Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

Similar documents
O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Matrix Factor Determination with the Waters Regulated Bioanalysis System Solution

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Jose Castro-Perez, Henry Shion, Kate Yu, John Shockcor, Emma Marsden-Edwards, Jeff Goshawk Waters Corporation, Milford, MA, U.S. and Manchester, UK

Meeting Challenging Requirements for the Quantitation of Regulated Growth Promoters Dexamethasone and Betamethasone in Liver and Milk

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

UPLC-MS/MS Analysis of Azole Antifungals in Serum for Clinical Research

Comparison of a UPLC Method across Multiple UHPLC Systems

A High Sensitivity UPLC/MS/MS Method for the Analysis of Clopidogrel and Clopidogrel Carboxylic Acid Metabolite in Human K 2 EDTA Plasma

INTRODUCTION CH 3 CH CH 3 3. C 37 H 48 N 6 O 5 S 2, molecular weight Figure 1. The Xevo QTof MS System.

Latest Innovations in LC/MS/MS from Waters for Metabolism and Bioanalytical Applications

Quantification of Budesonide Using UPLC and Xevo TQ-S

[ APPLICATION NOTE ] APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Rapid Lipid Profiling of Serum by Reverse Phase UPLC-Tandem Quadrupole MS

Probing for Packaging Migrants in a Pharmaceutical Impurities Assay Using UHPLC with UV and Mass Detection INTRODUCTION

[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

SEPARATION OF BRANCHED PFOS ISOMERS BY UPLC WITH MS/MS DETECTION

A UPLC/MS Approach for the Diagnosis of Inborn Errors of Metabolism

A RAPID AND SENSITIVE UPLC/UV/MS METHOD FOR SIMVASTATIN AND LOVA S TAT IN IN SU P P O RT O F C L E A NING VA L I DAT IO N S T U DIES

Profiling Flavonoid Isomers in Highly Complex Citrus Juice Samples Using UPLC Ion Mobility Time-of-Flight Mass Spectrometry

Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification

John Haselden 1, Gordon Dear 1, Jennifer H. Granger 2, and Robert S. Plumb 2. 1GlaxoSmithKline, Ware, UK; 2 Waters Corporation, Milford, MA, USA

Reducing Sample Volume and Increasing Sensitivity for the Quantification of Human Insulin and 5 Analogs in Human Plasma Using ionkey/ms

Analysis of the Non-Ionic Surfactant Triton-X Using UltraPerformance Convergence Chromatography (UPC 2 ) with MS and UV Detection

Development of a High Sensitivity SPE-LC-MS/MS Assay for the Quantification of Glucagon in Human Plasma Using the ionkey/ms System

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section

Method Development for the Analysis of Endogenous Steroids Using Convergence Chromatography with Mass Spectrometric Detection

The Application of UPLC/MS E for the Analysis of Bile Acids in Biological Fluids

[ APPLICATION NOTE ] Oasis PRiME HLB Cartridge for Cleanup of Infant Formula Extracts Prior to UPLC-MS/MS Multiresidue Veterinary Drugs Analysis

Rapid Analysis of Bisphenols A, B, and E in Baby Food and Infant Formula Using ACQUITY UPLC with the Xevo TQD

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Analysis of Rosuvastatin in Dried Blood Spot and Plasma Using ACQUITY UPLC with 2D Technology

PosterREPRINT RAPID, SELECTIVE SCREENING OF URINE SAMPLES FOR GLUCURONIDES BY LC/MS/MS INTRODUCTION ABSTRACT

Identification & Confirmation of Structurally Related Degradation Products of Simvastatin

Improving Selectivity in Quantitative Analysis Using MS 3 on a Hybrid Quadrupole-Linear Ion Trap Mass Spectrometer

Impurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF

Analysis of Food Sugars in Various Matrices Using UPLC with Refractive Index (RI) Detection

CHARACTERIZATION AND DETECTION OF OLIVE OIL ADULTERATIONS USING CHEMOMETRICS

Fast Separation of Triacylglycerols in Oils using UltraPerformance Convergence Chromatography (UPC 2 )

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

SYNAPT G2-S High Definition MS (HDMS) System

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

USP Method Transfer of Ziprasidone HCl from HPLC to UPLC

Chemical Analysis Business Operations Waters Corporation Milford MA

Analysis of N-Linked Glycans from Coagulation Factor IX, Recombinant and Plasma Derived, Using HILIC UPLC/FLR/QTof MS

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

Simultaneous Analysis of Intact Human Insulin and Five Analogs in Human Plasma Using μelution SPE and a CORTECS UPLC Column

[ APPLICATION NOTE ] Profiling Mono and Disaccharides in Milk and Infant Formula Using the ACQUITY Arc System and ACQUITY QDa Detector

Automated Lipid Identification Using UPLC/HDMS E in Combination with SimLipid

PosterREPRINT SIMULTANEOUS QUANTIFICATION OF PSYCHOTHERAPEUTIC DRUGS IN HUMAN PLASMA BY TANDEM MASS SPECTROMETRY

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Edgar Naegele. Abstract

Development of a Bioanalytical Method for Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid

UPLC/MS/MS Method for the Routine Quantification of Regulated and Non-Regulated Lipophilic Marine Biotoxins in Shellfish

PosterREPRINT EVALUATION OF A RF-ONLY STACKED RING BASED COLLISION CELL WITH AXIAL FIELD FOR THE LC-MS-MS ANALYSIS OF THE METABOLITES OF RABEPRAZOLE

A Metabolomics Approach to Profile Novel Chemical Markers for Identification and Authentification of Terminalia Species

Dr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Jody Dunstan, 1 Antonietta Gledhill, 1 Ailsa Hall, 2 Patrick Miller, 2 Christian Ramp 3. Waters Corporation, Manchester, UK 2

Direct Analysis of Urinary Opioids and Metabolites by Mixed-Mode µelution SPE Combined with UPLC/MS/MS for Forensic Toxicology

[ application note note ] ]

[Application Note] TOXICOLOGY SCREENING BY UPLC/PDA IN COMBINATION WITH AN EXTENSIVE COMPOUND LIBRARY

Polymer Additive Analysis by EI and APCI

LCMS-8030 Application Report. Steroids: Testosterone, Progesterone, Estradiol, Ethinylestradiol

Moving from targeted towards non-targeted approaches

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Increased Identification Coverage and Throughput for Complex Lipidomes

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Supplementary Information

Supporting information

Dienes Derivatization MaxSpec Kit

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Designer Cannabinoids

The Raptor HILIC-Si Column

Phospholipid characterization by a TQ-MS data based identification scheme

Agilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value

[ APPLICATION NOTE ] The Separation of 8 -THC, 9 -THC, and Their Enantiomers by UPC 2 Using Trefoil Chiral Columns INTRODUCTION APPLICATION BENEFITS

The use of mobile-device data to better estimate dynamic population size for wastewater-based epidemiology

Identification of Ginsenosides Using the SCIEX X500R QTOF System

Abstract. Introduction

[ APPLICATION NOTE ] UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research INTRODUCTION APPLICATION BENEFITS

More structural information with MS n

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

[application note] DIRECT TISSUE IMAGING AND CHARACTERIZATION OF PHOSPHOLIPIDS USING A MALDI SYNAPT HDMS SYSTEM

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

application Natural Food Colorants Analysis of Natural Food Colorants by Electrospray and Atmospheric Pressure Chemical Ionization LC/MS

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer

Challenges in Developing an Ultra-Sensitive Bioanalytical Method for Ethinylestradiol in Human Plasma

Lipid Class Separation Using UPC 2 /MS

Paul Silcock 1, Anna Karrman 2, and Bert van Bavel 2. Aim. Introduction. Sample preparation

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

LC/MS Analysis of Various Hydrophilic Compounds Using a Polymer-Based Amino Column - Shodex TM HILICpak TM VG-50 2D

1. Sample Introduction to MS Systems:

Transcription:

Robert Plumb, Michael D. Jones, and Marian Twohig Waters Corporation, Milford, MA, USA INTRODUCTION The detection and characterization of impurities and degradation products of an active pharmaceutical ingredient (API) are regulatory filing requirements. The detection and identification of impurities not only ensures medicine safety but can also be used as a fingerprint for patent protection and counterfeit drug analysis. Impurity characterization and identification is normally carried out using information-rich analytical techniques such as NMR and LC/MS. Analysis by LC/MS provides parent ion mass from full-scan MS and structural information from the fragments generated in MS/MS experiments. With traditional tandem quadrupole instrumentation, the generation of this data requires multiple experiments to obtain MS and MS/MS information. With these LIT MS systems, this would result in just 1 to 2 points across the peak, with the peaks either poorly defined or missed completely; thus slower, lower-resolution LC systems must be used, resulting in reduced throughput and lower data quality. The Waters Xevo TQ Mass Spectrometer is equipped with a novel collision cell design that is continuously filled with collision gas, allowing rapid switching between MS and MS/MS modes. The Xevo TQ MS is capable of operating at up to 1, Da/sec and can correctly define the very sharp peaks produced by UPLC, with more than 1 points across a 2-second-wide peak, even on a multi-scan experiment. This collision cell is capable of enhanced high-sensitivity operation in MS/MS mode. In this mode of operation, ions are constrained in the final third section of the collision cell using both DC and RF barriers. These ions are then ejected from the collision cell, in a controlled manner, from high to low m/z in synchronization with the scanning of the final resolving quadrupole. This increases the duty cycle of the instrument, resulting in enhanced sensitivity that is ideal for the detection and characterization of low-concentration impurities that may result in toxic effects. Modern Linear Ion Trap (LIT) mass spectrometers allow the collection of MS, multiple reaction monitoring (MRM), and MS/MS data in the same analytical run, allowing quantitative and qualitative data to be obtained simultaneously. However, the duty cycle of these instruments when switching between MS and MS/MS modes is typically 2 to 3 seconds. With modern high-resolution, sub-2-µm column particle chromatography such as UPLC, peak widths of 2 to 3 seconds are now commonplace. Figure 1. Xevo TQ Mass Spectrometer with the ACQUITY UPLC System. 1

E X P E R IM E N TA L To evaluate the performance of this system, the impurities of the common pharmaceutical drug quetiapine, used to treat biopolar disorder, was investigated using UPLC/MS/MS. LC /MS conditions LC system: ACQUITY UPLC Column: ACQUITY UPLC BEH C 18, 2.1 x 5 mm, 1.7 µm Column temp.: 65 C Flow rate: 8 µl/min Mobile phase A: 2 mm ammonium bicarbonate ph 1 Mobile phase B: Acetonitrile Gradient: 15 to 95 B/18 min MS system: Xevo TQ MS Ionization mode: ESI positive ion mode Capillary voltage: 3 V Collision energy: 15 ev RESULTS The unique collision cell design allows the Xevo TQ MS to be operated in several different modes of operation: full scan MS, MRM, as well as MS/MS mode. As the collision cell is continuously filled with collision gas, the instrument can rapidly switch between MS and MS/MS in the same analytical run. This allows MRM and MS scans to be performed in the same run. Combined with the high scan rate, this allows for rapid survey scans to be performed, such as MS neutral loss or parent ion, before switching to MS/MS. This high data-capture rate allows for the accurate definition of the peak, even with the very narrow peaks produced by UPLC. Figure 2 shows the UPLC/MS chromatogram produced in the analysis of an API batch of quetiapine at a concentration of 1 µg/ml. Here we can see that impurity peaks are 2 to 4 seconds wide at the base. The data shown in Figure 3 illustrates the number of scans achieved in MS and MS/MS modes. Maximizing LC peak definition In this example, the Xevo TQ MS was operated in ScanWave MS mode, switching to ScanWave DS (daughter ion scan) mode when a peak was detected above a user-defined threshold. In this mode of operation, the instrument selects the most intense peak in the MS spectrum and acquires MS/MS data on this peak before returning to MS mode. Since the collision cell is continuous filled with collision gas, there is a no delay in switching between MS and MS/MS modes. 9.81 17.26 1.9.29 9.24 17.5 17.1 17.66 15.79 5.65 8.37 13.88 18.7 19.38 5. 1. 15. 2. Figure 2. UPLC/MS analysis of quetiapine at 1 µg/ml. 2

We can see from this data that the instrument has acquired 9 points across the peak in MS mode, and 15 points across the peak in MS/MS mode despite the fact that the peak is only 2 seconds wide at the base. This high data-capture rate enables the Xevo TQ MS to perform high quality, data-dependent MS-to-MS/MS experiments in a UPLC timeframe with sufficient data points to accurately define the peak. This dual mode of operation can also be used to acquire full-scan MS data simultaneously with MRM data, or to detect a peak with precursor ion scanning before switching to MS/MS mode. 17.57 MS 17.23 16.63 17.78 15.61 34 36 38 4 42 44 46 48 17.58 MS/MS 17.49 Scan Scan 287 288 289 29 291 Precursor ion scanning The detection of new impurities, degradation products, or, in a DMPK study, drug metabolites, is often confounded by the signal from the matrix. To detect and visualize these analytes, the analytical chemist can use the specificity of the mass spectrometer. Since compounds can undergo fragmentation as a result of the degradation or metabolism process, the use of simple, predicted MRM transitions for common degradation/metabolism pathways may result in the non-detection of a potentially toxic impurity, degradation product, or metabolite. A more comprehensive way to detect these compounds is to monitor for the common fragment ions of the molecule of interest. The Xevo TQ MS s Survey Scan functionality utilizes the fast data-capture rate of the instrument to facilitate the collection of precursor ion data as well an MS/MS spectrum of the peaks detected. This functionality was used to evaluate a commercially-purchased API sample of quetiapine. The MS/MS spectra of quetiapine revealed that it gave rise to three major product ions having m/z values, 253, and 279. This data was used to detect drug-related impurities in the API batch by performing a Survey Scan analysis on each of these ions. The data collected for the parent ion chromatogram of m/z 279 is displayed in Figure 4. Here, we can see the presence of seven major peaks, six impurities and the quetiapine active peak, eluting with a retention time of 9.86 minutes. A similar analysis using the common fragment ion m/z to trigger data collection produced the chromatogram shown in Figure 5. With this fragment ion, a total of 12 peaks were detected and MS/MS data acquired. Figure 3. Rapid data collection is performed simultaneously in both MS and MS/MS modes. 2.5 5. 7.5 1. 12.5 15. 17.5 Figure 4. Survey Scan UPLC/MS analysis of quetiapine for ion m/z 279 in positive ion mode. 5.69 5.64 6.46 8.42 9.29 8.39 8.38 8.4 Figure 5. Survey Scan UPLC/MS analysis of quetiapine for ion m/z in positive ion mode. 9.25 9.86 9.84 1.94 9.89 9.9 1.12 1.72 12.28 13.99 15.1 17.63 17.91 17.15 17.25 18.28 2.5 5. 7.5 1. 12.5 15. 17.5 2. 3

The MS/MS spectra obtained from the peak eluting with a retention time of 5.6 minutes is displayed in Figure 6. This impurity has a m/z value of 4 amu and has been identified as the S-Oxide impurity of quetiapine. ScanWave Technology The detection of low-level impurities is becoming increasingly important, especially when monitoring potential genotoxins. Collection of the MS/MS spectrum from Survey Scan experiments, either precursor ion or common neutral loss, can be performed in two modes of operation: standard MS/MS or ScanWave MS/MS. As described previously, ScanWave Technology allows for increased sensitivity in the collection of MS/MS data. This increase in sensitivity is illustrated by the MS spectra obtained for the desthanol impurity of quetiapine (Figure 7). The top spectrum is obtained in standard MS/MS mode, while the lower spectrum is obtained in ScanWave MS/MS mode. In this example, we can see that the ScanWave MS/MS data is 13 times more sensitive than that in standard MS/MS mode. This increase is essential for the correct confirmation or identification of low-level impurities. 196 247 176 211 222 279 158 239 253 15 2 25 3 35 4 Figure 6. AutoScanWave MS/MS spectrum of S-Oxide of quetiapine eluting at 5.6 minutes. m/z 196 247 176 211 222 279 158 239 253 15 2 25 3 35 4 m/z Figure 7. Comparison of standard and ScanWave MS/MS sensitivity. 4

CONCLUSIONS The Xevo TQ MS provides unrivaled levels of sensitivity and functionality. The high data-capture rates of the instrument, and its unique collision cell design and ScanWave Technology, allows the maximum amount of data to be collected in one analytical run. This reduces the number of experiments needed to make a decision, allowing impurities to be detected and identified quicker, and making maximum use of instrumentation. The rapid switching between MS and MS/MS possible with the Xevo TQ MS allows the collection of qualitative data and quantitative data in the same analytical run. The instrument s high data-capture rate ensures that, even with the narrow peaks of 2 to 3 seconds produced by today s modern sub-2-µm particle LC systems, sufficient points can be collected across for accurate quantification. The use of ScanWave Technology ensures that even the lowest-level peaks are detected and MS/MS spectra acquired, ensuring comprehensive impurity detection. Waters, The Science of What s Possible, ACQUITY UPLC, Xevo, and UPLC are registered trademarks of Waters Corporation. ScanWave is a trademark of Waters Corporation. All other trademarks are the property of their respective owners. 28 Waters Corporation. Produced in the U.S.A. June 215 722831EN AG-PDF Waters Corporation 34 Maple Street Milford, MA 1757 U.S.A. T: 1 58 478 2 F: 1 58 872 199 www.waters.com