Validation of UV Spectrophotometric and HPLC Methods for Quantitative determination of Iloperidone in Pharmaceutical Dosage Form

Similar documents
METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE

Development and Validation for Simultaneous Estimation of Sitagliptin and Metformin in Pharmaceutical Dosage Form using RP-HPLC Method

RP-HPLC Method Development and Validation of Abacavir Sulphate in Bulk and Tablet Dosage Form

IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page:

Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page:

RP-HPLC Analysis of Temozolomide in Pharmaceutical Dosage Forms

Pelagia Research Library

DEVELOPMENT AND VALIDATION OF NEW HPLC METHOD FOR THE ESTIMATION OF PALIPERIDONE IN PHARMACEUTICAL DOSAGE FORMS

Journal of Chemical and Pharmaceutical Research, 2013, 5(1): Research Article

J Pharm Sci Bioscientific Res (4): ISSN NO

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Available online at Scholars Research Library

Pelagia Research Library

Development, Estimation and Validation of Lisinopril in Bulk and its Pharmaceutical Formulation by HPLC Method

International Journal of Medicine and Pharmaceutical Research. International Journal of Medicine and Pharmaceutical Research


Scholars Research Library. Der Pharmacia Lettre, 2015, 7 (5):44-49 (

Journal of Chemical and Pharmaceutical Research

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD ESTIMATION OF TOLVAPTAN IN BULK PHARMACEUTICAL FORMULATION

International Journal of Pharma and Bio Sciences DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR THE ESTIMATION OF STRONTIUM RANELATE IN SACHET

Simultaneous Estimation of Gemcitabine Hydrochloride and Capecitabine Hydrochloride in Combined Tablet Dosage Form by RP-HPLC Method

World Journal of Pharmaceutical Research

Available online Research Article

MEDAK DIST. ANDHRA PRADESH STATE, INDIA. Research Article RECEIVED ON ACCEPTED ON

ISSN (Print)

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

New RP-HPLC Method for the Determination of Fludarabine in Pharmaceutical Dosage Forms

Reverse Phase HPLC Analysis of Atomoxetine in Pharmaceutical Dosage Forms

A New Stability-Indicating and Validated RP-HPLC Method for the Estimation of Liraglutide in Bulk and Pharmaceutical Dosage Forms

Scholars Research Library

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR ESTIMATION OF LACOSAMIDE IN BULK AND ITS PHARMACEUTICAL FORMULATION

International Journal of Pharma and Bio Sciences

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS TOLBUTAMIDE IN PURE AND PHARMACEUTICAL FORMULATIONS

36 J App Pharm Vol. 6; Issue 1: 36-42; January, 2014 Rao et al., 2014

VALIDATED RP-HPLC METHOD FOR SIMULTANEOUS ESTIMATION OF DEXAMETHASONE AND GRANISETRON IN COMBINED DOSAGE FORMS

CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG

A HIGH PERFORMANCE LIQUID CHROMATOGRAPHIC ASSAY FOR LERCANIDIPINE HYDROCHLORIDE

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR SIMULTANEOUS ESTIMATION OF ETORICOXIB AND THIOCOLCHICOSIDE IN PHARMACEUTICAL DOSAGE FORMS

Simultaneous estimation of Metformin HCl and Sitagliptin in drug substance and drug products by RP-HPLC method

Development and Validation of a New Uv Method for the Analysis of Rebamipide

SIMULTANEOUS ESTIMATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN TABLETS BY RP-HPLC METHOD

ANALYTICAL TECHNIQUES FOR THE QUALITY OF ANTI- AIDS DRUGS

Journal of Chemical and Pharmaceutical Research, 2017, 9(9): Research Article

Validated UV Spectrophotometric Method Development And Stability Studies Of Acamprosate Calcium In Bulk And Tablet Dosage Form

Method Development and Validation for the Estimation of Saroglitazar in Bulk and Pharmaceutical Dosage Form by RP-HPLC

Tentu Nageswara Rao et al. / Int. Res J Pharm. App Sci., 2012; 2(4): 35-40

Corresponding Author:

Development of a Validated RP-HPLC Method for the Analysis of Citicoline Sodium in Pharmaceutical Dosage Form using Internal Standard Method

Research Article Derivative Spectrophotometric Method for Estimation of Metformin Hydrochloride in Bulk Drug and Dosage Form

Validated Spectrophotometric Method for Simultaneous Estimation of Atorvastatin and Nicotinic acid in Combined Pharmaceutical dosage form

IJRPC 2011, 1(4) Rohan et al. ISSN: INTERNATIONAL JOURNAL OF RESEARCH IN PHARMACY AND CHEMISTRY

Development and validation of related substances method for Varenicline and its impurities

Development and validation of RP-HPLC method for simultaneous estimation of gliclazide and metformin in pure and tablet dosage form

RP- HPLC and Visible Spectrophotometric methods for the Estimation of Meropenem in Pure and Pharmaceutical Formulations

S. G. Talele, D. V. Derle. Department of Pharmaceutics, N.D.M.V.P. College of Pharmacy, Nashik, Maharashtra, India

Pankti M. Shah et al, Asian Journal of Pharmaceutical Technology & Innovation, 04 (17); 2016; 07-16

Scholars Research Library. Der Pharmacia Lettre, 2016, 8 (6): (

210 J App Pharm Vol. 6; Issue 2: ; April, 2014 Mustafa et al, 2014

A stability indicating RP-HPLC method for simultaneous estimation of darunavir and cobicistat in bulk and tablet dosage form

Development and validation of UV spectrophotometric estimation of lisinopril dihydrate in bulk and tablet dosage form using area under curve method

Research Article. ISSN Available online at 746

Journal of Pharmacreations

ISSN: ; CODEN ECJHAO E-Journal of Chemistry 2011, 8(3),

Stability indicating RP-HPLC method development and validation of Etizolam and Propranolol hydrochloride in pharmaceutical dosage form

Available online Research Article

UV Spectrophotometric Estimation of Alprazolam by Area Under Curve And First Order Derivative Methods in Bulk and Pharmaceutical Dosage Form

Scholars Research Library

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Development and Validation of RP-HPLC Method for the Estimation of Gemigliptin

Research and Reviews: Journal of Pharmacy and Pharmaceutical Sciences

Volume 2 (6), 2014, Page CODEN (USA)-IJPRUR, e-issn: International Journal of Pharma Research and Health Sciences

PHARMA SCIENCE MONITOR AN INTERNATIONAL JOURNAL OF PHARMACEUTICAL SCIENCES

DEVELOPMENT OF RP-HPLC METHOD FOR ESTIMATION OF DROTAVERINE HYDROCHLORIDE IN PHARMACEUTICAL FORMULATIONS

Development and validation of UV and RP-HPLC method for estimation of Nepafenac in bulk drug and opthalmic formulation

Simultaneous UV Spectrophotometric Method For Estimation Of Ebastine And Montelukast Sodium In Tablet Dosage Form By Q-Ratio Method

Sanjog Ramdharane 1, Dr. Vinay Gaitonde 2

Ankit et al Journal of Drug Delivery & Therapeutics; 2013, 3(2), Available online at RESEARCH ARTICLE

Development and validation of UV-visible spectrophotometric method for estimation of rifapentine in bulk and dosage form

Spectrophotometric Method for Estimation of Sitagliptin Phosphate in Bulk...

Estimation of zolmitriptan by a new RP-HPLC method

International Journal of Advances in Pharmacy and Biotechnology Vol.3, Issue-1, 2017, 1-7 Research Article Open Access.

RP-HPLC METHOD DEVELOPMENT AND VALIDATION FOR CILOSTAZOL IN TABLET DOSAGE FORM

Comparative study of analytical method development of fluconazole in tablets and capsule by ultraviolet spectrophotometric method

Available Online through Research Article

Hyderabad, India. Department of Pharmaceutical Chemistry, Glocal University, Saharanpur, India.

pharmaceutical formulations. Anagliptin shows absorption maximum at 246 nm and obeys beer s law in the

REVERSE PHASE HPLC METHOD FOR THE ANALYSIS OF ALFUZOSIN HYDROCHLORIDE IN PHARMACEUTICAL DOSAGE FORMS

RP-HPLC METHOD DEVELOPMENT AND VALIDATION FOR THE ESTIMATION OF BACLOFEN IN BULK AND PHARMACEUTICAL DOSAGE FORMS

Scholars Research Library

Validation of Developed Analytical Method for Balofloxacin Floating Tablets by Reverse Phase High Performance Liquid Chromatography

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

Airo International Research Journal ISSN : Volume : 7 October 2015

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF TRAMADOL IN PURE AND PHARMACEUTICAL FORMULATIONS

Simple and stability indicating RP-HPLC assay method development and validation lisinopril dihydrate by RP-HPLC in bulk and dosage form

Dr. Dilip G. Maheshwariet al, Asian Journal of Pharmaceutical Technology & Innovation, 04 (17); 2016; Research Article

438 Full Text Available On

International Journal of Drug Research and Technology

Journal of Chemical and Pharmaceutical Research

SIMULTANEOUS DETERMINATION OF ATORVASTATIN AND EZETIMIBE BY RP-HPLC IN PURE AND PHARMACEUTICAL DOSAGE FORM

Transcription:

International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.4, No.2, pp 576-581, April-June 2012 Validation of UV Spectrophotometric and HPLC Methods for Quantitative determination of Iloperidone in Pharmaceutical Dosage Form A. S. Manjula Devi* and T. K. Ravi Department of Pharmaceutical Analysis, College of Pharmacy, SRIPMS, Coimbatore-641 044,India. *Corres.author: hari1509@gmail.com Phone: 9443776673 Abstract: A specific and sensitive RP-HPLC method with PDA detection and UV spectrophotometric method were developed for iloperidone in tablet dosage form. Chromatographic separation was achieved on a LiChrospher RP-18 HPLC column (5 μ particle size and 25 cm 4.6 mm internal diameter) using 0.1% trifluoroacetic acid:acetonitrile in the ratio of 50:50 v/v (ph 5.02) as mobile phase and paracetamol as the internal standard. The effluent was monitored at 275 nm. Two sharp peaks were obtained for internal standard and iloperidone at 2.8 and 7.6 min, respectively. UV spectrophotometric method was performed at 229 nm using methanol as the solvent. Linear range was 1-10 µg ml -1 (r 2 =0.9986±0.0009) for HPLC method and 2-20 μg -1 (r 2 =0.9988±0.0004) for UV spectrophotometric method. Validation as per ICH guidelines and statistical analysis showed that both the methods were precise, accurate, sensitive, and can be used for the routine quality control of iloperidone in pharmaceutical dosage forms. Key words: Iloperidone, RP-HPLC, UV-spectrophotometric, internal standard, ICH guidelines. INTRODUCTION Iloperidone is an antipsychotic drug used in the treatment of schizophrenia. Chemically it is designated as 1-[4-[3-[4-(6-Fluoro-1, 2-benzisoxazol-3-yl)-1- piperidinyl]propoxy]-3-methoxyphenyl]ethanone with a molecular formula C 24 H 27 FN 2 O 4 and molecular weight 426.48 1. A thorough literature survey revealed LC-MS method for the determination of iloperidone in human plasma by Mutlib and Strupczewski 2 and application of LC-MS to identify the human liver cytochrome P450 isoforms involved in the metabolism of iloperidone by Mutlib and Klein 3. Mandpe and Varsha has reported stress degradation behaviour of iloperidone and a stability indicating assay method by RP-HPLC 4. However, RP-HPLC method employing internal standard and UV spectrophotometric method for the quantification of iloperidone in pharmaceutical dosage form has not yet reported in the literature. Internal standard methods are preferred for quality control of drugs since it overcomes errors from sample preparation and run to run variations, thus improving the accuracy and precision of the method. Moreover, these methods are highly sensitive and specific 5. UV spectrophotometric method is rapid, very simple, economic, and it allows the quantification in pharmaceuticals with satisfactory reliability 6.

A.S. Manjula Devi et al /Int.J. PharmTech Res.2012,4(2) 577 Therefore the aim of the present study was development and validation of RP-HPLC method using a suitable internal standard and UVspectrophotometric method for the estimation of iloperidone from its tablet dosage form. MATERIALS AND METHODS Analytically pure sample of iloperidone and paracetamole (assigned purity >99%purity) were obtained as gift samples from Sun Pharmaceuticals Ltd, Chennai. HPLC grade methanol was obtained from Merck specialties Pvt. Ltd. (Mumbai, India). Water for RP-HPLC was prepared using Millipore purification system (Milli-Q, Bangalore, India) and double distilled water was used to prepare all the solutions for UV spectrophotometric experiments. Tablet formulation (Ilosure, Sun Pharmaceuticals Ltd, Chennai) containing 2 mg iloperidone was purchased from the local pharmacy. Instrumentation and analytical conditions Chromatography was performed under ambient conditions with liquid chromatograph mass spectrometer (Shimadzu LCMS-2010EV) equipment comprising of a binary gradient pump (LC-20AD), degasser (DGU-20A 3 ), and a variable wavelength programmable PDA detector (SPD-M20A) with auto sampler system (SIL-20AC). The instrumentation was controlled by Shimadzu LCMS Solution software. Chromatographic separation was achieved on a LiChrospher RP-18 HPLC column (5 μ particle size and 25 cm 4.6 mm internal diameter) using the mobile phase composed of 0.1% trifluoroacetic acid : acetonitrile in the ratio of 50:50 v/v (ph 5.02; adjusted with 1% triethylamine) at a flow rate of 1 ml/min. The mobile phase was filtered through 0.45 μ nylon filter and degassed by sonication prior to use. Column temperature was maintained at ambient and the run time was set at 10 min. The column was equilibrated for 30-40 min with mobile phase prior to injection of the analyte. The volume of injection was 20µL. Detection was performed at 275 nm since both the components showed reasonable absorbance at this wavelength. The UV spectrophotometric method was performed using a Jasco V-630 UV/VIS Spectrophotometer with Spectra Manager (Version 2) controlling software at 229 nm in 1 cm quartz cells. The samples were prepared using methanol. Preparation of standard solutions For HPLC method, 100µg ml -1 of iloperidone and paracetamol were prepared separately by dissolving 10 mg each in 100 ml methanol. Working standard solutions of iloperidone (1-10 µg ml -1 ) were prepared by transferring 0.1 to 1 ml of iloperidone stock solution to serially arranged 10 ml standard flasks and diluting to volume using the mobile phase. A constant volume (0.2 ml) of stock solution containing paracetamol was added as the internal standard before making the dilutions. For the UV spectrophotometric method, stock solution of 100 μg ml -1 of iloperidone was prepared in methanol. The working standard solutions were prepared by dilution of the stock solution with methanol to get a concentration range 2-20 μg ml -1. The absorbance was measured at 229 nm using methanol as blank. Preparation of the sample solutions HPLC method Twenty tablets (Ilosure) were weighed, powdered and a powder mass equivalent to 2 mg of iloperidone was transferred to a 50 ml volumetric flask containing methanol. The mixture was then sonicated for 20 min to dissolve the material completely and centrifuged at 3,000 rpm for 5 min. An aliquot of supernatant solution (1.25 ml) was transferred to 10 ml volumetric flask, fixed concentration of internal standard (0.2 ml) was added and diluted to 10 ml using the mobile phase to get a solution of 5μg ml -1 of iloperidone and 2 μg ml -1 of IS. The resulting solution was analyzed by the proposed method. All determinations were carried out in six replicates. UV Spectrophotometric method Twenty tablets (T.Ilosure), each containing 2 mg of iloperidone was weighed and finely powdered. A powder mass equivalent to 2 mg of iloperidone was transferred to 50 ml volumetric flask, dissolved with methanol and the volume was made up to get a concentration of 40 µg ml -1. The mixture was then sonicated for 20 min to dissolve the material completely and centrifuged at 3,000 rpm for 5 min. An aliquot of supernatant solution (2.5 ml) was transferred to 10 ml volumetric flask and made upto volume using methanol. The resulting solutions were then analyzed spectrophotometrically at 229 nm. Method Validation Both the methods were validated by following ICH recommendations for validation of analytical procedures 7. Linearity and range For the HPLC method, stock solution of iloperidone was suitably diluted with the mobile phase to get concentrations in the linear range of 1-10 μg ml -1. Then 2 μg ml -1 of paracetamol (IS) was added to each dilution and 20 μl was injected into the column, the peak area and retention times were recorded. The calibration curve for iloperidone was constructed by

A.S. Manjula Devi et al /Int.J. PharmTech Res.2012,4(2) 578 plotting the ratio of the peak area of iloperidone to that of internal standard (Y) against concentration (X) and linearity was evaluated by linear regression equation. The slope, intercept and correlation coefficient values were recorded. For the UV Spectrophotometric method, calibration graph was prepared with 2-20 µg ml -1 of iloperidone and absorbance was recorded at 229 nm. Each experiment was performed in six replicates. Accuracy Accuracy was determined by standard addition method. To a pre-analyzed sample formulation a known quantity of standard was added at three levels (50, 100 and 150% of the assay concentration). A constant concentration (2 µg ml -1 ) of IS was added to all the samples in HPLC experiment and the mixtures were analyzed by the developed method. Absorbance was measured directly for the UV spectrophotometric method. The experiment was performed in six replicates. %RSD and %recovery were calculated for all the concentrations. Precision The precision of the method was studied by repeatability (within-day) and intermediate precision (inter-day).the intra-day precision studies were carried out by estimating the response six times on the same day using three different concentrations (4, 5 and 6 µg ml -1 for HPLC method and 6, 10 and 14 µg ml -1 for UV spectrophotometric method) of iloperidone and inter-day precision studies were done by repeating the above procedure on three different days. A fixed concentration (2 µg ml -1 ) of the internal standard was added to all the samples in HPLC experiment. The results of precision studies were expressed as %RSD. LOD and LOQ Sensitivity of HPLC and UV methods were determined from limit of detection (LOD) and limit of quantitation (LOQ). The LOD and LOQ were calculated from the calibration curve using the following equations LOD = 3.3xσ/S and LOQ = 10 x σ/s, where, σ = standard deviation of y intercept of regression line S = slope of the calibration curve Specificity Purity of iloperidone and IS peaks were assessed in HPLC method by comparing the individual spectrum at three regions i.e. peak start, peak apex and peak end. The specificity of the method was further assessed by comparing the chromatograms obtained from standards and from placebo solutions prepared using the excipients most commonly present in pharmaceutical formulations, including lactose monohydrate, starch, hydroxypropyl methyl cellulose, and magnesium stearate. Specificity of UV spectrophotometric method was determined from the absorption spectra of iloperidone reference standard and that of formulation. Robustness Robustness of HPLC method was studied to evaluate the effect of small but deliberate variations of the chromatographic conditions on the method parameters. Robustness was determined by changing individually the flow rate (1±0.1ml/min), organic solvent (50±1%), ph (5.02±0.01) and evaluating their effects on peak parameters. System suitability tests The test was carried out by making six replicate injections of a standard solution containing 5 and 2 μg ml -1 of iloperidone and IS, respectively. The peak area of the sample and IS, number of theoretical plates (N), resolution(r s ), tailing factor (T),capacity factor (k ), and asymmetric factor (A s ) were analysed. Stability Studies A suitably diluted solution of iloperidone and IS in methanol was stored under laboratory bench conditions upto 48 hours, and under refrigeration (8±0.5ºC) for 5 days. The solutions were assayed by the proposed method. Figure 1: Overlay spectrum of iloperidone and internal standard[paracetamol]

A.S. Manjula Devi et al /Int.J. PharmTech Res.2012,4(2) 579 Figure 2: RP-HPLC chromatogram of standard solution of iloperidone (5 µg ml -1 ) with internal standard (2 µg ml -1 ) at 275 nm Table 1: Regression parameters and validation parameters of iloperidone by RP-HPLC and UV-spectrophotometric method Parameter RP-HPLC method UV-Spectrophotometric method Concentration range 1-10 2-20 Slope a 0.3338±0.01966 0.05535 ±0.00010 Intercept a -0.1138±0.01743-0.01602 ±0.00031 Correlation coefficient (r 2 ) 0.9986±0.0009 0.9988± 0.00040 LOD (µg ml -1 ) 0.1766 µg ml -1 0.0185 µg ml -1 LOQ (µg ml -1 ) 0.5889 µg ml -1 0.0562 µg ml -1 Recovery (%) a 100.193 ±0.4726 100.048±0.1352 Intra-day precision (RSD, %) a 0.35-0.60 0.04-09 Inter-day precision (RSD, %) a 0.37-1.14 0.07-0.14 a Mean ± SD, n = 6 Table 2: Assay of iloperidone in formulation by RP-HPLC and UV-spectrophotometric method Method Labeled amount Amount found Assay RSD (%) (mg/tab) (mg) a (%) RP-HPLC method 2 1.9977±0.0042 99.9133 0.2579 UV Spectrophotometric 2 2.0015±0.0019 100.0566 0.12712 method a Mean ± SD, n = 6 Table 3: System suitability parameters of the proposed RP-HPLC method Parameter Peak Drug IS Retention time (t R ) 7.619 2.849 Peak purity index 1 0.999 Asymmetry 1.120 0.99 Tailing factor (T) 1.245 0.97 Capacity factor (k ) 18.05 6.122 Theoretical plates (N) 8620 3140 Selectivity (α) 4.77 - Resolution (R s ) 8.5 - Figure 3: RP-HPLC chromatogram of formulation of iloperidone (5 µg ml -1 ) with internal standard (2 µg ml -1 ) at 275 nm

A.S. Manjula Devi et al /Int.J. PharmTech Res.2012,4(2) 580 RESULTS AND DISCUSSION HPLC Method: The current study describes a new RP-HPLC method and UV-spectrophotometric method for the determination of iloperidone in bulk and tablet dosage form. The retention times of IS and iloperidone was 2.8 and 7.6 min, respectively. Figure 1 shows the overlay spectrum of IS and iloperidone. The typical chromatograms of iloperidone with internal standard are shown in figure 2. The developed method was validated in terms of specificity, linearity and range, accuracy, intra-day and inter-day precision, LOD, LOQ, robustness and system suitability as per the recommendations of ICH guidelines. Specificity of the method was established by the complete separation of the peaks of iloperidone and internal standard in the presence of tablet excipients and their peak purity index which was found to be 1. The standard calibration curve was linear over the concentration range 1 to 10 μg ml -1 with a correlation coefficient (r 2 ) 0.9986, the corresponding linear regression equation being y = 0.3338x-0.11386. LOD and LOQ were found to be 0.1766 µg ml -1 and 0.5889 µg ml -1 respectively, indicating the sensitivity of the method. Recovery experiments demonstrated satisfactory accuracy with small relative standard deviations (RSD %). The mean recovery ranged from 99.92±0.3553 to 100.61±0.2579%. The developed RP-HPLC method was precise since the %RSD values were <2 for both repeatability and intermediate precision studies as recommended by ICH guidelines. RSD of repeatability (intra-day) and intermediate precision (inter-day) ranged from 0.3521 to 0.6085 and 0.3756 to 1.1486 respectively. Table 1 illustrates the regression parameters and validation parameters of iloperidone by RP-HPLC method. Deliberate changes in experimental conditions did not alter peak symmetry and there was no significant change in the retention time of IS and iloperidone during these experiments. The %RSD for each method parameter was calculated and was found to be <2. Present chromatographic conditions established the assay of commercial tablet of iloperidone and the results were in good agreement with the labelled claim, indicating the suitability of the method. The percentage amount of iloperidone present in the commercial formulation (T.Ilosure) was found to be 99.9133±0.2579 (table 2). Excipients present in the tablet did not interfere with the analyte peaks. Figure 3 shows the separation of analyte in the formulation and IS. Adequacy of the proposed RP-HPLC method for routine analysis of iloperidone was assured by system suitability tests. The capacity factor (k ) of IS and drug were found to be 6.122 and 18.05 respectively, signifying that both the peaks were well resolved with respect to the void volume. The tailing factor of 0.97 and 1.2 for IS and iloperidone peaks reflect good peak symmetry. The resolution (Rs) for the principle peak and internal standard was found to be 8.5, showing good separation. The theoretical plate number (N) was found to be 8620 for drug, thus demonstrating good column efficiency. The result of system suitability tests (table 3) shows that the newly developed method fulfils these requirements within the accepted limits. Solution stability studies demonstrated good stability of standard solution and sample solutions during all storage conditions and confirmed the applicability of the method for routine analysis. UV Spectrophotometric method: The solutions were scanned in the wavelength range of 200-400 nm after making suitable dilutions from the stock solutions. Iloperidone showed absorption maxima at 229 nm. The method was validated as per ICH guidelines. Iloperidone in methanol showed linear relationship in the concentration range of 2 to 20 µg ml -1 with a correlation coefficient (r 2 ) 0.9988±0.0004. The regression equation was found to be Y = 0.05537-0.01595. The recoveries of iloperidone estimated by standard addition method for the three concentrations ranged from 99.77 to 100.04% with acceptable %RSD values. LOD and LOQ were found to be 0.0185 µg ml -1 and 0.0561 µg ml -1 respectively. The % RSD values for intra-day and inter-day precision varied from 0.0417 to 0.0833 and 0.0793 to 0.1485 (table 1) respectively. The amount assayed was in good agreement with the label claim (table 2) with RSD values less than 2%. Chromatographic conditions were carefully optimized to get satisfactory resolution between analyte and IS.

A.S. Manjula Devi et al /Int.J. PharmTech Res.2012,4(2) 581 The final decision on mobile phase composition and flow rate was made on the basis of peak shape (peak area, asymmetry, tailing factor), baseline drift, time required for analysis, and cost of solvents. The optimized mobile phase was 0.1% trifluoroacetic acid:acetonitrile (ph 5.02) in the ratio of 50:50 v/v. Initial trial with water:methanol (20:80 v/v) produced no peak for iloperidone. Subsequently, 0.1% TFA:methanol (50:50 v/v) showed tailing factor 2.4 while 0.1% TFA:acetonitrile (50:50 v/v) gave a broad asymmetric peak at 9 minutes. Peak shape was found to improve when the ph was increased to 4. At ph 5.02, a good symmetric peak was obtained at 7.6 minutes. Resolution (8.5) between the analyte and IS was also satisfactory using this mobile phase. The value of resolution meets the acceptance criteria which should be greater than or equal to 2 5. Various IS like aceclofenac, diclofenac, paracetamol and ibuprofen were tried. Among these, paracetamol was chosen due to less run time (2.8 min), good peak shape, adequate resolution and easy availability. Diclofenac showed retention time greater than 19 minutes. The correlation between concentration and response was also excellent in both the established methods. The retention time observed at 2.8 and 7.6 min for drug and IS allowed rapid quantification of iloperidone which is important in routine analysis. The proposed methods were also found to be precise and accurate, as depicted by the statistical data of analysis. Recovery values met the acceptance criteria of 100±2% at the three different concentration levels evaluated. The current UV spectrophotometric method allowed rapid quantification of iloperidone in tablets. Methanol was chosen as the solvent because of good solubility and stability. The spectral characteristics were also good in this solvent. Though iloperidone showed two absorption maxima at 229 and 275 nm, 229 nm was selected as the detection wavelength since it showed better linearity and sensitivity at this wavelength with minimal interferences. Low values of %RSD signify the precision of the method. A good accuracy of the method was documented with mean recoveries close to 100%. Moreover, the method is highly sensitive as evidenced by the LOD and LOQ values. CONCLUSIONS The validated HPLC and UV methods were found to be accurate, precise and reliable. UV spectrophotometric method was simpler and sensitive than HPLC method and the same may be used as an alternative method when advanced instruments like HPLC are not available for routine quantification purpose. Both the methods can be employed for the routine quality control of iloperidone in tablet dosage forms. Acknowledgements: The authors thank Sun Pharmaceuticals Ltd, Chennai, India, for providing reference samples of iloperidone and the management, SNR Sons Charitable trust, Coimbatore for providing facilities to carry out the study. REFERENCES: 1. Thomson Reuters, Micromedex (R) Healthcare series, 2011, 148. 2. Mutlib AE and Strupczewski JT., Picogram determination of iloperidone in human plasma by solid-phase extraction and by high-performance liquid chromatography-selected-ion monitoring electrospray mass spectrometry, Journal of chromatography B., 1995, 669, 237-246. 3. Mutlib AE and Klein JT., Application of liquid chromatography/mass spectrometry in accelerating the identification of human liver cytochrome P450 isoforms Involved in the metabolism of iloperidone, JPET., 1998, 286, 1285-1293. 4. Mandpe LP and Varsha, BP., Stress degradation studies on iloperidone and development of a stability indicating HPLC method for bulk drug and pharmaceutical dosage form, Der Chemica Sinica.,2011, 2, 230-239. 5. Snyder LR, Kirkland JJ and Glajch JL, Practical HPLC Method Development,2 nd ed., John Wiley and Sons, New York, 1997, 165. 6. Willard HL, Merritt, LL Dean JA and Settle FA. Instrumental Methods of Analysis, 7 th ed, Wadsworth Publishing Company, California, 1988, 196. 7. International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use, ICH Harmonised Tripartite Guideline Validation of Analytical Procedures:Text and Methodology Q2(R1), Current Step 4 version, London 2005. *****