IMPROVED CEMENTING TECHNIQUES AND FEMORAL COMPONENT LOOSENING IN YOUNG PATIENTS WITH HIP ARTHROPLASTY

Similar documents
ORIGINAL PAPER. Department of Orthopedic Surgery, Hamamatsu Medical Center ABSTRACT

Charnley Total Hip Arthroplasty with Use of Improved Cementing Techniques A MINIMUM TWENTY-YEAR FOLLOW-UP STUDY

AGGRESSIVE GRANULOMATOUS LESIONS IN CEMENTLESS TOTAL HIP ARTHROPLASTY

Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists

AGGRESSIVE GRANULOMATOUS LESIONS AFTER HIP ARTHROPLASTY

INTRAMEDULLARY CANCELLOUS BONE BLOCK TO IMPROVE FEMORAL STEM FIXATION IN CHARNLEY LOW-FRICFION ARTHROPLASTY B. M. WROBLEWSKI, ADRIAN VAN DER RIJT

Long-term Outcome of Polished Stems in Total Hip Arthroplasty

Twenty-five-Year Results After Charnley Total Hip Arthroplasty in Patients Less than Fifty Years Old A CONCISE FOLLOW-UP OF A PREVIOUS REPORT*

Results of Conversion Total Hip Prosthesis Performed Following Painful Hemiarthroplasty

Optimizing function Maximizing survivorship Accelerating recovery

CIRCUMFERENTIAL PROXIMAL FEMORAL ALLOGRAFTS IN REVISION SURGERY ON TOTAL HIP ARTHROPLASTY: CASE REPORTS WITH A MINIMUM FOLLOW-UP OF 20 YEARS

Continuing the Tradition. VerSys Heritage Hip System

Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem

Revision hip arthroplasty with S-ROM prosthesis: a study of clinical outcomes and implant stability

Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty

111. ACETABULAR FIXATION

CAUTION: Ceramic liners are not approved for use in the United States.

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

Femoral Osteolysis Around the Unrevised Stem During Isolated Acetabular Revision

TEN YEARS EXPERIENCE WITH NONCEMENTED REVISION SOCKETS. Lawrence D. Dorr, MD Zhinian Wan, MD

Optimum implant geometry

Pregnancy After Total Hip Arthroplasty BY CATHY M. MCDOWELL, RN, AND PAUL F. LACHIEWICZ, MD

Tri-Lock Bone Preservation Stem

REVIEW ARTICLE The use of cemented and cementless femoral components in revision total hip arthroplasty: a review of the literature

Revision total hip arthroplasty with cemented femoral component

Centralization of the femoral component in cemented hip arthroplasty using guided stem insertion

Fracture of an Acetabular Component Inserted without Cement: A Case Report

SEVERE VARUS AND VALGUS DEFORMITIES TREATED BY TOTAL KNEE ARTHROPLASTY

The Fate of Revised Uncemented Acetabular Components in Patients With Rheumatoid Arthritis

Initial Evaluation on Subsidence of Cemented Collarless Polished Tapered Stem Applied to the Patients with Narrow Femoral Medullar Canal

Outcomes of Charnley total hip arthroplasty using improved cementing with so-called second- and third-generation techniques

N Plain films are usually sufficient to assess the state of. N Criteria for loosening depends on site and make of

Optimum implant geometry

The shape and size of femoral components in revision total hip arthroplasty among Chinese patients

Case Report Massive Femoral Osteolysis Secondary to Loosening of a Cemented Roughened Long Stem: A Case Report

The Exeter Trauma Stem: Early results of a new cemented Hemiarthroplasy for femoral neck fracture

Optimum implant geometry

Extensively Porous-coated Stems for Femoral Revision: Reliable Choice for Stem Revision in Paprosky Femoral Type III Defects

Hedrocel trabecular metal monoblock acetabular cups : mid-term results

Cementless Total Hip Arthroplasty with Medial Wall Osteotomy for the Sequelae of Septic Arthritis of the Hip

Socket wall addition device in the treatment of recurrent hip prosthesis dislocation

MINIMUM TEN-YEAR FOLLOW-UP OF COMPUTED TOMOGRAPHY BASED, CUSTOM CEMENTLESS STEM AFTER INTERTROCHANTERIC OSTEOTOMY FOR DYSPLASTIC HIPS

Conversion total hip arthroplasty Functional outcome in Egyptian population

Segmental Fracture Of A Cemented Femoral Stem - A Case Report And Review Of Litrature

CORAIL HIP SYSTEM DESIGN RATIONALE

Ito Hiroshi, Tanino Hiromasa, Yamanaka Yasuhiro, Minami Akio, Matsuno Tateo

After Total Hip Arthroplasty Comparison of a Traditional Disease-specific and a Quality-of-life Measurement of Outcome

A mechanical review of postoperative stem fractures of cemented total hip arthroplasty implants without femoral fracture

Function and Fixation of Total Hip Arthroplasty in Patients 25 Years of Age or Younger

ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY

treatment of the rheumatoid hip

Early results of Trabecular Metal augment for acetabular reconstruction in revision hip arthroplasty

Current attitudes to cementing techniques in British hip surgery

Ten-year survival of the cemented MS-30 femoral stem Increased revision rate in male patients

The science of simplicity

Shunsaku Nishihara 1, Nobuhiko Sugano 1, Takashi Nishii 1, Hisashi Tanaka 2, Hideki Yoshikawa 1, and Takahiro Ochi 1

The Journal of Arthroplasty

Cotyloplasty in Cementless Total Hip Arthroplasty for an Insufficient Acetabulum

Total Hip Arthroplasty around the Inception of the Interface Bioactive Bone Cement Technique

Midterm results of cemented Press Fit Condylar Sigma total knee arthroplasty system

HELIOS h i p s y s t e m

DESIGN RATIONALE AND SURGICAL TECHNIQUE

Mid Term Results of Total Hip Arthroplasty for Various Hip Disorders

Revision of the cemented femoral stem using a cement-in-cement technique

Reprint requests: Dr Brenda Dower CLINICAL ARTICLE SA ORTHOPAEDIC JOURNAL Spring 2012 Vol 11 No 3 / Page 29

Bone Bangalore

LONG TERM RESULTS OF REVISION TOTAL HIP ARTHROPLASTY

Bilateral total knee arthroplasty: One mobile-bearing and one fixed-bearing

Effect of Superior Placement of the Hip Center on Abductor Muscle Strength in Total Hip Arthroplasty

REVISING THE DEFICIENT PROXIMAL FEMUR

Femoral stem loosening rate of less than 0.16% per year at years 1. Exeter Hip

simplicity THE SCIENCE OF DESIGN RATIONALE

FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY

HYDROXYAPATITE IN REVISION OF TOTAL HIP REPLACEMENTS WITH MASSIVE ACETABULAR DEFECTS

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Management Of Acetabular Deficiency In Total Hip Arthroplasty: A Series Of 15 Cases

Thigh Pain in Primary Total Hip Arthroplasty

Poor mid-term results of total hip arthroplasty with use of a Hylamer liner

SUMMIT and DURALOC. Clinical Summary

Cementation Prof. Anil Arora MS (Ortho) DNB (Ortho) Dip SIROT (USA) FAPOA (Korea), FIGOF (Germany), FJOA (Japan) Associate Director

Stability without compromise. Epsilon Durasul Constrained Insert

TaperFit Cement-in-Cement Revision Femoral Stem Surgical technique

Furlong H-A.C. THR System

Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Metha Short Hip Stem System

EXTENDED TROCHANTERIC OSTEOTOMY SURGICAL TECHNIQUE FPO EXTENSIVELY COATED FIXATION

Arcos Modular Femoral Revision System

Cementless Acetabular Fixation With and Without Screws

The Exeter Universal stem

VERSYS HERITAGE REVISION HIP PROSTHESIS. Surgical Technique for Revision Hip Arthroplasty

TriboFit Hip System. Issue 3 - August 2013 English

ACTIS. TOTAL HIP SYSTEM Approach Active Patients with Confidence DESIGN RATIONALE

Knee Revision. Portfolio

Radiographic Wear Measurements in a Cementless Metal-Backed Modular Cobalt-Chromium Acetabular Component

Progeny Hip Stem. Surgical Protocol and Product Specifications

Design Rationale. ECHELON Primary Hip System

Principles of acetabular fixation in primary and revision hip arthroplasty Piotr Wojciechowski, Damian Kusz, Anna WAGNER

J of Evolution of Med and Dent Sci/ eissn , pissn / Vol. 4/ Issue 22/ Mar 16, 2015 Page 3785

Uncemented Total Hip Arthroplasty With Subtrochanteric Derotational Osteotomy for Severe Femoral Anteversion

Transcription:

IMPROVED CEMENTING TECHNIQUES AND FEMORAL COMPONENT LOOSENING IN YOUNG PATIENTS WITH HIP ARTHROPLASTY A 12-YEAR RADIOGRAPHIC REVIEW ROBERT L. BARRACK, RICHARD D. MULROY JR, WILLIAM H. HARRIS From the Massachusetts General Hospital and Harvard Medical School, Boston To assess the effect of improved methods of femoral cementing on the loosening rates in young patients, we reviewed 50 second-generation cemented hip arthroplasties in 44 patients aged 50 years or less. The femoral stems were all collared and rectangular in cross-section with rounded corners. The cement was delivered by a gun into a medullary canal occluded distally with a cement plug. A clinical and radiographic review was undertaken at an average of 12 years (10 to 14.8) and no patient was lost to follow-up. No femoral component was revised for aseptic loosening, and only one stem was definitely loose by radiographic criteria. By contrast, 11 patients had undergone revision for symptomatic aseptic loosening of the acetabular component and 11 more had radiographic signs of acetabular loosening. Aseptic loosening ofthe components is the most common long-term complication of cemented total hip arthroplasty. At five-year follow-up, radiographic loosening of the femora! component was reported in 20% to 24% of cases in which the cementing methods initially advocated by Charnley, which we call first-generation techniques, had been used (Amstutz et a! 1976 ; Beckenbaugh and Ilstrup 1978). The figures increased to 30% to 40% at ten years (Stauffer 1982 ; Sutherland et a! 1982). Improved cementing techniques can substantially reduce the mcidence of femora! loosening (Harris and McGann 1986; Roberts, Poss and Kelley 1986 ; Russotti, Coventry and Stauffer 1988) to as little as 3% at 1 1 years (Mulroy and Harris 1990). In young patients the incidence of failure of cemented hip replacements is reported to be even higher (Chandler et a! 1981 ; Dorr, Takei and Conaty 1983; R. L. Barrack, MD, Assistant Professor of Orthopaedic Surgery and Director Division of Reconstructive Surgery, Tulane University School of Medicine, 1430 Tulane Avenue, New Orleans, Louisiana 70112, USA. W. H. Harris, MD, Chief, Hip and Implant Surgery Unit Orthopaedic Biomechanics Laboratories, Massachusetts General Hospital, Boston, Massachusetts 02114, USA. R. D. Mulroy, Jr, MD, Director, Joint Replacement Unit Waltham Weston Medical Center, Waltham, Massachusetts 02154, USA. Correspondence should be sent to Dr W. H. Harris. 1992 British Editorial Society ofbone and Joint Surgery 030l-620X/92/3339 $2.00 J Bone Joint Surg [Br] 1992; 74-B : 385-9. Dorr, Luckett and Conaty 1990) and for this reason many surgeons changed to cementless femora! stems. The question remained, however, whether the improved cementing technique which had reduced the incidence of loosening in the elderly would be effective in younger patients. We now report the results of improved cementing techniques of 50 hips in 44 patients aged 50 years or younger at an average follow-up of 12 years (10 to 14.8). PATIENTS AND METHODS Between January 1, 1976 and January 1, 1979, 234 cemented total hip replacements were performed on 206 patients by the senior author (WHH), using secondgeneration cementing techniques and stem design. The specific improvements included the use of a methylmethacrylate femora! medullary plug, delivery of a doughy mix of cement (Simplex P ; Howmedica Inc, Rutherford, New Jersey) in retrograde fashion by a cement gun, and the use of a collared cobalt-chrome femoral stem with a rectangular cross-section and rounded corners. The medullary canal was irrigated and packed with sponges until immediately before cementing. Pulsati!e!avage and adrenaline-soaked sponges were not used, nor was a cement pressuriser. A large stem was preferred and the largest rasp which could be comfortably inserted was used to prepare the socket in the upper femur. The rasp was usually about 3 mm oversized compared with the stem. VOL. 74-B, No. 3, MAY 1992 385

386 There were 50 hips in 44 patients (21 women, 23 men) with an average age of 40.9 years (18 to 50) at the time of surgery. The patients average weight was 69 kg (45.5 to 107). The original diagnoses ofthe 50 hips were congenital dislocation and/or dysplasia (19), pistol-grip deformity causing osteoarthritis (5), osteoarthritis which was too far advanced to determine the cause (4), avascular Figure la - A postoperative radiograph shows excellent cement distribution ( white-out ) graded as A. Figure lb - Thirteen years later fixation is secure despite adaptive remodelling of the femur. Figure 2a - A postoperative radiograph shows slight radiolucencies at the cementbone interface, graded B. Figure 2b - Eleven years later the patient had an excellent clinical result and stable radiographic appearance. THE JOURNAL OF BONE AND JOINT SURGERY

IMPROVED CEMENTING AND FEMORAL LOOSENING IN YOUNG PATIENTS WITH HIP ARTHROPLASTY 387 tions were for revision of failed cemented components, two for failed femoral osteotomies, one for a failed cup arthroplasty, and one for a failed double-cup artjiropasty. All patients completed a questionnaire and 13 patients (15 hips) were examined by the authors. Current radiographs were obtained for all. These and the postoperative intermediate radiographs were reviewed for evidence of loosening. Definite loosening was defined as migration, or a change in position of the stem or the cement (Harris, McCarthy and O Neill 1982). This included fracture or bending of the stem, fracture of the cement, the appearance of a radiolucent line at the cement-stem interface not present on the immediate postoperative radiograph, and a shift in the position of the cement mantle relative to the femur. Radiographs that showed a continuous (100%) radiolucent line at the cement-bone interface without evidence of migration were graded as probably loose. If a radiolucent zone was present that was not complete, but involved between 50% and 99% of the interface, the component was classified as possibly loose. The radiographic appearance of the initial cementing was graded on the immediate postoperative radiograph. Complete filling of the medullary cavity by cement, a so-called white-out at the cement-bone interface was graded A (Fig. 1). Slight radiolucency of the cement-bone interface was defined as B (Fig. 2). Radiolucency involving 50% to 99% of the cement-bone interface or a defective or incomplete cement mantle was graded C. Radiolucency at the cement-bone interface of 100% in any projection, or a failure to ifil the canal with cement such that the tip ofthe stem was not covered, was classified D. Fixation of the acetabu!ar component was also assessed. Any cemented acetabular component which had migrated or had a continuous radio!ucent line regardless of its thickness, was diagnosed as definitely loose (Hodgkinson, Shelley and Wroblewski 1988). RESULTS The average duration of follow-up of the 50 hips was 12.0 years (10 to 14.8). No femoral component had been revised for loosening. By contrast, aseptic loosening was the reason for revision of 1 1 acetabular components. At 12 years, the average Harris hip score (Harris 1969) for the 39 hips that had not been revised was 88 points (41 to 100), 24 were rated excellent, six good, six fair, and three poor. Of the three hips with a poor result, one was due to unexplained pain with negative aspiration arthrogram, normal radiographs and a norma! bone scan. In the other two the low score was caused by functional limitations. One patient had recently been operated on for severe spinal stenosis and the other was paraplegic from rheumatoid cervical mye!opathy. Of the six hips rated fair, in three the low score was due to symptomatic aseptic loosening of their acetabular components with migration. One patient was being treated for osteomyelitis of the tibia, unrelated to his hip arthroplasty, with a resultant severe functional limitation which lowered his score. In the 50 current radiographs only one stem was judged to be definitelyboose (Fig. 3): a 0.5 mm radiolucent Figure 3a - The postoperative radiograph shows good cementing. Figure 3b - Twelve years later there is, however, a new radiolucent line at the cement-stem interface in zone one. This stem was classified radiographically as definitely loose. VOL. 74-B, No. 3, MAY 1992

388 R. L. BARRACK, R. D. MULROY JR, W. H. HARRIS line had developed in zone one between the stem and the cement mantle (Fig. 3b). No stem was classified as probably loose but nine hips were possibly loose. Six hips had a small region of focal osteolysis adjacent to the femoral component without evidence ofboosening (Jasty eta! 1986). Femoral cementing was classified on the postoperative radiographs as grade A in 32 cases and grade B in 18. There were no cases in grades C or D. In addition to the 1 1 acetabular components which had undergone revision for aseptic loosening, there were a further 1 1 acetabula classified as definitely loose radiographically. A total of 44% of the acetabular components had, therefore, loosened. During the course of acetabular revision, three of the femoral stems were changed. In each case the stem was secure in the bone, but was removed to alter its position to improve access or, in one case, because of massive bone resorption caused by the granuloma from the loose acetabulum. All three stems had been in situ for more than ten years. DISCUSSION In this group ofpatients, aged 50 years or less undergoing cemented total hip replacement by second-generation femoral cementing techniques, there had been no femoral stem revisions by 12 years and only one femora! component was loose on radiographic criteria. This is a striking change from the results reported using firstgeneration cementing techniques, particularly as five of the 50 arthroplasties were revisions for failed cemented implants. Chandler et al(l98l) reported revised or radiographically loose components in 57% of 29 patients aged 30 years or less. The incidence of radiolucent lines was 94% on the acetabular side and 33% on the femoral side. The numberofrevisions for acetabu!ar and femoral loosening, however, were equal (five of each in seven patients). Component migration or ominous radiolucency (progressive and more than 2.5 mm thick), was seen in nine acetabula, one femur, and in one case involved both components (Chandler et a! 1981). Dorr et a! (1983) described 81 patients aged 45 years or less and followed up for an average of4.5 years. There was radiographic evidence of impending failure in 29% and most of the problems were on the acetabular side where progressive radio!ucency was seen in 83% as against 42% in the femoral stem. Even in those hips in which the cementing technique wasjudged to be adequate (40%), there was progressive demarcation around 95% of the acetabula and 53% of the femoral stems. The cement technique was technically unsatisfactory in 60% of cases by the authors criteria. In a series of cemented total hip replacements in patients 49 years old or less using first-generation methods, Collis (1991) found at 14.9 years that 20% of the femora! stems had been revised and another 8% were loose radiographically. Cornell and Ranawat (1986), in their series of cemented total hips in patients under 55 years of age, found that by 13 years, survivorship of prostheses had decreased to 70%, mainly due to acetabular component failure. The predicted rate of success for the femoral stems was 93%. No stem had been revised for aseptic loosening but two acetabular components had been revised. One other acetabularcomponent and one femora! stem were radiographically loose. The results ofour series are similar. We had 22% acetabular revisions and a further 22% were radiographically loose ; on the femoral side the corresponding figures were 0% and 2%. Based on the poor results of some early series, many authors have advised against the use of cemented stems in young patients (Aronson 1986 ; Dorr et a! 1990) and many surgeons have changed to cementless femoral implants. Engh and Glassman (1990), however, found that there was a 2.6% failure rate at an average follow-up of only 4.75 years in 303 AML (DePuy Inc, Warsaw, Indiana) cementless femoral components in patients less than 50 years old. The use ofa PCA stem (Howmedica Inc, Rutherford, New Jersey) in patients aged 50 years or less, recently reported by Davies and Hedley (1991), resulted in 2% femora! revisions and 6% radiographically unstable femora! components in a series of 48 hips followed for only 38 months. We attribute our reduced loosening rate to the introduction of improved cementing techniques and better stem design. The cement gun allows more complete filling of the medullary canal, and reduces the voids and laminations in the cement mantle. The use of a methylmethacrylate medullary plug not only allows greater intrusion pressure of cement (Oh et a! 1978) with better filling and improved cement-bone interface strength, but also automatically extends the cement mantle 2 to 4 cm beyond the tip of the prosthesis. Beckenbaugh and Ilstrup (1978) noted that this was an important factor for improved fixation. The clear evidence for the improvement in technique is that we had no hips with grade C or D cementing. Wrob!ewski (1990) has recently reported a reduction of stem revision for loosening from 6% to 0.9% since he added an intramedullary bone plug to his cementing technique. Advances in cementing technique have greatly improved thelong-term survival offemora! stem implants in young patients. For cementless methods of fixation to compete they must now match these results, not those of first-generation cementing methods. This study was supported by the William H. Harris Foundation. One or more of the authors have received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this article. In addition benefits have also been or will be directed to a research fund, foundation, educational institution, or other non-profit institution with which one or more of the authors is associated. THE JOURNAL OF BONE AND JOINT SURGERY

IMPROVED CEMENTING AND FEMORAL LOOSENING IN YOUNG PATIENTS WITH HIP ARTHROPLASTY 389 REFERENCES Amstutz HC, MarkolfKL, McNeice GM, Gruen TA. Loosening of total hip components : cause and prevention. In : The Hip. Proc 4th meeting of The Hip Society. St Louis, etc : CV Mosby Co, 1976: 102-16. Aronson J. Osteoarthritis ofthe young adult hip : etiology and treatment. Instructional Course Lectures. AAOS 1986; 36:119-28. Beckenbaugh RD, llstrup DM. Total hip arthroplasty: a review of three hundred and thirty-three cases with long follow-up. J Bone Joint Surg[Am] 1978; 60-A :306-13. Chandler HP, Reineck Ff, Wixson RL, McCarthy JC. Total hip replacement in patients younger than thirty years old : a five-year follow-up study. J Bone Joint Surg [Am] 1981 ; 63-A :1426-34. Cohn DK. Long-term (twelve to eighteen-year) follow-up of cemented total hip replacements in patients who were less than fifty years old : a follow-up note. J Bone Joint Surg [Am] 1991 ; 73-A :593-7. Cornell CN, Ranawat S. Survivorship analysis of total hip replacements : results in a series of active patients who were less than fiftyfive years old. J Bone Joint Surg [Am] 1986; 68-A :1430-4. Davies JF, Hedley AK. Uncemented total hip replacement in the young patient. Trans AAOS, March, 1991 :147. Dorr LD, Luckett M, Conaty JP. Total hip arthroplasties in patients younger than 45 years : a nine to ten year follow-up study. Clin Orthop 1990; 260 :215-9. DolT LD, TakelGK, Conaty JP. Total hip arthroplasties in patients less than forty-five years old. J Bone Joint Surg [Am] 1983; 65-A: 474-9. Engh CA, Glassman AH. The influence of femoral component porous coating level in cementless total hip arthroplasty. 20th Annual Hip Course, October 1990. Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures : treatment by mold arthroplasty. J Bone Joint Surg[Am] 1969; 51-A :737-55. Harris WH, McCarthy JC Jr, O Neill DA. Femoral component loosening using contemporary techniques of femoral cement fixation. J Bone Joint Surg [Am] 1982; 64-A :1063-7. Harris WH, McGann WA. Loosening of the femoral component after use of the medullary-plug cementing technique. J Bone Joint Surg [Am] 1986; 68-A:1064-6. HOdgkIJISOII JP, Sbelley P, Wroblewski BM. The correlation between the roentgenographic appearance and operative findings at the bone-cement junction of the socket in Charnley low friction arthroplasties. C/in Orthop 1988 ; 228:105-9. Jasty MJ, Floyd WE III, Schiller AL, Goidring SR, Harris Wil. Localized osteolysis in stable, non-septic total hip replacement. J BoneJoint Surg [Am] 1986; 68-A :912-9. Mulroy RD Jr, Harris WH. The effect of improved cementing techniques on component loosening in total hip replacement. J Bone Joint Surg [Br] 1990; 72-B :757-60. Oh I, Carlson CE, Tomford WW, Harris WH. Improved fixation of the femoral component after total hip replacement using a methacrylate intramedullary plug. J Bone Joint Surg [Am] 1978 ; 60-A: 608-1 3. Roberts DW, Poss R, Kelley K. Radiographic comparison of cementing techniques in total hip arthroplasty. JArthrop/asty 1986; 1 :241-7. Russotti GM, Coventry MB, Stauffer RN. Cemented total hip arthroplasty with contemporary techniques : a five-year minimum follow-up study. Clin Orthop 1988 ; 235:141-7. Stauffer RN. Ten-year follow-up study oftotal hip replacement. J Bone JointSurg[Am] 1982; 64-A :983-90. Sutherland CJ, Wilde AR, Borden IS, Marks KE. A ten-year follow-up of one hundred consecutive Muller curved-stem total hip-replacement arthroplasties. J Bone Joint Surg [Am] 1982; 64-A :970-82. Wroblewski BM. Resultsofimproved cement distribution. 20th Annual Hip Course, October 1990. VOL. 74-B, No. 3, MAY 1992