The Influence of Exogenous Fat and Water on Lumbar Spine Bone Mineral Density in Healthy Volunteers

Similar documents
9 Quality Assurance in Bone Densitometry section

Comparison of Bone Density of Distal Radius With Hip and Spine Using DXA

Body composition analysis by dual energy X-ray absorptiometry in female diabetics differ between manufacturers

ASJ. How Many High Risk Korean Patients with Osteopenia Could Overlook Treatment Eligibility? Asian Spine Journal. Introduction

Dual-energy Vertebral Assessment

Osteoporosis International. Original Article. Bone Mineral Density and Vertebral Fractures in Men

Comparison of Dual-Energy X-Ray Absorptiometry and Dual Photon Absorptiometry for Bone Mineral Measurements of the Lumbar Spine

Effect of Precision Error on T-scores and the Diagnostic Classification of Bone Status

Measurement Uncertainty in Spine Bone Mineral Density by Dual Energy X-ray Absorptiometry

Medical Policy. MP Whole Body Dual X-Ray Absorptiometry to Determine Body Composition

Diagnostische Präzision von DXL im Vergleich zu DXA bei pmp Frauen mit Frakturen

Norland Densitometry A Tradition of Excellence

Osteoporotic Fracture Risk Assessment Using Bone Mineral Density in Korean: A Community-based Cohort Study

Cross-calibration of fat and lean measurements by dualenergy X-ray absorptiometry to pig carcass analysis in the pediatric body weight range13

Whole Body Dual X-Ray Absorptiometry to Determine Body Composition

Clinical Study Comparison of QCT and DXA: Osteoporosis Detection Rates in Postmenopausal Women

Quality Control of DXA System and Precision Test of Radio-technologists

Location of body fat and body size impacts DXA soft tissue measures: a simulation study

DXA When to order? How to interpret? Dr Nikhil Tandon Department of Endocrinology and Metabolism All India Institute of Medical Sciences New Delhi

Whole Body Dual X-Ray Absorptiometry (DXA) to Determine Body Composition

Screening points for a peripheral densitometer of the calcaneum for the diagnosis of osteoporosis

Obesity Increases Precision Errors in Dual X-ray Absorptiometry Measurements.

Suprailiac or Abdominal Skinfold Thickness Measured with a Skinfold Caliper as a Predictor of Body Density in Japanese Adults

In-vivo precision of the GE Lunar idxa for the measurement of visceral adipose tissue in

Bone Mineral and Body Composition Measurements: Cross-Calibration of Pencil-Beam and Fan-Beam Dual- Energy X-Ray Absorptiometers*

Documentation, Codebook, and Frequencies

PhenX Measure: Body Composition (#020300) PhenX Protocol: Body Composition - Body Composition by Dual-Energy X-Ray Absorptiometry (#020302)

Prevalence of Sarcopenia Adjusted Body Mass Index in the Korean Woman Based on the Korean National Health and Nutritional Examination Surveys

Purpose. Methods and Materials

Prodigy. from GE Healthcare. Most trusted, reliable and best-selling DXA system with one of the largest installed base in the world. gehealthcare.

Clinical Densitometry

Does standardized BMD still remove differences between Hologic and GE-Lunar state-of-the-art DXA systems?

High Dietary Sodium Intake Assessed by 24-hour Urine Specimen Increase Urinary Calcium Excretion and Bone Resorption Marker

Bone Densitometry Radiation dose: what you need to know

Validation Study of Multi-Frequency Bioelectrical Impedance with Dual-Energy X-ray Absorptiometry Among Obese Patients

Differences in whole body measurements by DXA-scanning using two Lunar DPX-L machines

Bone Mineral Density and Fatty Degeneration of Thigh Muscles Measured by Computed Tomography in Hip Fracture Patients

The DXL Calscan heel densitometer: evaluation and diagnostic thresholds

DUAL ENERGY X-RAY ABSORPTIOMETRY (DXA) GOLD STANDARD FOR BONE HEALTH AND BODY COMPOSITION ASSESSMENT

Prevalence of Osteoporosis p. 262 Consequences of Osteoporosis p. 263 Risk Factors for Osteoporosis p. 264 Attainment of Peak Bone Density p.

Interpreting DEXA Scan and. the New Fracture Risk. Assessment. Algorithm

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

Bone Densitometry. Total 30 Maximum CE 14. DXA Scanning (10) 7

Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition. Original Policy Date 12:2013

Accuracy of DEXA scanning & other methods for determining BMD.

Mineral Density Of Subchondral Bone May Be Quantitatively Evaluated Using A Clinical Cone Beam Computed Tomography Scanner

Are glucocorticoid-induced osteoporosis recommendations sufficient to determine antiosteoporotic treatment for patients with rheumatoid arthritis?

Bone Mineral Density and Its Associated Factors in Naresuan University Staff

Skeletal Manifestations

The effect of vertebral rotation of the lumbar spine on dual energy X-ray absorptiometry measurements: observational study

DEXA T-score Concordance and Discordance Between hip and Lumbar Spine

A FRAX Experience in Korea: Fracture Risk Probabilities with a Country-specific Versus a Surrogate Model

Relationship between Decrease in Serum Sodium Level and Bone Mineral Density in Osteoporotic Fracture Patients

Adina Alazraki, MD, FAAP Assistant Professor, Radiology and Pediatrics Emory University School of Medicine Children s Healthcare of Atlanta

Whole Body Dual X-Ray Absorptiometry (DEXA) to Determine Body Composition

Lateral Vertebral Analysis DXA Body Composition Quality Assurance in DXA

MP Whole Body Dual X-Ray Absorptiometry to Determine Body Composition

Concordance of a Self Assessment Tool and Measurement of Bone Mineral Density in Identifying the Risk of Osteoporosis in Elderly Taiwanese Women

Annotations Part III Vertebral Fracture Initiative. International Osteoporosis Foundation March 2011

Segmental Body Composition Assessment for Obese Japanese Adults by Single-Frequency Bioelectrical Impedance Analysis with 8-point Contact Electrodes

Validation of the Osteoporosis Self-Assessment Tool in US Male Veterans

The Bone Wellness Centre - Specialists in Dexa Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017

Prevalence of Osteoporosis in the Korean Population Based on Korea National Health and Nutrition Examination Survey (KNHANES),

Body Mass Index as Predictor of Bone Mineral Density in Postmenopausal Women in India

The Bone Densitometry Examination

LUMBAR IS IT IMPORTANT? S. Tantawy,, M.D.

The Bone Wellness Centre - Specialists in Dexa Total Body 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

Evaluation of Bone Mineral Status in Adolescent Idiopathic Scoliosis

Is Sarcopenia a Potential Risk Factor for Distal Radius Fracture? Analysis Using Propensity Score Matching

Efficacy of risedronate in men with primary and secondary osteoporosis: results of a 1-year study

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

DEVELOPMENT OF A RISK SCORING SYSTEM TO PREDICT A RISK OF OSTEOPOROTIC VERTEBRAL FRACTURES IN POSTMENOPAUSAL WOMEN

Bone mineral density in the normal Iranian population: a comparison with American reference data

Bone Mineral Densitometry with Dual Energy X-Ray Absorptiometry

Preoperative dual-energy X-ray absorptiometry and FRAX in patients with lumbar spinal stenosis

Quality Assurance and Control in Dual energy X ray Absorptiometry

Diagnosis of Vertebral Fractures by Vertebral Fracture Assessment

The Effect of Disc Degeneration in Osteoporotic Vertebral Fracture

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

An audit of bone densitometry practice with reference to ISCD, IOF and NOF guidelines

Clinical Application of Computed Radiography in Orthopedic Surgery

The Bone Wellness Centre - Specialists in DEXA Scanning 855 Broadview Avenue Suite # 305 Toronto, Ontario M4K 3Z1

Estimation of Stellate Ganglion Block Injection Point Using the Cricoid Cartilage as Landmark Through X-ray Review

Analysis of Clinical Features of Hip Fracture Patients with or without Prior Osteoporotic Spinal Compression Fractures

DXA IS THE MOST commonly used method of assessing

Lunar idxa. The intelligent DXA. gehealthcare.com

Development of a Viable Bedside Ultrasound Protocol to Accurately Predict Appendicular Lean Tissue Mass

Association between Seasonal Changes in Vitamin D and Bone Mineral Density

Available online at ScienceDirect. Osteoporosis and Sarcopenia 1 (2015) 109e114. Original article

Obesity Increases Precision Errors in Total Body Dual X-ray Absorptiometry Measurements.

O. Bruyère M. Fossi B. Zegels L. Leonori M. Hiligsmann A. Neuprez J.-Y. Reginster

2013 ISCD Official Positions Adult

CLINICAL DENSITOMETRY

TigerPrints. Clemson University. Nadia Marie Ghassan Najm Clemson University

Healthy aging. It s vital.

Use of DXA / Bone Density in the Care of Your Patients. Brenda Lee Holbert, M.D. Associate Professor Senior Staff Radiologist

Statistical data preparation: management of missing values and outliers

Densitometry Techniques

The effects of minodronate and activated vitamin D on bone mineral density and muscle mass in postmenopausal women with osteoporosis

Transcription:

Original Article http://dx.doi.org/10.3349/ymj.2012.53.2.289 pissn: 0513-5796, eissn: 1976-2437 Yonsei Med J 53(2):289-293, 2012 The Influence of Exogenous Fat and Water on Lumbar Spine Bone Mineral Density in Healthy Volunteers Kyu-Nam Kim, Bom-Taeck Kim, Kwang-Min Kim, Sat-Byul Park, Nam-Seok Joo, Sang Hyeon Je, Young-Sang Kim, and Duck-Joo Lee Department of Family Practice and Community Health, Ajou University School of Medicine, Suwon, Korea. Received: April 6, 2011 Revised: May 17, 2011 Accepted: May 26, 2011 Corresponding author: Dr. Duck-Joo Lee, Department of Family Practice and Community Health, Ajou University School of Medicine, 164 Worldcup-ro, Yeongtong-gu, Suwon 443-721, Korea. Tel: 82-31-219-5324, Fax: 82-31-217-2418 E-mail: djleemd@msn.com The authors have no financial conflicts of interest. Purpose: Changes in human body composition can affect the accuracy of spine bone mineral density (BMD) measurements. The purpose of this study was to evaluate whether fat and water in the soft tissue of the abdomen influence lumbar spine BMD measurements obtained using dual energy X-ray absorptiometry (DXA). Materials and Methods: Duplicate BMD measurements were carried out on healthy volunteers (10 men and 10 women) and the Hologic anthropomorphic spine phantom had on the same day before and after placement of following 3 materials in the abdominal area: lard 900 g, 1.5 cm thick; oil 1.4 liters in a vinyl bag; and water 1.2 liters in a vinyl bag. Results: In the case of human participants, following the placement of exogenous water to mimic extracellular fluid (ECF), there was a significant decrease in lumbar spine BMD (-0.012 g/cm 2, p=0.006), whereas the placement of exogenous lard and oil to mimic abdominal fat produced a slight increase in lumbar spine BMD (0.006 g/cm 2, p=0.301; 0.008 g/cm 2, p=0.250, respectively). The average percentage of lumbar spine BMD change with and without exogenous lard, oil, and water showed increase of 0.51%, and 0.67%, and decrease of 1.02%, respectively. Using the phantom, BMD decreased with the placement of both lard (-0.002 g/cm 2, p=0.699) and water (-0.006 g/cm 2, p=0.153); however, there was no difference in BMD after oil placement. Conclusion: These results suggest that in cases where changes in fat and ECF volume are similar, ECF exerts a greater influence than fat on DXA lumbar BMD measurements. Key Words: Bone mineral density, lard, extracellular fluid, dual energy X-ray absorptiometry (DXA) INTRODUCTION Copyright: Yonsei University College of Medicine 2012 This is an Open Access article distributed under the terms of the Creative Commons Attribution Non- Commercial License (http://creativecommons.org/ licenses/by-nc/3.0) which permits unrestricted noncommercial use, distribution, and reproduction in any medium, provided the original work is properly cited. Lumbar spine bone mineral density (BMD) measured by dual energy X-ray absorptiometry (DXA) is known to be the best predictor of fracture risk: 1-3 DXA beam is attenuated by bone and its surrounding soft tissue. However, it is not known how much of the attenuation is influenced by bone and soft tissue. DXA is based on the assumption of homogeneity of soft tissue composition in bone and its adjacent soft tissue. Thus, some inaccuracies can result from DXA measurements, because of Yonsei Med J http://www.eymj.org Volume 53 Number 2 March 2012 289

Kyu-Nam Kim, et al. it s assumptions on the homogeneity of soft tissue. The influence of ascites on lumbar spine bone density measurements was reported by some researchers, 4,5 and the fat in the adipose capsules of the kidney has also been found to lead to overestimation of the bone mineral content (BMC) in spinal BMD DXA measurements. 6 The shape of vertebrae as well as thickness and amount of its covering soft tissue varies not only between subjects but also within the same person. Therefore, identifying more influencing factors for DXA measurements are important for lumbar spine BMD interpretation in order to decrease DXA measurement error. To our best knowledge, no study has been carried out to examine the change in DXA measured lumbar spine BMC and BMD due to changes in the different soft tissue composition of the abdominal area. The purpose of this study was to evaluate how similar changes in fat and extracellular fluid (ECF) volume in the abdominal area influence lumbar spine BMC and BMD, measured with DXA, and differences due to the type of fat (solid versus liquid) were also evaluated. MATERIALS AND METHODS Study subjects and design Following informed consent, healthy volunteers (10 men and 10 women) were enrolled in the study. Control BMD of L1 to L4 was measured for each participant with a LU- NAR idxa scan (GE Healthcare, Waukesha, WI, USA). The scanning precision (coefficient of variance, CV%) was calculated from two repeated measurements with reposi- tioning, which was performed on 20 healthy adults. The precision (CV%) of these measurements was 0.9% for lumbar spine BMD. The following 3 materials were then placed consecutively on the abdominal area, and measurements were repeated: lard 900 g to mimic solid fat (20 15 1.5 cm), oil 1.2 liters (25 15 cm) to mimic liquid fat (purely 100% fat), and water 1.4 liters (25 15 cm) to mimic ECF volume. The same procedure was applied using the Hologic anthropomorphic spine phantom with BMD measured twice under each condition. The phantom which was made of calcium hydroxyapatite enclosed in an epoxy resin was composed of four vertebrae to simulate soft tissue. Four vertebrae have similar densities and areas. All measurements were performed between 3 and 6 pm. The Ethical Committee of Ajou University Hospital approved this study. Statistics Data were analyzed using the statistical software package SPSS for Windows 13.0 version. Paired t-tests were used to examine the significance between changes in BMD measurements before and after placing lard, oil, and water on the abdominal area. The independent t-test was used to check for a significant difference in BMD changes in lumbar spine with and without exogenous lard and oil. Results were considered significant at a 2-tailed level of 0.05. RESULTS The average participants age was 30.3±3.5 (mean±sd) years and the average body mass index (BMI) was 21.4±1.2 Table 1. BMC, Bone Area and BMD (Mean±SD) of Lumbar Spine and Phantom with and without Exogenous Lard, Oil, and Water (n=20) BMC (g) Bone area (cm 2 ) BMD (cm 2 /g) L1-4 human spine Control 70.469±10.95 59.969±7.33 1.170±0.21 Values after placing 3 materials Lard 70.634±10.71 59.901±7.49 1.176±0.10 Oil 70.497±9.90 59.750±7.08 1.178±0.11 Water 69.794±10.61* 59.996±7.28 1.158±0.10* L1-4 phantom spine Control 57.736±08.95 51.689±6.31 1.117±0.00 Values after placing 3 materials Lard 57.201±06.72 51.256±6.40 1.116±0.00 Oil 57.378±08.90 51.368±7.18 1.117±0.00 Water 57.473±09.90 51.685±6.98 1.112±0.00 BMC, bone mineral content; BMD, bone mineral density; SD, standard deviation. *p<0.05. 290 Yonsei Med J http://www.eymj.org Volume 53 Number 2 March 2012

Body Composition and Bone Mineral Density Table 2. BMD Difference of Lumbar Spine and Phantom with and without Exogenous Lard, Oil, and Water (Paired t-test, n=20) Materials 95% confidence interval of the difference Upper Lower p value L1-4 BMD of human (g/cm 2 ) Lard 0.006 0.003-0.012 0.301 Oil 0.008 0.007-0.020 0.250 Water -0.012* 0.021 0.006 0.006 L1-4 BMD of phantom (g/cm 2 ) Lard -0.002 0.015-0.012 0.699 Oil 0 0.008-0.007 0.870 Water -0.006 0.016-0.005 0.153 BMD, bone mineral density. *p value<0.05. kg/m 2. Tables 1 and 2 show BMC, bone area and BMD differences in lumbar spine and phantom with and without exogenous lard, oil, and water. Bone density of L1 to L4 in human subjects increased with the placement of lard and oil (0.006 g/cm 2, p=0.301; 0.008 g/cm 2, p=0.250, respectively); however, this was not statistically significant. In the case of water placement, BMC and BMD showed a statistically significant decrease (-0.675 g/cm 2, p=0.035; -0.012 g/cm 2, p=0.006, respectively). The average percentage of lumbar spine BMD change with and without exogenous lard, oil, and water showed increase of 0.51% and 0.67% and decrease of 1.02%, respectively. With the phantom, bone density decreased after placing both lard (-0.002 g/cm 2, p=0.699) and water (-0.006 g/cm 2, p=0.153) but there was no difference in bone density after placing oil (0 g/cm 2, p=0.870). In men, BMD decreased statistically significantly only when water was placed (-0.019 g/cm 2, p=0.040, data not shown). In women, when lard or oil was placed, BMD increased statistically significantly (0.016 g/cm 2, p=0.039; 0.03 g/cm 2, p=0.040 respectively, data not shown). Differences in lumbar spine BMD between lard and oil, representing solid fat and liquid fat respectively, were not statistically significant (p=0.607, data not shown). DISCUSSION The present results indicate that fat, oil, and water influence lumbar BMD values measured by DXA. Lumbar BMD showed a statistically significant decrease upon water placement on the abdominal area and tended to increase slightly with the placement of lard and oil. The average percentage of lumbar spine BMD change with and without exogenous lard, and oil, increased 0.51% and 0.67%, respectively, and decreased 1.02% with water. The difference in lumbar spine BMD change between lard and oil, representing solid fat and liquid fat, respectively, was not statistically significant. These materials showed similar influence on BMD, when measured with a DXA scanner made from Lunar Corp. (Madison, WI, USA) using a phantom particularly in the case of water placement. Consequently, these results suggest that if changes in fat and ECF volume of the human body are similar, ECF exerts a greater influence on lumbar BMD measurements by DXA than fat. A study showed that total-body bone mineral change, simulated by homogenously placing 8.8 kgs of porcine lard on the bodies of women, increased approximately 1.3%, 7 and an other study showed also that there were some changes in soft tissue heterogeneity with weight loss (11.3±6.9 kg), which falsely decreased the anterior-posterior (AP) spinal BMD 1-2% by DXA. 8 On the other hand, there was a significant increase in the spine BMD of 4.2% 4 after paracentesis (6.4±2.0 L) and there was a significant increase in the spine bone mineral content of 1.5% after drainage of peritoneal dialysate (1,908±508 ml), 5 in support of the present results. To our best knowledge, this is the first report on the impact of human body fluid changes, because of changes in abdominal soft tissue heterogeneity on the accuracy of BMD measurements, measured by DXA. A subtraction or addition of 1 kg of ECF to Reference Man would result in roughly 0.6% over- or under-estimation of total body fat mass in a theoretical model. 9 However, there was no significant difference in the data obtained by DXA before and after dialysis. 10,11 Of note is the fact that these studies measured body composition with DXA, but not BMD. Lumbar spine BMD could be measured in patients with fluid retention or removal. Significant ascites can cause falsely lower Yonsei Med J http://www.eymj.org Volume 53 Number 2 March 2012 291

Kyu-Nam Kim, et al. BMD measurements in the spine. 4 There was a significant difference in the BMC of the spine before and after the drainage of dialyzate. 5 ECF distorts tissue boundaries between bone and soft tissue, resulting in an increased measurement of soft tissue composition and hence a smaller difference between the soft tissue and bone compartments, and consequently underestimating spinal BMD. 4 In regards to gender, lumbar BMD was significantly decreased by placing water on the abdominal area in men, but not women. However, these findings should further be confirmed in future by multiple DXA scans of the same subjects as well as a larger numbers of subjects. In the case of fat and oil used to mimic abdominal fat, the present results show that spinal BMD measurements, following the placement of lard and oil on the abdominal area, were higher, though not significant, than without lard and oil. These results are supported by previous reports that heterogeneity in body fat distribution may theoretically cause inconsistency in the measurements of BMD by DXA due to extrapolation of thickness and fat % of soft tissue in nonbone pixels to bone pixels. 12 The present results are also in accordance with a previous cross-sectional study of postmenopausal women with a wide range of BMI, 13 which indicated that a difference in fat % between bone and nonebone pixels may lead to overestimation of AP spinal BMD. Higher BMD was obtained when an aluminum phantom surrounded by a 30 : 70 oil/water mixture was used to depict increased truncal thickness, 14 and a significant increase in BMD measurement error was observed when ashed bone surrounded by an oil/water mixture representative of 50% body fat was scanned. 15 The possible mechanism behind these results has been explained by the study, which showed, the greater the thickness of soft tissue, the greater the resulting X-ray attenuation and interference with bone edge detection, when layers of lard at a thickness of less than 9 cm were placed on a spine phantom. 16 However, a previous study demonstrated different BMD s, depending on the brands of DXA scanners used to study the changes of bone mineral measurements during weight change. 17 For example, with the Lunar instruments, the total-body BMD was reduced with weight loss, whereas it appeared to increase with the Hologic scanners, indicating that differences in the brands of DXA scanners used may influence BMD value, though there is no real bone change. As for the gender, lumbar BMD was significantly increased by placing lard and oil on the abdominal area of women. However, we found that BMD assessed by DXA using a phantom decreased with the place- ment of lard. It is quite possible that the thickness of lard used is responsible for these conflicting results. The limitation of this study is its relatively small sample size compared with other studies, and this might be the reason of why lumbar BMD failed to show a significant decrease after placing water on the abdominal area of women. Nevertheless, the present study is the first pilot study to examine the impact of human body composition change on the accuracy of BMD measurements by DXA: body composition change may result from simultaneous heterogenous changes in abdominal soft tissue, i.e. changes in fat % and volume of soft tissue and ECF of the human body as they correspond to bone and non-bone pixels detected by DXA. In conclusion, there are differences in spinal BMD measured by DXA before and after placement of oil and lard, especially water, on the abdominal area to mimic abdominal fat and ECF, respectively. If changes in fat and ECF volume are similar, ECF influences the changes in lumbar BMD measurements more than fat; therefore, lumbar BMD measurements need to be interpreted carefully when there is a change, especially in fluid composition, in the soft tissue of the abdominal area. REFERENCES 1. Marshall D, Johnell O, Wedel H. Meta-analysis of how well measures of bone mineral density predict occurrence of osteoporotic fractures. BMJ 1996;312:1254-9. 2. Cauley JA, Wu L, Wampler NS, Barnhart JM, Allison M, Chen Z, et al. Clinical risk factors for fractures in multi-ethnic women: the Women s Health Initiative. J Bone Miner Res 2007;22:1816-26. 3. Siris ES, Chen YT, Abbott TA, Barrett-Connor E, Miller PD, Wehren LE, et al. Bone mineral density thresholds for pharmacological intervention to prevent fractures. Arch Intern Med 2004; 164:1108-12. 4. Labio ED, Del Rosario DB, Strasser SI, McCaughan GW, Crawford BA. Effect of ascites on bone density measurement in cirrhosis. J Clin Densitom 2007;10:391-4. 5. Mann ML, Thornley-Brown D, Campbell R, Bell E, Burroughs L, Nunnally N, et al. The effect of peritoneal dialysate on DXA bone densitometry results in patients with end-stage renal disease. J Clin Densitom 2008;11:532-6. 6. Roos BO, Hansson TH, Sköldborn H. Dual photon absorptiometry in lumbar vertebrae. Evaluation of the baseline error. Acta Radiol Oncol 1980;19:111-4. 7. Svendsen OL, Haarbo J, Hassager C, Christiansen C. Accuracy of measurements of body composition by dual-energy x-ray absorptiometry in vivo. Am J Clin Nutr 1993;57:605-8. 8. Svendsen OL, Hendel HW, Gotfredsen A, Pedersen BH, Andersen T. Are soft tissue composition of bone and non-bone pixels in spinal bone mineral measurements by DXA similar? Impact of 292 Yonsei Med J http://www.eymj.org Volume 53 Number 2 March 2012

Body Composition and Bone Mineral Density weight loss. Clin Physiol Funct Imaging 2002;22:72-7. 9. Pietrobelli A, Formica C, Wang Z, Heymsfield SB. Dual-energy X-ray absorptiometry body composition model: review of physical concepts. Am J Physiol 1996;271:E941-51. 10. Formica C, Atkinson MG, Nyulasi I, McKay J, Heale W, Seeman E. Body composition following hemodialysis: studies using dualenergy X-ray absorptiometry and bioelectrical impedance analysis. Osteoporos Int 1993;3:192-7. 11. Stenver DI, Gotfredsen A, Hilsted J, Nielsen B. Body composition in hemodialysis patients measured by dual-energy X-ray absorptiometry. Am J Nephrol 1995;15:105-10. 12. Svendsen OL, Hassager C, Skødt V, Christiansen C. Impact of soft tissue on in vivo accuracy of bone mineral measurements in the spine, hip, and forearm: a human cadaver study. J Bone Miner Res 1995;10:868-73. 13. Svendsen OL, Hassager C, Bergmann I, Christiansen C. Measurement of abdominal and intra-abdominal fat in postmenopausal women by dual energy X-ray absorptiometry and anthropometry: comparison with computerized tomography. Int J Obes Relat Metab Disord 1993;17:45-51. 14. Johnson J, Dawson-Hughes B. Precision and stability of dual-energy X-ray absorptiometry measurements. Calcif Tissue Int 1991;49:174-8. 15. Wahner HW, Dunn WL, Brown ML, Morin RL, Riggs BL. Comparison of dual-energy x-ray absorptiometry and dual photon absorptiometry for bone mineral measurements of the lumbar spine. Mayo Clin Proc 1988;63:1075-84. 16. Compston JE, Laskey MA, Croucher PI, Coxon A, Kreitzman S. Effect of diet-induced weight loss on total body bone mass. Clin Sci (Lond) 1992;82:429-32. 17. Tothill P. Dual-energy x-ray absorptiometry measurements of total-body bone mineral during weight change. J Clin Densitom 2005;8:31-8. Yonsei Med J http://www.eymj.org Volume 53 Number 2 March 2012 293