Eileen Crimmins 1 Jianwei Hu 2 Peifeng Hu 3 Wei Huang 2 Jungki Kim 1 Yuhui Shi 2 John Strauss 1 Lu Zhang 1 Xiaohui Zhao 2 Yaohui Zhao 2

Similar documents
IFLS C-Reactive Protein Data User Guide

Results from the Health and Retirement Study Biomarker Validation Project

Venous Blood Based Biomarker Data from the China Health and Retirement Longitudinal Study (CHARLS)

Biomarkers of inflammation for population research: Stability of C-reactive protein and alpha 1 -acid glycoprotein in dried blood spots.

S150 KEEP Analytical Methods. American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153

HbA1c (Human) ELISA Kit

Documentation of Blood-Based Biomarkers in the 2014 Health and Retirement Study

EliKine Free Thyroxine (ft4) ELISA Kit

Is Knowing Half the Battle? Behavioral Responses to Risk Information from the National Health Screening Program in Korea

Human LDL ELISA Kit. Innovative Research, Inc.

Insulin ELISA. For the quantitative determination of insulin in serum and plasma

Mouse HBsAg(Hepatitis B Virus Surface Antigen) ELISA Kit

LDL (Human) ELISA Kit

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet

Preliminary Dried Blood Spot Biomarker Data from the Study on global AGEing and adult health (SAGE)

Prothrombin (Human) ELISA Kit

Human HIV (1+2) antigen&antibody ELISA Kit

Nori Rabbit IL-2 ELISA Kit DataSheet

300 Biomed Environ Sci, 2018; 31(4):

Bovine Insulin ELISA

Serum levels of galectin-1, galectin-3, and galectin-9 are associated with large artery atherosclerotic

Porcine/Canine Insulin ELISA

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

Chapter 2. High-sensitivity C-reactive protein methods examined

Insulin ELISA. For the quantitative determination of insulin in serum and plasma.

human Total Cathepsin B Catalog Number: DY2176

Mouse C-peptide ELISA

Supplementary Online Content

Analytical Methods: the Kidney Early Evaluation Program (KEEP) The Kidney Early Evaluation program (KEEP) is a free, community based health

Cortisol (Sheep) ELISA Kit

Depok-Indonesia STEPS Survey 2003

Toxoplasma gondii IgM ELISA Kit

Assessing Agreement Between Methods Of Clinical Measurement

Mouse Ultrasensitive Insulin ELISA

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures.

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

THE HEALTH AND RETIREMENT STUDY: INVESTIGATION OF DRIED BLOOD SPOT HBA1C MEASUREMENT ERRORS USING TIPTEMP STAMPS

Rubella virus IgG ELISA Kit

Insulin (Porcine/Canine) ELISA

Know Your Number Aggregate Report Single Analysis Compared to National Averages

Mouse C-peptide ELISA

Toxoplasma gondii IgM ELISA Kit

Figure S1. Comparison of fasting plasma lipoprotein levels between males (n=108) and females (n=130). Box plots represent the quartiles distribution

Urea Nitrogen (BUN) detection Kit

The Framingham Coronary Heart Disease Risk Score

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum

Cotinine (Mouse/Rat) ELISA Kit

Thyroid Stimulating Hormone (TSH) ELISA Catalog No. GWB , legacy id (96 Tests)

Rat Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA kit

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

IgG Antibodies To Toxoplasma Gondii ELISA Kit Protocol

Accuracy and Precision. Intra- and inter-assay accuracy and precision for both rifapentine

Online Supplementary Appendix

DetectX. Urinary Creatinine Detection Kit. Catalog Number K002-H1. Sample Types Validated: Human, Monkey, Dog, Rat and Mouse Urine

DKK-1 ELISA, Cat.No. BI For the quantitative determination of DKK-1 in human serum

Mouse C-Peptide ELISA Kit

Human Apolipoprotein A1 EIA Kit

ORIGINAL INVESTIGATION. C-Reactive Protein Concentration and Incident Hypertension in Young Adults

Association of serum adipose triglyceride lipase levels with obesity and diabetes

Apolipoprotein A1 (Apo A1) ELISA

Zhengtao Liu 1,2,3*, Shuping Que 4*, Lin Zhou 1,2,3 Author affiliation:

YK052 Mouse Leptin ELISA

Estrone EIA kit. For the quantitative determination of estrone in dried fecal extracts, urine and tissue culture media. Cat. No.

CoQ10(Coenzyme Q10) ELISA Kit

Procine sphingomyelin ELISA Kit

Instructions for use. TSH rat ELISA. Please use only the valid version of the Instructions for Use provided with the kit AR E-8600

A n aly tical m e t h o d s

Determination of hemoglobin is one of the most commonly

Triiodothyronine (T3) ELISA

Changes and clinical significance of serum vaspin levels in patients with type 2 diabetes

Instructions for use. TSH rat ELISA. Please use only the valid version of the Instructions for Use provided with the kit AR E-8600

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Supplementary Appendix

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

Human Creatinine Urinary Detection Kit

What a drop can(t) do: Methodological challenges associated with incorporating dried blood spot biomarkers into population-based research

pulmonary artery vasoreactivity in patients with idiopathic pulmonary arterial hypertension

IAPT: Regression. Regression analyses

Human Alpha 1 microglobulin ELISA Kit

Rat Hemoglobin A1c (HbA1c) Kit Instructions

Rat Proinsulin ELISA

Correlation and regression

RELATIONSHIP OF CLINICAL FACTORS WITH ADIPONECTIN AND LEPTIN IN CHILDREN WITH NEWLY DIAGNOSED TYPE 1 DIABETES. Yuan Gu

HPLC-based measurement of glycated hemoglobin using dried blood spots collected under adverse field conditions

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

Canine Thyroid Stimulating Hormone, TSH ELISA Kit

Human TSH ELISA Kit. User Manual

Human Troponin I ELISA Kit

Central pressures and prediction of cardiovascular events in erectile dysfunction patients

Add Health Biology & Birth Records Projects

colorimetric sandwich ELISA kit datasheet

Does Consumer Behaviour on Meat Consumption Increase Obesity? - Empirical Evidence from European Countries

Clincial Biostatistics. Regression

PKU (Phenylketonuria) Serum HPLC Analysis Kit User Manual

Population: All participants Number of Obs: Variable # Sas Name: Sas Label: Categories: Variable # Sas Name: F80VNUM. Sas Label: Categories:

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

LEPTIN AS A NOVEL PREDICTOR OF DEPRESSION IN PATIENTS WITH THE METABOLIC SYNDROME

colorimetric sandwich ELISA kit datasheet

Transcription:

CHARLS Pilot: Blood-based Biomarker Documentation Eileen Crimmins 1 Jianwei Hu 2 Peifeng Hu 3 Wei Huang 2 Jungki Kim 1 Yuhui Shi 2 John Strauss 1 Lu Zhang 1 Xiaohui Zhao 2 Yaohui Zhao 2 June 2011 1 University of Southern California 2 Peking University 3 University of California, Los Angeles

1 Biomarker Data in CHARLS Data collection in CHARLS was done using a medically trained staff. Each data collection team had two recent graduates of a nursing school who collected the biological measurements. These nurses were trained in collection techniques by staff of the Public Health program at Peking University. They also watched the videos of biomarker and performance data collection provided by HRS and used for training of HRS interviewers. Nurses collected information on blood pressure, pulse, measured height, weight, waist circumference, lower leg length, and arm length. Performance tests were given to provide information on lung peak flow rate, grip strength and a timed five repetitions of sit to stand. Cardiochek-PA meters were used in the field to provide information on total and HDL cholesterol. Like HRS, CHARLS used dried blood spot (DBS) technology to provide blood samples for biochemical assay. Two dried blood spots were collected from each respondent onto Whatman 903 Protein Saver Cards. After the spots were collected, the cards were dried on a special, covered rack for at least 4 hours, then inserted into a ziplock bag along with a desiccant and then mailed from the field to our headquarters at Peking University. There they were stored in deep freezers at -20C at the Peking University laboratory. About 25% of sample individuals did not participate in the biomarker data collection. The number of people with data differs for each marker. Analysis of who did not participate and associated sample weight corrections are described in the CHARLS User s Guide. Cardiochek Meter Readings of Cholesterol Cholesterol was measured at the time of interview without any request to fast prior to blood drop collection. The meter range for total cholesterol is 100 mg/dl as the bottom value and 400 mg/dl for the top value. For HDL, the meter range is from 15 mg/dl to 100 mg/dl. One value of total cholesterol was recorded as 40 mg/dl in the field and has been coded as missing. Method of DBS-based C-reactive protein (CRP) CRP concentrations in DBS specimens were measured using validated ELISA methods (McDade, 2004). The detection limit of this CRP assay is reported to be 0.028 mg/l. (There is an upper limit on the initial reading of the concentration off the OD curve such that the maximum is 10; on reanalysis with a diluted sample result could be multiplied by 2, so the upper limit would be 20). The intra-assay coefficient of variation (CV) is 5.8% and the inter-assay CV is 8.2% (McDade, 2004). In a number of cases where the coefficient of variation was above 10% a retest was done. In some cases these retests resulted in an indication of the level of concentration as below minimal value as redoing the assay often meant using a substandard blood spot. For these cases (N=40) we have entered the original CRP concentration value in a revised CRP variable. We have indicated these cases with a flag variable. A large number of cases are reported to be below minimum value (681). Initially cases without a detectable limit were retested but because the concentration was almost always still below the minimum level, this practice was stopped. The number of cases with undetectable levels of CRP was 28% in Gansu province and 44% in Zhejiang. Field reports indicate that spot quality may have been worse in Zhejiang province

2 because of the quality of the medical teams doing the collection. A test of whether the lab results changed over time indicated that this was not the case in Gansu province where we had variation in timing of assays (i.e. they were done from January to May in 2010 and then again from October through December in the same year). Assays were done in two time periods because of the need to acquire new reagent supplies. The first resupply was with defective material and several months passed before assays were started again with a new reagent supply. Assays on samples from Gansu were done in both periods but virtually all of the Zhejiang samples were measured in the later period. The CRP values produced with the DBS assays are quite different from those based on whole blood values; values are on a much lower scale. For this reason we do not recommend the use of conventional cutoffs for high CRP based on whole-blood values. Method of Hemoglobin (Hb) assays Hemoglobin (Hb) level was measured using an ELISA protocol developed by Dr. Thomas McDade at Northwestern University, based on the method by O Broin and Gunter (O Broin, 1999). Across the range of the assay, the between-assay CVs for low, normal, and high control concentrations are 7.7%, 4.8%, and 6.2%, respectively. Quality Control for DBS bioassays CRP assay Optic densities (OD) of all samples were measured using an automated ELISA reader at wave length of 490 nm. Every plate had 8 standard samples that were used to generate a standard curve to relate OD and standard CPR concentration. A standard curve for each microplate was generated by computer software program, based on 4-parameter method. On each plate there were also 3 control samples to ensure the stability of standard curves. All CHARLS samples, and CRP standards and controls were measured in duplicate. Individual samples with duplicate CV above 10% were rerun, unless both concentrations were less than 0.2 mg/l. When interassay variability exceeded 10%, which was based on control values, the entire microwell plate would be rerun. Hb assay OD of all samples were measured using an automated ELISA reader at wave length of 540 nm. All samples, standards, and controls were measured in duplicate. Individual samples with duplicate CV above 10% were rerun. Control samples (normal, low, and high levels) were used to ensure the stability of standard curves. When inter-assay variability exceeded 10%, the entire microwell plate would be rerun. In addition, all samples with initial hemoglobin concentrations below 8 g/dl or above 18 g/dl were also measured again to ensure the results were accurate. When low values were retested, all were fairly close to the original value except two where the difference was approximately 1 g/dl. In the cases where they were similar the retest value was used. In the two cases where there was some difference, the average was used. There was one high case which was retested twice, all readings were high (from 18 to 23 g/dl) and the average value of 21.6 g/dl was assigned.

3 CHARLS Validation Study To examine the validity of collection and lab procedures in China we performed validation tests on DBS samples from community volunteers (non-charls participants) collected at the UCLA Alhambra Research Center during the June through December 2010 period. Participants in the validation tests were adults 50 years of age or older. Participants provided signed consent to be part of the study and were paid $25 to participate. Participants provided a venous whole blood sample, a number of DBS, cholesterol measurements from a point of service Cardiochek PA meter, and hemoglobin readings from a point of service HemoCue meter (range 0-25.6 g/dl). The Cardiochek meter was the same model used in the field in China and the HemoCue Hb 301 meter was the same as that used in the field in Indonesia. Validation involved a calibration of DBS results collected in Los Angeles and then assayed in the lab in China against whole blood values of CRP assayed at the University of Vermont by Russell Tracy and DBS results assayed in the lab of Thomas McDade at Northwestern University. McDade provided the training in DBS assays for the CHARLS lab staff. The whole blood was separated and frozen in the UCLA lab and frozen serum was sent to the University of Vermont. The DBS collected in Los Angeles were frozen after drying and sent frozen to China and Northwestern for analysis. The validation study allows comparison of values of hemoglobin assayed from DBS in China with those collected using the Hemocue meter. Cholesterol values from Cardiochek meters collected in Los Angeles were compared to values collected from whole blood in Los Angeles. Below we provide information on paired assays for CRP, hemoglobin, total cholesterol and HDL cholesterol. For each pair of matched assays we provide descriptive statistics; a simple regression linking the two values and providing an R 2 ; a scatterplot of the two values plotted against each other; and a Bland-Altman plot which indicates the size of the differences against the average values of the two assays.

4 Descriptive Information on Assays for Matched Samples Cholesterol Total Cholesterol (TC) was assessed using the Cardiocheck PA meter and blood serum in a sample of 88 persons. The average for the sample using the Cardiocheck meter (172 mg/dl) is about 20 points less than that based on the venous blood (192 mg/dl) (Table 1). The differences between two assays are fairly similar across the range as indicated in the scatter plots and the Bland-Altman plot (Figure 1). The R 2 between the 2 assays is.65. Table 1. Total Cholesterol Cardiocheck Cardiocheck Meter vs. Vermont Blood Serum Total Cholesterol Average Range #Cases Median Interquartile Range Cardiocheck 172.49 (32.72) 104.0-263.0 88 174.0 145.5-192.0 Vermont 191.73 (34.17) 122.0-285.0 191.0 167.5-215.0 Equation Vermont = 46.43 + 0.84 Cardiocheck (11.68) (0.07) R 2.65

5 Figure 1: Validation Study of Cardiochek Meter: Comparison of Total Cholesterol Measured Using Cardiochek Meter and from Blood Serum at the University of Vermont Bland Altman Bland Altman Difference 30 20 10 0-10 -20-30 -40-50 -60-70 90 120 150 180 210 240 270 300 Average Difference (Cardiocheck Total Cholesterol - UVSerum Cholesterol)

6 HDL Cholesterol (HDL) was assessed for 89 people using the two approaches. Again, the Cardiocheck meter provides somewhat lower values than the venous blood sample: means of 45 mg/dl and 53 mg/dl respectively (Table 2). The R 2 between these two values (.76) is higher than that for total cholesterol. The scatterplots and the Bland-Altman plots indicate the measurement differences are relatively similar across the assay range (Figure 2). Table 2. HDL Cardiocheck Meter vs. Vermont Blood Serum HDL Average Range #Cases Median Interquartile Range Cardiocheck 44.79 (17.97) 17.0-100.0 89 40.00 31.00-58.00 Vermont 53.06 (14.62) 24.00-98.00 51.00 41.00-63.00 Equation Vermont = 21.28 + 0.71 Cardiocheck (2.06) (0.04) R 2.76 Figure 2: Validation Study of Cardiochek Meter: Comparison of HDL Cholesterol Measured Using Cardiochek Meter and from Blood Serum at the University of Vermont

7 Bland Altman Cardiocheck: HDL 120 110 100 90 80 70 60 50 40 30 20 10 0 0 10 20 30 40 50 60 70 80 90 100 110 120 Bland Altman Vermont: HDL (mg/dl)-serum Bland Altman Difference 40 30 20 10 0-10 -20-30 -40 0 10 20 30 40 50 60 70 80 90 100 110 120 Average Difference (Cardiocheck HDL - UVSerum HDL)

8 Hemoglobin Hemoglobin levels were compared for 33 cases where hemoglobin was assessed with a DBS assay done in China and a reading on a Hemocue meter in Los Angeles. The average Hemocue meter value (14.02 g/dl) exceeded that from the DBS (13.30 g/dl) (Table 3). The higher level of the Hemocue meter reading is seen throughout the range (Figure 3). The R 2 from the equation linking the China DBS and the HemoCue reading is.36. Because the cutoff values for low hemoglobin (< 12 for women and < 13 for men) are close to the median and the mean in both distributions, cutoff values should be used with caution. Table 3. Hemoglobin CHARLS DBS vs. Hemocue Meter Hemoglobin Average Range #Cases Median Interquartile Range China 13.30 (1.35) 11.00-16.46 33 13.13 12.26-13.95 Hemocue 14.02 (1.47) 11.20-16.90 13.80 13.00-15.30 Equation Hemocue= 5.35 + 0.65 CHARLS (2.08) (0.16) R 2.36

9 Figure 3: Validation Study of CHARLS Hemoglobin DBS Assay Against HemoCue Meter Values. CRP: McDade and China hbconcentration 20 19 18 17 16 15 14 13 12 11 10 9 8 7 7 8 9 10 11 12 13 14 15 16 17 18 19 20 Hemocue: Hemoglobin Bland Altman CRP: McDade and China Difference 3 2 1 0-1 -2-3 -4 7 8 9 10 11 12 13 14 15 16 17 18 19 20 Average Difference (CHARLS HB - Hemocue HB)

10 CRP Comparison of CRP DBS Values from CHARLS and McDade Assays As indicated above, there are two CRP comparisons in the validation sample, first we consider the differences between assays done for CHARLS in China and assays done in McDade s lab for 34 cases (Table 1). Because the distribution of CRP is not normal but highly skewed toward low values, we also examine logged values of CRP. The average value of the DBS assayed for CHARLS (1.12 mg/l) was lower than that reported from the McDade assay (1.42 mg/l), largely because the CHARLS assay did not have any values above 4.16 mg/l. McDade had only two values above 4.16 mg/l but these weight the mean. The median value was actually higher in the CHARLS data (.71 mg/l versus.59 mg/l) and the interquartile range was slightly greater. The R 2 of the equation linking the two is.43. The differences between the two assays are greater at higher values. The plots of the logged values of CRP are shown in Figure 5. Table 4. CRP CHARLS CHARLS DBS Assay vs. McDade DBS Assay CRP Average Range #Cases Median Interquartile Range China 1.12 (1.14) 0.12-4.16 34 0.71 0.34-1.58 McDade 1.42(2.38) 0.12-12.16 0.59 0.26-1.24 Equation McDade = -.10945 + 1.36577 CHARLS CRP (0.44) (0.28) R 2.43

11 Figure 4A: Comparison of DBS CRP Assays from McDade and CHARLS China: CRP 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 Bland-Altman Analysis 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Bland Altman McDade: CRP CRP: McDade and China Difference 7 5 3 1-1 -3-5 -7-9 -11 0 2 4 6 8 10 12 14 Average Difference (CHARLS CRP - McDade CRP)

12 Figure 4B: Comparison of Logged CRP from DBS Assays done by McDade and for CHARLS CRP: McDade and CHina Logged China CRP 5 4 3 2 1 0-1 -2-3 -4-5 -5-4 -3-2 -1 0 1 2 3 4 5 Bland Altman Logged McDade CRP CRP: McDade and China Difference 4 3 2 1 0-1 -2-3 -4-5 -6-5 -4-3 -2-1 0 1 2 3 4 5 Average Difference(logged CHARLS CRP-logged McDade CRP))

13 CHARLS DBS Assay and Vermont Blood Serum Assay The average CRP level from the venous blood is considerably higher than that from the DBS sample (2.39 mg/l versus 1.30 mg/l). The upper bound of the range is considerably higher (7.00 mg/l 20.10 mg/l), the middle and lower ranges much less so (Figure 6A), as a consequence the difference between the medians is not as great as for the means. In logs, however, the differences are much more stable across the range of CRP values (Figure 6B). The R 2 from the equation linking the two measures of CRP is.30, somewhat lower than the value for the McDade value. Table 5. CRP CHARLS CHARLS DBS Assay vs. Vermont Serum Assay CRP Average Range #Cases Median Interquartile Range CHARLS 1.30 (1.44) 0.12-7.00 42 0.79 0.34-1.74 Vermont 2.39 (3.74) 0.33-20.10 1.05 0.51-2.93 Equation Vermont Serum = 0.557 + 1.411 CHARLS CRP (0.66) (0.34) R 2.30

14 Figure 5A: Comparison of CHARLS DBS Assay with Venous Serum Blood in Vermont Bland-Altman Analysis China: CRP 24 21 18 15 12 9 6 3 0 0 3 6 9 12 15 18 21 24 Vermont: CRP - Serum (ug/ml) Bland Altman CRP: McDade and China Difference 7 4 1-2 -5-8 -11-14 -17-20 0 3 6 9 12 15 18 21 24 Average Difference (CHARLS CRP - UVSerum CRP)

15 Figure 5B: Comparison of Logged Values for CHARLS DBS Assay and Venous Serum Blood at the University of Vermont Logged China CRP CRP: McDade and CHina 4 3 2 1 0-1 -2-3 -4-4 -3-2 -1 0 1 2 3 4 Logged UV Serum CRP Bland Altman CRP: McDade and China Difference 8 6 4 2 0-2 -4-6 -8-10 -4-3 -2-1 0 1 2 3 4 Average Difference (Logged CHARLS CRP - Logged UVSerum CRP)

16 CHARLS data description Next we describe the CHARLS biomarker data. We start with a table of descriptive statistics, also showing population frequencies below and above standard cutoffs. We then plot empirical densities together with histograms. Systolic blood pressure has a mean of 135, with the median just below that. Some 36% of the CHARLS pilot population has systolic above the cutoff of 140. The diastolic mean is 77 (median almost the same) and only 15% has a reading over the cutoff of 90. Mean BMI is 23, but there are 33% of women and 19% of men with BMI 25 and over, indicating overweight. On the other hand only a trivial proportion has BMI 30 and over, so obesity is low. Underweight (under 18.5) is also low, 6-7%. Mean hemoglobin measure is 12.6, and the median is just under. Using the standard cutoffs of 12.0 for women and 13.0 for men, 51% of the sample have low hemoglobin levels. Note, however, that because a large part of the density is near the cutoffs, the fraction below is quite sensitive to where one sets the cutoff. Mean total cholesterol is 215 and for HDL it is 50. Some 33% of the sample has total cholesterol above the standard cutoff of 240, while 31% has HDL below the cutoff of 40. Finally, the mean CRP is 0.87, and the median 0.52. As seen above, in the validation analysis, the DBS measurements for CRP are lower than the venous blood measures. As a result, it is not a good idea to use the standard cutoff of above 3.0, as indicating high risk levels of CRP. However, the distribution of CRP, and log CRP look fairly standard, as indicated by the density plots below. We recommend using the data as a continuous variable. The density plots show considerable variation in the measures. Most of the distributions look reasonable, with very few outliers. The log of CRP looks far more symmetric than levels of CRP. For total and HDL cholesterol one can see the measured limits of the CariochekPA meter.

17 Table 6. Descriptive Statistics of Biomarkers in CHARLS Systolic Blood Pressure (mmhg) Diastolic Blood Pressure (mmhg) N Mean (SD) Median Interquartile Range 1967 134.83 (22.40) Ranges 131.33 29.00 82.66-231.33 1967 77.25 (11.96) 76.00 16.00 45.33-129.00 BMI (kg/m 2 ) 1946 22.99 (3.53) 22.70 4.60 13.93-62.50 CRP (mg/l) 1151 0.87 (1.25) 0.52 0.70 0.0007-13.97 Hemoglobin (g/dl) 1829 12.62 (1.74) 12.43 2.07 5.60-21.62 Total cholesterol (mg/dl) 1850 215.05 (66.80) 208.00 95.00 100.00-400.00 HDL cholesterol (mg/dl) 1842 50.22(18.70) 48.00 25.00 15.00-100.00 Table 7. %High Risk Biomarkers in CHARLS Cutoff Percent Systolic Blood Pressure (mmhg) >=140 mmhg 35.79% Diastolic Blood Pressure (mmhg) >=90 mmhg 14.18% Underweight (kg/m 2 ) <=18.5 kg/m 2 M: 6.39% F: 7.05% Overweight (kg/m 2 ) >=25 kg/m 2 M: 19.07% F: 32.77% Obese (kg/m 2 ) >=30 kg/m 2 M: 1.52% Hemoglobin (g/dl) <13 g/dl (male) <12 g/dl (female) F: 4.87% 50.55% Total cholesterol (mg/dl) >=240 mg/dl 33.14% HDL cholesterol (mg/dl) <40 mg/dl 30.67%

18

19

20

21

22

23 Males Females

24

25

26 Analysis of undetectable values of CRP A large fraction of CRP cases had undetectable levels, 36% overall, higher in Zhejiang province. HRS as a comparison had about 50% undetectable. It is of interest to explore the correlates of respondents who had undectable values. Table 9 shows logit regression results for the Gansu sample with age, gender and whether the assay was done after September 2010. The date is of interest because the reaction antibody for the CRP assay had to be reordered in May 2010 and a good batch did not arrive until October. Because different production runs of the reagent may have small random differences it is of interest to see if the probability of an undetectable level is correlated with the batch. However since all of the Zhejiang samples were assayed after October 2010 (see Table 8), it only makes sense to explore this with the Gansu samples, for which 22% were assayed after October and the rest before May. As one can see in the logit regression the time dummy is not significant at standard levels, indicating that this potential concern is not an issue. Also a regression of the CRP levels on age, gender and the time dummy shows no impact of the time dummy. Hence the difference between undetectable levels in Gansu and Zhejiang is due to other factors. One may be smaller blood spots for Zhejiang, which apparently was true, but there may be other reasons as well.

27 Table 8. CRP levels by province Total % freq Gansu % freq Zhejiang % freq Undetectable CRP N= 1873 N= 870 N= 1003 Detectable 63.64% (n=1192) 72.18% (n=628) 56.23% (n=564) Undetectable 36.36% (n=681) 27.82% (n=242) 43.77% (n=439) Time of Detection N= 1831 N= 857 N= 974 Before Sept 2010 36.21% (n=663) 77.25% (n=662) 0.10% (n=1) After Sept 2010 63.79% (n=1168) 22.75% (n=195) 99.90% (n=973) Table 9. Logistic regression on undetectable CRP (Gansu only) Estimate Standard Error Intercept 1.54 0.51 Age -0.05 0.01 Male 0.37 0.16 After Sept 2010 0.11 0.19 Table 10. Regression on CRP (N=1083) Parameter Estimate Standard Error Intercept -0.01 0.24 Age 0.02 0.00 Male -0.00 0.08 After Sept 2010-0.08 0.08 R 2 = 0.02

28 REFERENCES: Bland JM, Altman DG. (1986). Statistical methods for assessing agreement between two methods of clinical measurement. Lancet, i, 307-310. Bland JM, Altman DG. (1992) This week's citation classic: Comparing methods of clinical measurement. Current Contents, CM20(40) Oct 5, 8. McDade TW, Burhop J, Dohnal J. High-sensitivity enzyme immunoassay for C-reactive protein in dried blood spots. Clin Chem. 2004;50(3):652-654. O'Broin SD, Gunter EW. Screening of folate status with use of dried blood spots on filter paper. Am J Clin Nutr. 1999 Sep;70(3):359-367.