Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer

Similar documents
Workflow for Screening and Quantification of the SAMHSA (NIDA) Panel in Urine Using UHPLC-TOF

Workflow for Screening and Quantification of the SAMHSA (NIDA) Panel in Serum Using UHPLC-TOF

Rapid Screening and Quantitation of Postharvest Fungicides on Citrus Fruits Using AxION DSA/TOF and Flexar SQ MS

A Comprehensive Screening of Illicit and Pain Management Drugs from Whole Blood Using SPE and LC/MS/MS

A Simple and Accurate Method for the Rapid Quantitation of Drugs of Abuse in Urine Using Liquid Chromatography

Development of a Screening Analysis by LC Time-Of-Flight MS for Drugs of Abuse Application

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations

PerkinElmer talks at spring TCJJP Meeting Wednesday, Mar. 14, 2012

Application Note LCMS-108 Quantitation of benzodiazepines and Z-drugs in serum with the EVOQ TM LC triple quadrupole mass spectrometer

Fast and easy separation of 23 drugs of abuse. including high, stable resolution of isobaric opioids from human urine by UHPLC-MS/MS

4.5 Minute Analysis of Benzodiazepines in Urine and Whole Blood Using LC/MS/MS and an Ultra Biphenyl Column

Quantitative Analysis of Opiates in Urine Using RRHT LC/MS/MS. Application. Authors. Introduction. Abstract. Forensics

Analyze Barbiturates in Urine with Agilent 6430 LC/MS/MS and Poroshell 120 EC-C18

Analysis of Common Sweeteners and Additives in Beverages with the PerkinElmer Flexar FX-15 System Equipped with a PDA Detector

Comprehensive Forensic Toxicology Screening in Serum using On-Line SPE LC-MS/MS

SWATH Acquisition Enables the Ultra-Fast and Accurate Determination of Novel Synthetic Opioids

Quantification of lovastatin in human plasma by LC/ESI/MS/MS using the Agilent 6410 Triple Quadrupole LC/MS system

Analysis of Isoflavones with the PerkinElmer Flexar FX-15 UHPLC System Equipped with a PDA Detector

Automated Direct Sample Analysis (DSA/TOF) for the Rapid Screening and Confirmation of Illicit Street Drugs

Cannabinoid Profiling and Quantitation in Hemp Extracts using the Agilent 1290 Infinity II/6230B LC/TOF system

Removal of Triton X-100 from Plasma Samples Using Mixed-Mode Solid Phase Extraction (SPE)

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Comparison of Different Whole Blood Sample Pretreatment Methods for Targeted Analysis of Basic Drugs

Edgar Naegele. Abstract

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Determination of Selected Illegal Drugs and its Important Metabolites in Waste Water by Large Volume Direct Injection HPLC-MS/MS

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

Ultra High Definition Optimizing all Analytical Dimensions

Chemical Analysis Business Operations Waters Corporation Milford MA

The Development of LC/MS Methods for Determination of Polar Drugs of Abuse in Biological Samples

Analysis of Cannabinoids in Cannabis by UHPLC Using PDA Detection

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Determination of β2-agonists in Pork Using Agilent SampliQ SCX Solid-Phase Extraction Cartridges and Liquid Chromatography-Tandem Mass Spectrometry

Analysis of drugs of abuse in biological matrix using Time of Flight technology

Extraction of Aflatoxin M1 From Infant Formula Using ISOLUTE Myco SPE Columns prior to LC-MS/MS Analysis

Vitamin D3 and related compounds by ESI and APCI

[ APPLICATION NOTE ] High Sensitivity Intact Monoclonal Antibody (mab) HRMS Quantification APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Rapid, Simple Impurity Characterization with the Xevo TQ Mass Spectrometer

Modernizing the Forensic Lab with LC-MS/MS Technology

Extraction of Multiple Mycotoxins From Grain Using ISOLUTE Myco prior to LC-MS/MS Analysis

Application Note. Author. Abstract. Introduction. Food Safety

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid

Analysis of FAMEs Using Cold EI GC/MS for Enhanced Molecular Ion Selectivity

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

Determination of Amphetamine and Derivatives in Urine

Direct Analysis of Urinary Opioids and Metabolites by Mixed-Mode µelution SPE Combined with UPLC/MS/MS for Forensic Toxicology

Extraction of Multiple Mycotoxins From Nuts Using ISOLUTE Myco prior to LC-MS/MS Analysis

Analysis of Cannabinoids in Hemp Seed Oils by HPLC Using PDA Detection

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System

Characterization of an Unknown Compound Using the LTQ Orbitrap

Analysis of Phenolic Antioxidants in Edible Oil/Shortening Using the PerkinElmer Altus UPLC System with PDA Detection

Profiling of Endogenous Metabolites Using Time-of-Flight LC/MS with Ion Pair Reverse Phase Chromatography

Analysis of Cannabinoids in Hemp Seed Oils by HPLC Using PDA Detection

One Source Toxicology Laboratory, 1213 Genoa Red Bluff, Pasadena, Texas 77504

Robust extraction, separation, and quantitation of structural isomer steroids from human plasma by SPE-UHPLC-MS/MS

PosterREPRINT RAPID, SELECTIVE SCREENING OF URINE SAMPLES FOR GLUCURONIDES BY LC/MS/MS INTRODUCTION ABSTRACT

A RAPID AND SENSITIVE ANALYSIS METHOD OF SUDAN RED I, II, III & IV IN TOMATO SAUCE USING ULTRA PERFORMANCE LC MS/MS

Determination of Amantadine Residues in Chicken by LCMS-8040

Rapid and Accurate LC-MS/MS Analysis of Nicotine and Related Compounds in Urine Using Raptor Biphenyl LC Columns and MS-Friendly Mobile Phases

The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples

Identification of Ginsenosides Using the SCIEX X500R QTOF System

Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification

Determination of Opiates and Metabolites in Blood Using Electrospray LC/MS. Application Note

A Novel Platform of On-line Sample Pre-treatment and LC/MS/MS Analysis for Screening and Quantitation of Illicit Drugs in Urine

Application of LC/Electrospray Ion Trap Mass Spectrometry for Identification and Quantification of Pesticides in Complex Matrices

Application Note. Abstract. Authors. Pharmaceutical

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Dienes Derivatization MaxSpec Kit

LCMS-8050 Drugs of Abuse: 113 Analytes with Polarity Switching

Fast and simultaneous analysis of ethanol metabolites and barbiturates using the QTRAP 4500 LC-MS/MS system

High Throughput Extraction of Opiates from Urine and Analysis by GC/MS or LC/MS/MS)

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

A FORENSIC TOXICOLOGY METHOD FOR THE DETERMINATION OF DESOMORPHINE, HEROIN, METHADONE, BUPRENORPHINE AND METABOLITES IN URINE USING LC/MS QQQ

E XCEL LENCE JUST GOT BETTER UCT FORENSICS CLEAN SCREEN XCEL SPE COLUMNS

Performance of an ultra low elution volume 96-well plate

Authors. Abstract. Introduction. Environmental

Quantitative Analysis of -Hydroxybutyrate in Hair Using Target Analyte Finding Processing of Comprehensive GC-HRT Data

LC/MS/MS Analysis of Metabolites of Synthetic Cannabinoids JWH-018 and JWH-073 in Urine

Supporting Information

Determination of Benzodiazepines in Urine by CE-MS/MS

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

Overview. Introduction. Experimental. Cliquid Software for Routine LC/MS/MS Analysis

Analyze Drugs of Abuse with Agilent J&W Ultimate Plus Tubing in an Inert Flow Path

[ APPLICATION NOTE ] APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEYWORDS

4-Fluoroethamphetamine

Authors. Abstract. Forensic Toxicology. Irina Dioumaeva, John M. Hughes Agilent Technologies, Inc.

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Agilent s LC/MS Portfolio and Applications Examples

LC/MS Method for Comprehensive Analysis of Plasma Lipids

Qualitative and quantitative determination of cannabinoid profiles and potency in CBD hemp oil using LC/UV and Mass Selective Detection

Ultra-Fast Forensic Toxicological Screening and Quantitation under 3 Minutes using SCIEX X500R QTOF System and SCIEX OS 1.

Transcription:

application Note Liquid Chromatography/ Mass Spectrometry Authors Sharanya Reddy Blas Cerda PerkinElmer, Inc. Shelton, CT USA Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer Introduction Drugs of abuse are typically screened by immunoassay techniques (EMIT) and confirmed by GC/MS or LC/MS/MS assays. 1-3 We present an alternative route for analysis, quantitation and confirmation of these drugs using a Time-of-flight (TOF) mass spectrometer. The PerkinElmer AxION 2 TOF MS, fitted with ADC detector technology, provides a wide dynamic range and is suitable for quantitative analysis of target drugs. In contrast to quadrupole instruments that are not very sensitive in the scanning mode, TOF can be used for qualitative screening of unknown compounds. The accurate mass and isotope ratios provided by the AxION 2 TOF MS along with retention time matching is used to identify and confirm the presence of both known and unknown target analytes without the need for fragmentation information. The AxION 2 TOF MS fitted with the dual-probe Ultraspray 2 ESI source allows for simultaneous infusion of both internal calibrant mixture and the LC effluent so the mass accuracy is maintained throughout the analysis at low ppm level with continuous lock mass correction.

Experimental SPE extraction: Blank urine samples were spiked with varying concentrations of analyte drugs (1-10,000 ppb) and with internal standards (d5-diazepam and d3-doxepine at 50 ng/ ml each). Urine (300 µl) was extracted over PerkinElmer HLB solid phase cartridges (30 mg/1 ml cartridges, 30 micron particle size, Part Number N9306650). The cartridges were conditioned with methanol (1 ml) followed by water (1 ml). The urine samples were then loaded on the cartridges and extracted at a flow rate of 1 ml/min. The cartridges were washed with 5% methanol in water (1 ml) dried under vacuum for ~ 1 min and eluted with methanol (1 ml). The eluate was dried and reconstituted to 300 µl with water. LC conditions: Pump: PerkinElmer Flexar FX-10 pump Flow: 0.4 ml/min Mobile phase A: Water containing 0.1% formic acid Mobile phase B: Acetonitrile containing 0.1% formic acid Gradient conditions: 95% A/5% B for 1 min., followed by 95% A/5% B to 10% A/90% B in 5 min. (linear gradient), maintained at 10% A/90% B for next 1.5 min. Injection volume: 4 μl in partial fill mode Column used: PerkinElmer Brownlee SPP C-18, 2x50 mm, 2.7 μm, 25 C MS conditions: Mass spectrometer: PerkinElmer AxION 2 TOF MS Ionization source: PerkinElmer Ultraspray 2 (Dual ESI source) Ionization mode: Positive Spectral acquisition rate: 3 spectra/sec Capillary exit voltage: 100 V TrapPulse mode: 100-1000 m/z (D7:68, D8:91) Internal calibration was performed using m/z 118.08625 and 622.02896 as lock mass ions. Results The separation and analysis of 21 drugs of abuse belonging to several classes including opiates, benzodiazepines, amphetamines and cannabinoids was achieved within 7 min (Figure 1). The recovery of the drugs over SPE cartridges was estimated to be 80% for nearly all of the analytes (determined at 20 ppb concentration spiked in urine). Using the proprietary TrapPulse technology of the AxION 2 TOF MS, we observed excellent limits of detection (LODs, based on S/N 3) for a majority of the drugs (1-5 ppb, Table 1). In the TrapPulse mode of operation, the duty cycle of the TOF increases, resulting in significant improvement in signal. The LODs measured by the AxION 2 TOF MS were 100-400 times lower than the required cut-off levels by the EMIT assays for most classes of drugs of abuse showing the excellent sensitivity provided by the instrument. Figure 1. Analysis of different classes of drugs of abuse within 7 min. Table 1. LODs of some representative drugs of abuse in human urine. Drug LOD in urine (ng/ml or ppb) Methamphetamine 1 Amphetamine 5 Diazepam 1 Haloperidol 4 Alpralozam 1 Fluorazepam 5 Codeine 5 Morphine 5 Doxepine 1 Haloperidol 4 Cocaine 1 Benzoylecognine 1 Methadone 1 3, 4 MDA 1 MDEA 1 Oxycodone 1 The linearity of a representative drug, oxycodone, is shown in Figure 2. The analysis showed regression values (r 2 = 0.999) over ~4 orders of dynamic range thereby emphasizing the quantitative capabilities of the AxION 2 TOF MS. Figure 2. Calibration curve of oxycodone over a concentration range of 1-10,000 ppb (d 3 doxepine used as internal standard). 2

Table 2. Shows theoretical mass, observed mass and mass error of representative drugs of abuse. Monoisotopic Mass Analyte Formula [M+H] + Observed Mass Mass Error (ppm) Methamphetamine C 10H 15N 150.1277 150.1273-2.66 Amidarone C 25H 29I 2NO 3 646.0310 646.0312 0.39 Amphetamine C 19H 13N 136.1121 136.1113-5.88 Codeine C 18H 21NO 3 300.1594 300.1594-0.07 Diazepam C 16H 13ClN 2O 285.0789 285.0792 0.98 Doxepine C 19H 21NO 280.1696 280.1695-0.36 Haloperidol C 21H 23ClFNO 2 376.1474 376.1474 0.00 Morphine C 17H 19NO 3 286.1438 286.1434-1.40 Flurazepam C 21H 23ClFN 3O 388.1586 388.1593 1.80 Alprolazam C 17H 13ClN 4 309.0902 309.0900-0.65 Cocaine C 17H 21NO 4 304.1543 304.1550 2.30 EDDP perchlorate C 20H 24ClNO 4 278.1903 278.1906 1.08 (fragment, [M-O 4Cl] + ) Benzoylecognine C 16H 19NO 4 290.1387 290.1394 2.41 Methadone C 21H 27NO 310.2165 310.2171 1.93 3,4 MDA C 10H 13NO 2 180.1019 180.1016-1.67 MDEA C 12H 17NO 2 208.1332 208.1326-2.88 MDMA C 11H 15NO 2 194.1176 194.1169-3.61 Phenylpropylamine C 9H 13NO 152.1069 152.1060-5.92 Oxycodone C 18H 21NO 4 316.1543 316.1545 0.54 The major advantage of TOF MS is the accurate mass capability for confirming the presence of a given analyte. Table 2 shows examples of accurate mass of the drugs of abuse provided by AxION 2 TOF MS in comparison to the corresponding theoretical mass. The accurate mass, isotope ratios and retention time is used as confirmation of the presence of a given analyte. Analyte detection in urine at low parts-per-billion level is significantly improved by narrowing the extracted ion mass window as shown in Figure 3 (a-c) for alpralozam. The noise is reduced considerably with a narrower mass window improving overall S/N and thereby providing more confidence in analysis of a given compound. Figure 3. (a-c) Shows the effect of narrowing the extracted ion mass window on the S/N of 1 ppb of alpralozam in urine. Blue trace is of control urine and the black trace is of spiked urine. 3

In addition to the federally mandated workplace testing, which involves analysis of known compounds, forensic toxicologists have the challenge of often looking for unknown compounds in samples. TOF MS technology provides the user the ability to re-mine already acquired data for presence of these unknown compounds. As an example, we analyzed an unknown compound eluting at retention time (Rt) of ~2.5 mins (Figure 4). The identity of this unknown peak was determined with the AxION EC ID software using accurate mass and isotope ratios. The software searches against a database and provides a ranked summary of the potential target compounds for a given elemental composition. The compound was determined to be caffeine, one of the most commonly consumed stimulants. Similar analysis using AxION EC ID software suggested the unknown peaks eluting at Rt ~1 and ~1.8 mins were most likely theobromine/theophylline/paraxanthine, which are all isomer metabolites of caffeine. The presence of caffeine and its metabolites in urine was further confirmed by retention time matching with standards. Figure 4. Panel a: shows the elution profile of unknown targets that were identified as caffeine and metabolites of caffeine using accurate mass and isotope ratio obtained from the AxION 2 TOF MS along with the AxION EC ID software; Panel b: with the AxION EC ID software, the elemental composition C 7H 8N 4O 2 was the first hit for m/z 181.0718; Panel c: upon sorting for decreasing AIDS (chemical listed for most number of assays), the first possible candidate for C 7H 8N 4O 2 was theophylline and the second choice was paraxanthine, both of which are metabolites of caffeine. 4

Conclusions The study of drugs of abuse in urine presented in this application note shows the exceptional capabilities of the AxION 2 TOF MS for both quantitative analysis and for screening of additional components present in urine. The detection limits of several classes of drugs of abuse analyzed by the TOF were 100-400 times lower than those required by non-specific immunoassays. Besides the wide quantitative dynamic range of the AxION 2 TOF MS, which rivals capabilities of the triple quadrupole instruments, the TOF also provides full spectrum information which allows for screening of non-target compounds. This application note was developed for 21 drugs of abuse; however, it can easily be extended to a wider range of drugs of abuse used in screening analysis with the help of the easy to use quantitation and reporting software package provided with the instrument. References 1. Moeller, K.E., Lee, K.C., Kissack, J.C., Mayo Clin. Proc. 83 (2008) 66. 2. Thevis, M., Schanzer, W., Mass Spectrom. Rev. 26 (2007) 79. 3. Gallardo E., Barroso, M., Queiroz, J.A., Drug Test Anal. 1 (2009) 109. PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com For a complete listing of our global offices, visit www.perkinelmer.com/contactus Copyright 2012, PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 010198_01