A Comparative Study of Supreme LMA Vs I-gel: Two Supraglottic Airway Devices in Short Surgical Procedures

Similar documents
ISSN X (Print) Research Article. *Corresponding author Dr. Souvik Saha

ORIGINAL ARTICLE. Shashank Chitmulwar, MD, Charulata Deshpande, MD, DA ABSTRACT. ANAESTHESIA, PAIN & INTENSIVE CARE

I - Gel Versus Cuffed Tracheal Tube in Elective Laparoscopic Cholecystectomy A Clinical Comparative Study

A Comparative Study of Two Disposable Supraglottic Devices in Diagnostic Laparoscopy in Gynecology

Citation British journal of anaesthesia, 104. pp ; 2010 is available onlin

JMSCR Vol 4 Issue 02 Page February 2016

Randomised comparison of the LMA Supreme with the I-Gel in spontaneously breathing anaesthetised adult patients

Comparison of Ease of Insertion and Hemodynamic Response to Lma with Propofol and Thiopentone.

Deposited on: 11 February 2010

Comparison of I-gel with Baska Mask Airway for Controlled Ventilation in Obese Patients Undergoing Ambulatory Surgery: A Prospective Randomized Trial

A comparative study between i-gel and classical laryngeal mask airway in elective surgery under general anaesthesia

Comparison of clinical performance of i-gel with laryngeal mask airway pro-seal in elective surgery in adults

I-gel vs cuffed tracheal tube during volume controlled ventilation in elective laparoscopic cholecystectomy

Dr. Ranjeet Rana De 1, Dr. Saurav Shekhar 2, Dr. D G Pathak 3, Dr. Harshwardhan 4, Dr. Shashank Dhiraj 5 1,2,4,5

A Comparative Study of Classic LMA and Proseal LMA in Paralyzed Anaesthetized Patients

Deposited on: 24 February 2009

International Journal of Health Sciences and Research ISSN:

RESPIRATION AND THE AIRWAY Comparison of the i-gel with the cuffed tracheal tube during pressure-controlled ventilation

A randomised comparison between Cobra PLA and classic laryngeal mask airway and laryngeal tube during mechanical ventilation for general anaesthesia

LMA Supreme Second Seal. Maintain the airway. Manage gastric contents. Meet NAP4 recommendations.

Received 1 November 2009 Revised 8 December 2009 Accepted 10 December 2009

COMPARATIVE CLINICAL EVALUATION OF PROSEAL LMA WITH I - GEL AIRWAY IN PATIENTS UNDERGOING ELECTIVE SURGICAL PROCEDURES UNDER GENERAL ANAESTHESIA

Comparison of efficacy of the Laryngeal tube with the Laryngeal mask airway in securing the upper airway

Comparison of the LMA Supreme vs the i-gel TM in paralysed patients undergoing gynaecological laparoscopic surgery with controlled ventilation*

ISPUB.COM. M Roberts, M Mani, A Wilkes, E Flavell, N Goodwin INTRODUCTION

Similar oropharyngeal leak pressures during anaesthesia with i-gel TM, LMA-ProSeal TM and LMA-Supreme TM Laryngeal Masks

AIRWAY MANAGEMENT AND VENTILATION

Evaluation of Baska Mask Performance in Laparoscopic Cholecystectomy

Use of the Intubating Laryngeal Mask Airway

Other methods for maintaining the airway (not definitive airway as still unprotected):

The Laryngeal Mask and Other Supraglottic Airways: Application to Clinical Airway Management

Comparison of the Berman Intubating Airway and the Williams Airway Intubator for fibreoptic orotracheal intubation in anaesthetised patients.

Comparison of the i-gel and the LMA-Unique laryngeal mask airway in patients with mild to moderate obesity during elective short-term surgery

Anatomy and Physiology. The airways can be divided in to parts namely: The upper airway. The lower airway.

Comparison of the clinical performance of i-gel, LMA Supreme and LMA ProSeal in elective surgery

Comparison of the Baska â mask with the single-use laryngeal mask airway in low-risk female patients undergoing ambulatory surgery

Comparative evaluation of Ambu AuraGain with ProSeal laryngeal mask airway in patients undergoing laparoscopic cholecystectomy

DIFFICULT AIRWAY MANAGMENT. Dr.N.SANTHOSH KUMAR MD ANESTHESIA (2 nd Yr)

I-GEL TM VS. AURAONCE TM LARYNGEAL MASK FOR GENERAL ANAESTHESIA WITH CONTROLLED VENTILATION IN PARALYZED PATIENTS

Comparative study of various supraglottic devices with clinical and fiber optic assessment in elective laparoscopic procedures

The LMA CTrach TM, a new laryngeal mask airway for endotracheal intubation under vision: evaluation in 100 patients

Comparison of the Hemodynamic Responses with. with LMA vs Endotracheal Intubation

Airway Management. Teeradej Kuptanon, MD

Success Rate of Airway Devices Insertion: Laryngeal Mask Airway Versus Supraglottic Gel Device

Comparative Evaluation of Supraglottic Devices i-gel and Fastrach LMA as Conduit for Blind Endotracheal Intubation

Research Article. Shital S. Ahire 1 *, Shweta Mhambrey 1, Sambharana Nayak 2. Received: 22 July 2016 Accepted: 08 August 2016

Original Research Article. Amol P. Singam 1, Arpita A. Jaiswal 2 *, Ashok R. Chaudhari 1

Advanced Airway Management. University of Colorado Medical School Rural Track

Comparisonof Classic- Laryngeal Mask Airway with I-Gel Airway,during Anaesthesia with Controlled Ventilation in Routine Surgical Procedures

A Clinical Comparative Study Of Evaluation Of Proseal LMA V/S I-GEL For Ease Of Insertion And Hemodynamic Stability; A Study Of 60 Cases

Comparison of the efficacies of I-gel TM and LMA-ProSeal TM for airway management in pediatric patients

REVISTA BRASILEIRA DE ANESTESIOLOGIA

Equipment: NRP algorithm, MRSOPA table, medication chart, SpO 2 table Warm

Veena Mathur, Deepak Garg, Neena Jain, Vivek Singhal, Arvind Khare, Surendra K. Sethi*

MD (Anaesthesiology) Title (Plan of Thesis) (Session )

Journal of Anesthesia & Clinical

A prospective randomized study comparing the efficacy of the LMA Classic TM, the

Cricoid pressure: useful or dangerous?

Blind Insertion Airway Devices (BIAD)

Observation of ventilation effects of I-gel TM, Supreme TM and Ambu AuraOnce TM with respiratory dynamics monitoring in small children

CONTINUING EDUCATION IN HONOR OF NORMAN TRIEGER, DMD, MD

J. Brimacombe, 1 L. Holyoake, 2 C. Keller, 3 J. Barry, 4 D. Mecklem, 4 A. Blinco 5 and K. Weidmann 5

Pharyngolaryngeal Morbidity with the Laryngeal Mask Airway in Spontaneously Breathing Patients

Recent Advances in Airway Management HA Convention 2014

Journal of Society of Anesthesiologists of Nepal

J of Evolution of Med and Dent Sci/ eissn , pissn / Vol. 3/ Issue 64/Nov24, 2014 Page 13923

Original Article INTRODUCTION. Abstract

Comparision of Hemodynamic Changes after Insertion of Classic Lma and Proseal Lma

Comparison between i-gel airway an the proseal laryngeal mask airway in pediatric patients undergoing general anesthesia

Comparative Study of laryngeal mask airway Supreme and laryngeal mask airway Classic in paralyzed patients

Akhilesh Agrawal*, Smita S. Lele, Bharati A. Tendolkar

University of Groningen

Study of Comparison of I-gel with Classic Laryngeal Mask Airway Regarding the Ease of Use and Clinical Performance

ORIGINAL RESEARCH ARTICLE

Airway/Breathing. Chapter 5

Use of the Aintree Intubation Catheter with the Laryngeal Mask Airway and a Fiberoptic Bronchoscope in a Patient with an Unexpected Difficult Airway

Airway/Breathing. Chapter 5

Airway Management. DFMRT Casualty Care Examination Course. Revision notes for. January Les Gordon

Original Contributions

The Pro-Seal LMAtm And The Tracheal Tube: A Comparison Of Events At Insertion Of The Airway Device

Unanticipated difficult tracheal intubation - during routine induction of anaesthesia in an adult patient

Comparison of the air-q ILA and the LMA-Fastrach in airway management during general anaesthesia

Optimal Size AMBU Laryngeal Mask Airway Among Asian Adult Population

Ingo Bergmann *, Thomas Allen Crozier, Markus Roessler, Hanna Schotola, Ashham Mansur, Benedikt Büttner, José Maria Hinz and Martin Bauer

Positive pressure ventilation with the laryngeal mask airway in non-paralysed patients: comparison of sevoflurane and propofol maintenance techniques

Veena Patodi*, Maina Singh, Surendra K. Sethi, Vini N. Depal, Neena Jain, Vijay Kumar

Basic Airway Management

i-gel TM supraglottic airway in clinical practice: a prospective observational multicentre study

Dose-dependent effectiveness of ketamine nebulisation in preventing postoperative sore throat due to tracheal intubation

student handbook BARS handbook September 2012.indd Front Cover 27/11/12 12:08 PM

Airway Anatomy. Soft palate. Hard palate. Nasopharynx. Tongue. Oropharynx. Hypopharynx. Thyroid cartilage

Airway Workshop Lecture. University of Ottawa

Asai and Shingu Table 1 Size selection and recommended cuff volumes Size Patients Body size Recommended cuff volumes (ml) Connector colour 0 Newborn <

Clinical Study McGrath Video Laryngoscope May Take a Longer Intubation Time Than Macintosh Laryngoscope

Comparison of Three Disposable Extraglottic Airway Devices in Spontaneously Breathing Adults

Laryngeal mask airway (LMA-ProSeal) malfunction causing acute airway obstruction

We will not be using the King LTS-D in our system!

RESPIRATION AND THE AIRWAY. Editor s key points. L. Theiler 1,2 *, M. Kleine-Brueggeney 1,3, N. Urwyler 1,4,T.Graf 1, C. Luyet 1 and R.

Airway Management & Safety Concerns Experience from Bariatric Surgery

ISPUB.COM. The Use of LMA in Newborn Resuscitation. R Vadhera INTRODUCTION VENTILATION

Transcription:

Original Research Article MVP Journal of Medical Sciences, Vol 5(1), 5 11, January-June 2018 ISSN (Print) : 2348 263X ISSN (Online) : 2348-2648 DOI: 10.18311/mvpjms/2018/v5i1/20044 A Comparative Study of Supreme LMA Vs I-gel: Two Supraglottic Airway Devices in Short Surgical Procedures Balasaheb Tukaram Govardhane 1, Shantanu Kulkarni 2*, Mukesh Parmar 3 1,2 Assistant Professor, Department of Anaesthesia, Lokmanya Tilak Municipal Medical College & General Hospital, Mumbai - 400022, Maharashtra, India; drbgovardhane@gmail.com, drshan70@rediffmail.com 3 Resident, Department of Anaesthesia, Lokmanya Tilak Municipal Medical College & General Hospital, Mumbai - 400022, Maharashtra, India; dr.mukesh_parmar@indiatimes.com Abstract Background and Aims: The objective of this prospective, randomized trial was to compare I-Gel with Supreme LMA in anaesthetized spontaneously breathing patients for short surgical procedures. Material and Methods: Sixty patients of ASAI -ASAII of either gender between 18-60 years undergoing short surgical procedures were randomly assigned to I-gel (Group I) or LMA-Supreme (Group S). After induction with propofol the supraglottic airway device was inserted. We compared the ease and time required for insertion, airway sealing pressure and adverse events related to airway. Results: There were no significant differences in demographic and haemodynamic data. I-gel insertion was easier than LMA Supreme but statistically not significant (p > 0.05) (Chi square test). Numbers of attempts for successful insertions were comparable and in majority device was inserted in first attempt. Although the airway sealing pressure was significantly higher with Group S (25.73+2.21 cm of H 2 O), the airway sealing pressure of Group I (20.0+2.94 cm of H 2 O) was very well within normal limit (Student s t test). There was no evidence of airway trauma, regurgitation and aspiration. Conclusion: I-Gel with acceptable airway sealing pressure, easier to insert, less traumatic with lower incidence of sore throat. Hence I-Gel can be a good alternative to LMA-Supreme. Keywords: Adverse Events, I-gel, Supreme LMA, Seal Pressure, Time for Insertion 1. Introduction During the past decade, several Supraglottic Airway Devices (SAD) have been introduced for airway management as an alternative to tracheal intubation in general anaesthesia and in cardiopulmonary resuscitataion. These devices have become popular because of their ability to maintain airway without perturbing the trachea and can be used in patients without muscle relaxation. Both the i-gel (i-gel) and LMA Supreme (Supreme) are new, single-use, second-generation, Supraglottic Airway Devices (SAD). The i-gel (Intersurgical Ltd., Wokingham, Berkshire, UK) is a latex-free SAD with a non-inflatable cuff and a gastric drain tube. The Supreme (The Laryngeal Mask Company Ltd., St Helier, Jersey, UK) has a curved and rigid airway tube, a drain tube positioned within the center of the airway tube and a relatively large inflatable cuff made of polyvinyl chloride 1 5. There are many studies comparing the Supreme with other SADs that have shown effective clinical performances in adults and children 4 6. These two SADs which provide higher airway leak pressure than the classic LMA (clma) and can be used for spontaneous as well as Positive Pressure Ventilation (PPV). *Author for correspondence

A Comparative Study of Supreme LMA Vs I-gel: Two Supraglottic Airway Devices in Short Surgical Procedures Though studies have been done comparing i-gel with SLMA, we proposed to compare the two devices, especially with regards ease, time taken for insertion, number of attempts, airway sealing pressure and occurrence of airway complications. 2. Materials and Methods This prospective, randomized, comparative study was conducted after obtaining approval from the institutional ethical committee. American Society of Anesthesiologists physical status I and II of either gender aged 18-60 years, with body mass index 18-30 kgm 2 undergoing elective short surgical procedures (duration between 30 and 120 min) requiring general anesthesia without muscle relaxation were included. Patients with anticipated difficult airway, restricted mouth opening, pregnant females, cervical spine disease, obese with body mass index > 30 kg/m 2 and patients with history of regurgitation were excluded from the study. Sixty patients were randomly allocated into two groups each based on the computer generated codes. Group I in whom I-Gel was inserted and Group S in whom LMA-Supreme was inserted. Anesthesiologist in post-anesthesia care unit monitoring the post-operative parameters and incidence of sore throat and the patient were all blinded to the group assignment. All the patients were evaluated pre-operatively and all the baseline vitals were recorded. The randomly selected patients for the study were pre-medicated with the Inj. Glycopyrrolate 0.2 mg, Inj. Midazolam 0.02 mg/kg and Inj. Fentanyl 2 mcg/kg intravenously. All patients pre - oxygenated with 100% oxygen for 3 minutes. Each patient then received induction dose of Inj. Propofol (2-2.5 mg/kg iv) till the loss of eyelash reflex. The patient s head was placed in sniffing the morning air position, device was inserted by a qualified anesthesiologist with minimum 2 year s experience. The supraglottic airway device was inserted after lubricating posterior surface of the cuff with a water based jelly. LMA-Supreme cuff was inflated with half the recommended volume of air and in case of inadequate seal entire recommended volume of air was used to inflate the cuff. In case of further leak, the device was removed and one size bigger was inserted. Additional doses of propofol were used in case of reinsertion if required. The device was then connected to the breathing circuit and capnometer assembly and secured after confirming bilaterally equal air entry. Nasogastric tube was inserted. An effective airway was confirmed from the bilaterally symmetrical chest movements, square wave form on the capnograph and a normal saturation. Ease of insertion was defined as no resistance to insertion of device in the pharynx in single attempt. The time taken for the insertion of device was noted. It was the time from the end of the propofol bolus to the connection of the airway to the breathing circuit. The airway seal pressure was measured after closing the adjustable pressure limiting valve with a fresh gas flow of 3 lit/min, noting the airway pressure at equilibrium or when there was an audible air leak from the throat. The maximum pressure allowed was 40cm H 2 O. The epigastrium was also auscultated when measuring the oropharyngeal leak pressure to detect any air entrainment in the stomach. If an effective airway was not achieved then manipulations were done in the form of increasing the depth of insertion, giving jaw thrust or chin lift or changing the size of the device. The device insertion was abandoned after 3 unsuccessful attempts. Then patient was given muscle relaxant and was intubated with endotracheal tube. Anaesthesia was maintained on oxygen, nitrous oxide and propofol infusion with spontaneous respiration. At end of the procedure, all the patients were ventilated with 100% oxygen during emergence from anesthesia. The device was removed when the patient was able to open the mouth on command. The patient was inspected for any injury to lips, teeth or tongue and the device was inspected for the presence of any blood stains. The mask of the supraglottic device was inspected for the presence of any gastric contents to confirm regurgitation. All the patients were observed for a period of 24 h for any complaints of sore throat. Sore throat in the postoperative period was treated using warm saline nebulization and in patients with sore throat 24 h later warm saline gargles were advised. Laryngospasm was given standard treatment. We gave 100% oxygen followed by injection scoline 0.5 mg/kg. Hiccups were tackled by increasing the depth of anesthesia by increasing the maintenance dose of Injection propofol. 3. Statistical Analysis Sample size was calculated based on the results of previous study to detect a projected difference in airway 6 MVP Journal of Medical Sciences

Balasaheb Tukaram Govardhane, Shantanu Kulkarni, Mukesh Parmar sealing pressure of 30% between groups with 80% power and 5% alpha error and a reported difference in airway sealing pressure of 15% between two groups, a sample size 22 patients were required, which was rounded off to 30 patients in each group. Table 1. Patient characteristics: Data are expressed as mean ± standard deviation for age and body mass index; and absolute number for gender, MPC and ASA status Parameters Group I Group S *Age (yrs) 32.07 ± 11.42 33.20 ± 9.14 *Weight (kg) 51.73±7.77 53.23± 3.02 @Sex 13:17 20:10 @MPC 21:09 26:04 @ASA 26:04 26:04 *Student t test P > 0.05, Not Significant @Chi square test P > 0.05, Not Significant The two groups were compared with each other in terms of age, weight and sex. The statistical test used was Unpaired Students t test for age and weight. For qualitative data like the sex of the patient the statistical test employed was Pearson s Chi-square test. Hemodynamic parameters such as mean heart rate, blood pressure both systolic and diastolic, respiratory rate, SpO 2 and end tidal CO 2 were compared using Analysis of Variance ( ANOVA ). The mean time required for insertion and the mean seal pressure was compared using Unpaired Student s t test. The ease of insertion, attempts required for insertion, airway manipulations and the incidence of adverse events were compared using Pearson s Chi-Square test. In all the parameters, p < 0.05 was considered to be significant. Table 2. Comparison of Time for successful insertion, ease of insertion, insertion attempts, ease of RT insertion and airway sealing pressure Variables Group I Group S P value Mean time for Successful 29.53 ± 8.23 31.77 ± 2.38 > 0.05 insertion Ease of LMA insertion Yes No 28 2 27 3 > 0.05 Insertion Attempts 1 2 Ease of Ryle s Tube insertion Easy Difficult 28 2 30 0 27 3 30 0 > 0.05 > 0.05 Mean Airway Sealing Pressure 20.07 + 2.94 25.73 + 2.21 0.0000 Table 3. Profile of adverse events Events Group I Group S P value Coughing 02 01 > 0.05 Laryngospasm - 01 Leak 03 01 Regurgitation - - Injury to teeth, gum and - - lips Sore throat - 03 Aspiration - - Blood on device - - Gastric insufflation - - No. of patients 05 06 MVP Journal of Medical Sciences 7

A Comparative Study of Supreme LMA Vs I-gel: Two Supraglottic Airway Devices in Short Surgical Procedures cases was less as compared to 19.9% in Supreme LMA group but the difference was not statistically significant (Table 3). Out of these most common was sore throat and leak followed by coughing, laryngospasm. Figure 1. Comparison of changes in mean heart rate between two groups. Figure 3. Comparison of changes in mean ETCO 2 between two groups. 5. Discussion Figure 2. Comparison of changes in mean systolic pressure between two groups. 4. Results The demographic profile is comparable in two groups (Table 1). In both groups the mean heart rate (Figure 1), other hemodynamic parameters (Figure 2) and end-tidal CO 2 (Figure 3) were comparable. In both I-gel and Supreme-LMA groups, mean time for successful insertion, ease of insertion, insertion attempts and ease of ryle s tube insertion were comparable (Table 2). Higher airway sealing pressure was noted with Supreme LMA (25.73+2.21) as compared to I-gel (20.07+2.94) and difference was statistically significant (P<0.0001) (Table 2). This analysis reveals that 16.7% of the total cases among I-Gel group had an adverse event where the proportion of In both the groups, the mean heart rate was comparable and no statistically significant difference was observed. Also the systolic blood pressure and diastolic blood pressure difference was comparable in both groups and was not statistically as well as clinically significant. There was no episode of de-saturation and hypotension throughout the surgery. Our results of hemodynamic parameters were in concordance with those reported by Amr Helmy et al. and W. H. L. Teoh et al 6 8. The mean time required for insertion of I-GEL was 29.53+08.23 sec and for Supreme LMA was 31.77+02.38 sec which was statistically insignificant (p = 0.165). Tuoh et al, also found similar results. Both devices had similar first attempt insertion rates (LMAS 94% vs. I-GEL 91%) with similar ease and comparable times to achieve an effective airway, LMA Supreme 14.3 (4.7) versus I-GEL 15.4 (8.2), P = 0.4 6,9. In our study, insertion was easy in 28/30 cases (93.3%) as compared to 27/30 cases (90%) in the Supreme LMA. The difference was statistically as well as clinically insignificant. Similar results were demonstrated by W. H. L. Teoh et. al. 6 In our study ease of Ryle`s Tube insertion for both the devices were comparable and the difference was statistically 8 MVP Journal of Medical Sciences

Balasaheb Tukaram Govardhane, Shantanu Kulkarni, Mukesh Parmar as well as clinically insignificant. But Teoh et al. found that it was more difficult and took significantly longer to insert the gastric tube in the I-Gel group (78%) than Supreme LMA group (100%) 6. Ragazzi et al. and Suhitharann et al. also found similar results, I-Gel (77%)/Supreme LMA (100%) and I-Gel (73%)/Supreme LMA (87%) 5,9. The mean airway sealing pressure with I-Gel was 20.07+02.94 cm of H 2 O and with Supreme LMA 25.73 + 02.21cm of H 2 O which was statistically significant (p = 0.0000*) but it was not clinically significant. The higher values of EtCO 2 among the LMA-Supreme group can explain the high airway sealing pressure and a better seal provided by it. Though the seal pressure of I-gel was lower than that of LMA-Supreme, it was enough to provide optimum ventilation. Chew et al. also reported higher seal pressure with Supreme LMA (25.6 cm H 2 O) than I-Gel (20.7 cm of H 2 O) 10. Ragazzi R et al. also noted that the airway sealing pressure provide by Supreme LMA (28 cm of H 2 O) was higher than that with the I-GEL (24 cm of H 2 O) 5. However Teoh WH et al. found no difference in seal pressure, between the Supreme LMA and the I-Gel (mean (SD) 26.4 (5.1) vs 25.0 (5.7) cm H 2 O, respectively; p = 0.18) 6. Gabbot et al. also concluded that I-Gel provides a good airway sealing pressure which improved over time and may be due to the thermoplastic properties of gel cuff which forms a effective seal around the larynx after warming to body temperature 11. Richez et al. and Kannaujia et al. both had performed preliminary studies on I-Gel and determined the airway seal pressure to be (30±7 cm of H 2 O) and oropharyngeal seal pressure was 20 cm of H 2 O (range 16-40 cm of H 2 O) respectively. 12,13 We compared the incidence of adverse events intraoperatively, during emergence and in the postoperative period. Coughing was noted in 2/30 (6.7%) cases of I-Gel whereas there was only 1/30 (3.3%) case of coughing with Supreme LMA. The difference was statistically insignificant (p > 0.05) and the complaints were relieved after warm saline nebulisation. Cook et al. reported that 2 out of 100 patients developed coughing without a fall in arterial oxygen saturation 14. Laryngospasm was observed in 1/30 (3.3%) of the Supreme LMA case during emergence, though it was statistically insignificant (p > 0.05) it was clinically significant. The incidence of laryngospasm can be attributed to the lighter plane of anaesthesia during the end of the procedure. None of the patients in I-Gel group had laryngospasm 14. Intra-operatively leak was present in 3/30 (10.0%) cases of the I-Gel and 1/30 (3.3%) cases with the Supreme LMA. The difference was statistically as well as clinically insignificant. Hosten et al. reported reported the incidence of 2/30 patients with intra-operative oropharyngeal leak 15. Suhitharan et al. found that there was a significantly greater leak fraction with the I-Gel of 0.06 (0.03) versus 0.04 (0.02) with the LMAS, P = 0.013 9. There was no evidence of injury to lip, teeth, tongue or gums and blood on device with either device. Yao et al., reported 2.6% patients had on supreme on removal 16. Russo et al., reported the incidence of blood stains on Supreme LMA, I-Gel and Laryngeal Tube Suction-D 17. Suhitharan et al., reported 2.9% patients had mucosal injury 9. There was no incidence of gastric insufflation with either device probably due to a good seal around the laryngeal inlet and presence of nasogastric tube through the gastric channel. Hosten et al., reported the incidence of 3/30 patients with intra-operative gastric insufflation 15. In the Supreme LMA group 3/30 (10.0%) case complained of sore throat immediately in the postoperative period whereas in the I-Gel group 1/30 (3.3%) patients complained of sore throat. Though the difference was not statistically significant it was clinically significant. According to the 3-point scale all the patients had class 1 sore throat that is just throat discomfort, which was relieved to some extent by warm saline nebulisation. 24 hours later the same patient had throat discomfort in Supreme LMA group where as in the I-Gel group there was NO throat discomfort. Teoh et al. and Ragazzi et al. also found that the use of Supreme LMA produces more sore throat as compared to the I-Gel 5 7,13,18,19. The lower incidence of sore throat in our study can be attributed to the soft seal non inflatable mask of I-GEL. I-GEL being a supraglottic airway device without an inflatable mask has some potential advantage of easier insertion and minimal tissue compression whereas supraglottic airway device with inflatable cuff like the Supreme LMA in our study can absorb anaesthetic gases leading to increased mucosal pressure 20,21. Hosten T et al. conducted a randomized cross-over study of Proseal LMA and Supreme LMA in 60 adult patients. He observed that both Proseal LMA and Supreme LMA show significant increase in intra-cuff pressure (p < 0.05) 15. MVP Journal of Medical Sciences 9

A Comparative Study of Supreme LMA Vs I-gel: Two Supraglottic Airway Devices in Short Surgical Procedures 6. Conclusion Hence to conclude, I-Gel is a simple and safe supraglottic airway device made of a soft gel like material. It has potential advantage of providing an effective airway sealing pressure which was within normal limit and sufficient to prevent aspiration. It has a gastric channel providing a means to drain the gastric secretions. Being made of a soft gel like material and the presence of non inflatable mask makes it less irritant to the airway and hence there was no incidence of laryngospasm and comparatively lower incidence of sore throat. Hence I-Gel can be a good alternative to Supreme LMA and other supraglottic airway devices. 7. References 1. Uppal V, Gangaiah S, Fletcher G, Kinsella J. Randomized crossover comparison between the i-gel and the LMA-Unique in anaesthetized, paralysed adults. Br J Anaesth. 2009; 103:882 5. https://doi.org/10.1093/bja/aep292 PMid:19846406 2. Francksen H, Renner J, Hanss R, Scholz J, Doerges V, Bein B. A comparison of the i-gel with the LMA-Unique in nonparalysed anaesthetised adult patients. Anaesthesia. 2009; 64:1118 24. https://doi.org/10.1111/j.1365-2044.2009.06017.x PMid:19735404 3. Uppal V, Fletcher G, Kinsella J. Comparison of the i-gel with the cuffed tracheal tube during pressure-controlled ventilation. Br J Anaesth. 2009; 102:264 8. https://doi. org/10.1093/bja/aen366 PMid:19151051 4. heiler LG, Kleine-Brueggeney M, Kaiser D, Urwyler N, Luyet C, Vogt A, et al. Crossover comparison of the laryngeal mask supreme and the i-gel in simulated difficult airway scenario in anesthetized patients. Anesthesiology. 2009; 111:55 62. https://doi.org/10.1097/aln.0b013e3181a4c6b9 PMid:19512881 5. Ragazzi R, Finessi L, Farinelli I, Alvisi R, Volta CA. LMA SupremeTM vs i-geltm- a comparison of insertion success in novices. Anaesthesia. 2012; 67:384 8. https://doi. org/10.1111/j.1365-2044.2011.07002.x PMid:22329593 6. Teoh WH, Lee KM, Suhitharan T, Yahaya Z, Teo MM, Sia AT. Comparison of the LMA Supreme vs thei-gel in paralysed patients undergoing gynaecological laparoscopic surgery with controlled ventilation. Anaesthesia. 2010; 65:1173 9. https:// doi.org/10.1111/j.1365-2044.2010.06534.x PMid:20958278 7. Helmy A, Atef HM, El Taher E, Henidak A. Comparative study between I-gel, a new supraglottic airway device, and classical laryngeal mask airway in anesthetized spontaneously ventilated patients. Saudi J Anaesth. 2010; 4:131 6. https://doi.org/10.4103/1658-354x.71250 PMid:21189847 PMCid:PMC2980656 8. Gasteiger L, Brimacombe J, Perkhofer D, Kaufmann M, Keller C. Comparison of guided insertion of the LMA ProsealTM vs the i-gel TM. Anaesthesia. 2010; 65:913. https://doi. org/10.1111/j.1365-2044.2010.06422.x PMid:20645948 9. Suhitharan T, Teoh WH. Use of extraglottic airways in patients undergoing ambulatory laparoscopic surgery without the need for tracheal intubation. Saudi J Anaesth. 2013; 7:436 41. https://doi.org/10.4103/1658-354x.121081 PMid:24348297 PMCid:PMC3858696 10. Chew EE, Hashim NH, Wang CY. Randomised comparison of the Supreme LMA with the I-Gel in spontaneously breathing anaesthetised adult patients. Anaesth Intensive Care. 2010; 38:1018 22. PMid:21226431 11. Gabbot D, Beriger R. The I-gel supraglottic airway: A potential role for resuscitation. Resuscitation. 2007; 73:161 4. https:// doi.org/10.1016/j.resuscitation.2006.10.026 PMid:17289250 12. Richez B, Saltel L, Banchereau F, Torrielli R, Cros AM. A new single use supraglottic airway device with a noninflatable cuff and an esophageal vent: An observational study of the i-gel. Anesth Analg. 2008; 106:1137 9. https://doi. org/10.1213/ane.0b013e318164f062 PMid:18349185 13. Kannaujia A, Srivastava U, Saraswat N, Mishra A, Kumar A, Saxena S. A preliminary study of I-gel: A new supraglottic airway device. Indian J Anaesth. 2009; 53:52 6. PMid:20640078 PMCid:PMC2900034 14. Cook TM, Gatward JJ, Handel J, Hardy R, Thompson C, Srivastava R, Clarke PA. Evaluation of the supreme LMA in 100 non-paralysed patients. Anaesthesia. 2009; 64:555 62. https:// doi.org/10.1111/j.1365-2044.2008.05824.x PMid:19413827 15. Hosten T, Gurkan Y, Ozdamar D, Tekin M, Toker K, Solak M. A new supraglottic airway device: LMA-supreme, comparison with LMA-Proseal. Acta Anaesthesiol Scand. 2009; 53:852 7. https://doi.org/10.1111/j.1399-6576.2009.01986.x PMid:19426239 16. Yao WY, Li SY, Sng BL, Lim Y, Sia AT. The Supreme LMA in 700 parturients undergoing Cesarean delivery: An observational study. Can J Anaesth. 2012; 59:648 54. https://doi. org/10.1007/s12630-012-9718-4 PMid:22556005 17. Russo SG, Cremer S, Galli T, Eich C, Brauer A, Crozier TA, Bauer M, Strack M. Randomized comparison of the I-Gel, the supreme LMA, and the laryngeal tube suction-d using clinical and fibreoptic assessments in elective patients. BMC Anesthesiol. 2012 ; 7:12 8. https://doi.org/10.1186/1471-2253-12-18 18. Gatward JJ, Cook TM, Seller C, Handel J, Simpson T, Vanek V et al. Evaluation of the size 4 i-geltm airway in one hundred non-paralysed patients. Anaesthesia. 2008; 63:1124 30. https:// doi.org/10.1111/j.1365-2044.2008.05561.x PMid:18616521 10 MVP Journal of Medical Sciences

Balasaheb Tukaram Govardhane, Shantanu Kulkarni, Mukesh Parmar 19. Keijzer C, Buitelaar D, Efthymiou K, Sramek M, Cate J, Ronday M et al. A comparison of postoperative sore throat and neck complaints after the use of the I-gel and the La Premiere disposable laryngeal mask: A double-blinded, randomized, controlled trial. Anesth Analg. 2009; 109:1092 4. https://doi.org/10.1213/ane.0b013e3181b6496a PMid:19641052 20. Twing S, Brown JM, Williams R. Swelling and cyanosis of tongue associated with use of laryngeal mask airway. Anaesthesia and Intensive Care. 2000; 28:449 50. 21. Ouellete RG. The effect of nitrous oxide on laryngeal mask cuff pressure. American Association of Nurse Anesthetist Journal. 2000; 68:411 4. Cite this article as: Govardhane BT, Kulkarni S, Parmar M. A Comparative Study of Supreme LMA Vs I-gel: Two Supraglottic Airway Devices in Short Surgical Procedures. MVP Journal of Medical Sciences 2018; 5(1):5-11. MVP Journal of Medical Sciences 11