SIV p27 Antigen ELISA Catalog Number:

Similar documents
HIV-1 p24 Antigen ELISA Catalog Number:

HIV-1 p24 Antigen ELISA 2.0 Catalog Number:

HIV-1 ICx/CRx Kit Reagents for Immune Dissociation and Reactivity Confirmation

Human TSH ELISA Kit. User Manual

SIV p27 ANTIGEN CAPTURE ASSAY

ACTG Laboratory Technologist Committee Revised Version 2.0 ACTG Lab Man Coulter HIV-1 p24 ELISA May 21, 2004

HBeAg and HBeAg Ab ELISA Kit

HIV-1 p24 ANTIGEN CAPTURE ASSAY

HIV-1 p24 ELISA Kit. Catalog Number KA assays Version: 06. Intended for research use only.

Cotinine (Mouse/Rat) ELISA Kit

Procine sphingomyelin ELISA Kit

HIV-1 p24 ELISA Kit. Cat.No: DEIA10155 Lot. No. (See product label) Size. Storage. Principle Of The Test. Reagents And Materials Provided

Mouse C-Peptide ELISA Kit

Serum Amyloid A ELISA

Human Apolipoprotein A1 EIA Kit

Instructions for use. TSH rat ELISA. Please use only the valid version of the Instructions for Use provided with the kit AR E-8600

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit

DRG International, Inc., USA Fax: (908)

Insulin (Porcine/Canine) ELISA

EliKine Free Thyroxine (ft4) ELISA Kit

Human HIV (1+2) antigen&antibody ELISA Kit

Human Immunodeficiency Virus type 1 (HIV-1) gp120 / Glycoprotein 120 ELISA Pair Set

Canine Thyroid Stimulating Hormone, TSH ELISA Kit

Ultra-sensitive Human TSH ELISA Kit. MyBioSource.com

TECHNICAL BULLETIN. Catalog Number RAB0447 Storage Temperature 20 C

RayBio Human Granzyme B ELISA Kit

Apolipoprotein A-1 ELISA

Human Immunodeficiency Virus type 1 (HIV-1) p24 / Capsid Protein p24 ELISA Pair Set

TNF-alpha ELISA. For Research Use Only. Not For Use In Diagnostic Procedures.

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

TECHNICAL BULLETIN. GLP-1 EIA Kit for serum, plasma, culture supernatant, and cell lysates. Catalog Number RAB0201 Storage Temperature 20 C

TECHNICAL BULLETIN. Phospho-Akt (pser 473 ) ELISA Kit for detection of human, mouse, or rat phospho-akt (pser 473 ) in cell and tissue lysates

Influenza A H7N9 (A/Anhui/1/2013) Hemagglutinin / HA ELISA Pair Set

Human LDL Receptor / LDLR ELISA Pair Set

Mouse HBsAg(Hepatitis B Virus Surface Antigen) ELISA Kit

Hepatitis A virus IgM ELISA Kit

Porcine/Canine Insulin ELISA

Rat IL-2 ELISA Kit. Instructions for use. For research use only. Fast Track Your Research.

Bovine Insulin ELISA

Morinaga Mouse C-peptide ELISA Kit

Rat cholesterol ELISA Kit

Mouse C-peptide ELISA

Instructions for use. TSH rat ELISA. Please use only the valid version of the Instructions for Use provided with the kit AR E-8600

STAT1 (ps727) (Human/Mouse) ELISA Kit

Influenza B Hemagglutinin / HA ELISA Pair Set

RayBio Human ENA-78 ELISA Kit

Influenza A IgG ELISA

Mouse C-peptide ELISA

Human HBcAb IgM ELISA kit

EGFR (py1045)/ Pan EGFR (Human) ELISA Kit

Triiodothyronine (T3) ELISA

Mouse/Rat THYROXINE (T4) ELISA Catalog No (96 Tests)

HIV-1 p24 ELISA Pair Set Cat#: orb54951 (ELISA Manual)

Mouse Ultrasensitive Insulin ELISA

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum

IV2-113E Use by. Invitron Glargine ELISA Kit REF LOT IVD. Definitions. English. For in-vitro diagnostic use. Instructions for use.

RayBio Human Phosphotyrosine BTK ELISA Kit

Rubella virus IgG ELISA Kit

Prothrombin (Human) ELISA Kit

Mouse Leptin ELISA Kit Instructions

RayBio Human Phospho-DDR1 (Tyr792) ELISA Kit

ab Hemoglobin Human ELISA Kit

Mouse C3 (Complement Factor 3) ELISA Kit

Human Aldosterone ELISA Kit

RayBio Human Phospho-DDR1 (Tyr792) and Total DDR1 ELISA Kit

TECHNICAL BULLETIN. C-Peptide EIA Kit for serum, plasma, culture supernatant, and cell lysates. Catalog Number RAB0326 Storage Temperature 20 C

Influenza A H1N1 (Swine Flu 2009) Hemagglutinin / HA ELISA Pair Set

APOB (Human) ELISA Kit

Fish follicle-stimulating hormone, FSH ELISA Kit

STAT3 (py705) (Human/Mouse/Rat) ELISA Kit

TSH Receptor Autoantibody ELISA

STAT3 (py705)/ Pan STAT3 (Human/Mouse/Rat) ELISA Kit

RayBio Human, Mouse and Rat Phospho-NF-kB P65 (Ser536) and Total NF-kB P65 ELISA Kit

Influenza A H1N1 HA ELISA Pair Set

RayBio Human Phospho-DDR2 (Tyr740) and Total DDR2 ELISA Kit

Leishmania Antigen ELISA (VL ELISA)

RayBio Human Thyroglobulin ELISA Kit

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum. For Research Use Only. Not For Use In Diagnostic Procedures.

Helicobacter pylori IgA ELISA Kit

HbA1c (Human) ELISA Kit

Lenti-X p24 Rapid Titer Kit User Manual

GLP-2 (Rat) ELISA. For the quantitative determination of glucagon-like peptide 2 (GLP-2) in rat serum and plasma

Rat Proinsulin ELISA

Thyroid Stimulating Hormone (S-TSH) Thyroid Stimulating

Hepatitis B surface antigen Ab ELISA Kit

Human LDL ELISA Kit. Innovative Research, Inc.

Human IL-2 ELISA Antibody Pair

Lenti-X p24 Rapid Titer Kit User Manual

T3 (Total) (Mouse/Rat) ELISA Kit

Human Urokinase / PLAU / UPA ELISA Pair Set

2009 H1N1 Influenza ( Swine Flu ) Hemagglutinin ELISA kit

INSTRUCTION MANUAL FOR HBsAg SUBTYPE EIA

Mouse IFN gamma Coated ELISA Kit

Human Cytomegalovirus Virus (CMV) IgG ELISA Kit

INSTRUCTION FOR USE EIA-ANTI-HDV

Rabbit Glycated hemoglobin A1c (GHbA1c) ELISA Kit

Human Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA Kit

Human Cathepsin V ELISA Kit

NF-κB p65 (Phospho-Thr254)

Transcription:

INTENDED USE The RETRO-TEK SIV p27 Antigen ELISA is for research use only and is not intended for in vitro diagnostic use. The RETRO-TEK SIV p27 Antigen ELISA is an enzyme linked immunoassay used to detect Simian Immunodeficiency Virus (SIV) p27 antigen in research specimens, including cell culture media. It can also be used to monitor the purification and biochemical behavior of SIV. Furthermore, the assay may augment or supplant reverse transcriptase measurements traditionally employed to detect the presence of SIV. Such enzymatic measurements are not SIV specific. PRINCIPLE OF THE TEST Microwells are coated with a monoclonal antibody specific for the p27 gag gene product of SIV. Viral antigen in the specimen is specifically captured onto the immobilized antibody during specimen incubation. The captured antigen is then reacted with a high-titer polyclonal anti-siv antibody conjugated to biotin. Following a subsequent incubation with streptavidin peroxidase, color develops as the bound enzyme reacts with the substrate. Resultant optical density is proportional to the amount of SIV p27 antigen present in the specimen. REAGENTS Materials Supplied: SIV p27 Antibody Coated Microplate for 96 determinations,1 plate: 12x8 well strips. SIV p27 Detector Antibody, 1 vial, lyophilized: Contains biotin-labelled polyclonal antibody to SIV, milk, Tween, normal human serum and PBS. Detector Antibody Diluent: 25 ml. SIV p27 Antigen Standard, 0.5 ml: Contains detergent-disrupted, heat-inactivated viral antigen, goat serum, TritonX- 100 and sodium azide Lysing Buffer, 5 ml: Triton X-100 in PBS and 2-chloroacetamide. Streptavidin peroxidase, 0.3 ml: Contains streptavidin conjugated to horseradish peroxidase, PBS, goat serum, and 2- chloroacetamide. Assay Diluent, 100 ml: Contains goat serum, PBS, Triton X-100 and 2-chloroacetamide. 10X Plate Wash Buffer, 125 ml: Contains PBS, Tween 20 and 2-chloroacetamide Substrate, 0.3 ml: Contains tetramethylbenzidine (TMB) and dimethyl sulfoxide. Substrate Buffer, 50 ml: Contains citrate/acetate buffer, hydrogen peroxide and 2-chloroacetamide. Stop Solution, 12 ml: Proprietary formulation Plate Sealers, 10 Biological Risk

Resealable Plastic Bag. Triton X-100 is a registered trademark of Union Carbide Chemicals and Plastics Co., Inc. Tween 20 is a registered trademark of Imperial Chemical Industries. Materials required but not supplied: Disposable gloves Validated adjustable micropipettes, single and multichannel Test tubes and racks for preparing specimen and control dilutions Graduated cylinders and assorted beakers Validated automatic microplate washer or manual vacuum aspiration equipment Validated incubator for 37 C ±1 C Validated microplate reader Timer 1% sodium hypochlorite as disinfectant. May be prepared from household bleach Distilled or deionized water STORAGE Store all kit reagents at 2-8 C. DO NOT FREEZE. When stored properly the kit is stable until the date indicated on the box label. PRECAUTIONS FOR RESEARCH USE ONLY. Not For in vitro Diagnostic Use. Prior to performing the assay, carefully read all instructions. Use universal precautions when handling kit components and test specimens. * To avoid cross-contamination, use separate pipet tips for each specimen. Disposal: When testing potentially infectious human specimens, adhere to all applicable local, state and federal regulations regarding the disposal of biohazardous materials. Human source material used in the manufacture of the SIV Detector Antibody has been tested and found negative for Hepatitis B surface antigen. The viral lysate used to prepare the SIV-p27 Antigen Standard has been inactivated by chemical disruption and heating. Handle these reagents as if capable of transmitting infectious agents. The SIV p27 Antigen Standard contains sodium azide as a preservative. Sodium azide may react with lead or copper pipes to form explosive metal azides. Flush pipes with large quantities of water upon disposal to prevent azide buildup in drains. Stop Solution contains hydrochloric acid which may cause severe burns. In case of contact with eyes or skin, rinse immediately with water and seek medical assistance. Wear protective clothing and eyewear. * from MMWR, June 24, 1988, Vol. 37, No. 24, pp. 377-382, 387-388. Biological Risk

PREPARATION OF REAGENTS SIV p27 Detector Antibody: Add 11.0 ml of Detector Antibody Diluent to the bottle of SIV p27 Detector Antibody. Mix or rock gently for at least 1 hour. Reconstituted reagent can be stored for one week at 2-8 C. Alternatively, the reconstituted SIV p27 Detector Antibody can be divided into aliquots and frozen at -20 C. SIV p27 Antigen Standard: Prepare a series of six standards from the SIV p27 Antigen Standard. Use the dilution scheme in Table 1. Any diluted SIV p27 Antigen Standard remaining after the completion of the assay should be disinfected and discarded. Do not save diluted reagent. Table 1 Preparation of SIV p27 Antigen Standard Standard Number Concentration of SIV p27 (pg/ml) SIV p27 Antigen Standard (µl) Assay Diluent (µl) 1 2000.0 50 950 2 1000.0 500 of #1 500 3 500.0 500 of #2 500 4 250.0 500 of #3 500 5 125.0 500 of #4 500 6 62.5 500 of #5 500 7 0 0 500 Streptavidin peroxidase and Substrate: To prepare the Streptavidin peroxidase Working Solution and the Substrate Working Solution, use the dilution schemes in Table 2. Use Substrate Working Solution within 15 minutes of its preparation. Any diluted Substrate Working Solution and Streptavidin peroxidase Working Solution remaining after the completion of the assay must be discarded. Table 2 Preparation of Streptavidin peroxidase Working Solution and Substrate Working Solution Streptavidin peroxidase Substrate Working Solution Number Working Solution of Strips Streptavidin Assay Substrate Used peroxidase(l) Diluent (ml) Substrate(µl) Buffer (ml) 3 40 4.0 40 4.0 6 70 7.0 70 7.0 9 100 10.0 100 10.0 12 120 12.0 120 12.0 Plate Wash Buffer: Dilute 10X Plate Wash Buffer 1:10 in distilled or deionized water prior to use. 1X Plate Wash Buffer may be stored at 2-8 C for up to 1 week. Additional bottles of 10X Plate Wash Buffer (ZMC Catalog #: 0801060) may be ordered. Biological Risk

TEST PROCEDURE Allow all reagents to reach room temperature before use. Label test tubes to be used for the preparation of standards and specimens. Label each strip on its end tab to identify the strips should they become detached from the plate frame during the assay. If the entire 96 well plate is not used, remove surplus strips from the plate frame. Place surplus strips and desiccant into the Resealable Plastic Bag, seal and store at 2-8 C. Step 1: Treat specimens in a test tube by pipetting 50 µl of Lysing Buffer into 450 µl specimen and mix well. Step 2: Wash each well of the microplate or the selected number of strips 6 times with 300µl of 1X Plate Wash Buffer and aspirate. Thoroughly blot by striking inverted microplate or strips on a pad of absorbent towels. Continue striking until no droplets remain in the wells. Do not allow plates to dry completely prior to sample addition. Drying will adversely affect test results. Step 3: Leave one well of the microtiter plate empty during the assay. This well is used for a substrate blank. Step 4: Pipet 200 µl of standards #1-7 into duplicate wells. Step 5: Pipet 200 µl of each specimen, as prepared in Step 1, into duplicate wells. Step 6: Cover microplate with a plate sealer and incubate as follows, either: a. Overnight at 37 C ± 1 C or b. 2 hours at 37 C ± 1 C Step 7: Aspirate and wash plate as described in Step 2. Step 8: Pipet 100 µl of reconstituted SIV p27 Detector Antibody into each well, except the substrate blank. Cover the microplate with a sealer and incubate for 1 hour at 37 C ± 1 C. Step 9: Aspirate and wash plate as described in Step 2. Step 10:Pipet 100 µl of the Streptavidin Peroxidase Working Solution into each well except the substrate blank. Cover the microplate with a sealer and incubate for 30 minutes at 37 C ± 1 C. Step 11: Aspirate and wash plate as described in Step 2. Step 12:Pipet 100 µl of Substrate Working Solution into all wells and incubate uncovered for 30 minutes at room temperature (18-25 C). A blue color will develop in wells containing viral antigen. Step 13:Stop the reaction by pipetting 100 µl of Stop Solution into each well. A color change from blue to yellow will result. Step 14:Within 15 minutes, read the optical density of each well at 450 nm using a microplate reader. Biological Risk

TEST VALIDITY Determine the mean optical density values for each standard and specimen. For the test to be valid, it must meet the following criteria: The mean optical density of the 0 pg/ml standard and the substrate blank must be less than 0.200. The mean optical density of the 1000 pg/ml standard must be greater than or equal to 0.500 Cut-off Determination for a Qualitative Assay: Determine the mean optical density of the media or normal plasma control wells. Add a factor of 0.05 Alternatively, run a series of normal specimens and determine cut-off as the mean plus five standard deviations. CALCULATION OF RESULTS To Quantitate Levels of SIV p27: Using linear graph paper, plot the concentration of SIV p27 Antigen Standard (pg/ml) on the X-axis versus the mean optical densities for each standard on the Y-axis. Then determine the concentration of SIV p27 antigen in specimens by interpolation from the standard curve. Alternatively, the level of SIV p27 may be calculated by computer using a point-to-point algorithm. Be sure to correct for all dilutions, including the 1 to 1.1 dilution made during the addition of Lysing Buffer. Typical Standard Curve SIV p27 Antigen Concentration Optical Density at 450 nm 2000.0 pg/ml 1.247 1000.0 pg/ml 0.755 500.0 pg/ml 0.469 250.0 pg/ml 0.303 125.0 pg/ml 0.208 62.5 pg/ml 0.161 0.0 pg/ml 0.108 Biological Risk

PROCEDURAL FLOW CHART PREPARE REAGENTS WASH PLATES PIPET SPECIMENS, STANDARDS AND CONTROLS INCUBATE OVERNIGHT AT 37º + 1ºC or INCUBATE 2 HOURS AT 37º + 1ºC WASH PLATE PIPET DETECTOR ANTIBODY INCUBATE 1 HOUR AT 37 C ± 1 C. WASH PLATE PIPET STREPTAVIDIN PEROXIDASE SOLUTION INCUBATE 30 MINUTES AT 37 C ± 1 C. WASH PLATE PIPET SUBSTRATE SOLUTION INCUBATE 30 MINUTES AT ROOM TEMPERATURE ADD STOP SOLUTION AND READ AT 450 NM NOT for in vitro Diagnostic Use Biological Risk