Remote Magnetic Navigation guided Ventricular Tachycardia ablation with continuous flow Mechanical Circulatory Support

Similar documents
Clinical Evidence for the Stereotaxis Niobe Magnetic Navigation System: Advantages of Electrophysiology (EP) Ablation

Case Report Epicardial Ablation: Prevention of Phrenic Nerve Damage by Pericardial Injection of Saline and the Use of a Steerable Sheath

Peri-Mitral Atrial Flutter with Partial Conduction Block between Left Atrium and Coronary Sinus

Case Report Substrate Based Ablation of Ventricular Tachycardia Through An Epicardial Approach

Complex Left Atrial (CLA) Best Practices for Navigant

Case Report What Next After Failed Septal Ventricular Tachycardia Ablation?

The first human experience of a contact force sensing catheter for epicardial ablation of ventricular tachycardia

Catheter ablation of monomorphic ventricular tachycardia. Department of Cardiology, IKEM, Prague, Czech Republic

A rare case of acute myocardial infarction during extraction of a septally placed implantable cardioverter-defibrillator lead

Antiarrhythmic Drugs and Ablation in Patients with ICD and Shocks

Impella Ins & Outs. CarVasz November :45 12:15

Ventricular Arrhythmias

Title. CitationJournal of Electrocardiology, 43(5): Issue Date Doc URL. Type. File Information.

Girish M Nair. MBBS, Jeffrey S Healey. MD, Elaine Gordon. MD, Syamkumar Divakaramenon MBBS, Carlos A Morillo. MD.

ΔΠΔΜΒΑΣΙΚΗ ΘΔΡΑΠΔΙΑ ΚΟΙΛΙΑΚΩΝ ΑΡΡΤΘΜΙΩΝ

EHRA EUROPACE How to perform epicardial ventricular tachycardia mapping and ablation

Reversible mechanical atrioventricular block during cryoablation for paroxysmal atrial fibrillation with a 28 mm balloon

Journal of the American College of Cardiology Vol. 37, No. 2, by the American College of Cardiology ISSN /01/$20.

Recovering Hearts. Saving Lives.

Use of Catheter Ablation in the Treatment of Ventricular Tachycardia Triggered by Premature Ventricular Contraction

Impella Versus Intra-Aortic Balloon Pump For Treatment Of Cardiogenic Shock: A Meta-Analysis of Randomized Controlled Trials

Map-Guided Ablation of Non-ischemic VT. Takashi Nitta Cardiovascular Surgery, Nippon Medical School Tokyo, JAPAN

Ventricular Tachycardia in Structurally Normal Hearts (Idiopathic VT) Patient Information

Mini-Electrodes Help Identify Hidden Slow Conduction. during Ventricular Tachycardia Substrate Ablation

Ventricular tachycardia and ischemia. Martin Jan Schalij Department of Cardiology Leiden University Medical Center

Focus on the role of Catheter Ablation: Simple cases Intermediate level Difficult cases (and patients) Impossible (almost )

Atrial fibrillation (AF) is a disorder seen

Non-Invasive Ablation of Ventricular Tachycardia

Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm

2010 Canadian Cardiovascular Society/ Canadian Heart Rhythm Society Training and Maintenance of Competency in Adult Clinical Cardiac

TRIALS. Hendriks et al. Trials (2015) 16:487 DOI /s

Advances in Ablation Therapy for Ventricular Tachycardia

Catheter ablation of atrial fibrillation: Indications and tools for improvement of the success rate of the method. Konstantinos P.

Alcohol Septal Ablation for Hypertrophic Obstructive Cardiomyopathy. CardioVascular Research Foundation

Epicardial substrate ablation for Brugada syndrome

Rationale for Prophylactic Support During Percutaneous Coronary Intervention

Title. CitationJournal of electrocardiology, 46(6): Issue Date Doc URL. Type. File Information

Visualization of Critical Isthmus by Delayed Potential Mapping

Mechanics of Cath Lab Support Devices

Natale A, Tung R, Szili-Torok T, Di Biase L xxx xxx on behalf of Magnetic VT investigators

Summary, conclusions and future perspectives

Ablative Therapy for Ventricular Tachycardia

Mechanics of Cath Lab Support Devices

UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS

Treatment of VT of Purkinje fiber origin: ablation targets and outcome

SUPPLEMENTAL MATERIAL

Prophylactic ablation

Raphael Rosso MD, Yuval Levi Med. Eng., Sami Viskin MD Tel Aviv Sourasky Medical Center

Tachycardia Devices Indications and Basic Trouble Shooting

Catheter Ablation of Recurrent Ventricular Tachycardia Should Be Done Before Antiarrhythmic Therapy with Amiodarone is Tried CONTRA

The implantable cardioverter defibrillator is not enough: Ventricular Tachycardia Catheter Ablation in Patients with Structural Heart Disease

IHCP bulletin INDIANA HEALTH COVERAGE PROGRAMS BT JANUARY 24, 2012

Intracardiac Echocardiography in Premature Ventricular Complex/Ventricular Tachycardia Ablation

Prevention of sudden cardiac death: With an emphasis on sudden cardiac death from ventricular arrhythmias

Urgent VT Ablation in a Patient with Presumed ARVC

Case Report Catheter ablation of ventricular tachycardia related to a septo-apical left ventricular aneurysm

Medicine. Dynamic Changes of QRS Morphology of Premature Ventricular Contractions During Ablation in the Right Ventricular Outflow Tract

Πρώτης γραμμή θεραπεία η κατάλυση κοιλιακής ταχυκαρδίας στην ισχαιμική μυοκαρδιοπάθεια

Dialysis-Dependent Cardiomyopathy Patients Demonstrate Poor Survival Despite Reverse Remodeling With Cardiac Resynchronization Therapy

VT Ablation in Structural Heart Disease Patient Information

Ablation of Ventricular Tachycardia in Non-Ischemic Cardiomyopathy

When to implant an ICD in systemic right ventricle?

Reentrant Ventricular Tachycardia Originating in the Right Ventricular Outflow Tract

Biventricular Pacing: A Promising Therapeutic Alternative for Patients with Severe Congestive Heart Failure

Commentary:Right Ventricular-Tricuspid Valve Interdependance And The Challenges For Structural Heart Valve Therapy

file://c:\documents and Settings\admin\My Documents\CV\92.htm

Disclosures 8/29/2016. VT Ablation 2016: Indications and Expected Outcomes. Medtronic: advisory board, review panel. St Jude Medical: speakers bureau

Case Report Catheter Ablation of Long-Lasting Accelerated Idioventricular Rhythm in a Patient with Mild Left Ventricular Dysfunction

Substrate modification or ventricular tachycardia induction, mapping, and ablation as the first step? A randomized study

Dual-Chamber Implantable Cardioverter-Defibrillator

Magnetic Guidance System for Cardiac Electrophysiology A Prospective Trial of Safety and Efficacy in Humans

Ventricular Tachycardia Substrate. For the ablationist. Stanley Tung, MD FRCPC Arrhythmia Service/St Paul Hospital University of British Columbia

Need to Know: Implantable Devices. Carolyn Brown RN, MN, CCRN Education Coordinator Emory Healthcare Atlanta, Georgia

Novel Approaches to VT Management Glenn M Polin MD

Impact of radiofrequency ablation of frequent post-infarction premature ventricular complexes on left ventricular ejection fraction

Original Article Pulmonary vein isolation without left atrial mapping

Journal of the American College of Cardiology Vol. 61, No. 20, by the American College of Cardiology Foundation ISSN /$36.

HEART OF THE MATTER: cardiac issues in safe endoscopy & sedation

Accepted Manuscript. Improving Survival in Cardiogenic shock: Is Impella the Answer?,, James J Glazier MD, Amir Kaki MD S (18)

Epicardial Approach to Mapping and Ablation of VT: Clinical and ECG Predictors of Epicardial Location

Accepted Manuscript. Inadvertent Atrial Dissociation Following Catheter Ablation: A Demonstration of Cardiac Anisotropy and Functional Block

Sustained monomorphic ventricular tachycardia (VT) due to

Successful treatment of tachycardia-induced cardiomyopathy secondary to dual atrioventricular nodal nonreentrant tachycardia using cryoablation

Accepted Manuscript. Tisa Saha, MD, Robert Norris, MD, Jeffrey Luebbert, MD. S (17) DOI: /j.hrcr Reference: HRCR 456

Ventricular Tachycardia Radiofrequency Ablation with Extracorporeal Membrane Oxygenation

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia

PVCs: Do they cause Cardiomyopathy? Raed Abu Sham a, M.D.

- Special VT Cases - Idiopathic Dilated Cardiomyopathy. D. Bänsch

Indications for catheter ablation in 2010: Ventricular Tachycardia

MEDICAL POLICY Cardioverter Defibrillators

Electroanatomic Substrate and Outcome of Catheter Ablative Therapy for Ventricular Tachycardia in Setting of Right Ventricular Cardiomyopathy

Different indications for pacemaker implantation are the following:

Surgical Ablation of Refractory Ventricular Tachycardia in Patients With Nonischemic Cardiomyopathy

New technology to minimize radiation

Dottorato di ricerca in Immunoreumatologia e Oncologia Clinica e Sperimentale, Bioetica ed Epidemiologia dei tumori XXVII ciclo

Catheter Ablation of Ventricular Arrhythmias via the Radial Artery in a Patient With Prior Myocardial Infarction and Peripheral Vascular Disease

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

Automatic External Defibrillators

Ventricular Assisting Devices in the Cathlab. Unrestricted

CARDIOINSIGHT TM NONINVASIVE 3D MAPPING SYSTEM CLINICAL EVIDENCE SUMMARY

Transcription:

Accepted Manuscript Remote Magnetic Navigation guided Ventricular Tachycardia ablation with continuous flow Mechanical Circulatory Support Anna M.E. Noten, MD, Nicolas M. Van Mieghem, MD PhD, Tamas Szili-Torok, MD PhD FHRS PII: S2214-0271(19)30003-X DOI: https://doi.org/10.1016/j.hrcr.2019.01.002 Reference: HRCR 655 To appear in: HeartRhythm Case Reports Received Date: 7 November 2018 Revised Date: 20 December 2018 Accepted Date: 5 January 2019 Please cite this article as: Noten AME, Van Mieghem NM, Szili-Torok T, Remote Magnetic Navigation guided Ventricular Tachycardia ablation with continuous flow Mechanical Circulatory Support, HeartRhythm Case Reports (2019), doi: https://doi.org/10.1016/j.hrcr.2019.01.002. This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

1 2 Remote Magnetic Navigation guided Ventricular Tachycardia ablation with continuous flow Mechanical Circulatory Support 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 Anna M.E. Noten MD 1, Nicolas M. Van Mieghem MD PhD 1, Tamas Szili-Torok MD PhD FHRS 1 1: Department of Cardiology, Erasmus MC, University Medical Center, Rotterdam, the Netherlands Short title: Robotic Navigation and Mechanical Circulatory support Total Word count: 2401 Corresponding author: Tamas Szili-Torok, MD, PhD, FHRS Thoraxcenter, Department of Cardiology, Erasmus MC Postbus 2040, 3000CA Rotterdam, The Netherlands E-mail to: t.szilitorok@erasmusmc.nl Tel: +31-10-703-5018 Declarations of interest: none 1

21 22 Keywords Robotic navigation 23 24 25 26 27 Remote magnetic navigation Ventricular tachycardia Continuous flow mechanical circulatory support Mechanical circulatory support Catheter ablation 2

28 29 Introduction Catheter ablation (CA) is an important treatment option for patients with ischemic heart disease 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 presenting with ventricular tachycardia (VT). 1, 2 Various CA techniques are currently available, including remote magnetic navigation (RMN) guided ablation. 2 Most published studies reported superiority of RMN guided VT ablation over manual ablation, with respect to acute success, recurrence, procedure time, fluoroscopy time and complications. 4-6 The RMN system (Niobe Epoch, Stereotaxis Inc., St. Louis, MO) utilizes two permanent magnets mounted on pivoting arms, one magnet on either side of the patient, creating a computer controlled steerable magnetic field to remotely guide the movement of a magnetically enabled ablation catheter. 3 Unfortunately, many VTs are not suitable for mapping during VT ablation, largely due to their hemodynamic instability. 7 It is well known that hemodynamic unstable VT might be successfully ablated with the aid of mechanical circulatory support (MCS), using a variety of devices. 8-10 However, most of the currently available continuous flow MCS devices operate using a metal pump as core of the technology, which limits their use in a magnetic environment. The risk of electromagnetic interference has resulted in restraint in the use of percutaneous continuous flow MCS during RMN guided VT ablation. 11, 12 Especially in centers with a preference for RMN guided VT ablation, this would be a major limitation. For the first time, we report a case, where with careful planning, RMN guided VT ablation was successfully combined with hemodynamic support using the Impella continuous flow MCS. 3

48 49 Case report A 67-year-old male, with a past medical history of coronary artery disease causing ischemic 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 cardiomyopathy, was admitted to our hospital because of recurrent VTs. This patient underwent a successful resuscitation after ventricular fibrillation (VF) caused by an acute anterior myocardial infarction in 2006. Rescue percutaneous intervention of the left anterior descending (LAD) coronary artery was performed. Chronic total occlusion of the mid right coronary artery was noted. The distal right coronary system was supplied by collateral circulation from the LAD. In 2008, the patient was readmitted with myocardial infarction due to stent occlusion. The left ventricular (LV) function was poor (LVEF = 16%). Magnetic resonance imaging (MRI) showed a transmural anterior infarction. A VVI implantable cardioverter defibrillator (ICD) was implanted for secondary prevention. His ICD was upgraded to a cardiac resynchronization therapy-icd in 2018 because of progressive heart failure. The patient developed slow VTs in 2016, which had a cycle length under the programmed detection zone of the ICD. These VTs caused rapid hemodynamic deterioration requiring immediate basic life support. Initially medical treatment was started (amiodarone). The patient did not experience recurrences until 2018, when he was admitted again with slow VTs causing rapid hemodynamic deterioration. A manually guided VT ablation (Navistar Smarttouch D-curve catheter (Biosense Webster Inc., Diamond Bar, CA, USA) was undertaken. The ablation procedure could not be completed because of difficulty in maneuvering in the extremely dilated LV and frequent spontaneous induction of unstable VTs during mapping causing severe 69 70 hemodynamic deterioration. A repeat procedure with hemodynamic support was proposed. However, because of the difficulties encountered during the first ablation attempt, the operator 4

71 72 could not prioritize between RMN guided ablation, to facilitate maneuvering, and hemodynamic support by continuous flow MCS. Patient provided written informed consent for the procedure, 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 as well as presentation of his medical history. RMN guided VT ablation with hemodynamic support In July 2018, a VT ablation was scheduled with hemodynamic support using a continuous flow MCS device in combination with RMN. The procedure was performed under local anesthesia. The left and right femoral arteries and veins were punctured. A 5Fr sheath was introduced in the left femoral artery for hemodynamic monitoring. The Impella CP percutaneous continuous flow MCS (Abiomed Inc. Danvers, MA, USA) was inserted via a 14Fr sheath in the right femoral artery, crossing the aortic valve. A quadripolar catheter and ICE transducer were inserted in the right ventricular (RV) apex and right atrium respectively. After the MCS was activated, programmed ventricular stimulation resulted in fast VTs which were not mappable. The mean arterial pressure (MAP) remained adequate with help of the MCS. Next, ICE guided transseptal puncture was performed using an 8.5Fr SL1 sheath. After transseptal puncture, a bipolar voltage map of the LV was created using CARTO 3D electroanatomical mapping system (Biosense Webster Inc., Diamond Bar, CA, USA) together with the RMN system (Stereotaxis Inc., St. Louis, MO, USA) and the Navistar RMT Thermocool ablation catheter (Biosense Webster Inc., CA, USA). After a very dense voltage map was completed, scar homogenization was performed using the following radiofrequency settings: continuous ablation with 50W, 43 C, flow 20ml/min. Because of MCS, it was hemodynamically well tolerated to ablate during VT for more 92 93 than 35 minutes in total (Figure 1). The mapping and ablation were performed using a NaviStar RMT ThermoCool catheter (Biosense Webster Inc., CA, USA). The very extensive anterior scar 5

94 95 became unexcitable as proven by pacing manoeuvers. Only (non-clinical) fast VTs and VF were inducible after ablation, which were also well tolerated because of the continuous flow MCS. 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 Total procedure time was 354 minutes, total fluoroscopy time 27 minutes and total ablation time 2971 seconds. Continuous rhythm observation during 48-hours post-procedure showed no VT recurrences. Kidney function remained stable (CKD-EPI egfr pre-procedural 51 ml/min, postprocedural 45ml/min). The patient was discharged home the second day after the procedure. Procedural precautionary measures. Pre-procedure several precautionary measures were taken to safely combine the Impella continuous flow MCS with RMN guided ablation. Pre-procedure, a step-by-step procedural approach was designed by the electrophysiologist, interventional cardiologist and operating team, along with technical support from engineers from Abiomed and Stereotaxis. At the start of the procedure, Carto patches were positioned carefully on the chest and back of the patient, with the Impella motor lying as far out of their field as possible to prevent from EMI (Figure 2). Subsequently, the Automated Impella Controller (AIC) module, the Impella user control interface, was positioned carefully outside the magnetic field (5-Gauss zone). The continuous flow MCS was positioned and activated before the RMN magnets were put in navigate position, to prevent eventual motor stop. The Impella s flow rate can be adjusted by performance levels (P-levels), corresponding to a fixed rate of motor rotations per minute. Instead of using the automatic flow mode, the Impella was switched to manual P-control mode during this procedure. Level P8 was chosen as the start level, which corresponds to a high flow of ±3.5L/min. A flow of 115 116 1.5L/min was chosen as lower limit to prevent from aortic regurgitation. In manual P-control mode, the P-levels have to be downregulated manually when suction alarms appear. During this 6

117 118 procedure, no suction alarms occurred. No dislocation, nor interference due to the magnetic field was noted. However, the motor power and flow rate displayed at the AIC monitor, appeared to 119 120 121 122 123 124 125 126 127 128 129 130 131 132 133 134 135 136 137 be falsely elevated when the magnets were in navigation mode (an average false elevation of flow of 0.5L/min). As MAP remained constant, a technical origin was suspected. The motor power and flow rate returned to their baseline values when the magnets were moved to the stowed position. Discussion To the best of our knowledge, this is the first reported case where hemodynamic support by the Impella continuous flow MCS was used during Remote Magnetic Navigation guided VT ablation. Despite the magnetic field, the Impella functioned normally during this procedure. There are different approaches for VT ablation. An ablation approach based on substrate modification, does not require routine use of MCS. However, in patients with structural heart disease undergoing VT ablation, the numerous comorbidities, the complexity of underlying substrates and procedural factors (such as fluid overload and the use of anesthesia) might lead to acute hemodynamic decompensation. 13, 14 Moreover, only an average of 30% of patients can hemodynamically tolerate VT. 7 In the context of manual guided VT ablation, many studies reported on the safe use of continuous flow MCS for peri-procedural hemodynamic support. 8-10, 13 Furthermore, emergent rescue MCS insertion during VT ablation due to hemodynamic collapse, is associated with a high 30-day mortality compared to pre-emptive MCS insertion. 15 138 139 Therefore, it is important to identify high-risk patients undergoing CA of scar-related VT for prophylactic MCS. 14 7

140 141 Many studies compared RMN guided with manual VT ablation, 4-6 and reported superiority of 142 143 144 145 146 147 148 149 150 151 152 153 154 155 156 157 158 159 160 RMN with respect to acute success, recurrence, procedure time, fluoroscopy time and complications. Compatibility of hemodynamic support devices with the preferential VT ablation technique is desirable. The case illustrates that it s possible to safely combine the Impella CP continuous flow MCS with RMN guided VT ablation. However, there are several hazards to overcome when combining MCS with the strong magnetic fields used in RMN guided ablation. The primary concern is the risk of EMI. 11 Theoretically interference will be particularly seen when ablating in the LV outflow tract, RV outflow tract and septal wall, due to proximity to the MCS motor. Based on the early clinical experiences, 12 we advise careful positioning of Carto patches, AIC monitor and catheters to reduce EMI as much as possible. In case of EMI, a possible resolution is lowering the Impella s P-level until the EMI resolves, as proposed by Vaidya et al (e.g. from P8 to P6). 11 Fortunately, this was not necessary in our case. When the magnets are in navigation position before the MCS is activated, there s the theoretical hazard of motor stop because of its alignment with the magnetic field. Therefore, it s recommended to position and start the MCS prior to activation of the magnets. In automatic Impella motor control mode, measurements of the rotational speed of the motor could become inaccurate because of the magnetic field and therefore the flow rates could become uncontrollable. We observed that in navigation mode, Impella motor power and flow-rate displayed on the AIC were falsely elevated and should be interpreted with care. Consequently, it s recommended to monitor MAP and use the manual P-control mode instead of the automatic mode. Our experience is based on a single 161 162 case and further research is warranted to establish the safety of combining the Impella CP continuous flow MCS with magnetic guided ablation. However, the two techniques seem to be 8

163 164 compatible, which extends treatment options for patients suffering from hemodynamic unstable VTs. 165 166 167 168 Conclusion The Impella CP continuous flow MCS can be used to provide hemodynamic support during RMN guided VT ablation. 9

169 170 References 1. Priori SG, Blomstrom-Lundqvist C, Mazzanti A, et al. 2015 ESC Guidelines for the 171 172 173 174 175 176 177 178 179 180 181 182 183 184 185 186 187 188 189 management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Eur Heart J 2015;36:2793-2867. 2. Aliot EM, Stevenson WG, Almendral-Garrote JM, et al. EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias. Europace 2009;11:771-817. 3. Faddis MN, Blume W, Finney J, et al. Novel, magnetically guided catheter for endocardial mapping and radiofrequency catheter ablation. Circulation 2002;106:2980-2985. 4. Hendriks AA, Akca F, Dabiri Abkenari L, et al. Safety and Clinical Outcome of Catheter Ablation of Ventricular Arrhythmias Using Contact Force Sensing: Consecutive Case Series. J Cardiovasc Electrophysiol 2015;26:1224-1229. 5. Szili-Torok T, Schwagten B, Akca F, et al. Catheter ablation of ventricular tachycardias using remote magnetic navigation: a consecutive case-control study. J Cardiovasc Electrophysiol 2012;23:948-954. 6. Aagaard P, Natale A, Briceno D, et al. Remote Magnetic Navigation: A Focus on Catheter Ablation of Ventricular Arrhythmias. J Cardiovasc Electrophysiol 2016;27:38-44. 7. Zeppenfeld K, Stevenson WG. Ablation of ventricular tachycardia in patients with structural heart disease. Pacing Clin Electrophysiol 2008;31:358-374. 8. Lu F, Eckman PM, Liao KK, et al. Catheter ablation of hemodynamically unstable 190 191 ventricular tachycardia with mechanical circulatory support. Int J Cardiol 2013;168:3859-3865. 10

192 193 9. Miller MA, Dukkipati SR, Chinitz JS, et al. Percutaneous hemodynamic support with Impella 2.5 during scar-related ventricular tachycardia ablation (PERMIT 1). Circ 194 195 196 197 198 199 200 201 202 203 204 205 206 207 208 209 210 211 212 Arrhythm Electrophysiol 2013;6:151-159. 10. Reddy YM, Chinitz L, Mansour M, et al. Percutaneous left ventricular assist devices in ventricular tachycardia ablation: multicenter experience. Circ Arrhythm Electrophysiol 2014;7:244-250. 11. Vaidya VR, Desimone CV, Madhavan M, et al. Compatibility of electroanatomical mapping systems with a concurrent percutaneous axial flow ventricular assist device. J Cardiovasc Electrophysiol 2014;25:781-786. 12. Miller AC, Evans AC, Revenaugh J, Weiss JP, Reid BB, Kfoury AG. Combined use of TandemHeart percutaneous ventricular assist device and Stereotaxis magnetic navigation during cardiac ablation procedure. HeartRhythm Case Rep 2016;2:14-16. 13. Santangeli P, Muser D, Zado ES, et al. Acute hemodynamic decompensation during catheter ablation of scar-related ventricular tachycardia: incidence, predictors, and impact on mortality. Circ Arrhythm Electrophysiol 2015;8:68-75. 14. Muser D, Liang JJ, Castro SA, et al. Outcomes with prophylactic use of percutaneous left ventricular assist devices in high-risk patients undergoing catheter ablation of scar-related ventricular tachycardia: A propensity-matched analysis. Heart rhythm 2018;15:1500-1506. 15. Mathuria N, Wu G, Rojas-Delgado F, et al. Outcomes of pre-emptive and rescue use of percutaneous left ventricular assist device in patients with structural heart disease 213 214 undergoing catheter ablation of ventricular tachycardia. J Interv Card Electrophysiol 2017;48:27-34. 11

215 12

216 Figure 1. Ablation procedure 217 218 219 220 221 222 223 224 225 Figure 1 demonstrates several procedural recordings from the RMN guided VT ablation procedure. A: Fluoroscopic image of the Impella and catheters in RAO view. B: Intracardiac electrogram of a VT with a cycle length of 300msec recorded during ablation. C: Surface electrocardiogram and MAP curve on which the induction of a VT as seen. The pulsatile MAP curve changes into a flat curve around 60mmHg because of the hemodynamic support provided by the Impella. D: Surface electrocardiogram and MAP pressure recorded a few minutes later, on which MAP remains constant around 60mmHg during sustained VT. E: Image of the Carto map where all ablation points on the anterior wall of the left ventricle are visualized. 226 227 RA=right atrium, RV=right ventricle, ICD=implantable cardioverter defibrillator, RMN=remote magnetic navigation, VT=ventricular tachycardia, MAP=mean arterial pressure 13

228 Figure 2. Schematic image of Impella and Carto patch positioning 229 230 231 232 233 Figure 2 displays a schematic PA overview of positioning of the Impella CP continuous flow MCS device and the Carto patches. The Carto patches were positioned anterior and posterior on the chest and back of the patient, so that the Impella CP motor was lying as far out of their field as possible. A=anterior, P=posterior. 14

Key teaching points - Hemodynamic support by the Impella continuous flow mechanical circulatory support device seems to be compatible with remote magnetic navigation guided ablation and did not result in electromagnetic interference - To prevent electromagnetic interference of Impella with the strong magnetic fields used in remote magnetic navigation guided ablation, we advise careful positioning of materials as well as starting the Impella in advance to the magnetic fields - Precautionary use of the manual mode on the Impella instead of the automatic mode is recommended during remote magnetic navigation guided ablation