Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene Liner

Similar documents
Wear Analysis of Second-generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty

Wear and Osteolysis of Highly Crosslinked Polyethylene at 10 to 14 Years: The Effect of Femoral Head Size

Investigation performed at the Anderson Orthopaedic Research Institute, Alexandria, Virginia

Comparative Study of Peripheral Rim Fixation Using Jumbo Cup in Revisional Hip Arthroplasty

Outcomes of Surgical Treatment of Periprosthetic Femoral Fractures in Cementless Hip Arthroplasty

Bone Bangalore

Revision Total Hip Arthroplasty Using Tantalum Augment in Patients with Paprosky III or IV Acetabular Bone Defects: A Minimum 2-year Follow Up Study

Are Portable Imaging Intraoperative Radiographs Helpful for Assessing Adequate Acetabular Cup Positioning in Total Hip Arthroplasty?

111. ACETABULAR FIXATION

Measuring the Impact of Femoral Head Size on Dislocation Rates Following Total Hip Arthroplasty

FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY

Poor mid-term results of total hip arthroplasty with use of a Hylamer liner

MARATHON Cross-Linked Polyethylene

The Usefulness of Three-dimensional Computed Tomography as an Assessment of Periacetabular Osteolysis in Revision Total Hip Arthroplasty

The Effect of Hydroxyapatite Coating on Long-term Results of Total Hip Arthroplasty with Hydroxyapatite-coated Anatomic Femoral Stem

ARTICLE IN PRESS. Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction. Effects on Polyethylene Wear

Templating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5.

Increased Rate Of Dislocation Of Total Hip Arthroplasty With Cementless Implants: Accuracy Of Acetabular Inclination.

Acetabular Cup System. Clinical Summary

Femoral Osteolysis Around the Unrevised Stem During Isolated Acetabular Revision

DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574)

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

Dissociation of Polyethylene Liner of the Dual Mobility Acetabular Component after Closed Reduction of Dislocation: A Case Report

Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces

Radiographic Wear Measurements in a Cementless Metal-Backed Modular Cobalt-Chromium Acetabular Component

Zimmer TM. Continuum Acetabular System. Bringing together proven technologies.

Reprint requests: Dr Brenda Dower CLINICAL ARTICLE SA ORTHOPAEDIC JOURNAL Spring 2012 Vol 11 No 3 / Page 29

The Challenge of Durability in Hip Replacement

Treatment of Periprosthetic Femoral Fractures Following Hip Arthroplasty

PINNACLE Acetabular Cup System

Medium- to Long-term Results of Strut Allografts Treating Periprosthetic Bone Defects

Cementless Acetabular Fixation With and Without Screws

Metal-on-Metal Bearings in Cementless Primary Total Hip Arthroplasty

Disclosure. Direct Anterior THA via Extension Table. History 11/3/2015

Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem

This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

Zimmer Longevity Highly Cross-linked Polyethylene. Clinical Value Dossier - APAC

Hedrocel trabecular metal monoblock acetabular cups : mid-term results

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

ACETABULAR CUP SURGICAL TECHNIQUE

The optimal timing of baseline radiostereometric analysis of uncemented press fit cups

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head

Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo?

Department of Orthopaedic Surgery, Graduate School of Medical Sciences, Kyushu University, Maidashi, Higashi-ku, Fukuoka , Japan

Fracture of an Acetabular Component Inserted without Cement: A Case Report

1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe

Optimum implant geometry

Formosan Journal of Musculoskeletal Disorders

Stability without compromise. Epsilon Durasul Constrained Insert

Reaching new heights. Comprehensive. Efficient. Simple.

CAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician.

Impingement in THA. Georgi I. Wassilew. Orthopaedic Department, Centrum für Muskuloskeletale Chirurgie Univ.-Prof. Dr. med. C.

Pseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008)

RM Philosophy Innovation based on clinical evidence.

Furlong H-A.C. THR System

Avascular necrosis of the hip treated with Metal-on-Metal. surface hip arthroplasty

The Spatial Location of Impingement in Total Hip Arthroplasty

The Journal of Arthroplasty

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY

A 42-year-old patient presenting with femoral

Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital)

Active Articulation E1

Revision Total Hip Arthroplasty after Ceramic Bearing Fractures in Patients Under 60-years Old; Mid-term Results

Enhanced Stability Constrained Liners. Design Rationale Surgical Technique

CAUTION: Ceramic liners are not approved for use in the United States.

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

DURALOC. Acetabular Cup System. Surgical Technique

Tri-Lock Bone Preservation Stem

PRODUCT RATIONALE & SURGICAL TECHNIQUE

Featuring. Technology. Product Rationale

Surface Roughness of Femoral Head Prostheses After Dislocation

Product Information. *smith&nephew POLARCUP Dual Mobility System

Optimum implant geometry

Scandinavian Journal of Surgery 103: 54 59, 2013

CASE 56 Hip Pain. ? Diagnosis

The shape and size of femoral components in revision total hip arthroplasty among Chinese patients

Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs: Report of Two Cases

Effect of alumina femoral heads on polyethylene wear in cemented total hip arthroplasty

Breakage of the Tail Portion of the Lag Screw during Removal of Proximal Femoral Zimmer Natural Nail: Report of Two Cases with Technical Notes

High wear rates and extensive osteolysis in 3 types of uncemented total hip arthroplasty

Metasul LDH Large Diameter Head

TriboFit Hip System. Issue 3 - August 2013 English

Femoral neck fracture during physical therapy following surface replacement arthroplasty: a preventable complication?

CLINICAL PAPER / ORTHOPEDIC

Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists

Duraloc CONSTRAINED LINER

The proven, simple solution. CPT 12/14 Hip System

University of Groningen. Metal-on-metal total hip arthroplasty Zijlstra, Wierd Pieter

AGGRESSIVE GRANULOMATOUS LESIONS IN CEMENTLESS TOTAL HIP ARTHROPLASTY

Painful Metal-on-Metal Total Hip Arthroplasty

Granulomatous Lung Disease: A Novel Complication following Metallosis from Hip Arthroplasty

Polyethylene wear rate and osteolysis: critical threshold versus continuous dose-response relationship

Optimum implant geometry

The information contained in this document is intended for healthcare professionals only.

Extensively Porous-coated Stems for Femoral Revision: Reliable Choice for Stem Revision in Paprosky Femoral Type III Defects

Ceramic Femoral Heads and Acetabular Cup Liners

TOTAL HIP REPLACEMENT:

Versatility in a proven design. Trilogy Acetabular System

The information contained in this document is intended for healthcare professionals only.

Transcription:

ORIGINAL ARTICLE Hip Pelvis 27(4): 216-222, 2015 http://dx.doi.org/10.5371/hp.2015.27.4.216 Print ISSN 2287-3260 Online ISSN 2287-3279 Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene Liner Min-Yook Kim, MD, Young-Yool Chung, MD, Ji-Hoon Park, MD, Jung-Ho Lee, MD Department of Orthopaedic Surgery, Kwangju Christian Hospital, Gwangju, Korea Purpose: This retrospective study was performed to evaluate the clinical results and measure polyethylene liner wear in total hip arthroplasty (THA) with highly cross-linked polyethylene. Materials and Methods: Except for patients who had died or were unable to have follow-up at least 2 years, 60 of 78 hips that underwent THA were included this study. The mean age was 64.5 years (range, 25-81 years) and the mean body mass index (BMI) was 23.0 kg/m 2 (18.1-32.3 kg/m 2 ). Diagnosis at the time of the operation was osteonecrois of the femoral head in 28 hips, primary osteoarthritis in 14, hip fracture in 13, and other diseases in 5. The mean follow-up period was 3.8 years (2.1-7.1 years). Harris hip score (HHS) was reviewed before THA and at the last follow-up. On the anteroposterior pelvic radiographs, acetabular cup inclination and ante-version were also measured. The annual linear wear rate was measured using Livermore s method on the radiographs. Results: The mean HHS was 60.1 (28-94) before operation and 90.4 (47-100) at the last follow-up. In the immediate post-operation, the average inclination and anteversion angles of the acetabular cups were 46.3 (standard deviation, ±6.7 ) and, 21.4 (±10.1 ) respectively. The mean of the annual linear polyethylene wear was 0.079 mm/year (0.001-0.291 mm/year). Age, gender and BMI were not statistically related to linear polyethylene wear but the period of follow-up and the acetabular cup s inclination showed significant negative and positive correlation respectively. Conclusion: The wear rate of a highly cross-linked polyethylene was shown to correlate negatively with duration of follow-up. However, our study was based on a short-term follow-up, so a long-term follow-up study is necessary in the future. Key Words:Total hip arthroplasty, Highly cross-linked polyethylene, Linear wear rate Submitted: July 1, 2015 1st revision: September 1, 2015 2nd revision: October 24, 2015 Final acceptance: October 25, 2015 Address reprint request to Young-Yool Chung, MD Department of Orthopaedic Surgery, Kwangju Christian Hospital, 37 Yangnim-ro, Nam-gu, Gwangju 61661, Korea TEL: +82-62-650-5064 FAX: +82-62-650-5066 E-mail: Paedic@chol.com This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons. org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. INTRODUCTION The use of a metal head with polyethylene in total hip arthroplasty (THA) is the oldest articular surface combination and is in wide use nowadays 1). However, wear particles of polyethylene were produced from the articular surface when metal to polyethylene was used in THA and seen in long-term follow-up. The osteolysis due to the wear particles is the main cause of revision because it causes aseptic loosening of the prosthesis 2-6). In order to increase the survival rate of the prosthesis in 216 Copyright c 2015 by Korean Hip Society

Min-Yook Kim et al. THA Using Metal Head on a Highly Cross-linked Polyethylene Liner long-term follow-up, several bearing surfaces were attempted using ceramic to ceramic or metal to metal articular surfaces instead of a metal to polyethylene articular surface. Even though these articular surfaces showed good results with respect to wear, there were the complications of ceramic breakage, increased serum metal ions, and hypersensitivity to metal ions 7). At the same time work continued to reduce the wear in polyethylene. Highly cross-linked polyethylene improves its wear resistance by increasing cross-linking and reducing free radicals and inhibiting its oxidation. This bearing surface has excellent wear durability in comparison with conventional polyethylene 8). In particular, highly cross-linked polyethylene was not significantly influenced by the activity level of the patient, body mass index (BMI), age, or gender. So it is possible to use a large diameter head in THA to prevent dislocation and increase the range of motion 9,10). The purpose of this study is to analyze the wear rate in THA with highly cross-linked polyethylene and validate the factors that affect the wear rate. MATERIALS AND METHODS From January 2006 to May 2010, 60 cases in 55 patients from 78 cases in 71 patients with primary THA using highly cross-linked polyethylene and a 28 mm metal head (Zimmer, Warsaw, IN, USA) were studied after at least 2 years of follow-up. Five patients underwent arthroplasty on their both hips. This excludes patients who died (11 cases) and those who were unable to follow-up (7 cases). The mean age at the time of surgery was 64.5 years (25-81 years). There were 36 males and 24 females. The average BMI was 23.0 kg/m 2 (range, 18.1-32.3 kg/m 2 ). There were 28 cases of avascular necrosis (46.7%), 14 cases of osteoarthritis (23.3%), 13 cases of hip fractures (21.7%), and 5 cases of other diseases such as ankylosing spondylitis (8.3%) as the causes leading to THA. Follow-up period was an average of 3.8 years (2.1-7.1 years). Surgery in all cases was performed with a posterolateral approach by one surgeon (C.Y.Y), used cementless acetabular cups (Zimmer), and the diameter ranged from 48 mm to 62 mm. For press-fit fixation, the acetabular cups were inserted 1-2 mm bigger than the last reaming diameter. From one to three metal screws were added to fix the acetabular cups. The polyethylene used was Longevity (Zimmer) which is the first generation for highly crosslinked polyethylene and its average thickness was 7.7 mm (6.2-11.3 mm). Metal femoral heads of cobalt-chrome with a diameter of 28 mm and cementless femoral stems for proximal fixation were used in all cases. After the removal of wound drainage, patients practiced walking with crutches for up to two months after surgery. Clinical follow-up was done every 3 months for the first year. For every follow-up, patients were evaluated using anteroposterior and lateral views of pelvic radiographs. During the follow-up period, the anteversion and inclination angles of the acetabular cups were compared with the immediate post-operation results. Osteolysis and stress shielding around the acetabular cup or the femoral stem were analyzed on the follow-up radiographs. Loosening of acetabular cup was defined as change of cup angle more than 5 or displacement more than 2 mm in the follow-up radiographs compared to that of immediate postoperation 11). Liner wear rate of the polyethylene was measured using the Livermore s method on the anteroposterior views of the pelvic radiographs 3 months after surgery and at the last follow-up 12). Enlargement or reduction of the measured radiologic parameters, except the anteversion and inclination angles, were corrected through the ratio between measured diameter on radiograph and actual diameter of femoral head 28 mm in diameter. The annual wear rates of polyethylene were calculated on the basis of this method. Three orthopedic surgeons measured the radiologic parameters and calculated the annual wear rates of polyethylene by the same technique in order to increase the accuracy of the measurement. Average values of the calculated wear rates were defined as the final wear rate of the polyethylene. And we found that moderate reliability was confirmed (P<0.05) with 0.695 reliability in measured values between the observers using this method. For clinical outcome evaluation, Harris hip scores (HSS) of the pre-operation and last follow-up were assessed and grouped as excellent for >90 out of 100 points, good for >80, average for >70, and poor for <70. Statistical analysis was done using SPSS version 18.0 for Windows (SPSS, Chicago, Illinois) program. The Pearson correlation test was used to analyze correlation of multiple continuous variables with the annual wear rate of polyethylene. Based on the results of the Pearson correlation test, we used the Univariate linear regression test between each continuous variable and the annual www.hipandpelvis.or.kr 217

Hip Pelvis 27(4): 216-222, 2015 Table 1. A Correlation Analysis of Influential Factors for the Annual Wear Rate of Polyethylene Annual wear rate Age F/U BMI Acetabular Liner Last F/U Postop. Postop. period cup size thickness HHS inclination anteversion Pearson correlation coefficient 0.022 0.433* 0.002 0.012 0.009 0.131 0.280* 0.020 P-value 0.869 0.001* 0.988 0.929 0.946 0.318 0.030* 0.882 F/U: follow-up, BMI: body mass index, HHS: Harris hip score, Postop.: postoperative. * Significant correlation, P < 0.05. wear rate of polyethylene. Intra class correlation analysis was done to determine the inter-observer reliability of the calculated wear rate. This study was approved by the institutional ethics committee of our hospital and informed consent was obtained from each patient. There was no external funding source for this study. RESULTS The annual wear rate of polyethylene was 0.079 mm/year (0.001-0.291 mm/year) measured using the Livermore s method. From the correlation analysis, the annual wear rate of polyethylene did not have a significant correlation with age, gender, BMI, acetabular cup size, thickness of polyethylene and anteversion angles. But there was a negative correlation between the annual wear rate of polyethylene and the follow-up period (P=0.001), and a positive correlation between the inclination angles of the acetabular cups and the annual wear rate of polyethylene (P=0.030) (Table 1). Significantly, above two factors were shown the linear relationship with the annual wear rate of polyethylene from univariate linear regression analysis (Table 2, Fig. 1). Divided into several groups of the above two factors, the more the period of follow up increase, the more annual wear rate of polyethylene decrease. In contrast, the more the inclination angles of the acetabular cups, the more annual wear rate of polyethylene increase (Fig. 2). Harris hip score (HHS) significantly increased from an average of 50.1 points (28-64 points) before surgery to 90.4 points (77-100 points) at the last follow-up. Satisfactory clinical results were obtained in 56 cases (93.3%) with more than a good grade in HHS. There were no cases of osteolysis at the last follow-up in pelvic radiographs and there were no cases of revision surgery during the follow-up period. However there were 4 cases of dislocation so those were treated with abduction orthosis. Table 2. The Association of the Annual Wear Rate of Polyethylene with Variables by Univariate Linear Regression Analysis Variable β±se P-value Age 0.000±0.001 0.8690 F/U period 0.19±0.005 0.001* BMI 0.000±0.003 0.9880 Liner thickness 0.001±0.0070 0.9460 Inclination 0.003±0.001 0.030* Anteversion 0.000±0.001 0.8820 SE: standard error, F/U: follow-up, BMI: body mass index. * Significant association of linear regression analysis with the annual wear rate of polyethylene. DISCUSSION Polyethylene is the most commonly used articular surface in THA. However, in conventional polyethylene, free radicals remain through the production and packaging process, so oxidative reactions which cause wear in polyethylene occur due to the oxygen in its environment 13). Wear in the polyethylene eventually causes osteolysis around the implants and is a major cause of revision arthroplasty 14-16). Therefore, what was needed was a more durable polyethylene for increasing the survival of the implant in the long-term follow-up. In order to increase the durability, polyethylene was irradiated with gamma radiation to increase crosslinking; moreover, other procedures have been introduced to eliminate the remaining free radicals 17). The methods to eliminate the remaining free radicals after polyethylene processing were the remelting method by heating the polyethylene to its melting point and the annealing method by heating it to a temperature a little lower than the melting point 18). In this study, we used Longevity which is highly cross-linked by irradiation with an electron beam of 10 Mega rad. And the polyethylene was processed using the remeltannealing method for elimination of any remaining free radicals and finally the implant was packaged with the 218 www.hipandpelvis.or.kr

Min-Yook Kim et al. THA Using Metal Head on a Highly Cross-linked Polyethylene Liner gas-plasma method in order to minimize the wear of the polyethylene 9,18). McCalden et al. 19) used highly cross-linked polyethylene to carry out THA and reported the average annual wear rate to be 0.003 mm/year from an average 6.8 year follow-up. It was reported that wear rates were reduced more than 95% compared to the wear rate of conventional polyethylene and osteolysis was not observed during the follow-ups. Lee et al. 20) carried out the retrospective study of THA with the same polyethylene which used our study for follow-ups of more than 7 years. They reported that the average annual wear rate was 0.006 mm/year and 10.6% of all patients had osteolysis. The average HHS was 90.4 points at the last follow-up and the annual wear rate of the polyethylene was 0.079 mm/year in this study. A B Fig. 1. A scatter plots show linear correlation the annual wear rate of polyethylene versus the period of follow-up (A) and inclination angles of the acetabular cups (B). The y-axis is the annual wear rate of polyethylene (mm/year), x-axis are the period of follow-up (years) and inclination angles of the acetabular cups ( ). The dash lines show the regression lines of linear correlation the annual wear rate of polyethylene versus the period of follow-up and inclination angles of the acetabular cups. www.hipandpelvis.or.kr 219

Hip Pelvis 27(4): 216-222, 2015 Osteolytic lesions could not be observed in this study, it was considered that the average follow-up period was a little short at 3.8 years. But the annual wear rate of polyethylene did not increase in cases with follow up more than 5 years. Therefore it is estimated that good clinical and radiologic results could be obtained from mid-term follow-ups after several years. Sychterz et al. 21,22) described two processes where the femoral head penetrates the polyethylene liner. The bedding-in state lasts for 1-2 years after surgery and the true-wear state continues thereafter. The bedding-in state is a setting in process where the polyethylene liner moves slightly in the acetabular cup. The true-wear state was reported to be a process of production of wear particles. In the study of Lee et al. 20), the annual wear rate of polyethylene increased up to 2 years after surgery and was drastically reduced thereafter. In our study, there was a negative correlation of the annual wear rate of polyethylene with the follow-up period, in other words, the longer the follow-up period, the more the annual wear rate of polyethylene is reduced. Therefore we considered that our annual wear rate for polyethylene includes more of the period of the bedding-in state rather than the true-wear state as Sychterz et al. 22) described. So the wear rate of highly cross-linked polyethylene is actually thought to be much A B Fig. 2. Each divided into groups of the period of follow-up (A) and inclination angles of the acetabular cups (B), a flow plots show the means of annual wear rate of polyethylene and the total wear during follow up. The y-axis is the means of annual wear rate of polyethylene (mm/year), x-axis are different groups by the period of follow-up (years) and inclination angles of the acetabular cups ( ). The dash lines represent the regression lines of linear correlation. 220 www.hipandpelvis.or.kr

Min-Yook Kim et al. THA Using Metal Head on a Highly Cross-linked Polyethylene Liner smaller than what has been calculated as the wear rates of polyethylene are drastically reduced in cases after 2 years of surgery. Also, the annual wear rate of polyethylene is thought to reduce as the overall followup period increases (Fig. 2A). It is known that a patient s activity, the modular cementless acetabular cup, higher inclination angles of the acetabular cup, a larger metallic femoral head, and a thinner thickness of polyethylene are negative factors which affect the wear of polyethylene 23,24). But the wear of polyethylene is not affected in THA with highly cross-linked polyethylene by these factors. Manning et al. 25) reported that the wear rate of highly cross-linked polyethylene does not have significant relations with gender, age, BMI, or a patient s activity. Hermida et al. 26) reported no significant difference in wear rate between using a 28 mm-diameter femoral head and a 32 mmdiameter femoral head. Lee et al. 20) reported that there were no significant correlations of other factors including the thickness of the polyethylene liner, only a significant correlation with the inclination angle of the acetabular cups. Our study also showed a positive correlation between the wear rates of polyethylene and the inclination angle of the acetabular cups. The wear rates of polyethylene were significantly higher for angles greater than 45 than for those less than 45 of inclination in the acetabular cups on radiographs after surgery (Fig. 2B). Therefore, it is very important to correctly insert the acetabular cups so they have a stable position, in order to reduce the wear in highly crosslinked polyethylene and increase the survival rate for long-term follow-up. A limitation of our study was that the follow-up period was short. So the wear rates of polyethylene were calculated to be higher than the actual wear rate of polyethylene because it was measured on radiographs which had taken at three months after surgery that is in bedding-in period. But it is indirectly confirmed that the actual period of the bedding-in state is longer than three months from a negative correlation between the annual wear rate of polyethylene and follow-up period. It is necessary to study the clinical results of the wear in polyethylene with long-term follow-ups in the future. CONCLUSION The wear rate of a highly cross-linked polyethylene was shown to correlate negatively with duration of follow-up. It also shows excellent clinical results for the overall period of follow-up. However, our study was based on a short-term follow-up, so a long-term followup study is necessary in the future. REFERENCES 01.Laupacis A, Bourne R, Rorabeck C, et al. The effect of elective total hip replacement on health-related quality of life. J Bone Joint Surg Am. 1993;75:1619-26. 02.Malchau H, Herberts P. Prognosis of total hip replacement. Int J Risk Saf Med. 1996;8:27-45. 03.Clohisy JC, Calvert G, Tull F, McDonald D, Maloney WJ. Reasons for revision hip surgery: a retrospective review. Clin Orthop Relat Res. 2004;(429):188-92. 04.Harris WH. Conquest of a worldwide human disease: particle-induced periprosthetic osteolysis. Clin Orthop Relat Res. 2004;(429):39-42. 05.Bauer TW. Particles and periimplant bone resorption. Clin Orthop Relat Res. 2002;(405):138-43. 06.Böhler M, Kanz F, Schwarz B, et al. Adverse tissue reactions to wear particles from Co-alloy articulations, increased by alumina-blasting particle contamination from cementless Ti-based total hip implants. A report of seven revisions with early failure. J Bone Joint Surg Br. 2002; 84:128-36. 07.van der Weegen W, Sijbesma T, Hoekstra HJ, Brakel K, Pilot P, Nelissen RG. Treatment of pseudotumors after metal-on-metal hip resurfacing based on magnetic resonance imaging, metal ion levels and symptoms. J Arthroplasty. 2014;29:416-21. 08.Kurtz SM, Gawel HA, Patel JD. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:2262-77. 09. Jacobs CA, Christensen CP, Greenwald AS, McKellop H. Clinical performance of highly cross-linked polyethylenes in total hip arthroplasty. J Bone Joint Surg Am. 2007;89: 2779-86. 10.García-Rey E, García-Cimbrelo E. Polyethylene in total hip arthroplasty: half a century in the limelight. J Orthop Traumatol. 2010;11:67-72. 11.Onsten I, Carlsson AS, Ohlin A, Nilsson JA. Migration of acetabular components, inserted with and without cement, in one-stage bilateral hip arthroplasty. A controlled, randomized study using roentgenstereophotogrammetric analysis. J Bone Joint Surg Am. 1994;76:185-94. 12.Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518-28. 13. Schmalzried TP, Kwong LM, Jasty M, et al. The mechanism of loosening of cemented acetabular components in total hip arthroplasty. Analysis of specimens retrieved at autopsy. Clin Orthop Relat Res. 1992;(274):60-78. 14.Maloney WJ, Galante JO, Anderson M, et al. Fixation, polyethylene wear, and pelvic osteolysis in primary total hip replacement. Clin Orthop Relat Res. 1999;(369):157-64. 15. Sochart DH. Relationship of acetabular wear to osteolysis www.hipandpelvis.or.kr 221

Hip Pelvis 27(4): 216-222, 2015 and loosening in total hip arthroplasty. Clin Orthop Relat Res. 1999;(363):135-50. 16.Oparaugo PC, Clarke IC, Malchau H, Herberts P. Correlation of wear debris-induced osteolysis and revision with volumetric wear-rates of polyethylene: a survey of 8 reports in the literature. Acta Orthop Scand. 2001;72:22-8. 17.McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res 1999;17:157-67. 18.Kang JS, Kim RS. Metal on polyethylene in total hip arthroplasty. J Korean Hip Soc. 2010;22:241-6. 19. McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty. A randomized controlled trial. J Bone Joint Surg Am. 2009;91:773-82. 20.Lee JH, Lee BW, Lee BJ, Kim SY. Midterm results of primary total hip arthroplasty using highly cross-linked polyethylene: minimum 7-year follow-up study. J Arthroplasty. 2011;26:1014-9. 21. Sychterz CJ, Engh CA Jr, Shah N, Engh CA Sr. Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J Bone Joint Surg Am. 1997; 79:1040-6. 22.Sychterz CJ, Engh CA Jr, Yang A, Engh CA. Analysis of temporal wear patterns of porous-coated acetabular components: distinguishing between true wear and socalled bedding-in. J Bone Joint Surg Am. 1999;81:821-30. 23.Devane PA, Horne JG, Martin K, Coldham G, Krause B. Three-dimensional polyethylene wear of a press-fit titanium prosthesis. Factors influencing generation of polyethylene debris. J Arthroplasty. 1997;12:256-66. 24.Young AM, Sychterz CJ, Hopper RH Jr, Engh CA. Effect of acetabular modularity on polyethylene wear and osteolysis in total hip arthroplasty. J Bone Joint Surg Am. 2002;84-A:58-63. 25. Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH. In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty. J Arthroplasty. 2005;20(7):880-6. 26.Hermida JC, Bergula A, Chen P, Colwell CW Jr, D Lima DD. Comparison of the wear rates of twenty-eight and thirty-two-millimeter femoral heads on cross-linked polyethylene acetabular cups in a wear simulator. J Bone Joint Surg Am. 2003;85-A:2325-31. 222 www.hipandpelvis.or.kr