The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates

Similar documents
IMRT QA: Point Dose Measurements or 2D Array?

EPID dosimetry for pretreatment quality assurance with two commercial systems

Volumetric Modulated Arc Therapy - Patient Specific QA

Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell

A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Confidence limits for patient-specific IMRT dose QA: a multi-institutional study in Korea

Leila E. A. Nichol Royal Surrey County Hospital

Statistical process control analysis for patient-specific IMRT and VMAT QA

Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times

Gamma analysis dependence on specified low-dose thresholds for VMAT QA

PMP. Gamma Evaluation with Portal Dosimetry for Volumetric Modulated Arc Therapy and Intensity-Modulated Radiation Therapy.

Reliability of the gamma index analysis as a verification method of volumetric modulated arc therapy plans

Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

MEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Agenda. 1.1 Why pre-treament dosimetry in IMRT/VMAT? 1.2 Verification systems in IMRT/VMAT: from 0D to 3D (4D) dosimetry to real time dosimetry

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors

The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT

Small field diode dosimetry

EPID-based dosimetry to verify IMRT planar dose distribution for the as1200 EPID and FFF beams

EPID dosimetry confi guration and pre-treatment IMRT verifi cation

Dosimetric verification and quality assurance of runningstart-stop (RSS) delivery in tomotherapy

IMRT pre-treatment verification basic approaches

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Patient QA Recommended Publications

A new geometric and mechanical verification device for medical LINACs

Abdulhamid Chaikh 1,2, Jacques Balosso 1,2. Introduction

Nuclear Associates

Eric E. Klein, Ph.D. Chair of TG-142

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Limits of Precision and Accuracy of Radiation Delivery Systems

biij Patient-specific quality assurance of RapidArc treatments: Portal prediction dosimetry compared with phantom studies Krishna Murthy K *

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

EORTC Member Facility Questionnaire

Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

ORIGINAL ARTICLE Oncology & Hematology INTRODUCTION

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Trajectory Modulated Arc Therapy: Application to Partial Breast Irradiation. Research and development to advance radiotherapy

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter

Learning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy

VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

PlanIQ. Plan Smarter, Not Harder. Your Most Valuable QA and Dosimetry Tools

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

SunCHECK Patient Comprehensive Patient QA

Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Assessing the shift of radiobiological metrics in lung radiotherapy plans using 2D gamma index

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements

A multi-institution evaluation of MLC log files and performance in IMRT delivery

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

The effect of body contouring on the dose distribution delivered with volumetric-modulated arc therapy technique

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

IMRT Planning Basics AAMD Student Webinar

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Normal tissue dose in pediatric VMAT Piotr Zygmanski

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments

Verification of Advanced Radiotherapy Techniques

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

Comparison of 3D anatomical dose verification and 2D phantom dose verification of IMRT/VMAT treatments for nasopharyngeal carcinoma

Evaluation of triple channel correction acquisition method for radiochromic film dosimetry

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

Quantification of interplay and gradient effects for lung stereotactic ablative radiotherapy (SABR) treatments

Neutron dose evaluation in radiotherapy

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

7/10/2015. Acknowledgments. Institution-specific TG-142? AAPM:Task Group-142. Failure-Mode & Effects Analysis

Tuning of AcurosXB source size setting for small intracranial targets

Many vendors are beginning to allow couch motion during radiation delivery.

The Radixact System Experience at Miami Cancer Institute. Tino Romaguera, D.Sc. Senior Physicist

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Transcription:

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates Daniel W. Bailey, 1,2 a Jason D. Spaans, 2 Lalith K. Kumaraswamy, 2 and Matthew B. Podgorsak 2,3 Department of Radiation Oncology, 1 Northside Hospital Cancer Institute, Atlanta, GA, USA; Department of Radiation Medicine, 2 Roswell Park Cancer Institute, Buffalo, NY, USA; Department of Physiology and Biophysics, 3 State University of New York at Buffalo, Buffalo, NY, USA daniel.bailey@northside.com Received 17 August, 2015; accepted 13 November, 2015 Our study aimed to quantify the effect of the Measurement Uncertainty function on planar dosimetry pass rates, as measured and analyzed with the Sun Nuclear Corporation MapCHECK 2 array and its associated software. This optional function is toggled in the program preferences of the software (though turned on by default upon installation), and automatically increases the dose difference tolerance defined by the user for each planar dose comparison. Dose planes from 109 static-gantry IMRT fields and 40 VMAT arcs, of varying modulation complexity, were measured at 5 cm water-equivalent depth in the MapCHECK 2 diode array, and respective calculated dose planes were exported from a commercial treatment planning system. Planar dose comparison pass rates were calculated within the Sun Nuclear Corporation analytic software using a number of calculation parameters, including Measurement Uncertainty on and off. By varying the percent difference (%Diff) criterion for similar analyses performed with Measurement Uncertainty turned off, an effective %Diff criterion was defined for each field/arc corresponding to the pass rate achieved with Measurement Uncertainty turned on. On average, the Measurement Uncertainty function increases the user-defined %Diff criterion by 0.8% 1.1% for 3%/3 mm analysis, depending on plan type and calculation technique (corresponding to an average change in pass rate of 1.0% 3.5%, and a maximum change of 8.7%). At the 2%/2 mm level, the Measurement Uncertainty function increases the user-defined %Diff criterion by 0.7% 1.2% on average, again depending on plan type and calculation technique (corresponding to an average change in pass rate of 3.5% 8.1%, and a maximum change of 14.2%). The largest increases in pass rate due to the Measurement Uncertainty function are generally seen with poorly matched planar dose comparisons, while the function has a notably smaller effect as pass rates approach 100%. The Measurement Uncertainty function, then, may substantially increase the pass rates for planar dose comparisons. Meanwhile, the types of uncertainties incorporated into the function (and their associated quantitative estimates, as described in the software user s manual) may not be an accurate estimation of actual measurement uncertainty, depending on the user s measurement conditions. Pass rates listed in published reports, comparisons between institutions or simply separate workstations, or comparisons with the calculation methods of other vendors, should clearly indicate whether or not the Measurement Uncertainty function is used, since it has the potential to substantially inflate pass rates for typical IMRT and VMAT dose planes. PACS number(s): 87.55.Qr, 87.56.Fc Key words: radiation therapy, IMRT QA, quality assurance, dosimetry, dose verification, diode array, planar dose a Corresponding author: Daniel W Bailey, Department of Radiation Oncology, Northside Hospital Cancer Institute, Atlanta, GA 30342, USA; phone (404) 851 6842; fax: (404) 851 6010; email: daniel.bailey@northside.com 165 165

166 Bailey et al.: MapCHECK uncertainty 166 I. INTRODUCTION Planar diode arrays are a standard tool for myriad external beam quality assurance tests, most popularly for patient-specific dosimetric verification of intensity-modulated radiation therapy fields, (1,2) for both static (3,4) and rotational (5,6) delivery techniques. One of the diode arrays most commonly used for planar dose measurement is the MapCHECK (7,8) device (and its upgrade, the MapCHECK 2) (9-11) by Sun Nuclear Corporation of Melbourne, FL, accompanied by its analytical software component (i.e., MapCHECK up to version 5.2; SNC Patient thereafter). The MapCHECK and SNC Patient softwares allow the user to quantitatively compare measured and calculated dose planes, utilizing the gamma analysis and distance-to-agreement (DTA) techniques described elsewhere in the literature. (12-14) In a previous report, it was demonstrated that, rather than describing such a dosimetric comparison with a single pass rate, the comparison of any two respective measured and calculated planar dose distributions is actually characterized by a spectrum of pass rates dependent on parameters such as calculation method, percent difference (%Diff) and DTA tolerances, choice of percent difference normalization value, detector spatial resolution and individual diode position with respect to the beam s eye view, absolute or relative dose comparison, and the user s definition of region of interest and/or choice of applied dose threshold (i.e., dose level below which individual diode responses are not incorporated into the pass rate). (14) For this reason, pass rates should be reported (especially between institutions, e.g., in publications) with a comprehensive description of exactly how said pass rates are calculated. An additional calculation parameter within the MapCHECK and SNC Patient software, which is often applied but seldom reported in the literature, is the Measurement Uncertainty function, which is turned on by default upon software installation and can be toggled in program preferences. (15) The software manual describes this function as a supplement to the user-defined acceptance criteria added to the percentage acceptance criterion defined by the user to account for various uncertainties in measurement. (15) In other words, this optional function automatically increases the dose difference tolerance selected by the user for DTA or gamma comparison of planar dose distributions, in an attempt to compensate for various sources of presumed uncertainty in MapCHECK or ArcCHECK array measurements. Because the Measurement Uncertainty value is added to the %Diff criterion chosen for gamma or DTA analysis, this function always increases the resultant pass rate. Task Group 119 of the American Association of Physicists in Medicine utilized the Measurement Uncertainty function in the analysis of their cross-institutional data, stating that they applied measurement uncertainty (a presumed measurement error of about 1% is included in the analysis, so that a nominal 3% dose difference can be 4%). (16) In the absence of quantitative data demonstrating the effect of the Measurement Uncertainty function on absolute dose pass rates, we determined to study a number of static-gantry IMRT and RapidArc fields at various %Diff/DTA criteria levels and with/without Measurement Uncertainty to evaluate this parameter that is unique to the Sun Nuclear Corporation software. II. MATERIALS AND METHODS The following treatment plans were used in this study, calculated with the Eclipse treatment planning system (TPS; AAA 10, Varian Medical Systems, Palo Alto, CA): five static-gantry IMRT plans for head/neck (49 fields), nine static-gantry IMRT plans for prostate (60 fields), ten RapidArc plans for head/neck (20 arcs), and ten RapidArc plans for prostate (20 arcs). Each field or arc was delivered on a Varian Trilogy linear accelerator system, and planar dose data were measured via MapCHECK 2 placed inside the MapPHAN (9-11) water-equivalent phantom with source-to-detector distance of 100 cm and water-equivalent depth of 5 cm. Respective inphantom verification plans were created for each field or arc within the TPS and cross-sectional

167 Bailey et al.: MapCHECK uncertainty 167 dose planes (at 5 cm water-equivalent depth) were exported in DICOM format for analysis with the SNC Patient software. For large IMRT fields requiring split carriages, each carriage was measured separately with the diode array, then merged and analyzed as one treatment field. For brevity and clarity, we will refer to MapCHECK Measurement Uncertainty on and off as MCU on and, respectively. Quantitative comparison of measured versus calculated dose planes was conducted within the SNC Patient software (version 6.2.0), first with the DTA method and then with the gamma method, using absolute dose mode, global percent difference normalization to the maximum planned point dose, and a 10% dose threshold (effectively limiting the analysis to within the collimator jaws). The Calculate Shift function was not used, so no geometric optimization of pass rate was employed. In addition, pass rates (i.e., percentage of measured points with gamma value less than unity) were recorded with DTA criterion of 2 mm while the %Diff criterion was varied from 1% to 4% in half-percent increments, and again with DTA criterion of 3 mm with the same variation in %Diff. To quantify the effect of Measurement Uncertainty, pass rates with each selection of %Diff/DTA criterion were calculated both with MCU on and. These data were used to approximate an effective percent difference (%Diff eff ) for each pass rate calculated with MCU on : for a given DTA criterion and, the pass rates for varying %Diff were linearly interpolated, and the %Diff value corresponding to the pass rate calculated with MCU on was defined as the effective percent difference criterion for that comparison. III. RESULTS AND DISCUSSION Tables 1 and 2 summarize the results for prostate and head/neck static-gantry IMRT using DTA and gamma analysis, respectively, both with 3%/3 mm criteria. Tables 3 and 4 present the same analysis but at the 2%/2 mm level. Similarly, Tables 5 through 8 present the results for RapidArc plans. These tables all indicate similar results: at the 3%/3 mm analysis level, the Measurement Uncertainty function increases user-defined %Diff criterion to an effective %Diff between 3.8% to 4.1% on average, depending on plan type and calculation technique. Meanwhile, at the 2%/2 mm analysis level, the Measurement Uncertainty function yields an effective %Diff of approximately 2.7% 3.2% on average. In other words, the 3%/3 mm pass rate with MCU on is approximately equivalent to a 4%/3 mm pass rate calculated with. For individual fields, the effective %Diff can be as high as 6% when analyzing fields at the 3%/3 mm level, and as high as 5.8% when using the 2%/2 mm criteria. The maximum observed change in pass rate for static-gantry IMRT fields due to the Measurement Uncertainty function was +8.7% at 3%/3 mm and +14.2% at the 2%/2 mm level. For RapidArc plans, the maximum observed change in pass rate due to the Measurement Uncertainty function was +8.7% at 3%/3 mm and +13.1% at the 2%/2 mm level. Table 1. MapCHECK Uncertainty results for IMRT with DTA 3%/3 mm. 3%/3 mm 3%/3 mm 3%/3 mm %Diff eff (on - off ) for 3%/3 mm Prostate average 95.5 92.3 3.3 4.1 minimum 84.4 77.1 0.0 3.0 maximum 100.0 99.5 8.7 6.0 SD 4.0 5.8 2.2 0.5 H&N average 96.1 94.2 1.9 4.1 minimum 90.8 85.9 0.0 3.2 maximum 99.3 98.7 5.3 6.0 SD 2.1 2.7 1.1 0.5

168 Bailey et al.: MapCHECK uncertainty 168 Table 2. MapCHECK Uncertainty results for IMRT with gamma 3%/3 mm. 3%/3 mm 3%/3 mm 3%/3 mm %Diff eff (on - off ) for 3%/3 mm Prostate average 97.2 95.5 1.7 3.8 minimum 89.8 83.8 0.0 3.1 maximum 100.0 100.0 7.8 4.0 SD 2.8 4.1 1.8 0.3 H&N average 97.4 96.4 1.0 3.8 minimum 92.6 90.3 0.0 3.0 maximum 100.0 99.7 4.0 6.1 SD 1.7 2.3 1.0 0.5 Table 3. MapCHECK Uncertainty results for IMRT with DTA 2%/2 mm. 2%/2 mm 2%/2 mm 2%/2 mm %Diff eff (on - off ) for 2%/2 mm Prostate average 84.9 76.8 8.1 3.2 minimum 64.4 53.6 2.2 2.4 maximum 98.7 92.7 14.2 5.8 SD 9.8 11.0 3.1 0.5 H&N average 89.5 84.4 5.0 3.0 minimum 81.0 73.3 1.9 2.4 maximum 95.8 92.6 8.7 5.6 SD 3.3 4.5 1.8 0.5 Table 4. MapCHECK Uncertainty results for IMRT with gamma 2%/2 mm. 2%/2 mm 2%/2 mm 2%/2 mm %Diff eff (on - off ) for 2%/2 mm Prostate average 87.1 81.6 5.5 2.8 minimum 69.9 60.7 0.7 2.3 maximum 99.4 96.8 11.8 3.0 SD 8.5 10.1 2.7 0.2 H&N average 91.3 87.8 3.5 2.7 minimum 83.0 77.7 1.2 2.3 maximum 97.4 95.1 8.3 3.2 SD 3.1 4.0 1.7 0.2 Table 5. MapCHECK Uncertainty results for VMAT with DTA 3%/3 mm. 3%/3 mm 3%/3 mm 3%/3 mm %Diff eff (on - off ) for 3%/3 mm Prostate average 96.2 92.8 3.4 3.8 minimum 92.7 87.3 1.6 3.6 maximum 98.7 96.5 5.5 4.1 SD 2.0 2.7 1.1 0.2 H&N average 95.8 92.2 3.5 3.8 minimum 90.4 86.2 1.3 3.5 maximum 99.5 97.7 8.7 4.1 SD 2.4 3.4 1.7 0.1

169 Bailey et al.: MapCHECK uncertainty 169 Table 6. MapCHECK Uncertainty results for VMAT with gamma 3%/3 mm. 3%/3 mm 3%/3 mm 3%/3 mm %Diff eff (on - off ) for 3%/3 mm Prostate average 97.8 95.7 2.2 3.8 minimum 94.6 89.7 0.8 3.3 maximum 100.0 98.6 4.9 4.0 SD 1.7 2.2 0.9 0.2 H&N average 97.4 95.1 2.3 3.8 minimum 92.9 90.8 0.7 3.4 maximum 100.0 99.2 5.1 4.0 SD 1.7 2.5 1.1 0.2 Table 7. MapCHECK Uncertainty results for VMAT with DTA 2%/2 mm. 2%/2 mm 2%/2 mm 2%/2 mm %Diff eff (on - off ) for 2%/2 mm Prostate average 87.8 81.2 6.5 2.7 minimum 78.4 73.0 2.6 2.4 maximum 92.1 86.5 9.4 3.2 SD 3.5 4.1 1.8 0.2 H&N average 87.5 79.8 7.7 2.8 minimum 80.3 69.8 2.6 2.6 maximum 94.1 91.5 13.1 3.0 SD 4.2 6.1 2.6 0.1 Table 8. MapCHECK Uncertainty results for VMAT with gamma 2%/2 mm. 2%/2 mm 2%/2 mm 2%/2 mm %Diff eff (on - off ) for 2%/2 mm Prostate average 90.6 85.2 5.3 2.7 minimum 83.8 77.9 2.8 2.4 maximum 95.2 91.8 9.3 2.9 SD 3.5 4.3 1.5 0.1 H&N average 90.2 84.5 5.8 2.7 minimum 83.1 73.4 2.6 2.5 maximum 96.4 93.8 10.3 2.9 SD 4.2 5.5 1.9 0.1 Figures 1 and 2 allow visualization of the spread of pass rates for different gamma criterion (namely, 3%/3 mm and 2%/2 mm), as well as the effect of toggling the Measurement Uncertainty function. In each of the figures, the x markers indicate data collected with the MCU on while the dot markers indicate data collected with. Figure 1 (top) shows the average pass rates for a static-gantry IMRT head/neck case (14 fields total) and Fig. 1 (bottom) shows the average pass rates for a static-gantry IMRT prostate case (10 fields total). For these calculations, the percent difference criterion was varied from 1% to 4%, as indicated on the horizontal axes, while the distance-to-agreement criterion was fixed at 3 mm. Meanwhile, Fig. 2 displays the individual pass rates for all 49 static-gantry IMRT head/neck fields tested for this report, first with gamma analysis of 3%/3 mm (top) and secondly with 2%/2 mm criteria (bottom). Each vertical column displays a pair of pass rates for an individual planar dose comparison (measured vs. TPS), one with MCU on (x) and the other with (dot) (i.e., the only change is made by toggling the Measurement Uncertainty function). From these figures,

170 Bailey et al.: MapCHECK uncertainty 170 Fig. 1. Static-gantry IMRT planar dose comparisons for a head/neck (top) and prostate (bottom) case. Absolute dose gamma evaluations performed with variable percent difference criterion as indicated on the horizontal axes, and fixed distance-to-agreement criterion of 3 mm. Average pass rates calculated with Measurement Uncertainty on are indicated by x markers, while similar averages calculated with Measurement Uncertainty off are indicated by dot markers. Fig. 2. Static-gantry IMRT planar dose comparisons for all 49 head/neck fields tested in this report, first with gamma analysis of 3%/3 mm (top) and secondly with 2%/2 mm criteria (bottom). Each vertical column represents a single planar dose comparison of a measured field vs. its respective dose plane calculated in the TPS, resulting in one pass rate with MCU on (x) and the other with (dot).

171 Bailey et al.: MapCHECK uncertainty 171 two general trends are noteworthy and complimentary to the tabular data discussed above: the Measurement Uncertainty function may result in large increases in pass rate for poorly matching planar dose comparisons; however, the Measurement Uncertainty function has a smaller effect on already high pass rates, evidently due to the asymptotic behavior of the gamma calculation when approaching 100%. At the time of final submission of this manuscript, the most current versions of the SNC Patient software (beginning with version 6.2.3, released after the data in this work were collected) include an updated Measurement Uncertainty function, as detailed identically in the current versions of the Sun Nuclear Corporation help manuals (either MapCHECK Help, MapCHECK 2 Help, or ArcCHECK Help, depending on the user s device). While the basic functionality of the Measurement Uncertainty feature remains consistent with the discussion in this work, a constant value was assumed for the magnitude of the correction applied directly to (i.e., increasing) the user-selected dose difference percentage: 0.8% for absolute dose comparison and 0.5% for relative dose comparison with MapCHECK and MapCHECK 2. According to these vendor manuals, the Measurement Uncertainty evidently behaves similarly when applied to ArcCHECK measurements, with slightly higher constant values of 1.0% for absolute dose comparison and 0.7% for relative dose comparison, added directly to the userdefined dose difference comparison criterion. Since the Measurement Uncertainty function in previous versions of the software was allowed to take on variable values depending on each measured data point, the magnitude of the function in the latest version of the software, or in any future update to the software, may potentially be different. Users of the MapCHECK and SNC Patient software tools should be aware that the Measurement Uncertainty function can result in substantially inflated pass rates for planar dose comparisons. While each institution included in the TG-119 data pool employed Measurement Uncertainty for the sake of consistency (and not as an endorsement), it should be noted that this function adds another level of difficulty in comparing the pass rates calculated by the SNC software to pass rates calculated by other software analysis tools (e.g., from other vendors), or pass rates calculated by other institutions (or even other workstations within the same institution) that do not have the SNC software set up to employ the Measurement Uncertainty function. At the same time, several recent reports (17-20) indicate that inflated pass rates may actually mask errors in treatment planning and delivery, and this information should be weighed in the decision to use the Measurement Uncertainty function. A typical indication of measurement uncertainty would tend to provide an interval of confidence in the given measurement. Namely, the actual uncertainty in any measured dose plane comparison indicates a pass rate potentially higher or lower than the value calculated without accounting for such uncertainties. However, because of the definition of the Measurement Uncertainty function, as an addition to the user-defined %Diff tolerance level, it can only increase the calculated pass rate. (20) Specifically, due to this application of estimated uncertainty, a measured dose point outside the user-defined tolerance might be included as a passing point, while a dose point inside the tolerance criteria but with uncertainty enough to potentially be a failing point is never actually counted as a failing point. The SNC user s manual lists several uncertainties approximately quantified and included in the Measurement Uncertainty function. These include an array calibration correction (approximated from a study dealing with the original MapCHECK device), (21) a correction for assumed accelerator output and room temperature differences between the time of array calibration and actual measurement, and a correction for an assumed error in setup source-to-surface distance (SSD). The user should decide whether or not these enumerated uncertainties (and their approximate values used to supplement the %Diff criteria) are applicable to one s own measurement conditions. For example, a default 2 mm SSD correction may not be appropriate for a carefully set up device coupled to a quality-controlled accelerator, while an automatic pass rate adjustment to account for assumed output fluctuation and assumed temperature difference between the time of measurement and the time of calibration may not be accurate for a diode array that is frequently calibrated. If

172 Bailey et al.: MapCHECK uncertainty 172 the actual use of the diode array does not match the assumed corrections incorporated into the Measurement Uncertainty function, then the use of the function does not enhance the accuracy of subsequent calculated pass rates. From the current literature on the subject, planar dose QA pass rates are quite variable even for a single treatment field, and are dependent on device used for measurement, resolution of detector and detector position (i.e., sampling of measured points within the field), region of interest, and dose/dta tolerance criteria, along with all of the physical uncertainties that accompany these types of measurements. However, this work demonstrates that even identical measurements may yield significantly different pass rates depending on the specifics of the vendor s algorithm utilized to implement the gamma evaluation. It would be useful to all users if each vendor would provide an analysis of expected returned pass rate values based upon a universally employed dataset, much like the dataset provided alongside the TG-119 report, giving the user an indication of any pass rate inflation or deflation inherent to the vendor s calculation algorithm. V. CONCLUSION This study demonstrates that the Measurement Uncertainty function within the SNC planar dose comparison software may substantially inflate the pass rates for planar dose comparisons, by as much as 9% 14% depending on plan type and calculation technique. Since this function is unique to the MapCHECK and SNC Patient software tools, use of the Measurement Uncertainty function should be clearly indicated, especially in published reports. Meanwhile, the types of uncertainties incorporated into the function (and their associated quantitative estimates, as described in the software user s manual) may not be an accurate estimation of actual measurement uncertainty, depending on the user s measurement conditions. Consequently, each user should carefully consider whether or not the application of the Measurement Uncertainty function matches his or her own measurement techniques and conditions. COPYRIGHT This work is licensed under a Creative Commons Attribution 4.0 International License. REFERENCES 1. Low DA, Moran JM, Dempsey JF, Dong L, Oldham M. Dosimetry tools and techniques for IMRT. Med Phys. 2011;38(3):1313 38. 2. Li JG, Yan G, Liu C. Comparison of two commercial detector arrays for IMRT quality assurance. J Appl Clin Med Phys. 2009;10(2):62 74. 3. Buonamici FB, Compagnucci A, Marrazzo L, Russo S, Bucciolini M. An intercomparison between film dosimetry and diode matrix for IMRT quality assurance. Med Phys. 2007;34(4):1372 79. 4. Létourneau D, Gulam M, Yan D, Oldham M, Wong JW. Evaluation of a 2D diode array for IMRT quality assurance. Radiother Oncol. 2004;70(2):199 206. 5. Jursinic PA, Sharma R, Reuter J. MapCHECK used for rotational IMRT measurements: step-and-shoot, TomoTherapy, RapidArc. Med Phys. 2010;37(6):2837 46. 6. Gloi AM, Buchana RE, Zuge CL, Goettler AM. RapidArc quality assurance through MapCHECK. J Appl Clin Med Phys. 2011;12(2):39 47. 7. Jursinic PA and Nelms BE. A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery. Med Phys. 2003;30(5):870 79. 8. Nelms BE and Simon JA. A survey on planar IMRT QA analysis. J Appl Clin Med Phys. 2007;8(3):76 90. 9. Keeling VP, Ahmad S, Jin H. A comprehensive comparison study of three different planar IMRT QA techniques using MapCHECK 2. J Appl Clin Med Phys. 2013;14(6):222 33. 10. Rinaldin G, Perna L, Agnello G, et al. Quality assurance of rapid arc treatments: performances and pre-clinical verifications of a planar detector (MapCHECK2). Phys Med. 2014;30(2):184 90.

173 Bailey et al.: MapCHECK uncertainty 173 11. Jin H, Keeling VP, Johnson DA, Ahmad S. Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance. J Appl Clin Med Phys. 2014;15(3):4638. 12. Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25(5):656 61. 13. Low DA and Dempsey JF. Evaluation of the gamma dose distribution comparison method. Med Phys. 2003;30(9):2455 64. 14. Bailey DW, Nelms BE, Attwood K, Kumaraswamy L, Podgorsak MB. Statistical variability and confidence intervals for planar dose QA pass rates. Med Phys. 2011;38(11):6053 64. 15. Sun Nuclear Corporation. MapCHECK 2 reference guide: raising the bar in IMRT QA. Document 1177011, Rev I-5. Melbourne, FL: Sun Nuclear; 2014. 16. Ezzell GA, Burmeister JW, Dogan N, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009;36(11):5359 73. 17. Nelms BE, Zhen H, Tomé WA. Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors. Med Phys. 2011;38(2):1037 44. 18. Zhen H, Nelms BE, Tome WA. Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA. Med Phys. 2011;38(10):5477 89. 19. Nelms BE, Chan MF, Jarry G, et al. Evaluating IMRT and VMAT dose accuracy: practical examples of failure to detect systematic errors when applying a commonly used metric and action levels. Med Phys. 2013;40(11):111722. 20. Nelms B, Jarry G, Chan M, Hampton C, Watanabe Y, FeygelmanV. Real-world examples of sensitivity failures of the 3%/3mm pass rate metric and published action levels when used in IMRT/VMAT system commissioning. J Phys Conf Ser. Vol. 444. Bristol, UK: IOP Publishing; 2013. 21. Létourneau D, Gulam M, Yan D, Oldham M, Wong JW. Evaluation of a 2D diode array for IMRT quality assurance. Radiother Oncol. 2004;70(2):199 206.