Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Similar documents
IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

QA for Clinical Dosimetry with Emphasis on Clinical Trials

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

Anthropomorphic breast phantoms for quality assurance and dose verification

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

Quality assurance in external radiotherapy

A new geometric and mechanical verification device for medical LINACs

Brachytherapy Planning and Quality Assurance

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Leila E. A. Nichol Royal Surrey County Hospital

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET

Protection of the contralateral breast during radiation therapy for breast cancer

Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS

Measure the Errors of Treatment Set-Ups of Prostate Cancer Patient Using Electronic Portal Imaging Device (EPID)

The Critical Importance of High Quality in Radiation Therapy

SBRT Credentialing: Understanding the Process from Inquiry to Approval

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Defining Target Volumes and Organs at Risk: a common language

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Patient dosimetry for total body irradiation using single-use MOSFET detectors

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

Commission, Ganakbari, Savar, Dhaka, Bangladesh

Medical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center

Effectiveness of compensating filters in the presence of tissue inhomogeneities

Diode calibration for dose determination in total body irradiation

Radiotherapy In-vivo Treatment Verification: Multi-center analysis for more than 6 million fractions delivered

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

Inhomogeneity correction and the analytic anisotropic algorithm

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Limits of Precision and Accuracy of Radiation Delivery Systems

Additional Questions for Review 2D & 3D

Preventing treatment errors in radiotherapy by identifying and evaluating near misses and actual incidents

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Hampton University Proton Therapy Institute

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

TREATMENT PLANNING: CLINICAL ELECTRONS John A. Antolak, Ph.D. Prepared for the AAPM Therapy Physics Review Course Austin, TX July 19 20, 2014 PURPOSE

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

A translational couch technique for total body irradiation

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague

Computerized Treatment Planning in Radiation Therapy of Intact Breast: Influence of Number of CT-Cuts

Dose prescription, reporting and recording in intensity-modulated radiation therapy: a digest of the ICRU Report 83

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Review of TG-186 recommendations

The determination of timer error and its role in the administration of specified doses

Heterogeneity Corrections in Clinical Trials

Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Abdulhamid Chaikh 1,2, Jacques Balosso 1,2. Introduction

International Practice Accreditation

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy

Transition to Heterogeneity Corrections. Why have accurate dose algorithms?

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

SRS Uncertainty: Linac and CyberKnife Uncertainties

T R E A T M E N T D E L I V E R Y V E R I F I C A T I O N U S I N G A P E L V I C A N T H R O P O R M O P H I C P H A N T O M

Nuclear Associates

Ritu Raj Upreti, S. Dayananda, R. L. Bhalawat*, Girish N. Bedre*, D. D. Deshpande

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Predicting Maximum Pacemaker/ICD Dose in SAVI HDR Brachytherapy

Treatment of exceptionally large prostate cancer patients with low-energy intensity-modulated photons

Dosisverifikation mit Delta 4 Discover während der Behandlung

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a

REPLACING PRETREATMENT VERIFICATION WITH IN VIVO EPID DOSIMETRY FOR PROSTATE IMRT

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

Original Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

Transcription:

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and Peter B. Dunscombe Department of Medical Physics, Northeastern Ontario Regional Cancer Centre, 41 Ramsey Lake Road, Sudbury, P3E 5J1, Canada, Department of Radiology, University of Ottawa, Ottawa K1N 6N5, Canada, and Department of Physics, Laurentian University, Sudbury P3E 2C6, Canada Received 17 May 2000; accepted for publication 10 August 2000 The checking of monitor unit calculations is recognized as a vital component of quality assurance in radiotherapy. Using straightforward but detailed computerbased verification calculations it is possible to achieve a precision of 1% when compared with a three-dimensional 3D treatment planning system monitor unit calculation. The method is sufficiently sensitive to identify significant errors and is consistent with current recommendations on the magnitude of uncertainties in clinical dosimetry. Moreover, the approach is accurate in the sense of being highly consistent with the validated 3D treatment planning system s calculations. 2000 American College of Medical Physics. PACS number s : 87.53. j, 87.52. g Key words: monitor unit calculations, quality assurance, treatment planning INTRODUCTION The success of radiation therapy crucially depends on the accuracy with which the prescribed dose is delivered to the tumor volume. Quantitative assessments of this relationship have led to the development of recommendations as to the accuracy required in dose delivery. According to the generally accepted recommendations made by the International Commission on Radiation Units and Measurements 1 the dose delivered should not deviate by more than 5% from the dose prescribed. More recently Mijnheer et al. 2 and Wambersie et al. 3 proposed that the standard deviation of the uncertainty in the delivered dose should not be greater than 3.5%. As only a part of the overall uncertainty arises from the process of dose calculations in treatment planning, the tolerances for the accuracy of treatment planning systems TPS have to be appropriately smaller. Uncertainties in dose delivery may be introduced at the treatment phase including machine calibration or during the process of deriving monitor unit or timer settings from the dose prescription determined by the radiation oncologist treatment preparation. Dose errors of the latter type, arising at the treatment planning phase, could potentially affect the whole course of treatment and therefore are of particular concern. For computerized calculations of monitor units MU, whether or not accompanied by a dose distribution, uncertainties may be further categorized as arising from the input beam data, the calculation algorithm, incorrect use of the system, and data transfer to the treatment sheet. Although there is a possibility that a significant dose error may arise as a result of beam data or the algorithm, commissioning is designed to minimize this risk. 4 In routine clinical practice, more likely sources of systematic dose error for individual patients result from misuse of the system e.g., through inadequate understanding of normalization protocols, misinterpretation of the system output and data transfer errors. Currently, MU settings required to deliver the prescribed dose are often calculated by a computerized treatment planning system using methods and quantities different from those used in manual MU calculations. This is particularly true of those planning systems, which employ the convolution-superposition method in their calculations. Published recommendations for quality 120 1526-9914Õ2000Õ1 4 Õ120Õ6Õ$17.00 2000 Am. Coll. Med. Phys. 120

121 K. W. Leszczynski and P. B. Dunscombe: Independent corroboration of monitor... 121 assurance QA in radiation therapy stipulate routine checking of MU calculations using means independent from the original calculator. 5 7 These checks are primarily intended to identify dose errors resulting from the three sources listed above and prior to the start of treatment. Independent checking, as we shall show below, can also enhance confidence in the accuracy of the algorithm and integrity of the beam data used although this is not generally the main intention. Previously published reports on independent checks of MU calculations 8 10 confirm the usefulness of this QA procedure in promoting the accurate delivery of the prescribed dose. They also provide an indication of the limitations of applicability of conventional dose calculation algorithms employed in computerized treatment planning systems. 10 In the selection of a method for checking monitor unit calculations it is clearly necessary to establish that the method has sufficient sensitivity, quantified as accuracy and precision, to perform its intended function. In this paper we describe our experience with an approach to the routine verification of the monitor units calculated by a three-dimensional 3D computerized treatment planning system Helax-TMS, Helax AB, Uppsala, Sweden. The approach was computer based and incorporated almost all the factors affecting photon dosimetry, which are features of the full 3D system. The method has been employed for the verification of the monitor unit calculations associated with close to 500 computerized treatment plans. Our analysis enables us to quantify the accuracy and precision of our method, by treatment site, and hence discuss its applicability in routine quality control of computerized monitor unit calculations. METHODS AND MATERIALS The quality control protocol at our cancer center requires all computerized treatment plans to be checked independently by a medical physicist. The study described here is based on timer setting and MU calculation checks performed for 497 treatment plans for patients who were treated with Cobalt-60, 6-MV, and 23-MV photon beams. All common treatment sites were represented in this sample. Our approach to the checking of Helax-TMS MU calculations has been based on the standard system of dosimetric calculations, 11 using output factor and tissue phantom ratio TPR tables and wedge and tray transmission factors. These factors were acquired totally independently of, and several years prior to, the data used by Helax-TMS and their applicability verified by analysis of routine QA on our treatment units, which indicated no significant change in radiation beam characteristics throughout the years. Certain simplifications were applied such as the use of a single scatter output factor without separating it into the collimator and phantom scatter components. All cases in this study involve isocentric dose calculations, and the reference normalization point for which the MU calculations were checked was defined according to the International Commission on Radiation Units and Measurements ICRU 50 recommendations. 12 In the great majority of cases the reference point was placed at the isocenter. In order to calculate monitor units the Helax-TMS planning system requires the user to specify the prescribed dose and the percentage isodose level to which it applies. Helax-TMS will then compute the monitor units for each beam taking into account beam modifiers such as blocks, trays, and wedges. The monitor units so calculated are entered into the patient s chart. The MU checking method employed an MS-Excel spreadsheet for the dose calculations. The depths for TPRs were measured independently, with a ruler, in the computed technology CT section or contours used in planning. The nominal field size was adjusted in the direction parallel to the central section of the treated volume to take into account additional shielding and/or the fact that a part of the field might have fallen outside the patient s body. Density corrections based on radiological depths and TPR ratios were applied only for lung and major bone e.g., pelvis inhomogeneities. Dose calculations for the majority of treatment sites were based on CT data. In these cases the Helax-TMS system derived electron densities from CT numbers through an appropriate calibration. Three major treatment sites for which CT data were not used in dose calcu-

122 K. W. Leszczynski and P. B. Dunscombe: Independent corroboration of monitor... 122 FIG. 1. A histogram showing the distribution of the ratios of independently evaluated reference point doses to those calculated by Helax-TMS. lations included the breast/chest wall, supraclavicular node region in breast cancer patients, and rectum. The only tissue inhomogeneity considered in these sites was due to lungs in tangential breast irradiations. A standard value of 0.26 was assigned as the relative electron density within contours representing the lung. In manual calculations always standard rather than CT-derived values for relative electron densities were used for the major inhomogeneities that were considered. Lungs were assigned the same relative electron density as in non-ct based TPS calculations and bone inhomogeneities were assumed to have a uniform density of 1.5 which falls within the range defined by the densities of cancellous spongy and compact bones. For a wedged field where the reference point was placed off the central axis a correction based on the previously measured dose profiles along the wedged direction was applied. The appropriate dosimetric factors were read from tables, interpolated for the actual field size, and entered manually into the spreadsheet. Thus the general formula used to calculate the dose contribution, D, from one field with the monitor unit setting, MU, was as follows: D K MU S c,p TPR ISqF TF WTF OAF CF. In the above equation, K is the output factor cgy/mu under calibration conditions; S c,p denotes the scatter output factor; ISqF is an inverse-square correction factor applied when nonstandard treatment distance was used; TF and WTF are the shielding/compensator tray and wedge transmission factors, respectively; OAF is the off-axis correction factor, used in off-axis calculations for wedged fields only; and CF represents a correction factor for tissue inhomogeneities. The analysis below consisted of taking the ratio of the independently evaluated dose to the reference point to that specified by the user in Helax-TMS and generating the means and standard deviations by site for the check method. RESULTS AND DISCUSSION The histogram in Fig. 1 shows frequency counts of the ratios of independently evaluated reference point doses to those calculated by Helax-TMS. Table I contains the results of the statistical analysis of these data. The 497 cases included have been divided into eight anatomical sites for which the averages and standard deviations of the dose ratios are tabulated. The different 1

123 K. W. Leszczynski and P. B. Dunscombe: Independent corroboration of monitor... 123 TABLE I. The averages and standard deviations SD of the ratios of independently evaluated reference point doses to those calculated by Helax- TMS. The results are grouped according to the major anatomical sites. Site group n Average SD Brain 23 1.007 0.006 Head & neck 83 1.006 0.012 Lung/esophagus 44 1.009 0.014 Breast/chest wall 117 1.001 0.009 Supraclavicular 24 1.013 0.010 Abdomen/pelvis a 37 1.003 0.013 Prostate 110 1.002 0.008 Rectum 59 0.990 0.009 All sites 497 1.002 0.012 a Excluding prostate and rectum, which are analyzed separately. anatomical sites represent different degrees of complexity of dose calculations performed both manually and by Helax-TMS, in terms of inhomogeneities, contour and field shape irregularities, and off-axis corrections. From the results shown in Table I the limiting standard deviation appears to be of the order of 1%. The sites for which standard deviations are smaller are those where the dose calculation geometry, in terms of external contours, internal inhomogeneities, and beam orientations, is very consistent from case to case. This is especially true for the two most common treatment sites, i.e., breast and prostate as well as some less common ones such as brain and rectum. Geometric variability in internal tissue inhomogeneities, beam orientations with respect to external contours and in the position of the dose reference points lead to somewhat poorer precision for manual calculations performed for lung, abdomen/pelvis, and head and neck cases, respectively. The accuracy of the MU check method can be inferred from the agreement between the manual calculations and those performed by the 3D system. The Helax-TMS treatment planning system has been extensively validated using several techniques including thermoluminescence dosimetry TLD measurements in an anthropomorphic phantom 4 and is believed to be accurate. In nearly 500 plans the reference point dose calculated is, on the average, only 0.2% different from that specified in the computerized treatment plan for MU calculations. This overall consistency between the MU check method and full 3D calculations leads us to believe that the check method is also highly accurate. It should be noted, however, that the accuracy of the manual MU calculations varies somewhat with the treatment site. The largest numerically and statistically most significant systematic deviations exist in calculations for supraclavicular and rectum treatment plans. In the case of supraclavicular calculations this is attributable to not taking into account off-axis dose profile variations in the half-blocked supraclavicular fields the only off-axis correction applied was due to wedge. The systematic deviation of 1% in the case of rectum plans is primarily due to ignoring in manual MU checks the beam hardening effect produced by wedge filters. This effect is not pronounced at shallower depths in other treatment sites where wedges are used e.g., breast. The dose underestimation due to neglecting the beam hardening effect is also not present in calculations for treatment sites of similar geometry to that of the rectum, but where wedge filters were not used e.g., prostate. The third significant systematic inaccuracy that can be identified in the results occurred in the calculations for lung and esophagus. It can be explained by the fact that the manual calculation method did not take into account scatter dose perturbations caused by tissue inhomogeneities. These systematic effects are very small and, in the opinion of the authors, are insignificant in the context of an MU checking procedure. However, if desired, they could be simply corrected for by applying a site dependent offset deduced from the results presented in Table I. The appropriateness of these methods for routine quality control of monitor units calculated by

124 K. W. Leszczynski and P. B. Dunscombe: Independent corroboration of monitor... 124 3D computerized treatment planning computers can be assessed by comparing the accuracy and precision results of Table I with available recommendations and analyses. Clearly if these two statistical characteristics are inconsistent with the sensitivity required of a checking procedure then the method is of little value. Van Dyk et al. 13 suggest that the accuracy of a treatment planning system should be 4% in a low dose gradient region with all corrections invoked. Mijnheer et al. 2 have quantitatively analyzed the sources of uncertainty in clinical dosimetry. Their analysis suggests that the treatment planning process may introduce an uncertainty of 2 3%. Both of these recommendations presumably refer to the agreement between the calculated dose at a point with that that would be measured under the same irradiation conditions at the same point. As such, these recommendations are not directly transferable to monitor unit check procedures, which compare one method of calculation with another, although they do provide a context within which to suggest the minimum sensitivity. Ideally, a monitor unit check procedure should provide a reasonable guarantee that the recommended uncertainties 2,13 are the limiting uncertainties in the treatment preparation phase of a course of radiotherapy. Thus errors which the checking procedure can identify, viz. system misuse, output misinterpretation, and data transfer, should not, if present, result in a significant increase in the potential discrepancy between intended and delivered dose. Such a criterion will be met if, as with the checking procedure described here, the accuracy and precision of the method are numerically less than the same intrinsic characteristics of the treatment planning system. 2,13 The usefulness of independent checks of MU calculations is generally recognized and this study indicated that in typical situations it is possible to corroborate the calculations of dose to a reference point, performed by an advanced 3D planning system, using standard methods of manual dose calculation. At the same time one expects that with the advent of modern treatment techniques involving complex noncoplanar treatment geometries, intensity-modulated and dynamic irradiations, it will become increasingly difficult to check MU calculations using the traditional manual approach. As a possible solution, development of a standardized computerized system for verification of MU calculations has been suggested by the European Society for Therapeutic Radiology and Oncology ESTRO MU working group. 14 It is also expected that in vivo dosimetry will be playing an increasing role in verification of dose calculation in advanced 3D conformal treatments 15 where accurate manual calculations are not possible. CONCLUSION Acceptable sensitivity for the verification of monitor unit calculations performed by a 3D treatment planning system can be achieved, with care, using the standard system of dosimetric calculations. ACKNOWLEDGMENTS The authors wish to thank Sonja Desjardins for her assistance. The generous support of the Northern Cancer Research Foundation is also gratefully acknowledged. *Electronic address: kleszczynski@neorcc.on.ca 1 International Commission on Radiation Units and Measurements ICRU, Determination of absorbed dose in a patient irradiated by beams of X or gamma rays in radiotherapy procedures, ICRU Report 24 ICRU Publications, Washington, D.C., 1976. 2 B. J. Mijnheer, J. J. Battermann, and A. Wambersie, What degree of accuracy is required and can be achieved in photon and neutron therapy?, Radiother. Oncol. 8, 237 252 1987. 3 A. Wambersie, J. Van Dam, G. Hanks, B. J. Mijnheer, and J. J. Battermann, What accuracy is needed in dosimetry?, IAEA TECDOC 734, 11 35 1994. 4 P. Dunscombe, P. McGhee, and E. Lederer, Anthropomorphic phantom measurements for the validation of a treatment planning system, Phys. Med. Biol. 41, 399 411 1996. 5 G. Kutcher et al., Comprehensive QA for radiation oncology: Report of AAPM Radiation Therapy Committee Task Group 40, Med. Phys. 21, 581 618 1994.

125 K. W. Leszczynski and P. B. Dunscombe: Independent corroboration of monitor... 125 6 B. Fraass et al., AAPM Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning, Med. Phys. 25, 1773 1829 1998. 7 A. Dutreix, B. R. Bjarngard, A. Bridier, B. Mijnheer, J. E. Shaw, and H. Svensson, Monitor Unit Calculation for High Energy Photon Beams Garant, Leuven, 1997. 8 R. A. Dahl, E. C. McCullough, and D. E. Mellenberg, A quality assurance program for monitor unit calculators, Med. Phys. 17, 103 105 1990. 9 L. Duggan, T. Kron, S. Howlett, A. Skov, and P. O Brien, An independent check of treatment plan, prescription and dose calculation as a QA procedure, Radiother. Oncol. 42, 297 301 1997. 10 R. F. Hill, M. D. Perez, and W. A. Beckham, An independent check method of radiotherapy computer plan derived units, Australas. Phys. Eng. Sci. Med. 21, 79 84 1998. 11 F. M. Khan, The Physics of Radiation Therapy, 2nd ed. Williams and Wilkins, Baltimore, 1994. 12 International Commission on Radiation Units and Measurements ICRU, Prescribing, recording, and reporting photon beam therapy, ICRU Report 50 ICRU, Bethesda, MD, 1993. 13 J. Van Dyk, R. B. Barnett, J. E. Cygler, and P. C. Schragge, Commissioning and quality assurance of treatment planning computers, Int. J. Radiat. Oncol., Biol., Phys. 26, 261 273 1993. 14 J. Olofsson, D. Georg, M. Karlsson, A. Dutreix, and H. Svensson, Quality assurance of monitor unit calculations abstract, Radiother. Oncol. 51, Suppl. 1, 26 1999. 15 M. Essers and B. J. Mijnheer, In vivo dosimetry during external photon beam radiotherapy, Int. J. Radiat. Oncol., Biol., Phys. 43, 245 259 1999.