Comparative evaluation of Ambu AuraGain with ProSeal laryngeal mask airway in patients undergoing laparoscopic cholecystectomy

Similar documents
I - Gel Versus Cuffed Tracheal Tube in Elective Laparoscopic Cholecystectomy A Clinical Comparative Study

LMA Supreme Second Seal. Maintain the airway. Manage gastric contents. Meet NAP4 recommendations.

A randomised comparison between Cobra PLA and classic laryngeal mask airway and laryngeal tube during mechanical ventilation for general anaesthesia

Citation British journal of anaesthesia, 104. pp ; 2010 is available onlin

ISSN X (Print) Research Article. *Corresponding author Dr. Souvik Saha

Comparison of clinical performance of i-gel with laryngeal mask airway pro-seal in elective surgery in adults

I-gel vs cuffed tracheal tube during volume controlled ventilation in elective laparoscopic cholecystectomy

Dr. Ranjeet Rana De 1, Dr. Saurav Shekhar 2, Dr. D G Pathak 3, Dr. Harshwardhan 4, Dr. Shashank Dhiraj 5 1,2,4,5

International Journal of Health Sciences and Research ISSN:

A Comparative Study of Classic LMA and Proseal LMA in Paralyzed Anaesthetized Patients

Original Contributions

British Journal of Anaesthesia 82 (5): (1999)

ORIGINAL ARTICLE. Shashank Chitmulwar, MD, Charulata Deshpande, MD, DA ABSTRACT. ANAESTHESIA, PAIN & INTENSIVE CARE

J of Evolution of Med and Dent Sci/ eissn , pissn / Vol. 3/ Issue 64/Nov24, 2014 Page 13923

Evaluation of Baska Mask Performance in Laparoscopic Cholecystectomy

The Laryngeal Mask and Other Supraglottic Airways: Application to Clinical Airway Management

Veena Patodi*, Maina Singh, Surendra K. Sethi, Vini N. Depal, Neena Jain, Vijay Kumar

The LMA CTrach TM, a new laryngeal mask airway for endotracheal intubation under vision: evaluation in 100 patients

Use of the Intubating Laryngeal Mask Airway

Tibe ProSeal Laryngeal Mask Airway

LMA Unique Airway Portfolio

Received 1 November 2009 Revised 8 December 2009 Accepted 10 December 2009

MD (Anaesthesiology) Title (Plan of Thesis) (Session )

Recent Advances in Airway Management HA Convention 2014

Randomised comparison of the LMA Supreme with the I-Gel in spontaneously breathing anaesthetised adult patients

Original Research Article. Amol P. Singam 1, Arpita A. Jaiswal 2 *, Ashok R. Chaudhari 1

Deposited on: 11 February 2010

ISPUB.COM. M Roberts, M Mani, A Wilkes, E Flavell, N Goodwin INTRODUCTION

Laryngeal mask airway (LMA-ProSeal) malfunction causing acute airway obstruction

Comparison of the Hemodynamic Responses with. with LMA vs Endotracheal Intubation

Comparison of the Berman Intubating Airway and the Williams Airway Intubator for fibreoptic orotracheal intubation in anaesthetised patients.

Comparison of Ease of Insertion and Hemodynamic Response to Lma with Propofol and Thiopentone.

University of Groningen

Other methods for maintaining the airway (not definitive airway as still unprotected):

Unanticipated difficult tracheal intubation - during routine induction of anaesthesia in an adult patient

A Comparative Study of Two Disposable Supraglottic Devices in Diagnostic Laparoscopy in Gynecology

Comparative study of various supraglottic devices with clinical and fiber optic assessment in elective laparoscopic procedures

COMPARATIVE CLINICAL EVALUATION OF PROSEAL LMA WITH I - GEL AIRWAY IN PATIENTS UNDERGOING ELECTIVE SURGICAL PROCEDURES UNDER GENERAL ANAESTHESIA

Comparison of I-gel with Baska Mask Airway for Controlled Ventilation in Obese Patients Undergoing Ambulatory Surgery: A Prospective Randomized Trial

Anatomy and Physiology. The airways can be divided in to parts namely: The upper airway. The lower airway.

A cohort evaluation of clinical use and performance characteristics of Ambu AuraGain : A prospective observational study

DIFFICULT AIRWAY MANAGMENT. Dr.N.SANTHOSH KUMAR MD ANESTHESIA (2 nd Yr)

Optimal Size AMBU Laryngeal Mask Airway Among Asian Adult Population

Comparison of efficacy of the Laryngeal tube with the Laryngeal mask airway in securing the upper airway

We will not be using the King LTS-D in our system!

The Pro-Seal LMAtm And The Tracheal Tube: A Comparison Of Events At Insertion Of The Airway Device

Materials and Methods

EUROANESTHESIA 2007 Munich, Germany, 9-12 June 2007

Veena Mathur, Deepak Garg, Neena Jain, Vivek Singhal, Arvind Khare, Surendra K. Sethi*

The Laryngeal Tube. An Evaluation of the Laryngeal Tube During General Anesthesia Using Mechanical Ventilation

Cricoid pressure: useful or dangerous?

LMA CTrach TM in patients with anticipated difficult airway: A retrospective study

J. Brimacombe, 1 L. Holyoake, 2 C. Keller, 3 J. Barry, 4 D. Mecklem, 4 A. Blinco 5 and K. Weidmann 5

Pharyngolaryngeal Morbidity with the Laryngeal Mask Airway in Spontaneously Breathing Patients

CONTINUING EDUCATION IN HONOR OF NORMAN TRIEGER, DMD, MD

P V Praveen Kumar 1*, P. Archana 2. Original Research Article. Abstract

Airway/Breathing. Chapter 5

REVISTA BRASILEIRA DE ANESTESIOLOGIA

A Clinical Comparative Study Of Evaluation Of Proseal LMA V/S I-GEL For Ease Of Insertion And Hemodynamic Stability; A Study Of 60 Cases

Nitrous oxide diffusion into the cuffs of disposable laryngeal mask airways

Observation of ventilation effects of I-gel TM, Supreme TM and Ambu AuraOnce TM with respiratory dynamics monitoring in small children

Journal of Anesthesia & Clinical

Asai and Shingu Table 1 Size selection and recommended cuff volumes Size Patients Body size Recommended cuff volumes (ml) Connector colour 0 Newborn <

Success of Tracheal Intubation with Intubating Laryngeal Mask Airways

LMA for Laparoscopic Surgery

Original Article INTRODUCTION. Abstract

All I need is an LMA

Comparison of the air-q ILA and the LMA-Fastrach in airway management during general anaesthesia

A comparative study between i-gel and classical laryngeal mask airway in elective surgery under general anaesthesia

Comparison of hemodynamic and metabolic stress responses caused by endotracheal tube and Proseal laryngeal mask airway in laparoscopic cholecystectomy

The effect of head rotation on efficiency of ventilation and cuff pressure using the PLMA in pediatric patients

ORIGINAL RESEARCH ARTICLE

The flexible laryngeal mask airway (FLMA) was

Airway Management. Teeradej Kuptanon, MD

Comparison of intracuff dexamethasone, lignocaine and normal saline on post extubation response - A prospective study

JMSCR Vol 4 Issue 02 Page February 2016

Airway Management. Key points. Rapid Sequence Intubation. Rapid Sequence Intubation Recognizing difficult airway Managing difficult airway

Comparisonof Classic- Laryngeal Mask Airway with I-Gel Airway,during Anaesthesia with Controlled Ventilation in Routine Surgical Procedures

THE laryngeal mask airway (LMA) classic (Vitaid Ltd.,

Blind Insertion Airway Devices (BIAD)

Airway/Breathing. Chapter 5

A Comparative Study Between Proseal Laryngeal Mask Airway And Endotracheal Intubation.

GENERAL ANAESTHESIA AND FAILED INTUBATION

RESPIRATION AND THE AIRWAY Comparison of the i-gel with the cuffed tracheal tube during pressure-controlled ventilation

Clinical Study McGrath Video Laryngoscope May Take a Longer Intubation Time Than Macintosh Laryngoscope

Saumya Biswas *, Soumen Mandal **, Tapobrata Mitra ***, Suprio De Ray ****, Ranabir Chanda ***** and Debajyoti Sur ******

Haemodynamic response to orotracheal intubation: direct laryngoscopy versus fiberoptic bronchoscopy

Clinical Study Performance of Size 1 I-Gel Compared with Size 1 ProSeal Laryngeal Mask in Anesthetized Infants and Neonates

Comparison of the clinical performance of i-gel, LMA Supreme and LMA ProSeal in elective surgery

Cuffed or uncuffed ETT in pediatric anesthesia? Dr. Renata Haghedooren Dr. Sophie Chullikal Dr. Julie Lauweryns

Clinical Study Randomised Comparison of the AMBU AuraOnce Laryngeal Mask and the LMA Unique Laryngeal Mask Airway in Spontaneously Breathing Adults

Supra-glottic Airway Devices

Nicolette Mosinski MPAS, PA-C

Use of the Aintree Intubation Catheter with the Laryngeal Mask Airway and a Fiberoptic Bronchoscope in a Patient with an Unexpected Difficult Airway

Comparative study of effective-site target controlled infusion with standard bolus induction of propofol for laryngeal mask airway insertion

Deposited on: 24 February 2009

AIRWAY MANAGEMENT AND VENTILATION

Comparative Evaluation of Performance of Videolaryngoscope vs Fastrach Intubating Laryngeal Mask Airway

Comparison of Three Disposable Extraglottic Airway Devices in Spontaneously Breathing Adults

Comparative Evaluation of Supraglottic Devices i-gel and Fastrach LMA as Conduit for Blind Endotracheal Intubation

Transcription:

Original Article Comparative evaluation of Ambu AuraGain with ProSeal laryngeal mask airway in patients undergoing laparoscopic cholecystectomy Address for correspondence: Dr. Kriti Singh, Department of Anaesthesia and Critical Care, Batra Hospital and Medical Research Centre, 1, Tughlakabad Institutional Area, MB Road, New Delhi 110 062, India. E mail: drkritisingh@gmail.com Access this article online Website: www.ijaweb.org DOI: 10.4103/ija.IJA_163_17 Quick response code Kriti Singh, Pavan Gurha Department of Anaesthesia and Critical Care, Batra Hospital and Medical Research Centre, New Delhi, India ABSTRACT Background and Aims: Second generation supraglottic airways are increasingly being used in surgical patients undergoing laparoscopic surgery. Preventing aspiration at higher airway pressures may be at the expense of a higher cuff pressure which can impair mucosal perfusion. We attempted to elucidate whether Ambu AuraGain (AAU) would provide a higher oropharyngeal leak pressure (OLP) with a lower mucosal pressure in comparison to ProSeal laryngeal mask airway (PLMA). Methods: This was a prospective randomised study involving sixty patients undergoing laparoscopic cholecystectomy under general anaesthesia, using either AAU (Group AAU [n = 30]) or PLMA (Group PLMA [n = 30]) for elective ventilation. Primary outcome measure was the OLP. Number of insertion attempts, ease of insertion, time required for placement and calculated pharyngeal mucosal pressure were the secondary outcome measures. Data were analysed using Student s t test and Chi square test. Results: No significant difference in the OLP was noted in both groups. The ease of insertion and success rate at first attempt was similar between the groups. Time taken for insertion in Group AAU was longer than Group PLMA (13.57 ± 1.94 vs. 11.60 ± 2.22 s). The calculated pharyngeal mucosal pressures were lower with Group AAU than Group PLMA for all 3 sizes. The minimum cuff pressure and minimum cuff volume required to prevent leak were found similar in both groups. Conclusion: AAU provides adequate sealing pressures and effective ventilation with lower calculated pharyngeal mucosal pressure, compared to PLMA. Key words: Airway Management, Ambu AuraGain, oropharyngeal leak pressure, ProSeal laryngeal mask airway, supraglottic devices INTRODUCTION Supraglottic Airways (SGAs) have revolutionised the airway management. [1] Besides serving as a rescue device in the difficult airway, and as a conduit for the endotracheal tube insertion, SGAs provide a less invasive and less traumatic means of securing the airway in surgical patients. [2,3] The introduction of the ProSeal laryngeal mask airway (PLMA) (Teleflex, NC, USA), a second generation SGA, led to its clinical use in patients undergoing laparoscopic surgery. [4] The ability of an SGA to prevent aspiration at higher airway pressures is desirable, but it may be at the expense of a higher cuff pressure which may impair mucosal perfusion. This may result in greater airway morbidities. Thus, an ideal SGA would provide a high oropharyngeal leak pressure (OLP) with a low pharyngeal mucosal pressure. The Ambu AuraGain (Ambu, DK) is a disposable, preformed second generation SGA with integrated gastric access and intubation capability, introduced some time back. [5] We designed this study to compare This is an open access article distributed under the terms of the Creative Commons Attribution NonCommercial ShareAlike 3.0 License, which allows others to remix, tweak, and build upon the work non commercially, as long as the author is credited and the new creations are licensed under the identical terms. For reprints contact: reprints@medknow.com How to cite this article: Singh K, Gurha P. Comparative evaluation of Ambu AuraGain with ProSeal laryngeal mask airway in patients undergoing laparoscopic cholecystectomy. Indian J Anaesth 2017;61:469-74. 2017 Indian Journal of Anaesthesia Published by Wolters Kluwer - Medknow 469 Page no. 27

the Ambu AuraGain (AAU) with the ProSeal LMA (PLMA) for patients undergoing laparoscopic cholecystectomy in terms of the OLP and pharyngeal mucosal pressure. METHODS This randomised, prospective study was conducted in a super specialty hospital from February 2016 to November 2016 and was approved by the Institutional Ethics Committee. We recruited sixty patients aged 18 65 years with the American Society of Anesthesiologists physical status I II, who were to undergo elective laparoscopic cholecystectomy under general anaesthesia. Written informed consent was obtained from all patients. The exclusion criteria were obesity (body mass index 30 kg/m 2 ), pregnancy, known or predicted difficult airway, reduced lung compliance, mouth opening <2.5 cm and high risk for pulmonary aspiration (nonfasted, gastroesophageal reflux disease). Patients were wheeled in the operation theatre, monitors including heart rate, noninvasive blood pressure, electrocardiogram, oxygen saturation and temperature probe were connected, intravenous (i.v.) cannula placed and a maintenance fluid (Ringer lactate) was started at the rate of 4 ml/kg/h. Pre medication was done with midazolam 0.03 mg/kg i.v. Induction of anaesthesia was done with fentanyl 2 µg/kg i.v. and propofol 2 2.5 mg/kg i.v. and maintenance with 50% oxygen and 50% air with 1 minimum alveolar concentration of isoflurane. Muscle relaxation was achieved with atracurium 0.5 mg/kg i.v. The patients were ventilated for 3 min. with a face mask till adequate jaw relaxation was achieved. Before device insertion, the patients were divided into two groups (30 patients each) and were randomised to receive either the conventional reusable PLMA (Group PLMA) or single use disposable AAU (Group AAU). Randomisation of patients was done according to computer generated random codes that were maintained in sequentially numbered opaque envelopes. Before inserting the device, the cuff was inflated with maximum permissible air (as per manufacturer s recommendation), and the cuff pressure (ex vivo) was measured with a cuff pressure monitor (Mallinckrodt VBM Medizintechnik GmbH DE). All SGAs were fully deflated and lubricated with a water soluble gel. After the adequate depth of anaesthesia and muscle relaxation was achieved, insertion of both devices was performed by an anesthesiologist with more than 10 years experience in using the two SGA devices. The patient s head was placed in a sniffing position (neck flexion and head extension). In Group AAU, AAU of appropriate size was inserted as per manufacturers instructions, and in Group PLMA, ProSeal LMA of appropriate size was inserted by introducer tool technique as per manufacturers instructions. A size 3, 4 or 5 of either device was selected according to the patient s weight. [6] After insertion, the device was inflated with the maximum permissible volume of air, and the air entry was checked on both sides of the chest. The in vivo cuff pressure was measured. The pharyngeal mucosal pressure was calculated as: P = P in vivo P ex vivo. [7] We recorded the number of insertion attempts and the time needed for SGA placement, (measured from when the SGA was picked up until the connection to the breathing circuit). The ease of insertion was graded as 1 easy or 2 difficult (when deep rotation and jaw thrust or a second attempt was used for proper device insertion). [8] An effective airway was judged by an absence of leak, normal thoracoabdominal movement and square wave capnograph trace. If there was a leak, the SGA was repositioned. If leak persisted after a total of three attempts, the patient was intubated with an endotracheal cuffed tube and recorded as failed insertions. After confirmation of adequacy of ventilation, airway pressure was maintained at an arbitrarily determined pressure of 18 cm H 2 O by squeezing the reservoir bag for 10 s. Inflation pressure was measured using an airway pressure gauge. The incidence of airleak around the mask was recorded by detection of an audible noise over the mouth. If there was no leak and ventilation was adequate, air from the cuff was aspirated slowly till leak appeared. At this point, 1cc aliquots of air were reinjected into the cuff till leak disappeared. The residual volume of air left behind in the cuff was considered as the minimum effective volume of air. The minimum effective volume of air and the minimum pressure in the cuff (at this cuff volume) to prevent leak at airway pressure of 18 cm H 2 O were noted. [7] Before commencement of the surgery, the cuff was inflated with a manometer and cuff pressure was set to 60 cm of H 2 O. OLP was measured by closing the expiratory valve of the breathing circuit and 470 Indian Journal of Anaesthesia Volume 61 Issue 6 June 2017 Page no. 28

delivering a gas flow of 3 L/min until the seal pressure or a maximum pressure of 40 cm H 2 O was reached. A lubricated 16 Fr gastric tube was introduced via the drain tube of both the devices to aspirate gastric contents. If resistance was felt during the introduction, a 14 Fr gastric tube was used instead. Peak inspiratory airway pressure was measured before and after the creation of carboperitoneum keeping a constant intra abdominal pressure of 14 mm Hg in the two groups. The primary outcome was the OLP. With the help of a pilot study of ten patients per group, the mean OLP was calculated as 29 cm H 2 O in the PLMA group and 34 cm H 2 O in the AAU group. The total sample size was calculated as 38 calculated from an effect size of 0.83, a power of 80% and α of 0.05 where the standard deviation of two groups was 6 and 5, respectively. For better validation of our result, the total sample size was set at 60 (a minimum of 30 per group). Statistical testing was conducted with the International Business Machines Corporation SPSS Statistics for windows (version 17.0, Armonk, NY, USA). Demographic data were analysed using the Student s unpaired t test and Chi square test. Ease of insertion, time taken for insertion and OLP were categorised and compared using Mann Whitney U test and Chi square test. For all statistical tests, a P < 0.05 was taken to indicate a significant difference. RESULTS Our study enrolled a total of sixty participants who were analysed for the primary outcome. Since there were no exclusions or losses, all participants completed the study [Figure 1]. There were no statistically significantly differences between the groups regarding age, gender, weight, height and ASA grade [Table 1]. The distribution of the size of SGA used in each group was similar. The ease of insertion and success rate at first attempt was similar amongst both groups. The time taken for insertion of AAU was longer than PLMA. No significant difference in the OLPs was noted in both groups. The peak inspiratory airway pressures before and after insufflation in both groups were comparable. In Group PLMA, a size 14 Fr Ryles Tube was inserted in all the patients, whereas in group AAU, size 16 Fr Ryles Tube Figure 1: CONSORT flowchart Indian Journal of Anaesthesia Volume 61 Issue 6 June 2017 471 Page no. 29

was used in 70% of the patients [Table 2]. The calculated pharyngeal mucosal pressures were lower with AAU than PLMA for all 3 sizes of the SGAs. However, the difference was statistically significantly only for a size 4 AAU [Table 3]. The minimum cuff volume required to prevent leak (at 18 cm H 2 O airway pressure) was found similar in both groups, notably, these values were lower than the maximum recommended values. In size 3 AAU, the minimum cuff pressure (at the minimum cuff volume) required to prevent leak was significantly lower than for a size 3 PLMA [Table 4]. DISCUSSION PLMA is an established airway device for use in laparoscopic surgeries. AAU has been compared to LMA supreme in a few studies, [9 11] but to the best of our knowledge, there is no literature available that compares the clinical efficacy of AAU to PLMA in laparoscopic cholecystectomy. Hence, this study was undertaken to evaluate the safety profile of AAU in a clinical setting. In our study, we found that both the AAU and PLMA are easy to insert, with a similar successful insertion rate on the first attempt. However, the time for insertion of AAU was significantly higher, and this may be attributed to its rigid PVC preformed structure in contrast to the flexible silicone structure of the PLMA. Gastric tubes were successfully inserted on the first attempt in both the devices. However, AAU allowed Table 1: Patient characteristics Variables Group PLMA (n=30) Group AAU (n=30) Age (years) 43.93±12.28 37.93±11.72 Gender (male/female) 11/19 11/19 Weight (kg) 64.30±11.00 63.50±11.83 Height (cm) 163.23±8.26 161.17±7.61 ASA grade (I/II) 19/11 21/9 Data are expressed as number of patients or mean±sd. SD Standard deviation; PLMA ProSeal laryngeal mask airway; AAU Ambu AuraGain ; ASA American Society of Anesthesiologists the passage of a larger bore gastric tube, making it favourable in situations where a larger tube may help in better gastric decompression. In laparoscopic surgeries, the incidence of suboptimal and failed ventilation is often high with SGAs owing to the high peak airway pressures required during carboperitoneum. Second generation SGAs allow higher airway pressures due to their effective seal. In our study, we observed that the OLP of AAU was comparable to that of PLMA. A recent study reports higher OLP of AAU in comparison to LMA Supreme in patients undergoing gynaecologic laparoscopy. [9] Another study reports similar OLP of AAU and LMA Supreme in children during controlled ventilation under general anaesthesia. [10] However, the OLP of AAU was comparable to that of other SGAs such as i gel, PLMA and LMA Supreme as reported in some other studies. [8,12,13] In our study, the OLP of both AAU and PLMA was higher than the peak airway pressure and was sufficient to prevent aspiration while ventilating the study patients during carboperitoneum. A study reports a progressive reduction in the pharyngeal mucosal perfusion when mucosal pressure increases from 25 to 60 mmhg while using a cuffed oropharyngeal airway. [14] The cuffs exert pressure on the pharyngeal mucosa which causes a concomitant decrease of pharyngeal perfusion and increase in the incidence of post operative complications including sore throat, dysphonia and nerve damage. [15] We found that the calculated pharyngeal mucosal pressure exerted by AAU was less than PLMA and this may lead to a lower incidence of airway morbidities. However, we did not study the post operative airway morbidity which may be a limitation of our study. A study showed that intracuff pressure increased as the cuff volume increased. The pressure exerted on the Table 2: Airway management details Variables Group PLMA (n=30) Group AAU (n=30) P Size of SGA used (3/4/5) 3/19/8 4/17/9 0.855 Ease of insertion easy/difficult 22/8 18/12 0.273 Number of attempts (1/2/3) 24/6/0 18/12/0 0.091 Time for insertion (s) 11.60±2.22 13.57±1.94* 0.001 Oropharyngeal leak pressure (cm H 2 O) 27.17±16.91 (18 40) 28.77±4.82 (21 38) 0.303 Ryle s tube size (14/16 Fr) 30/0 9/21* 0.001 Initial peak airway pressure (cm H 2 O) 14.53±2.98 14.63±2.86 0.895 Peak airway pressure after insufflations (Cm H 2 O) 20.83±2.90 20.93±3.11 0.898 Data are expressed as number or mean±sd. *P 0.05. SGA Supraglottic airway; PLMA ProSeal laryngeal mask airway; AAU Ambu AuraGain ; SD Standard deviation 472 Indian Journal of Anaesthesia Volume 61 Issue 6 June 2017 Page no. 30

Table 3: Ex vivo pressure, in vivo pressure and calculated pharyngeal mucosal pressure Variables Group PLMA Group AAU P (n=30) (n=30) EVP Size 3 36.67±15.28 56.00±32.17 0.390 Size 4 34.84±5.18 72.47±29.19* 0.001 Size 5 41.00±10.69 68.11±31.58* 0.036 IVP Size 3 100.67±25.33 89.00±35.80 0.653 Size 4 102.42±17.49 117.76±5.24* 0.001 Size 5 114.50±5.93 118.89±3.33 0.075 Pharyngeal mucosal pressure (IVP EVP) Size 3 64.00±14.42 33.00±18.22 0.060 Size 4 67.68±15.68 45.29±27.30* 0.004 Size 5 73.50±10.18 50.78±30.22 0.061 Data are expressed mean±sd. *P 0.05. SD Standard deviation; EVP Ex vivo pressure; IVP In vivo pressure; PLMA ProSeal Laryngeal Mask Airway; AAU Ambu AuraGain Table 4: The minimum cuff volume required to prevent a leak, and cuff pressure at that volume Variables Group PLMA Group AAU P (n=30) (n=30) Minimum cuff volume (ml) Size 3 8.33±0.58 3.50±4.04 0.101 Size 4 4.16±5.72 3.18±5.54 0.605 Size 5 2.75±3.33 3.22±3.96 0.795 Minimum cuff pressure (cm H 2 O) Size3 11.33±2.31 0.00±0.00* <0.001 Size 4 6.11±11.11 8.35±19.68 0.671 Size 5 13.50±16.86 9.33±11.87 0.561 Data are expressed as mean±sd. *P 0.05. SD Standard deviation; PLMA ProSeal Laryngeal Mask Airway; AAU Ambu AuraGain pharynx by the SGA is usually higher than the mucosal capillary perfusion pressure when the cuff is inflated with the recommended maximum volume of air. [15] However, if the cuff is deflated excessively, it may not protect the airway from soiling, due to the regurgitated fluid from the stomach. [16] Therefore, it is desirable to inflate the cuff of the SGA with a minimum volume of air which provides a seal around the mask. In our study, there was no statistically significant difference between the PLMA and AAU in terms of the minimum effective cuff volume for all three sizes. However, these values were much less than the company recommended maximum values. Hence, removal of air from the cuff of the SGA (from the recommended maximum volume to the effective minimum volume to prevent leak) reduces the intracuff pressure and pressure exerted on the pharynx. Since these values were obtained at an airway pressure of 18 cm H 2 O, this cuff pressure and volume can be used only for mechanical ventilation in non-laparoscopic surgeries. Indian Journal of Anaesthesia Volume 61 Issue 6 June 2017 Our study has some limitations. First, our study was a single-centre study with a relatively small sample size. More studies are required to confirm the overall safety profile of AAU. Second, we did not measure the mucosal pressures in the pharynx directly with a microchip sensor but used an indirect derived value. Furthermore, the pharyngeal mucosal pressure at the cuff pressure of 60 cm H 2 O is not known, since it was measured by inflating the cuff with the maximum volume of air recommended. Our study involved patients with a normal airway and whether the same outcome can be extrapolated to patients with difficult airway needs further study. Finally, after repeated use, PLMA may lose its elasticity and cuff permeability may increase leading to higher cuff pressures and mucosal pressures. CONCLUSION Ambu AuraGain provides adequate sealing pressures and effective ventilation, hence, it may be considered for use in clinical practice. However, further studies are required to assess how this device compares to other widely utilised SGAs Acknowledgement We acknowledge the assistance and cooperation received from our departmental colleagues. Financial support and sponsorship Nil. Conflicts of interest There are no conflicts of interest. REFERENCES 1. Pennant JH, White PF. The laryngeal mask airway. Its uses in anesthesiology. Anesthesiology 1993;79:144 63. 2. Kovac AL. Controlling the hemodynamic response to laryngoscopy and endotracheal intubation. J Clin Anesth 1996;8:63 79. 3. Prys Roberts C, Greene LT, Meloche R, Foëx P. Studies of anaesthesia in relation to hypertension. II. Haemodynamic consequences of induction and endotracheal intubation. Br J Anaesth 1971;43:531 47. 4. Maltby JR, Beriault MT, Watson NC, Liepert D, Fick GH. The LMA ProSeal is an effective alternative to tracheal intubation for laparoscopic cholecystectomy. Can J Anaesth 2002;49:857 62. 5. AMBU. Baltorpbakken (DK): ABMU A/S. AuraGain Disposable Laryngeal Mask. Available from: http://www.ambu.com/ corp/products/anaesthesia/product/auragain%e2%84%a2_ disposable_laryngeal_mask prod18315.aspx. [Last updated on 2015 Nov 13; Last cited on 2017 May 15]. 6. Dorsch JA, Dorsch SE. Understanding Anesthesia Equipment. 5 th ed. Philadelphia: Lippincott Williams & Wilkins; 2007. p. 474 5. 473 Page no. 31

7. Asai T, Howell TK, Koga K, Morris S. Appropriate size and inflation of the laryngeal mask airway. Br J Anaesth 1998;80:470 4. 8. Singh I, Gupta M, Tandon M. Comparison of clinical performance of I Gel with LMA Proseal TM in elective surgeries. Indian J Anaesth 2009;53:302 5. 9. Lopez AM, Agusti M, Gambus P, Pons M, Anglada T, Valero R. A randomized comparison of the Ambu AuraGain versus the LMA supreme in patients undergoing gynaecologic laparoscopic surgery. J Clin Monit Comput 2016. Epub 2016 Nov 26. 10. Jagannathan N, Hajduk J, Sohn L, Huang A, Sawardekar A, Gebhardt ER, et al. A randomised comparison of the Ambu AuraGain and the LMA supreme in infants and children. Anaesthesia 2016;71:205 12. 11. Lopez AM, Sala Blanch X, Valero R, Prats A. Cross over assessment of the AmbuAuraGain, LMA supreme new cuff and intersurgical I Gel in fresh cadavers. Open J Anesthesiol 2014;4:332 9. 12. López AM, Valero R, Hurtado P, Gambús P, Pons M, Anglada T. Comparison of the LMA Supreme with the LMA Proseal for airway management in patients anaesthetized in prone position. Br J Anaesth 2011;107:265 71. 13. Gasteiger L, Brimacombe J, Perkhofer D, Kaufmann M, Keller C. Comparison of guided insertion of the LMA ProSeal TM vs. the i gel TM. Anaesthesia 2010;65:913 6. 14. Brimacombe J, Keller C, Pühringer F. Pharyngeal mucosal pressure and perfusion: A fiberoptic evaluation of the posterior pharynx in anesthetized adult patients with a modified cuffed oropharyngeal airway. Anesthesiology 1999;91:1661 5. 15. Marjot R. Pressure exerted by the laryngeal mask airway cuff upon the pharyngeal mucosa. Br J Anaesth 1993;70:25 9. 16. Brimacombe J, Berry A, Brain AI. Optimal intracuff pressures with the laryngeal mask. Br J Anaesth 1996;77:295 6. Announcement Northern Journal of ISA Now! Opportunity for our members to submit their articles to the Northern Journal of ISA (NJISA)! The NJISA, launched by ISA covering the northern zone of ISA, solicits articles in Anaesthesiology, Critical care, Pain and Palliative Medicine. Visit http://www.njisa.org for details. Dr. Sukhminder Jit Singh Bajwa, Patiala Editor In Chief Dr. Zulfiqar Ali, Srinagar Co-Editor 474 Indian Journal of Anaesthesia Volume 61 Issue 6 June 2017 Page no. 32