Article. Reference. Comparison of organ doses and image quality between CT and flat panel XperCT scans in wrist and inner ear examinations

Similar documents
X-Ray & CT Physics / Clinical CT

Computed tomography Acceptance testing and dose measurements

Clinical Study Imaging of Electrode Position after Cochlear Implantation with Flat Panel CT

Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method

CT Quality Control Manual FAQs

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

Toshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery

Managing Radiation Risk in Pediatric CT Imaging

GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT

Combined Anatomical and Functional Imaging with Revolution * CT

Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer

Original Research THE USE OF REFORMATTED CONE BEAM CT IMAGES IN ASSESSING MID-FACE TRAUMA, WITH A FOCUS ON THE ORBITAL FLOOR FRACTURES

Patient dose assessment of CT perfusion scanning at the RSCH

Austin Radiological Association Nuclear Medicine Procedure PET SODIUM FLUORIDE BONE SCAN (F-18 NaF)

Bone Densitometry Radiation dose: what you need to know

Assessment of effective dose in paediatric CT examinations

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Why is CT Dose of Interest?

R/F. Clinical Experience Using the SONIALVISION safire II Utility of Tomosynthesis in Orthopedic Surgery

Paediatric Dose Reduction and Image Quality

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

Dental Cone Beam CT. What is Dental Cone Beam CT?

Hi RES Extremity - (04/18/2011) CTDI: ~13 mgy per acquisition Used for evaluation of: Ankle Elbow Hand Wrist Foot /Calcaneous Toes Fingers

Imaging of the Paranasal Sinuses

CT Optimisation for Paediatric SPECT/CT Examinations. Sarah Bell

Unrivaled, End-to-End

True Dual Energy. Dr. Stefan Ulzheimer, Siemens Healthcare GmbH. DEfinitely Siemens

Computed tomography. Department of Radiology, University Medical School, Szeged

CT NUMBER ACCURACY ANALYSIS FOR RADIOTHERAPY TREATMENT PLANNING IMAGING

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Study of Mastoid Air Cells Diseases using Spiral CT

Doses from Cervical Spine Computed Tomography (CT) examinations in the UK. John Holroyd and Sue Edyvean

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Optimizing radiation dose by varying age at pediatric temporal bone CT

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Cone-Beam CT for MSK Extremities

PACS: ERGONOMIC CONSIDERATIONS 1

Bringing the power of imaging to your patient. Portable full body 32-slice CT scanner

The value of weight-bearing functional CT scans

Chief Radiographer TEI Clinical Associate 2016

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Studies in both the United States and Europe have revealed that computed tomographic (CT) examinations account for only up to 15% of all imaging exami

Accounting for Imaging Dose

Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose)

BHARAT SEVAK SAMAJ NATIONAL DEVELOPMENT AGENCY, PROMOTED BY GOVERNMENT OF INDIA CT SCAN TECHNOLOGY COURSE ONE YEAR EXAMINATION

ESTABLISHING DRLs in PEDIATRIC CT. Keith Strauss, MSc, FAAPM, FACR Cincinnati Children s Hospital University of Cincinnati College of Medicine

Dual-Energy CT: The Technological Approaches

HI-Res Extremity Sensation 16

Ultrasound. Computed tomography. Case studies. Utility of IQon Spectral CT in. cardiac imaging

Radiologic Imaging Magnetic Resonance Imaging (MRI)

A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Breast CT and Dosimetry

Austin Radiological Association Ga-68 NETSPOT (Ga-68 dotatate)

Disclosures. Outline. Learning Objectives. Introduction. Introduction. Stereotactic Breast Biopsy vs Mammography: Image Quality and Dose.

Computed Tomography (CT) - Head

Dosimetric Consideration in Diagnostic Radiology

JMSCR Vol 06 Issue 10 Page October 2018

An Introduction to Dual Energy Computed Tomography

CT Imaging at the Point-of-Care

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

STRUCTURED EDUCATION REQUIREMENTS EFFECTIVE: JANUARY 1, 2016

B. CT protocols for the spine

Preparing for Medical Physics Components of the ABR Core Examination

Estimating Iodine Concentration from CT Number Enhancement

Managing the imaging dose during image-guided radiation therapy

Practical CT and MRI Anthony J. Fischetti, DVM, MS, DACVR Department Head of Diagnostic Imaging The Animal Medical Center, New York OBJECTIVE:

CURRENT CT DOSE METRICS: MAKING CTDI SIZE-SPECIFIC

ACR MRI Accreditation: Medical Physicist Role in the Application Process

SUMMARY AND EXTRACTS FROM THE 2010 GUIDANCE ON THE SAFE USE OF DENTAL CONE BEAM CT (COMPUTED TOMOGRAPHY) EQUIPMENT

Vascular. Experiences Using an Angiography System with a Inch FPD - CT-Like Imaging of the Neck and Abdomen - 1.

Multislice CT versus cone beam CT: dosimetric and image quality comparision.

BioMedical quantitative X-Ray Imaging. Emmanuel Brun Researcher Inserm Université Grenoble Alpes

Eye Doses in Head CT; Sequential Vs Spiral. CT Users Group October 2010

Clinical management of metallic orbital. Helical and conventional CT in the imaging of metallic foreign bodies in the orbit. Material and Methods

CT Dosimetry in the Clinical Environment: Methods and Analysis

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Full ultrasound breast volumes. Faster scans. Streamlined workflow. ACUSON S2000 Automated Breast Volume Scanner. Answers for life.

Translating Protocols Across Patient Size: Babies to Bariatric

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Ask EuroSafe Imaging. Tips & Tricks. Paediatric Imaging Working Group. Shielding in pediatric CT

Radiology Review Course Hotel del Coronado Coronado, California

Cone Beam Computed Tomography: Clinical Applications in Musculoskeletal Imaging.

1. Axial view, left temporal bone. Plane through the upper antrum (A), superior semicircular canal (SSC) and IAC.

Radiographic myelography and postmyelographic CT using

CT examination is a high-radiation-dose imaging technique

Emerging Applications in Musculoskeletal CT Imaging

Radiation Dose Reduction: Should You Use a Bismuth Breast Shield?

Skyscan 1076 in vivo scanning: X-ray dosimetry

The print quality of this copy is not an accurate representation of the original.

Dosimetric Evaluation of Multislice CT Using Anthropomorphic Head Phantom

A feasibility study of multislice X-ray CT imaging of gel dosimeters using the zero scan method

CT Dose Estimation. John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor and Vice Chair of Radiology. University of California Davis Medical Center

btaining a minimally adequate radiographic

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

ph fax

Transcription:

Article Comparison of organ doses and image quality between CT and flat panel XperCT scans in wrist and inner ear examinations DAMET, J, et al. Abstract The aim of this study was to evaluate and compare organ doses delivered to patients in wrist and petrous bone examinations using a multislice spiral computed tomography (CT) and a C-arm cone-beam CT equipped with a flat-panel detector (XperCT). For this purpose, doses to the target organ, i.e. wrist or petrous bone, together with those to the most radiosensitive nearby organs, i.e. thyroid and eye lens, were measured and compared. Furthermore, image quality was compared for both imaging systems and different acquisition modes using a Catphan phantom. Results show that both systems guarantee adequate accuracy for diagnostic purposes for wrist and petrous bone examinations. Compared with the CT scanner, the XperCT system slightly reduces the dose to target organs and shortens the overall duration of the wrist examination. In addition, using the XperCT enables a reduction of the dose to the eye lens during head scans (skull base and ear examinations). Reference DAMET, J, et al. Comparison of organ doses and image quality between CT and flat panel XperCT scans in wrist and inner ear examinations. Radiation Protection Dosimetry, 2010, vol. 139, no. 1-3, p. 164-8 DOI : 10.1093/rpd/ncq062 PMID : 20200104 Available at: http://archive-ouverte.unige.ch/unige:25669 Disclaimer: layout of this document may differ from the published version.

Radiation Protection Dosimetry (2010), Vol. 139, No. 1 3, pp. 164 168 Advance Access publication 3 March 2010 doi:10.1093/rpd/ncq062 COMPARISON OF ORGAN DOSES AND IMAGE QUALITY BETWEEN CT AND FLAT PANEL XPERCT SCANS IN WRIST AND INNER EAR EXAMINATIONS J. Damet 1, *, M. Sans-Merce 1, F. Miéville 1, M. Becker 2, P. A. Poletti 2, F. R. Verdun 1 and S. Baechler 1 1 Institute of Radiation Physics, University Hospital Center and University of Lausanne, Grand Pré 1, 1007 Lausanne, Switzerland 2 Department of Radiology, Radiodiagnostic Service, University Hospital of Geneva, Rue Gabrielle-Perret- Gentil 4, 1211 Geneva 14, Switzerland *Corresponding author: jerome.damet@chuv.ch The aim of this study was to evaluate and compare organ doses delivered to patients in wrist and petrous bone examinations using a multislice spiral computed tomography (CT) and a C-arm cone-beam CT equipped with a flat-panel detector (XperCT). For this purpose, doses to the target organ, i.e. wrist or petrous bone, together with those to the most radiosensitive nearby organs, i.e. thyroid and eye lens, were measured and compared. Furthermore, image quality was compared for both imaging systems and different acquisition modes using a Catphan phantom. Results show that both systems guarantee adequate accuracy for diagnostic purposes for wrist and petrous bone examinations. Compared with the CT scanner, the XperCT system slightly reduces the dose to target organs and shortens the overall duration of the wrist examination. In addition, using the XperCT enables a reduction of the dose to the eye lens during head scans (skull base and ear examinations). INTRODUCTION Diagnostic examinations for the identification of carpal ligamentous disruptions (e.g. rupture of the scapholunate ligament) are usually 2-fold and require the use of an X-ray installation and a computed tomography (CT) scan. Contrast enhancing agent is first injected in the wrist to assess the presence of dye leakage through the scapholunate joint with CT arthrography images (1). The flat-panelbased XperCT Allura FD20 system (2) (Philips Healthcare, Best, The Netherlands) enables performing the whole examination at once, thus reducing the overall duration of the examination and avoiding the need to move patients into another diagnostic room. Diagnostic examinations for the identification of lesions or diseases in the petrous bone or for performing CT-guided interventional procedures of the skull base, such as fine needle aspiration biopsies, are also 2-fold (3). The first step is an X-ray radiography for screening purposes followed by a high-resolution multislice spiral CT examination. It is well known that head CT scans result in significant radiation exposure to the lens of the eye (4). Using the XperCT system may reduce the dose delivered to the eye lens. The present study compares both techniques, for wrist and head scans, with regard to dose assessment and image quality. The aim was to evaluate and compare the quality of reconstructed images as well as doses delivered to patients using a Siemens Somaton Sensation 64 multislice spiral CT (Siemens Healthcare, Erlangen, Germany) and a detector angiographic system, Allura FD 20(Philips Health Care, Best, The Netherlands). Measurements were performed under different XperCT operating modes. The dose to the target organ, i.e. the wrist or the petrous bone, together with those to most radiosensitive nearby organs, i.e. the thyroid and the eye lens, were measured and compared. METHODS Dose measurements Measurements of radiation dose were performed on in-house-developed head and wrist phantoms using thermoluminescent detectors (TLDs) and a Radcal 3036 dosemeter with an external 11-cm 3 ion chamber. The hand phantom consisted of a human hand skeleton embedded in resin and a 4-cm thick Plexiglas 10 10 cm 2 plate. Two phantoms were used for the head: one with a skull embedded in transparent resin for the measurements of doses from scatter radiation to the thyroid and the eye lens and one made of an empty skull that can be opened to place detectors precisely on the petrous bone. For the XperCT system, the X-ray tube and the flat-panel detector were mounted on a C-arm structure to provide radiographic projections over an angle of 2408. Several operating modes were considered: 300 images ation, 600 images ation and 3D rotational angiography (3DRA) mode using 120 # The Author 2010. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org

images ation. For the multislice spiral CT, all measurements of radiation exposure on phantoms were performed using standard protocols for the wrist (120kV/250mAs, 0.18mm reconstruction thickness on XperCT and 120kV/195mAs/pitch of 0.9, 0.4 0.2mm reconstruction thickness on CT) and petrous bone (120kV/250mAs, 0.18mm reconstruction thickness on XperCT and 120kV/430mAs/pitch of 0.8, 0.4 0.2mm reconstruction thickness on CT) examinations. ORGAN DOSES AND IMAGE QUALITY Wrist examination Measurements were performed placing the Radcal ion chamber on the 4-cm thick Plexiglas plate. Further measurements were done using TLD bands all over the human hand phantom. Petrous bone examination For the XperCT system, measurements were performed with both head phantoms by positioning the Radcal ion chamber on the ear and inside the skull as close as possible to the petrous bone (Figure 1). For the CT scanner, measurements were done using the resin head phantom. As for the case of the wrist examination, dose measurements to nearby radiosensitive organs, i.e. lens and thyroid, were performed. Image quality measurements The comparison of the image quality of both systems was based on measurements with the Catphan 600 phantom (5) (The Phantom Laboratory, Salem, NY, USA). Images of the 0.28-mm tungsten carbide bead of the CTP528 high-resolution module were used to evaluate the spatial resolution and calculate the modulation transfer function (MTF) using an inhouse-developed Matlab77 (The MathWorks, Natick, MA, USA) program. In addition, two attending radiologists subjectively evaluated the image quality of patients examined with both systems. Parameters are detailed in Figures 4 and 5. RESULTS Dose measurements Doses delivered to target and nearby organs are given in Table 1 for wrist CT and XperCT scans and in Table 2 for head CT and XperCT scans. Figure 1. Reconstructed 3D image of the empty skull phantom with the XperCT system showing the position of the Radcal ion chamber close to the petrous bone. Table 1. Dose measurements (in mgy) for wrist examinations. Operational modes 300 images XperCT 600 images 3DRA Multislice CT Wrist (4 cm plate) 23.9 47.8 0.5 62.3 Lens 0.24 0.47 0.14 Thyroid 0.04 0.08 0.006 Wrist (phantom) 22.1 44.1 0.6 62.3 Table 2. Dose measurements (in mgy) for petrous bone examinations. Operational modes 300 images XperCT 600 images 3DRA Multislice CT Petrous bone 17.6 35.2 3.4 95.6 Lens 9.3 18.4 1.5 39.5 Thyroid 1.0 1.9 0.2 2.0 Image quality measurements Reconstructed images of the Catphan point source are shown in Figure 2. The images from both systems are reconstructed in the axial, coronal and sagittal planes to obtain a 3D visualisation of Figure 2. Reconstructed images from data acquired with the Siemens multislice CT (a) and the Philips XperCT (b) using standard protocols for wrist examinations. 165

J. DAMET ET AL. organs. The axial projection was used to calculate the in-plane MTF (Figure 3). Images of a patient who underwent an arthrography of a post-traumatic wrist with the XperCT system, and shortly after, with the CT scan, were evaluated (Figure 4). Axial CT and XperCT reformation images of a patient admitted after a facial trauma (Lefort III fracture) were obtained from the initial full head acquisition to rule out a committed fracture of the petrous bone (Figure 5). In both cases, the Ultra High Resolution Mode for bone was used. Complete diagnostic could be performed using either system. Figure 3. MTF of both systems for wrist and petrous bone examination protocols with different acquisition modes. DISCUSSION Dose comparison For both types of examination, the measured dose delivered to all organs using the XperCT system is Figure 4. Images of a patient admitted after a wrist trauma. A rupture of the scapholunate ligament can be identified on CT (120kV/127mAs/MPR reconstruction) (left) and XperCT (120kV/250mAs) (right) images. Figure 5. CT (120kV/300mAs) (left) and XperCT (120kV/250mAs) (right) reconstructed images obtained from the raw data of the initial full head scan. 166

multiplied by two when doubling the number of images ation (from 300 to 600 images per rotation). Using the 3DRA fluoroscopy mode of the XperCT (120 images ation, 63 kv and 59 mas) drastically reduces the doses, up to a factor of 100 for the wrist when comparing with the 600 images ation mode (Table 1). Using the human hand phantom shows a slightly lower dose to the wrist compared with the 4-cm thick Plexiglas plate. The dose to the wrist based on the CT scan is higher by a factor of 1.3 and 2.6 compared with the XperCT scan based on the 600 and 300 images ation mode, respectively. Concerning nearby radiosensitive organs in case of wrist examinations, the dose to the eye lens and the thyroid is lower when using the CT scanner. Considering the 300 images ation mode, a reduction factor of 1.7 and 6.7 is observed for the eye lens and the thyroid, respectively. Dose differences to nearby organs are explained by the distinct beam collimation of the CT scanner (fan beam) and the XperCT system (cone beam). This effect is even more pronounced for the thyroid, which is located at a larger distance from the region under examination. The measured dose to the petrous bones using the resin embedded skull phantom is higher than doses measured with the empty skull phantom. This is due to the back scattering produced by the resin and the fact that the detector is placed on the surface of the skull. The dose delivered to the petrous bone is 2.7 and 5.4 times higher for the CT examination compared with the XperCT examination realised with the 600 and 300 images ation mode, respectively. Doses to the eye lens and the thyroid, which were obtained with the resin embedded skull phantom, are higher for the CT scan. Indeed, those organs are in the direct vicinity of the beam during the petrous bone examination. The dose delivered with the CT scan examination is increased by a factor of 4.3 for the eye lens and 2.0 for the thyroid in comparison with the examination done with the XperCT system using the 300 images ation mode. The results show that for both examinations the dose to the target organ is lower when using the XperCT system compared with the CT scan. The dose is at least reduced by a factor of 1.3 for the wrist and 2.6 for the petrous bone. Image quality comparison The reconstructed images of the Catphan 600 phantom shown in Figure 2 present more artefacts with the XperCT system than with the CT scanner. This can be explained by the incomplete rotation of the C-arm structure (2408 vs. 3608 for CT scan) and by the fact that it has significantly more beam ORGAN DOSES AND IMAGE QUALITY hardening due to specific filters designed to compensate for absorption differences in objects and enhance the image quality for cerebral and lower peripheral angiography. The energy shift towards the highest values of the spectrum induces a diminution of the attenuation for high-density objects like bones. The spatial resolution depends on the resolution matrix and the field of view. Two matrices are used with the XperCT system. The 256 reconstruction mode default mode (XperCT600-256 curve in Figure 3) is appropriate for visualising soft tissue structures and low-contrast regions. A higher spatial resolution is available using the 512 reconstruction mode (XperCT600-512 curve on Figure 3) of the XperCT system that would be more suitable for objects with high contrast since the noise level is higher. The spatial resolution of the CT scanner is limited by the matrix size, which is fixed at 512 512. The size of the pixels is therefore larger for a small region of interest. The MTF of the images of the petrous bones (Figure 3) is thus below the MTF of the images of the wrist (Figure 3). The reconstructed images for examinations done with the CT scanner have a lower spatial resolution in the axial axis but a higher contrast resolution than images acquired with the XperCT system. According to the subjective evaluation of attending radiologists, both systems provide equivalent diagnostic image quality and information for wrist examinations. Concerning petrous bone examinations, although the ossicular chain and the inner ear were depicted more clearly on the CT scanner, the XperCT system depicted all essential details related to an examination of the petrous bone, such as incudo-stapedial joint, round and oval windows, cochlear turns, canal of the intermediate nerve and the tegmen tympani. CONCLUSION The measured dose to the target organs, i.e. the wrist and the petrous bones, is higher when the examination is performed with the CT scanner compared with the dose for the same examination performed with the XperCT system. For wrist examinations, the dose to the nearby critical radiosensitive organs is higher with the XperCT system. On the contrary, for petrous bone examinations, the dose to nearby organs, mainly the eye lens, is lower with the XperCT system. For wrist examinations, using the XperCT scan shortens the time between the arthrography and the XperCT scan, and thus reduces diffusion of the contrast agent in the soft tissues. The high-spatial resolution of the XperCT system enables high quality imaging of small structures, but the reconstructed images include artefacts and a high level of noise compared with the images 167

obtained with the multislice CT scanner. The quality of reconstructed images from both systems guarantees adequate accuracy for diagnostic purposes for wrist and petrous bone examinations. ACKNOWLEDGMENTS The authors are grateful to Christel Elandoy, Olivier Brina and Patrice Bregis for their contribution to measurements and fruitful discussions. REFERENCES 1. Lanz, U. and Schmitt, R. Diagnostic Imaging of the Hand, p.25, 2008 Thieme Publishing Group, Stuttgart, J. DAMET ET AL. Germany and Thieme New York, NY, USA, ISBN 978-3-13-140581-4. 2. Söderman, M., Babic, D., Holmin, S. and Andersson, T. Brain imaging with a flat detector C-arm technique and clinical interest of XperCT. Neuroradiology 50, 863 868 (2008). 3. Greess, H., Baum, U., Römer, W., Tomandl, B. and Bautz, W. CT und MRT des Felsenbeins. HNO 50, 906 919 (2002). 4. Moström, U., Ytterbergh, C. and Bergström, K. Eye lens dose in cranial computed tomography with reference to the technical development of CT scanners. Acta. Radiol. Diagn. 27, 599 606 (1986). 5. Catphanw 500 and 600 Manual. (2009), The Phantom Laboratory Incorporated, Salem, NY 12865-0511, USA. 168