Frictional Resistance in Various Ceramic Brackets using Archwires of different Alloys, Sizes and Cross sections: A Comparative Study

Similar documents
There has been increased use of self-ligating. Friction between various self-ligating brackets and archwire couples during sliding mechanics

In Vitro Frictional Forces Generated by Three Different Ligation Methods

ABSTRACT INTRODUCTION /jp-journals

Frictional Properties of Self-Ligating Brackets and Low-Friction Ligatures

An in vitro investigation of the influence of self-ligating brackets, low friction ligatures, and archwire on frictional resistance

Coated, uncoated stainless steel ligatures versus Self ligation- In the perspective of friction

Effect of passive self-ligating bracket placement on the posterior teeth on reduction of frictional force in sliding mechanics

A Comparison of Frictional Resistance between Active and Passive Self-ligating Brackets with Conventional Bracket Systems

Introduction. Formerly known as Susan David

Frictional resistance in monocrystalline ceramic brackets with conventional and nonconventional elastomeric ligatures

A New Low-Friction Ligation System

The Study of Frictional Force in Self-Ligating Orthodontic Brackets

Comparison of Frictional Resistance in Conventional Brackets with Different Stainless Steel Wires

Tooth movement can occur when the applied

Third-Order Torque and Self-Ligating Orthodontic Bracket Type Effects on Sliding Friction

Analysis of the characteristics of slot design affecting resistance to sliding during active archwire configurations

An Evaluation of Efficiency and Effectiveness of Self-ligating Bracket Systems: A Prospective Clinical Study

Evaluation of methods of archwire ligation on frictional resistance

The Comparison of Space Closure Rate between Conventional and Passive Self-ligating System Using Elastomeric Chain in Maxilla

Friction between different wire bracket combinations in vitro evaluation

Effect of Sodium Fluoride Mouthwash on the Frictional Resistance of Orthodontic Wires

A study of the frictional characteristics of four commercially available self-ligating bracket systems

저작권법에따른이용자의권리는위의내용에의하여영향을받지않습니다.

The role of friction in orthodontics

Role of lubricants on friction between self-ligating brackets and archwires

In vitro comparative study on the friction of stainless steel wires with and without Orthospeed (JAL 90458) on an inclined plane

FRICTION PRODUCED BY ESTHETIC BRACKETS WITH VARYING LIGATION CHRISTINE KNOX ABENOJA

Plus Combines Traditional Twin Bracket Treatment With Self-Ligation Convenience

HYCON DEVICE: A PRECISE AND CONTROLLED METHOD OF SPACE CLOSURE A CASE REPORT

Characterizing constraining forces in the alignment phase of orthodontic treatment

Evaluation of frictional forces generated by different brackets and orthodontic wires

The management of impacted

Is nickel titanium superior to multistranded stainless steel wire in aligning crowded lower anteriors? A comparative in-vivo study.

INDICATIONS. Fixed Appliances are indicated when precise tooth movements are required

Frictional properties of aesthetic brackets

A comparative study of the static and kinetic frictional resistance of titanium molybdenum alloy archwires in stainless steel brackets

Key words: Friction, orthodontic brackets, finite element analysis.

Dual Force Cuspid Retractor

Advantages. Fantastic. Treatment. Finishes. 3M Self-Ligating Appliances Lateral Development System. Unitek Lateral Development Archwires

Advantages. Fantastic. systems. Together with Unitek Lateral Development Archwires, this system puts you in control to

Experience with Contemporary Tip-Edge plus Technique A Case Report.

The Loading And Unloading Properties Of Various Arch Wires As A Function Of Cross Sectional Dimension And Inter Bracket Span Width System

Measurements of the torque moment in various archwire bracket ligation combinations

RESEARCH ARTICLE /jp-journals

Tahereh Hosseinzadeh Nik 1, Tabassom Hooshmand 2*, Habibeh Farazdaghi 3, Arash Mehrabi 1 and Elham S Emadian Razavi 1

Frictional resistance exerted by different lingual and labial brackets: an in vitro study

Introducing the Clarity SL Self-Ligating Appliance System

The dynamic frictional resistance between orthodontic brackets and arch wires

In vitro evaluation of resistance to sliding in selfligating and conventional bracket systems during dental alignment

Archwire seating forces produced by different ligation methods and their effect on frictional resistance

Sliding mechanics in an extraction case treated with 3M Clarity ADVANCED Ceramic Brackets and the 3M MBT Appliance System.

Integrative Orthodontics with the Ribbon Arch By Larry W. White, D.D.S., M.S.D.

The Tip-Edge appliance and

A review of the orthodontic literature

Dental Materials Journal 2015; 34(3): Keisuke TOCHIGI, Souichiro ODA and Kazuhito ARAI INTRODUCTION

Evaluation of Torsional Strength of Ceramic Brackets Produced by Arch Wire Twisting Moment

Canine Extrusion Technique with SmartClip Self-Ligating Brackets

Gentle-Jumper- Non-compliance Class II corrector

Use of a Tip-Edge Stage-1 Wire to Enhance Vertical Control During Straight Wire Treatment: Two Case Reports

wiirre e a a n n d d w wiirre e ffo orrms ms

With judicious treatment planning, the clinical

Controlled Space Closure with a Statically Determinate Retraction System

ANTERIOR AND CANINE RETRACTION: BIOMECHANIC CONSIDERATIONS. Part One

Evaluation of surface roughness of the bracket slot floor a 3D perspective study

AL-AZHAR. Dental Journal. Print ISSN Online ISSN ADJ-for Girls, Vol. 5, No. 1, January (2018) PP. 9:14

ORTHOdontics SLIDING MECHANICS

Victory Series Active Self-Ligating Brackets. Reliable. and. Effective

Thermal Nitanium. Archwires. Reduced Friction. Accelerated Results.

Comparative Biomechanics of Labial versus Lingual Fixed Appliances - A Review

LIMITING FACTORS AFFECTING STAGE I TREATMENT: A BIOMECHANICAL PERSPECTIVE. Christopher G. Gibson. Chapel Hill 2016

GAC Archwire Collection

The effect of surface treatment and clinical use on friction in NiTi orthodontic wires

SnapLink Re-Convertible Tube

Delta Force. Bracket System. Putting you in the driver s seat for ultimate control

Mollenhauer Aligning Auxiliary for Bodily Alignment of Blocked-out Lateral Incisors in Preadjusted Edgewise Appliance Therapy

SE NiTi Heat Activated NiTi Ultra T Heat Activated NiTi Stainless Steel Beta Titanium. Ultra T Archwires

Study Model-based Evaluation of Built-in Tip, Torque, and In out Characteristics of a Third-generation Preadjusted Edgewise Appliance

Developing lingual orthodontics

.::...;;;,:..=.::..==;::..;=== RESEARCH ARTICLE -

Effect of Brackets Types on the Amount of Movement and Rotation of Canine during Retraction: A Simulated In vitro Study

A comparative in vivo evaluation of the alignment efficiency of 5 ligation methods: A prospective randomized clinical trial

Aesthetic Brackets and Accessories

Effects of a diamond-like carbon coating on the frictional properties of orthodontic wires

In the last decade, there has been a significant increase

Self-ligating versus Invisalign: analysis of dento-alveolar effects

Treatment of an open bite case with 3M Clarity ADVANCED Ceramic Brackets and miniscrews.

MATERIAL CHARACTERISTICS OF THE ORTHODONTIC ARCHWIRES

Correlation Between Frictional Force and Surface Roughness of Orthodontic Archwires

Bracket Systems. A good bracket is the foundation of a good treatment plan... Go ahead, choose the best. Metal Bracket.

Molar anchorage loss has been shown to occur

Comparison of mechanical properties of beta-titanium wires between leveled and unleveled brackets: an in vitro study

6. Timing for orthodontic force

An Accurate Methodology to detect Leaching of Nickel and Chromium Ions in the Initial Phase of Orthodontic Treatment: An in vivo Study

The Science Behind The System Clinical Abstracts, Volume 2

(12) Patent Application Publication (10) Pub. No.: US 2005/ A1

The Effect of Using Self-ligating Brackets on Maxillary Canine Retraction: A Split-mouth Design Randomized Controlled Trial

A comparative assessment of torque generated by lingual and conventional brackets

Simply Great Wire! Great prices.

Evaluation of space closure rate during canine retraction with nickel titanium closed coil spring and elastomeric chain.

04 Inserting the HYCON TUBE

Transcription:

Research Article Frictional Resistance in Various Ceramic Brackets using Archwires of different Alloys, Sizes and Cross sections: Dr Yash Raj Bahadur. 1 Dr Deepak Kumar Agarwal, 2 Dr Ankur Gupta, 3 Dr P Narayana Prasad 4 1 Senior Lecturer, 4 Professor, Dept of Orthodontics, Seema Dental College, Rishikesh 2 Professor, 3 Reader, Dept of Orthodontics, Institute of Dental Sciences, Bareilly, India Correspondence: Dr Yash Raj Bahadur; Email: dr.yash.ortho@gmail.com ABSTRACT Introduction: Friction at the bracket-archwire interface has been observed as one of the most important factors affecting tooth movement. Hence it is importance to assess the friction generated during tooth movement to bring about optimal treatment results. Objective: To compare the frictional resistance of various ceramic brackets using different archwires, and to compare the static and kinetic frictional force of various ceramic brackets using different archwires. Materials & Method: The present study evaluated and compared the friction generated at the bracket archwire interface when.18 and.19 x.25 stainless steel archwires and.19 x.25 teflon coated stainless steel archwires were moved through conventionally ligated, passive self-ligating and interactive self-ligating ceramic brackets. The static and kinetic frictional forces were also evaluated and compared. Result: Highly significant differences in kinetic (p<.1) and static (p<.1) frictional forces were observed in all three groups when used with the different archwires. On comparing the static and kinetic frictional forces significant differences were observed among all three groups (p<.5). Conclusion: The passive self-ligating brackets produce the least frictional forces when compared to interactive self-ligating and conventionally ligated brackets. Also, the static frictional forces were found to be more as compared to kinetic frictional forces Key words: archwire, ceramic brackets, friction, self-ligating bracket INTRODUCTION Friction is defined as a force tangential to the common boundary of two bodies in contact that resists the motion or tendency to motion of one relative to the other. 1 It exists in two forms: static friction, which is the resistance that prevents actual motion, and dynamic (kinetic) friction, which exists during motion. 2 The efficiency of fixed appliance therapy depends on the fraction of force delivered with respect to the force applied; hence, high frictional forces due to the interaction between the bracket and the guiding archwire affect treatment outcomes and duration in a negative manner. 3 When sliding mechanics are used, friction occurs at the bracket-wire interface. Some of the applied force is therefore dissipated as friction, and the remainder is transferred to supporting structures of the tooth to mediate tooth movement. 4 Schumacher et al stated that friction is determined mostly by nature of ligation. Self-ligating brackets were introduced in the mid-193s in the form of Russell attachment. Selfligating brackets are ligatureless bracket systems that have a mechanical device built into the bracket to close the slot. 5 Self-ligating brackets can be classified as active or passive; based on the mode of self-ligation. Active appliances have a spring clip that functions as the fourth wall of the bracket slot, hence making positive contact with the archwire. Passive appliances have a movable labial side that creates a hollow tube inside the bracket during closure. 6 Contemporary orthodontics provides service to great number of adults, especially women, thus the need for optimum cosmetic appearance of orthodontic appliances has been emphasized. As a result, ceramic brackets were introduction in 1986. 7 Thus the present study was conducted to compare the frictional forces between various ceramic brackets and archwires and also to compare the static and kinetic frictional forces respectively. 18

Figure 1: Universal Testing Machine (Instron) Figure 2: Custom made apparatus with test sample Figure 3: Test sample attached to Instron machine MATERIALS AND METHOD The present study was performed to evaluate the friction between different types of orthodontic ceramic brackets and wire of different sizes, cross-sections and alloys in the Department of Orthodontics, Institute of Dental Sciences, Bareilly in association with Northern India Textile Research Association, Ghaziabad. All testing was done on Universal Testing Machine (Instron) (Figure 1). The study consisted of eight millimeter long.18 stainless steel (3M Unitek),.19 x.25 stainless steel (3M Unitek) and.19 x.25 Teflon coated stainless steel (Spectra, GAC Dentsply) working archwires which were moved through three types of maxillary premolar brackets which were grouped as: Group 1 : Interactive self-ligating ceramic bracket (InOvation C, GAC Dentsply) Group 2 : Passive self-ligating ceramic bracket (Damon Clear, Ormco) Group 3 : Conventionally ligated ceramic bracket (Transcend, 3M Unitek) The brackets and archwires were cleaned with alcohol wipe before performing the tests. Lower end of each test unit was attached to a custom made apparatus which was itself attached to the lower jaw of the testing machine. The custom made apparatus was made as a heavy acrylic block (2mm length x 2mm width x 2mm height) with a hollow sheath (having.36 diameter and 21mm length), which was constructed to hold the wire parallel to the vertical framework of the universal Testing Machine (Figure 2). This apparatus was made to be free to undergo rotational movement by making a 2mm right-angle bend in the archwire at the lower end. It was ensured that the bracket and archwire specimens could self-align during a test run, allowing tip and torque to be effectively eliminated as variables so that the effect of ligation method on friction could be studied in isolation. The sample was arranged parallel to the vertical framework of the machine. The bracket was then pulled in a vertical direction by a loop of.18 stainless steel wire which was fixed at upper jaw of the machine (Figure 3). The force required for initiating and maintaining the movement of the bracket over an 8mm test distance was recorded. Each bracket and archwire was used for 1 tests and after 1 tests the sample was changed. 8 The static frictional force was measured as the initial rise or peak force required to initiate movement of the wire through the bracket and then halved. The drawing force required to maintain the movement beyond the point of initial displacement was averaged and then halved and recorded as the kinetic friction. Data were obtained at a rate of five readings per second. 9 Descriptive statistics were determined for all groups, which were used for comparison between the groups. The data obtained were analyzed using MS Excel and SPSS. The comparison of data obtained for various bracket archwire ligature groups was done by Kruskal-Wallis H test and Wilcoxon signed rank test. The comparison of the frictional forces between the three groups was done using Kruskal- Wallis H test. 19

Brackets Table 1: Comparison of kinetic frictional force between three groups Archwires Mean (gm) SD Mean (gm) SD Mean (gm) SD Group 1 43.4 4.4 32.39 9.27 Group 2 3.58 2.38 3.35 2.1 Group 3 37.92 7.23 83.37 8.54 65.81 7.27 1 Vs 2 Vs 3 (Kruskal Wallis H test) *Statistically significant at p<.5 2 groups empty, test not performed H(2) = 123.373, df= 18, p=.* H(2)=117.828, df= 18, p=.* RESULT Comparison of static frictional forces among three groups showed highly significant differences with the maximum static frictional forces being generated in conventionally ligated ceramic and the least being produced in passive self-ligating ceramic brackets. Both the ceramic self-ligating brackets showed no static frictional forces in conjunction with.18 stainless steel wires (Table 2). Similar results were observed when the kinetic frictional forces were evaluated (Table 1). Comparison of the static and kinetic frictional forces in each of the three groups showed significant differences in all groups with the static frictional force being more in comparison to the kinetic frictional forces (Graph 1,2,3). DISCUSSION Friction is the force that resists the movement of one surface past another and acts in a direction opposite the direction of movement. It is a function of the relative roughness of two surfaces in contact. 1 Frictional force has two components; the initial friction between the arch wire and the bracket when a force is applied is termed as static friction and must be overcome to initiate tooth movement. As the tooth is moving, the second component to friction termed kinetic friction occurs with the arch wire moving in the direction of the applied force as it is guided through the molar and premolar bracket slots. 11 Ceramic brackets were introduced in orthodontics to meet increasing aesthetic demand, along with concerns about the effect of corrosion and reaction to Nickel that leaches out of stainless steel-based material in oral cavity. These brackets are aesthetically superior and more biocompatible in nature. 12 The present day scenario in orthodontics emphasizes the aesthetic utility of ceramic brackets. However, the increased frictional resistance generated by ceramic brackets diminishes its effectiveness in clinical use. Hence the ceramic self-ligating brackets were introduced to combine the superior aesthetics of ceramic brackets with the reduced friction of self-ligating brackets. 13 The.18 round and.19 x.25 rectangular wires are used in the present study to assess the effect of wires of different cross-sections and sizes on the frictional forces generated in different brackets. Earlier the stainless steel wires were the mainstay for orthodontic usage, but the aesthetic concerns have resulted in the evolution of newer aesthetic wires. Teflon or polytetrafluoroethylene is a material characterized by a completely fluoridated chain. It is an anti-adherent and aesthetic material that has excellent chemical inertia as well as good mechanical stability. 14 In the present study, when considering the.18 SS archwire, no frictional forces were generated with Group 1 and 2 whereas considerable static and kinetic frictional forces were generated with Group 3. Our findings are in agreement with studies conducted by Farronato et al, 14 Singh et al, 15 Brauchli et al 16 who stated that the active and interactive self-ligating brackets did not demonstrate any frictional forces with.18 SS wires and their conventional counterparts generated considerable frictional forces. Brackets Table 2: Comparison of static frictional force between three groups Archwires Mean (gm) SD Mean (gm) SD Mean (gm) SD Group 1 45.87 3.92 34.9 3.62 Group 2 5.9 2.24 5.31 2.19 Group 3 39.4 3.71 85.16 4.34 68.1 3.96 1 Vs 2 Vs 3 (Kruskal Wallis H test) *Statistically significant at p<.5 2 groups empty, test not performed H(2) = 141.783, df= 18, p=.* H(2) = 125.119, df= 18, p=.* 2

Group 2 demonstrated levels of friction that were considerably lower than that of the Group 1 during sliding mechanics with a.19 x.25 rectangular wire. The static and kinetic frictional forces exerted by Group 2 were minimal when compared to Group 1. Similarly the static and kinetic frictional forces exerted by Group 1 and 2 were lesser as compared to Group 3 in the present study. These results are in accordance with the studies conducted by Cacciafesta et al, 5 Stefanos et al, 9 Singh et al, 15 Brauchli et al, 16 Kannan et al, 17 Oliver et al, 18 Voudaris et al 19 who stated that interactive self-ligating brackets have higher static and kinetic frictional forces compared to passive selfligating brackets when coupled with.19 x.25 SS wires, which is lesser than that generated by the conventionally ligated brackets. When frictional forces were assessed during sliding mechanics with.19 x.25 teflon coated stainless steel wire, it was found that, Group 2 demonstrated the least frictional force followed by Group 1. The maximum force was generated when Group 3 were used. These findings are in agreement with the findings reported by Farronato et al 14 who reported that Teflon coated archwires produced lower frictional levels than their corresponding uncoated counterparts. The friction values in.19 x.25 SS archwires were more for all groups when compared to the values with.18 SS wires. This finding confirms the studies of Singh et al 15 and Tecco et al 2 that all brackets show higher frictional forces as the wire size is increased. Graph 1: Comparison of static and kinetic frictional forces in Group 1 5 4 43.4 45.87 3 2 32.39 34.9 Kinetic 1 7.18 SS.19 x.25 SS.19x.25 teflon coated SS Graph 2: Comparison of static and kinetic frictional forces in Group 2 6 5 5.9 5.31 4 3 3.58 3.35 Kinetic 2 1 Graph 3: Comparison of static and kinetic frictional forces in Group 3 9 8 7 6 83.37 85.16 68.1 65.1 Kinetic 5 4 3 37.2 39.4 2 1 21

The present study showed that the static friction was higher than kinetic friction in all brackets and archwire combinations. Highly significant differences were seen in the results of the present study when comparing the static and kinetic frictional force in Group 1, Group 2 and Group 3. This is in accordance with the studies conducted by Cacciafesta et al, 5 Krishnan et al, 6 Stefanos et al, 9 Downing et al, 21 Kusy et al 22 who stated that the static friction is always greater than kinetic friction irrespective of the bracket or archwire combination used (Graph 1,2,3). Since the present study was conducted in vitro, it has certain limitations. The test conditions may not really correspond to orthodontic tooth movement. Also the study was not designed to replicate the sliding mechanics according to the dental arch and the effect of saliva was not considered as no wet conditions were simulated. CONCLUSION The present study enables to conclude the following: Self-ligating brackets produce less friction as compared to conventionally ligating brackets. Passive self-ligating brackets produce lesser friction than interactive selfligating brackets. As the archwire size and cross-section changes, the frictional resistance also changes. When.18 stainless steel wires are used with self-ligating brackets, no friction was generated and the frictional resistance increased when.19 x.25 stainless steel wires are used. frictional force is always greater than kinetic friction, irrespective of bracket type or archwire used. OJN REFERENCES 1. Drescher D, Bourauel C, Schumacher HA. Frictional forces between bracket and arch wire. Am J Orthod Dentofacial Orthop. 1989; 96:397-44. 2. Tselepis M, Brockhurst P, West VC. The dynamic frictional resistance between orthodontic brackets and arch wires. Am J Orthod Dentofacial Orthop. 1994; 16:131-8. 3. Baccetti T, Franchi L. Friction produced by types of elastomeric ligatures in treatment mechanics with the preadjusted appliance. Angle Orthod. 26; 76:211 6. 4. Nanda R, Gosh J. Biomechanical considerations in sliding mechanics. W.B. Saunders Company, 1996. 5. Cacciafesta V, Sfondrini MF, Ricciardi A, Scribante A, Klersy C, Auricchio F. Evaluation of friction of conventional and metal-insert ceramic brackets in various bracket-arch wire combinations. Am J Orthod Dentofacial Orthop. 23; 124:43-9. 6. Krishnan M, Kalathil S, Abraham KM. Comparative evaluation of frictional forces in active and passive self-ligating brackets with various archwire alloys. Am J Orthod Dentofacial Orthop. 29; 136:675-82. 7. Karamouzos A, Athanasiou A, Papadopoulos M. Clinical characteristics and properties of ceramic brackets: A comprehensive review. Am J Orthod Dentofacial Orthop. 1997; 112:34-4. 8. Hain M, Dhopatkar A, Rock P. The effect of ligation method on friction in sliding mechanics. Am J Orthod Dentofacial Orthop. 23; 123:416-22. 9. Stefanos S, Secchi AG, Coby G, Tanna N, Mante FK. Friction between various self-ligating brackets and archwire couples during sliding mechanics. Am J Orthod Dentofacial Orthop 21; 138:463-7. 1. Kapila S, Angolkar PV, Duncanson MG Jr., Nanada RS. Evaluation of friction between edgewise stainless steel brackets and orthodontic wires of four alloys. Am J Orthod Dentofacial Orthop. 199; 98:117-26. 11. Bednar JR, Gruendeman GW. The influence of bracket design on moment production during axial rotation. Am J Orthod Dentofacial Orthop. 1993; 14:254-61. 12. Nishio C, Foneca A, Nelson J. In-vitro evaluation of frictional forces between arch wires and ceramic brackets, Am J Orthod Dentofacial Orthop. 24; 125:56-64. 13. Oliver CL, Daskalogiannakis J, Tomson BD. Archwire depth is a significant parameter in the frictional resistance of active and interactive, but not passive self-ligating brackets. Angle Orthod. 211; 81:136-44. 14. Farronato G, Maijer R, Carìa MP, Esposito L, Alberzoni D and Cacciatore G. The effect of Teflon coating on the resistance to sliding of orthodontic archwires. Eur J Orthod. 212; 34:41 7. 15. Singh S, Hamsa PRR, Shetty P, Desai M. Comparison of friction between conventional and self-ligating brackets using various archwire combination: An in vitro study. Int J Dent Clinics. 212; 4:27-31. 16. Brauchli LM, Senn C, Wichelhaus. Active and passive self-ligation: A myth? Angle Orthod. 211; 81:312-8. 17. Kannan MS, Murali RV, Jayanth V, Annamalai PR. Comparison of frictional resistance of aesthetic and semi aesthetic self-ligating brackets. Ind J Multidisciplin Dent. 212; 2:551-5. 18. Oliver CL, Daskalogiannakis J, Tomson BD. Archwire depth is a significant parameter in the frictional resistance of active and interactive, but not passive self-ligating brackets. Angle Orthod. 211; 81:136-44. 19. Voudaris JC, Schismenos C, Lackovic K, Kuftinec MM. Self-ligation aesthetic brackets with low frictional resistance. Angle Orthod. 21; 8:188-94. 2. Tecco S, Festa F, Caputi S, Traini T, Di Iorio D, D Attilio M. Friction of conventional and self-ligating brackets using a 1 bracket model. Angle Orthod. 25; 75:141-5. 21. Downing A, McCabe J, Gordon P. A study of frictional forces between orthodontic brackets and archwires. Br J Orthod. 1994; 21:349-57. 22. Kusy RP, Whitley JQ. Friction between different wire-bracket configurations and materials. Semin Orthod 1997; 3:166-77. 22