Radiographic Evaluation of Dorsal Screw Penetration After Volar Fixed-Angle Plating of the Distal Radius: A Cadaveric Study

Similar documents
Technique Guide. 2.4 mm Variable Angle LCP Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

The Utility of the Fluoroscopic Skyline View During Volar Locking Plate Fixation of Distal Radius Fractures

ORIGINAL ARTICLE TREATMENT DISTAL RADIUS FRACTURE WITH VOLAR BUTTRESS TECHNIQUE- A CLINICAL STUDY

Distal Radius Plate Instrument and Implant Set. Discontinued December 2017 DSUS/TRM/0916/1063(1)

Disclosures. Distal Radius Fractures 5/16/2017. Distal Radius Fractures: Complications & Limitations of the Volar Approach

Disclosure. 85% Dorsal. 15% Volar. Distal Radius Fractures- Volar Plating for All. Skeletal Dynamics- Consultant

CASE REPORT. Distal radius nonunion after volar locking plate fixation of a distal radius fracture: a case report

Treatment of unstable distal radius fractures with the volar locking plate

Complications of Distal Radius Fractures. How to Treat a Distal Radius Fx 11/13/2017. Michael S. Bednar, M.D. Loyola University Chicago

AcUMEDr. FoREARM ROD SYSTEM

Acu-Loc Wrist Plating System. Surgical Technique

Chapter 4: Forearm 4.3 Forearm shaft fractures, transverse (12-D/4)

MINIMALLY INVASIVE PLATE OSTEOSYNTHESIS FOR DISTAL RADIUS FRACTURES: SURGICAL TECHNIQUE M. TOBE 1, K. MIZUTANI 1, Y. TSUBUKU 1, Y.

Emile N. Brown, MD, and Scott D. Lifchez, MD

Distal Radius and Distal Ulna Plates System Self-Tapping Spherical Locking Screw Self-Tapping Conical Locking Screw Cortex Screw

Surgical Technique Guide

SURGICAL TECHNIQUE GUIDE PROTEAN. radial head plate. As described by: Jorge L. Orbay, M.D. Miami Hand & Upper Extremity Institute Miami, Florida

Variable Angle LCP Volar Rim Distal Radius Plate 2.4. For fragment-specific fracture fixation with variable angle locking technology.

Instrument and Implant for wrist fracture

6/5/2018. DISCLOSURES Hassan R. Mir, MD, MBA, FACS. Evolution of Distal Radius Fracture Treatment [Chung Hand Clinics 2012]

Mark VanDer Kaag 1, Ajmal Ikram 2. Hand Unit, Tygerberg Hospital University of Stellenbosch

The NBX Non-Bridging External Fixator A Non-Bridging External Fixator/Locking Plate capturing a series of.062mm K-wires and 3mm half-pins that are

Locking Radial Head Plates

MICRONAIL. Intramedullary Distal Radius System SURGICAL TECHNIQUE

Acu-Loc 2 Wrist Plating System. Surgical Technique

Acu-Loc 2 Wrist Plating System

Wrist Fixation System

Radiographic Evaluation and Classification of Distal Radius Fractures

Surgical Technique. Wrist Plating System

SURGICAL TECHNIQUE GUIDE PROTEAN. radial head plate. As described by: Jorge L. Orbay, M.D. Miami Hand & Upper Extremity Institute Miami, Florida

Long-term Results of Dorsally Displaced Distal Radius Fractures Treated With the Pi-Plate: Is Hardware Removal Necessary?

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

2.4 mm LCP Volar Column Distal Radius Plates. Part of the 2.4 mm LCP Distal Radius System.

journal ORIGINAL RESEARCH

Distal Radius Plate 2.4/2.7 dorsal and volar

Clinical Study Dorsally Comminuted Fractures of the Distal End of the Radius: Osteosynthesis with Volar Fixed Angle Locking Plates

WINSTA-R. Distal Radius System

Surgical Technique. Forearm Fracture Solutions

Acu-Loc Wrist Spanning Plate System. Surgical Technique

2.4 mm LCP Distal Radius System

A STUDY OF THE FUNCTIONAL OUTCOME OF LOW PROFILE DORSAL PLATING IN DISTAL END RADIUS FRACTURES

VariAx TM Distal Radius Locking Plate System

Hardware-Related Complications After Dorsal Plating for Displaced Distal Radius Fractures

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

Complications and functional outcome following fixation of complex, intra-articular fractures of the distal radius with the AO Pi-Plate

Fractures of the distal radius have occupied a storied history

Reproducibility of Three Dimensional Digital Preoperative Planning for the Osteosynthesis of Distal Radius Fractures

Distal Ulnar Locking Plate

Volar Distal Radius Plating System Surgical Technique

Index. Springer International Publishing Switzerland 2016 J.N. Lawton (ed.), Distal Radius Fractures, DOI /

Assessment of penetration of dorsal screws after fixation of the distal radius using ultrasound: cadaveric study

of intramedullary nailing for correction of posttraumatic

Volar Distal Radius Plating System Surgical Technique

VariAx Distal Radius Locking Plate System

Volar fixed-angle plating for distal radius fractures 刀圭会協立病院 津村敬

Factors Predicting Late Collapse of Distal Radius Fractures

Distal radius fractures raises considerable interest and

Hand and wrist emergencies

COMPARATIVE STUDY OF FUNCTIONAL OUTCOME OF EXTERNAL AND INTERNAL FIXATION IN TREATMENT OF COMMINUTED DISTAL RADIUS FRACTURES

Malunited extra-articular distal radius fractures: corrective osteotomies using volar locking plate

Degrees Of Volar Angulation In Distal Radius Fracture Effects Distal Radioulnar Joint Stability:a Biomechanical Study

QUICK REFERENCE GUIDE. The Pennig Dynamic Wrist Fixator. Part A: Trans-articular application

DVR Crosslock Distal Radius Plating System. Product Brochure

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

Wright Medical Technology, Inc Cherry Road Memphis, TN

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Technique Guide. VA-Locking Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Long Volar Plates for Diaphyseal-Metaphyseal Radius Fractures LCP. Dia-Meta Volar Distal Radius Plates. Surgical Technique

The Wrist Fusion Set. Stainless Steel and Titanium TECHNIQUE GUIDE. Instruments and implants approved by the AO Foundation

Essential Radiographic Evaluation for Distal Radius Fractures

DVR Anatomic Plate. Surgical Technique

Index. Note: Page numbers of article titles are in boldface type. Hand Clin 21 (2005)

Angle Stable Distal Radial Plate System WINSTA-R

LCP Distal Humerus Plates

Comparative Study between Bridging External Fixation vs Volar Plating (Ellis-T Plate) for Comminuted Fracture of the Distal End Radius

Conventus CAGE PH Surgical Techniques

FUNCTIONAL AND RADIOLOGICAL OUTCOMES AFTER VOLAR LOCKING PLATE FIXATION IN TYPE C DISTAL RADIUS FRACTURES A RETROSPECTIVE STUDY

Surgical Technique Carpal Fusion

AcUMEDr. LoCKING CLAVICLE PLATE SYSTEM

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

D-RAD SMART PACK Plating System. A series of case studies

Zimmer Small Fragment Universal Locking System. Surgical Technique

FOOSH It sounded like a fun thing at the time!

AcUMEDr. Anatomic Midshaft Forearm Plates

2.4 mm Variable Angle LCP Intercarpal Fusion System. Variable angle locking technology for mediocarpal partial arthrodesis.

Integra. Spider and Mini Spider Limited Wrist Fusion System SURGICAL TECHNIQUE

The Flower Medial Column Fusion Plate

ORIGINAL ARTICLE. possible. Accurate assessment of standard radiographs is essential for appropriate 3. management. And includes true posterior- 4

Distal radius fractures are common injuries encountered

Hoffmann II Compact External Fixation System. Modular System for

5/9/2017. Distal Radius Fractures: What s New? What s New? Or Maybe New to you. Single Pak Sterile distal radius kits

INTERNAL FIXATION OF THE METACARPALS AND PHALANGES P. BURGE

Disclosures. Epidemiology 10/12/2015. Distal Radius Fractures: Staying Out of Trouble

Plating of the Distal Radius

Fractures of the Radial and Ulnar Shafts In the Pediatric Patient

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

Injury, Int. J. Care Injured 42 (2011) Contents lists available at ScienceDirect. Injury. journal homepage:

Failed Extensor Indicis Proprius Tendon Transfer for Extensor Pollicis Longus Tendon Rupture after Distal Radial Fracture

Transcription:

HAND (2007) 2:144 150 DOI 10.1007/s11552-007-9038-2 Radiographic Evaluation of Dorsal Screw Penetration After Volar Fixed-Angle Plating of the Distal Radius: A Cadaveric Study Steven D. Maschke & Peter J. Evans & David Schub & Richard Drake & Jeffrey N. Lawton Received: 17 March 2007 /Accepted: 22 March 2007 /Published online: 26 April 2007 # American Association for Hand Surgery 2007 Abstract Introduction Extensor tendon irritation and attritional tendon ruptures are potentially serious complications after open reduction and internal fixation of distal radius fractures. These complications are well recognized after dorsal plating of distal radii; and these are now being reported after errant screw placement during volar fixedangle plating. Intraoperative detection of improper screw placement is critical, as corrective action can be taken before completion of the operative procedure. The purpose of this study was to define the extensor tendon compartments at risk secondary to dorsal screw penetration and to compare pronation and supination fluoroscopic images with standard lateral images in demonstrating dorsal screw prominence during volar locked plating. S. D. Maschke Department of Orthopedic Surgery, Cleveland Clinic, 9500 Euclid Avenue A41, Cleveland, OH 44195, USA e-mail: maschks@ccf.org P. J. Evans : J. N. Lawton (*) Section of Hand & Upper Extremity Surgery, Department of Orthopedic Surgery, Cleveland Clinic, 9500 Euclid Avenue A40, Cleveland, OH 44195, USA e-mail: lawtonj@ccf.org D. Schub Case Western Reserve, University School of Medicine, Cleveland, OH, USA R. Drake Lerner College of Medicine, NA24, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH 44195, USA Methods Eight fresh-frozen human cadaveric upper extremities underwent fixation with a volar, fixed-angle distal radius locked plate (Wright Medical Technology, Arlington, TN). Three fluoroscopic views (lateral, supinated, and pronated) followed by dorsal wrist dissections were compared to determine accuracy in detecting dorsal screw prominence and extensor tendon compartment violation. Subsequently, screws measuring 2, 4, 6, 8, and 10(mm longer than the measured depths were sequentially inserted into each distal locking screw, with each image deemed either in (completely inside the bone) or out (prominent screw tip dorsally-would typically be exchanged for a shorter screw intraoperatively). Results The radial most distal locking screw (position 1) violated either the first (25%) or second (75%) extensor tendon compartments. The average screw prominence required for radiographic detection was: 6.5(mm for lateral views and 2(mm for supinated views. Pronated views did not identify prominent screws. Screws occupying plate position 2 consistently entered Lister s tubercle, with 5/8 exiting the apex and 3/8 exiting the radial base. The average screw prominences for radiographic detection were: 2.75(mm-lateral views and 3.0(mm-supinated views. Although the screws entered the second dorsal compartment, they did not encroach upon either of the tendons. Screws occupying plate position 3 violated the third extensor tendon compartment in 7/8 specimens with 1/8 exiting the Ulan base of Lister s tubercle. The average screw prominences for radiographic detection were: 3.5 (mm-lateral views and 2.5(mm-pronated views. Supinated views did not identify prominent hardware. Screws occupying plate position 4 all violated the IV dorsal extensor compartment-2/8 screws were noted to tent the posterior interosseous nerve. The average screw prominences required for radiographic detection were: 4.0(mm-lateral

HAND (2007) 2:144 150 145 Figure 1 Sequelae of missed dorsal screw prominence: (a) 45 pronation view long ulnar most distal locking screw (arrow): (b) clinical photo: extensor tenosynovitis (arrow) required hardware removal and tenosynovectomy. views and 2.5(mm-pronated views. The supinated views did not identify prominent screws. Conclusions Volar fixed-angle plating has shown great promise in the advancement of distal radius fracture management. We have seen in our referral practices and in the literature an increase in the number of extensor tendon complications arising from unrecognized dorsally prominent screws, pegs, or tines. Standard PA and lateral radiographs cannot adequately visualize screw position and length secondary to the complex geometry of the dorsal cortex. We believe this study supports the routine application of intraoperative, oblique pronosupination fluoroscopic imaging for enhanced confirmation of distal locking screw position and length. complications and poor clinical results despite satisfactory fracture reduction and fixation (Fig. 1) [11]. Intraoperative detection of improper screw placement is critical, as corrective action can be taken before completion of the procedure. The anatomy of the distal forearm and wrist accounts for the high frequency of soft-tissue complications arising from the insertion of metallic implants. The volar surface of the distal radius is relatively flat and separated from the overlying flexor tendons by the interposed muscle belly of the pronator quadratus (PQ) (Fig. 2). Re-approximating the muscle belly of the PQ over volarly inserted plates and screws maintains a low friction gliding surface for the flexor tendons. Additionally, instrumentation confined to the PQ fossa are remote from direct contact with the flexor tendons. Conversely, the dorsal surface of the distal radius has a complex geometry with tendon directly opposed to bone (Fig. 2). Dorsal hardware disrupts this intimate bone tendon relationship, placing the extensor tendons at risk for irritation and attritional rupture. The purpose of this study was to define the extensor tendon compartments at risk secondary to dorsal screw penetration and determine the accuracy of fluoroscopic evaluation in demonstrating dorsal screw prominence during volar locked plating. Materials and Methods Eight fresh frozen human cadaveric upper extremities were obtained from the Anatomic Gift Registry. Each extremity had a volar, fixed-angle distal radius locking plate Keywords Distal radius fractures. Dorsal screw penetration. Volar fixed-angle plating Introduction Extensor tendon irritation and attritional tendon ruptures are serious complications after open reduction and internal fixation of distal radius fractures. These complications are well recognized after dorsal plating of distal radius fractures; they are now being reported with increasing frequency secondary to errant screw placement during volar fixed-angle plating [1, 5, 9, 12, 14, 16, 18]. Volar fixedangle plating has gained popularity in the treatment of unstable fractures of the distal radius by providing stable fracture fixation while avoiding surgical exposure and violation of the dorsal soft-tissues. However, inserting screws of improper length and/or orientation, can lead to Figure 2 T1-weighted axial MRI of the wrist DRUJ: pronator quadratus interposed between the flexor tendons (dashed arrow) and the volar cortex of the distal radius; extensor tendons (solid arrow) coursing directly over the dorsal cortex of the distal radius. DRUJ= distal radioulnar joint.

146 HAND (2007) 2:144 150 Figure 3 Screw Position 1 (screw is 6 mm proud): (a) neutral lateral screw in (arrow); (b) 45 supination screw out ; (c) direct visualization screw violation of extensor compartment II (arrow). Screws occupying the remaining distal locking holes have been backed out: isolating plate position 1 (dashed arrows). (LOCON, Wright Medical Technology, Arlington, TN) inserted via a standard volar approach. All procedures were performed by two fellowship-trained hand surgeons. Coronal plane alignment was confirmed via fluoroscopic PA images; and the plates were secured to the radial shafts with 3.5-mm cortical screws. Next, all four of the distal locking screw holes were drilled utilizing the distal locking guide and sequentially numbered 1 4 (radial to ulnar). The bone was intentionally drilled in bicortical fashion, and the depths were measured and recorded using the standard depth gauge; 2.7-mm distal locking screws of the measured length were inserted into each (numbered) plate position. Violation of the dorsal cortex was evaluated fluoroscopically in three planes: (1) neutral lateral (utilizing scaphopisocapitate alignment), (2) supinated view, and (3) pronated view [19] (Figs. 3 and 4). The angles of the pronation and supination fluoroscopic images were not standardized. This was intentionally done to further simulate the intra-operative environment and allow for live time imaging in these multiple planes. Four observers (two attending upper extremity surgeons, one orthopedic surgery resident, and one medical student) evaluated the images and came to a consensus whether each screw was deemed in (completely inside the bone) or out (prominent screw tip dorsally would be typically exchanged for a shorter screw intra-operatively). After hardware placement and fluoroscopic evaluation, the dorsal cortex of the distal radius and all extensor tendon compartments over the distal radius were exposed via a standard dorsal approach to the wrist. Screw prominence was evaluated under direct visualization and the extensor compartment violated by the drill hole was photographed and recorded for each distal-locking screw position. Evaluation progressed from screw position 1 sequentially through position 4. To allow discrete evaluation of screw position 1, the remaining screws were backed out 5 mm; this was repeated for each successive position. After the depthgauge-measured screws were visually evaluated, the screw occupying plate position 1 (radial most distal-locking screw) was removed and exchanged with a screw measuring 2 mm longer. Direct visualization confirmed screw tip violation of the dorsal cortex. The series of three fluoroscopic images Figure 4 Screw position 4 (screw is 4 mm proud): (a) neutral lateral screw in (arrow); (b) 45 pronation screw out (arrow); (c) direct visualization screw violation of extensor compartment IV (arrow). Screws occupying the remaining distal locking holes have been backed out: isolating plate position 4 (dashed arrows).

HAND (2007) 2:144 150 147 Table 1 Average length (mm) of screw prominence for radiographic detection. Srew Position were repeated for this longer screw and a consensus agreement was made regarding screw prominence, in or out, for all views. This sequence was repeated for plate position 1, with screws increasing in length by 2-mm increments, until the screw tip was visualized out on all radiographic views. Distal-locking screw positions 2, 3, and 4 underwent the identical sequence of screw exchange and fluoroscopic/visual evaluation. Results Supination Pronation Lateral Plate Lateral 1 2.0 (2 2) N/A 6.75 (6 10) 6.0 (4 8) 2 2.5 (2 4) N/A 2.75 (2 4) 2.5 (2 4) 3 N/A 2.5 (2 4) 3.75 (2 6) 3.75 (2 6) 4 N/A 2.25 (2 4) 4.0 (4 4) 4.5 (4 6) Accurate fluoroscopic determination of proper screw length was variable and dependent upon which screw position was being evaluated and by which view. Table 1 summarizes the average screw prominence (shortest screw length deemed out starting screw length) required for fluoroscopic evidence of cortical penetration for each screw position and for each view. The shortest value for initial detection of dorsal screw prominence is 2 mm, as this is the length of the first intentionally long screw inserted. Table 2 summarizes the extensor tendon compartment violated by each distal-locking screw position. The radial most distal-locking screws (position 1) violated either the first (25%) or second (75%) extensor tendon compartments. Supination views align the x-ray beam with the radial aspect of the dorsal cortex and identified all prominent screws occupying position 1 (defined as 2 mm proud in this study). In contrast, the average screw prominence required before fluoroscopic detection utilizing a lateral view was 6.5 mm (i.e., a screw that appeared in on a lateral view was in reality 6.5 mm proud; Figs. 5 and 6). Likewise, the ulnar most distal-locking screws (position 4) always violated the IV dorsal extensor compartment 2/ 8 screws were actually noted to tent the posterior interosseous nerve. Pronation views align the x ray beam with the ulnar aspect of the dorsal cortex and identified prominent screws with an average threshold of 2.5 mm. The average screw prominence required for fluoroscopic detection was 4.0 mm on lateral views (Fig. 7). The more central distal locking screw positions (positions 2 and 3, respectively) exit the dorsal cortex either within or close to Lister s tubercle. These plate positions displayed less variability in identifying prominent screws, utilizing the different fluoroscopic views. Discussion Distal radius fractures account for approximately 15% of all extremity fractures [12]. The goals of treatment are to reestablish the osseous anatomy and restore painless wrist motion and strength. Open reduction and internal fixation with plate and screw constructs has enhanced the surgeon s ability to achieve stable fracture fixation, thus reducing immobilization time and expediting return to function [1, 2, 5, 6, 8, 10, 14, 15]. The implantation of plates and screws into the distal radius requires careful consideration and handling of the adjacent soft tissues and precise placement of the hardware to avoid both short and long-term complications. The most frequent complications after plate fixation of distal radius fractures are tendon irritation and attritional tendon ruptures occurring in up to 50% of reported cases [1, 9, 18]. Friction over dorsally implanted plates and screws can lead to tendon fraying and subsequent tendon rupture. Low-profile dorsal plating systems were introduced to better manage the dorsal soft-tissues, but relatively high rates of extensor tendon complications continue to be reported, requiring frequent implant removal [9]. Volar fixed-angle plating has been developed to enhance stable fracture fixation while avoiding surgical exposure and violation of the dorsal soft-tissues [8, 10, 14, 15]. Commercially available volar fixed-angle plating systems utilize fixed-angle pegs, screws, or tines to Table 2 Dorsal extensor tendon compartment violated. Srew Position Compartment I Compartment II Lister Tubercle Compartment III Compartment IV 1 2/8 6/8 0 0 0 2 0 0 8/8 0 0 3 0 0 1/8 7/8 0 4 0 0 0 0 8/8

148 HAND (2007) 2:144 150 Figure 5 Cartoon depiction of the lateral radiographic projection when distal locking screw in position 1 is too long. The lateral radiograph in the bottom right corner is of a screw in position 1 protruding 6 mm out of the cortex. Reprinted with permission of the Cleveland Clinic. support the articular fragments and subchondral bone. The volar surface of the distal radius is relatively flat and well suited for the application of plates and screws. The overlying flexor tendons course along the volar wrist, protected from the underlying bone and implanted hardware by the interposed muscle belly of the pronator quadratus. Orbay and colleagues report stable fixation, rapid fracture healing, and a low incidence of complications and/or subsequent operative procedures after volar fixation of distal radius fractures [14, 15]. However, as Figure 6 Cartoon depiction of the same screw length and position as in Fig. 5, but with the radiograph taken in the supinated view. The x-ray beam is aligned with the radial aspect of the dorsal cortex and easily detects the proud (6 mm) screw. Reprinted with permission of the Cleveland Clinic.

HAND (2007) 2:144 150 149 Figure 7 Cartoon depiction of the same screw length and position as in Fig. 5, but with the radiograph taken in the pronated view. The x- ray beam is aligned with the radial aspect of the dorsal cortex and easily detects the proud (6 mm) screw. Reprinted with permission of the Cleveland Clinic. these implants have gained in popularity, complications are being increasingly reported. Improper placement of volar plates and screws can obviously lead to intraarticular penetration, flexor tendon irritation, and/or neurovascular compromise [3, 4, 13, 15, 16]. Recently, extensor tendon complications arising from the unrecognized insertion of distal-locking screws of inappropriate length are being reported [12]. Comminution of the dorsal cortex of the distal radius challenges accurate depth-gauge measurement of screw length. Thus, intra-operative, radiographic/fluoroscopic confirmation of proper screw trajectory and length, with critical attention focused on both articular and dorsal cortical violation, can minimize the occurrence of poor operative results. Ironically, volar-locked constructs, which were innovative in addressing extensor tendon complications, are now causing dorsal soft-tissue complications by a completely different mechanism. This study sought to evaluate the risk to the extensor compartments secondary to errant screw placement, and to determine if routine intra-operative fluoroscopic evaluation could detect screws of improper length. Fluoroscopic/radiographic evaluation of hardware placement into the distal radius is challenging given the complex angular geometry of the bone. Standard PA and lateral images of the wrist do not account for the multiplanar tilts and inclinations of the distal radius and cannot reliably confirm proper screw position and length. Several investigators have evaluated the efficacy of different oblique radiographic views for assessment of the radiocarpal joint; aligning the x-ray beam with the normal radial and volar axes of the articular surface [4, 17]. These anatomic views have proven accurate and clinically useful for identifying screw penetration of the articular surface [4, 17]. Similarly, the dorsal cortex of the distal radius is inclined in more than one plane, prohibiting evaluation of the dorsal surface by a single fluoroscopic/radiographic projection. Our study evaluated the reliability of pronation and supination radiographic views in identifying dorsal screw prominence after volar fixed-angle plating and compared these results with a standard lateral radiographic view. The supination view was developed to better assess the radial portion of the dorsal cortex, encompassing the first and second dorsal extensor tendon compartments, which slope away from Lister s tubercle. Likewise, the pronation view was developed to improve fluoroscopic visualization of the ulnar portion of the dorsal cortex, containing extensor tendon compartments III and IV. The addition of these two oblique views improved our ability to detect screw penetration of the dorsal cortex compared with lateral images alone. Direct visualization of the dorsal cortex and adjacent soft tissues confirmed violation of the extensor tendons by prominent screws frequently missed on standard lateral fluoroscopic views. We believe that this finding supports the routine application of intra-operative pronation and supination fluoroscopic imaging for accurate confirmation of appropriate and safe distal-locking screw length. We attempted to simulate the intra-operative environment and utilize readily available surgical and fluoroscopic equipment for this study. Fluoroscopic imaging allows for live time imaging and immediate evaluation of the multiple views. The benefits of minimal interruption of the operative procedure, immediate image evaluation and live time imaging, far outweigh, in our opinion, the modestly diminished quality of the images when compared with standard radiographs. The angles of the pronation and supination radiographs were not standardized. This was intentionally done given the intra-operative environment for which these radiographs are intended. All images were taken by two fellowship-trained hand surgeons, both of whom already incorporate oblique intra-operative fluoroscopic imaging in practice. Also, a single volar-locked plating system was utilized for this study (LOCON, Wright Medical Technology, Arlington, TN). All current volar-locked plating systems incorporate multiple distal locking screws, tines, or pegs to support the articular surface. Although not evaluated in this study, different implant designs may impact the severity of soft-tissue compromise arising secondary to dorsally prominent hardware. Volar fixed-angle plating has shown great promise in the advancement of distal radius fracture management. Unfortunately, we have seen in both our referral practice, and in the literature, a steady increase in the number of extensor tendon complications arising from dorsally prominent hardware [7, 12]. The advancement of new technologies requires critical evaluation of outcomes, allowing the identification of complications and development of solutions to increase the number of favorable outcomes. We believe this study supports the routine application of intra-operative, oblique pronation and supi-

150 HAND (2007) 2:144 150 nation fluoroscopic images for confirmation of final screw position and length during volar fixed-angle plating of the distal radius. References 1. Axelrod TS, McMurtry RY. Open reduction and internal fixation of comminuted, intra-articular fractures of the distal radius. J Hand Surg 1990;15A:1 11. 2. Beharrie Aw, Beredjiklian PK, Bozentka DJ. Functional outcomes after open reduction and internal fixation for treatment of displaced distal radius fractures in patients over 60 years of age. J Orthop Trauma 2004;18(10):680 6. 3. Bell JSP, Wollstein R, Citron ND. Rupture of the flexor pollicis longus tendon: a complication of volar plating of the distal radius. J Bone Jt Surg Br 1998;80B:225 6. 4. Boyer MI, Korcek KJ, Gelberman RH, Gilula LA, Ditsios K, Evanoff BA. Anatomic, tilt x-rays of the distal radius: an ex vivo analysis of surgical fixation. J Hand Surg 2004;29A:116 22. 5. Jakob M, Rikli D, Regazzoni P. Fractures of the distal radius treated by internal fixation and early function: a prospective study of 73 consecutive patients. J Bone Jt Surg Br 2000;82B (3):340 4. 6. Jupiter JB. Plate fixation of the distal aspect of the radius: relative indications. J Orthop Trauma 1999;13(8):559 69. 7. Jupiter JB, Fernandez DL, Choon-Lai T, Fellman T, Ring D. Operative treatment of volar intra-articular fractures of the distal end of the radius. J Bone Jt Surg Am 1996;78A(12):1817 28. 8. Kamano M, Honda Y, Kazuki K, Yasuda M. Palmar plating for dorsally displaced, fractures of the distal radius. Clin Orthop 2002;397:403 8. 9. Kambouroglou GK, Axelrod TS. Complications of the AO/ASIF titanium distal radius plate system in internal fixation of the distal radius: a brief report. J Hand Surg. 1998;23A:737 41. 10. Liporace FA, Gupta S, Jeong GK, Gerard K, Stracher M, Kummer F, et al. A biomechanical comparison of a dorsal 3.5 mm T-plate and a volar, fixed-angle plate in a model of dorsally unstable distal radius fractures. J Orthop Trauma 2005;19(3):187 91. 11. McKay SD, MacDermid JC, Roth JH, Richards RS. Assessment of complications of distal radius fractures and development of a complications checklist. J Hand Surg 2001;26A:916 22. 12. Nana AD, Joshi A, Lichtman DM. Plating of the distal radius. J Am Acad Orthop Surg. 2005;13:159 71. 13. Nunley JA, Rowan PR. Delayed rupture of the flexor pollicis longus tendon after, inappropriate placement of the : plate on the volar surface of the distal radius. J Hand Surg 1999;24A:1279 80. 14. Orbay JL, Fernandez DL. Volar fixation of dorsally displaced fractures of the distal radius: a preliminary report. J Hand Surg 2002;27A:205 15. 15. Orbay JL, Fernandez DL. Volar fixed angle plate fixation for unstable distal radius, fractures in the elderly patient. J Hand Surg 2004;29A:96 102. 16. Rozental TD, Beredjiklian PK, Bozentka DJ. Functional outcomes and complications following two types of dorsal plating for unstable fractures of the distal part of the radius. J Bone Jt Surg Am 2003;85A:1956 60. 17. Smith DW, Henry MH. The 45 pronated oblique view for volar fixed-angle plating, of distal radius fractures. J Hand Surg 2004;29A:703 6. 18. Weiland AJ. External fixation, not ORIF, as the treatment of choice for fractures of the distal radius. J Orthop Trauma 1999;13 (8):570 2. 19. Yang Z, Mann FA, Gilula LA, Haerr C, Larsen CF. Scaphopisocapitate aligment: criterion to establish a neutral lateral view of the wrist. Radiology 1997;205:865 9.