TRACE ELEMENTS IN SERUM. Event #3, 2011

Similar documents
TRACE ELEMENTS IN SERUM

TRACE ELEMENTS IN SERUM. Event #2, 2010

TRACE ELEMENTS IN SERUM

TRACE ELEMENTS IN SERUM. Educational Event. Proficiency Test Report. Event #2, 2013

TRACE ELEMENTS IN SERUM

TRACE ELEMENTS IN SERUM

TRACE ELEMENTS IN URINE. Event #1, 2010

OFFICIAL METHODS FOR CHEMICAL ELEMENTS IN FOOD OF ANIMAL ORIGIN

STANDARD METHODS FOR CHEMICAL ELEMENTS IN FOOD OF ANIMAL ORIGIN

Multi Analyte Custom Grade Solution

Bottom Ash Data Week 38

Bottom Ash Data Week 49

Bottom Ash Data Week 30

Bottom Ash Data Week 8

DETERMINATION OF TRACE ELEMENTS IN WATERS AND WASTES BY INDUCTIVELY COUPLED PLASMA - MASS SPECTROMETRY EPA Method (Revision 5.

Bottom Ash Data Week 37

Bottom Ash Data Week 12

Bottom Ash Data Week 17

Bottom Ash Data Week 9

Bottom Ash Data Week 40

Bottom Ash Data Week 1

of human hair and nails. Part I. Analytical methodology, Sci. Tot. Environ. 2000, 250/1-3,

Analyte Proficiency From All Labs # Analytes: 26 Sample # Statistical Summary # Labs Reporting: 82 Urea Issue Date : 06/30/2016

CERTIFICATE OF ANALYSIS. tel: fax:

AAFCO Check Sample 2015 Minerals Program. A Targeted and Engineered Concentration Program

TOXIC AND ESSENTIAL ELEMENTS

Test Report No.T JP Date: FEB 22, 2017 Page 1 of 6

Product Stewardship Information Sheet CH350LN

Test Report No.T JP Date: FEB 24, 2017 Page 1 of 6

Product Stewardship Information Sheet CH200LN-02

Toxicology Blood Lead Comprehensive Testing

Toxicology Blood Lead Comprehensive Testing

MULTI-COMPONENT ANALYSIS OF HEAVY METALS

AAP Lead Testing Webinar Series

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

Actual Excipient Test Data on Metal Impurities Submitted to IPEC-Americas from Industry

Routine Analysis of Fortified Foods using the Agilent 7800 ICP-MS

Toxicology Blood Lead Comprehensive Testing

Test Report No.T TC Date: JUN 29, 2018 Page 1 of 5

!!"#$%&'!())*+,$-.&(.*/*+,)0&1-$2

Alkosel /Selsaf D08/FSQ/CVH/GS/(2006) D 421

Application Note novaa 800 D. Determination of Macro and Trace Minerals as well as Toxic Trace Metals in Powdered Milk. Challenge.

Landolt-Börnstein: Numerical Data and Functional Relationships in Science and Technology - New Series 12A. Ac-ag...Au-Zr. Bearbeitet von B Predel

POTENTIALLY g Head Brown Dye Normal V010.08

HAIR ELEMENTS. PATIENT: Number 74 SEX: Female AGE: 49. POTENTIALLY TOXIC ELEMENTS TOXIC RESULT REFERENCE PERCENTILE ELEMENTS µg/g RANGE 68 th 95 th

water & dialysate testing services

Edward V. Krizhanovsky, Ph.D., Kamila B. Tursunova

Matrix Interferences in ICP-MS: Causes, Effects, and Strategies to Reduce or Eliminate Them

Micro Trace Minerals Laboratory

CERTIFICATE OF ANALYSIS

!!"#$%&'!())*+,$-.&(.*/*+,)0&1-$2

HAIR ELEMENTS. DOCTOR S DATA, INC. ADDRESS: 3755 Illinois Avenue, St. Charles, IL CLIA ID NO: 14D MEDICARE PROVIDER NO:

PT-2017-NRL-TE-FASFC Determination of As, Asi, Cd, Pb, Cu and Zn in chocolate

Trace Elements in Manure

Test Report No.: GZHL IP Date: Sep 01, 2016 Page 1 of 12

HAIR ELEMENTS. COMMENTS: results checked

Chinese Zinc Sulfate Monohydrate testing. Dick Camp Kronos Micronutrients

TEST REPORT Job No./Report No TR RV1 Date: 25 June 2015 Page 1 of 10

Test Report No.: T TY Date: DEC 19, 2018 Page 1 of 10

Wolf Creek Community Alliance 2009 Soil Sampling Report By BJ Schmitt January 26, 2010

Magruder Check Sample: 2016 Year End Summary

Test Report No.: T TY Date: DEC 19, 2018 Page 1 of 12

SAMPLE. Control of Preanalytical Variation in Trace Element Determinations; Approved Guideline

CERTIFICATE OF COMPLIANCE SGS United Kingdom A UKAS Accredited Test House

Uncertainty from sample preparation in the laboratory on the example of various feeds

Appendix D: Aerobic and Anaerobic Leaching Tests

Figure 1. Location of 43 benchmark sites across Alberta.

A Water Proficiency Testing Scheme to ensure safe drinking Water in Africa

ESSENTIAL TRACE ELEMENTS PREPARED BY B.KIRUTHIGA LECTURER DEPT OF PHARMACEUTICAL CHEMISTRY

HAIR ELEMENTS. LAB#: H PATIENT: Franklin C. Cook ID: COOK-F SEX: Male AGE: 66

Summary p. 1 What Are Dietary Reference Intakes? p. 2 Approach for Setting Dietary Reference Intakes p. 7 Nutrient Functions and the Indicators Used

TEST REPORT. LAB LOCATION:TURKEY LAB NO. : (7216) SERVICE TYPE: Regular DATE IN: November 29 th, 2016 DATE OUT: December 01 th, 2016

Matrix Reference Materials - SCP SCIENCE

Kelp and Dairy Cattle - a Pilot Study

TEST REPORT Job No./Report No TR Date:24 June 2015 Page 1 of 14

!!"#$%&'&())*+,$-.&(.*/*+,)0&12".*&3.""4

NUTRIENT AND HEAVY METAL CONTENTS OF HOG MANURE - EFFECT ON SOIL QUALITY AND PRODUCTIVITY

Greg Patterson C.C.A. President A&L Canada Laboratories

Element concentrations in livers and kidneys of ranch mink

Analytical methods and laboratory choice for food composition, an introduction

LAB LOCATION: HONG KONG ISSUE DATE: JUN 05, 2017 REPORT NUMBER: PAGE:

Ca Mg Na K Cu Zn P Fe Mn Cr Se B Co Mo S Sb U As Be Hg Cd Pb Al

SAMPLE TEST Toxic Elements Analysis P P P Detection of Amines Derived From Azocolourants and Azodyes NA P P

Analyses no.: Telephone: Fax: Telephone: Fax:

Test Report. No. SHATY Date: 21 Mar Page 1 of 6.

Photochemical investigation of elements in Oryza sativa

(FAD ; CRL/100149) (FAD ; CRL/100303) (FAD ; CRL/100296)

Determination of available nutrients in soil using the Agilent 4200 MP-AES

!!"#$%&'!())*+,$-.&(.*/*+,)0&1-$2

Chemical Tests: How to understand your measurement method and your result

GUIDANCE FOR THE ACCREDITATION OF GENERIC METHODS & METHODS OF FLEXIBLE SCOPE OF DETERMINATION OF METALS IN FOODS

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Chemical Test: Understand your

AOAC Metals Subgroup. Chemical Contaminants and Residues in Food Community

V. LAB REPORT. PART I. ICP-AES (section IVA)

Determination of Copper in Green Olives using ICP-OES

EUROPEAN COMMISSION JOINT RESEARCH CENTRE Institute for Reference Materials and Measurements European Union Reference Laboratory for Feed Additives

Transcription:

TRACE ELEMENTS IN SERUM Event #, 0 November 5, 0

Event #, 0 Serum Aluminum The test materials for serum Al were prepared from human serum obtained from Tennessee Blood Services, Inc. The units were tested by FDA approved methods and found to be Non-reactive for Anti-HIV-/, Anti-HCV.0 and HBsAg. The serum has also been found to be STS (RPR) Non-reactive and Negative for HIV- and HCV by PCR. Serum units were dispensed into acid-washed 500-mL polypropylene containers to make up five (5) serum pools. Each pool was spiked with a suite of additional trace elements including aluminum as Al + at various concentrations. The Target Value assigned for each PT material is the robust mean of the results reported by all participants in this event. The robust statistics were obtained utilizing algorithms based on those presented in ISO 58:005E Statistical methods for use in proficiency testing by interlaboratory comparisons. Values for serum aluminum range from 9 µg/l (.0 µmol/l) to 00 µg/l (.4 µmol/l). Acceptable ranges for serum aluminum are based on fixed criteria of ±0%, or ±5 µg/l below 5 µg/l. These criteria are based on consensus recommendations from several EQAS organizers (). Discussion. Based on the above criteria, 9.% of test results reported were judged as satisfactory, with two out of 6 participant laboratories (.%) reporting or more of the 5 results outside the acceptable ranges.. Taylor, A., Angerer, J., Claeys, F., Kristiansen, J., Mazarrasa, O., Menditto, A., Patriarca, M., Pineau, A., Schoeters, I., Sykes, C., Valkonen, S. and Weykamp, C. Comparison of procedures for evaluating laboratory performance in external quality assessment schemes for lead in blood and aluminum in serum demonstrates the need for common quality specifications. Clinical Chemistry 00 48 000-00. Trace Elements PT Program www.wadsworth.org/testing/lead/index.htm

Serum Aluminum Test Results, 0 Event # ROBUST STATISTICAL SUMMARY TARGET VALUE ASSIGNMENT AND STATISTICS SE- SE- SE- SE-4 SE-5 Robust Mean Robust Standard Deviation Standard Uncertainty RSD (%) 9 0 5 00 88.9.4 6.. 4.8.0 0.8.5.8..6 4.9 4.6 5. 5.5 Acceptable Range: Upper Limit: 5 84 6 40 06 Lower Limit: 56 08 60 0 notes: Results reported as less than the method detection limit are excluded from statistical calculations.

Serum Aluminum Test Results, 0 Event # PERFORMANCE OF PARTICIPATING LABORATORIES Lab Code Method SE- SE- SE- SE-4 SE-5 Info Only Target Values: 9 0 5 00 88 0 ETAAS-Z 0 4 48 8 9 4 ETAAS-Z >00 >00 89 4 ETAAS-Z 9 68 9 89 Info 56 ICP-MS 5 0 40 88 59 ETAAS-Z 4 8 9 60 ETAAS-Z 69 9 96 88 64 ICP-MS 6 66 9 85 9 DRC/CC-ICP-MS 8 5 0 9 ICP-MS 9 8 9 09 00 DRC/CC-ICP-MS 40 8 5 90 6 Info 06 ICP-MS 5 6 6 9 88 8 ETAAS-Z 4 66 8 9 80 9 ICP-MS 99 89 Info 05 ICP-MS 8 0 00 89 4 HR-ICP-MS 45 5 89 Info 5 ETAAS-Z 4 59 84 4 Info 55 ICP-MS 9 0 9 5 9 5 ICP-MS 9 5 05 9 58 ICP-MS 6 65 00 85 6 ICP-MS 4 00 8 6 ICP-MS 8 68 4 96 84 66 ETAAS-Z 0 90 9 Info 6 ETAAS-Z 9 0 4 9 85 Info 40 ICP-AES/OES 6 6 05 9 Info 458 ETAAS Other 5 6 4 05 6 465 ICP-MS 68 9 04 8 Percent satisfactory results for all participants: 9. % notes: reported outside upper limit Info only: results included for informational purposes only. reported outside lower limit notes: Results reported as less than the method detection limit are excluded from statistical calculations.

DRC/CC-ICP-MS New York State Department of Health Serum Aluminum Test Results, 0 Event # STATISTICAL SUMMARY BY METHOD ETAAS Other ETAAS-Z HR-ICP-MS ICP-AES/OES ICP-MS All Laboratories SE- SE- SE- SE-4 SE-5 6 6 5 9 8 6. 9 9.9 6 9 5 6.9 80 6 9 66 5.4 6 69 4.8 6 68 0 4.0 9 9 4 8 8 8. 45 6 4 4. 5.6 0 05 8 86 4. 5 05 0 8 4. 5 94 6.4 8 8 6 9 84 6 9. 89 9 89. 6 86 4. notes: Insufficient data for calculation.

Event #, 0 Serum Copper The test materials for serum Cu were prepared from human serum obtained from Tennessee Blood Services, Inc. The units were tested by FDA approved methods and found to be Non-reactive for Anti-HIV-/, Anti-HCV.0 and HBsAg. The serum has also been found to be STS (RPR) Non-reactive and Negative for HIV- and HCV by PCR. Serum units were dispensed into acid-washed 500-mL polypropylene containers to make up five (5) serum pools. Each pool was spiked with a suite of additional trace elements including copper as Cu + at various concentrations. The Target Value assigned for each PT material is the robust mean of the results reported by all participants in this event. The robust statistics were obtained utilizing algorithms based on those presented in ISO 58:005E Statistical methods for use in proficiency testing by interlaboratory comparisons. Values for serum copper range from 956 µg/l (5.04 µmol/l) to 45 µg/l (8.59 µmol/l). Acceptable ranges for serum copper are based on fixed criteria of ±5%, or ±95 µg/l below 65 µg/l. These criteria are consistent with those proposed by the OELM Network of EQAS organizers (, ) for trace elements in serum, and are slightly less stringent than those previously suggested for NYS (±0%). Discussion. Based on the above criteria, 9.4% of test results reported were judged as satisfactory, with two out of participant laboratories (9.5%) reporting or more of the 5 results outside the acceptable ranges.. A. Taylor, J. Angerer, J. Arnaud, F. Claeys, R.L. Jones, O. Mazarrasa, E. Mairiaux, A. Menditto, P.J. Parsons, M. Patriarca, A. Pineau, S. Valkonen, J.-P. Weber and C. Weykamp Accreditation and Quality Assurance 006 440-445.. J. Arnaud, J.-P. Weber, C.W. Weykamp, P.J. Parsons, J. Angerer, E. Mairiaux, O. Mazarrasa, S. Valkonen, A. Menditto, M. Patriarca, and A. Taylor Clinical Chemistry 008 54 89-899. Trace Elements PT Program www.wadsworth.org/testing/lead/index.htm

Serum Copper Test Results, 0 Event # ROBUST STATISTICAL SUMMARY TARGET VALUE ASSIGNMENT AND STATISTICS SE- SE- SE- SE-4 SE-5 Robust Mean Robust Standard Deviation Standard Uncertainty RSD (%) 45 5 956 54 44 55 04 4 6 00 4 8 8 6..9 4. 5..0 Acceptable Range: Upper Limit: 80 5 099 44 649 Lower Limit: 084 8 8 066 9 notes: Results reported as less than the method detection limit are excluded from statistical calculations.

Serum Copper Test Results, 0 Event # PERFORMANCE OF PARTICIPATING LABORATORIES Lab Code Method SE- SE- SE- SE-4 SE-5 Info Only Target Values: 45 5 956 54 44 0 DRC/CC-ICP-MS 86 549 46 490 659 Info 0 ICP-MS 6 40 0 5 5 4 ICP-MS 0 90 880 40 00 4 ICP-MS 5 60 9 96 49 Info 56 FAAS 460 90 000 90 50 59 ICP-AES/OES 40 60 940 0 90 60 ETAAS-Z 560 0 90 0 440 64 ICP-MS 6 45 9 60 9 DRC/CC-ICP-MS 50 50 990 00 480 9 ICP-MS 40 00 940 40 40 00 ICP-MS 504 9 95 84 46 Info 06 ICP-MS 40 50 90 80 0 9 ICP-MS 45 4 9 5 46 Info 05 ICP-MS 465 948 9 4 HR-ICP-MS 46 9 965 5 44 Info 5 FAAS 80 50 0 50 650 Info 60 FAAS 00 50 90 00 40 6 ICP-MS 40 00 90 60 440 66 ETAAS other 00 4 905 6 68 Info 40 DRC/CC-ICP-MS 8 85 4 59 Info 45 ICP-AES/OES 6 44 059 8 588 Info Percent satisfactory results for all participants: 9.4 % notes: reported outside upper limit Info only: results included for informational purposes only. reported outside lower limit notes: Results reported as less than the method detection limit are excluded from statistical calculations.

DRC/CC-ICP-MS New York State Department of Health Serum Copper Test Results, 0 Event # STATISTICAL SUMMARY BY METHOD ETAAS other ETAAS-Z FAAS HR-ICP-MS ICP-AES/OES ICP-MS All Laboratories SE- SE- SE- SE-4 SE-5 556 6 00 560 5 66 46 50 5 0 48 4 4. 465 66 6. 46 06 4 0 90 40 9 5 8 0 09 5.6 0 8. 996 4 905 90 04 65 965 000 84 0 95 9 4. 94 86 8.8 4 6 0 40 5 96 0 4 6 5.0 99. 466 00 68 440 500 55 44 489 40 0 4 68 4.8 440 05. notes: Insufficient data for calculation.

Event #, 0 Serum Selenium The test materials for serum Se were prepared from human serum obtained from Tennessee Blood Services, Inc. The units were tested by FDA approved methods and found to be Non-reactive for Anti-HIV-/, Anti-HCV.0 and HBsAg. The serum has also been found to be STS (RPR) Non-reactive and Negative for HIV- and HCV by PCR. Serum units were dispensed into acid-washed 500-mL polypropylene containers to make up five (5) serum pools. Each pool was spiked with a suite of additional trace elements including selenium as Se 4+ at various concentrations. The Target Value assigned for each PT material is the robust mean of the results reported by all participants in this event. The robust statistics were obtained utilizing algorithms based on those presented in ISO 58:005E Statistical methods for use in proficiency testing by interlaboratory comparisons. Values for serum selenium range from 0 µg/l (.5 µmol/l) to 64 µg/l (.4 µmol/l). Acceptable ranges for serum selenium are based on fixed criteria of ±0%, or ± µg/l below 0 µg/l. These criteria are a little less stringent than those proposed by the OELM Network of EQAS organizers (±5% or ±8 µg/l below 55 µg/l) (, ) for trace elements in serum. As performance for serum Se improves among NYS-permit laboratories, consideration will be given to adopting the OELM criteria. Discussion. Based on the above criteria, 00% of test results reported were judged as satisfactory, with none of the participant laboratories reporting or more of the 5 results outside the acceptable ranges.. A. Taylor, J. Angerer, J. Arnaud, F. Claeys, R.L. Jones, O. Mazarrasa, E. Mairiaux, A. Menditto, P.J. Parsons, M. Patriarca, A. Pineau, S. Valkonen, J.-P. Weber and C. Weykamp Accreditation and Quality Assurance 006 440-445.. J. Arnaud, J.-P. Weber, C.W. Weykamp, P.J. Parsons, J. Angerer, E. Mairiaux, O. Mazarrasa, S. Valkonen, A. Menditto, M. Patriarca, and A. Taylor Clinical Chemistry 008 54 89-899. Trace Elements PT Program www.wadsworth.org/testing/lead/index.htm

Serum Selenium Test Results, 0 Event # ROBUST STATISTICAL SUMMARY TARGET VALUE ASSIGNMENT AND STATISTICS SE- SE- SE- SE-4 SE-5 Robust Mean Robust Standard Deviation Standard Uncertainty RSD (%) 8 90 64 0 5.0 8.5 8.0.0 4..5.6.4...8.6 4. 4..4 Acceptable Range: Upper Limit: 5 86 8 44 Lower Limit: 05 90 5 96 notes: Results reported as less than the method detection limit are excluded from statistical calculations.

Serum Selenium Test Results, 0 Event # PERFORMANCE OF PARTICIPATING LABORATORIES Lab Code Method SE- SE- SE- SE-4 SE-5 Info Only Target Values: 8 90 64 0 0 DRC/CC-ICP-MS 0 6 4 0 Info 0 DRC/CC-ICP-MS 5 8 59 8 4 ICP-MS 6 49 95 69 6 4 ICP-MS 4 5 99 6 Info 56 ICP-MS 8 9 80 54 6 59 ETAAS-Z 9 84 48 64 ICP-MS 5 4 94 6 9 9 DRC/CC-ICP-MS 44 99 5 4 9 ICP-MS 9 4 9 6 00 DRC/CC-ICP-MS 4 8 89 Info 06 ICP-MS 4 8 55 9 9 DRC/CC-ICP-MS 5 46 94 6 Info 05 ICP-MS 4 8 5 0 4 HR-ICP-MS 9 40 96 68 Info 66 ETAAS-Z 5 44 0 8 Info 6 DRC/CC-ICP-MS 5 4 9 69 0 Info 40 DRC/CC-ICP-MS 4 6 89 68 6 Info Percent satisfactory results for all participants: 00.0 % notes: reported outside upper limit Info only: results included for informational purposes only. reported outside lower limit notes: Results reported as less than the method detection limit are excluded from statistical calculations.

DRC/CC-ICP-MS New York State Department of Health Serum Selenium Test Results, 0 Event # STATISTICAL SUMMARY BY METHOD ETAAS-Z HR-ICP-MS ICP-MS All Laboratories SE- SE- SE- SE-4 SE-5 5. 9 5 4. 6 4. 6 5. 6 40 8 9.6 0 4. 88 0 5.4 9 8 96 89 8 4.0 90 0 5.0 6 4.6 65 68 60 5.0 64 4.0 8 8 6.4 5. 5.9 notes: Insufficient data for calculation.

Event #, 0 Serum Zinc The test materials for serum Zn were prepared from human serum obtained from Tennessee Blood Services, Inc. The units were tested by FDA approved methods and found to be Non-reactive for Anti-HIV-/, Anti-HCV.0 and HBsAg. The serum has also been found to be STS (RPR) Non-reactive and Negative for HIV- and HCV by PCR. Serum units were dispensed into acid-washed 500-mL polypropylene containers to make up five (5) serum pools. Each pool was spiked with a suite of additional trace elements including zinc as Zn + at various concentrations. The Target Value assigned for each PT material is the robust mean of the results reported by all participants in this event. The robust statistics were obtained utilizing algorithms based on those presented in ISO 58:005E Statistical methods for use in proficiency testing by interlaboratory comparisons. Values for serum zinc range from 606 µg/l (9. µmol/l) to 5 µg/l (.8 µmol/l). Acceptable ranges for serum zinc are based on fixed criteria of ±5%, or ±5 µg/l below 00 µg/l. These criteria are consistent with those proposed by the OELM network of EQAS organizers () for trace elements in serum. Discussion. Based on the above criteria, 95.6% of test results reported were judged as satisfactory, with one out of participant laboratories (.%) reporting or more of the 5 results outside the acceptable ranges.. A. Taylor, J. Angerer, J. Arnaud, F. Claeys, R.L. Jones, O. Mazarrasa, E. Mairiaux, A. Menditto, P.J. Parsons, M. Patriarca, A. Pineau, S. Valkonen, J.-P. Weber and C. Weykamp Accreditation and Quality Assurance 006 440-445.. J. Arnaud, J.-P. Weber, C.W. Weykamp, P.J. Parsons, J. Angerer, E. Mairiaux, O. Mazarrasa, S. Valkonen, A. Menditto, M. Patriarca, and A. Taylor Clinical Chemistry 008 54 89-899. Trace Elements PT Program www.wadsworth.org/testing/lead/index.htm

Serum Zinc Test Results, 0 Event # ROBUST STATISTICAL SUMMARY TARGET VALUE ASSIGNMENT AND STATISTICS SE- SE- SE- SE-4 SE-5 Robust Mean Robust Standard Deviation Standard Uncertainty RSD (%) 606 64 5 884 66 9 89 48 9 8 9 6.4 6.5 5.0 5.5 4.8 Acceptable Range: Upper Limit: 69 569 50 0 88 Lower Limit: 55 59 94 5 65 notes: Results reported as less than the method detection limit are excluded from statistical calculations.

Serum Zinc Test Results, 0 Event # PERFORMANCE OF PARTICIPATING LABORATORIES Lab Code Method SE- SE- SE- SE-4 SE-5 Info Only Target Values: 606 64 5 884 66 0 DRC/CC-ICP-MS 646 49 66 9 86 Info 0 ICP-MS 64 448 6 96 8 4 ICP-MS 50 60 490 80 0 4 ICP-MS 580 5 59 88 58 Info 56 ICP-MS 566 86 40 80 699 59 ICP-AES/OES 590 0 490 860 60 60 FAAS 60 40 560 90 90 64 ICP-MS 560 6 450 84 9 DRC/CC-ICP-MS 640 460 600 940 80 9 ICP-MS 550 50 50 850 90 00 ICP-MS 56 4 49 80 9 Info 06 ICP-MS 640 40 490 880 60 8 FAAS 50 50 450 850 00 9 ICP-MS 595 80 5 889 Info 05 ICP-MS 58 98 90 8 00 4 HR-ICP-MS 606 4 5 88 Info 5 FAAS 00 040 0 500 0 Info 55 ICP-MS 65 9 50 94 88 5 ICP-MS 60 0 50 880 50 58 ICP-MS 64 44 608 99 89 60 FAAS 560 0 490 840 0 6 ICP-MS 600 90 40 90 690 6 ICP-MS 650 40 50 90 80 66 FAAS 604 8 466 88 5 Info 40 DRC/CC-ICP-MS 588 68 4 90 06 Info 45 ICP-AES/OES 585 4 60 900 8 Info 458 FAAS 60 40 60 900 90 Percent satisfactory results for all participants: 95.6 % notes: reported outside upper limit Info only: results included for informational purposes only. reported outside lower limit notes: Results reported as less than the method detection limit are excluded from statistical calculations.

DRC/CC-ICP-MS New York State Department of Health Serum Zinc Test Results, 0 Event # STATISTICAL SUMMARY BY METHOD FAAS HR-ICP-MS ICP-AES/OES ICP-MS All Laboratories SE- SE- SE- SE-4 SE-5 65 6 66 5 6. 606 588 4 5 66 46.9 6 4.0 40 6 488 8.6 4 64 6 5 9 8 6. 8 54. 565 9 6 68 58 6.0 5 546 8 5 50 4.8 58 9 9.0 940 8 6 980 56 6. 88 880 8 5 8 48 5.5 904 4. 6 6 89 0.6 69 5 5 4 5.6 4 88.4 notes: Insufficient data for calculation.

Event #, 0 Additional Trace Elements Reported in Serum Participant laboratories reported their analytical results for any additional trace elements (other than Al, Cu, Se and Zn) that are routinely reported so that a more complete characterization can be recorded for these PT materials. Results for additional trace elements are reported here, but no target value is implied nor are any acceptable ranges provided. These data are provided solely for educational and informational purposes. In addition to Al, Cu, Se and Zn, the serum pools were supplemented with additional trace elements as indicated below. Additional Elements Mn, Cr, V, Tl, Co Trace Elements PT Program www.wadsworth.org/testing/lead/index.htm

Serum Additional Elements, 0Event # Page Serum Antimony (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS <0.058 <0.058 <0.058 <0.058 <0.058 Serum Arsenic (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS 0.54 0.86.48 0.44 0.95 9 DRC/CC-ICP-MS <0 <0 <0 <0 <0 Serum Barium (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS 0.9.9.8.44.09 9 ICP-MS <.0 <.0 <.0 <.0 <.0 Serum Beryllium (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS <.4 <.4 <.4 <.4 <.4 9 ICP-MS <0. <0. <0. <0. <0. Serum Bismuth (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS 0.5 0.09 0.05 0.056 0.05 Serum Cadmium (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS <0.04 <0.04 <0.04 <0.04 <0.04 9 DRC/CC-ICP-MS <0.5 <0.5 <0.5 <0.5 <0.5 Serum Chromium (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS 0.8.90 0.64 4..5 64 ICP-MS 0.9. 0.8 4..5 9 DRC/CC-ICP-MS 0.8.9 0. 4.4.4 9 ICP-MS... 4..8 05 ICP-MS 0.8. 0.6 4.5.4 Arithmetic mean (n=5) 0.9. 0.8 4.5.5 SD 0. 0. 0. 0. 0. Serum Cobalt (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS 0..64 0. 0.9.48 9 ICP-MS 0.8.8 0. 0.5. 9 ICP-MS.0.0 <.0 <.0. Arithmetic mean (n=) 0.8.8 0. 0.4.5 SD 0. 0. 0. 0. 0. Serum Iodine (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS 5.899 55.06 5.59 55.696 9.8

Serum Additional Elements, 0Event # Page Serum Lead (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS <4.44 <4.44 <4.44 <4.44 <4.44 9 DRC/CC-ICP-MS <0.4 <0.4 <0.4 <0.4 <0.4 Serum Lithium (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS.409 0.6 0.84.46 0.49 Serum Manganese (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 0 DRC/CC-ICP-MS 0.0.8.0. 4.60 4 ICP-MS 0.94.4 6.9.8 4.9 9 DRC/CC-ICP-MS 0.8.4 6.8. 4.6 9 ICP-MS.4. 8.0. 5.4 05 ICP-MS.0.5 6..4 4.8 Arithmetic mean (n=5).0.5..4 4.8 SD 0. 0. 0.5 0.4 0.4 Serum Molybdenum (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS.956 0.54.86 0.9 0.6 Serum Nickel (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS 0.44 0.55 0.44 <0.05 <0.05 Serum Silver (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS 0.6 0.6 0.8 0.9 0. Serum Tin (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS <0.50 <0.50 <0.50 <0.50 <0.50 9 ICP-MS <5.0 <5.0 <5.0 <5.0 <5.0 Serum Tellurium (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS <0.4 <0.4 <0.4 <0.4 <0.4 9 ICP-MS <.0 <.0 <.0 <.0 <.0 Serum Thallium (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS 0.5.6 0.6 4.456.04 9 ICP-MS <.0.9 <.0 4.. Serum Uranium (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 4 ICP-MS <0.06 <0.06 <0.06 <0.06 <0.06 Serum Vanadium (µg/l) Lab Code Method SE- SE- SE- SE-4 SE-5 9 DRC/CC-ICP-MS..5.5.8 0.6

ATOMIC SPECTROMETRY METHODS New York State Department of Health Trace Elements in Serum METHOD NOTES A- ETAAS-Z (Electrothermal atomic absorption spectrometry with Zeeman background correction) A- ETAAS other (i.e., D, S-H background correction) A- FAAS (Flame atomic absorption spectrometry) A-4 CV-AAS (Cold vapor atomic absorption spectrometry) A-5 HG-AAS (Hydride generation atomic absorption spectrometry) A-6 AFS (Atomic fluorescence spectrometry) A- Other INDUCTIVELY COUPLED PLASMA P- ICP-MS (Inductively coupled plasma - mass spectrometry) P- DRC/CC-ICP-MS (ICP-MS used in the Dynamic Reaction Cell or Collision Cell mode) P- ICP-AES/OES (ICP atomic/optical emission spectrometry) P-4 HR-ICP-MS (High resolution ICP-MS) P-5 ETV-ICP-MS (Electrothermal vaporization ICP-MS) P-6 ID-ICP-MS (Isotope dilution ICP-MS) P- Other ELECTROCHEMICAL METHODS E- ASV (Anodic stripping voltammetry without digestion) E- ASV-LeadCare (Anodic stripping voltammetry using the ESA LeadCare system) E- Fluoride specific electrode E-4 Other MOLECULAR FLUORIMETRY F- EtOAc (Ethyl acetate-acetic acid extraction method for determination of erythrocyte protoporphyrin) F- Aviv hematofluorometry (for determination of EP at hematocrit 5) F- Helena ZPP (for determination of zinc protoporphyrin in µmol ZPP/mol heme) F-4 Other OTHER METHODS If your method is not listed in the above list, please describe it briefly.