeneralHospital,Boston,MA02114,USA Correspondence should be addressed to Shaun P. Patel; shaun.p.patel@gmail.com Received 18 September 2015; Revised 11 January 2016; Accepted 13 January 2016 Academic Editor: Elke R. Ahlmann Copyright 2016 Shaun P. Patel et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Arthroplasty implant fracture is a rare but critical complication that requires difficult revision surgery, often with poor results, patient disability, and significant cost. Several reports show component fracture either at the stem or at the neck interface after a relatively short postoperative course. We report such failure after 12 years, suggesting no safe period after which femoral implant fracture does not occur. 1. Introduction Arthroplastyhasbeenrankedinthetop10listofmedical innovations, improving quality of life and prolonging activity into old age [1]. However, complications such as femoral component failure, even though rare, can have significant sequelae, including death. A retrospective survey performed by the American Association of Hip and Knee Surgeons estimated the prevalence of femoral stem fractures to be 0.27% [2]. Previous case reports have documented the failure of femoral neck and stems of JRI-Furlong hydroxyapatiteceramic coated femoral prostheses (Joint Replacement Instrumentation Limited, London, UK) [3]. Failures occurred in patients with a BMI over 25 and were attributed to fatigue secondary to high stress over time. Another case report series described failures of the Omnifit cobalt chrome femoral component (Osteonics, Allendale, NJ) occurring in the femoral neck adjacent to the modular head. Macroscopic and microstructural analysis revealed that the cause of failure had been initiated by intergranular corrosion, crack growth through fatigue, and resultant failure from tensile stress. An annealing step that would prevent intergranular carbide precipitation could prevent such failure and would represent a viable solution to repeat occurrence [4]. We now report on the failure of the Exactech Opteon femoral component (Exactech, Gainesville, FL) that may have subsequently led to death. A previous case report series documented failure of these components at an average of 43 months, with early fatigue and subsequent fracture thought to be resulting from laser etching and marking of the metal [5]. Interestingly, unlike previously reported, the component failure we observed occurred much later at 12 years after implantation. 2. Case Presentation An 82-year-old female, with a BMI of 33.9, was transferred to our hospital from an outside facility with a diagnosis of a failedlefthipprosthesis.theimplanthadfracturedinthearea of the femoral neck adjacent to the modular head (Figure 1). The patient had undergone left total hip arthroplasty twelve years prior to presentation for osteoarthritis. The femoral component was a cemented Exactech Opteon stem, size 2, with a 28 mm, +5 mm femoral head. The acetabular component was a 54 mm Duraloc cup with an Enduron polyethylene liner (DePuy Inc., Warsaw, IN). The patient had a total knee arthroplasty performed in the ipsilateral extremity three years prior to the total hip arthroplasty also for osteoarthritis. Three months after the total hip arthroplasty, the patient developed a Staphylococcus epidermidis periprosthetic infection in her knee and underwent exchange arthroplasty later that same year. Only the total knee implant had been affected and revised, whereas the total hip implant

2 Case Reports in Orthopedics Figure 1: Admission radiographs show femoral component fracture at the neck-shoulder junction at 12 years after implantation. was unaffected. The patient subsequently underwent successful rehabilitation and had been an independent community ambulator thereafter. Upon presentation, the patient denied shortness of breath, chest pain, fevers, chills, night sweats, or recent travel. Five month prior to presentation, the patient had dental work with appropriate antibiotic prophylaxis. The critical event most likely occurred several days prior to presentation whenthepatientslippedandfellinthebathroom.plain radiographs revealed a fracture of the femoral component at the area of the femoral neck adjacent to the modular head (Figure 1). The stem appeared well seated with no evidence of infection, loosening, malalignment, or periprosthetic fracture. Examination and imaging of the ipsilateral knee were unremarkable. Surgery was subsequently planned for component retrieval and revision arthroplasty. Under general anesthesia, the patient was placed in the lateral decubitus position. The previous incision was opened, the subcutaneous tissues were dissected, and the plane beneath the tensor fascia lata was elevated. The sciatic nerve was identified and noted to be in its normal position. The short external rotators were scarred and adhered to the joint capsule. This layer was dissected as a full thickness flap. The quadratus femoris muscle appeared healthy distally. The capsule was then incised, revealing a small amount of serous fluid within the joint. The posterior capsule was removed and the remainder of the anterior rim of the cup was exposed. Tissue samples were sent for frozen section. The top of the fractured femoral stem was then located (Figure 2(a)). The soft tissues were removed from the area of thegreatertrochanterandthecementaroundtheposterior shoulderandaroundthecollaroftheprosthesiswasremoved. A carbide punch was used to drive the prosthesis out of the cement mantle. Examination of the cement mantle after removal of the prosthesis showed a small crack proximally which was suspected to proceed into the calcar. To stabilize this crack, a 16-gauge doubled wire was passed around the femur, proximal to the lesser trochanter, and tightened into place (Figure 2(b)). The remainder of the capsule was removed around the rim of the acetabulum. There was some suspicion for an infection based on frozen sections sent earlier and tissue cultures were therefore taken. Once the rim of the cup had been exposed, a drill hole was placed into the cup and a screw was used to assist with removal of the polyethylene liner. As documented previously, this was a Duraloc cup. The old locking ring was removed and the acetabulum was then irrigated with bacitracin solution. A new locking ring was placed in the well-fixed Duraloc cup and a new polyethylene liner was placed. A Duraloc Marathon cup acetabular liner was used with a 32 mm inner diameter and a 54 mm outer diameter with a neutral face (Figure 2(b)). The femoral canal was irrigated and dried. The cement mantle was intact. A size 9 Heritage stem was then cemented in place. Once the cement was hardened, the hip was trialed. The hip pistoned in an unacceptable amount even with a 10.5 mm neck length with a 32 mm head. Cement was therefore removed from around the top of the prosthesis and the prosthesis was driven out of the cement mantle without difficulty. Cement was then again placed within the previous cement mantle and the prosthesis was cemented one centimeter higher than the previous attempt. With the cement hardened, the prosthesis was trialed again and the hip was stable with a +7 mm neck (Figure 2(c)). The wound was then copiously irrigated. The quadratus femoris and short external rotators were reapproximated to the greater trochanter along with some capsule to provide hipstability.thefascialatawasclosedandtheremainderof the wound was closed in layers. The implant was examined closely for signs of mechanical or chemical wear (Figure 3). No specific factor was identified in the femoral stem fracture. Intraoperative cultures ultimately returned negative. 3. Discussion Early analysis of femoral component failure after total hip arthroplasty cited a number of possible causes including design configuration, improper metallic alloy, quality of cement, surgical technique, undersizing of prosthesis, and patient characteristics [6 9]. Intergranular corrosion has beenshowntocausefailureincobaltalloyfemoralstems in two different case report series [4, 10]. Implant handling during implantation is also important and may result in notches or other material defects that can weaken the implant structure and ultimately lead to failure. In our current case, we did not observe any specific notching, although we cannot confirm that such a defect was not present before the implant fracture. As the obesity epidemic propagates through the arthroplasty population, it raises the question of mechanical overload of implant materials and their long term survival, particularly in conjunction with smaller size implants which may be necessary in these patients. Although the American Academy of Orthopaedic Surgery (AAOS) recommends a

Case Reports in Orthopedics 3 (a) (b) (c) Figure 2: (a) Intraoperative femoral component fracture and separation. (b) A new locking ring and liner were placed in the preexisting well-fixed Duraloc cup. (c) Postoperative imaging confirms stable implant positioning. Figure 3: Implant examination from different perspectives shows no gross mechanical or chemical degradation at the fracture site. BMI cut-off of 40, this is still a morbidly obese population and such a body habitus may overly stress the smaller size implants and lead to implant fracture. One such case report series noted that failure of femoral components occurred in patients with a BMI greater than normal [3], while another study found no significant difference in the complication rate between obese and nonobese patients [11]. In addition to increased BMI as a potential risk factor for implant failure, implant and material analysis studies have demonstrated decreased failure loads with increased loading cycles. In particular, Bergmann et al. have even shown that peak forces generated while stumbling, as occurred in this case, can be nearly three times as high as those generated by walking [12]. Implant design and postmanufacturing processing are also crucial factors that determine the success of the implant. In a six-year period at one institution, 113 primary and revision total hip arthroplasties were performed using the Exactech Opteon femoral component. Two cases were found to have undergone early fatigue fractures secondary to laser etching at the neck-shoulder junction. Similarly, another case report series also noted failure of femoral components associated with laser etching of Precoat femoral component prostheses (Precoat and Precoat Plus; Zimmer, Warsaw, IN) [9]. The Exactech manufacturer had recalled all stems that had undergone laser etching in this area and no other reports of implant failure have since surfaced [5]. We have closely analyzed the implant and fracture surfaces in this report and the failure seems to be similarly related to the laser

4 Case Reports in Orthopedics etching at the interface, which, in turn, created a stress riser and fatigue fracture. It is important to note, however, that patients in previous studies showed failure much earlier, 43 months on average, while we currently report on the possibility of fracture much later, even after 12 years of normal use. In this case report, we show a unique circumstance of implant failure over 12 years after implantation. The primary cause of implant failure was likely the laser etching at the fracture interface, though the patient also had risk factors of an elevated BMI and increased implant loading cycles which could contribute to implant failure. Even though revision arthroplasty is a frequent and appropriate occurrence in large volume institutions, it is important to remember that repeat surgery has associated risks, many of which are significantly increased compared to the index procedure. Our patient ultimately had a technically successful revision surgery. However, due to the immobilization in the perioperative period, and despite appropriate deep venous thrombosis prophylaxis, thepatientdevelopedamassivepulmonaryembolismatthe postoperative rehabilitation facility, unfortunately resulting in death. Consent The authors have obtained informed written consent from the patient s next of kin for the publication of this case report. cases, The Bone & Joint Surgery American Volume, vol.63,no.7,pp.1078 1094,1981. [8] L.M.Jazrawi,C.J.DellaValle,F.J.Kummer,E.M.Adler,and P. E. Di Cesare, Catastrophic failure of a cemented, collarless, polished, tapered cobalt-chromium femoral stem used with impaction bone-grafting, The Bone & Joint Surgery American Volume,vol.81,no.6,pp.844 847,1999. [9]S.T.Woolson,J.P.Milbauer,J.D.Bobyn,S.Yue,andW. J. Maloney, Fatigue fracture of a forged cobalt-chromiummolybdenum femoral component inserted with cement. A report of ten cases, The Bone & Joint Surgery American Volume,vol.79,no.12,pp.1842 1848,1997. [10] J. L. Gilbert, C. A. Buckley, J. J. Jacobs, K. C. Bertin, and M. R. Zernich, Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants, The Journal of Bone & Joint Surgery American Volume, vol.76,no.1,pp. 110 115, 1994. [11] J. R. McLaughlin and K. R. Lee, The outcome of total hip replacement in obese and non-obese patients at 10- to 18-years, The Bone & Joint Surgery British Volume,vol.88,no. 10, pp. 1286 1292, 2006. [12] G. Bergmann, F. Graichen, A. Rohlmann et al., Realistic loads for testing hip implants, Bio-Medical Materials and Engineering,vol.20,no.2,pp.65 75,2010. Conflict of Interests The authors declare that there is no conflict of interests regarding the publication of this paper. References [1] V. R. Fuchs and H. C. Sox Jr., Physicians views of the relative importance of thirty medical innovations, Health Affairs, vol. 20, no. 5, pp. 30 42, 2001. [2] D.A.Heck,C.M.Partridge,J.D.Reuben,W.L.Lanzer,C.G. Lewis, and E. Michael Keating, Prosthetic component failures in hip arthroplasty surgery, Arthroplasty, vol. 10, no. 5, pp. 575 580, 1995. [3] P. Harvie, M. Haroon, N. Henderson, and M. El-Guindi, Fracture of the hydroxyapatite-ceramic-coated JRI-Furlong femoral component: body mass index and implications for selection of the implant, The Bone & Joint Surgery British Volume,vol.89,no.6,pp.742 745,2007. [4] L.-O. Lam, K. Stoffel, A. Kop, and E. Swarts, Catastrophic failure of 4 cobalt-alloy Omnifit hip arthroplasty femoral components, Acta Orthopaedica,vol.79,no.1,pp. 18 21,2008. [5]E.W.LeeandH.T.Kim, Earlyfatiguefailuresofcemented, forged, cobalt-chromium femoral stems at the neck-shoulder junction, Arthroplasty, vol. 16, no. 2, pp. 236 238, 2001. [6] J. O. Galante, Causes of fractures of the femoral component in total hip replacement, TheJournalofBone&JointSurgery American Volume,vol.62,no.4,pp.670 673,1980. [7]E.Y.S.ChaoandM.B.Coventry, Fractureofthefemoral component after total hip replacement. An analysis of fifty-eight

MEDIATORS of INFLAMMATION The Scientific World Journal Gastroenterology Research and Practice Diabetes Research International Endocrinology Immunology Research Disease Markers Submit your manuscripts at BioMed Research International PPAR Research Obesity Ophthalmology Evidence-Based Complementary and Alternative Medicine Stem Cells International Oncology Parkinson s Disease Computational and Mathematical Methods in Medicine AIDS Behavioural Neurology Research and Treatment Oxidative Medicine and Cellular Longevity