The diagnostic and treatment challenge of type IIIb endoleaks

Similar documents
Type IIIb Endoleak Is an Important Cause of Failure Following Endovascular Aneurysm Repair

Outcomes of endovascular repair of isolated iliac artery aneurysms. A. Stella

Disclosures. EVAR follow-up: actual recommendation. EVAR follow-up: critical issues

Management of Endoleaks

Report of Regional Single Center Experience of Endovascular Abdominal Aortic Aneurysm Repair at King Hussein Medical Center

Endovascular Repair o Abdominal. Aortic Aneurysms. Cesar E. Mendoza, M.D. Jackson Memorial Hospital Miami, Florida

Increased Flexibility of AneuRx Stent-Graft Reduces Need for Secondary Intervention Following Endovascular Aneurysm Repair

NEW INNOVATIONS IN ENDOLEAK MANAGEMENT

From 1996 to 1999, a total of 1,193 patients with

MODERN METHODS FOR TREATING ABDOMINAL ANEURYSMS AND THORACIC AORTIC DISEASE

EVAR replaced standard repair in most cases. Why?

Experience of endovascular procedures on abdominal and thoracic aorta in CA region

Durable outcomes. Proven performance.

Case Report Early and Late Endograft Limb Proximal Migration with Resulting Type 1b Endoleak following an EVAR for Ruptured AAA

An Overview of Post-EVAR Endoleaks: Imaging Findings and Management. Ravi Shergill BSc Sean A. Kennedy MD Mark O. Baerlocher MD FRCPC

History of the Powerlink System Design and Clinical Results. Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ

SANWICH TECHNIQUE TO REDUCE COMPLICATIONS WHEN TREATING BILATERAL INTERNAL ILIAC ARTERY

Bilateral use of the Gore IBE device for bilateral CIA aneurysms and a first interim analysis of the prospective Iceberg registry

Internal iliac artery aneurysms: When to intervene and outcomes of EVAR

Management of Endoleaks. Michael Meuse, M.D Vascular and Interventional Radiology 12/14/09

GORE EXCLUDER AAA Endoprosthesis demonstrates long-term durability. Michel Reijnen Rijnstate Hospital Arnhem, The Netherlands

Talent Abdominal Stent Graft

Nellix Endovascular System: Clinical Outcomes and Device Overview

Anatomy-Driven Endograft Selection for Abdominal Aortic Aneurysm Repair S. Jay Mathews, MD, MS, FACC

Abdominal and thoracic aneurysm repair

Current Status of EVAR for Infrarenal AAA. 31 st Annual Florida Vascular Society. PENN Surgery

Endoleak Visualized With Carbon Dioxide Angiography During Endovascular Aneurysm Repair Using the Endurant Stent-Graft

Eccentric stent graft compression: An indicator of insecure proximal fixation of aortic stent graft

Role of Gender in TEVAR and EVAR results from the GREAT registry

Analysis of Type IIIb Endoleaks Encountered with Endologix Endografts

Chungbuk Regional Cardiovascular Center, Division of Cardiology, Departments of Internal Medicine, Chungbuk National University Hospital Sangmin Kim

Accessi Iliaci Ostili

THE ENDURANT STENT GRAFT IN HOSTILE ANEURYSM NECK ANATOMY

The Distal Seal Zone in AAA Repair A facet of EVAR that is not to be overlooked.

Eight Year Experience with Type I Endoleaks at a Tertiary Care Center

Considerations for a Durable Repair

ENCORE, a Study to Investigate the Durability of Polymer EVAR with Ovation A Contemporary Review of 1296 Patients

LOWERING THE PROFILE RAISING THE BAR

Intravascular Ultrasound in the Treatment of Complex Aortic Pathologies. Naixin Kang, M.D. Vascular Surgery Fellow April 26 th, 2018

Degeneration of the Neck Post Implementation - a New Era of AAA Stent

Retrograde Embolization of a Symptomatic Hypogastric Artery Aneurysm

Why Nellix? Treating Concomitant Common Iliac Aneurysms

Robert F. Cuff, MD FACS SHMG Vascular Surgery

Stanford Division of Vascular Surgery

Bell-bottoms, trouser grafts and crossovers: Maintaining pelvic perfusion with endovascular repair of aorto bi-iliac aneurysms

Management of Endoleaks

Tips for Delayed Open Conversion in Patients with a Type III Endoleak after Endovascular Aortic Aneurysm Repair

symptomatic aneurysms or aneurysms that grow >1cm/yr

Ultrasound Evaluation after EVAR: (Trying to) Let the CAT Scan Out of the Bag

An endoleak is radiographic or ultrasonic evidence

Faculty Disclosure. Glue, Particulates, Thrombin, Coils and the Kitchen Sink for Type II Endoleak Management. Background.

Modified candy-plug technique for chronic type B aortic dissection with aneurysmal dilatation: a case report

Case Report Late Type 3b Endoleak with an Endurant Endograft

A New EVAR Device for Infrarenal AAAs

Aortic Neck Issues Associated Clinical Sequelae/Implications for Graft Choice

Endoleaks after F-BEVAR How to Assess & Treat? Gustavo S. Oderich, MD Mayo Clinic Rochester, MN

RadRx Your Prescription for Accurate Coding & Reimbursement Copyright All Rights Reserved.

Performance of the conformable GORE TAG device in Type B aortic dissection from the GORE GREAT real world registry

Type IIIb endoleak and relining: a mathematical model on distraction forces

The Management and Treatment of Ruptured Abdominal Aortic Aneurysm (RAAA)

FEVAR FIFTEEN YEARS OF EFFICIENCY E.DUCASSE MD PHD FEBVS CHU DE BORDEAUX

From the Western Vascular Society

Current Status of Abdominal Aortic Stent Grafts. John R. Laird Professor of Medicine Director of the Vascular Center UC Davis Medical Center

Important Update to Field Safety Notice Nellix EndoVascular Aneurysm Sealing System Updated Instructions for Use (IFU)

Case Report A Case of Successful Coil Embolization for a Late-Onset Type Ia Endoleak after Endovascular Aneurysm Repair with the Chimney Technique

Percutaneous Approaches to Aortic Disease in 2018

Type II Endoleak Embolization Choice of Materials: EVOH, Glue, Thrombin & Coils. Michael S. Rosenberg, MD Assistant Professor of Radiology

Type IIIb endoleak and relining : a mathematical model of distraction forces

Type III B endoleak leading to aortic rupture after endovascular repair: analysis of errors in follow up and treatment

When and where EVAR patients should be discharged?

TriVascular Ovation Prime Abdominal Stent Graft System

Taming The Aorta. David Minion, MD Program Director, Vascular Surgery University of Kentucky Medical Center Lexington, Kentucky, USA

My personal experience with INCRAFT in standard and challenging cases

Clinical trial and real-world outcomes of an endovascular iliac aneurysm repair with the GORE Iliac Branch Endoprosthesis (IBE)

Treatment options for endoleaks: stents, embolizations and conversions

How to Categorize the Infrarenal Neck Properly? I Van Herzeele Dept. Thoracic and Vascular Surgery, Ghent University, Belgium

Type-II Endoleaks Following Endovascular AAA Repair: Preoperative Predictors and Long-term Effects

Volume 3 (2017) Issue 1 ISSN Abstract. Keywords fenestrated after open repair; Pararenal abdominal aortic aneurysm.

Improving Endograft Durability with EndoAnchors

BC Vascular Day. Contents. November 3, Abdominal Aortic Aneurysm 2 3. Peripheral Arterial Disease 4 6. Deep Venous Thrombosis 7 8

Technique and Tips for Complicated AAA Cases with Stent Graft

Treatment options of late failures of EVAS. Michel Reijnen Rijnstate Arnhem The Netherlands

Access More Patients. Customize Each Seal.

Abdominal Aortic Aneurysm (AAA)

The Ventana Off-the-Shelf Graft for Pararenal AAA. Andrew Holden Associate Professor of Radiology Auckland Hospital

Conversion to Open Repair After Endografting for Abdominal Aortic Aneurysm: Causes, Incidence and Results

The Auckland Experience with the Nellix EVAS System. Andrew Holden, MBChB, FRANZCR

Aortic stents, types, selection, tricks in deployment.

Treating very short necks ( 4mm <10mm) using the Endurant stent graft + EndoAnchors: 1-year results and current insights

Mid-term results from ANCHOR: How does this data influence the treatment algorithm for hostile EVAR anatomies

Reduction in cardiovascular related adverse events following active sac management with Nellix vs. EVAR: Are there biological advantages?

Anatomical challenges in EVAR

Il trattamento endovascolare dell aorta sottorenale (EVAR) mediante anidride carbonica (AI-CO 2 )

Original Excluder component overlap from proximal or distal extension during initial repair not correlated with aneurysm sac shrinkage

Abdominal Aortic Aneurysms. A Surgeons Perspective Dr. Derek D. Muehrcke

Endoleak Sealing after AAA Endovascular Repair. When and How?

How effective is preservation when viewed through a clinical and economic lens?

3. Endoluminal Treatment of Infrarenal Abdominal Aortic Aneurysm

Durability of The Endurant Stent-Graft through 5 Years

Expert Discussion: When to Preserve the Hypogastric Artery

Transcription:

The diagnostic and treatment challenge of type IIIb endoleaks Rodolfo Pini, MD, Gianluca Faggioli, MD, Chiara Mascoli, MD, Antonio Freyrie, MD, Mauro Gargiulo, MD, and Andrea Stella, MD, Bologna, Italy Type IIIb endoleak is a rare complication of endovascular aortic repair caused by endoprosthesis deterioration, leading to aneurysm pressurization and potential rupture. Because of its rarity, few cases are published. We report six cases of type IIIb endoleak in a 15-year period. Appropriate preoperative diagnosis was achieved in five cases; duplex ultrasonography, computed tomography angiography, and contrast-enhanced ultrasonography were useful diagnostic tools in one case, and angiography led to the correct diagnosis in four cases. In the remaining case, only surgical exploration identified the type IIIb endoleak. Successful treatment was achieved by endovascular relining in five cases and by surgical conversion in one case. (J Vasc Surg Cases 2015;1:249-53.) Late type IIIb endoleaks are due to structural failure of the endoprosthesis after endovascular aortic aneurysm repair (EVAR). 1 Although they are exceedingly rare, their identification is crucial because of aneurysm growth and the risk of rupture determined by direct sac pressurization. 2 In a literature search, only 15 cases have been identified so far; therefore, there is no standardized approach for diagnosis and treatment. 3 Type IIIb endoleak can also be associated with other types of endoleaks, leading to further difficulties in diagnosis and treatment choice. The aim of our work was to report a case series of type IIIb endoleaks identified and treated in the same institution. Each patient has consented to the publication of his data. CASE REPORT Patient 1. A 70-year-old man was treated (in another institution) with Vanguard (Boston Scientific, Oakland, NJ) in 1998. During follow-up, he maintained a stable aneurysm sac (60 mm); after 8 years, a new endoleak was identified by duplex ultrasound (DUS) at annual control, with a sac increase of 8 mm. The endoleak was detected between the endoprosthesis and the sac wall, with no communication with the proximal or distal landing zone. Computed tomography angiography (CTA) and contrastenhanced ultrasonography (CEUS) were also performed, but a secure endoleak diagnosis was not achieved. Diagnosis was made by digital subtraction angiography (DSA), identifying double type IIIb endoleaks from both iliac legs. A relining procedure was performed with two Anaconda (Vascutek, Glasgow, United From the Department of Vascular Surgery, University of Bologna. Author conflict of interest: none. Correspondence: Gianluca Faggioli, MD, Policlinico S. Orsola, via Massarenti 9, 40138 Bologna, Italy (e-mail: gianluca.faggioli@unibo.it). The editors and reviewers of this article have no relevant financial relationships to disclose per the Journal policy that requires reviewers to decline review of any manuscript for which they may have a conflict of interest. 2352-667X Copyright Ó 2015 The Authors. Published by Elsevier Inc. on behalf of the Society for Vascular Surgery. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/ 4.0/). http://dx.doi.org/10.1016/j.jvsc.2015.07.009 Kingdom) iliac legs. The procedure was completed without complication, and exclusion of the endoleaks was confirmed at intraoperative DSA and at postoperative and 1-year CTA (Fig 1). Patient 2. A 70-year-old man was treated with Vanguard (Boston Scientific) in 1999. During follow-up, the aneurysm sac shrunk from 65 mm to 38 mm. The DUS control after 9 years identified a new high-flow endoleak and a sac expansion of 9 mm. The endoleak type was not categorized by CTA and CEUS; only DSA revealed a type IIIb endoleak from the left iliac leg. Subsequently, one Anaconda (Vascutek) iliac leg was implanted. The procedure was completed without complication, and exclusion of the endoleaks was confirmed at intraoperative DSA and at postoperative and 1-year CTA. Patient 3. A 71-year-old man was treated with Anaconda (Vascutek) in 2005 for a 60-mm abdominal aortic aneurysm (AAA). The AAA remained stable in diameter with no endoleaks for 8 years; after that, a new high-flow endoleak was detected at DUS annual control with a sac growth of 11 mm. The endoleak type was not categorized by CTA and CEUS; only DSA revealed a type IIIb endoleak from the main body (Fig 1, C and D). Subsequently, a Cook (Cook Medical, Bloomington, Ind) proximal cuff was implanted. The procedure was completed without complication, and exclusion of the endoleaks was confirmed at intraoperative DSA and at postoperative and 1-year CTA (Fig 1). Patient 4. A 67-year-old man was treated for an AAA of 55 mm with an Anaconda (Vascutek) stent graft in 2007. After 7 years of DUS follow-up, he presented with a new endoleak in the distal portion of the right iliac limb with an increase in diameter of both common iliac arteries from 20 mm to 40 mm. A type Ib endoleak from the right iliac limb was suspected after CTA. To resolve the type Ib endoleak and to cover the left common iliac artery more (considering the growth of the iliac aneurysm), an extension in both the iliac arteries was planned with two Gore Excluder (W. L. Gore & Associates, Flagstaff, Ariz) limbs. During the procedure, retrograde DSA from the right femoral access failed to identify a type Ib endoleak, showing instead a type IIIb endoleak from the right iliac limb. The procedure was performed as planned, however, with deployment of the two iliac limbs and resolution of the endoleak, as confirmed by DUS and CTA postoperative control. Patient 5. A 79-year-old man who was treated (in another institution) in 2009 for an AAA of 55 mm with a Talent 249

250 Pini et al December 2015 Fig 1. Patient 1 (A and B) and patient 3 (C and D). A, Preoperative digital subtraction angiography (DSA) of a type IIIb endoleak from both iliac limbs (arrows). B, Postoperative computed tomography angiography (CTA) reconstruction after treatment with two iliac limbs. C, Preoperative DSA of a type IIIb endoleak from the main body (arrow). D, Postoperative CTA reconstruction after treatment with a proximal cuff. (Medtronic, Minneapolis, Minn) endoprosthesis and embolization of the left hypogastric artery, landing the left iliac limb in the external iliac artery, was discharged with a type II endoleak. The aneurysm sac slowly grew in the first 3 years after the procedure from 55 mm to 62 mm in diameter. At 4-year follow-up, the diameter increased to 70 mm, and CTA and DUS were performed. The type II endoleak was confirmed, and a possible type Ib endoleak from the right iliac limb was suspected. During

Volume 1, Number 4 Pini et al 251 Fig 2. Patient 5. Preoperative computed tomography (A) and volume rendering reconstruction of the endoleak (B, dotted line) with the identification of type II endoleak. The incomplete sealing of the right iliac limb led to an incorrect diagnosis of type Ib endoleak (C). D-F, Intraoperative angiography with exclusion of the type Ib endoleak from the right iliac limb (D), angiography of the nidus through cross-limb catheterization (E), and identification of the type IIIb endoleak (F, arrow). G and H, Intraoperative angiography of trans-limb sac embolization (G) and deployment of aortouni-iliac endoprosthesis with the exclusion of the type IIIb endoleak. I, Postoperative volume rendering reconstruction of the aortouni-iliac endoprosthesis and femoral-femoral crossover showing the resolution of the endoleaks. intraoperative DSA, the presence of type Ib endoleak was excluded, and a type IIIb endoleak from the endograft body was identified. The aortic sac was therefore embolized with a transright-limb approach (five MReye-Cook coils, Cook Medical) to treat the type II endoleak, and an aortouni-iliac endoprosthesis (Medtronic Endurant) was placed to exclude the type IIIb endoleak. The procedure was completed with femoral-femoral crossover and occlusion of the left external iliac artery. Intraoperative

252 Pini et al December 2015 Fig 3. Patient 6. A, Preoperative computed tomography angiography (CTA) with possible type II endoleak. B, Intraoperative identification of an endograft tear (arrow). DSA and postoperative CTA and CEUS confirmed the resolution of the endoleaks (Fig 2). Patient 6. A 77-year-old man was treated in 2006 for a 60- mm infrarenal AAA with a Talent (Medtronic) endoprosthesis. The aneurysm sac remained stable for 5 years. In 2011, DUS examination identified a 5-mm increase in the aneurysm sac, with possible late type II endoleak. After 6 months (in 2012), CTA and CEUS showed further 10-mm-diameter increase, with maximum diameter now reaching 75 mm. The patient was scheduled for sacculotomy and surgical lumbar artery ligation; however, a contained rupture occurred in the preoperative period. The patient underwent emergency surgical treatment, and a tear in the endoprosthesis was detected intraoperatively, with subsequent bifurcated aortoiliac grafting. The patient was discharged without complications after 10 days (Fig 3). DISCUSSION The six patients in this series were treated during a 16- year period from the original procedure; since that time, 958 standard EVAR procedures were performed in our high-volume center. Type IIIb endoleak occurrence should be considered an extremely rare complication (only four cases were treated initially in our hospital), developing 4 to 9 years after EVAR; moreover, no predictive elements were identified in the first CTA after the procedures. All the cases presented herein belong to first-generation endografts; in particular, Vanguard (Boston Scientific) and Talent (Medtronic) devices are reported in the literature with a high rate of structural damage. 4,5 Although the new-generation devices might be more durable in follow-up, no data are currently available in this regard. The high volume of EVAR procedures performed worldwide may lead to significant numbers of type IIIb endoleak, with subsequent necessity of careful follow-up and treatment. Type IIIb endoleak is a severe complication, leading to sac pressurization and potential rupture of the aneurysm, similar to type I endoleak. 6 Moreover, the diagnosis is particularly difficult, as shown by our series. In five of six cases, the diagnosis was achieved by DSA, either preoperatively or during the relining procedure. CTA, DUS, and CEUS are helpful in endoleak analysis and follow-up; however, in our experience, DSA was the only tool able to confirm the presence of the type IIIb endoleak. The use of different, expensive, and invasive diagnostic methods led us to consider type IIIb endoleak a difficult diagnosis. Other case reports underline the difficulties of a correct type IIIb endoleak diagnosis. 7,8 Candell et al, 8 in a retrospective analysis of 1700 EVAR procedures, reported a similar experience with an incorrect diagnosis of type I endoleak, with the identification of a structural tear only during the surgical conversion. CONCLUSIONS Similar to other experiences, a relining procedure was the preferred approach in our series. The structural characteristics of the endoprosthesis, in particular, the main body length, are important elements to consider during the relining procedure. Even if in our series (patient 3) no complications occurred, certainly endoprostheses with a long main body are associated with easy relining procedures. Other authors reported surgical correction of type IIIb endoleak with partial endograft substitution with glue or pledged surgical repair of the tear, with satisfactory results. 3 Although structural deterioration of the fabric is a rare complication, Ueda et al 9 identified a high percentage (18%) of metal ring fractures in AneuRx (Medtronic) endoprostheses at 5-year follow-up with three-dimensional CTA, with a significant type IIIb endoleak development. The long-term durability of the new devices is therefore exceedingly important to keep the occurrence of type IIIb endoleak at a minimum, particularly considering the increasing number of EVAR procedures. REFERENCES 1. Cao P, De Rango P, Verzini F, Parlani G. Endoleak after endovascular aortic repair: classification, diagnosis and management following

Volume 1, Number 4 Pini et al 253 endovascular thoracic and abdominal aortic repair. J Cardiovasc Surg (Torino) 2010;51:53-69. 2. Zhou W, Blay E Jr, Varu V, Ali S, Jin MQ, Sun L, et al. Outcome and clinical significance of delayed endoleaks after endovascular aneurysm repair. J Vasc Surg 2014;59:915-20. 3. Jones SM, Vallabhaneni SR, McWilliams RG, Naik J, Nicholas T, Fisher RK. Type IIIb endoleak is an important cause of failure following endovascular aneurysm repair. J Endovasc Ther 2014;21:723-7. 4. Espinosa G, Ribeiro Alves M, Ferreira Caramalho M, Dzieciuchowicz L, Santos SR. A 10-year single-center prospective study of endovascular abdominal aortic aneurysm repair with the talent stent-graft. J Endovasc Ther 2009;16:125-35. 5. Chakfé N, Dieval F, Riepe G, Mathieu D, Zbali I, Thaveau F, et al. Influence of the textile structure on the degradation of explanted aortic endoprostheses. Eur J Vasc Endovasc Surg 2004;27:33-41. 6. Lawrence-Brown MM, Sun Z, Semmens JB, Liffman K, Sutalo ID, Hartley DB. Type II endoleaks: when is intervention indicated and what is the index of suspicion for types I or III? J Endovasc Ther 2009;16(Suppl 1):I106-18. 7. Yoshitake A, Hachiya T, Itoh T, Kitahara H, Kasai M, Kawaguchi S, et al. Nonvisualized type III endoleak masquerading as endotension: a case report. Ann Vasc Surg 2015;29:595.e15-7. 8. Candell L, Tucker LY, Goodney P, Walker J, Okuhn S, Hill B, et al. Early and delayed rupture after endovascular abdominal aortic aneurysm repair in a 10-year multicenter registry. J Vasc Surg 2014;60: 1146-52. 9. Ueda T, Takaoka H, Petrovitch I, Rubin GD. Detection of broken sutures and metal-ring fractures in AneuRx stent-grafts by using threedimensional CT angiography after endovascular abdominal aortic aneurysm repair: association with late endoleak development and device migration. Radiology 2014;272:275-83. Submitted Apr 2, 2015; accepted Jul 24, 2015.