The noninvasive evaluation of forward flow areas for

Similar documents
PISA Evaluation of Mitral Regurgitation. Raymond Graber, MD Cardiac Anesthesia Group University Hospitals Case Medical Center 4/07/2011

Assessment of the severity of aortic stenosis depends on measurement of. Aortic Stenosis: Physics and Physiology What Do the Numbers Really Mean?

In the clinical management of patients with ventricular

Journal of the American College of Cardiology Vol. 33, No. 6, by the American College of Cardiology ISSN /99/$20.

aortic regurgitation, vena contracta area, vena contracta width, live three-dimensional echocardiography

Quantifying Aortic Regurgitation

Non-invasive estimation of pressure gradients in regurgitant jets: an overdue consideration

An understanding of the many factors involved in the

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics

The flow convergence method has recently been

A Magnetic Resonance Imaging Method for

Comparison of Vena Contracta Width by Multiplane Transesophageal Echocardiography With Quantitative Doppler Assessment of Mitral Regurgitation

New 3D Quantification of Mitral Regurgitation Severity. Judy Hung, MD Cardiac Ultrasound Laboratory Massachusetts General Hospital Boston, MA

Key Words Blood pressure pressure recovery Aortic valve stenosis subaortic stenosis Echocardiography, transesophageal, transthoracic

Flow-Dependent Changes in Doppler-Derived Aortic Valve Effective Orifice Area Are Real and Not Due to Artifact

In Vitro Two-Dimensional Echocardiographic Imaging of a Stented Porcine Bioprosthetic Valve: The Bent Strut Artifact

Mitral regurgitation (MR) is a

Policy #: 222 Latest Review Date: March 2009

Disclosures Rebecca T. Hahn, MD, FASE

Left atrial function. Aliakbar Arvandi MD

Non-invasive estimation of the mean pressure

How to assess ischaemic MR?

3D Printing & Echocardiography

Valvular Regurgitation: Can We Do Better Than Colour Doppler?

Improvements in outcomes for heart valve surgery, in particular mitral valve repair, have dictated

Assessment of Hemodynamics Properties of a New-Type Artificial Heart Valve Prosthesis Using Catheterization and Echocardiography

Transvalvular pressure gradients (TPG) and valve effective

Georgios C. Bompotis Cardiologist, Director of Cardiological Department, Papageorgiou Hospital,

Comprehensive Echo Assessment of Aortic Stenosis

Appendix II: ECHOCARDIOGRAPHY ANALYSIS

Tissue Doppler Imaging in Congenital Heart Disease

Prosthesis-Patient Mismatch or Prosthetic Valve Stenosis?

A Practical Approach to Prosthetic Valves

3D Printing & Echocardiography

Course Learning Objectives Sunday, February 17 Friday, February 22

Outline. EuroScore II. Society of Thoracic Surgeons Score. EuroScore II

Reproducibility of Proximal Isovelocity Surface Area, Vena Contracta, and Regurgitant Jet Area for Assessment of Mitral Regurgitation Severity

The Effect of Mitral Valve on Left Ventricular Flow

Doppler echocardiography is currently the

Echocardiographic Evaluation of Aortic Valve Prosthesis

Determinants of symptoms and exercise capacity in aortic stenosis: a comparison of resting haemodynamics and valve compliance during dobutamine stress

What are the best diagnostic tools to quantify aortic regurgitation?

Journal of the American College of Cardiology Vol. 41, No. 3, by the American College of Cardiology Foundation ISSN /03/$30.

Prof. JL Zamorano Hospital Universitario Ramón y Cajal

DOPPLER HEMODYNAMICS (1) QUANTIFICATION OF PRESSURE GRADIENTS and INTRACARDIAC PRESSURES

Direct Planimetry of Mitral Valve Regurgitation Orifice Area by Real-time 3D Transesophageal Echocardiography

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03

Methods XIAODONG ZHOU, MD, MICHAEL JONES, MD,* TAKAHIRO SHIOTA, MD, IZUMI YAMADA, MD,* DAG TEIEN, MD, DAVID J. SAHN, MD, FACC

Doppler-Echocardiographic Assessment of Carbomedics Prosthetic Valves in the Mitral Position

Aortic stenosis aetiology: morphology of calcific AS,

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Uncommon Doppler Echocardiographic Findings of Severe Pulmonic Insufficiency

Patterns of Normal Transvalvular Regurgitation in Mechanical Valve Prostheses

Team 19 Kaitlyn Clarke Brittany DePoi Andrew Reynolds Adarsha Selvachandran. Client: Dr. Wei Sun

ADoppler echocardiographic investigation is now frequently used

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

A Validation Study of Aortic Stroke Volume Using Dynamic 4-Dimensional Color Doppler: An In Vivo Study

Medical device design using Computational Fluid Dynamics (CFD)

Prosthetic valve dysfunction: stenosis or regurgitation

When Does 3D Echo Make A Difference?

Methods Forty patients representing all consecutive patients referred to. our laboratory on randomly selected days were chosen for

Journal of the American College of Cardiology Vol. 40, No. 8, by the American College of Cardiology Foundation ISSN /02/$22.

Aortic stenosis (AS) is common with the aging population.

Left Ventricular Volumes from Three-Dimensional. Echocardiography by Rapid Freehand Scanning using

ACUSON Sequoia C512 Echocardiography System The ultimate in imaging

Effects of Bileaflet Mechanical Mitral Valve Rotational Orientation on Left Ventricular Flow Conditions

Ultrasound 10/1/2014. Basic Echocardiography for the Internist. Mechanical (sector) transducer Piezoelectric crystal moved through a sector sweep

Back to Basics: Common Errors In Quantitation In Everyday Practice

PART II ECHOCARDIOGRAPHY LABORATORY OPERATIONS ADULT TRANSTHORACIC ECHOCARDIOGRAPHY TESTING

Color Doppler Echocardiographic Assessment of Valvular Regurgitation in Normal Infants

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction

Part II: Fundamentals of 3D Echocardiography: Acquisition and Application

Journal of the American College of Cardiology Vol. 44, No. 9, by the American College of Cardiology Foundation ISSN /04/$30.

Professors Carpentier and McGoon Mechanism, resulting from the disease Severity of regurgitation, resulting from the mechanism Echo

left heart catheterization

Diagnostic approach to heart disease

Imaging Assessment of Aortic Stenosis/Aortic Regurgitation

25 different brand names >44 different models Sizes mm

Comments restricted to Sapien and Corevalve 9/12/2016. Disclosures: Core Lab contracts with Edwards Lifesciences, Middlepeak, Medtronic

Echocardiographic evaluation of mitral stenosis

CAS SECTION 3 PASSPORT

TAVR: Echo Measurements Pre, Post And Intra Procedure

IS PVR THE RIGHT METRIC FOR RV AFTERLOAD?

MITRAL STENOSIS. Joanne Cusack

Chapter 20. JACC Cardiovasc Imaging 2009;2:

Three-dimensional echocardiography in the clinical world

Correlation Between Vortex Ring Formation and Mitral Annulus Dynamics During Ventricular Rapid Filling

2/2/2011. Strain and Strain Rate Imaging How, Why and When? Movement vs Deformation. Doppler Myocardial Velocities. Movement. Deformation.

ICE: Echo Core Lab-CRF

True morphology of mitral regurgitant flow assessed by three- dimensional transesophageal echocardiography

Quantification of Mitral Regurgitation by Automated Cardiac Output Measurement: Experimental and Clinical Validation

The Doppler Examination. Katie Twomley, MD Wake Forest Baptist Health - Lexington

By: Stephanie Bendtsen, Joseph Calderan, and Celeste Dupont Team 17 Client: Dr. Sun

Image Analysis and Cytometry in Three-Dimensional Digital Reconstruction of Porcine Native Aortic Valve Leaflets

Effect of Heart Rate on Tissue Doppler Measures of E/E

Surgery for Acquired Cardiovascular Disease

Revealing new insights. irotate electronic rotation and xplane adjustable biplane imaging. Ultrasound cardiology. irotate and xplane

Planimetric and continuity equation assessment of aortic valve area (AVA): comparison between cardiac magnetic resonance (cmr) and echocardiography

Prof. JL Zamorano Hospital Universitario Ramón y Cajal

Transcription:

A Method for Determining the Reference Effective Flow Areas for Mechanical Heart Valve Prostheses In Vitro Validation Studies Robin Shandas, PhD; Jeffrey Kwon, MSEE; Lilliam Valdes-Cruz, MD Background The anatomic opening area (AOA) is usually reported as the primary index of mechanical heart valve function. Because flow contracts immediately distal to an orifice as a result of the vena contracta effect, AOA may not be a good measure of true effective flow area. Methods and Results Laser flow imaging was used to visualize the contraction in the jet flow stream as it passed through bileaflet mechanical valves under steady and pulsatile conditions. Such visualization allowed clear measurement of the individual vena contracta areas (VCAs) of the 3 valve orifices. VCAs for side orifices were larger (94 2% of AOA) than those through the central orifice (34 8%). Formation of large radial vortices around the leaflet tips constricted the central orifice flow stream and appeared to be the main reason for smaller central VCA. Total VCA remained constant until 0.5 orifice diameters ( 1.0 cm) downstream, beyond which cross-sectional area increased as a result of entrainment of receiving chamber flow. Total VCA was larger for steady flow (89.6 2.7% of AOA) than for pulsatile flow (76.3 5.0% of AOA). Conclusions This study further clarifies flow dynamics through bileaflet mechanical valves and provides previously unavailable reference information on VCAs for these valves. Such information should aid clinicians in explaining Doppler-derived and catheter-measured pressure discrepancies, validating clinical techniques for quantifying effective flow areas, and optimizing valve size for implantation. The method should also be useful for comparative studies of different valve designs. (Circulation. 2000;101:1953-1959.) Key Words: blood flow valves echocardiography The noninvasive evaluation of forward flow areas for prosthetic heart valves remains an important goal for cardiologists because such information is directly related to valve function. Echocardiographic determination of anatomic orifice area may not represent the available flow area through the valve because of the vena contracta effect, which is a constriction in the flow stream and flow area as it passes through an orifice. In this regard, it is the effective orifice area, ie, the cross-sectional area of the flow at the vena contracta, that truly reflects the area available for flow. There is much confusion regarding how best to measure vena contracta area, whether vena contracta area changes with flow, and the exact definition of vena contracta area. Accurate information on the vena contracta areas for bileaflet mechanical valves has not been available to date. A clear understanding of vena contracta dynamics and information on true vena contracta areas would help clarify discrepancies between catheter-measured and Doppler-derived pressure drops through these valves, 1,2 refine existing clinical techniques such as the Gorlin and Gorlin formula and the Doppler continuity equation for effective flow area measurement, 3 5 provide reference data for validating new flow quantification techniques such as color Doppler and MRI, aid in optimizing prosthetic valve size during valve replacement surgery, and establish an important clinical database for comparison with clinically measured areas, similar to the tables available for normal pressure drops for various valve sizes and models. 6 The purpose of this study, therefore, was to elucidate vena contracta dynamics and areas for bileaflet mechanical valves with precise laser flow visualization (LFV) techniques. There were 4 aims for this study: (1) to validate the LFV method for measurement of vena contracta areas (effective flow areas), (2) to visualize the flow dynamics distal to bileaflet mechanical valves, (3) to determine how far downstream the vena contracta extends for such valves, and (4) to examine the effect of flow rate on vena contracta area. Methods Experimental Setup Figure 1 illustrates the experimental setup. A dual-chambered transparent acrylic in vitro model was used for these studies. Flow media used was a mixture of deionized water and 30% (by volume) Received June 14, 1999; revision received November 3, 1999; accepted December 1, 1999. From The Children s Hospital, Denver, Colo. Correspondence to Robin Shandas, PhD, Cardiovascular Flow Research Laboratory, The Children s Hospital, 1056 E 19th. Ave, B-100, Denver, CO 80218. E-mail Shandas.Robin@tchden.org 2000 American Heart Association, Inc. Circulation is available at http://www.circulationaha.org 1953

1954 Circulation April 25, 2000 Figure 1. Experimental setup. Dual-chambered in vitro flow model was used. Valves were mounted into central plate. LFV was performed by illuminating distal jet core region with thin laser sheet. glycerin mixed with neutrally buoyant silver-coated glass microparticles (mean diameter, 8 m; fluid viscosity, 3.5 cp; density, 1.04 g/ml). Steady (30 to 100 ml/s in increments of 8 ml/s) and pulsatile (60 bpm; 40 to 80 ml/beat) flows were directed through the model. Bileaflet mechanical valves (St Jude Medical Inc) were placed in the central plate of the model. Valves had nominal diameters of 25, 29, and 33 mm (cross-sectional areas, 2.45, 3.03, and 4.06 cm 2 ). Instantaneous flow rates were measured with a calibrated ultrasonic flowmeter (model T109, Transonics Inc). Stroke volumes were obtained by digital integration of instantaneous flows. Proximal and distal chamber pressures were measured with solid-state pressure transducers (2.5F, model SPR 524, Millar Instruments Inc). Pressure and flow information was digitized with a data acquisition system (model NB MIO 16X, National Instruments Inc) and displayed on a Macintosh microcomputer with LabView (National Instruments Inc) software. Mean pressure gradients between proximal and distal chambers varied from 1.5 to 4 mm Hg with the valve in the open position and from 10 to 24 mm Hg in the closed position. These gradients correspond to previously published data for St Jude valves of the sizes studied. 2,5 LFV of the Vena Contracta An Argon-Ion (5 W) laser (model 909, American Laser Co) was used to illuminate perpendicular cross sections of the jet immediately distal to the valve (Figure 1). A cylindrical lens, mounted at the end of a fiberoptic cable, expanded the laser beam into a thin (0.5 mm) sheet of finely focused light. The lens was mounted onto a horizontal positioning system that could be moved in increments of 0.1 cm, thereby allowing any perpendicular cross section of the jet core to be illuminated. Cross sections of the jet core region from the tip of the valve leaflets to 2.0-cm diameter downstream were visualized. A digital video camera recorded the illuminated cross-sectional flow images from an orthogonal window, and these images were recorded onto super VHS videotape for subsequent measurements. Resolution of the video camera was 580 412 pixels, which corresponds to 0.1-mm resolution in the vertical axis and 0.07-mm resolution in the horizontal axis. Temporal resolution of the camera was standard video rate (33 ms). Measurement of Distal Jet Core Areas of Bileaflet Valves With LFV An offline measurement system (ImageVue Workstation, Nova Microsonics Inc) was used for measurement of distal jet core areas. The jet core region was defined as the region immediately downstream of the orifice and comprising the fresh flow entering through the orifice. Therefore, the cross-sectional area of the flow at this point represented the distal jet core area. Figure 2A shows a typical jet core image, and Figure 2B shows the planimetry technique. The jet core region was clearly distinguished for each orifice from the surrounding flow by the strong particle reflections within the core. Within the immediate distal region, this en face view also revealed the border of the valve housing in the background, revealing that the jet core area is indeed smaller than the anatomic valve area. Calibration of length scales with the known dimensions of the valve ring was performed before each experiment. For each frame, the jet core area for each orifice was measured 3 times (Figure 2B). The region of vortical flow (Figure 2A) was not included in the area measurement because this flow is radial and does not contribute to the transorifice flow component. Vortical flow was identified on the video images by the existence of swirling flow within the image plane. Such flow patterns are imaged as straight or curved streaks; by contrast, reflective particles within the axial jet core flow are imaged only at the instant they cross the illumination plane, so they are displayed as densely packed particles with no streaking. Total jet core area was obtained by summation of the individual core areas. Validation of LFV Method to Measure Vena Contracta Areas To ensure that the LFV method provided true vena contracta areas, we performed a validation study using 3 circular orifices with areas comparable to the areas of the bileaflet valves tested (1.78, 3.14, and 4.90 cm 2 ) and steady flow (25 to 100 ml/s, n 5) conditions. Because the vena contracta area represents the cross-sectional area of the fresh flow entering through the orifice, no entrainment flow, ie, suction of receiving chamber fluid into the jet, should be present. Therefore, measurement of the velocities across the cross section of the vena contracta should allow differentiation of the vena contracta flow region, where velocities would be high, from the background flow region, where velocities would be very low. We have used this approach to identify and measure the vena contracta for computer modeling studies of flow through stenotic valves. 7 This method has also been used to identify the vena contracta width on color Doppler images of regurgitant flows. 8 Because of the relatively poor lateral resolution of color Doppler, 9 we used the high-resolution technique Figure 2. Typical LFV image (A) and planimetry technique (B). Flow direction is out of page. Flow through side (S) and central (C) orifices can be clearly seen (A). Swirling vortex flow (V) can also be seen at top and bottom portions of vena contracta. B, Measurement of vena contracta area. H indicates valve housing.

Shandas et al Flow Areas of Mechanical Valves 1955 Figure 3. DPIV was performed on steady flow through circular orifices to validate LFV method. A, Typical DPIV image of flow through 1.5-cm-diameter orifice (flow is from left to right). Solid white lines show orifice location and size. Flow stream contraction can be clearly seen, and width of vena contracta (VCW) can be measured. B, Excellent agreement was found between 2 techniques. of digital particle image velocimetry (DPIV) to obtain crosssectional velocity profiles of the jet immediately distal to the orifice. DPIV is now a well-established, highly robust technique to measure whole-field velocity vectors, with excellent spatial resolution ( 1 mm), high velocity range (0.01 to 10 m/s), and quantitative output of results. Because of the limited scope of this paper, the reader is referred elsewhere for details on DPIV. 10 Validation study data are given in the Results section. Statistical Analysis All values given are mean SD. Linear regression of jet core area versus downstream distance was performed to examine whether jet core area changed significantly with downstream distance. Vena contracta area was also analyzed against flow rate (stroke volume for pulsatile flow conditions) through linear regression. To facilitate analysis among valve sizes, the contraction coefficient, defined as vena contracta area divided by manufacturer s anatomic area, was also calculated. Contraction coefficients for different valve sizes were compared by use of 1-way ANOVA. Intraobserver variability was calculated by comparing the differences in 3 consecutive measurements of vena contracta area performed by the main observer. A subset of the measurements (25%) was repeated by a second observer, blinded to the previous measurements, to calculate interobserver variability. The level of significance for all statistical tests was set at 5%. Results Validation of LFV Method: Circular Orifices Figure 3 displays the results from the DPIV validation studies, showing a typical DPIV image of the vena contracta

1956 Circulation April 25, 2000 Figure 4. Dynamic images of vena contracta area over pulsatile cycle for 25-mm valve (60 bpm; stroke volume 70 ml/beat). At cycle initiation, side orifices fill first, followed by central orifice flow. Swirling vortex structures can be seen extending into central orifice flow region. Flow area through side orifices is more stable, ie, it retains its initial structure over majority of pulsatile, than flow through the central orifice. region (Figure 3A) and linear regression between DPIV- and LFV-measured vena contracta areas (Figure 3B). Velocity magnitude in Figure 3A is represented by both vector length and vector color. The reduction in flow area as it passes through the orifice can be clearly seen, and the vena contracta width can be easily identified. It is only after 1 orifice diameter (15 mm here) that we see entrainment flow (shown in dark blue) entering the jet. This suggests that vena contracta area can be accurately measured until 1 diameter downstream of the orifice. Excellent agreement between the 2 methods was also found (Figure 3B). For each orifice size, the close clustering of data points indicates that flow rate does not significantly affect DPIV- or LFV-measured areas. Mean percent difference between the 2 methods was 0.34 1.9%. Mean contraction coefficients (vena contracta area/anatomic area) for these orifices ranged from 0.81 to 0.79 and agree with previously published engineering standards for orifice flow. 11 Measurement of Distal Jet Core Areas: Bileaflet Valves Figure 4 displays the dynamic change in vena contracta area over the pulsatile cycle for the 29-mm valve obtained with the LFV technique. Time increments are shown on each image. The flow is directed out of the page. The 3 individual jet core areas and vortex structures at the top and bottom edges of side and central flow areas can be clearly seen. At the beginning of the pulsatile cycle, flow is initially directed through the side orifices (Figure 4B). The central orifice begins to fill up 66 ms into the cycle (Figure 4C), after which side and central orifice vena contracta areas remain constant (Figure 4D through 4F) until the leaflets begin to close (t 528 to 594 ms; Figure 4G through 4I). After the initial filling and for most of the pulsatile cycle, the side orifices fill to nearmaximal capacity, while the vortex structures emanating from the leaflets substantially constrict flow area through the

Shandas et al Flow Areas of Mechanical Valves 1957 Figure 5. Total LFV measured jet core area vs downstream distance for 25-mm valve at steady flow rate of 70 ml/s. Jet core area remained constant until 0.5 valve diameters (1.0 cm) distal. Beyond this, entrainment into jet began expanding total jet core area. Values are mean SD. central orifice (see frames from 99 to 330 ms, Figure 4D through 4F). Three separate jet cores could be visualized until 1.8 cm downstream of the orifice, after which the 3 jets coalesced into 1. Total vena contracta area remained constant until 0.5 valve diameters downstream of the leaflet tips (Figure 5). This corresponded to 1 cm downstream of the leaflet tips for all valve sizes tested. Linear regression analysis revealed no significant change in the minimum jet core area up to 1 cm downstream over all flow and orifice conditions (P NS). Beyond this point, flow within the receiving chamber began mixing with flow in the jet core by the process of entrainment. This caused the cross-sectional area of the jet to increase and an obliteration of orifice shape. These findings are similar to DPIV results seen for the simple circular orifice studies, except that for the circular orifices, jet core area remains constant for a greater distance (1 orifice diameter, 1.5 to 2.5 cm) downstream. The difference is presumably due to the complex 3-dimensional (3D) nature of the bileaflet mechanical valve orifices. No significant change in vena contracta area with steady flow rate was observed (Figure 6A). Figure 6B shows instantaneous vena contracta area measured over the pulsatile cycle for the 25-mm valve at all stroke volumes. Similar results were seen for the other valve sizes. Very little change in vena contracta area was observed over most of the pulsatile cycle. Figure 6C shows mean contraction coefficient (averaged over the pulsatile cycle) as a function of stroke volume. As for the steady flow conditions, vena contracta areas did not vary significantly with flow rate. Mean vena contracta areas for pulsatile flow were smaller (mean C c 0.763 0.05) than for steady flow (mean C c 0.896 0.027, P 0.0001). Interobserver and intraobserver variabilities for vena contracta area measurements were 3.8% and 4.2% for steady flow and 5.7% and 6.3% for pulsatile conditions, respectively. Discussion In this investigation, we first validated the use of a LFV technique to measure actual vena contracta areas and then Figure 6. A, Contraction coefficient as function of steady flow rate, showing no significant change in vena contracta area with flow. B and C, Vena contracta areas during pulsatile flow. As in steady flow conditions, no significant change was seen in vena contracta area either over pulsatile cycle or with stroke volume. used this method to visualize and measure vena contracta areas for bileaflet mechanical valves. The vena contracta area represents a contraction in the edges of the flow streamlines as they move through the orifice. 12 For orifices without a smoothly tapering proximal geometry, inertia prevents proximal streamlines entering the orifice from the side from changing direction instantly. As the flow passes through the orifice, the streamlines change direction to run parallel to the

1958 Circulation April 25, 2000 main flow direction but not before squeezing the main flow and causing a constriction (vena contracta) in the crosssectional area of flow immediately distal to the orifice. The vena contracta area thus represents the most direct measure of the flow area for a particular orifice or valve. Many attempts have been made to use the color Doppler image of the jet region immediately distal to the orifice (proximal jet region) as the clinical measure of vena contracta width for regurgitant and stenotic lesions. 7 9,13,14 One problem with the use of color Doppler is that the poor lateral resolution may not accurately delineate the region separating vena contracta and receiving chamber flow. Our DPIV validation technique provides significantly higher spatial resolution and clearly separates vena contracta flow from background flow (Figure 3). The DPIV technique could not be used to determine vena contracta areas for the bileaflet valves because of the complex flow field, which would require the highly labor- and time-intensive task of full 3D reconstruction of DPIV velocities. The validation studies show that the LFV method measures vena contracta areas accurately and contains no flow dependence. Additionally, the DPIV results reveal that the vena contracta region remains relatively constant until 1 orifice diameter downstream. LFV was then performed on the bileaflet valves. Several observations were noted. First, for pulsatile conditions, flow entered through the side orifices before the central orifice. Second, side orifices fill to nearmaximal capacity for steady and pulsatile flows, whereas central orifice flow is constricted severely because of the presence of vortex structures. Third, total vena contracta area remained constant until 0.5 valve diameters downstream of the orifice for all conditions. Fourth, there was no significant change found in vena contracta area with flow rate for steady or pulsatile flow state. Finally, steady-flow vena contracta areas were higher than mean pulsatile flow areas. The dynamics of the vena contracta region have been studied in the design of orifice flow meters, regurgitant and stenotic cardiac lesions, and various jet flow investigations. 7 9,11,13,14 Fluid mechanics predicts that the vena contracta region for an orifice should remain constant until a short distance downstream; this is called the jet core region. 15 Within this region, the viscous shear layer developed at the jet boundary has not penetrated into the jet core. The result is that the shape and size of the jet core mimic orifice shape and size. Our LFV method takes advantage of this fact by visualizing the flow cross section at the jet core region. The length of the vena contracta region for bileaflet mechanical valves is even smaller ( 0.5 diameter) than that for planar orifices ( 1 diameter). This could be due to the 3D nature of the valve orifice, which allows the initial development of the shear layer within the valve rather than immediately downstream. This does not present a practical problem, however, because for most valve sizes ( 19 mm), the region for vena contracta measurements should extend to 1 cm. The question of whether vena contracta area varies with flow rate remains controversial. Investigators have found variations in Doppler- and Gorlin-calculated valve areas in in vitro and clinical studies. 16 18 Our results for these prosthetic valves reveal no significant variation in vena contracta area with mean flow rate or stroke volume. We have observed a similar lack of flow dependence in previous studies using computer modeling and in vitro experimentation of flow through valvular stenosis. 7,9 This was also confirmed by the DPIV data. Our results provide evidence that variations in Doppler or Gorlin areas with flow may be a consequence of methodology limitation rather than true vena contracta variation. It has been known that flow dynamics and downstream velocities are different for the side versus central orifices for the bileaflet mechanical valve. 2,19,20 However, no direct visualization of the actual vena contracta flow area through the valves has been reported. We observed stable flow through the side orifices but disturbed flow through the central orifice. Central flow disturbances are presumably due to the 3D expansion-type shape of the central orifice, which, like a flow diffuser, causes an increase in pressure and the potential for adverse pressure gradients, flow separation, and vortex formation to occur. 15 In fact, we observed severe vortex-type disturbances within central orifice flow. The 2 side orifices are primarily responsible for transporting most of the flow through the valve. Study Limitations There are several limitations that should be acknowledged in this study. This is an in vitro study, and as with any experimental study, it was not possible to reproduce all physiological variables. For example, we did not simulate the complex physiological loading of the valve in open and closed states. The valve was mounted in a rigid position, and the working fluid simulated blood under high shear (constant viscosity). The pulsatile flow conditions may not absolutely mimic physiological pressure and flow waveforms. Clinical Implications The question of how best to evaluate mechanical valve function remains controversial. Although measurement of transvalvular pressure gradients with Doppler remains the method of choice, there is still confusion as to the meaning behind this parameter. For example, investigators have shown that the pressure gradient differs between side and central orifices for the bileaflet mechanical valve. 2,20 Maximal pressure drop and hence maximum velocity have been shown to occur through the central orifice 2,20 ; however, sampling this gradient may produce a false estimate of overall pressure drop through the valve because the central orifice contains the smallest vena contracta area. Our results point to the side orifices as the most important conduit for flow and indicate that pressure gradients should be measured through these orifices. This is echoed in a clinical study by Vandervoort et al, 20 who suggested sampling the side orifice pressure gradient as the most reliable reflector of overall transvalvular dynamics. The measurement of gradients, however, can be problematic in the presence of changing preload and/or afterload in which the pressure gradient may vary with no alteration in valve function. Therefore, we believe that effective flow area, if accurately measured, provides the most accurate means of clinically evaluating valve function. Recently, Leung et al 21 used proximal flow convergence methods to estimate effective orifice area for bileaflet mechanical

Shandas et al Flow Areas of Mechanical Valves 1959 valves in patients. These authors point to the lack of a universal gold standard for comparing clinically measured effective orifice areas as a continuing limitation for clinical studies, especially because results from Gorlin, pressure half-time, and continuity may vary widely for the same valve. The vena contracta results presented here can be used as reference data for comparing effective orifice areas measured with existing clinical methods, especially because we have shown that actual vena contracta areas exhibit no flow dependence over a wide range of flow rates. Finally, our results should lay groundwork for the development of new clinical imaging techniques for vena contracta visualization and measurement. We have shown using in vitro studies that 3D echocardiography coupled with contrast provides a novel yet accurate means of noninvasively visualizing and quantifying vena contracta area for stenotic lesions. 22 The contrast method is a direct parallel of our LFV technique in that the ultrasound sheet is projected at the short axis of flow (perpendicular cross section) immediately distal to the valve and that the contrast agent acts as the reflective particles that will be imaged echocardiographically at the instant they cross the viewing plane. Such echo-based flow visualization methods should allow direct visualization and quantification of the vena contracta in the clinical situation. Acknowledgments This study was supported in part by grants from the National Institutes of Health (NHLBI HL-55243) and The Children s Hospital Research Institute. References 1. Burstow DJ, Nishimura RA, Bailey KR, Reeder GS, Holmes DR Jr, Seward JB, Tajik AJ. Continuous wave Doppler echocardiographic measurement of prosthetic valve gradients: a simultaneous Doppler-catheter correlative study. Circulation. 1989;80:504 514. 2. Solowiejczyk DE, Yamada I, Cape EG, Manduley RA, Gersony WM, Jones M, Valdes-Cruz LM. Simultaneous Doppler and catheter transvalvular pressure gradients across St Jude bileaflet mitral valve prosthesis: in vivo study in a chronic animal model with pediatric valve sizes. JAmSoc Echocardiogr. 1998;11:1145 1154. 3. Chambers JB, Cochrane T, Black MM, Jackson G. The Gorlin formula validated against directly observed orifice area in porcine mitral bioprostheses. J Am Coll Cardiol. 1989;13:348 353. 4. Nanda NC, Cooper JW, Mahan EF, Fan PH. Echocardiographic assessment of prosthetic valves. Circulation. 1991;84(suppl I):I- 228 I-239. 5. Bitar JN, Lechin ME, Salazar G, Zoghbi WA. Doppler echocardiographic assessment with the continuity equation of St. Jude medical mechanical prostheses in the mitral valve position. Am J Cardiol. 1995;76:287 293. 6. Feigenbaum H. Echocardiography. 5 th ed. Baltimore, Md: Williams & Wilkins; 1993:678 681. 7. DeGroff C, Shandas R, Valdes-Cruz LM. Analysis of the effect of flow rate on the Doppler continuity equation for stenotic orifice area calculations: a numerical study. Circulation. 1998;97:1597 1605. 8. Hall SA, Brickner E, Willet DL, Irani WN, Afridi I, Grayburn PA. Assessment of mitral regurgitation severity by Doppler color flow mapping of the vena contracta. Circulation. 1997;95:636 642. 9. Shandas R, DeGroff C, Kwon J, Trujillo N, Gill E, Valdes-Cruz LM. Utility of three dimensional ultrasound Doppler flow reconstruction of the proximal jet to quantify effective orifice area: in vitro steady and pulsatile flow studies. J Am Soc Echocardiogr. 1998;11:313 321. 10. Willert C, Gharib M. Digital particle image velocimetry. Experiments Fluids. 1991;10:181 187. 11. Fried E, Idelchik IE. Flow Resistance: A Design Guide for Engineers. New York, NY: Hemisphere Publishing Corp; 1989. 12. Fox RW, McDonald AT. Introduction to Fluid Mechanics. 4th ed. New York, NY: John Wiley & Sons; 1992:354. 13. Baumgartner H, Schima H, Kuhn P. Value and limitations of proximal jet dimensions for the quantitation of valvular regurgitation. Circulation. 1993;87:841 848. 14. Ishii M, Jones M, Shiota T, Yamada I, Heinrich R, Holcomb SR, Yoganthan AP, Sahn DJ. Quantifying aortic regurgitation by using the color Doppler imaged vena contracta. Circulation. 1997;96:2009 2015. 15. Batchelor GK. An Introduction to Fluid Dynamics. New York, NY: Cambridge University Press; 1992:372 377. 16. Burwash IG, Thomas DD, Sadahiro M, Pearlman AS, Verrier ED, Thomas R, Kraft CD, Otto CM. Dependence of Gorlin formula and continuity equation valve areas on transvalvular volume flow rate in valvular aortic stenosis. Circulation. 1994;89:827 835. 17. Voelker W, Reul H, Nienhaus G, Stelzer T, Schmitz B, Steegers A, Karsch KR. Comparison of valvular resistance, stroke work loss, and Gorlin valve area for quantification of aortic stenosis: an in vitro study in a pulsatile aortic flow model. Circulation. 1995;91:1196 1204. 18. Tardiff J-C, Rodrigues AG, Hardy J-F, Leclerc Y, Petitclerc R, Mongrain R, Mercier L-A. Simultaneous determination of aortic valve area by the Gorlin formula and transesophageal echocardiography under different transvalvular flow conditions. J Am Coll Cardiol. 1997;29:1296 1302. 19. Chandran KB, Cabell GN, Khaligi B, Chen CJ. Laser anemometry measurements of pulsatile flow past aortic valve prostheses. J Biomech. 1983;16:865 873. 20. Vandervoort PM, Greenberg NL, Pu M, Powell KA, Cosgrove DM, Thomas JD. Pressure recovery in bileaflet heart valve prostheses: localized high velocities and gradients in central and side orifices with implications for Doppler-catheter gradient relation in aortic and mitral position. Circulation. 1995;92:3464 3472. 21. Leung DY, Wong J, Rodriguez L, Pu M, Vandervoort PM, Thomas JD. Application of color Doppler flow mapping to calculate orifice area of St. Jude mitral valve. Circulation. 1998;98:1205 1211. 22. Shandas R, Kwon J, Valdes-Cruz LM. Real time three dimensional volumetric ultrasound imaging of the vena contracta for stenotic valves using echo contrast: in vitro pulsatile flow studies. J Am Soc Echocardiogr. 1999;12:541 550.