The Vista System: A New Bipolar Resection Device for Endourological Procedures: Comparison with Conventional Resectoscope

Similar documents
980-nm Diode Laser: A Novel Laser Technology for Vaporization of the Prostate

Comparison of outpatient versus inpatient transurethral prostate resection for benign prostatic hyperplasia: Comparative, prospective bi-centre study

Early-Stage Clinical Experiences of Holmium Laser Enucleation of the Prostate (HoLEP)

Transurethral Resection of Prostate (TURP) Through The Decades A Comparison of Results Over the Last Thirty Years in a Single Institution in Asia

Original Article - Lasers in Urology. Min Ho Lee, Hee Jo Yang, Doo Sang Kim, Chang Ho Lee, Youn Soo Jeon

Voiding Dysfunction. Joon Seok Kwon, Jung Woo Lee 1, Seung Wook Lee, Hong Yong Choi, Hong Sang Moon. DOI: /kju

TRANSURETHRAL PLASMA VAPORIZATION OF THE PROSTATE Procedure Guide

IS IRRIGATION NECESSARY AFTER MONOPOLAR TURP? OUR 11 YEARS EXPERIENCE

Benign prostatic hyperplasia (BPH) is one of the

Experience the Innovative Therapy for Benign Prostate Enlargement

Executive Summary. Non-drug local procedures for treatment of benign prostatic hyperplasia 1. IQWiG Reports - Commission No.

Benign enlargement of prostate (BEP), which is. Early Experiences with HoLEP. Original Article ABSTRACT INTRODUCTION. Bhandari BB*

Index. urologic.theclinics.com. Note: Page numbers of article titles are in boldface type.

Original Policy Date

ISPUB.COM. Photoselective Vaporisation of the Prostate - Independent Surgical Experience Following Comprehensive Resident Training in the Technique.

NOTE: This policy is not effective until April 1, Transurethral Water Vapor Thermal Therapy of the Prostate

Control & confidence. You deserve both. YOUR GUIDE TO THE TREATMENT OF BPH

Control & confidence. You deserve both.

RFA is an attractive, heat energy based ablation modality

Submitted: August 15, Accepted: September 10, Published: October 15, 2018.

Index. urologic.theclinics.com. Note: Page numbers of article titles are in boldface type.

Bipolar Resectoscopes GYN /2014/EW-E

Rezūm procedure for the Prostate

Surgical Treatment of LUTS in Men with BPE

Initial Experience of High Power Diode Laser for Vaporization of Prostate Muhammad Rafiq Zaki, Mujahid Hussain, Tahir Mehmood, Murtaza Hiraj

Bipolar Transurethral Resection in Saline An Alternative Surgical Treatment for Bladder Outlet Obstruction?

Name of Policy: Transurethral Microwave Thermotherapy

Changes in Surgical Strategy for Patients with Benign Prostatic Hyperplasia: 12-Year Single-Center Experience

BENIGN PROSTATIC HYPERPLASIA (BPH) affects 70%

European Urology 44 (2003) 89 93

Yong Taec Lee, Young Woo Ryu, Dong Min Lee, Sang Wook Park, Seung Hee Yum, June Hyun Han

Surgical procedures for benign prostatic hyperplasia: A nationwide survey in Japan

Combination of Thulium Laser Incision and Bipolar Resection Offers Higher Resection. Velocity than Bipolar Resection Alone in Large Prostates

Thulium Laser versus Standard Transurethral Resection of the Prostate: A Randomized Prospective Trial

POLICIES AND PROCEDURE MANUAL

Title:Transurethral Cystolitholapaxy with the AH -1 Stone Removal System for the Treatment of Bladder Stones of Variable Size

Outcomes of the Holmium Laser Enucleation of the Prostate for Patients With Prior Benign Prostatic Hyperplasia Surgery

Treating BPH: Comparing Rezum UroLift and HoLEP

Cheng Liu, Xuejun Liu, Duo Liu, Jiagui Mu, Dongsheng Zhu, Li Fan, Qun Song, Dongwei Yao *

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

Benign Prostatic Hyperplasia (BPH)

Novel low-cost prostate resection trainer description and preliminary evaluation

What is Benign Prostatic Hyperplasia (BPH)?

W J C U. World Journal of Clinical Urology. Transurethral bipolar prostatectomy: Where do we stand now? INTRODUCTION. Abstract MINIREVIEWS

Increasing life expectancy has resulted in an increase in

Minimally InvasiveTreatment of BPH: An Update

MODULE 3: BENIGN PROSTATIC HYPERTROPHY

ABSTRACT KEY WORDS. Original Article. Page 163. Background

Early outcome of transurethral enucleation and resection of the prostate versus transurethral resection of the prostate

Can Bipolar Vaporization be Considered an Alternative Energy Source in the Endoscopic Treatment of Urethral Strictures and Bladder Neck Contracture?

Wednesday 25 June Poster Session 9: BPH 1 Chairmen: M. Speakman and D. Rosario

Complications of Diode Laser in Endourological Procedures in Co-morbid Patients

Transurethral incision versus transurethral resection of the prostate in small prostatic adenoma: Long-term follow-up

PROSTATIC ARTERY EMBOLISATION (PAE) FOR BENIGN PROSTATIC HYPERPLASIA. A Minimally Invasive Innovative Treatment

Bipolar Transurethral Resection of the Prostate: Perioperative Safety and Functional Outcome: Prospective Randomized Study

EUROPEAN UROLOGY 58 (2010)

Prostatic Diseases and Male Voiding Dysfunction

Lasers in Urology. Hyeon Jun Kim, Han Yi Lee, Sang Hun Song 1, Jae-Seung Paick.

GUIDELINES ON NON-NEUROGENIC MALE LUTS INCLUDING BENIGN PROSTATIC OBSTRUCTION (BPO)

TURP: How Much Should Be Resected? International Braz J Urol Vol. 35 (6): , November - December, 2009 doi: /S

Original Article Holmium laser enucleation of the prostate prevents postoperative stress incontinence in patients with benign prostate hyperplasia

Citation Acta medica Nagasakiensia. 1997, 42

A new experience of speed

TRANSURETHRAL RESECTION

Randomized, controlled trial of laser vs. bipolar plasma vaporization treatment of benign prostatic hyperplasia

Agenda of Product Presenta0on. I. Introduc*on II. Laser Physics III. Key Advantages IV. Clinical Data V. Key Takeaways

INJINTERNATIONAL. Original Article INTRODUCTION. Int Neurourol J 2017;21: pissn eissn

Clinical Study Long-Term Followup after Electrocautery Transurethral Resection of the Prostate for Benign Prostatic Hyperplasia

TURP Complications & Treatments. G. Testa

An Anteriorly Positioned Midline Prostatic Cyst Resulting in Lower Urinary Tract Symptoms

SURGICAL MANAGEMENT OF BPH IN GHANA: A NEED TO IMPROVE ACCESS TO TRANSURETHRAL RESECTION OF THE PROSTATE

The Enlarged Prostate Symptoms, Diagnosis and Treatment

Nikolaos Mertziotis, 1 Diomidis Kozyrakis, 2 Christos Kyratsas, 1 and Andreas Konandreas Introduction

Electrolyte Changes in Monopolar and Bipolar Transurethral Resection of Prostate (TURP) A Prospective Randomized Study

HOLMIUM:YAG LASER WHEN POWERFUL VERSATILITY INTEGRATES UNMATCHED PRECISION.

Public Statement: Medical Policy Statement:

UROLOGY COMPARISION OF EFFECTIVENESS OF IRRIGATION FLUID MANNITOL 5% WITH THAT OF GLYCINE 1.5% IN PREVENTING POST TURP HYPONATREMIA MEDICAL CHANNEL

Management of LUTS after TURP and MIT

EAU GUIDELINES POCKET EDITION 3

Transurethral Laser Technology: Treatments for BPH

PROSTATIC EMBOLIZATION FOR BENIGN HYPERPLASIA

Kathmandu University Medical Journal (2008), Vol. 6, No. 3, Issue 23,

Safety and efficacy of photoselective vaporization of the prostate using the 180-W GreenLight XPS laser system in patients taking oral anticoagulants

AMERIHEALTH / INDEPENDENCE BLUE CROSS

Transurethral Resection of Prostate (TURP)

Can 80 W KTP Laser Vaporization Effectively Relieve the Obstruction in Benign Prostatic Hyperplasia?: A Nonrandomized Trial

Overview. Urology Dine and Learn: Erectile Dysfunction & Benign Prostatic Hyperplasia. Iain McAuley September 15, 2014

HOLMIUM:YAG LASER WHEN POWERFUL VERSATILITY INTEGRATES UNMATCHED PRECISION. Main Application Fields UROLOGY GENERAL SURGERY ENT

Effect of Transurethral Resection of the Prostate Based on the Degree of Obstruction Seen in Urodynamic Study

Biomedical Research 2017; 28 (12):

Roberto Giulianelli, Barbara Gentile, Luca Albanesi, Paola Tariciotti, and Gabriella Mirabile

UROLOGY INSTRUMENTS. - / Web:- Maryland, USA : +1(202) /

Holmium Laser Enucleation of the Prostate: Modified Morcellation Technique and Results

Int J Clin Exp Med 2016;9(4): /ISSN: /IJCEM

TITLE: Plasma Vaporization of the Prostate for Treatment of Benign Prostatic Hypertrophy: A Review of Clinical and Cost-Effectiveness, and Safety

Elective Hemi Transurethral Resection of Prostate: A Safe and Effective Method of Treating Huge Benign Prostatic Hyperplasia

Chapter 2: Methodology

Long-term Follow-up of Transurethral Enucleation Resection of the Prostate for Symptomatic Benign Prostatic Hyperplasia

Transcription:

European Urology European Urology 46 (2004) 586 590 The Vista System: A New Bipolar Resection Device for Endourological Procedures: Comparison with Conventional Resectoscope Gunnar Wendt-Nordahl a, *,AxelHäcker a, Oliver Reich b, Bob Djavan c, Peter Alken a, Maurice Stephan Michel a a Department of Urology, University Hospital Mannheim, Theodor-Kutzer-Ufer 1 3, 68135 Mannheim, Germany b Department of Urology, University Hospital Klinikum Grosshadern, 81377 Munich, Germany c Department of Urology, University of Vienna, Wahringer Gurtel 18 20, 1090 Vienna, Austria Accepted 22 July 2004 Available online 12 August 2004 Abstract Objective: Conventional transurethral resection of the prostate (TURP) uses a monopolar electrocautery system in which the current passes from the active electrode through the patient s body towards the return plate and may cause distant negative effects. In this study a new developed resection device, the Vista system, using a bipolar electrocautery system and 0.9% sodium chloride solution for irrigation, was evaluated in an ex-vivo model. Methods: The modified model of the isolated blood perfused kidney was used to determine cutting qualities, ablation rate, blood loss and coagulation depth of the bipolar resectoscope. After ablating the renal tissue of a perfused kidney in a surface area, blood loss was semiquantitatively determined. Afterwards samples were taken and processed for histological evaluation of the coagulation depth. We compared the new bipolar resection device against a conventional monopolar resectoscope. Results: We found good cutting qualities of the bipolar resectoscope although it is more difficult to start a cut. The ablation rate is determined by the width of the electrode and is similar to the standard device (30 cm 2 /min). The bleeding is reduced with increasing output powers (26.13 6.15 g/min (level 5); 20.49 5,47 g/min (level 6); 13.16 5,47 g/min (level 7); 10.43 4.76 g/min (level 8) and lower compared to a conventional monopolar resectoscope (17.08 4.47 g/min). The coagulation depth increases with higher output powers but is reduced compared to the standard device (118 22 mm (level 5); 121 23 mm (level 6); 141 62 mm (level 7); 163 30 mm (level 8) versus 287 57 (monopolar resectoscope)). Conclusion: Our results with the bipolar resection device for TURP suggest that it may offer an alternative to conventional TURP. As active and the return electrode are placed on the resectoscope, high current densities are achieved locally and complications caused by distant negative effects of the current are theoretically reduced in vivo. Furthermore the risk of TUR syndrome is theoretically eliminated by using physiological sodium chloride solution for irrigation. To prove the clinical significance of our ex-vivo findings, clinical studies including large numbers of patients have to be performed. # 2004 Elsevier B.V. All rights reserved. Keywords: Benign prostatic enlargement; TURP; Bipolar resectoscope; Bleeding 1. Introduction * Corresponding author. Tel. +49 621 383 2629; Fax: +49 621 383 2192. E-mail address: gunnar.wendt-nordahl@uro.ma.uni-heidelberg.de (G. Wendt-Nordahl). At present, transurethral resection of the prostate (TURP) is predominant in the surgical therapy of 0302-2838/$ see front matter # 2004 Elsevier B.V. All rights reserved. doi:10.1016/j.eururo.2004.07.018

G. Wendt-Nordahl et al. / European Urology 46 (2004) 586 590 587 benign prostatic enlargement. It is characterized by immediate treatment success due to removal of the infravesical obstruction and produces a long-lasting improvement of symptoms and voiding parameters [1]. However TURP is an invasive surgical method and high perioperative morbidity such as blood loss with the related risk of infection from blood transfusions and TUR-syndrome due to irrigation fluid absorption have been reported [2 4]. In the search for new technologies to reduce morbidity various alternative treatment devices, e.g. interstitial laser therapy, needle ablation, high intensity focused ultrasound, microwave therapy or electrovaporization, have been introduced [5 9]. However standard loop TURP is still regarded as the reference standard method in which urologists have been well trained and remain confident in its use. Contemporary TURP uses a monopolar electrocautery system in which the current passes through the patient s body from the active electrode, placed on the resectoscope, towards the return plate, normally placed on the patient s leg. This has several disadvantages such as heating of deeper tissue, nervous or muscle stimulation and possible malfunction of cardiac pacemakers [10,11]. In the face of these disadvantages, resectoscopes using bipolar electrocautery offer an alternative. As active and return electrode are placed on the same axis on the resectoscope high current densities are achieved locally and distant negative effects are reduced. Another risk of standard monopolar TURP is the absorption of hypoosmolar irrigation fluid during extended procedures causing the TUR syndrome. As bipolar resectoscopes use physiological sodium chloride solution for irrigation, the risk of TUR syndrome is theoretically eliminated. The first bipolar device for endourological procedures was the Gyrus device using bipolar electrocautery to electrovaporize the prostate [12]. In this study we investigated the Vista system (ACMI Corporation, Southborough, MA, USA), a new developed resection device using bipolar electrocautery and saline solution for irrigation. The bipolar resection system comprises the resectoscope and the matching generator (Vista CTR, ACMI Corporation) by which it is activated. The generator works with fixed output voltage levels in the ablation and coagulation mode. The output voltages in the ablation mode at different levels are indicated in Table 1. The bipolar resectoscope (VISTA system, ACMI Corporation) uses a cutting loop (5 mm loop wire diameter) and a second loop, bent in the same direction, working as return electrode. Both are placed on the same axis on the resectoscope and the current flows Table 1 Output voltage of the bipolar HF-generator at different fixed levels in the ablation mode according to the manufacturer Level 1 100 2 125 3 150 4 175 5 200 6 225 7 250 8 275 between them, providing a localized energy field (Fig. 1). The high-frequency energy passes through the conductive irrigation solution (0.9% sodium chloride solution), which is in contact with the tissue to be resected, from the active to the return electrode. The irrigation solution forms a thin layer between the electrodes. In the ablation mode, if sufficient energy is applied, the conductive solution is converted into a vapour layer (plasma) containing energy charged particles. When these high-energy charged particles come in contact with the tissue, they cause its disintegration through molecular dissociation. Compared to conventional monopolar resection systems this leads to lower temperatures at the treatment site, so that the thermal damage of the surrounding tissue is reduced. In this study the Vista system was investigated in an ex-vivo model to determine its cutting qualities, ablation rate, blood loss and coagulation depth and compared it to a standard monopolar resectoscope. 2. Material and methods Output voltage (V rms 10%) 2.1. Equipment The bipolar resection system (Vista system) was activated by the matching generator (Vista CTR, ACMI Corporation). 0.9% sodium chloride solution was used as irrigation fluid. The bipolar resection device was compared to a standard monopolar resectoscope with 5 mm loop wire diameter (Karl Storz, Tuttlingen, Germany) using Purisole SM (standard monnitol/ Fig. 1. Bipolar resection device: the Vista system with tip (enlarged).

588 G. Wendt-Nordahl et al. / European Urology 46 (2004) 586 590 sorbiol solution for TUR; Fresenius, Bad Homburg, Germany) for irrigation. The monopolar resectoscope was activated by the Autocon 400 II high-frequency generator (Karl Storz) using an output power of 160 Watt, coagulation degree 4. 2.2. Ex-vivo experiments To determine cutting qualities, ablation rate, blood loss and coagulation depth we used a modified model of the isolated bloodperfused porcine kidney as described previously [13]. The kidneys were removed from pigs within the first 5 minutes after slaughtering (Mannheim city slaughterhouse). After catheterization of the renal artery with a 10 F catheter the kidneys were perfused with 0.9% sodium chloride solution until the effluent was clear. The kidneys were stored at 4 8C before commencing the trial. During the experiments the kidneys were perfused by a roller pump. To achieve a constant perfusion pressure the pump delivered the perfusion solution over an air chamber via the 10 F catheter, placed in the renal artery. The perfusion pressure was set to 110 130 cmh 2 O resulting in a perfusion rate of 60 100 ml/min. The kidneys were placed on a neutral electrode in an acryl glass basin and fixed. After removing the renal capsule a surface area of 9 cm 2 (3 3 cm) was ablated with a drag speed of 1 cm/s (timed with a stop watch) under permanent perfusion. All experiments were carried out at room temperature. The blood loss was semi quantitatively determined by weighing a swab before and after it was placed on the bleeding surface for 60 seconds. n = 5 measurements were performed per output power level. The weight difference marked the amount of blood loss per minute. Afterwards n = 5 samples per output power level were taken, fixed in 10% formalin and processed for histological evaluation. Hematoxylin-eosin staining was used and the depths of the coagulation zones were determined under the microscope. Five measurements per slide were performed. Statistical data is presented as mean standard deviation. Statistical significance was evaluated using the unpaired student t-test. A p-value <0.05 was considered to be statistically significant. 3. Results We used the modified model of the isolated blood perfused porcine kidney to test the cutting qualities of the new bipolar resection device. In our ex-vivo model satisfactory cutting is only possible using the pre-set levels 5 to 8. It is especially difficult to start a cut. Once the cutting electrode is submerged in the tissue, cutting becomes easier so that accurate and effective cuts are possible. The ablation rate is determined by the loop wire diameter. A loop wire diameter of 5 mm and a drag speed of 1 cm/sec results in an ablation rate of 30 cm 2 / min. This is similar to the ablation rate reached with the monopolar resectoscope, which also has a loop wire diameter of 5 mm. Next, the amount of blood loss was determined using the new bipolar resection device at different HF-generator parameter settings and compared against the conventional monopolar resectoscope. Fig. 2 displays the bleeding rates of the bipolar resection device at different HF-generator output power Fig. 2. Comparison of bleeding rate between the monopolar resectoscope (160 Watt, coagulation degree 4) and the bipolar resectoscope at different output power levels (levels 5 8). Using level 7 and level 8 bleeding is significantly reduced (p < 0.05) compared to the monopolar device. levels and compares them to the standard monopolar resectoscope (n = 5 per output power level). As shown, bipolar resection at level 7 and level 8 exhibits a statistically significantly lower bleeding rate ( p < 0.05) than the conventional resection. Using an output power of 160 Watt; coagulation degree 4 the bleeding rate is 17.08 4.57 g/min for the monopolar resectoscope. The bleeding rate is significantly reduced for the bipolar device resulting in 13.16 5.47 g/min (level 7) and 10.43 4.76 g/min (level 8). Finally samples of the ablated renal surface area were taken and processed histologically (n = 5 per output power level). After HE-staining the depths of the coagulation zones were measured under the microscope. The results are displayed in Fig. 3. It becomes clear that with increasing output powers the depths of the coagulation zones increase when the bipolar resection device is used. Compared to the measurements with the Fig. 3. Comparison of coagulation depth caused by the monopolar resectoscope (160 Watt, coagulation degree 4) and the bipolar resectoscope at different output power levels (levels 5 8). Coagulation depth is significantly smaller ( p < 0.05) using the bipolar resection system compared to the monopolar resectoscope.

G. Wendt-Nordahl et al. / European Urology 46 (2004) 586 590 589 conventional monopolar resectoscope, the depths of the coagulation zones are significantly smaller ( p < 0.05) indicating a reduced effect on deeper tissue. The coagulation depths were 118 22 mm (level 5); 121 23 mm (level 6); 141 62 mm (level 7) and 163 30 mm (level 8) for the bipolar resectoscope and 287 57 mm for the monopolar resectoscope. 4. Discussion Although in the last two decades various alternative treatment devices have been introduced, TURP remains the gold standard in surgical treatment of the benign prostatic enlargement. In the search to optimize the standard method, a new resection device has been developed using bipolar electrocautery rather than the conventional monopolar electrocautery and 0.9% sodium chloride solution as irrigation fluid. The aim of our study was to test this new resection device concerning cutting qualities, ablation rate, bleeding rate and coagulation depth. Our ex-vivo findings indicate that the new device is well suitable to ablate tissue. Although the ex-vivo experiments suggest slightly worse cutting qualities compared to a standard monopolar resectoscope, precise and effective cutting is possible. The tissue ablation rate, defined by drag speed and wire loop diameter, was found to be similar to the standard device using the same loop wire diameter. This is an advantage over other alternative surgical treatment devices, which often exhibit a reduced tissue ablation rate resulting in prolonged procedures [13,14]. Our findings suggest that intraoperative bleeding may be reduced by using the bipolar resectoscope. In the exvivo experiments the bleeding rate is significantly reduced compared to the standard monopolar resection device when the high output voltage levels are used. In contrast to this the coagulation zones are found to be smaller when the bipolar resectoscope is used. This indicates that the bipolar electrocautery produces indeed a locally limited energy field in which high power densities are achieved between the electrodes at the resectoscope. This high energy field effectively coagulates bleeding vessels at the surface of the resected tissue. Deeper tissue layers are affected less indicating that no energy is wasted to them. As the maximum energy field is limited to the resected tissue, distant negative effects, such as deep tissue heating, nervous or muscle stimulation and interfering with cardiac pacemakers, are theoretically reduced in-vivo. Another advantage of bipolar versus monopolar electrocautery lays in the use of 0.9% sodium chloride solution for irrigation. Therefore the risk of TUR syndrome, resulting from the absorption of large amounts of irrigation fluid during prolonged procedures, is theoretically eliminated. So far no clinical studies have been performed comparing the Vista system to conventional monopolar TURP. First clinical results with the Gyrus device, a device using bipolar plasma kinetic technology to electrovaporize the prostate are promising. As the Vista system the Gyrus uses bipolar electrocautery and sodium chloride solution as irrigation fluid. Botto et al. found no significant intraoperative bleeding in 42 treated patients and the reported duration of the procedure appeared similar to that of TURP [12]. Eaton and collegues reported about transurethral prostatectomy using bipolar plasma kinetic vaporization of the prostate on a day-case basis. They found no electrolyte disturbances intraoperatively and bleeding was minimal, so that no patient required a blood transfusion [15]. Boths trials provide only short follow-up information (3 months and 4 months follow-up) which show an improvement in symptoms (IPSS) and voiding parameters (peak flow rate). Dunsmuir and collegues report about a randomized prospective study comparing bipolar electrovaporization of the prostate with the Gyrus device to conventional TURP. After a one year follow-up symptom scores, quality of life, flow rates and post-void residual volumes were similar. The reported re-catheterization rate was higher (30% vs. 5%) in the bipolar vaporization group, the rate of postoperative clot-evacuation was higher in the conventional TURP group. They reach the conclusion that bipolar electrovaporization of the prostate with the Gyrus device is safe and produces similar results to TURP after one year [16]. All these clinical trials emphasize on the low intraoperative and postoperative bleeding when the bipolar device is used. No incidence of hyponatiemia or TUR syndrome is reported as this is theoretically prevented by the use of 0.9% sodium chloride solution. As these trial were performed with a device to electrovaporize the prostate they cannot be directly transferred to the Vista system which performs a resection of the prostate. Unlike other alternative surgical treatments, such as electrovaporization, microwave therapy, needle ablation, radiofrequency ablation or high intensity focussed ultrasound, bipolar TURP offers the possibility of gathering samples of the resected prostatic tissue for histological evaluation. Another treatment option offering the possibility to gather the resected prostatic tissue for histological analysis is the holmium laser enucleation of the prostate when the mushroom technique is used. In this technique the prostatic lobes are

590 G. Wendt-Nordahl et al. / European Urology 46 (2004) 586 590 enucleated with a holmium laser yust short of completion until attached to the capsule by a narrow mushroom-like stalk. As the vascular supply is almost interrupted at that point, they can be electroresected into small pieces without bleeding. This technique combines the advantages of the bloodless holmium laser enucleation and the electroresection of the prostate [17]. In conclusion transurethral resection of the prostate using a bipolar electrocautery system offers a promising alternative to conventional monopolar resectoscopes. Our ex-vivo findings indicate a reduced intraoperative blood loss whereas cutting qualities and ablation rate are comparable to the monopolar device. Clinical studies are necessary to prove if there is a significant benefit for the patient. References [1] Holtgrewe HL, Mebust WK, Dowd JB, Cockett AT, Peters PC, Proctor C. Transurethral prostatectomy: practice aspects of the dominant operation in American urology. J Urol 1989;141(2): 248 53. [2] Horninger W, Unterlechner H, Strasser H, Bartsch G. Transurethral prostatectomy: mortality and morbidity. Prostate 1996;28:195 200. [3] Uchida T, Ohori M, Soh S, Sato T, Iwamura M, Ao T, et al. Factors influencing morbidity in patients undergoing transurethral resection of the prostate. Urology 1999;53(1):98 105. [4] Neal DE. The National Prostatectomy Audit. Br J Urol 1997;79(Suppl 2):69 75. [5] Kursh ED, Concepcion R, Chan S, Hudson P, Ratner M, Eyre R. Interstitial laser coagulation versus transurethral prostate resection for treating benign prostatic obstruction: a randomized trial with 2-year follow-up. Urology 2003;61:573 8. [6] Zlotta AR, Giannakopoulos X, Maehlum O, Ostrem T, Schulman CC. Long-term evalutation of transurethral needle ablation of the prostate (TUNA) for treatment of symptomatic benign prostatic hyperplasia: clinical outcome up to 5 years from 3 centers. Eur Urol 2003;44: 89 93. [7] Schatzl G, Madersbacher S, Djavan B, Lang T, Marberger M. Twoyear results of transurethral resection of the prostate versus four lessinvasive treatment options. Eur Urol 2000;37:695 701. [8] Thalmann GN, Mattei A, Treuthardt C, Burkhard FC, Studer UE. Transurethral microwave therapy in 200 patients with minimumfollow-up of 2 years: urodynamic and clinical results. J Urol 2002;167:2496 501. [9] Hammadeh MY, Fowlis GA, Singh M, Philp T. Transurethral electrovaporization of the prostate a possible alternative to transurethral resection: a one-year follow-up of a prospective randomized trial. BJU 1998;81:721 5. [10] Kellow NH. Pacemaker failure during transurethral resection of the prostate. Anaesthesia 1993;48(2):136 8. [11] Batra YK, Bali IM. Effect of coagulating and cutting current on a demand pacemaker during transurethral resection of the prostate. A case report Can Anaesth Soc J 1978;25(1):65 6. [12] Botto H, Lebret T, Barre P, Orsoni JL, Herve JM, Lugagne PM. Electrovaporization of the prostate with the Gyrus device. J Endourol 2001;15(3):313 6. [13] Michel MS, Köhrmann KU, Weber A, Krautschick AW, Alken P. Rotoresect: new technique for resection of the prostate: experimental phase. J Endourol 1996;10(5):473 8. [14] Tunuguntla HS, Evans CP. Minimally invasive therapies for benign prostatic hyperplasia. World J Urol 2002;20(4):197 206. [15] Eaton AC, Francis RN. The provision of transurethral prostatectomy on a day-case basis using bipolar plasma kinetic technology. BJU 2002;89(6):534 7. [16] Dunsmuir WD, McFarlane JP, Tan A, Dowling C, Downie J, Kourambas J, et al. Gyrus bipolar electrovaporization vs transurethral resection of the prostate: a randomized prospective single-blind trial with 1 y follow up. Prostate Cancer Prostatic Dis 2003;6(2):182 6. [17] Hochreiter WW, Thalmann GN, Burkhard FC, Studer UE. Holmium laser enucleation of the prostate combined with electrocautery resection: the mushroom-technique. J Urol 2002;168(4 Pt 1):1470 4.