Patricia C. Fallest-Strobl, 1 Elin Olafsdottir, 2 Donald A. Wiebe, 1 and James O. Westgard 1 * Lipoproteins

Similar documents
Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips

Technical Bulletin. Cholesterol Reference Method Laboratory Network (CRMLN) Overview. TB Rev. 0

Individual Lab Report Ci-Trol Nov,2016. Abnormal Fibrinogen (mg/dl) Abnormal Fbg Control - Lot# LFC Your Lab

HDL CHOLESTEROL. 01 English - Ref.: 13. Ref.:13. Insert

Rethinking Traditional Lab QA/QC:

METHOD VALIDATION: WHY, HOW AND WHEN?

Rethinking Tradi.onal QA/QC: Weaknesses of Current Prac.ces; Six Sigma Metrics and Tools for Improvement

LDL . (LDL) Downloaded from ijdld.tums.ac.ir at 12:38 IRST on Friday February 1st 2019 LDL LDL NCEP-ATP-III (LDL) :

General Chemistry Scheme Guide

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

Quality Control in the Chiropractic Clinical Setting Utilizing Thermography Instrumentation as a Model

Farouk AL Quadan lecture 3 Implentation of

The analytical phase

1. Reagent Store at 2-8 ºC.

Current Challenges in CardioMetabolic Testing. Kenneth French, Director of Clinical Operations

Calibration of High-Density Lipoprotein Cholesterol Values From the Korea National Health and Nutrition Examination Survey Data, 2008 to 2015

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

ARTICLE. Utility of Direct Measurement of Low-Density Lipoprotein Cholesterol in Dyslipidemic Pediatric Patients

Evaluation of Calculated Low-Density Lipoprotein Against a Direct Assay

Analytical Goal-Setting for Monitoring Patients When Two Analytical Methods Are Used

The Virtues and Pitfalls of Implementing a New Test

Method Comparison Report Semi-Annual 1/5/2018

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Comparison of two assays for measuring LDL cholesterol

Separation of HDL Particles by Immunoprecipitation

Friedewald formula. ATP Adult Treatment Panel III L D L Friedewald formula L D L = T- C H O - H D L - T G / 5. Friedewald formula. Friedewald formula

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

Quantitative results play an important role in diagnosing

LDL LD. 01 English - Ref.: 129. Ref.:129. Insert. Intended use. Methodology. Reagents. Test principle. Summary. Precautions and warnings

Placement of Biomarker Analysis in a CLIA or Bioanalytical Laboratory

4/10/2017. Placement of Biomarker Analysis in a CLIA or Bioanalytical Laboratory

Biological Variation The logical source for analytical goals

Guide to Fulfillment of Measurement Uncertainty

Non-fasting Lipid Profile Getting to the Heart of the Matter! Medimail Dec 2017

Shinichi Usui, 1 Masakazu Nakamura, 2 Kazuhiro Jitsukata, 3 Masayuki Nara, 4 Seijin Hosaki, 1 and Mitsuyo Okazaki 4*

SCIENTIFIC STUDY REPORT

Customer Bulletin January 2017

Setting of quality standards

ORIGINAL ARTICLE INTRODUCTION. Jongseok Lee 1, Sungok Jang 2, Haemin Jeong 3, and Ohk-Hyun Ryu 3

Evaluation of a Rapid Homogeneous Method for Direct Measurement of High-Density Lipoprotein Cholesterol

Quantitative Method for Amphetamines, Phentermine, and Designer Stimulants Using an Agilent 6430 LC/MS/MS

College of American Pathologists

Katsuyuki Nakajima, PhD. Member of JCCLS International Committee

INTERNAL QUALITY CONTROL ( Q I C) QC)

Comparison of Friedewald Formula and Modified Friedewald Formula with Direct Homogeneous Assay for Low Density Lipoprotein Cholesterol Estimation

Serum LDL- and HDL-cholesterol determined by ultracentrifugation and HPLC

Serum low-density lipoprotein and high-density lipoprotein cholesterol. determined by ultracentrifugation and high-performance liquid

QUALITROL 2H Insert. 01 Inglês - Ref.: 72. Ref.:72. Precautions and warnings

10 Intraclass Correlations under the Mixed Factorial Design

Customer Bulletin. Questions/Comments? Please call Customer Service at (Direct) or (Toll-free inside USA).

Piccolo Lipid Panel Reagent Disc

QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES

The Integrated Cardiovascular Clinical Network CHSA

Two Important Statistics

1 PROTOCOL. Comparison Study Summary

What information on measurement uncertainty should be communicated to clinicians, and how? Mario Plebani

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009

Biological Variability of Cholesterol, Triglyceride, Low- and High-Density Lipoprotein Cholesterol, Lipoprotein(a), and Apolipoproteins A-I and B

In Vitro Diagnostic Glucose Test System

Abaxis Piccolo xpress TM

Analytic Bias of Thyroid Function Tests. Analysis of a College of American Pathologists Fresh Frozen Serum Pool by 3900 Clinical Laboratories

Ingmar Jungner, 1* Santica M. Marcovina, 2 Göran Walldius, 3 Ingar Holme, 4 Werner Kolar, 1 and Eugen Steiner 1. Lipoproteins

Journal of Assisted Reproduction and Genetics Plenum Publishing Corporation All Rights Reserved.

Determination of hemoglobin is one of the most commonly

LIPID-ATHEROSCLEROSIS LABORATORY SERVICES GUIDE

The Adiponectin Turbidimetric Immunoassay Reagent Kit

Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer

Martin/Hopkins Estimation, Friedewald and Beta- Quantification of LDL-C in Patients in FOURIER

Clinical Policy: Pitavastatin (Livalo), Ezetimibe/Simvastatin (Vytorin 10/10 mg) Reference Number: CP.CPA.62 Effective Date:

6to. Congreso Virtual de Cardiología - 6th Virtual Congress of Cardiology

The Framingham Coronary Heart Disease Risk Score

NICO /14 Nipro Diagnostics, Inc.

It s a Gas! Issues in the Blood Gas Laboratory. D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC

Metabolic Monitoring Point of Care Testing

New immunoseparation-based homogeneous assay for HDL-cholesterol compared with three homogeneous and two heterogeneous methods for HDL-cholesterol

Quality Controls in Allergy Diagnosis

Accuracy of Three Dry-Chemistry Methods for Lipid Profiling and Risk Factor Classification

Analytical Performance in the The Test Evaluation Cycle

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

UREA CE Insert. 01 English - Ref.: 27. Ref.:27

Lipid Management 2013 Statin Benefit Groups

The determination of timer error and its role in the administration of specified doses

Rosuvastatin: An Effective Lipid Lowering Drug against Hypercholesterolemia

Why do we need SD LipidoCare?

Lipid Profile Analysis of Aircrew

Plasma Lipids and Global Cardiovascular Risk

Understanding Quality Control for Infectious Disease Testing. Wayne Dimech ASLM Quality Control Workshop Abuja, Nigeria 9 th December 2018

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

Reducing low-density lipoprotein cholesterol treating to target and meeting new European goals

Direct HbA1c testing capabilities on the RX modena

Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users. David Ducroq.

QUALITATIVE EVALUATION OF IN-HOUSE IMMUNOASSAYS OF T3, 14 AND TSH BASED ON BULK REAGENTS

INSTRUCTIONS FOR USE

Research Article. National Cerebral and Cardiovascular Center, Osaka, Japan 2 Public Health, Department of Social Medicine, Graduate School of

Lipoprotein Particle Profile

Transcription:

Clinical Chemistry 43:11 2164 2168 (1997) Lipids and Lipoproteins Comparison of NCEP performance specifications for triglycerides, HDL-, and LDL-cholesterol with operating specifications based on NCEP clinical and analytical goals Patricia C. Fallest-Strobl, 1 Elin Olafsdottir, 2 Donald A. Wiebe, 1 and James O. Westgard 1 * The National Cholesterol Education Program (NCEP) performance specifications for methods that measure triglycerides, HDL-cholesterol, and LDL-cholesterol have been evaluated by deriving operating specifications from the NCEP analytical total error requirements and the clinical requirements for interpretation of the tests. We determined the maximum imprecision and inaccuracy that would be allowable to control routine methods with commonly used single and multirule quality-control procedures having 2 and 4 control measurements per run, and then compared these estimates with the NCEP guidelines. The NCEP imprecision specifications meet the operating imprecision necessary to assure meeting the NCEP clinical quality requirements for triglycerides and HDL-cholesterol but not for LDLcholesterol. More importantly, the NCEP imprecision specifications are not adequate to assure meeting the NCEP analytical total error requirements for any of these three tests. Our findings indicate that the NCEP recommendations fail to adequately consider the quality-control requirements necessary to detect medically important systematic errors. Three different types of performance criteria have been established for lipid tests performed in healthcare laboratories. For example, cholesterol tests are regulated by the Clinical Laboratory Improvement Act (CLIA-88), which requires laboratories to provide test values that are correct within 10% of target values for proficiency testing materials (a total error criterion). The US National Cholesterol Education Program (NCEP) 3 provides physicians with explicit guidelines to interpret a patient s cholesterol values (a clinical criterion) and also provides laboratories with specifications for the imprecision and inaccuracy of their routine cholesterol methods (analytical criteria). Guidelines published by NCEP recommended that by 1992 the performance of cholesterol tests should have a 3.0% or less CV (or imprecision) and 3.0% or less bias (inaccuracy) [1]. We evaluated these recommendations in earlier studies [2 4] and expressed concern that QC issues had not been adequately considered, in that a CV of 2.4% or less was actually required to assure that routine methods would achieve the performance required for the clinical use of the test. Because cholesterol is a key risk factor for cardiovascular disease, the importance for establishing both clinical and analytical goals cannot be argued. However, these goals are interdependent (i.e., a method s analytical performance will often determine the clinical usefulness of a given test and, likewise, clinical demands often dictate the stringency required of a method s analytical performance). The two groups that established these guidelines, the Adult Treatment Panel and the Laboratory Standardization Panel, failed to interconnect the different performance criteria and address them together. Nonetheless, the primary goals of NCEP have been overwhelmingly 1 Department of Pathology and Laboratory Medicine, University of Wisconsin Medical School, Madison, WI 53792. 2 Department of Clinical Biochemistry, University Hospital, Reykjavik, Iceland. *Author for correspondence. Fax (608)263-1568; e-mail jo.westgard@ hosp.wisc.edu. Received February 25, 1997; revision accepted June 26, 1997. 3 Nonstandard abbreviations: NCEP, National Cholesterol Education Program; OPSpecs, operating specifications for allowable imprecision (CV), allowable inaccuracy (bias), and necessary QC; TE a, total allowable analytical error; D int, clinical decision interval; s wsub, within-subject biological variation; TRIG, triglycerides; LDL-C, low-density lipoprotein cholesterol; HDL-C, high-density lipoprotein cholesterol. 2164

Clinical Chemistry 43, No. 11, 1997 2165 achieved. Family physicians appropriately treat hypercholesterolemic patients with aggressive therapy to lower cholesterol to lessen their cardiovascular risk. Laboratories have improved their performance and are providing more-accurate cholesterol test values. The laboratory success was driven in part by manufacturers that provided improved instruments and techniques designed to meet the NCEP guidelines. Cholesterol, however, is only one of several lipid components of clinical importance for which appropriate guidelines are needed. In 1995, the NCEP Laboratory Standardization Panel published recommendations of performance specifications for triglycerides (TRIG), HDLcholesterol (HDL-C), and LDL-cholesterol (LDL-C) [5 7]. These recommendations define specifications for the imprecision and inaccuracy of the respective methods, the total analytical error that is allowable, and the medical decision values. However, lack of guidance for the routine operation and QC of these methods prompted us to evaluate the operating specifications necessary to assure that the NCEP clinical and analytical goals could be achieved in routine lipid testing, taking into account the QC performance that would be necessary to detect analytical runs in which method performance is unstable. Detailed specifications for QC are essential for these methods, which are prone to many analytical variables that could affect routine laboratory performance and result in inappropriate patient data transmitted to the patient. Materials and Methods OPSpecs charts [8] were prepared by using the QC Validator computer program [9] (Westgard Quality Corp., Ogunquit, ME). The principles of the OPSpecs methodology have been described in detail elsewhere [10]. OPSpecs charts plot the allowable inaccuracy on the y-axis vs the allowable imprecision on the x-axis for a defined quality requirement and a specified requirement of error detection, or analytical quality assurance. The observed imprecision and inaccuracy of an individual method can be plotted to locate the operating point of the method. Appropriate QC procedures are provided by the control rules and number of control measurements per run (N), the operating limits of which are above the method s observed operating point [11]. Estimates of maximum allowable imprecision can be obtained from the x-intercepts of the operating limits [12]. In this study, OPSpecs charts were prepared to show the imprecision and inaccuracy that are allowable for common QC procedures with N 2 to 4, 90% analytical quality assurance, and the stated NCEP analytical total error requirements and the clinical decision interval (D int ) requirements implied by the guidelines for test interpretation [5 7]. We calculated D int for each of the tests according to the medical decision values for treatment of patients as recommended by the Adult Treatment Panel. For TRIG, concentrations for healthy subjects are 2000 mg/l and high concentrations are 4000 mg/l; therefore, the D int is 100%: (2000/2000) 100%. For HDL-C, desirable concentrations are 600 mg/l and major risk concentrations are 350 mg/l; therefore, D int is 42%: (250/600) 100%. For LDL-C, desirable values are 1300 mg/l and high-risk concentrations are 1600 mg/l; therefore, D int 23.1%: (300/1300) 100%. Estimates of the within-subject biological variation were taken as 23.7% for TRIG [6], 7.5% for HDL-C [7], and 8.2% for LDL-C [5], as recommended in the NCEP documents. The range of allowable imprecision (operating imprecision in Table 1) was determined at the bias stated for each test. Results Clinical quality requirements. The OPSpecs charts for the NCEP clinical quality requirements for TRIG, HDL-C, and LDL-C are shown in Fig. 1. For TRIG, the operating point that represents the NCEP-recommended specifications (inaccuracy of 5%, imprecision of 5%) is well below the allowable range of imprecision (14.0 22.5%) at the stated value of 5% for inaccuracy, as denoted by the doubleheaded arrow in Fig. 1A. Therefore, the TRIG specifications for imprecision and inaccuracy will assure the quality defined by the NCEP clinical requirement (D int 100%; s wsub 23.7%). The operating point for HDL-C falls within the allowable range of imprecision (5.0 7.8%) for the control rules with N 2 4 (Fig. 1B); thus, the HDL-C specifications for accuracy (10.0%) and precision (6.0%) will also assure the quality defined by the NCEP clinical requirement (D int 42.0%, s wsub 7.5%). However, the LDL-C operating point of 4.0% imprecision and 4.0% inaccuracy exceeds the allowable range of imprecision (1.5 2.6%), as shown in Fig. 1C. Therefore, the NCEP Table 1. Comparison of NCEP clinical and analytical requirements with operating imprecision derived from OPSpecs charts. Test and requirements NCEP bias NCEP CV, % Operating imprecision Clinical requirements TRIG 5.0 5.0 14.0 22.5 D int 100%, s wsub 23.7% 0 5.0 15.5 24.0 HDL-C 10.0 6.0 5.0 7.8 D int 42.0%, s wsub 7.5% 0 6.0 7.0 11.0 LDL-C 4.0 4.0* 1.5 2.6 D int 23.1%, s wsub 8.2% 0 4.0 2.5 4.2 Analytical requirements TRIG 5.0 5.0* 2.2 3.2 TE a 15.0% 0 5.0* 2.9 4.2 HDL-C 10.0 6.0* 2.2 3.7 TE a 22.0% 0 6.0 4.2 6.2 LDL-C 4.0 4.0* 1.6 2.3 TE a 12.0% 0 4.0* 2.3 3.5 D int clinical decision interval, s wsub within-subject biological variation, TE a total allowable error, bias allowable inaccuracy, CV coefficient of variation or allowable imprecision, * NCEP recommendation exceeds the operating imprecision.

2166 Fallest-Strobl et al.: NCEP performance specifications vs clinical/analytical goals method performance specifications for LDL-C will not assure the quality defined by the NCEP clinical requirement (D int 23.1%, s wsub 8.2%). Analytical quality requirements. The operating point for TRIG (inaccuracy 5%, imprecision 5%) exceeds the allowable range of imprecision (2.2 3.2%) for analytical quality requirements (TE a 15.0%), as depicted by the OPSpecs chart in Fig. 2A. Therefore, common QC procedures with N 2 4 will not be able to detect unstable performance. Similarly, the OPSpecs charts for HDL-C and LDL-C (Fig. 2B and C) show operating points that also exceed the allowable range of imprecision for detection of unstable performance. Therefore, the analytical quality requirements for precision of test methods for TRIG, HDL-C, and LDL-C will not assure the analytical quality defined by NCEP. NCEP s recommendations for HDL-C [7] also included performance specifications (TE a 13%, CV 4.0%, bias 5.0%) to be achieved by 1998. Although these specifications are more stringent than the current criteria for HDL-C, analytical quality requirements are still not satisfied. OPSpecs charts prepared for use of common QC procedures (not shown) indicate that the allowable imprecision for HDL-C methods still needs to be 4.0% (within the range 1.5 2.3%) at a bias of 5.0% and TE a 13.0%. Table 1 summarizes the ranges of allowable imprecision derived from OPSpecs charts, based on the NCEP clinical and analytical quality requirements for all of these tests. Also shown are the allowable ranges of imprecision for each test if bias were eliminated as a consideration (bias 0). In this scenario, the allowable ranges of imprecision are increased for all of the tests, but NCEP analytical quality requirements are still not assured for TRIG or LDL-C (Table 1). Fig. 1. OPSpecs chart for clinical quality requirement for (A) triglycerides (D int 100%, s wsub 23.7%), (B) HDL-cholesterol (D int 42.0%, s wsub 7.5%), and (C) LDL-cholesterol (D int 23.1%, s wsub 8.2%). Curves (top to bottom in each panel) represent control procedures for the following rules: 1 2s with N 4, 1 3s /2 2s /R 4s /4 1s with N 4, 1 2.5 s with N 4, 1 2s with N 2, 1 3s with N 4, 1 2.5 s with N 2, 1 3s /2 2s /R 4s with N 2, and 1 3s with N 2). 7, range of imprecision at the bias allowed by NCEP. Discussion Methods for determination of TRIG, HDL-C, and LDL-C are complex and can be problematic, which means it is important to select optimal QC procedures capable of detecting unstable performance. TRIG measurements are often automated, whereas HDL-C and LDL-C methods usually include manual steps in test protocols. LDL-C test results may be the most problematic because common practice is to report estimated LDL-C by calculation with the Friedewald equation [estimated LDL total cholesterol HDL (TRIG/5)]. Therefore, each LDL-C result relies on accurate and precise measurements of three other lipids: cholesterol, HDL-C, and TRIG. The ability to detect unstable performance in routine laboratory testing for lipids through use of statistical QC procedures is crucial to avoid further propagation of error in generating LDL-C test results. Each laboratory should pay special attention to the operating specifications needed to assure that the desired analytical or clinical quality is achieved in routine service. OPSpecs charts provide a graphical display of the impre-

Clinical Chemistry 43, No. 11, 1997 2167 cision and inaccuracy that are allowable when different QC procedures are used. In this approach, statistical QC is factored in as a component of the error budget for a test, thus providing a more realistic assessment of the imprecision and inaccuracy limits that are necessary if the QC procedure is to detect medically important errors. OP- Specs charts, therefore, are a very useful tool for defining the specifications needed for lipid tests in a routine service laboratory. The NCEP analytical recommendations for imprecision and inaccuracy seem to be based on the method acceptance criterion of TE a bias 2 s. For methods that just meet this criterion, commonly used control procedures are not expected to provide reliable detection of medically important errors [13]. Even if bias could be eliminated from lipid test methods, the derived data for operating precision (when bias 0) show that NCEP-recommended precision goals for TRIG and LDL-C are not satisfactory to meet the stated analytical total error requirements. Laboratories should select methods for these tests for which operating imprecision is within the ranges indicated in Table 1 to assure adequate detection of unstable performance. In general, further reducing by a factor of 2 the limits set by NCEP for allowable imprecision would be desirable to provide more-controllable testing processes. In conclusion, improvements in lipid test performance and interpretation have continued over the years because of efforts by NCEP, but our findings indicate that additional guidelines are needed for statistical QC to assure detection of unstable performance. Although NCEP s 1998 goals for HDL-C are a step in a right direction, they will still fail to provide a high degree of analytical quality assurance. In the future, development of performance specifications for analytical methods should take into account the performance characteristics of the statistical QC procedures that are expected to be used in monitoring routine laboratory performance. The NCEP Adult Treatment Panel and Laboratory Standardization Panel independently generated clinical and analytical goals that have conflicting or unobtainable targets. Clearly, laboratorians will need guidance to establish acceptable performance standards for these routine, yet demanding lipid tests. However, if performance goals result in unrealistic objectives, the final outcome could be misuse or inappropriate handling of proficiency testing material to ensure that the analytical goals are met, while patients needs are ignored. These activities must be coordinated with clearer QC operations for laboratories to properly manage these tests. Fig. 2. OPSpecs chart for analytical quality requirements for (A) triglycerides (TE a 15.0%), (B) HDL-cholesterol (TE a 22.0%), and (C) LDL-cholesterol (TE a 12.0%). See Fig. 1 for QC rules. The Department of Pathology and Laboratory Medicine, University of Wisconsin Medical School, provided partial support for E.O. during a sabbatical period and full support for P.F.-S. during a Clinical Chemistry Fellowship.

2168 Fallest-Strobl et al.: NCEP performance specifications vs clinical/analytical goals References 1. National Cholesterol Education Program Laboratory Standardization Panel. Current status of blood cholesterol measurement in clinical laboratories in the United States. Clin Chem 1988;34: 193 201. 2. Westgard JO, Hyltoft Petersen P, Wiebe DA. Laboratory process specifications for assuring quality in the US National Cholesterol Education Program. Clin Chem 1991;37:656 61. 3. Westgard JO, Wiebe DA. Cholesterol operational process specifications for assuring quality required by CLIA proficiency testing. Clin Chem 1991;37:1938 44. 4. Wiebe DA, Westgard JO. Cholesterol a model system to relate medical needs with analytical performance. Clin Chem 1993;39: 1504 13. 5. Bachorik PS, Ross JW, for the NCEP Working Group on Lipoprotein for measurement of low-density lipoprotein cholesterol: executive summary. Clin Chem 1995;41:1414 20. 6. Stein EA, Myers GL, for the NCEP Working Group on Lipoprotein for triglyceride measurement: executive summary. Clin Chem 1995;41:1421 6. 7. Warnick GR, Wood PD, for the NCEP Working Group on Lipoprotein for measurement of high-density lipoprotein cholesterol: executive summary. Clin Chem 1995;41:1427 33. 8. Westgard JO. Charts of operational process specifications ( OP- Specs charts ) for assessing the precision, accuracy, and quality control needed to satisfy proficiency testing criteria. Clin Chem 1992;38:1226 33. 9. Westgard JO, Stein B. Automated selection of statistical qualitycontrol procedures to assure meeting clinical or analytical quality requirements [Tech Brief]. Clin Chem 1997;43:400 3. 10. Westgard JO. Error budgets for quality management: practical tools for planning and assuring the analytical quality of laboratory testing processes. Clin Lab Manage Rev 1996;10:377 403. 11. Mugan K, Carlson IH, Westgard JO. Planning QC procedures for immunoassays. J Clin Immunoassay 1994;17:216 22. 12. Westgard JO, Seehafer JJ, Barry PL. European specifications for imprecision and inaccuracy compared with operating specifications that assure the quality required by US CLIA proficiencytesting criteria. Clin Chem 1994;40:1228 32. 13. Westgard JO, Burnett RW. Precision requirements for cost-effective operation of analytical processes. Clin Chem 1990;35:1629 32.