MAMMOSITE BRACHYTHERAPY DOSIMETRY EFFECT OF CONTRAST AND AIR INTERFACE ON SKIN DOSE. A RESEARCH PAPER SUBMITTED TO THE GRADUATE SCHOOL

Similar documents
Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w

Brachytherapy Planning and Quality Assurance

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations

Dose distribution and dosimetry parameters calculation of MED3633 palladium-103 source in water phantom using MCNP

Online in vivo dosimetry in conformal radiotherapies with MOSkin detectors

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Predicting Maximum Pacemaker/ICD Dose in SAVI HDR Brachytherapy

Review of TG-186 recommendations

Implantable MOSFET dosimeter response to 192 Ir HDR radiation

Physics. Copyright 2010 Xoft, Inc.

Definitions. Brachytherapy in treatment of cancer. Implantation Techniques and Methods of Dose Specifications. Importance of Brachytherapy in GYN

Research Article Dose Distributions of an 192 Ir Brachytherapy Source in Different Media

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

Regulatory Guidelines and Computational Methods for Safe Release of Radioactive Patients II. Brachytherapy

ADVANCED TECHNOLOGY CONSORTIUM (ATC) CREDENTIALING PROCEDURES FOR LUNG BRACHYTHERAPY IMPLANT PROTOCOLS

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Ritu Raj Upreti, S. Dayananda, R. L. Bhalawat*, Girish N. Bedre*, D. D. Deshpande

A commissioning procedure for breast intracavitary electronic brachytherapy systems

AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams

AAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

HDR Brachytherapy Dosimetry at the NPL

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Quality management for Breast Brachytherapy.

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

Chapters from Clinical Oncology

Evaluation of three APBI techniques under NSABP B-39 guidelines

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 2, SPRING 2005

Verification of Balloon Integrity by Ultrasound Imaging in Accelerated Partial Breast Irradiation

Recent proceedings in Brachytherapy Physics

Dosimetric Characteristics of the Brachytherapy Sources Based on Monte Carlo Method

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

Commissioning and Radiobiology of the INTRABEAM System

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

NSABP PROTOCOL B-39B RTOG PROTOCOL 0413

Application of Implanted Markers in Proton Therapy. Course Outline. McLaren Proton Therapy Center Karmanos Cancer Institute McLaren - Flint

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

Hampton University Proton Therapy Institute

U.S. Nuclear Regulatory Commission

Transperineal Interstitial Permanent Prostate Brachytherapy (TIPPB) Quality Assurance Guidelines

Monte Carlo Simulation Study on Dose Enhancement by Gold Nanoparticles in Brachytherapy

Accurate Accurate boost or Simply Accuboost

Pulsed Dose Rate for GYN Brachytherapy

Wireless In-Vivo Dosimetry of High Dose Rate Brachytherapy for Prostate Cancer using MOSkin Detectors

Brachytherapy. What is brachytherapy and how is it used?

The Physics of Oesophageal Cancer Radiotherapy

Accelerated Partial Breast Irradiation. Dr Patricia Lillis MD, MHA,MSS Marshfield Clinic Radiation Oncology

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

EORTC Member Facility Questionnaire

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

Calculation of Dose Distribution Around a Clinical 252. Cf Source for Neutron Therapy Based on AAPM, TG-43 Protocol

Brachytherapy, Radionuclide Therapy Medical Physics in the Clinic. Raymond K. Wu, PhD Chairman AAPM Exchange Scientist Program

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?

Clinical Applications of Brachytherapy Radiobiology. Radiobiology is Essential

Transition to Heterogeneity Corrections. Why have accurate dose algorithms?

Metrology Laboratory of Ionizing Radiation

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

The Future of Brachytherapy. Alex Rijnders Europe Hospitals Brussels, Belgium

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

Diode calibration for dose determination in total body irradiation

Improving clinical brachytherapy: dosimetry & verification

3D CONFORMATIONAL INTERSTITIAL BRACHYTHERAPY PLANNING FOR SOFT TISSUE SARCOMA

Radiotherapy physics & Equipments

Current Status of Electronic Brachytherapy Dosimetry

Therapeutic Medical Physics. Stephen J. Amadon Jr., Ph.D., DABR

Brachytherapy an Overview

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Brachytherapy The use of radioactive sources in close proximity to the target area for radiotherapy

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy

Comparative Study on Homogeneous and Inhomogeneous Photon Dose Distribution from a High Dose Rate 192 Ir Brachytherapy Source using MCNP5

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Monte Carlo and experimental relative dose determination for an Iridium-192 source in water phantom

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Real-time brachytherapy for prostate cancer implant analysis

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

PRINCIPLES and PRACTICE of RADIATION ONCOLOGY. Matthew B. Podgorsak, PhD, FAAPM Department of Radiation Oncology

Medical Use of Radioisotopes

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Applications in brachytherapy

Advanced film dosimetry for a UK brachytherapy audit

I. Equipments for external beam radiotherapy

ph fax

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Larry A. DeWerd Accredited Dosimetry and Calibration Laboratory, University of Wisconsin, Madison, Wisconsin, 53706

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

Skyscan 1076 in vivo scanning: X-ray dosimetry

Transcription:

MAMMOSITE BRACHYTHERAPY DOSIMETRY EFFECT OF CONTRAST AND AIR INTERFACE ON SKIN DOSE. A RESEARCH PAPER SUBMITTED TO THE GRADUATE SCHOOL IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE MASTER OF ARTS BY IMENDRA PADMASRI RANATUNGA ADVISOR RANJITH WIJESINGHE, PH.D CO-SUPERVISOR ALVIS FOSTER, PH.D BALL STATE UNIVERSITY JULY 2014 i

TABLE OF CONTENT PAGE DEDICATION... TABLES OF CONTENT... LIST OF FIGURES... LIST OF TABLES... i ii iv vi CHAPTER ONE: Introduction 1.1 Brest cancer statistics... 1 1.2 Treatments... 2 1.2.1 External beam radiation therapy 2 1.2.2 Brachytherapy 3 1.2.3 Treatment Planning 4 1.3 Statement of the problem 6 1.3.1 Effect of contrast material.. 6 1.3.2 Effect of breast air interface 7 1.4 Significant of this study.. 8 CHAPTER TWO: Brachytherapy dosimetry 2.1 Dose rate calculation.. 9 2.1.1 Air kerma strength.. 10 2.1.2 Dose rate constant... 11 2.1.3 Geometry function.. 11 2.1.4 Radial dose function 12 2.1.5 Anisotropy function. 12 ii

2.2 Dose calculation.. 12 CHAPTER THREE: Material and method 3.1 Effect of Contrast material. 14 3.1.1 Phantom.. 14 3.1.2 CT simulation. 14 3.1.3 Treatment planning. 16 3.1.4 Radiation Delivery.. 16 3.2 Effect of Breast air interface... 17 3.2.1 Phantom.. 17 3.2.2 CT simulation. 17 3.2.3 Treatment planning. 18 3.2.4 Radiation Delivery.. 19 3.3 MammoSite Balloon Catheter. 19 3.4 Ion chamber... 20 CHAPTER FOUR: Results 4.1 Effect of Contrast material... 22 4.2 Effect of Breast air interface... 24 CHAPTER FIVE: Conclusion 5.1 Effect of Contrast material.. 27 5.2 Effects of breast-air interface.. 27 REFERENCE. 29 iii

LIST OF FIGURES Figure 1.1 External beam radiation therapy setup 2 Figure 1.2 Mammosite balloon and afterloader setup 4 Figure 1.3 Definition of PTV of a MammoSite balloon 5 Figure 1.4 A CT image of isodose surfaces of the MammoSite balloon 6 Figure 1.5 Dose contribution from over-lying tissues. A) Dose from full 7 scatter (including backscatter contribution). B) Dose without backscatter contribution. ( ) Figure 2.1 Schematic diagrams of a seed and the point of interest P r,θ 10 used in TG 43. Figure 3.1 Cross sectional diagram of the phantom setup 15 Figure 3.2 Actual phantom setup 15 Figure 3.3 Schematic diagram of a 100cGy isodose curve appear in 16 treatment planning system Figure 3.4 Radiation delivery system from afterloader to MammoSite 17 catheter. Figure 3.5 Cross sectional diagram of the phantom setup 18 Figure 3.6 Schematic diagram of a 100cGy isodose curve appear in 18 treatment planning system Figure 3.7 Mammosite Balloon 19 Figure 3.8 0.6 cc PTW Farmer Chamber 20 Figure 4.1 Comparison of the normalized dose as a function of contrast 23 concentration for the MammoSite balloon placed in cubical water phantom, at 5 cm from catheter axis. a) Delivered dose measurements from Oncentra treatment planning system. b) Measured dose measurements from 6cc Farmer chamber. iv

Figure 4.2 Comparison of the normalized dose as a function of overline 26 water thickness for the MammoSite balloon placed in cubical water phantom, at 5 cm from catheter. a) Delivered dose measurements from Oncentra treatment planning system. b) Measured dose from 6cc Farmer chamber. v

LIST OF TABLES Table 1.1 Relative survival rates from breast cancer 1 Table 4.1. Recorded dwell times from Nucletron Oncentra treatment 22 planning system for different contrast concentrations Table 4.2. Recorded ion chamber readings for different contrast 23 concentrations Table 4.3 Recorded dwell times from Nucletron Oncentra treatment 24 planning system for different over-lying water thicknesses. Table 4.4. Recorded ion chamber readings for different overline water 25 thicknesses. vi

CHAPTER 01 Introduction 1.1 Breast Cancer statistics Among the many types of cancer, breast cancer is the type that causes the most deaths among women worldwide [1]. In 2013, the United States alone reported 232,340 new breast cancer cases and 39,620 deaths due to breast cancer [2]. It can be inferred using the rates from 2007-2009 that 12.38% of women born today will be diagnosed with breast cancer during their lifetime [3]. This means that one in eight women are going to be diagnosed with cancer during their life [4]. Table 1.1 shows increasing five-year relative survival rates in the United States [5]. The increasing rate of survival is thought to be due to advanced treatment options and early diagnosis. Year of diagnosis 5 year relative survival rate % 1975-1977 74.8 1978-1980 74.4 1981-1983 76.1 1984-1986 78.9 1987-1989 84.0 1990-1992 85.2 1993-1995 86.4 1996-1998 88.2 1999-2003 89.9 2004-2010 90.5 Table 1.1 Relative survival rates from breast cancer 1

1.2 Treatments There are mainly three types of cancer treatment techniques; radiation therapy, surgical procedures and chemotherapy. Radiation therapy is divided into two categories as external beam radiation therapy and brachytherapy. 1.2.1 External beam radiation therapy External beam radiation therapy (EBRT) is one of the oldest and effective techniques used to treat cancer. EBRT is a method of delivering one or multiple beams of high-energy x- rays from an external source to a patient's tumor. Here, beams are generated outside the patient and are targeted at the tumor site [6]. The treatment time could vary from hours to several weeks depending on the type as well as the stage of the cancer. Fig 1.1 External beam radiation therapy setup 2

1.2.2 Brachytherapy Short distance treatment of cancer using radiation from small, an encapsulated radionuclide source is called Brachytherapy [7]. It is done by placing sources of radiation directly in or near the area to be treated. The dose of radiation has to be delivered continuously over a short period of time when temporary implants are used or over the lifetime of the source when permanent implants are used. The most common radiation modality used in brachytherapy is gamma photons. There are two main types of brachytherapy treatments: Intracavitary (where the sources are placed adjacent to the treatment volume, e.g., gynecology HDR treatments) and Interstitial (where the sources are implanted inside the treatment volume, e.g., prostate LDR treatments) [7]. In this study we focus on an Intracavitary treatment method called MammoSite brachytherapy. Radiation may be used to treat breast cancer after the tumor has been removed with a lumpectomy. This method of treatment is used to kill residual cancer cells surrounding the lumpectomy cavity. The technique is known as accelerated partial breast irradiation (APBI) [1]. The device is called MammoSite catheter, which consists of a small silicon balloon connected to an external inflation lumen and a internal lumen for the passage of a high dose rate Ir-192 brachytherapy source [8], as shown in figure 1.2. The device is placed into the lumpectomy cavity and inflated with a mixture of saline and radiographic contrast agent to fill the cavity [8]. The Ir-192 brachytherapy source is placed at the center of the balloon through a remote afterloader to deliver dose to the treated volume. 3

Figure 1.2 Mammosite balloon and afterloader setup 1.2.3 Treatment planning Treatment planning is the process of determining the target (tumor volume) and optimal technique for radiation delivery. After implantation of the MammoSite balloon catheter, the patient undergoes a CT scan. These CT imagers are used by the Nucletron treatment planning software to determine the area to be treated and those areas to be avoided. First, CT scans are transferred digitally to the treatment planning software to determine gross tumor volume (GTV), clinical tumor volume (CTV), and planning tumor volume (PTV). In Mammosite brachytherapy, the excision of the lumpectomy cavity is defined as the gross tumor volume [9]; PTV is defined as the breast tissue immediately surrounding the balloon to a measured 1 cm distance from the balloon surface [10]. 4

PTV (1cm from balloon surface) 60 cc of Saline + Contrast Balloon diameter (4cm) 192 Ir Source Mammosite catheter Figure 1.3 Definition of PTV of a MammoSite balloon. In MammoSite technique, the PTV is considered to be the same as the CTV because, the implanted balloon moves with the target, therefore, compensation for variability of breathing motion is not needed. Contouring starts with definition of the GTV (Also, the planning target balloon surface) on a central slice of the CT scan, then on each axial CT slice moving superiorly and then inferiorly [11]. Once all of the contours are drawn, then CTV and prescribed dose limits are defined using isodose surfaces. 5

Figure 1.4 A CT image of isodose surfaces of the MammoSite balloon Finally, using dose calculation algorithms, the treatment planning software is used to design a source position sequence in various locations within the implant applicator to build a matching dose distribution according to the isodose surfaces. The amount of dwell time in one location determines the amount of dose delivered. 1.3 Statement of the problem 1.3.1 Effect of Contrast material Computer tomography (CT) simulation is used for treatment planning [11]. Therefore, the balloon is inflated with saline solution and contrast media so it can be imaged through CT. But, the dosimetry of commercially available brachytherapy treatment planning systems like Nucletron Oncentra is based on the assumption that the HDR brachytherapy source is located in a unit density water medium [12]. Because the contrast medium contains elements with high atomic number (I-53) [8, 13], therefore, the balloon and it s contents are no longer equivalent to water. 6

1.3.2 Effects of breast-air interface Most of the treatment planning systems for brachytherapy procedures including Nucletron Oncentra use water-based dosimetry. A homogeneous medium is assumed in this process. But, this may result in errors, because the human body is heterogeneous. Therefore, these treatment planning systems assume that full scatter exists regardless of the amount of tissue density existing beyond and below the prescription line. This assumption is not true, especially when the tissue beyond the prescription line is thin [14]. As an example, when the balloon is closer to breast-air interface, dose contribution due to backscatter radiation is less (Figure 1.5), Therefore, it is important that treatment planning systems consider patient anatomy and heterogeneity when it calculates the target dose rate. Otherwise, it will over estimate the dose in the areas of breast- air and breast- lung interfaces. Figure 1.5 Dose contributions from over-laying tissues. A) Dose from full scatter (including backscatter contribution). B) Dose without backscatter contribution. 7

1.4 Significance of this study Current radiation techniques do not have perfect efficacy. Also, healthy cells are killed as well as malignant cells [15]. This can result in significant patient morbidity. Delivering correct doses with an accuracy of +/- 5% is important in order to successfully treat cancer with radiation [1]. Some treatment options have been created to spare healthy cells while still having enough radiation to damage cancerous cells [1]. However, if the cancer cells are not completely destroyed then they may continue to grow and cause the tumor to grow [15]. This is why it is critical to understand how patient doses change due to contrast and breast air interface so that more accurate doses can be determined [15]. 8

CHAPTER 02 Brachytherapy Dosimetry 2.1 Dose Rate calculation Most of the treatment planning systems used TG-43U1 model for calculation of dose distributions around a brachytherapy source [16]. TG-43 is the first report on the dosimetry of sources used in interstitial brachytherapy [16,17]. Since then due to the improved dosimetry methodologies and dosimetric characterization of particular source models, American Association of Physicists in Medicine (AAPM) introduced TG-43U1 the updated version of TG- 43 [17]. Both of these reports introduce new parameters like air kerma strength ( ) S k, dose rate ( ) ( G L ( r,θ )) ( g L ( r) ) constant Λ, geometry function, radial dose function, and anisotropy function ( F( r,θ )) to calculate brachytherapy dose [16, 17]. These dosimetry parameters consider geometry, encapsulation and self-filtration of the source, the spatial distribution of radioactivity within the source, and scattering in water surrounding the source [16,17]. According to this protocol, the absorbed dose rate distribution around a sealed brachytherapy source at point P (as shown in figure 2.1) with polar coordinates the following formalism [16,17] ( r,θ ) can be determined using 9

( ) Figure 2.1 Schematic diagram of a seed and the point of interest P r,θ used in TG 43. D ( r,θ ) = S k i Λ i ( ) ( ) i g L r G L r,θ G L r = 1,θ = π / 2 ( )i F r,θ ( ) (2.1) 2.1.1 Air kerma strength Air kerma strength, S k, this is the term that represents brachytherapy source strength. It is defined as the product of air kerma rate at a calibration distance ( d) in free space (which is usually chosen as 1 m) along the perpendicular bisector of the source and the square of the ( d 2 ) distance [16,17], S k = K ( d)i d 2 ( ) ( ) S k U 1U = 1cGy cm 2 h 1 Where, K d, is the air kerma rate. The unit of is [16]. (2.2) 10

2.1.2 Dose rate constant The dose rate constant, Λ, for the source and surrounding medium is defined at ( ) θ 0 = π 2 P r 0,θ 0 (1 cm away from the source on its perpendicular bisector where, ). The Λ is expressed in units of cgy hr 1 U 1 [18]. Therefore, dose rate constant, Λ, given as follows [16, 17]: D Λ = ( r 0,θ 0 ) S k (2.3) The value of the dose rate constant Λ depends on the medium surrounding the radiation source, because it indicates the rate of the energy is absorbed by the medium [17, 18]. Therefore, It is taking into account factors like, source geometry, the spatial distribution of radioactivity within the source, encapsulation, and self-filtration within the source and scattering in the water surrounding the source [16, 17]. 2.1.3 Geometry function The Geometry function, r,θ, with units of, describes the decrease in dose with distance from the source [16, 18]. For point source [16, 17]: G L ( ) cm 2 G L ( r,θ ) = r 2 (2.4) For line source [17]: G L ( ) ( r,θ ) = θ θ 2 1 L r Sinθ (2.5) 11

2.1.4 Radial dose function The radial dose function, g L ( r), takes in to account absorption and scatter along the transverse axis of the source; and normalized to the value at 1cm from the source. g L ( r) is determined from the depth dose measurements along the perpendicular axis of the source [18, 17]. g L D ( r) = ( r,θ 0 )i G L r 0,θ 0 D ( r 0,θ 0 )i G L r,θ 0 ( ) ( ) (2.6) 2.1.5 Anisotropy function Anisotropy function, F( r,θ ), accounts for the angular variation of photon absorption and scattering in the medium and in the source encapsulation [16, 17]. This function is expressed as a relative dose measurement, and it is normalized to the measurement at θ 0 = π 2 for each value of r [16, 17]. ( ) = D r,θ F r,θ ( ) ( ) ( )i G L r,θ 0 D ( r,θ 0 )i G L r,θ (2.7) 2.2 Dose calculation Section 2.1 described the estimation of the dose rate value D i ( r,θ ) at a point of interest ( ) D( r,θ ) P( r,θ ) P r,θ. In calculating the absorbed dose value at the point of interest, it was assumed that the source strength expressed as air kerma strength. Also; the dose rate is obtained by considering t 0 = 0 [19]. If the dwell time of the 192 Ir source is T, then the total dose delivered at the point of interest is obtained by using equation 2.6 [19]. Where, λ is the decay constant of the radioisotope. 12

( ) = D i r,θ,t 0 D r,θ t=t ( )e λ (t t 0 ) dt = D i 1 ( r,θ )i λ i 1 e λt t=t 0 =0 ( ) (2.8) But, mean life, τ, is equal to 1 λ D( r,θ ) = D i r,θ,t 0 ( )iτ i 1 e T τ (2.9) When T τ, exponential part of equation 2.9 is e T τ 1 T τ (2.10) Therefore, equation 2.9 can be expressed as D( r,θ ) = D i r,θ,t 0 ( )iτ i 1 1 T τ (2.11) D( r,θ ) = D i ( r,θ,t 0 )it (2.12) ( ) According to the equation 2.12, it can be assumed that the dose, D r,θ, at the particular point of interest, P( r,θ ), is proportional to the dwell time, T. As a result, dwell time, T, can be used as an indicator for dose. 13

CHAPTER 3 Material and method 3.1 Effect of Contrast Material 3.1.1 Phantom In order to conduct this investigation, a clinical MammoSite balloon of 4 cm diameter was used. The MammoSite balloon was placed in the middle of the 30cm X 30cm x 30cm water phantom. A 0.6cc ion chamber was placed parallel to the catheter axis between the water surface and the balloon surface. The distance from water surface to the middle of the chamber axis was set to 7cm to assure full scatter (remove the uncertainty from breast air interference). We also set the distance between the middle of the chamber surface to the MammoSite catheter axis to 5 cm. Both of the holders that are used to hold the MammoSite catheter and ion chamber were made with tissue equivalent materials (similar electron density). These tissue equivalent materials and water simulate the attenuation and scattering of soft tissues. 3.1.2 CT simulation The phantom was filled with water and aligned with positioning lasers on the General Electric Lite Speed 4 slice Computed Tomography (CT) scanner. Then, the MammoSite balloon was then filled with 60 cm 3 of water, which representing 0% contrast. Then, helical CT scans were acquired throughout the phantom with a 1mm slice width. 10% of the water in the balloon was then replaced with an equal volume of a contrast solution and a repeat CT scan was acquired. The contrast medium used in this experiment is Iohexol, which is sold under the trade name called Omnipaque 300 [20]. It has a molecular composition of C 19 H 26 I 3 N 3 O 9 [20]. 14

Omnipaque 300 contains 647 mg of Iohexol equivalent to 300 mg of organic iodine per milliliter [13]. The filling and scanning process of the MammoSite balloon catheter was repeated for the balloon filled with saline and 20% of radiographic contrast concentration, and for the balloon filled with saline and 30% radiographic contrast concentration. Figure 3.1 Cross sectional diagram of the phantom setup for Figure 3.2 Actual phantom setup 15

3.1.3 Treatment planning The CT images of the phantom, with the MammoSite balloon as described in section 3.1.2, were imported to the Nucletron Oncentra treatment planning system for contouring. After the countering, a 100 cgy isodose curve was defined at 5cm from the center of the balloon. The dose of 100 cgy is arbitrary and was chosen for convenience of measurement confirmation. This whole process was repeated for all of the contrast concentrations. 7cm Over-lying water thickness Ion chamber 100cGy 5cm Contoured MammoSite balloon. Isodose curve Figure 3.3 Schematic diagram of a 100 cgy isodose curve appear in treatment planning system 3.1.4 Radiation Delivery The treatment planning data from Nucletron Oncentra treatment planning system was exported to the afterloader. The phantom was placed on a stable counter top, and the MammoSite catheter was connected to the afterloader using a transfer tube. Treatment was delivered to the phantom. This was done by twice to get an average measurement; both dwell times as well as ionization chamber readings were recorded. The same procedure was repeated for the rest of the contrast concentration setups. 16

Figure 3.4 Radiation delivery system from afterloader to MammoSite catheter 3.2 Effect of Breast Air Interface 3.2.1 Phantom The same MammoSite balloon and phantom setup were used as in section 3.1.1. The distance from the center axis of the ion chamber to the center axis of the MammoSite catheter was kept at 5cm. 3.2.2 CT simulation The phantom was filled with water up to the center axis of the ion chamber, where d is equal to zero. Then, helical CT scans were acquired throughout the phantom with 1mm slice width using the same CT scanner described in section 3.12. The filling and scanning process was repeated for different d values, where d was equal to 1cm, 2 cm, 5 cm and 7 cm. This represents the different tissue thicknesses beyond the point of measurements. 17

Figure 3.5 Cross sectional diagram of the phantom setup 3.2.3 Treatment Planning The CT images of the phantom, with the MammoSite balloon as described in section 3.2.2, were imported to the Nucletron Oncentra treatment planning system for contouring. Like section 3.1.3, the 100 cgy isodose curve was defined at 5 cm from the center of the balloon. This was repeated for all of the over-lying water levels. d Over-lying water thickness Ion chamber 5cm 100cGy Contoured MammoSite balloon. Isodose curve Figure 3.6 Schematic diagram of a 100 cgy isodose curve appear in treatment planning system 18

3.2.4 Radiation Delivery The treatment planning data from the Nucletron Oncentra treatment planning system was exported to the afterloader. As described in section 3.1.4, the planned treatment was delivered to the phantom. This was done twice to get an average measurement; both dwell times as well as ionization chamber readings were recorded. The same procedure was repeated for the rest of the over-lying water thickness setups. 3.3 MammoSite Balloon Catheter Central lumen port for Radiation source External lumen port for inflation liquids 4cm diameter balloon Figure 3.7 Mammosite Balloon The MammoSite device used in this experiment consists of a single-lumen catheter and a silicone balloon. The external lumen is used as an inflation channel; and the central lumen accommodates the radioisotope high (e.g., 192 Ir) [21]. The diameter of the spherical balloon is 4 cm when the fill volume equals to 60 cm 3 [21]. 19

3.4 Ion chamber Figure 3.8 0.6 cc PTW Farmer Chamber The 0.6 cm 3 PTW Farmer ion chamber that was used in this experiment is the most convenient as well as accurate dosimeter for measure radiation dose in radiotherapy. Ionization chambers do not measure dose directly, so it is necessary to convert measured ionization to absorbed dose by the chamber [22]. This is done by multiplying the ion chamber reading with, N D,W, which called the dose to water calibration coefficient [22]. Before multiplying by, the raw reading from the ion chamber must be corrected for environmental conditions as well as some aspects of operation of instrument as indicated in the equation 3.1 [22]. N D,W M = M raw i P TP i P ion i P pol i P elec (3.1) P TP Where, is a correction for the temperature (T) and pressure (P) [22]. is a P ion correction factor for the incomplete collection of ions at electrodes in the ion chamber [22]. P pol 20

is the polarity correction factor for differences in the sensitivity of the ionization chamber when the operating voltage is reversed [22]. P elec is the electrometer correction factor for the response of the electrometer to the connected ion chamber [22]. Then this corrected ionization reading, M, is converted to dose in the water at the point of reference by using equation 3.2 [22]: D W = M i N D,W i k Q (3.2) Where, k Q is called the quality conversion factor which depending on the type of ion chamber, and the energy of the radiation [22]. Therefore, using equations 3.1 and 3.2: D W = M raw i( P TP i P ion i P pol i P elec i N D,W i k Q ) (3.3) If we assume temperature and pressure are constant through out the experiment then equation 3.3 becomes, D W = M raw i Constant (3.4) Because, P ion,p pol,p elec, N D,W, and k Q all depend on the ion chamber, electrometer s operating voltage, and radiation energy, which were constant throughout the experiment. Therefore, in this experiment, we can use the direct ion chamber reading ( for radiation dose. M raw ) as an indicator 21

CHAPTER 04 Results 4.1 Effect of Contrast Material The dwell times calculated with the treatment planning system for 100 cgy radiation dose at 5 cm from the MammoSite catheter axis were recorded for different contrast concentrations; and then the dwell time data were normalized to 0% contrast. At the same time readings from 0.6cc Farmer ion chamber were also recorded and normalized to 0% contrast concentration. Contrast concentration (%) Dwell time to deliver 100cGY Reading 1 Reading 2 Average (Seconds) (Seconds) (Seconds) Normalize to 0% concentration 0 266.7 266.7 266.7 1.000 10 268.8 268.8 268.8 1.008 20 268.8 268.8 268.8 1.008 30 268.8 268.8 268.8 1.008 Table 4.1. Recorded dwell times from Nucletron Oncentra treatment planning systems for different contrast concentrations 22

Contrast Ion chamber reading for 100cGy Normalize to 0% concentration (%) Reading 1 Reading 2 Average concentration (nc) (nc) (nc) 0 2.845 2.850 2.848 1.000 10 2.798 2.801 2.800 0.983 20 2.691 2.688 2.690 0.944 30 2.670 2.670 2.670 0.938 Table 4.2. Recorded ion chamber readings for different contrast concentrations Normalized Dose Vs. Contrast Concentration Normalized dose 1.020 1.000 0.980 0.960 0.940 0.920 0.900 0 10 20 30 Contrast Concentration, % (a) TPS (b) Ion chamber Figure 4.1. Comparison of the normalized dose as a function of contrast concentration for the MammoSite balloon placed in cubical water phantom, at 5 cm from catheter axis. a) Delivered dose measurements from Nucletron Oncentra treatment planning system. b) Measured dose measurements from 6cc Farmer chamber. 23

4.2 Effect of Breast Air Interface The dwell times calculated by the treatment planning system for 100 cgy radiation dose at 5c m from the MammoSite catheter axis were recorded for different over-lying water thicknesses; and then data were normalized to over-lying water thickness d, where d is equal to 7cm. At the same time 0.6cc Farmer ion chamber was used to measure the dose at 5 cm from the MammoSite catheter axis for different over-lying water thicknesses; and normalized to d equal to 7 cm (full scatter state). d, over-lying water Dwell time to deliver 100cGy Normalize to full thickness (cm) Reading 1 Reading 2 Average scatter (d = 7cm) (Seconds) (Seconds) (Seconds) 0.0 266.7 266.7 266.7 1.000 0.5 266.7 266.7 266.7 1.000 1.0 266.7 266.7 266.7 1.000 2.0 266.7 266.7 266.7 1.000 5.0 266.7 266.7 266.7 1.000 7.0 266.7 266.7 266.7 1.000 Table 4.3 Recorded dwell times from Nucletron Oncentra treatment planning systems for different over-lying water thicknesses. 24

d, over-lying water Dwell time to deliver 100cGy Normalize to full thickness (cm) Reading 1 Reading 2 Average scatter (d = 7cm) (nc) (nc) (nc) 0.0 2.629 2.631 2.630 0.923 0.5 2.684 2.683 2.684 0.942 1.0 2.696 2.696 2.696 0.947 2.0 2.773 2.793 2.783 0.977 5.0 2.842 2.841 2.842 0.998 7.0 2.845 2.850 2.848 1 Table 4.4. Recorded ion chamber readings for different over-lying water thicknesses 25

Normalized dose Vs. Over- lying water thickness Normalized dose 1.02 1 0.98 0.96 0.94 0.92 0.9 0.88 0 0.5 1 2 5 7 Overline water thickness,d, (cm) (a) TPS (b) Ion chamber Figure 4.2 Comparison of the normalized dose as a function of over-lying water thickness for the MammoSite balloon placed in cubical water phantom, at 5 cm from catheter. a) Delivered dose measurements from Nucletron Oncentra treatment planning system. b) Measured dose from 6cc Farmer chamber. 26

CHAPTER 05 Conclusion 5.1 Effect of Contrast Material Both of these experiments were carried out using the same MammoSite balloon, where diameter was equal to 4 cm. In first experiment, the contrast was changed from a concentration of 0% to 30% while keeping the same phantom setup. According to table 4.2, we observed a prescribed dose reduction from 1.7% to 6.2% at 5 cm away from the balloon axis relative to dose obtained without contrast material. The largest dose reduction occurred when contrast concentration is at its greatest. According to the table 4.1 and figure 4.1, we observed the predicted dose form the treatment planning system was constant with the different contrast concentrations. Therefore, using these observations we can conclude that treatment-planning system does not take into account dose perturbation as a function of different contrast concentrations. Dose perturbations due to contrast concentration has minimal affect on the MammoSite dosimetry as long as the concentration is below 20%. 5.2 Effects of Breast-Air Interface In the second experiment, we changed the over-lying water thickness from 0 cm to 7 cm while keeping the other parameters the same. Also In this experiment we assumed 7cm of overlying water thickness is more than sufficient for full scatter. According to table 4.4, we observed the prescribed dose reduction from 0.21% to 7.7% at 5 cm away from the balloon axis relative to dose obtained with full scatter (when d is equal to 7cm). According to the table 4.3 and figure 27

4.2, we observed that the predicted dose form the treatment planning system was constant with the different over-lying water thicknesses (d). Therefore, the same as section 5.1, we can conclude that the treatment-planning system does not take into account dose perturbation due to the effect of the breast air interface. According to the figure 4.2, when the over-lying water thickness d is less than 1 cm, there is a considerable dose reduction to PTV. Therefore, we can recommend that caution should be used when planning a MammoSite technique to treat lumpectomy cavities close to the skin surface as the predicted dose may be clinically different than that predicted. According to the above results, it is important to develop a treatment planning systems that performs heterogeneity correction for breast air interface, and different contrast materials as well as different contrast concentrations. In this experiment we only considered two variables, therefore, to quantify the uncertainties from the MammoSite brachytherapy technique completely, work needs to be done regarding the uncertainty due to balloon deformation, and source position [8]. 28

REFERENCE [1] Richard T. Hoppe, Theodore Locke Phillips, Mack Roach. Leibel and Phillips Textbook of Radiation Oncology. 3 ed. Philadelphia: Elsevier Saunder, 2010. [2] http://www.cancer.org/research/cancerfactsstatistics/breast-cancer-facts-figures [3] http://mbcc.org/breast-cancer-prevention/index.php/be-informed/breast-cancer-statistics/ [4] http://www.cancer.gov/cancertopics/factsheet/detection/probability-breast-cancer [5] http://www.seer.cancer.gov/archive/csr/1975_2010/results_merged/sect_04_breast. pdf [6] http://www.radiologyinfo.org/en/info.cfm?pg=ebt [7] http://www-naweb.iaea.org/nahu/dmrp/documents/chapter13.pdf [8] Saleh Bensaleh, Eva Bezak. The impact of uncertainties associated with MammoSite brachytherapy on the dose distribution in the breast. Journal of Applied Clinical Medical Physics 2011 Vol 9, No 4. [9] Ritesh Kumar, Suresh Chander Sharma, Rakesh Kapoor, Rajender Singh, Anup Bhardawaj. Dosimetric evaluation of 3Dconformal acceleratedpartial-breast irradiation vs. wholebreast irradiation: A comparative study. International Journal of Applied and Basic Medical Research 2012 Volume 2, Issue 1, Page: 52-57. [10] http://www.aboutcancer.com/rtog0413_details.htm [11] Ann Barrett, Jane Dobbs, Tom Roques. Practical Radiotherapy Planning. 4ed. 2009 29

[12] http://www.elekta.com/healthcare-professionals/products/elekta-brachytherapy /oncentra-brachy.html [13] http://www.medicineonline.com/drugs/o/2754/omnipaque-iohexol-injection-240-300- 350.html [14] Bassel Kassas, Firas Mourtada, John L. Horton, Richard G. Lane, Thomas A. Buchholz, Eric A. Strom. Dose modification factors for 192Ir high-dose-rate irradiation using Monte Carlo simulation. Journal of Applied Clinical Medical Physics 2006 Vol 7, No 3. [15] Eric J. Hall, Amato J. Giaccia. Radiobiology For The Radiologist. 7 ed. Philadelphia: Lippincott Williams & Wilkins, 2012. [16] Brachytherapy Edited by Kazushi Kishi [17] Mark J. Rivard, Bert M. Coursey, Larry A. DeWerd, William F. Hanson, M. Saiful Huq, Geoffrey S. Ibbott Michael G. Mitch, Ravinder Nath, Jeffrey F. Williamson. Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. American Association of Physicists in Medicine 2004. [18] William R. Hendee, Geoffrey S. Ibbott, Eric G. Hendee. Radiation Therapy Physics, 3 rd. 2005. [19] Dimos Baltas, Loukas Sakelliou, Nikolaos Zamboglou. The Physics of Modern Brachytherapy for Oncology. Taylor & Francis 2006 [20] http://en.wikipedia.org/wiki/iohexol [21] Jadwiga B. Wojcicka, a Donette E. Lasher, Ronald Malcom, and Gregory Fortier. Clinical and dosimetric experience with MammoSite -based brachytherapy under the RTOG 0413 protocol. Journal of Applied Clinical Medical Physics 2007 Vol 8, No 4. 30

[22] Peter R. Almond, Peter J. Biggs, B. M. Coursey, W. F. Hanson, M. Saiful Huq, Ravinder Nath, D. W. O. Rogers. AAPM s TG-51 protocol for clinical reference dosimetry of highenergy photon and electron beams. Med Phys. 1999; 26(9): 1847 70TG 51 31