Extended Mass Range Triple Quadrupole for Routine Analysis of High Mass-to-charge Peptide Ions

Similar documents
Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Quantitation of Protein Phosphorylation Using Multiple Reaction Monitoring

LC/MS/MS of Trichothecenes and Zearalenone in Wheat Using Different Sample Prep Methods

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

LC/MS/MS Separation of Cholesterol and Related Sterols in Plasma on an Agilent InfinityLab Poroshell 120 EC C18 Column

Mycotoxin Analysis in Peanut Butter Using Captiva EMR Lipid Cleanup and LC/MS/MS

Reduced Ion Suppression and Improved LC/MS Sensitivity with Agilent Bond Elut Plexa

LC-MS/MS analysis of Chlorates in Milk and Whey Powder using the Agilent 6470 QQQ

Analysis of mycotoxins in food matrices using the Agilent Ultivo Triple Quadrupole LC/MS

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ

Authors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc.

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts

Quantitative Analysis of Underivatized Amino Acids in Plant Matrix by Hydrophilic Interaction Chromatography (HILIC) with LC/MS Detection

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System

Vitamin D3 and related compounds by ESI and APCI

Agilent 6490 Triple Quadrupole LC/MS System with ifunnel Technology. Ultra sensitive quantitative performance

Determination of Benzodiazepines in Urine by CE-MS/MS

Mycotoxin Analysis in Infant Formula Using Captiva EMR Lipid Cleanup and LC/MS/MS

Application Note. Author. Abstract. Introduction. Food Safety

Author. Introduction. Small Molecule Pharmaceuticals & Generics

Analysis of Vitamins Using an SFC/UHPLC Hybrid System with a Triple Quadrupole LC/MS for Quantification

Breaking the Sensitivity Barrier in Triple Quad LC/MS Performance

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography

2D-LC as an Automated Desalting Tool for MSD Analysis

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Determination of B-vitamins in Energy Drinks by CE/MS/MS

Analysis of Cholesterol-Lowering Drugs (Statins) Using Dried Matrix Spot Technology

The New 6495 Triple Quadrupole LC/MS

4-Plex itraq Based Quantitative Proteomic Analysis Using an Agilent Accurate -Mass Q-TOF

Multiclass Mycotoxin Analysis in Cheese Using Agilent Captiva EMR Lipid Cleanup and LC/MS/MS

[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

Supporting Information

Modified QuEChERS for HILIC LC/MS/MS Analysis of Nicotine and Its Metabolites in Fish

Using Software Tools to Improve the Detection of Impurities by LC/MS. Application Note. Christine Miller Agilent Technologies.

Separation and Quantitation of Oxysterol Compounds Using LC-MS/MS. Liquid Chromatography Mass Spectrometry SSI-LCMS-082

High Resolution Glycopeptide Mapping of EPO Using an Agilent AdvanceBio Peptide Mapping Column

Fat-Soluble Vitamins Analysis on an Agilent ZORBAX Eclipse PAH Polymeric C18 Bonded Column

Determination of Amantadine Residues in Chicken by LCMS-8040

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Improved CE/MS Sensitivity by Operating the Triple-Tube Coaxial Sheath-Flow Sprayer Without Appling Nebulizing Gas

Efficient Quantitative Analysis of THC and Metabolites in Human Plasma Using Agilent Captiva EMR Lipid and LC-MS/MS

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

Detection, Confirmation, and Quantification of Chloramphenicol in Honey, Shrimp and Chicken Using the Agilent 6410 LC/MS Triple Quadrupole

Determination of Amphetamine and Derivatives in Urine

MS/MS Library Creation of Q-TOF LC/MS Data for MassHunter PCDL Manager

Analysis of Acrylamide in French Fries using Agilent Bond Elut QuEChERS AOAC kit and LC/MS/MS

Agilent 6460 LC-QQQ Highly Sensitive and Robust Analysis for Lipophilic Marine Toxins in Shellfish

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Cannabinoid Quantitation Using an Agilent 6430 LC/MS/MS

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

Agilent 6410 Triple Quadrupole LC/MS. Sensitivity, Reliability, Value

Comparison of Relative Quantification of Monoclonal Antibody N-glycans Using Fluorescence and MS Detection

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Conflict of Interest Statement

Extraction of a Comprehensive Steroid Panel from Human Serum Using ISOLUTE. SLE+ Prior to LC/MS-MS Analysis

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Simultaneous Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots

Application Note # LCMS-89 High quantification efficiency in plasma targeted proteomics with a full-capability discovery Q-TOF platform

PTM Discovery Method for Automated Identification and Sequencing of Phosphopeptides Using the Q TRAP LC/MS/MS System

Identification of Steroids in Water by Ion Trap LC/MS/MS Application

Analyze Biogenic Amines in Salmon and Shrimp by Capillary Electrophoresis-Tandem MS

Determination of red blood cell fatty acid profiles in clinical research

Determination of Eight Estrogens in Milk by UHPLC and the Agilent 6495 Triple Quadrupole Mass Spectrometer

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Opiates, Opioids and Benzodiazepines, Amphetamines & Illicit Drug Forensic Analysis by LC/MS

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

Two-Dimensional LC/MS/MS to Reduce Ion Suppression in the Determination of Cannabinoids in Blood Plasma

ANALYTICAL SCALE AGILENT LCMS SYSTEMS OPERATIONAL QUALIFICATION

Author. Introduction. Abstract

Nature Biotechnology: doi: /nbt Supplementary Figure 1

Quantitative Analysis of Amphetamine-Type Drugs by Extractive Benzoylation and LC/MS/MS. Application. Introduction. Authors. Abstract.

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

LC/MS/MS of Buprenorphine and Norbuprenorphine in Whole Blood Using Agilent Bond Elut Plexa PCX and an Agilent Poroshell 120 Column

UPLC/MS Monitoring of Water-Soluble Vitamin Bs in Cell Culture Media in Minutes

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Highly Repeatable Ultra Low Detection of Estradiol Using Triple Quadrupole GC/MS in NCI Mode

Investigating Antibiotic Mode of Action Using Targeted Metabolomics with Ion-pairing Reversed-phase LC/Q-TOF

Food Analysis Applications with Triple Quadrupole LC/MS and TOF LC/MS

Profiling the Distribution of N-Glycosylation in Therapeutic Antibodies using the QTRAP 6500 System

Determination of Isoflavones in Soybean by LC/MS/MS

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

Quantitative Method for Amphetamines, Phentermine, and Designer Stimulants Using an Agilent 6430 LC/MS/MS

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

Application Note. Small Molecule Pharmaceuticals. 2-Ethyl-2-phenylmalonamide Ethosuximide Primidone Carbamazepine. Carbamazepine-

Separation of 15 Underivatized Saccharide and Sialic Acid USP Standards

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

All stocks and calibration levels were prepared in water: methanol (50:50) v/v to cover range of all steroid concentrations (refer Table 1).

Accumulation and transformation of inorganic and organic arsenic in rice and role of

Quantitative Determination of Drugs of Abuse in Human Plasma and Serum by LC/MS/MS Using Agilent Captiva EMR Lipid Cleanup

Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System

Edgar Naegele. Abstract

AppNote 4/2015. Automated Desorption, SPE Extraction, and LC/MS/MS Analysis of Dried Blood Spots KEYWORDS ABSTRACT

Transcription:

Extended Mass Range Triple Quadrupole for Routine Analysis of High Mass-to-charge Peptide Ions Application Note Targeted Proteomics Authors Linfeng Wu, Christine A. Miller, Jordy Hsiao, Te-wei Chu, Behrooz Zekavat, and Anabel Fandino Introduction Routine, high-throughput protein quantitation by LC/Triple Quadrupole analysis uses peptides as surrogates for the corresponding proteins. A critical step to successful analysis is to select unique peptides and suitable MRM transitions for each targeted peptide. Peptides with high m/z precursor or product ions are often not selected for analysis due to instrument mass range limitations. However, to address biological questions these peptides may provide critical information or be the only analytical choice. Some examples where this may arise are membrane proteins with large hydrophobic transmembrane domains, proteins with extensive post-translational modifications, and endogenously produced peptides. This Application Note demonstrates the performance of the Agilent 6495B Triple Quadrupole for peptides with high m/z product ions, showing the benefits of having a mass range up to m/z 3,. Tryptically digested monoclonal antibody (mab) standard and a HeLa cell membrane-enriched extract were used as simple and complex samples, respectively.

Experimental Method Protein/peptide selection Using an Agilent 655 Q-TOF Mass Spectrometer, LC/MS data were acquired in a data dependent mode on anti-il8 mab digest and a HeLa membrane enriched protein digest. Raw data were searched against the appropriate database using Agilent Spectrum Mill (Rev B.5..8 SP) to identify peptides producing high m/z product ions for testing the performance of the Agilent 6495B Triple Quadrupole Mass Spectrometer. LC/MS/MS system An Agilent 9 Infinity UHPLC system was interfaced to a 6495B Triple Quadrupole with a mass range of 3, u. The LC/MS experiments were performed using an Agilent Jet Stream (AJS) source, and an Agilent Poroshell EC C8,. mm.7 µm column (p/n 695775 9). For the mab digest, a 3.5 minute MRM method was used and for the HeLa membrane digest, a minute MRM method was used. Table. Agilent 6495 Triple Quadrupole MS method. Parameter Ion mode Setting Gas temperature 5 C Drying gas flow Nebulizer gas AJS, Positive 5 L/min 3 psi Sheath gas temperature C Sheath gas flow Capillary voltage Nozzle voltage High/Low pressure RF voltage Delta EMV Q and Q3 resolution Fragmentor Cell accelerator voltage Cycle time L/min 3,5 V V / V V Wide/Unit 38 V 4 V 67 369 ms For glycopeptide quantification, the mab digest was diluted in series, and the signal response for each injection level was plotted against the total mab digest amount on-column. For the HeLa membrane peptide quantification, the signal response for each product ion was plotted against the relative injection amount on-column. Table. Selected peptides and ions for LC/MS/MS test. Protein name Peptide Modification m/z Precursor m/z MRM Quantifier m/z MRM Qualifier Anti-IL8 mab antibody EEQYN[+66.6]STYR N-glycan GF 93.7 4. 366.,39.6,65.8,7.9 Sterol O-acyltransferase SSTVPIPTVNQYLYFLFAPTLIYR none,4.3,5. 76.5,.,9.,49.3 ATP Synthase subunit d, mitochondrial NLIPFDQMTIEDLNEAFPETK none,33.,63. 8. 34. 474.3,5. ADP/ATP Translocase YFAGNLASGGAAGATSLC[+57]FVYPLDFAR Carbamidomethylation,398.7 78.4 75. 88.5,95.9,6.

Results and Discussion Peptide identification and selection Anti-IL8 mab digest and HeLa membrane digest were used for peptide identification and MRM method development with the following results: One GF N-glycopeptide from the mab digest, and three peptides from the HeLa membrane digest with product ions ranging from m/z 4. to m/z,49.3 were selected for a LC/MS/MS test (Table ). Figure and Figure show the MS/MS spectrum and the optimized MRM chromatography for two of the peptides. The product ions selected for the MRM method are marked with * in the MS/MS plot. Glycopeptides produce high m/z product ions, which have large glycan chains. These product ions are informative for interpreting glycan structure. Product ions with higher m/z tend to have less background noise or higher signal-to-noise ratios in complex samples. B 3 9 8 7 6 5 4 3 A 4 6 8,,,4,6,8,, Mass-to-charge (m/z) 4 3.9.. Acquisition time (min) Figure. MS/MS spectrum and the optimized MRM chromatography for GF glycopeptide EEQYN[+66.6]STYR. A precursor B 3 3..8.4..6..8 b3-h O b4 b5 b6 y6 y7 y + y8 + y8 y y3 y8 y5 y 5.4 6,,4,8,,6 3, Mass-to-charge (m/z) 6.7 6.8 6.9 7. 7. 7. 7.3 7.4 Acquisition time (min) Figure. MS/MS spectrum and the optimized MRM chromatography for peptide SSTVPIPTVNQYLYFLFAPTLIYR. 3

Glycopeptide quantification An 9 Infinity UHPLC system coupled with Agilent Jet Stream Technology, and a 6495B Triple Quadrupole LC/MS system with a mass range of 3, u were used for glycopeptide quantitation with the following results (Figure 3): Excellent precision and accuracy at all levels including injection amount as low as.97 ng (3 amol) of total mab digest on-column (8.5 % RSD, 97.5 % accuracy). This target peptide is estimated to represent about % of the total protein so LOQ for this glycopeptide is estimated at about 3 amol on-column Reproducible responses for product ions with extended mass range (m/z 4. ~ m/z,7.9) Excellent retention time (RT) reproducibility (RSD% =.47 % for n = 9) Excellent linearity at low injection levels with small fold changes (.97 ng ~ 9.7 ng, R =.998).48.46.48.46.5.6.6 C A Blank.98 ng.97 ng.394 ng.59 ng.9.95..5. Acquisition time (min) B Amount of total mab digest on-column (ng) %RSD (n = ) % Accuracy.97 8.5 97.5.394 8.3 4..59.5 8.3.985 9.3.4.97 3. 95.6.394 5.9 84.8.985 5.7..97.6 5.5 9.7.9 99.7, 93.7 & 4. 8, 93.7 &,7.9,4 6, 93.7 &,65.8, 4, 93.7 &,39.6,, 93.7 & 366., 8 8, 6 6, 4 4,, Zoom 3 4 5 6 7 8 9 3 4 5 6 7 8 9 4.9.8.7.6.5.4.3....9.8.7.6.5.4.3.. -. D EEQYN(+66.58676)STYR y = 943.67754*x 5.93398 R =.99843 Zoom 3..6..8.4..6. -..4.8..6..4.8-3 4 5 6 7 8 9 3 4 5 6 7 8 9 Concentration (ng of mab digest on-column) Figure 3. Quantification for GF glycopeptide EEQYN[+66.6]STYR. 4

HeLa membrane peptide quantification Three peptides from the HeLa membrane enriched sample digest were quantified following the same protocol as the glycopeptide to test high m/z product ion behavior (Figure 4). The quantifier response for each of these three peptides were plotted against the relative injection amount on column (left plots) as the stoichiometry of these peptides were unknown in the sample. Good linearity was observed (R ranges from.989 to.997). Reproducible responses for product ions with extended mass range (up to m/z 49.3) were observed (right plots). 4 YFAGNLASGGAAGATSLCFVYPLDFAR. y =.88488*x 54.476396.9 R =.9957687.8.7.6.5.4.3.. -. 4 8 6 4 8 3 36 4 44 48 5 Concentration (relative amount on-column),, 8, 6, 4,, 3 4 5,398.7 & 78.4,398.7 &,6.,398.7 &,95.9,398.7 & 88.5,398.7 & 75. 4...9.8.7.6.5.4.3.. -. NLIPFDQMTIEDLNEAFPETK y = 5.55859*x 3.485689 R =.997434 4 8 6 4 8 3 36 4 44 48 5 Concentration (relative amount on-column),, 8, 6, 4,, 3 4 5,33. &,63.,33. &,5.,33. & 474.3,33. & 34.,33. & 8. 3.5.4.3....9.8.7.6.5.4.3.. -. SSTVPIPTVNQYLYFLFAPTLIYR y = 9.685*x.58556 R =.9898869 4 8 6 4 8 3 36 4 44 48 5 Concentration (relative amount on-column),6,4,, 8 6 4 3 4 5,4.3 &,5.,4.3 &,49.3,4.3 &,9.,4.3 &,.,4.3 & 76.5 Figure 4. Quantification for peptides from a HeLa membrane digest. 5

Conclusions The Agilent 6495B Triple Quadrupole LC/MS system has a mass range of 3, u, which is useful for detecting high m/z peptide ions. One benefit of using high m/z precursor or product ions, particularly in a complex digest, is these ions have less background noise compared to those with low m/z. Some high m/z product ions may provide important biological information such as location and size of post-translational modifications, for example glycosylation. This application notes demonstrates the excellent reproducibility for MRM transitions with high m/z product ions achieved on the Agilent 6495B Triple Quadrupole LC/MS system. The work shown here uses Agilent Jet Stream Technology at standard flow rates and resulted in outstanding precision and accuracy for both glycopeptide and HeLa membrane peptide quantification. www.agilent.com For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc., 6, 8 Published in the USA, March 6, 8 599-798EN