Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times

Similar documents
Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

Leila E. A. Nichol Royal Surrey County Hospital

Introduction of RapidArc TM : an example of commissioning and implementing a QA programme for a new technology

PMP. Gamma Evaluation with Portal Dosimetry for Volumetric Modulated Arc Therapy and Intensity-Modulated Radiation Therapy.

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

biij Patient-specific quality assurance of RapidArc treatments: Portal prediction dosimetry compared with phantom studies Krishna Murthy K *

A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.

Trajectory Modulated Arc Therapy: Application to Partial Breast Irradiation. Research and development to advance radiotherapy

Minimizing dose variation from the interplay effect in stereotactic radiation therapy using volumetric modulated arc therapy for lung cancer

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Evaluating Pre-Treatment IMRT & VMAT QA Techniques Using Receiver Operating. Characteristic (ROC) Analysis. Allison Lorraine Mitchell

MEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO


Small field diode dosimetry

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

EPID dosimetry for pretreatment quality assurance with two commercial systems

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Debate Motion. Theoretical benefits of VMAT: Real Costs of VMAT. "Arc Based Techniques Will Make Conventional IMRT Obsolete

Quantification of interplay and gradient effects for lung stereotactic ablative radiotherapy (SABR) treatments

Statistical process control analysis for patient-specific IMRT and VMAT QA

IMRT QA: Point Dose Measurements or 2D Array?

IMRT Implementation And Patient Specific Dose Verification With Film And Ion Chamber Array Detectors

Reliability of the gamma index analysis as a verification method of volumetric modulated arc therapy plans

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments

EPID dosimetry confi guration and pre-treatment IMRT verifi cation

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

Confidence limits for patient-specific IMRT dose QA: a multi-institutional study in Korea

I. Equipments for external beam radiotherapy

Comparing phase and amplitude gated volumetric modulated arc therapy for stereotactic body radiation therapy using 3D printed lung phantom

A multi-institution evaluation of MLC log files and performance in IMRT delivery

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

Learning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy

International Multispecialty Journal of Health (IMJH) ISSN: [ ] [Vol-3, Issue-9, September- 2017]

Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry

Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Volumetric Modulated Arc Therapy - Patient Specific QA

A dosimetric evaluation of VMAT for the treatment of non-small cell lung cancer

The impact of continuously-variable dose rate VMAT on beam stability, MLC positioning, and overall plan dosimetry

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment

Technique For Plan Quality and Efficiency Using VMAT Radiosurgery For Patients with Multiple Brain Metastases

Original Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor

Electron Beam ET - Reloaded Therapy - Reloaded

Baotian Huang 1, Lili Wu 1, Peixian Lin 2 and Chuangzhen Chen 1*

A dosimetric comparison between volumetric-modulated arc therapy and dynamic conformal arc therapy in SBRT

Quality assurance in external radiotherapy

Image Guided Stereotactic Radiotherapy of the Lung

Implementation of advanced RT Techniques

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Mitsubishi Heavy Industries Technical Review Vol. 51 No. 1 (March 2014)

The effect of body contouring on the dose distribution delivered with volumetric-modulated arc therapy technique

EPID-based dosimetry to verify IMRT planar dose distribution for the as1200 EPID and FFF beams

Gamma analysis dependence on specified low-dose thresholds for VMAT QA

Dosisverifikation mit Delta 4 Discover während der Behandlung

PMP. Original Article. Introduction. Hee Jung Kim, Sung Yong Park, Young Hee Park, Ah Ram Chang

Asynchronization. (aka MLC interplay effect with tumor motion)

Research Article An IMRT/VMAT Technique for Nonsmall Cell Lung Cancer

VMAT plans for treatment prostate cancer: Dosimetric verifications and comparison with 3D-CRT and IMRT

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

IMRT Planning Basics AAMD Student Webinar

Eric E. Klein, Ph.D. Chair of TG-142

Cell survival following high dose rate flattening filter free (FFF) and conventional dose rate irradiation

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Radiobiological Impact of Planning Techniques for Prostate Cancer in Terms of Tumor Control Probability and Normal Tissue Complication Probability

8/2/2012. Transitioning from 3D IMRT to 4D IMRT and the Role of Image Guidance. Part II: Thoracic. Peter Balter, Ph.D.

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements

The performance of the progressive resolution optimizer (PRO) for RapidArc planning in targets with low-density media

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Potential conflicts-of-interest. Respiratory Gated and Four-Dimensional Tumor Tracking Radiotherapy. Educational objectives. Overview.

SunCHECK Patient Comprehensive Patient QA

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

Optimal collimator rotation based on the outline of multiple brain targets in VMAT

Transcription:

Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.87 Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times Mee Sun Yoon*, Yong-Hyeob Kim, Jae-Uk Jeong, Taek-Keun Nam*, Sung-Ja Ahn*, Woong-Ki Chung*, Ju-Young Song* *Department of Radiation Oncology, Chonnam National University Medical School, Gwangju, Department of Radiation Oncology, Chonnam National University Hwasun Hospital, Hwasun, Korea The gated RapidArc may produce a dosimetric error due to the stop-and-go motion of heavy gantry which can misalign the gantry restart position and reduce the accuracy of important factors in RapidArc delivery such as MLC movement and gantry speed. In this study, the effect of stop-and-go motion in gated RapidArc was analyzed with varying gating window time, which determines the total number of stop-and-go motions. Total 10 RapidArc plans for treatment of liver cancer were prepared. The RPM gating system and the moving phantom were used to set up the accurate gating window time. Two different delivery quality assurance (DQA) plans were created for each RapidArc plan. One is the portal dosimetry plan and the other is MapCHECK2 plan. The respiratory cycle was set to 4 sec and DQA plans were delivered with three different gating conditions: no gating, 1-sec gating window, and 2-sec gating window. The error between calculated dose and measured dose was evaluated based on the pass rate calculated using the gamma evaluation method with 3%/3 mm criteria. The average pass rates in the portal dosimetry plans were 98.72±0.82%, 94.91±1.64%, and 98.23±0.97% for no gating, 1-sec gating, and 2-sec gating, respectively. The average pass rates in MapCHECK2 plans were 97.80±0.91%, 95.38±1.31%, and 97.50±0.96% for no gating, 1-sec gating, and 2-sec gating, respectively. We verified that the dosimetric accuracy of gated RapidArc increases as gating window time increases and efforts should be made to increase gating window time during the RapidArc treatment process. Key Words: Gated RapidArc, Portal dosimetry, MapCHECK2, Gamma evaluation Introduction Many studies have been performed to minimize respiratory organ motion and consequent dosimetric errors in radiation treatment. The real-time position management (RPM) gating system (Varian Medical Systems, Palo Alto, CA) has been used effectively in many clinical sites. This system only irradiates tumors in a specified respiratory phase region and uses an infrared reflective marker block placed on the patient s abdomen as an indicator of tumor motion. It has also been used in intensity Received 16 May 2015, Revised 18 June 2015, Accepted 19 June 2015 Correspondence: Ju-Young Song (jysong@jnu.ac.kr) Tel: 82-61-379-7225, Fax: 82-61-379-7249 cc This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. modulated radiation therapy (IMRT), 1-4) and many clinics perform the gating method in volumetric modulated arc therapy (VMAT) with RapidArc (Varian Medical Systems, Palo Alto, CA) to reduce the respiratory motion-induced error in the complex treatment process. RapidArc optimizes dose distribution via the combination of gantry rotation speed, dose rate, and the dynamic motion of a multi-leaf collimator (MLC). The gated RapidArc has great potential for dosimetric error than applying the gating method to three-dimensional conformal radiation therapy (3D CRT) and IMRT delivered with fixed gantry fields. 5-9) Many studies have been examined the errors that can potentially occur in gated RapidArc process and verified the effective delivery could be performed under a stable respiratory pattern. 10) The respiratory pattern has many factors to keep a stable condition and a significant factor is gating window time length, which determines the total number of beam-on gates during - 87 -

Mee Sun Yoon, et al:dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times treatment beam delivery. More beam-on gates during heavy gantry rotation can compromise the dosimetric accuracy of gated RapidArc. In this study, the relationship between dosimetric errors in gated RapidArc and the number of stop-and-go gantry motions, which are inversely proportional to gating window time, was assessed. Errors were analyzed to establish the proper criteria for setting gating window time. a MapCHECK2 two-dimensional diode detector array (SunNuclear, Melbourne, FL) in a water-equivalent MapPHAN phantom (SunNuclear, Melbourne, FL) (the MapCHECK2 plan). After delivery, measured dos distributions were compared with calculated distributions in the two DQA plans. Error evaluation Materials and Methods Our study included 10 patients previously treated via gated IMRT for liver cancer. RapidArc plans were created for each patient using the Eclipse treatment planning system (Varian Medical Systems, Palo Alto, CA). Each plan contained two independent delivery quality assurance (DQA) plans for the measurement of integrated dose distribution during delivery. One DQA plan used an electronic portal imaging device (EPID) for portal dosimetry (the portal dosimetry plan) and the other used Fig. 1. System for the DQA process with the MapCHECK2 measurement and the dynamic moving phantom for the accurate control of gating window time. Fig. 2. Example of the DQA analysis with portal dosimetry plan in three different gating cases. - 88 -

PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 was based on the pass rate calculated using the gamma evaluation method with a 3% dose difference, and a 3-mm distance to agreement criteria. The Dynamic Platform Model 008PL system (CIRS Inc., Norfolk, VA), which can simulate respiratory motion, was used to set up accurate gating window times. The simulated respiratory cycle in this study was 4 sec, and beams were delivered and measured in three different conditions: no gating, a 1-sec gating window, and a 2-sec gating window. Our study used the Novalis Tx linear accelerator (Varian Medical Systems, Palo Alto, CA) and, for gating, the Varian RPM system. The overall setup for measurements using the MapCHECK2 plan is shown in Fig. 1. Results plan and the MapCHECK2 plan are shown in Fig. 2 and Fig. 3, respectively. The dosimetric accuracy of RapidArc with no gating and 1-sec and 2-sec gating windows are shown in Table 1. The pass rate worsened as gating window time decreased. Fig. 4 shows the results of the portal dosimetry plan. No gating produced the highest average pass rate; pass rates were 98.72±0.82%, 94.91±1.64%, and 98.23±0.97% for no gating, 1-sec gating, and 2-sec gating, respectively. Fig. 5 shows the results of the MapCHECK2 plan. No gating again produced the highest average pass rate; pass rates were 97.80±0.91%, 95.38±1.31%, and 97.50±0.96% for no gating, 1-sec gating, and 2-sec gating, respectively. The mismatch between the measurement data and the planning data was lower when gating window time was longer in both DQA plans. Example of the results obtained using the portal dosimetry Fig. 3. Example of the DQA analysis with MapCHECK2 plan in three different gating cases. - 89 -

Mee Sun Yoon, et al:dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times Table 1. The pass rates calculated using the gamma evaluation in the DQA process of gated RapidArc plans with a portal dosimetry and a MapCHECK2 measurement. Patient Portal dosimetry MapCHECK2 with MapPHAN No gating Gating (1 sec) Gating (2 sec) No gating Gating (1 sec) Gating (2 sec) A 96.5% 95.2% 95.6% 99.5% 97.1% 99.0% B 98.8% 95.4% 98.2% 98.3% 96.5% 98.2% C 98.5% 94.6% 98.3% 97.5% 95.4% 97.4% D 99.0% 91.7% 98.8% 97.7% 96.7% 98.1% E 99.1% 93.0% 98.5% 97.2% 95.4% 97.8% F 98.8% 95.3% 98.3% 96.9% 93.1% 96.5% G 99.2% 95.5% 98.7% 98.6% 95.7% 98.2% H 99.4% 96.6% 99.0% 96.5% 93.4% 96.0% I 99.1% 94.4% 98.8% 98.5% 95.5% 97.5% J 98.8% 97.4% 98.1% 97.3% 95.0% 96.3% Fig. 4. The comparison of DQA analysis results with portal dosimetry plan in three different gating cases. Fig. 5. The comparison of DQA analysis results with MapCHECK2 plan in three different gating cases. Discussion Gated RapidArc delivers VMAT beams only when the respiration phase is within a specified gating window and stops the gantry rotation when the respiration phase leaves the gating window. The stop-and-go motion of heavy gantry can offset the gantry restart position owing to momentum effects, which can also reduce the accuracy of MLC position, dose rate, and other important factors in RapidArc delivery. Our data support this premise: dosimetric error was greater in gated RapidArc delivery than continuous RapidArc delivery (no gating). Analysis of how the number of stop-and-go motions during RapidArc affects dosimetric error showed that the errors increased as the number of motion increased (i.e., as gating window time decreased). The average pass rate calculated using the gamma evaluation method was significantly lower (P<10 4 ) for gated compared with continuous beam delivery without gating in both the portal dosimetry and MapCHECK2 plans. In the MapCHECK2 plan, 2-sec gating had a lower average pass rate than no gating, but the difference was not significant (P=0.052). The average pass rate for 1-sec gating, however, was significantly lower than that of no gating (P<10 5 ), as well as that of 2-sec gating (P<10 5 ). In the portal dosimetry plan, the average pass rates of both - 90 -

PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 1-sec and 2-sec gating were significantly lower than no gating (P<10 4 ). The low pass rate in the 2-sec gating window may reflect the inevitable effects of stop-and-go motions on the EPID-attached gantry structure. Unlike the MapCHECK2 device, which is static, the EPID may introduce detector position errors. The average pass rate for 1-sec gating was significantly lower than that for 2-sec gating (P<10 3 ) in the portal dosimetry plan as in the MapCHECK2 plan. Our results confirm that dosimetric accuracy in gated RapidArc increases as gating window (beam-on) time increases and thus suggest more stop-and-go motions result in more dosimetric errors. The difference between the dose calculated in a plan and the dose delivered to the patient via a gated RapidArc is mainly due to the interplay effects of organ motion during treatment. 11,12) Court et al. 13) suggest that errors based on interplay effects cancel each other out when doses are delivered in many fractions and thus do not affect overall treatment accuracy. Although interplay effects produce random errors owing to differences in the organ motions of different patients, our study shows that gating window time length is a determinant of systematic error and thus should be considered when planning gated RapidArc treatments. A stable respiratory pattern is a basic prerequisite of accurate gated RapidArc treatment, and gating window time, which determines the number of stop-and-go motions, should be as long as possible. The suitable respiratory pattern for gated RapidArc includes a long flat respiratory phase in the region of end exhalation rather than a steep sinusoidal phase pattern. Proper training of patients will enable stable, flat respiratory patterns that result in longer gating window times during gated RapidArc treatment. Conclusion Because the dosimetric accuracy of gated RapidArc increases as the number of gantry stop-and-go motions decreases, efforts should be made to increase gating window time during the RapidArc treatment process. Training patients to maintain a flat respiratory pattern during end exhalation will help increase gating window time in gated RapidArc. References 1. Ahmed RS, Shen S, Ove R, Duan J, Fiveash JB, Russo SM: Intensity modulated with respiratory gating for radiotherapy of the pleural space. Med Dosim 32(1):16-22 (2007). 2. van der Geld YG, van Triest B, Verbakel WF, et al: Evaluation of four-dimensional computed tomography-based intensity-modulated and respiratory-gated radiotherapy techniques for pancreatic carcinoma. Int J Radiat Oncol Biol Phys 72(4): 1215-20 (2008). 3. Chen H, Wu A, Brandner ED, et al: Dosimetric evaluations of the interplay effect in respiratory-gated intensity-modulated radiation therapy. Med Phys 36(3):893-903 (2009). 4. Cheong KH, Kang SK, Lee M, et al: Evaluation of delivered monitor unit accuracy of gated step-and-shoot IMRT using a two-dimensional detector array. Med Phys 37(3):1146-51 (2010). 5. Ling CC, Zhang P, Archambault Y, Bocanek J, Tang G, Losasso T: Commissioning and quality assurance of RapidArc radiotherapy delivery system. Int J Radiat Oncol Biol Phys 72(2): 575-81 (2008). 6. Nicoloni G, Vanetti E, Clivio A, Fogliata A, Cozzi L: Pre-clinical evaluation of respiratory-gated delivery of volumetric modulated arc therapy with RapidArc. Phys Med Biol 55(12): N347-57 (2010). 7. Van Esch A, Huyskens DP, Behrens CF, et al: Implementing RapidArc into clinical routine: a comprehensive program from machine QA to TPS validation and patient QA. Med Phys 38(9): 5146-66 (2011). 8. Choi K, Xing L, Koong A, Li, R: First study of on-treatment volumetric imaging during respiratory gated VMAT. Med Phys 40(4):040701 (2013). 9. Coleman A, Skourou C: Sensitivity of volumetric modulated arc therapy patient specific QA results to multileaf collimator errors and correlation to dose volume histogram based metrics. Med Phys 40(11):111715 (2013). 10. Riley C, Yang Y, Li T, Zhang Y, Heron DE, Huq MS: Dosimetric evaluation of the interplay effect in respiratory-gated RapidArc radiation therapy. Med Phys 41(1):011715 (2014). 11. Ong C, Verbakel WF, Cuijpers JP, Slotman BJ, Senan S: Dosimetric impact of interplay effect on RapidArc lung stereotactic treatment delivery. Int J Radiat Oncol Biol Phys 79(1): 305-11 (2011). 12. Li R, Mok E, Han B, Koong A, Xing L: Evaluation of the geometric accuracy of surrogate-based gated VMAT using intrafraction kilovoltage x-ray images. Med Phys 39(5):2686-93 (2012). 13. Court L, Wagar M, Berbeco R, et al: Evaluation of the interplay effect when using RapidArc to treat targets moving in the craniocaudal or right-left direction. Med Phys 37(1):4-11 (2010). - 91 -

Mee Sun Yoon, et al:dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times 호흡연동래피드아크치료시빔조사구간설정에따른선량변화분석 * 전남대학교의과대학방사선종양학교실, 화순전남대학교병원방사선종양학과 윤미선 * ㆍ김용협 ㆍ정재욱 ㆍ남택근 * ㆍ안성자 * ㆍ정웅기 * ㆍ송주영 * 호흡연동래피드아크치료는무건운하중의선형가속기갠트리의회전과정지의반복과정에서갠트리회전재시작점의오차와다엽조리개의정확한움직임및갠트리속도와같은래피드아크의선량정확도를결정하는요소들의오차가능성으로인한선량오류가발생할수있다. 본연구에서는이러한갠트리의회전과정지의반복적인동작이호흡연동래피드아크치료의정확도에어떠한영향을끼치는지치료시총회전과정지동작수를결정하는빔조사구간길이의변화를통해분석하였다. 총 10명의간암환자를대상으로래피드아크치료계획을수립하였고, RPM 호흡연동장치와정확한빔조사구간길이를설정하기위해동적팬텀을사용하였다. 각래피드아크치료계획당 EPID를사용한 portal dosimetry delivery quality assurance (DQA) 계획과이차원다이오드검출기배열장치인 MapCHECK2를사용한 DQA 계획을수립하여호흡연동방사선치료과정에서누적된선량분포의정확도를분석하였다. 모든환자의호흡주기는 4초로설정하였고, 수립한 DQA 계획들을호흡연동없이연속적으로조사하는것과, 1초의빔조사구간과 2초의빔조사구간, 총 3가지의경우에대해실제방사선량측정과감마평가를통해선량의정확도를분석하였다. Portal dosimetry DQA 경우평균감마평가의합격률은호흡연동없이연속적일때 98.72±0.82% 였고, 1초의빔조사구간의경우 94.91±1.64%, 2초의빔조사구간의경우 98.23±0.97% 이었다. MapCHECK2 DQA경우평균감마평가의합격률은호흡연동없이연속적일때 97.80±0.91% 였고, 1초의빔조사구간의경우 95.38±1.31%, 2초의빔조사구간의경우 97.50±0.96% 였다. 본연구결과를통해빔조사구간의길이가증가하여갠트리정지동작수가감소할수록호흡연동래피드아크의선량정확도가증가함을확인할수있었으며, 이러한특성을호흡연동방사선치료대상환자의선정및호흡방식에대한교육과정에고려되어야할것으로판단되었다. 중심단어 : 호흡연동래피드아크, 포탈선량측정, MapCHECK2, 감마평가 - 92 -