PLR. Proximal Loading Revision Hip System

Similar documents
Integral 180 Surgical Technique

RECOVERY. P r o t r u s i o

Arcos Modular Femoral Revision System

Bone Preservation Stem

Arcos Interlocking Distal Stem. Surgical Technique Addendum to the Arcos Modular Femoral Revision System

9800 Metric Blvd. Austin, Texas

Answer. Hip System Design Features:

Arcos Modular Femoral Revision System. Broach and Calcar Proximal Bodies Surgical Technique

AML Hip System. Design Rationale/ Surgical Technique

ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY

Approach Patients with Confidence

Preoperative Planning. The primary objectives of preoperative planning are to:

CAUTION: Ceramic liners are not approved for use in the United States.

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

Optimum implant geometry

FLH /11

Design Rationale. ECHELON Primary Hip System

SURGICAL TECHNIQUE. Alpine Cementless Hip Stem

Clinical Evaluation Surgical Technique

Revision. Hip Stem. Surgical Protocol

Following a tradition of success. VerSys Heritage Primary Hip Prosthesis Surgical Technique

Approach Patients with Confidence

ZMR Crossover Instruments. Abbreviated Surgical Technique

Surgical Technique. Hip System

Encina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Zimmer Segmental System

ACETABULAR CUP SURGICAL TECHNIQUE

Hip Product Brochure

Biomet Large Cannulated Screw System

operative technique Kent Hip

The Leader in Orthopaedic Innovation

Common Instruments for Biomet Hip Systems

Restoration Modular. Revision Hip System Surgical Protocol

Optimum implant geometry

EXTENDED TROCHANTERIC OSTEOTOMY SURGICAL TECHNIQUE FPO EXTENSIVELY COATED FIXATION

VerSys LD/Fx Cemented and Press-Fit Hip Prostheses. Surgical Technique IMAGE TO COME. Versatile solutions for total and partial hip replacement

RESTORATION MODULAR. Surgical Protocol REVISION HIP SYSTEM

HIP SYSTEM SURGICAL TECHNIQUE

AUTOBLOQUANTE AUTOBLOQUANTE. Product Rationale Surgical Technique

Taperloc Complete Hip System. Surgical Technique

Zimmer Segmental System. Surgical Technique for Proximal Femoral Replacement and Intercalary Segments IMAGE TO COME

TaperFill. Surgical Technique

TOTAL HIP REPLACEMENT:

Optimum implant geometry

ZMR Over-the-Junction Instruments for Revision Hip Arthroplasty. Surgical Technique IMAGE TO COME

comprehensive Design Rationale shoulder system Knees Hips Extremities Cement and Accessories PMI TEchnology

ZMR CROSSOVER INSTRUMENTS AND SURGICAL TECHNIQUE. Surgical Technique for Revision Hip Arthroplasty

Specifications guide ECHELON. Hip System

VerSys Fiber Metal Taper Hip Prosthesis. Surgical Technique

RECLAIM REVISION SOLUTIONS

Anterior Approach Surgical Technique. Paragon Stem System. enabling people to enjoy life

Manza Cup HA SURGICAL TECHNIQUE.

Absolut TM Cemented Stem. Surgical Technique

Resurfacing Distal Femur. Orthopaedic Salvage System

Trabecular Metal Primary Hip Prosthesis

Rx90 Total Hip System Acetabular Series

S U R G I C A L T E C H N I Q U E

The bi-metric XR series of Hip implants blends biomet s proven philosophy and tradition with Advanced technology. porous primary hip series

RESTORATION MODULAR. Surgical Protocol REVISION HIP SYSTEM

ZMR Porous Revision Hip Prosthesis. Surgical Technique IMAGE TO COME

Cementless Tapered Femoral Stem Surgical technique

NEXGEN COMPLETE KNEE SOLUTION S A. Tibial Stem Extension & Augmentation Surgical. ATechnique

Featuring. Technology. Product Rationale

Tradition Hip Primary Surgical Technique

JOINT RULER. Surgical Technique For Knee Joint JRReplacement

One system for all revision types

Cement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter

RESTORATION MODULAR. Surgical Technique REVISION HIP SYSTEM

RESTORATION MODULAR. Surgical Protocol REVISION HIP SYSTEM

Versys Advocate V-Lign and Non V-Lign Cemented Hip Prosthesis

THE NATURAL FIT. Surgical Technique. Hip Knee Spine Navigation

VerSys 6 Beaded FullCoat Plus Hip Prosthesis

Inspiring people, Enriching lives

CORAIL HIP SYSTEM SURGICAL TECHNIQUE

Regenerex Porous Titanium Construct. Knees Hips Extremities Cement and Accessories PMI Trauma Technology

AVANTEON. Operative Technique & Catalogue Information AVANTEON

Zimmer NexGen Tibial Stem Extension & Augmentation. Surgical Technique IMAGE TO COME. Stem Extensions and Augments

LPS SYSTEM POCKET GUIDE

Cementless Tapered Femoral Stem Surgical technique

RECLAIM REVISION HIP SYSTEM

SURGICAL TECHNIQUE. Alpine Cemented Hip Stem

PROFEMUR Total Hip System P R O D U C T C A T A L O G

Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology. Surgical Technique

Trinity. Advanced Bearing Acetabular System Surgical technique

Progeny Hip Stem. Surgical Protocol and Product Specifications

EXACTECH HIP. Operative Technique. Tapered and Splined Press-Fit Stems. Cemented Femoral Stems. Renewing Innovations. Enduring Solutions.

Distal Cut First Femoral Preparation

Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA

Surgical Technique. Cup System

Restoration Modular Revision Hip System. Choice Matters

Surgical Technique r5.indd 1 12/8/10 10:36 AM

Zimmer M/L Taper Hip Prosthesis. Surgical Technique

SURGICAL TECHNIQUE. Protrusio Cage A COMPREHENSIVE ACETABULAR REVISION SYSTEM

Optimizing function Maximizing survivorship Accelerating recovery

DOWNLOAD OR READ : TECHNIQUES IN REVISION HIP AND KNEE ARTHROPLASTY PDF EBOOK EPUB MOBI

Bone Bangalore

Resurfacing Patellar Preparation

TaperFit. Cemented Total Hip Replacement Surgical technique

Bone Conserving Hip Replacement Surgical technique

28 Surgical Technique

Transcription:

PLR Proximal Loading Revision Hip System The PLR splined revision stem is designed to recreate the natural stresses in the revised femur, where proximal bone may be compromised.

PLR Hip System Design Considerations The Proximal Loading Revision (PLR ) splined stem is designed to bridge proximal defects, while achieving a precise distal fit in the revision femoral bone. Materials The choice of titanium in a press fit application allows for greater biocompatibility, and a modulus of elasticity closer to that of bone compared to other implant materials. The surface of the implant is roughened by a 30 grit-blast along the entire stem length to create an opportunity for potential long term stability through bone on-growth.1,2 Advanced Stem Geometry The PLR stem is gradually tapered 1.5 proximal to distal. The metal core exhibits a slightly different taper than the outer (spline) dimension to gradually reduce the stiffness of the implant from proximal to distal. The splines are deeper at the distal end (2 2.5mm) and slightly shallower proximally (1.5 2mm). The advanced spline geometry of the PLR stem allows for an increased bite into the distal bone, while using less metal than a solid stem. This provides approximately twice the rotational stability of fully porous cylindrical stems, while making the stem considerably less stiff.3 This unique spline geometry also provides resistance to subsidence. As an added benefit, the surface area of the stem is increased due to its splined geometry providing more contact with the host bone, thus maximizing the opportunity for bone on-growth. The distal tip of the long stems 270mm is tapered to reduce the possibility of spline cutout through the anterior cortex of the femur (Figure A). biomechanics through the use of 28mm and 32mm modular heads in 7 neck lengths ranging from 28 46mm. A wide variety of Biomet acetabular shells and RingLoc cup liners are used to achieve correct leg length and physiological abductor muscle tension. Lateral Soft Tissue Reattachment A three hole flange placed on the proximal lateral portion of the PLR revision stem allows the surgeon to create additional stability by suturing soft tissue to the stem. Precise, Simple Instrumentation and Sizing The PLR splined revision stem is available in 270mm length measured from the medial resection level to the stem tip. Each stem length is available in 1mm increments from 14mm to 24mm. Instrumentation for the PLR stem provides simple, reproducible, and accurate placement of the stem. The flow from templating to accurate implant insertion is facilitated by wellmarked, simple instrumentation. Size specific reamers and trials are used to implant the PLR revision stem. Each reamer and trial is sized 1mm smaller than the outer diameter of the corresponding prosthesis, allowing each spline to cut 1/2mm into virgin bone. Each template, reamer and corresponding trial are etched with depth markings to assure precise positioning. Implants are marked and correspond to the trial to allow for precise recreation of femoral anteversion. Figure A Recreation of Anatomic Offset The neck shaft angle of the PLR splined revision stem is 135, allowing the surgeon to recreate the patient s normal anatomic offset, which reduces the chance for dislocation and subluxation. The surgeon has added flexibility in recreating normal hip Preoperative X-ray Postoperative X-ray

The goal of the PLR splined revision stem is to create stable long-term fixation in hip reconstructive situations where proximal bone is missing or cavitated due to the presence of osteolytic lesions or stress shielding. Surgical Technique X-ray Templating X-ray templating is crucial to operative sizing, reaming, trialing, and final implant position. Templates (Figure 1) are placed such that the center of the head is recreated with the template overlay. The correct implant size is chosen by the template that fills the metaphysis and diaphysis best on both A/P and lateral views. Once this is accomplished the appropriate template is placed over the x-ray, an easily accessible bony landmark is identified (possibly the lesser trochanter or a bony proximal landmark that can be found intraoperatively). The depth line on the template corresponding to the bony landmark chosen is identified and noted by the surgical team or on the x-ray (Figure 2). Note: This is important because reamer and trial markings corresponding to this line will be used in surgery. Figure 1 Step One: Removal of Implant Removal of implant, cement and soft tissue is performed, while maintaining intact cortical walls and preserving as much host bone as possible. Care should be taken to remove all cement, soft tissue and debris. Step Two: Reaming Reaming is initiated with the smaller sized PLR tapered reamers to create a channel for subsequent reamers. Reaming is sequentially accomplished in 1mm increments to the templated diameter or until cortical chatter of the reamer is felt and heard. Each reamer should be positioned with the templated depth mark located at the bony landmark. This represents the appropriate diameter trial and prostheses to use (Figure 3). Figure 2 The PLR System was designed in conjuction with Chris Peters, M.D. Biomet, as the manufacturer of this device, does not practice medicine and does not recommend this or any other surgical technique for use on a specific patient. The surgeon who performs any implant procedure is responsible for determining and using the appropriate techniques for implanting the prosthesis in each individual patient. Biomet is not responsible for selection of the appropriate surgical technique to be used for an individual patient. Figure 3

Step Three: Trial The correct size trial is placed into the canal and lightly tapped into position noting that the trial depth marking is placed at the bony landmark as was identified in templating, and in the reaming process (Figure 4). Care should also be taken to place the trial stem in the correct amount of anteversion as required to make the hip stable and allow for proper range of motion (note the anteversion marking on the trial vertical line anteriorly and posteriorly) (Figure 5). A trial head may be placed on the trial and a trial reduction completed at this time. If a change in version is required, the trial may be changed and reset at this point. Once trial reduction is completed, the trial head is removed and a mark (surgical marker or Bovie) is made on the bone at the location of the anteversion control line running vertically along the length of the trial (Figure 6). This line will reference the anteversion of the prosthesis. Figure 4 Step Four: Prosthesis Implantation Make a mark on the implant correlating to the depth marking from the trial as to the depth to drive the implant. The implant is placed into the canal such that the anteversion control line on bone is placed in the groove between the two splines on the anterior portion of the prosthesis (the centerline Figure 7a) (Figure 7b). The prosthesis is then lightly tapped into place. A snug fit is expected as the splines are allowed to cut into the bone. The prosthesis is driven into the canal to the correct level as noted on the template, the reamer, and the trial. The head trial can be placed onto the neck of the prosthesis and another trial reduction can be completed to recheck stability, range of motion, and leg length. The appropriate head is then placed onto the trunion of the prosthesis and lightly tapped to secure the taper junction. Figure 5 Anteversion Marking Limb length can be adjusted using trials and implants one size over or under the reamed diameter. This will change leg length by approximately 3cm longer or shorter. It is also possible to change leg length in small increments by driving the implant deeper or letting the implant sit slightly proud. Figure 6 Figure 7a Centerline Figure 7b

The PLR splined revision stem is a one piece, fully tapered stem with an anatomical offset to resist subluxation. Closeup of Spline

Ordering Information Implants Instrumentation PLR Revision Stems PLR Template 11-158325 Implants Trials Reamers Part No. Part No. Part No. Description 11-158314 31-158314 31-158334 14mm x 270mm 11-158315 31-158315 31-158335 15mm x 270mm 11-158316 31-158316 31-158336 16mm x 270mm 11-158317 31-158317 31-158337 17mm x 270mm 11-158318 31-158318 31-158338 18mm x 270mm 11-158319 31-158319 31-158339 19mm x 270mm 11-158320 31-158320 31-158340 20mm x 270mm 11-158321 31-158321 31-158341 21mm x 270mm 11-158322 31-158322 31-158342 22mm x 270mm 11-158323 31-158323 31-158343 23mm x 270mm 11-158324 31-158324 31-158344 24mm x 270mm References 1 Goldberg VM, Stevenson S, Feighan J, Davy D: Biology of Grit-blasted Titanium Alloy Implants, Clinical Orthopedics, 319:122 9, October, 1995. 2 Hacking SA, Bobyn JD, Tanzer M, Krygier JJ: The Osseous Response to Corundum Blasted Implant Surfaces in a Canine Hip Model, Clinical Orthopedics, 364:240 53, July, 1999. 3 Data on file at Biomet RingLoc is a registered trademark of Biomet, Inc. PLR is a trademark of Biomet, Inc. P.O. Box 587, Warsaw, IN 46581-0587 219.267.6639 2000 Biomet Orthopedics, Inc. All Rights Reserved web site: http://www.biomet.com email: biomet@biomet.com Form No. Y-BMT-691/123100/K