OSS Orthopedic Salvage System Compress Device. Surgical Technique

Similar documents
Arcos Interlocking Distal Stem. Surgical Technique Addendum to the Arcos Modular Femoral Revision System

Arcos Modular Femoral Revision System. Broach and Calcar Proximal Bodies Surgical Technique

OSS Modular Arthrodesis System. Assembly Guide

Tibial Fixation. with TunneLoc Device. Surgical Technique by Mark J. Albritton, M.D. and Sherwin Ho, M.D.

JuggerLoc Bone-to-Bone System for Ankle Syndesmosis Fixation. Surgical Technique

Resurfacing Distal Femur. Orthopaedic Salvage System

MIS Cemented Tibial Component

FastFrame External Fixation System. Damage Control Surgical Technique

The AperFix II System

Correction System. Surgical Technique

ZMR Over-the-Junction Instruments for Revision Hip Arthroplasty. Surgical Technique IMAGE TO COME

Zimmer NexGen MIS Tibial Component. Cemented Surgical Technique IMAGE TO COME

3.0 mm Cannulated Screw System. Surgical Technique

N-Force Fixation System. Surgical Technique

Zimmer Trabecular Metal Ankle Interpositional Spacer and Trabecular Metal Ankle Fusion Spacer

Trabecular Metal Primary Patella

Taperloc Complete Hip System. Surgical Technique

Correction System. Surgical Technique

Zimmer ITST Intertrochanteric/ Subtrochanteric Fixation System. Abbreviated Surgical Technique

NexGen Trabecular Metal Tibial Tray. Surgical Technique

A novel cementless option. Zimmer NexGen Trabecular Metal Primary Patella Surgical Technique

OPTIONAL/ADDITIONAL INSTRUMENTS

MSP TM Metatarsal Shortening System. Surgical Technique

operative technique Kent Hip

U2 PSA. Revision Knee. Surgical Protocol

Orthopedic Bone Nail System - Distal Femoral Nail Surgical Technique Manual

PRIMARY POROUS PATELLA WITH TRABECULAR METAL. Surgical Technique

3. PATIENT POSITIONING & FRACTURE REDUCTION 3 8. DISTAL GUIDED LOCKING FOR PROXIMAL NAIL PROXIMAL LOCKING FOR LONG NAIL 13

Pre-Operative Planning. Positioning of the Patient

Valencia Pedicle Screw Surgical Technique

Constrained Posterior Stabilized (CPS)

Constrained Posterior Stabilized (CPS) Surgical Technique

System. Humeral Nail. Surgical Technique

Zimmer NexGen Trabecular Metal Tibial Tray

Tibial Replacements. Orthopaedic Salvage System

Trilogy Acetabular System

SureLock All-Suture Anchor System

Trilogy Acetabular System

EXACTECH KNEE. Operative Technique ADDENDUM. Metaphyseal Cones. Surgeon focused. Patient driven. TM

VerSys LD/Fx Cemented and Press-Fit Hip Prostheses. Surgical Technique IMAGE TO COME. Versatile solutions for total and partial hip replacement

LCP Distal Tibia Plate

S U R G I C A L T E C H N I Q U E David A. McQueen, MD Return to Menu

Clinical Evaluation Surgical Technique

AcUMEDr. FoREARM ROD SYSTEM

TRK REVISION KNEE Surgical Technique

InCoreTM Lapidus System

Wichita Fusion Nail Surgical Technique. David A. McQueen, MD

Surgical Technique.

FLH /11

ComposiTCP Anchor with BroadBand Tape

Distal Cut First Femoral Preparation

CrossFix II. All-Suture, All-Inside Meniscal Repair System. Surgical Technique

Zimmer Natural Nail System

BTB ACL Reconstruction with the ToggleLoc Fixation Device with ZipLoop Technology. Surgical Technique by James R. Andrews, M.D.

StageOne. Shoulder Cement Spacer Molds. Surgical Technique

Femoral / Tibial Augmentation. Orthopaedic Salvage System

Scan Bi-Polar 22/28. Operative Technique

JuggerKnot Soft Anchor 1.0 mm Mini

Trilogy Acetabular System

Surgical Technique Guide

Peanut Growth Control Plating System

NeoGen Tibia Nail System

Surgical Technique. Unisyn Hip System

TOTAL KNEE ARTHROPLASTY SYSTEM

3.5 mm Locking Attachment Plate

Persona. The Personalized Knee. Trabecular Metal Tibia. Surgical Technique

Femoral Fixation for ACL Reconstruction. Surgical Protocol by Mark Gittins, D.O.

JuggerKnot Soft Anchor 1.0 mm Mini

Signature Personalized Patient Care

Biomet Large Cannulated Screw System

Comprehensive Reverse Shoulder System Augmented Baseplate

Surgical Technique Guide

Technique Guide. *smith&nephew N8TIVE ACL Anatomic ACL Reconstruction System

SURGICAL TECHNIQUE GUIDE

External Skeletal Fixation (ESF)

NCB Distal Femur System. Surgical Technique

Approach Patients with Confidence

QUICK REFERENCE GUIDE. The XCaliber Meta-Diaphyseal Fixator

Correction System. Surgical Technique

QUICK REFERENCE GUIDE. The PreFix Fixator (92000 Series) ALWAYS INNOVATING

Biologically-Assisted ACL Reconstruction. Surgical Technique

NeoGen Femoral Nail System

HP KNEE EXTRACTION Instrumentation. Product Overview

3. Insert Tocar Sleeves Insert the NCB tissue protection sleeve assembly 1.6 to 10mm through a skin incision (Fig. 38).

BAK/C Cervical Anterior Interbody Fusion System

Phoenix Tibial Nail System

AxSOS Locking Plate System

OptiLock Periarticular Plating System For Distal Tibial Fractures. Surgical Technique

ZMR Porous Revision Hip Prosthesis. Surgical Technique IMAGE TO COME

Conventus CAGE PH Surgical Techniques

Biomet Multi-Axial Correction (MAC) System Correction Atlas

ANATOMIC SURGICAL TECHNIQUE. 5 in 1. Conventional instrumentation 07/11/2013

Monolateral External Fixation System for Trauma and Orthopaedics

Surgical Technique. Targeter Systems Overview

Ease of Use. Have the instrument you need when you need it.

PAL Pelvic Alignment Level

CANNULINK. Intraossous Fixation System SURGICAL TECHNIQUE

Surgical Technique. VISIONAIRE Disposable Instruments for the LEGION Total Knee System

Bone Preservation Stem

Transcription:

OSS Orthopedic Salvage System Compress Device Surgical Technique

Table of Contents Quick Reference Guide... 2 Device Description... 3 Resection... 4 No Face Adapter Provisional Construct... 4 Canal Preparation... 5 Anchor Plug... 9 Instrumentation Assembly... 9 Instrumentation Assembly Alignment... 13 Anchor Plug Implantation... 15 Transverse Pins... 17 Transverse Pin Measurement... 17 Transverse Pin Selection... 20 Transverse Pin Insertion... 23 Spindle... 28 Diameter Selection... 28 Face Reaming Assembly... 31 Face Reaming... 33 Centering Sleeve/Spindle Assembly... 35 Spindle Implantation... 37 Standard Distal Femoral Trialing... 40 Reduced Resection Distal Femoral Trialing... 40 Realignment of Spindle... 42 Pin Drilling and Insertion... 42 Final Construct Assembly... 45 Without Locking Cap and Screw... 45 Optional Locking Cap and Screw Assembly... 46 Screw Placement Guide... 48 Utilizing Reduced Resection for a Segmental Distal Femur Replacement... 48 Utilizing Taper Adapter for a Distal Femoral Replacement... 49 Utilizing Diaphyseal Segment and Taper Adapter for a Distal Femoral Replacement... 50 Utilizing Reduced Resection for a Proximal Femur Replacement... 51 Utilizing Taper Adapter for a Proximal Femoral Replacement... 52 Utilizing Short Spindle for a Proximal Femoral Replacement... 53 Utilizing Short Spindle for a Low Profile Proximal Femur Replacement... 54 Tibial Preparation... 55 The Compress Device utilizes the OSS Orthopedic Salvage System.

2 Compress Device Surgical Technique Quick Reference Guide Compress Resection Chart Total Construct Length Distal or Proximal Femoral Component (cm) Standard, RS or Reduced Resection Segment(s) 3 cm 4 cm 5cm 7cm 9 cm 11 cm 13 cm 15 cm 17 cm Taper Adapter 5 cm Spindle Collar 1 cm 30 7 17 5 1 29.5 8.5 15 5 1 29 7 7 9 5 1 28.5 8.5 (7+7) 5 1 28 7 15 5 1 27.5 8.5 13 5 1 27 7 (7+7) 5 1 26.5 8.5 5 7 5 1 26 7 13 5 1 25.5 8.5 11 5 1 25 7 5 7 5 1 24.5 8.5 (5+5) 5 1 24 7 11 5 1 23.5 8.5 9 5 1 23 7 (5+5) 5 1 22.5 8.5 (4+4) 5 1 22 7 9 5 1 21.5 8.5 7 5 1 21 7 (4+4) 5 1 20.5 8.5 (3+3) 5 1 20 7 7 5 1 19.5 8.5 5 5 1 19 7 (3+3) 5 1 18.5 8.5 4 5 1 18 7 5 5 1 17.5 8.5 3 5 1 17 7 4 5 1 16 7 3 5 1 14.5 8.5 5 1 13 7 5 1 11 10 Reduced Resection 1 9.5 8.5 Reduced Resection 1 8 7 Reduced Resection 1

3 Compress Device Surgical Technique Immediate Postoperative 12 Years Postoperative Short Compress Spindle Standard Compress Spindle Figure 1 Figure 2 Device Description The Compress Device was developed as an alternative method for fixating a segmental construct to host bone, while creating a stable bone-prosthesis interface. Unlike a stemmed construct, the Compress Device was designed to prevent bone resorption by allowing stress sharing via the implant and bone while providing a motion free environment suitable for biologic fixation (Figure 1). The system is available in two different anchor plug/spindle options: the short Compress Device requiring 45 mm of medullary placement (particularly advantageous for bone loss) and the standard Compress Device requiring 80 mm of medullary placement (Figure 2). The following surgical technique will focus on the standard 80 mm anchor plug/spindle offering. Note: Please reference the IFU for indications and contraindications relating to the Compress Device and OSS implants. X-rays compliments of: James O. Johnston, M.D. Kaiser Permanente Medical Center Department of Orthopedic Surgery South San Francisco, CA. Note: While the 9 and 13 hole anchor plugs exist, this technique focuses on 5 hole anchor plugs. If using the 9 or 13 hole anchor plug, additional instrumentation may be required.

4 Compress Device Surgical Technique For use with the Standard or Short Implants For use with the Reduced Resection Implants Part Number 32-481114 Figure 3 Figure 4 Figure 5 Resection Utilize the Compress Resection Chart on page 2 to determine the correct implant components needed to recreate the residual femur. To achieve a proximal or distal femoral resection, begin by making a resection mark and a rotational alignment mark with a cautery device or methylene blue (Figure 3). Using a freehand transverse cut, resect the femur at the reference resection mark. No Face Adapter Provisional Construct Following the transverse resection, it is recommended that the no face adapter provisional be attached to segmental provisionals in order to ascertain some semblance of the construct length (Figures 4 and 5). This is suggested at the onset of the surgery because once the holes for the transverse pins are drilled and the anchor plug is secure within the canal, no additional bone may be resected without dramatically altering the original replacement length. Note: If preparing the bone for use with the Compress reduced resection implants, utilize the no face adapter provisional with the threaded connection rather than the quick connect (32-481114).

5 Compress Device Surgical Technique Figure 6 Figure 7 Canal Preparation After a starter reamer has opened the canal, prepare the femoral canal for implant insertion by using the Compress triple reamers (Figure 6). When used in conjunction with the modular planers, the triple reamers are preset to the desired reaming depth and prepare for both the anchor plug and centering sleeve diameters (Figure 7). This allows the femoral osteotomy site to be perpendicular to the longitudinal axis of the femoral shaft. Note: The triple reamer tips protrude past the medullary placement to a depth of 65 mm on the short reamers and 100 mm on the standard reamers.

6 Compress Device Surgical Technique Figure 8 Canal Preparation (cont.) Sequentially ream with gentle force against the femoral osteotomy site until full circumferential cortical contact at the osteotomy site and centering sleeve location is achieved (Figure 8).

7 Compress Device Surgical Technique Anchor Plug A Diameter Centering Sleeve B Diameter Figure 9 Canal Preparation (cont.) The A diameter of the final triple reamer will determine the anchor plug size and the B diameter will determine the centering sleeve size (Figure 9).

8 Compress Device Surgical Technique 20 mm Figure 10 Canal Preparation (cont.) Note: If the triple reamer provides cortical contact for the anchor plug, but not the centering sleeve, reintroduce sequential triple reamers to a depth of 20 mm at the resection level until cortical contact is achieved (Figure 10). Refer to the Triple Reamer Table for the appropriate anchor plug and centering sleeve diameters. Triple Reamer Table Size A (mm) Diameters B (mm) Use Anchor Plug Diameter Use Centering Sleeve Diameter 10 12 10 12 10 * 11 13 11 13 10 13 12 14 12 14 12 14 13 15 13 15 12 15 14 16 14 16 14 16 15 17 15 17 14 17 16 18 16 18 16 18 17-19 17 19 16 19 18 20 18 20 18 20 19 21 19 21 18 21 20 22 20 22 20 22 21 23 21 23 20 23 22 24 22 24 22 24 *The diameter of the spindle shaft is 12 mm, no centering sleeve necessary

9 Compress Device Surgical Technique A 13 mm only B Figure 11 Figure 12 Anchor Plug Instrumentation Assembly Important: There are two distinct anchor plug holders in the Compress instrumentation. The first (Figure 11A) is used without a centering sleeve provisional when the final reamer at the resection level is 12 mm. It is also used with centering sleeve provisionals when the final reamer measures 14 mm through 28 mm. The second anchor plug holder (Figure 11B) is used only when the final reamer at the resection level measures 13 mm. This 13 mm anchor plug holder is used without a centering sleeve provisional. Slide the anchor plug holder through the cross bar (Figure 12).

10 Compress Device Surgical Technique Figure 13 Figure 14 Anchor Plug (cont.) Slide the centering sleeve provisional (based on the final reamer diameter B in the Canal Preparation step) onto the anchor plug holder. Be sure to align the exposed metal nub inside the provisional with the female keyway located on the holder. When the provisional meets the cross bar, lock into place with a clockwise turning motion (Figures 13 and 14). Confirm that the centering sleeve provisional fits snugly within the canal at the osteotomy site.

11 Compress Device Surgical Technique Figure 15 Figure 16 Anchor Plug (cont.) Slide the anchor plug implant (based on the final reamer diameter A in the Canal Preparation step) into the anchor plug holder. Be sure the male tab of the holder fits into the female tab of the anchor plug head (Figure 15). Tighten into position with a clockwise motion of the anchor plug holder knob (Figure 16).

12 Compress Device Surgical Technique The small drill guide is used with the 10 mm-12 mm anchor plugs. The large drill guide is used with 14 mm-24 mm anchor plugs. Figure 17 Figure 18 Anchor Plug (cont.) Choose the appropriate drill guide to match the selected anchor plug (Figure 17). Place the cross bar through the drill guide (Figure 18).

13 Compress Device Surgical Technique Figure 19 Figure 20 Anchor Plug (cont.) Instrumentation Assembly Alignment Note: All three drill bits and the pin inserter are packaged with each anchor plug. Place the other short drill bit through the far upper left hole of both the anchor plug and drill guide (Figure 20). Align the drill guide with the anchor plug by placing one of the short drill bits through the far upper right hole of both the anchor plug and the drill guide (Figure 19).

14 Compress Device Surgical Technique Figure 22 Figure 21 Figure 23 Anchor Plug (cont.) The long drill bit is positioned through either of the two lower holes in the anchor plug and drill guide (Figure 21). Once the drill bits are in place, minimize the lateral distance between the drill guide and the anchor plug to optimize targeting once the anchor plug is inserted into the bone (Figure 22). Alternate tightening of both adjustment knobs located on top with a clockwise motion until snug (Figure 23).

15 Compress Device Surgical Technique Firm Contact Figure 24 Figure 25 Anchor Plug (cont.) When the knobs are tight, all three drill bits should slide freely through the alignment holes (Figure 24). Note: If the drill bits do not slide freely, loosen the top two adjustment knobs and begin the tightening process again. Anchor Plug Implantation With the drill guide properly aligned with the anchor plug, remove the three drill bits and position the anchor plug into the femoral canal. Make certain that the cross bar is in firm and even contact with the femoral osteotomy site and that the centering sleeve provisional is snug within the canal (Figure 25). Placement of the anchor plug within the medullary canal is non-directional; location is typically dictated by the best exposure and ease of access to the drill guide.

16 Compress Device Surgical Technique Hole #1 Hole #5 Figure 26 Figure 27 Hold the cross bar and drill guide securely and, without power, run a short drill bit through hole #1 of the drill guide until contacting the bone. At full power drill through both cortices (Figure 26). Leave the short drill bit in place. Repeat this procedure with the second short drill bit through hole #5, leaving the short drill bit in place (Figure 27).

17 Compress Device Surgical Technique Figure 28 Figure 29 Anchor Plug (cont.) Using the long drill bit, sequentially drill each of the remaining three placement holes. Remove the long drill bit when finished (Figure 28). Transverse Pins Transverse Pin Measurement The hook depth gauge is used to measure the bicortical distance for determining the transverse pin length (Figure 29). Start by loosening knobs 1 and 2 and collapsing the sleeves together.

18 Compress Device Surgical Technique Figure 30 Figure 31 Transverse Pins (cont.) Tighten knob 2 (Figure 30). Push the hook through a pre-drilled hole and engage the outer aspect of the far cortex (Figure 31).

19 Compress Device Surgical Technique Figure 32 Figure 33 Transverse Pins (cont.) Collapse the sleeves against the drill guide and tighten knob 1 (Figure 32). Release the hook from the far cortex, and position the tip firmly against the outer aspect of the near cortex (Figure 33).

20 Compress Device Surgical Technique Figure 34 Figure 35 Transverse Pins (cont.) Loosen knob 2, slide the sleeve against the drill guide and tighten knob 2 (Figure 34). Transverse Pin Selection Remove the hook depth gauge from the drill guide and read the bi-cortical distance that appears on the sliding scale (Figure 35).

21 Compress Device Surgical Technique 20 mm 24 mm 28 mm 32 mm 36 mm 40 mm 44 mm 48 mm 52 mm 56 mm 60 mm 64 mm 68 mm 72 mm 76 mm Figure 36 Figure 37 Transverse Pins (cont.) Based on the depth measurement shown on the sliding scale, add an additional 4 mm to the measurement, and select the transverse pin length that will provide the necessary 2 mm overhang on each side of the cortex (Figure 36). Note: Transverse pins are available from 20-76 mm, in 4 mm increments (Figure 37).

22 Compress Device Surgical Technique A B Figure 38 Figure 39 Transverse Pins (cont.) The pin inserter handle has two components: the impaction head (Figure 38A) and the adjustment knob (Figure 38B). Place the threaded end of the pin inserter into the pin inserter handle and tighten with a clockwise turn (Figure 39). Note: Ensure the adjustment knob is loose by turning in a counterclockwise motion; this will ensure that the handle will accept the pin inserter.

23 Compress Device Surgical Technique Figure 40 Figure 41 Transverse Pins (cont.) Tighten the adjustment knob with a clockwise turn until the pin inserter is secure (Figure 40). Transverse Pin Insertion Place a transverse pin onto the pin inserter (Figure 41).

24 Compress Device Surgical Technique Figure 42 Figure 43 Transverse Pins (cont.) Guide the pin inserter into one of the three available holes in the drill guide (Figure 42). Gently tap the impaction head until the pin is properly seated. The pin is properly seated when the tabs on the pin inserter come into contact with the cortex (Figure 43) and a 2 mm overhang is visible on both the inner/outer cortex.

25 Compress Device Surgical Technique Figure 44 Figure 45 Transverse Pins (cont.) Pull to remove the pin inserter (Figure 44) and proceed with seating the next two transverse pins (Figure 45).

26 Compress Device Surgical Technique Figure 46 Figure 47 Transverse Pins (cont.) With the first three transverse pins in place, remove one of the short drill bits and seat the fourth transverse pin (Figure 46). Remove the final short drill bit and seat the fifth and final transverse pin (Figure 47). Note: The transverse pins come packaged with one additional pin that is not required for implantation. The anchor plug only accepts a total of five pins.

27 Compress Device Surgical Technique Figure 48 Figure 49 Transverse Pins (cont.) Loosen the anchor plug holder knob (Figure 48) and remove the drill guide assembly (Figure 49).

28 Compress Device Surgical Technique A B Figure 50 Figure 51 Transverse Pins (cont.) Remove the centering sleeve provisional from the anchor plug holder (Figure 50). Set the provisional aside, as it will be used later on in the technique. Spindle Diameter Selection Measure the maximum diameter of the resected bone with a ruler (Figure 51A) or the spindle sizer (Figure 51B). Note: The ruler may be used to measure for all four spindle sizes (30 mm, 38 mm, 44 mm and 37 x 49 mm elliptical). The spindle sizer measures for the 38 mm spindle (small) and the 44 mm spindle (large).

29 Compress Device Surgical Technique Figure 53 A B C D Figure 52 Figure 54 Spindle (cont.) Place the spindle sizer over the exposed anchor plug traction bar and select the spindle that comes closest to one of the inscribed markings (Figure 52). The selected implant must be closest to, but larger than, the bone measured in this step (in this instance, a 38 mm spindle would be selected). The implant collar of the spindle should overhang the bone (Figure 53). Select a spindle based upon the prepared diameter of the osteotomy site. 30 mm Spindle: Anatomies up to 26 mm in diameter (Figure 54A). 38 mm Spindle: Anatomies up to 34 mm in diameter (Figure 54B). 44 mm Spindle: Anatomies up to 40 mm in diameter (Figure 54C). 49 mm x 37 mm Elliptical Spindle: Anatomies in the metaphyseal flare where the bicortal measurement does not exceed 45 mm (Figure 54D).

30 Compress Device Surgical Technique Figure 55 Figure 56 Spindle (cont.) Measure the cortical thickness along the perimeter of the resection level using a ruler, calipers or the cortical gauge of the spindle sizer (Figure 55). Based on the measurement, reference the Compress Device Force Table to select the appropriate force level that corresponds to the measured thickness. Note: If using the cortical gauge of the spindle sizer, place the gauge onto the thinnest portion of the cortical wall to determine the appropriate force level (400 lb/600 lb/800 lb) (Figure 56). Compress Device Force Table for Standard Compress Cortical* Thickness (mm) Force Level (lb) 0.0 2.4 Not Indicated* 2.5 3.9 400 4.0 5.4 600 5.5 and above 800 *Cortical thickness less than 2.5 mm is a contraindication for this system. Note: Cortical thickness should not exceed range anywhere between the anchor plug and the interface. Note: Any cortical defect between the anchor plug and interface is a contraindication for this system.

31 Compress Device Surgical Technique A B Figure 57 Figure 58 Spindle (cont.) Face Reaming Assembly Select the appropriate face reamer blade based on the spindle diameter that was chosen in the previous step (Figure 57). The small face reamer blade addresses the 30 mm and 38 mm spindles (Figure 58A). The large face reamer blade is used for the 44 mm and elliptical 49 mm x 37 mm spindles (Figure 58B).

32 Compress Device Surgical Technique A B Figure 59 Figure 60 Spindle (cont.) Important: There are two distinct reamer pilots in the Compress instrumentation. The first (Figure 59A) is used without a centering sleeve provisional when the final triple reamer diameter at the resection level measures 12 mm. It is also used with centering sleeve provisionals when the final triple reamer diameter measures 14 mm through 24 mm. Place the face reamer blade over the two retaining tabs located on the reamer body and secure with a clockwise turning motion. Insert the reamer pilot into the face reamer and tighten the side set screw (Figure 60). Important: The 13 mm reamer pilot (Figure 59B) is used only when the final triple reamer diameter at the resection level measures 13 mm. The reamer pilot is used without a centering sleeve provisional.

33 Compress Device Surgical Technique Figure 61 Figure 62 Spindle (cont.) Slide the centering sleeve provisional that was previously used on the anchor plug holder and lock it into place with a clockwise turning motion (Figure 61). Face Reaming Slide the face reamer assembly over the exposed anchor plug traction bar and ream the osteotomy site (Figure 62). Important: Start power and run at full speed prior to the face reamer contacting the bone. Ream to the mechanical stop.

34 Compress Device Surgical Technique Figure 63 Figure 64 Spindle (cont.) The bone is contoured to match the spindle and is now ready to accept the face adapter provisional if required (Figure 63). If trialing is desired, insert the reamer pilot into the end of the face adapter provisional (Figure 64). Place the reamer pilot over the exposed traction bar and assemble the necessary provisionals. Once trialing is complete, remove provisionals.

35 Compress Device Surgical Technique Figure 65 Figure 66 Centering Sleeve/Spindle Assembly Important: All of the spindles have a 12 mm diameter designed into the standard implant. The centering sleeve implants are added to increase this diameter as needed. Place the centering sleeve implant onto the spindle (Figure 65). Impact the centering sleeve with the sleeve impactor (Figure 66). Note: The centering sleeve implant should be sized accordingly to fit snug within the canal.

36 Compress Device Surgical Technique Figure 67 Figure 68 Figure 69 Centering Sleeve/Spindle (cont.) Slide the spindle over the anchor plug traction bar until it fits snugly against the prepared femoral osteotomy site. Note: The centering sleeve must be completely seated within the prepared canal at the boneimplant interface. Position the anti-torque wrench into two of the female spindle slots and tighten the locking knob (Figures 68 and 69). The spindle should be aligned anteriorly so that the rotation alignment mark and one of the female spindle slots lines up with the vertical rotational alignment mark made earlier (Figure 67).

37 Compress Device Surgical Technique Figure 70 Figure 71 Spindle Implantation Place the nut into the nut driver and insert onto the exposed anchor plug traction bar in the spindle (Figure 70 and 71).

38 Compress Device Surgical Technique Figure 72 Figure 73 Spindle Implantation (cont.) Tighten in a clockwise direction with the nut driver, while holding the anti-torque wrench in place until resistance is felt (this is the initial contact of the nut with the washers of the spindle). Continue to slowly turn the nut driver in a clockwise direction until the compression cap can be manually released or until an additional one-half turn (180 degrees) is achieved, then stop turning. The compression cap should release by hand in a counterclockwise motion (Figures 72 and 73).

39 Compress Device Surgical Technique Figure 74 Figure 75 Spindle Implantation (cont.) Note: If the compression cap cannot be removed manually after one 180 degree turn, remove the nut driver and place the spanner wrench (Figure 74) into the surface holes of the compression cap and twist off in a counterclockwise direction (Figure 75). Important: After the compression cap is removed from the spindle, DO NOT DISCARD. Place the compression cap in reserve on the back table until the surgery is completed.

40 Compress Device Surgical Technique Figure 76 Figure 77 Figure 78 Spindle Implantation (cont.) Standard Distal Femoral Trialing Place the 5 cm taper adapter provisional over the spindle taper and assemble the additional provisionals to achieve the correct construct length (Figures 76 and 77). Reduced Resection Distal Femoral Trialing When trialing for the Compress 7 cm, 8.5 cm or 10 cm reduced resection implants (proximal or distal) (Figure 78), it is necessary to place the Compress 5 cm taper adapter provisional with the threaded connection (32-481075) over the exposed spindle taper (Figure 79). Place the Compress reduced resection provisional over the no face adapter provisional until the magnets inside the blue provisional engage with the taper adapter provisional (Figure 80).

Figure 79 Figure 80 41 Compress Device Surgical Technique

42 Compress Device Surgical Technique Figure 81 Figure 82 Figure 83 Spindle Implantation (cont.) Realignment of Spindle If the spindle alignment needs to be changed, disassociate the provisional and replace the compression cap with firm clockwise manual rotation (Figure 81 and 82). Reintroduce the nut driver and in a counterclockwise direction and make one full turn (360 degrees) to loosen the nut. Change the rotation of the spindle (Figure 83) and re-start the locking process (beginning with Figure 71). Pin Drilling and Insertion The Pin Placement Table specifies which array of holes may be used based on the centering sleeve that was previously selected (Figure 84). Pin Placement Table Spindle Size 30 mm Inner Array 13-14 mm Sleeve Middle Array Outer Array 38 mm 13 mm Sleeve 13-18 mm Sleeve 13-22 mm Sleeve 44 mm 13 mm Sleeve 13-20 mm Sleeve 13-27 mm Sleeve Elliptical (37x49 mm) 13-15 mm Sleeve 13-18 mm Sleeve 13-28 mm Sleeve

43 Compress Device Surgical Technique Figure 84 Figure 85 Figure 86 Figure 87 Spindle Implantation (cont.) Thread the anti-rotation drill guides into the selected holes of the spindle collar until fully seated (Figure 85). Insert the specially designed drill into each respective drill guide. Start at full power prior to contact and drill until the guides prevent further advancement. Drill to the mechanical stop (Figures 86 and 87).

44 Compress Device Surgical Technique Figure 88 Figure 89 Spindle Implantation (cont.) Remove the first drill guide, leaving the others in place (Figure 88). Using the pin driver, insert a pin into the first predrilled location and thread the screw until it is fully seated (Figure 89). Repeat for remaining pin locations. Note: It is recommended, whenever possible, that a minimum of 3 pins are placed in the spindle collar.

45 Compress Device Surgical Technique Figure 90 Final Construct Assembly Without Locking Cap and Screw After the anti-rotation pins are in place, impact the femoral construct (taper adapter, diaphyseal segment and distal femoral construct) together on the back table. Once the femoral construct is assembled, impact over the spindle taper (Figure 90). Refer to the Screw Placement Guide for proper placement beginning on page 48.

46 Compress Device Surgical Technique Figure 92 Figure 91 Figure 93 Final Construct Assembly (cont.) Optional Locking Cap and Screw A locking cap and screw are available to mechanically secure the taper adapter or reduced resection femoral components to the spindle. Thread the locking cap into the spindle taper using the nut driver to firmly seat the locking cap (Figures 92 and 93). With the spindle in place, unscrew the locking screw from the locking cap and set the screw aside (Figure 91).

47 Compress Device Surgical Technique Figure 94 Figure 95 Final Construct Assembly (cont.) Impact the 5 cm taper adapter over the spindle taper and secure the construct with the locking cap screw (Figure 94). After the locking cap and screw have been assembled, impact each implant component one at a time, and secure with the corresponding screw (Figure 95). Refer to the Screw Placement Guide for proper placement beginning on page 48.

48 Compress Device Surgical Technique Screw Placement Guide Utilizing Reduced Resection Segmental Distal Femur Replacement Compress Spindle (1 cm) Compress Locking Cap (optional) 1 cm Compress Reduced Resection Segmental Distal Femur (7 cm) 7 cm OSS Small Head / Small Thread Locking Screw (optional) (8 cm Replacement Length is Illustrated)

49 Compress Device Surgical Technique Screw Placement Guide Utilizing Taper Adapter for a Distal Femoral Replacement Compress Spindle (1 cm) Compress Locking Cap (optional) OSS / Compress Taper Adapter (5 cm) 1 cm OSS Small Head / Small Thread Locking Screw (optional) 5 cm OSS Segmental Distal Femur (7 cm) 7 cm OSS Large Head / Large Thread Locking Screw (13 cm Replacement Length is Illustrated)

50 Compress Device Surgical Technique Screw Placement Guide Utilizing Diaphyseal Segment and Taper Adapter for a Distal Femoral Replacement Compress Spindle (1 cm) Compress Locking Cap (optional) OSS / Compress Taper Adapter (5 cm) OSS Small Head / Small Thread Locking Screw (optional) OSS Diaphyseal Stacking Adapter 1 cm 5 cm OSS Diaphyseal Segment (5 cm) 5 cm OSS Small Head / Small Thread Locking Screw OSS Segmental Distal Femur (7 cm) 7 cm OSS Large Head / Large Thread Locking Screw (18 cm Replacement Length is Illustrated)

51 Compress Device Surgical Technique Screw Placement Guide Utilizing Reduced Resection Proximal Femur for a Proximal Femoral Replacement OSS Small Head / Small Thread Locking Screw (optional) Compress Reduced Resection Proximal Femur (7 cm) 7 cm 1 cm Compress Locking Cap (optional) (8 cm Replacement Length is Illustrated) Compress Spindle (1 cm) Compress Anchor Plug

52 Compress Device Surgical Technique Screw Placement Guide Utilizing Taper Adapter for a Proximal Femoral Replacement OSS Large Head / Large Thread Locking Screw 7 cm OSS Finn Style Proximal Femur (7 cm) OSS Small Head / Small Thread Locking Screw (optional) 5 cm 1 cm OSS / Compress Taper Adapter (5 cm) (13 cm Replacement Length is Illustrated) Compress Locking Cap (optional) Compress Spindle (1 cm) Compress Anchor Plug

53 Compress Device Surgical Technique Screw Placement Guide Utilizing Short Spindle for a Proximal Femoral Replacement OSS Large Head / Large Thread Locking Screw 7 cm OSS Finn-Style Proximal Femur (7 cm) OSS Small Head / Small Thread Locking Screw 5 cm OSS Diaphyseal Segment (5 cm) OSS Diaphyseal Stacking Adapter OSS Small Head / Small Thread Locking Screw (optional) 5 cm 1 cm (18 cm Replacement Length is Illustrated) OSS / Compress Taper Adapter (5 cm) Compress Locking Cap (optional) Compress Spindle (1 cm) Compress Anchor Plug

54 Compress Device Surgical Technique Screw Placement Guide Utilizing Short Spindle for a Low Profile Proximal Femoral Replacement OSS Small Head / Small Thread Locking Screw 7 cm OSS Low Profile Proximal Femur (7 cm) OSS Small Head / Small Thread Locking Screw 5 cm OSS Diaphyseal Segment (5 cm) OSS Diaphyseal Stacking Adapter OSS Small Head / Small Thread Locking Screw (optional) 5 cm 1 cm (18 cm Replacement Length is Illustrated) OSS / Compress Taper Adapter (5 cm) Compress Locking Cap (optional) Compress Spindle (1 cm) Compress Anchor Plug

55 Compress Device Surgical Technique Tibial Preparation To prepare the tibia, reference the surgical steps utilized in the OSS Surgical Techniques.

56 Compress Device Surgical Technique Notes

All content herein is protected by copyright, trademarks and other intellectual property rights, as applicable, owned by or licensed to Zimmer Biomet or its affiliates unless otherwise indicated, and must not be redistributed, duplicated or disclosed, in whole or in part, without the express written consent of Zimmer Biomet. This material is intended for health care professionals. Distribution to any other recipient is prohibited. For product information, including indications, contraindications, warnings, precautions, potential adverse effects and patient counseling information, see the package insert and www.zimmerbiomet.com. Zimmer Biomet does not practice medicine. This technique was developed in conjunction with health care professionals. This document is intended for surgeons and is not intended for laypersons. Each surgeon should exercise his or her own independent judgment in the diagnosis and treatment of an individual patient, and this information does not purport to replace the comprehensive training surgeons have received. As with all surgical procedures, the technique used in each case will depend on the surgeon s medical judgment as the best treatment for each patient. Results will vary based on health, weight, activity and other variables. Not all patients are candidates for this product and/or procedure. Caution: Federal (USA) law restricts this device to sale by or on the order of a surgeon. Rx only. Check for country product clearances and reference product specific instructions for use. Not intended for surgeons practicing medicine in France. 2018 Zimmer Biomet Authorized Representative Biomet UK Ltd. Waterton Industrial Estate Bridgend, South Wales CF31 3XA UK 1110.2-GLBL-en REV0818 Legal Manufacturer Biomet Orthopedics P.O. Box 587 56 E. Bell Drive Warsaw, Indiana 46581-0587 USA www.zimmerbiomet.com 0086 CE Mark on a surgical technique is not valid unless there is a CE mark on the product label.