ROLE OF LOCKING COMPRESSION PLATES IN THE TREATMENT OF FRACTURES DISTAL END OF RADIUS

Similar documents
COMPARATIVE STUDY OF FUNCTIONAL OUTCOME OF EXTERNAL AND INTERNAL FIXATION IN TREATMENT OF COMMINUTED DISTAL RADIUS FRACTURES

ORIGINAL ARTICLE. possible. Accurate assessment of standard radiographs is essential for appropriate 3. management. And includes true posterior- 4

Comparison between Distractor Application on Both Radial & Ulnar Side and Radial Side Only for Fracture Distal Radius with Ulnar Styloid Fracture

Index. Note: Page numbers of article titles are in boldface type. Hand Clin 21 (2005)

journal ORIGINAL RESEARCH

Surgical Management of Distal end Radius Fractures by Various Methods: A Prospective Study

Factors Predicting Late Collapse of Distal Radius Fractures

Mark VanDer Kaag 1, Ajmal Ikram 2. Hand Unit, Tygerberg Hospital University of Stellenbosch

Clinical Study Rate of Improvement following Volar Plate Open Reduction and Internal Fixation of Distal Radius Fractures

Abd Ali Muhsin FICMS.

A prospective study of surgical management of distal radius fracture in adult with plate fixation

University of Groningen. Fracture of the distal radius Oskam, Jacob

Do Radiographic parameters predict Functional Outcome in Distal end Radius Fracture?

Hand and wrist emergencies

ORIGINAL ARTICLE TREATMENT DISTAL RADIUS FRACTURE WITH VOLAR BUTTRESS TECHNIQUE- A CLINICAL STUDY

A Patient s Guide to Adult Distal Radius (Wrist) Fractures

Acomparison of percutaneous and pin-and-plaster techniques in distal radius fracture

FOOSH It sounded like a fun thing at the time!

MINIMALLY INVASIVE PLATE OSTEOSYNTHESIS FOR DISTAL RADIUS FRACTURES: SURGICAL TECHNIQUE M. TOBE 1, K. MIZUTANI 1, Y. TSUBUKU 1, Y.

Angle Stable Distal Radial Plate System WINSTA-R

Common Limb Fractures. Mr Sheraz Malik MB BS MRCS Instructor Mr Paul Ofori-Atta Mb ChB FRCS President Motc Life UK April 2009

Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Evaluation of instability factors in distal radius fractures

Comparison of Below The Elbow Cast with Above The Elbow Cast in Treating Distal Third Forearm Fractures in Children

FOOSH It sounded like a fun thing at the time!

Breakage of a volar locking plate after delayed union of a distal radius fracture

Disclosures. Distal Radius Fractures 5/16/2017. Distal Radius Fractures: Complications & Limitations of the Volar Approach

Effects of Malunited Fractured Distal End of Radius on the Morphometric Parameters of Distal Radioulnar Joint in Old Age Group

Evaluation of volar locking plate fixation for management of intraarticular fractures of distal end of radius

Treatment of unstable distal radius fractures with the volar locking plate

Fractures of distal end of Radius: A study on fracture reduction and stable fixation

Comparative Study between Bridging External Fixation vs Volar Plating (Ellis-T Plate) for Comminuted Fracture of the Distal End Radius

The NBX Non-Bridging External Fixator A Non-Bridging External Fixator/Locking Plate capturing a series of.062mm K-wires and 3mm half-pins that are

Cover Page. The handle holds various files of this Leiden University dissertation.

Distal radius fractures raises considerable interest and

Clinical Study Dorsally Comminuted Fractures of the Distal End of the Radius: Osteosynthesis with Volar Fixed Angle Locking Plates

Type III Supracondylar Fractures of the Humerus in Children Straight-Arm Treatment

6/5/2018. DISCLOSURES Hassan R. Mir, MD, MBA, FACS. Evolution of Distal Radius Fracture Treatment [Chung Hand Clinics 2012]

Functional Outcome in Distal Radius Fractures Treated with Locking Compression Plate

University of Groningen. Fracture of the distal radius Oskam, Jacob

D-RAD SMART PACK Plating System. A series of case studies

Union Surgical. T-Pin. Fixation System for Distal Radius & Distal Ulna Fractures. Surgical Technique

JMSCR Vol 05 Issue 04 Page April 2017

8 Recovering From HAND FRACTURE SURGERY

Degrees Of Volar Angulation In Distal Radius Fracture Effects Distal Radioulnar Joint Stability:a Biomechanical Study

Management of intra-articular fractures of distal end radius in adults

CASE REPORT. Distal radius nonunion after volar locking plate fixation of a distal radius fracture: a case report

Surgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE

University of Groningen. Fracture of the distal radius Oskam, Jacob

A N D R E W I R W I N, F R C S E D ( O R T H ) C O N S U L T A N T O R T H O P A E D I C S U R G E O N W E S T H E R T S N H S T R U S T, U K

GALEAZZI FRACTURE. Galeazzi fracture-dislocations can be difficult to recognize and are often not initially appreciated.

JMSCR Vol 05 Issue 06 Page June 2017

Functional Outcome of Displaced Extraarticular Distal Radius Fracture by Kapandji Intrafocal Fixation

OUTCOME OF MANAGEMENT OF CLOSED PROXIMAL TIBIA FRACTURES IN TERTIARY HOSPITAL OF SURAT Karan Mehta 1, Prashanth G 2, Shiblee Siddiqui 3

VOLAR LOCKING COMPRESSION SYSTEM DISTAL AND RADIUS FRACTURE DISSERTATION

Rehabilitation after Total Elbow Arthroplasty

Primary internal fixation of fractures of both bones forearm by intramedullary nailing

IS CASTING AN ACCEPTABLE TREATMENT APPROACH FOR A DISTAL RADIUS FRACTURE THAT HAS UNDERGONE A SATISFACTORY CLOSED REDUCTION?

Recurrent subluxation or dislocation after surgical

UZZAMAN KS 1, AWAL KA 2, ALAM MK 3

The functional importance of malunion in distal radius fractures

Non-comminuted extra-articular fractures of distal radius treated with Percutaneous pinning

The Kienböck disease and scaphoid fractures. Mariusz Bonczar

Surgical management of fracture both bone forearm in adult using limited contact dynamic compression plate

Is Closed Manipulative Reduction and Percutaneous Kirschner Wiring of Supracondylar Humeral Fracture in Children as Day-Care Surgery a Safe Procedure?

Client centered approach to distal radius fracture management. Jared Rasmussen OTR

Distal radius fracture Golden standard and breaking news

An assessment of the three-point index in predicting the redisplacement of distal radial fractures in children

RECOVERY OF ISOMETRIC GRIP STRENGTH AFTER COLLES FRACTURE: A PROSPECTIVE TWO-YEAR STUDY

Additional Ulno-Radial Pinning for Prevention of Metaphyseal Collapse in Distal Radius Fracture: A Comparative Study

Orthopedics in Motion Tristan Hartzell, MD January 27, 2016

Fractures of the Radial and Ulnar Shafts In the Pediatric Patient

Scaphoid Fractures. Mohammed Alasmari. Orthopaedic Surgery Demonstrator Majmaah University

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

Percutaneous Scaphoid Fixation: A Volar Approach

Differentiated approach to repair of displaced distal radial metaepiphyseal fractures. R.P. Matveev, S.V. Bragina, A.M. Shneiveis

Interesting Case Series. Ulnolunate Impaction Syndrome

Distal Femur Fractures in The Elderly The Ideal Construct

University of Groningen. Fracture of the distal radius Oskam, Jacob

Complications of Distal Radius Fractures. How to Treat a Distal Radius Fx 11/13/2017. Michael S. Bednar, M.D. Loyola University Chicago

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

A Patient s Guide to Adult Forearm Fractures

Operative Indications

Top 10 Ortho Urgent Care Injuries. J.C. Clark, M.D. ORA Orthopedics

E ORIGINAL ARTICLE Low extra-articular (transcondylar) fractures of the distal humerus

A Patient s Guide to Adult Radial Head (Elbow) Fractures

Pediatric Fractures. Objectives. Epiphyseal Complex. Anatomy and Physiology. Ligaments. Bony matrix

Paediatric fractures in the Emergency Department. October 2012

JMSCR Volume 03 Issue 02 Page February 2015

Operative Treatment of Intra-articular Distal Radius Fractures Using the Small AO External Fixation Device

Conservative treatment of the distal radius fracture using thermoplastic splint: pilot study results

Dieter Marquardt Medizintechnik GmbH 10 Hand System

Upper Extremity Injury Management. Jonathan Pirie MD, Med, FRCPC, FAAP

Disclosures. Epidemiology 10/12/2015. Distal Radius Fractures: Staying Out of Trouble

Int J Clin Exp Med 2017;10(1): /ISSN: /IJCEM

Common Forearm Fractures in Adults

Redisplacement of Distal Radius Fracture after Initial Closed Reduction: Analysis of Prognostic Factors

Results of lateral pin fixation for the displaced supracondylar fracture of humerus in children

Volar fixed-angle plating for distal radius fractures 刀圭会協立病院 津村敬

Transcription:

Clinical Article Orthopaedics ROLE OF LOCKING COMPRESSION PLATES IN THE TREATMENT OF FRACTURES DISTAL END OF RADIUS Simranjit Singh 1, Rajan Sharma 2 1 - Senior Rent, Department of Orthopaedics, Government medical college Amritsar 2 - Associate professor, Department of Orthopedics, Sri guru ram Das institute of medical sciences and research Amritsar Corresponding Author: Dr Rajan sharma Dept. of orthopaedics SGRD Medical college, Amritsar Email: sharmarajan29@yahoo.com Article submitted on: 13 June 2017 Article Accepted on: 27 June 2017 Abstract: Background: Distal Radius fractures are the most common fractures of the human skeleton. The distal radius fracture is almost always within an inch of the wrist joint and fracture lines may extend into the wrist joint. Dorsal angulation in distal radial fractures is common and there is tendency towards secondary displacement after conservative management. Hence, open reduction of the articular surface,stable reduction, restoration of the radial length, volar angulation and radial inclination are the prerequisite for good clinical outcome. The volar locking compression plate can provide a safe and effective implant for treatment of dorsally displaced fractures of distal radius. The present study is proposed to evaluate the role of locking compression plate in treatment of fractures of distal end of radius. Material and Methods: Twenty five patients, coming to Sri Guru Ram Das Institute of Medical Sciences and Research (SGRDIMS&R), Sri Amritsar, having fractures of distal end of radius grouped as per A. O Classification were treated by locking compression plates. Results: Most of the functional and clinical outcomes showed continued improvement between the three and six months periods. The grip strength showed score of 53% at 3months and 69% at 6 months as compared to. Lateral pinch strength improved to 81% of the value for the normal by three months and 89% till 6 months. The Jebsen-Taylor test was exceptional and showed that the scores on the injured were close to 87% of those on the uninjured even at three months and continued to improve. Conclusion: locking compression plate system for the management of distal radius fractures showed significantly good results as shown by various functional outcome scores Key words: locking compression plate, distal radius 617

Introduction Distal Radius fractures are the most common fractures of the human skeleton. 1 The percentage of these fractures that are conred unstable and require surgical fixation has been reported to be as high as 40 to 49%. 2,3 The distal radius fracture is almost always within an inch of the wrist joint and fracture lines may extend into the wrist joint. Traditionally, the fracture has all along been treated conservatively by closed reduction and POP cast immobilization. The assumption has always been that a good result will be achieved whatever the appearance of reduction. 4 This may be acceptable in the elderly, osteoporotic patients with low demands on the wrist but there is a clear distinction between this group and those with higher demands, bearing in mind that these may not necessarily be younger. 5, 6 Dorsal angulation in distal radial fractures is common and there is tendency towards secondary displacement after conservative management. Distal radius is important in the kinematics of radiocarpal and radioulnar joints. Hence, open reduction of the articular surface,stable reduction, restoration of the radial length, volar angulation and radial inclination are the prerequisite for good clinical outcome. All this reduces the incidence of post-traumatic osteo-arthritis and allow early fun ction biologicalosteosynthesis a terminology introduced to indicate a new type of osteosynthesis leading to a sufficiently stable fixation of bone fragments allowing early mobilization, but without major disturbance of the vascularisation. 7 To help accomplish this new concept of biological plate fixation, new implant such as locking compression plate (LCP) was introduced. 8 Locking minimizes the compressive forces exerted by the plate on the bone. This method of screw plate fixation s that the plate does not need to touch the bone at all. Precise anatomical contouring of a plate is no longer necessary because the plate does not need to be pressed on the bone to achieve stability. This prevents the loss of primary reduction of fracture fragments caused by inadequate contouring of a plate al rehabilitation. 9 The volar locking compression plate can provide a safe and effective implant for treatment of dorsally displaced fractures of distal radius. 10 The present study is proposed to evaluate the role of locking compression plate in treatment of fractures of distal end of radius. Material And Methods Twenty five patients, coming to Sri Guru Ram Das Institute of Medical Sciences and Research (SGRDIMS&R), Sri Amritsar, having fractures of distal end of radius grouped as per A. O Classification were treated by locking compression plates. The study had prospective component only. While certain fracture pattern in this series were amenable to simple techniques such as percutaneous pinning, we used volar locking plate system in all cases to better understand the properties and outcomes of this new technology. Inclusion criteria All patients having fracture of distal end of radius were grouped as per A. O. classification. 11 Patients over 18yrs were included in this study. Type A: Extra-articular fractures involving neither the radiocarpal nor the radioulnar joint. Type B: Simple articular fracture affecting a portion of the articular surface, but with the continuity of the metaphysis and epiphysis intact. Type C: Complex articular fracture affecting the joint surface and the metaphyseal area. Exclusion criteria y Patients with concomitant upper extremity injuries (carpal bone, proximal forearm, elbow, or humeral fracture). y Patients with systemic, multiple-organ, or head injuries. y Patients who were managed conservatively for more than three weeks i.e. old fractures. y Patients with bilateral fractures. y Patients having fractures associated with neurovascular injuries, inflammatory arthritis, open fractures and malunited fractures. Operative protocol Informed consent was taken as per proforma. Type of anaesthesia to be used was decided by anesthesiologist. The patient placed in the supine position with affected limb positioned to expose the surgical site. Following standard operative procedures for painting and draping of the part, volar fixation of distal radius was performed through Henry s Approach with locking compression plate. 20 Proper placement of locking impression plate was guided by C-arm image intensifier. Removable volar plaster splint was given for six weeks. All patients was allowed to perform activities of daily living while wearing the splint. They were given once a week structured hand therapy programme that included 618

active and passive finger movements, hand and wrist edema control, and active wrist motion exercises. Strengthening exercises were initiated six weeks after the surgical procedure. Postoperatively, patients were assessed clinically and radiologically (at one week to ten days) and then at six, nine and twelve weeks as well as six months after surgery. The commonly used hand function test included test of hand strength (grip test), lateral pinch, test range of motion and Jebson Taylor Test. Range of motion of wrist was evaluated by recording flexion, extension and radioulnar deviation with standard goniometer. Grip strength test was measured with dynamometer and value was compared with extremity. Anteroposteior and lateral radiograph were taken at each visit. Fracture was conred united when osseous bridging across the fracture site was seen in radiographic views. Radial height, radial inclination, and volar tilt was measured in the immediate post-operative period (ten days after surgery) and at the time of the last follow up (six months after treatment). Activities of daily living and general postoperative quality of life were assessed with the Michigan hand outcomes questionnaire scores. 12,13,14,15 The collected data and results obtained was be subjected to standard statistical analysis. Observations And Results 25 procedure were performed, 18 in males and 7 in females with use of the volar locking plating system for fracture distal radius at SGRDIMS&R Amritsar from 2009 to 2011. Most of the patient were in the age group of 31-50years (48%) followed by group of 18-30 years (36%). Most common affected was right radius (52%). Fracture was of closed variety in all cases and in most case was the result of road traffic accident (RTA) followed by fall on outstretched hand. The patients were approached for participation in the study, usually at the time of the three and six months after operation. Functional outcomes (grip strength, lateral pinch strength and the Jebsen Taylor test), clinical outcomes (wrist flexion, wrist extension, ulnar deviation, radial deviation, supination, and pronation), and radiographic measurements of distal radial alignment (radial height, radial inclination, volar tilt and ulnar variance) are shown in tables 1,2,3. Most of the functional and clinical outcomes showed continued improvement between the three and six months periods. The grip strength showed score of 53% at 3months and 69% at 6 months as compared to. Lateral pinch strength improved to 81% of the value for the normal by three months and 89% till 6 months. The Jebsen-Taylor test was exceptional and showed that the scores on the injured were close to 87% of those on the uninjured even at three months and continued to improve. The six months period is noted in terms of grip strength, lateral pinch strength, Jebsen-taylor test, flexion, extension, ulnar deviation, Test Grip strength (kg) radial deviation, pronation, supination (p < 0.05) adjusting for the outcomes values for the three month period. When we assessed the Michigan hand outcomes questionnaire score over the two measurement times, significant improvements were detected between three and six months in all domains. In summary at six months, the injured still had worse outcomes than the uninjured, in all domains. At three months, however, patients had a activities of daily living score of 83%, indicating excellent ability to perform most daily tasks. Patients appeared to improve in terms of their functional recovery at six months. Complications Three patients had short term complications (defined as complications occurring during the first thirty days postoperatively): two patients had painful wrist which resolved after physiotherapy and one had a suture abscess, that resolved with suture removal. Sudeck sosteodystrophy and deformity present noted in one case which was operated late (after 2 weeks). Two plate related complications were present, in one a screw back tracked (pulled out) and in the other one screw violated the articular surface of distal radius. Table 1 Postoperative Hand Function On Injured Side 12 weeks (n=25) 6 months (n=25) 19.52 + 4.7 53.2 25.36+ 5.7 69.3 Lateral pinch (kg) 8.16+1.2 81.2 9.0+1.2 89.7 Jebsen-Taylor test (s) 34.88 + 2.5 86.8 32.8+2.3 91.46 619

Table 2 Postoperative Wrist Range Of Motion On Injured Side 12 weeks (n=25) 6 months (n=25) Test Flexion 53.4+ 6.5 73.1 56.8+7.3 77.6 Extension 55.8+ 5.5 79.3 59.6+6.2 84.6 Ulnar deviation 29.4+ 4.6 83.1 31.6+3.7 89.7 Radial deviation 17.2+ 4.8 83.4 18.2+4.05 89 Pronation 71.6+ 3.7 92.6 73.2+4.0 94.7 Supination 70.4+ 4.3 87.6 75+ 4.1 93.4 Radial height (mm) Immediate Postoperative Radial inclination (deg) Immediate Postoperative Volar tilt (deg) Immediate postoperative Ulnar variance Immediate postoperative Conclusion There has been a distinct shift in the surgical techniques for distal radial fracture fixation over a period of time. There are many possible influences, and the change is probably multifactorial. This dramatic shift may have been due to introduction and reported good results for volar locking plates for the treatment of distal radial fractures. Surgeons believe that open reduction with locked volar locking plates provides more stable fixation and allows earlier range of motion than percutaneous fixation does. 16 The literature, on outcome after distal radial fractures, is controversial with only few comparative studies. Wright Table 3 Radiographic Outcomes 10.96 + 0.84 10.84 + 0.85 0.12 + 0.33 21.52 + 1.4 20.52 + 1.6 0.1 + 1.35 2.84 + 1.1 3 + 0.95 0.16 + 0.37-0.84 + 0.88-0.8 + 0.89 0.04 + 0.01 et al. performed a case control study of 32 patients in which use of volar fixed angle plate was compared with external fixation; significantly better radiographic outcome were found in association with use of plates but no difference between the groups in terms of patient based outcome. Leung et al. reported that those who had been managed with plate fixation had a better cosmetic result and function at two years 17. We would like to emphasize the importance of careful case selection, surgery in fresh cases & early physiotherapy to give better functional outcome. In summary, continued use of the volar locking plating system is supported by the excellent outcomes data presented in the current study. References 1. Chung KC, Spilson SV. The frequency and epidemiology of hand and forearm fractures in the United States. J Hand Surg (Am). 2001; 26: 908-15. 2. Kakarlapudi TK, Santini A, Shahane SA, Douglas D. The cost of treatment of distal radius fractures. Injury, 2000; 31: 229-32. 3. MacDermid JC, Roth JH, Richards RS. Pain and disability reported in the year following a distal radius fracture: a cohort study, BMC MusculoskeletDisord. 2003; 4: 24. 4. Colles A. On the fracture of the carpal extremity of the radius. Edinb Med Surg J 1914; 10: 181. 620

ClinOrthop 2006; 445: 5-7. 5. Jaremko JL, Lambert RG, Rowe BR, Johnson JA, Majumdar SR. Do radiographic indices of distal radius fracture reduction predict outcomes in older adults receiving conservative treatment? Clin Radiol2007; 62: 65-72. 6. Young BT, Rayan GM. Outcome following non-operative treatment of displaced distal radius fractures in low demand patients older than 60 years. J Hand Surg (Am) 7. Broos PL, Serman A. From unstable internal fixation to biological osteosynthesis. A historical overview of operative treatment. ActaChir Belgium 2004; 104(4): 396-400. 8. Sommer C, Gautier E, Muller M, Helfet DL, Wagner M. First clinical results of the locking compression plate (LCP). Injury 2003; 34(2): 43-54 9. Wong KK, Chan KW, Kwok TK, Mak KH. Volar fixation of dorsally displaced distal radial fractures using LCP. J Orthop Surgery 2005:13(2):153-157 10. Schutz M, Kolbeck S, Spranger A, Arndt KM, Hass NP. Palmer plating with the locking compression plate for dorsally displaced fractures of the distal radius - First clinical experiences. ZentralblChir 2003; 128(12): 997-1002. 11. Rockwood CA Jr, David Green P. Fractures in adults. 5thedn. 2001;2:815-863.1 12. Jebsen RH, N. Taylor, RB Triechmann, MJ Trotter, LA Howard. An objective and standardized test of hand function. Arch Phip Med Rehdill 1969; 50: 311-319. 13. Baxter Petralia, PL Blackmore, PM McEntec. Physical capacity evaluation. In Hunter J., et al: Rehabilitation of Hand: Surgery and Therapy. St. Louis CV Mosby Co. 1990. 14. Fess EE. Documentation: Essential elements of on upper extremity assessment battery. In Hunter J et al: Rehabilitation of the Hand: surgery and Therapy. St. Louis CV Mosby Co. 1990. 15. Blair SJ, McCormick E, Bear- Lehman J, Fess EE, Roder E. Evaluation of impairement of upper extremity. Clinical Orthopedic Related Res Arch 1987; 221: 42-58. 16. KK Wong, KW Chan, TK Kwok, KH Mak. Volar fixation of dorsally displaced distal radial fractures using LCP. J Orthop Surgery 2005:13(2):153-157. 17. Leung F, Tu YK, Chew WYC, Chow SP. Comparison of External and Percutaneous Pin fixation with plate fixation for intra-articular distal radial fractures. Brit Journal of Bone and Joint Surgery 2008; 90: 16-22. 621