DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

Similar documents
MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

RADIATION MONITORING EXPERIMENT USING THERMOLUMINESCENT DOSIMETER FOR THE TR 19 CYCLOTRON AREA IN NUCLEAR RESEARCH INSTITUTE

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

DOSE ASSESSMENT WITH PASSIVE PERSONAL DOSIMETERS EXPOSED TO X-RAY GENERATOR USING THE 241 Am CALIBRATION CURVE

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Radiation Protection Services in R.Macedonia

IMAGES THAT SPEAK ABOUT ERRORS IN PERSONAL DOSIMETER HANDLING AND PARTICULAR SOLUTIONS

THERMOLUMINESCENT DOSIMETRIC SYSTEM USED FOR CARACTERIZATION OF THE UNDERGROUND ENVIRONMENT

IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS

The determination of timer error and its role in the administration of specified doses

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

INTRODUCTION. Material and Methods

TLD system for the monitoring of the environmental radioactivity

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

Anthropomorphic breast phantoms for quality assurance and dose verification

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

5th ADAMAS Workshop at GSI December 15-16, 2016, Darmstadt, Germany

Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

Male Breast Cancer-Dosimetry Evaluation in Teletherapy

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS

TLD SYSTEM FOR THE MONITORING OF THE ENVIRONMENTAL RADIOACTIVITY

Implementation of personnel dosimetry systems using thermoluminescence dosimetry

Diode calibration for dose determination in total body irradiation

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

Calibration of two 90 Sr+ 90 Y dermatological applicators

Neutron Measurements for Intensity Modulated Radiation Therapy

Zakithi Msimang SAAPMB/SARPA 2011

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Basic dosimetric data for treatment planning system

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

U.S. DEPARTMENT OF ENERGY LABORATORY ACCREDITATION PROGRAM (DOELAP) FOR PERSONNEL DOSIMETRY SYSTEMS

Quality assurance in external radiotherapy

Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters


Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

PMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.

Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom

Protection of the contralateral breast during radiation therapy for breast cancer

Neutron dose evaluation in radiotherapy

Radiochromic film dosimetry in water phantoms

International Practice Accreditation

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

Radiation Monitoring Instruments

BELdART-2: Moving towards safer radiotherapy. Steven Lelie 1,2, Wouter Schroeyers 1, Brigitte Reniers 1, Sonja Schreurs 1

Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

Accreditation of Dosimetry Services

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

Patient dosimetry for total body irradiation using single-use MOSFET detectors

SSDL Newsletter. Contents. From the editor. No. 51 September IAEA/WHO Network of Secondary Standards Dosimetry Laboratories

Digital film dosimetry in radiotherapy and the development of analytical applications software

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

EDITORIAL NOTE. factor, which has been used instead of the chamber factor N D

SRS/SBRT Errors and Causes

Commissioning and Radiobiology of the INTRABEAM System

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

Dosisverifikation mit Delta 4 Discover während der Behandlung

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

GAMMA DOSE DISTRIBUTION EVALUATION OF XiO TREATMENT PLANNING SYSTEM FOR STATIC FIELD IMRT, USING AAPM TG-119

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

Ali S. Meigooni, Kai Dou, Navid J. Meigooni, Michael Gnaster, Shahid Awan, Sharifeh Dini, and Ellis L. Johnson

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland


Transcription:

Romanian Reports in Physics, Vol. 63, No. 3, P. 700 706, 2011 DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS ZOE GHITULESCU 1, ANA STOCHIOIU 2, MIHAI DUMITRACHE 3, 1 CNCAN- National Commission for Nuclear Activities Control, Bucharest, E-mail:ghitulescuzoe@yahoo.com 2 IFIN-HH - IFIN-HH (Horia Hulubei National Institute of Physics and Nuclear Engineering, Bucharest, E-mail:stoc@ifin.nipne.ro 3 Research and Developing Army Centre, Bucharest, E-mail: mihai149@yahoo.com Received April 27, 2010 Abstract. The tissue equivalence (LiF) of the thermoluminescent dosimeters enables their use within teletherapy in measuring the absorbed dose on phantoms. In order to set up and compare treatment doses of the inner skull tumors using a photon beam, a number of 22 thermally pretreated and calibrated TLD 100 was used and the measurements were made using the small water phantom type 4322 PTW and the antropomorphical phantom Rando Alderson. The two phantoms are exposed at the isocenter (accelerator axis intersection) and respectively to a lateral 6 MeV (6 MV) photon beam under the reference conditions from the Varian s medical accelerator type Clinac 2300 C/D. We choose a low value of 15 MU in order to avoid an overexposure of the TLDs. The corresponding dose values for 15 MU established with the aid of the Farmer ionization chamber and of the TLD s average dose value are very close. The absorbed dose values measured for the z 3 depth (in Rando phantom) obtained with the TLD s method, analytically and by simulation are comparable with the considered reference value measured with the Farmer ionization chamber. Thus, their relative deviations against the value measured with the Farmer chamber are less than 0.45% and the TLD s response is good in assessing the prescribed treatment dose. Key words: thermoluminescent detectors, absorbed dose, medical accelerator, isocenter, reference conditions, Rando-Alderson head phantom. 1. INTRODUCTION A good method for assessing and monitoring the absorbed dose to patient and also for the personal dosimetry is to use the thermoluminescent dosimeters. The thermoluminescent dosimeters are usually used for patient doses estimation within clinical dosimetry and furthermore can be useful tools in establishing the own national Diagnostic Reference Levels (DRLs), according to the actual nuclear legislation. It is well known that the thermoluminescent intensity is proportional to the irradiation dose.

2 Dose measurements in teletherapy 701 The aim of this study is to use the TL dosimetry within teletherapy in measuring the absorbed dose on phantoms exposed to 6 MeV (6 MV) photon beam under the reference conditions from a medical accelerator. The TLD-100 components and the tissue have also close values of mass energy-absobtion coefficients for 6 MeV energy according to NIST tables. Due to their small size and to their linear response the TL dosimeters can be placed easily into the antropomorphic phantom and thus widely used in the quality assurance procedure for the treatment planning systems. The TL dosimeters can be used in checking the absorbed dose inside the target volume. In teletherapy using a medical accelerator, it is necessary to establish the prescribed dose for the clinical target volume inside the patient, while reducing to a minimum the exposure of other organs within its vicinity. The check of the absorbed dose using TL dosimeters and a phantom head exposed to a photon beam was not performed before in Romania but the measurements uncertainty obtained less than 5% could express a good response of the TLDs [1, 6]. It is also considered acceptable a deviation of ±3% between the stated dose and the TLD measured dose [2, 7]. 2. MATERIALS AND METHOD The TL dosimeters used in the following experiments described below are LiF: Mg, Ti detectors also known as TLD-100. There were also used two phantoms and the medical accelerator type Clinac 2300 C/D produced by Varian. In order to be used in clinical dosimetry TL dosimeters have to be thermally pretreated at 100 o C for 10 minutes. The measurement of the TL detectors was performed at the Laboratory for Environment and Personnel Dosimetry from IFIN- HH, Bucharest. A TLD Reader-Analyser Model 770A was used. It works at up to temperatures of 240 o C. A reading factor expressed in impulses was found by taking readings from each TLD on the 800 V scale, taking into account the reading factor for which the device was calibrated, the background value of the tray holder as well as the TLD calibration dose. For the TLD calibration a source of 137 Cs was used. The TLD were panoramically irradiated. According to the IFIN-HH calibration certificate the source had the dose rate of 4.765 mgy/h ± 3% at a distance of one meter on 20.11.2006. It was found at a distance of 0.5 m from the source, the value of the dose rate are of 18 mgy/h ± 3% including the tissue (water) correction corresponding to the calibration procedure. Thereby to get the three TLD lots calibrated at 10 msv dose it was necessary their iradiation twice for 27 47 and for 27 50. Usually the individual dosimeters are calibrated using a 137 Cs source and the TLD s under-response is 5% less than that obtained from their calibration at the 60 Co source [3].

702 Zoe Ghitulescu, Ana Stochioiu, Mihai Dumitrache 3 In order to set up and compare treatment doses of the brain tumors using a photon beam, a lot of TLD 100 was used and two types of measurements were made. The first one was made with the small water phantom type 4322 PTW, and the second one with the antropomorphical phantom Rando Alderson. 2.1. REFERENCE MEASUREMENTS WITH FARMER IC AND UNIDOSE PTW First the absorbed dose was checked with a Farmer ionization chamber of 0.6 cm 3 type 30001 and a Universal dosimeter PTW Unidos which were previously calibrated by a Secondary Standard Laboratory of Institute of Public Health in Bucharest. The beam parameters were verified thus corresponding to the acceptance certificate. The water phantom mentioned above was positioned at the isocenter using a Laser system and checked with the mechanical pointer. Measurements on the water phantom were made at the isocenter under the reference conditions set out by IAEA TRS 398/2004. The reference conditions used for finding the absorbed dose in the water consist of a 10cm 10cm isocenter field and 100 cm source to ionization chamber distance at a depth of 10 cm [4, 5]. The pressure was p = 0.1005 MPa and the temperature t = 21.8 o C inside the room where the measurements were carried out. While this was taking place the background radiation measured with the ratemeter type FH 40 was in the range of 150 200 nsv/h. The value read with the PTW Unidos was 81.13 cgy for 100 MU (Monitor Units) according to the acceptance tests, and an average of 12.17 cgy for 15 MU was found after three readings. The average value of the dose (according to TRS 398) has to be corrected by the energy correction factor K Q for the Farmer chamber [8]: D F = D read K Q = 12.17 0.9913 = 12.064 cgy, (1) where: D F is the value of corrected dose in water measured with Farmer IC at 10 cm; D read is the value of uncorrected dose in water measured with Farmer IC. The PTW dosimeter reading error is ± 0.001cGy and it is provided by Manufacturer. 2.2. TLD MEASUREMENTS The first set of measurements with a number of 22 TLD-100 were used for two kinds of measurements on the antropomorphic phantom Rando Alderson and on the previously mentioned water phantom. When the ionization chamber was placed at 10 cm depth, source to skin distance (SSD) being of 90 cm, the TLDs were inserted at the same depth of 10 cm and the beam data were inputted into the accelerator console (field size, energy, monitor units).

4 Dose measurements in teletherapy 703 The two exposures at the isocenter were made each time using 5 TLDs placed inside the Farmer chamber holder under a photon beam with an energy of 6 MeV (the displayed value being 6 MV). After the TLD s readings were taken, the resulting absorbed doses in water at a distance d = 10 cm are shown in Table 1. Table 1 Comparison of TLD measurements with the reference absorbed doses, determined with the Farmer IC D F D 100 u A= TLD D TLD D D F σ n D F σ No. [cgy] [cgy] [cgy] [cgy] DiTLD n [%] 1 2 3 4 5 6 7 1 13.213 2 11.788 3 12.281 12.148 12.064 0.805-0.69 2.9 4 10.832 5 12.627 1 2 3 4 5 10.804 11.571 14.051 12.031 11.482 11.987 12.064 1.103 + 0.64 4.1 *100[%] For the second set of measurements, four TLDs were placed inside the second head slice of the Rando phantom, in the first row from the parallel bars (Fig. 1). Fig. 1 TLD s positions inside the Rando phantom on the central photons beam. z i represent the distances between the patient skin and the TLD positions. The phantom head is irradiated three times with 15 MU under a lateral photon beam of 6 MV, keeping SSD = 90 cm and using 12 TLDs. The TLD s read values are in Table 2. Using the following specified values the calculation of the dose at the depth z 3 =10 cm gives [4,5]: D = MU F z cal F OUTPUT η = 15 1.025 1 0.784 = 12.054 cgy, (2) 3

704 Zoe Ghitulescu, Ana Stochioiu, Mihai Dumitrache 5 where: F cal represents the value of colimator factor measured in water and F OUTPUT is the ratio of MU/cGy and η is the 6 MeV efficiency. These values are taken from Clinacs acceptance measurements. The calculated dose value at the depth z 4 = 12.8 cm can be obtained as follows: D = D ( f R / f ) 2 TMR (z 4, A z 4, t) / TMR (z 3, A z 3, t) (3) z4 z3 TMR (z 4, A z 4, t) and TMR (z 3, A z 3, t) represent the values for tissue maxim ratio for the z 4 z 3 and were measured during the acceptance procedures [4, 5]. The f R and f are the distances from the tungsten target to the isocenter, respectively to the fourth TLD. We can obtain the dose for a point outside the isocenter using the equation (3). In order to estimate TMR (z 4, A z 4, t) = 0.709 and TMR (z 3, A z 3, t) = 0.784 the software Mephysto was used to trace the required curves thus enabling the linear interpolation method. Using the independent calculation program of the monitor units, MUCalc 1.2.7 created by Melvin A. Abraham we found that the 15 MU for a 10cm 10cm field at the depth z 3 = 10 cm corresponded to a dose D = 12.01cGy. z 3 TL D No. Table 2 The absorbed dose values inside the planning target volume u A= σ. 100 DiTLD n D TLD [cgy] Dcalc D itld z i D [mm] D i TLD D MUC 100 calc σn Dcalc calc [cgy] [cgy] i=1,2,3,4 [cgy] [cgy] [%] [%] 1 2 3 4 5 6 7 8 9 1 18.104 5 18.213 z 1 = 23 18.060 18.031 17.94 0.146-0.16 0.4 9 17.863 2 6 10 3 7 11 4 8 12 14.510 14.745 15.020 12.151 12.281 11.788 10.040 10.530 10.217 z 2 = 70 14.758 14.350 14.21 0.208-2.76 0.8 z 3 = 100 12.073 12.053 12.01 0.208-0.17 0.9 z 4 = 128 10.262 10.322 10.28 0.203 0.58 1.1 Table 2 enables us the comparison between the average TLD doses (the fourth column) for the established depths and the calculated dose values (the fifth column). The sixth column represents the dose values calculated with an independent software. The measured dose values inside Rando phantom and the calculated dose values using ionising chamber measurements in water phantom are close related.

6 Dose measurements in teletherapy 705 The displayed results (Fig. 2), taking into account the beam parameters (TMR, F OUT, F calc ) consists of isodoses graphs and the dose values calculated on the central axis for several depths. Fig. 2 The isodose graphs and their values for Rando phantom exposure at 6 MeV photon beam and for 10 cm 10 cm field. 3. DISCUSSION The absorbed dose values taken from TLDs placed on the Rando phantom for the depths z 1, z 2, z 3, z 4 are close to the analytical dose values and to those obtained through MUCalc program simulation. The absorbed dose values of 12.07 cgy, 12.05 cgy, 12.01cGy for the z 3 depth obtained with the TLD method, analytically and by simulation are comparable with the value of 12.06 cgy measured with the Farmer ionization chamber. Thus, their relative deviations against the considered reference value measured with the Farmer chamber are less than 0.45%. Therefore the use of TLDs for checking the absorbed dose calculated with the Treatment Planing Systems (TPS) is proved to be a very convenient tool in the begining of the treatment session for each patient. The relative deviations of the TLD dose values determined for the depth z 3 in the water phantom, against the value of 12.073 cgy measured with the help of Rando phantom are not greater than 0.7% [1]. This is due to the placing of 5 TLDs

706 Zoe Ghitulescu, Ana Stochioiu, Mihai Dumitrache 7 in each of the two different plastic capsules and their handling with ordinary tweezers when positioning them into the Farmer chamber s holder. According to the percentage relative deviations from Tables 1 and 2, the values of the ratios D i,tld / D F and D i,tld / D calc are greater than the calculated value of 0.9936 and less than the calculated value of 1.028 showing acceptable results. The uncertainties values u A which estimate the standard deviation of the mean value are less than 5% [5]. The deviations are in the range of +/ 3% and are smaller than the acceptance limit of 5% adopted by most TLD networks [1, 2]. The SDs (standard deviations) of the doses against their average values are small showing that TLD s response is good in assessing the prescribed treatment dose. Acknowledgements. We express our gratitude for the given assistance in the TLDs calibration to physicist Adrian Leordeanu, Quality Management Director at Nuclear &Vacuum and lecturer at Nuclear Training Center of IFIN-HH Horia Hulubei National Institute of Physics and Nuclear Engineering, Magurele-Bucharest. Our warmest thanks for the scientific talks, practical advices and also for providing of some TL dosimeters are due to PhD physicist Constantin Milu, president of the Romanian Society of Radioprotection, to Professor Alexandru Jipa, Dean of Physics Faculty of the Bucharest University, and to medical physicist Teodor Sandulescu from the Oncological Institute Alexandru Trestioreanu, Bucharest. REFERENCES 1. *** IAEA/WHO TLD audit service Vienna, IAEA-SSDL Newsletter, No. 46, January, 9 14, (2002). 2. D. Kroutilíková and H. Zácková, Radiation Protection Dosimetry-TLD audit in radiotherapy, 85, 393 396 (1999). 3. M.L. Maiello, J.F. Gulbin, G. de Planque and H.T. Heaton, International Mini-Intercomparison Investigating an Under-Response of TLDs Calibration with 137 Cs, Radiation Protection Journal, 1990. 4. Subramaria Jayaraman, Lawrence H. Lanzl, Clinical Radiotherapy Physics, Springer, pp. 190 191, 205 206, 2002. 5. E.B. Podgorsak, Radiation Oncology Physics, A Handbook for Teachers and Students Vienna, pp. 78, 86 88, 171 172, 192, 212 (2005). 6. *** Physical aspects of quality assurance in radiation therapy, AAPM Report No.13, New York, Vol. 1& 2, 25 28 November 1994. 7. Ans Swinnen, Quality assurance in radiotherapy: Development and validation of mailed dosimetry procedure for external audits using a multipurpose phantom and in vivo dosimetry, Leuven, 2005. 8. IAEA -TRS 398, Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water, Vienna, 2004.