Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Similar documents
Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

Elekta Synergy Digital accelerator for advanced IGRT

Case Study. Institution Farrer Park Hospital

Elekta Infinity Digital accelerator for advanced treatments

Intensity modulation techniques for spinal treatments and on-line VolumeView TM guidance

Verification of Advanced Radiotherapy Techniques

Unrivaled, End-to-End

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Elekta Infinity. Digital accelerator for advanced treatments. Redefining treatment precision, speed and control

ABAS Atlas-based Autosegmentation

Dosimetric Analysis Report

A VMAT PLANNING SOLUTION FOR NECK CANCER PATIENTS USING THE PINNACLE 3 PLANNING SYSTEM *

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

ph fax

Activity report from JCOG physics group

Variable Dose Rate Dynamic Conformal Arc Therapy (DCAT) for SABR Lung: From static fields to dynamic arcs using Monaco 5.10

Clinical Precision for Best Cancer Care. Dee Mathieson Senior Vice President Oncology Business Line Management

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

SRS Uncertainty: Linac and CyberKnife Uncertainties

I. Equipments for external beam radiotherapy

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Leila E. A. Nichol Royal Surrey County Hospital

FROM ICARO1 TO ICARO2: THE MEDICAL PHYSICS PERSPECTIVE. Geoffrey S. Ibbott, Ph.D. June 20, 2017

Elekta - a partner and world-leading supplier

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

Radiosurgery. Most Important! 8/2/2012. Stereotactic Radiosurgery: State of the Art Technology and Implementation Linear Accelerator Radiosurgery

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

EORTC Member Facility Questionnaire

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Quality assurance of volumetric modulated arc therapy using Elekta Synergy

SunCHECK Patient Comprehensive Patient QA

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

Medical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center

Future upcoming technologies and what audit needs to address

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

IMRT Planning Basics AAMD Student Webinar

Radiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague

Lung SBRT in a patient with poor pulmonary function

Elekta Innovation and Growth. ESTRO 2012 Capital Markets Presentation

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

A model for assessing VMAT pre-treatment verification systems and VMAT optimization algorithms

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

CyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

Ingenia MR-RT. MR Systems. The comprehensive MR-sim solution to fit your planning

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

IMRT/IGRT Patient Treatment: A Community Hospital Experience. Charles M. Able, Assistant Professor

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

Work partially supported by VisionRT

Learning objectives. What kind of motions? 3D Dosimetry in the Clinic: Motion Interplay Effects in Dynamic Radiotherapy

IMRT for Prostate Cancer

Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc

iviewdose Confidence and assurance in dose delivery

Advanced Technology Consortium (ATC) Credentialing Procedures for 3D Conformal Therapy Protocols 3D CRT Benchmark*

Herlev radiation oncology team explains what MRI can bring

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Small field diode dosimetry

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

brain SPINE 2 SRS Matures into breast lung spine LUNG Dr. Robert Timmerman Discusses SBRT for Inoperable Lung Cancer BRAIN

MOSAIQ TM Image Enabled EMR. managing the spectrum of cancer care. Filmless and paperless workflow Universitätsklinikum Mannheim

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

Stereotactic Radiosurgery. Extracranial Stereotactic Radiosurgery. Linear accelerators. Basic technique. Indications of SRS

Image Guided Stereotactic Radiotherapy of the Lung

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?

Leksell Gamma Knife Icon contributes to significant growth, greater efficiency and an enhanced patient experience

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom

Varian Treatment. Streamlined Treatment Delivery Management Application. Specifications

Can we deliver the dose distribution we plan in HDR-Brachytherapy of Prostate Cancer?

RPC s Credentialing Programs for Clinical Trials

Pre-treatment and in-vivo dosimetry of Helical Tomotherapy treatment plans using the Dosimetry Check system

Varian Acuity BrachyTherapy Suite One Room Integrated Image-Guided Brachytherapy

Cyberknife Stereotactic Treatment

Quality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.

PRECISE, ROBOTIC TREATMENT AS INDIVIDUAL AS EVERY PATIENT

1 : : Medical Physics, Città della Salute e della Scienza, Torino, Italy

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

A method to predict patient specific table coordinates for quality assurance in external beam radiation therapy

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient

Development and evaluation of an end-to-end test for head and neck IMRT with a novel multiple-dosimetric modality phantom

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Limits of Precision and Accuracy of Radiation Delivery Systems

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Eric E. Klein, Ph.D. Chair of TG-142

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Transcription:

Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution University of Texas Health San Antonio Location San Antonio, Texas Medical Staff Daniel Saenz PhD Niko Papanikolaou PhD

Summary Phantom details: RTsafe PseudoPatient phantom filled with dosimetric gel constructed based on a real patient s CT scan RTsafe PseudoPatient phantoms of same size and shape as above but with ionization chamber and film inserts at location of QA target Simulated plan for: 6 brain metastases targets (6 25 mm) 1 larger QA target close to brain stem Isocenter at centroid of targets HDRS treatment details: Dose to targets = 8 Gy 5 non-coplanar VMAT arcs (lateral, vertex and 45 ) Monaco TPS Versa HD Agility XVI HexaPOD 1. Introduction The ability of Elekta Versa HD, equipped with the Agility MLC and in conjunction with the Monaco treatment planning system (TPS), to deliver accurate, precise and efficient radiosurgery to patients with multiple brain lesions was investigated in this study. The combined use of these technologies with image guidance and patient positioning (enabled by XVI and HexaPOD) allows high definition dynamic radiosurgery (HDRS). HDRS is characterized by the integration of optimization, planning, imaging and dose delivery techniques. The Monte Carlo-based TPS takes advantage of Agility MLC capabilities to create an optimized delivery and is able to place MLC leaves at 1 mm increments rather than to a course 5 mm grid. The 5 mm leaf width can be reduced in the jaw direction by positioning the Y-jaws within leaf widths on either end of the target in steps of 1 mm (Figure 1). Since stereotactic targets only require a few open leaf pairs, this jaw positioning accuracy can significantly improve target conformality. Fast leaf motion (up to 6.5 cm/sec) produces plans with sufficient modulation without compromising accuracy. 2

Figure 1. Illustration of conformality capabilities with jaw positioning The chain of uncertainty in radiosurgery consists of a number of steps, many of which are independent of dose planning and delivery. Therefore, when validating the accuracy of an integrated technique such as HDRS, end-to-end tests to examine both the localization and dose delivery accuracy of the workflow are essential. In this study, such validation is performed with a patient-specific, anatomically realistic phantom filled with a dosimetric gel, in conjunction with similar phantoms equipped with ion chamber and film inserts, to evaluate the dose distributions delivered by the HDRS workflow. This validation project was performed at the Mays Cancer Center at UT Health San Antonio. As a National Cancer Institutedesignated cancer center, our institution offers a wide array of radiation therapy modalities including VMAT, IMRT, linacbased SRS/SRT/SBRT and conventional radiotherapy. A range of HDR/LDR brachytherapy procedures is also offered with the Nucletron microselectron system. The department is equipped with three Elekta Versa HD, a Novalis Tx, two Varian 23EX linear accelerators and a GE LightSpeed CT Scanner for simulation purposes. MOSAIQ is the primary oncology information system (OIS), coordinating data transfer to the linear accelerators from the treatment planning systems (Pinnacle3, Monaco, Eclipse, Oncentra ). 3

2. Materials and Methods Planning A CT data set from a radiosurgery patient with multiple brain metastases was chosen as the model for the study. The metastases were artificially expanded and/or contracted in the CT data set so that a range of targets with diameters from 6 25 mm could be treated. Six targets were distributed in the brain to represent a range of possible target locations. This included targets close to the periphery of the head to test the effects of rotation on localization accuracy across the brain. The effects of small rotations would be the largest for such peripheral targets. In addition to these six targets, a larger target near the brainstem was also devised to be used for quality assurance of the dose delivery. The endpoint of this study was accurate localization and dosimetry in an anatomically realistic measurement. Therefore, it was crucial to perform 3D dosimetry, which is only possible with a gel dosimeter. The RTsafe PseudoPatient gel phantom (RTsafe P.C, Athens, Greece) was used in this study. This is the only 3D dosimeter that can be cast in nearly any form, allowing for measurement in a patientspecific geometry. The primary advantage of gel dosimetry in an anthropomorphic phantom is that unlike most patient-specific QA it does not rely on a recalculation of the plan on a phantom. Rather, the measurement in the phantom can be directly compared with the patient s calculated dose distribution. Three phantoms were produced by RTsafe based on the actual CT data set bony anatomy and external contour for the purpose of this study. The first one was filled with a dosimetric gel (also produced by RTsafe P.C. Athens, Greece) so that 3D dose measurements could be obtained. In addition, the other two nearly identical phantoms were made with minor modifications to accommodate other detectors. One had an insert for an A16 ionization chamber and the other had a holder for a film cassette. The ion chamber and film were situated to coincide with the location of the larger QA target. In making ion chamber and film measurements, we could validate the dosimetric accuracy in a larger target size, which is less prone to errors. Validation of the dose in this region allows for the accuracy to be transferred to the six targets in the brain via the gel phantom. A treatment plan for HDRS delivery was created in Monaco. The CT data set and structures were imported into the planning system, and the isocenter was set based on the centroid of the targets. An arc geometry of five non-coplanar VMAT arcs was used (lateral, vertex and 45 ). Since the gel saturates at a dose of 12 Gy, the plan was generated to deliver 8 Gy to the periphery of the six targets, with peaking doses remaining under 12 Gy. In the QA target, a homogeneous dose of 8 Gy was planned. Target penalty, quadratic overdose and conformality objectives were used for IMRT constraints. A 1 mm dose calculation grid spacing was used with one percent statistical uncertainty per calculation of dose to medium. In the five single arcs, 180 control points were allowed with 0.5 cm minimum segment width. Medium fluence smoothing was used. Once the phantoms were received and scanned, the film and ion chamber phantoms were fused with the patient-derived CT data set to determine the location of the ionization chamber sensitive volume and film plane. The mean dose to the ion chamber sensitive volume was recorded and the DICOM-RT dose was exported for gel and film analysis. 4

Figure 2. RTSafe PseudoPatient gel phantom photograph (left) and placement for dose delivery (right) Delivery For delivery of the HDRS plan, the patientderived CT data set was sent to MOSAIQ record-and-verify, and then imported to XVI for preparation. The isocenter was confirmed and the registration clipbox was set to cover the entire skull. The gel phantom was set up on the table with a mask and other standard immobilization devices on the SRS base plate. An XVI VolumeView cone beam CT (CBCT) scan was then performed and the resulting reconstructed scan was fused with the reference CT data set. Corrections were made in 6D, and the translations and rotations provided were executed by the HexaPOD table. The five arcs were then delivered. The process was repeated for the ionization chamber and film measurements. The A16 ionization chamber was cross-calibrated prior to use. A Gafchromic EBT3 film piece from a batch calibrated at a secondary standard laboratory was used for the film phantom measurement. Forty-eight hours after the phantom irradiation, the gel phantom was scanned on a 1.5 T MRI unit. A 2D, multi-slice, multi-echo, Half Fourier Single Shot Turbo Spin Echo (HASTE) PD to T2-weighted sequence was implemented sequentially using the head coil. The number of averages was set to 14 in order to increase signal-to-noise ratio, while the bandwidth was set to 1220 Hz/pixel in order to minimize MR-related geometric distortion. The MR protocol used is the one developed and recommended by the gel producer (RTsafe P.C.). The resulting scan was fused with the patient-derived CT data set for analysis. Quantitative analysis was performed with 3D gamma analysis and dose profile analysis. The film was scanned days later and compared with the dose distribution from the patient-derived CT. 5

Figure 3a. Sample profile and isodose curves from the film phantom analysis 3. Results Ion Chamber With the A16 ion chamber measurement in the phantom, the measured dose from the plan to the QA target was 844.8 cgy. By comparison, the mean dose in the ion chamber region of the patient-derived CT dose calculation was 829 cgy. This represents an agreement within two percent, which is a sufficient level of agreement for an end-to-end test at our institution. Film Analysis The film was read out two days later and compared with the exported dose distribution from the patientderived CT in Monaco. Sample profiles and isodose curves are shown in the following figures. Gamma analysis was also performed with various passing criteria. With 2%/2 mm criteria, the passing rate of gamma index < 1 was 94.2 percent. 6

Figure 3b. Sample profile and isodose curves from the film phantom analysis Figure 4. 2D gamma analysis (2%/2 mm) between TPS and film measured dose 7

Figures 5 6: (left) slice of the derived T2 maps of the irradiated phantom. High dose regions correspond to darker areas; (right) 1D profile comparison between calculated (TPS) and measured (RTsafe) dose distributions at the location depicted by the red line; error bars correspond to ±1 mm spatial uncertainty; 1D gamma index calculations are also given using passing criteria 5%/2 mm Gel Phantom Analysis Analysis of the gel was performed by registering the MR scan with the patient CT, and then comparing profiles and 3D gamma analyses between the TPS and measured dose distributions. Figures 5 8 show the dose profile comparisons and 1D gamma analysis along the profiles (5%, 2 mm). In the Figures, the target contours are also shown so the high dose region (in black) can be seen as the dose conforms to the target contours. In the QA target, 3D gamma analysis with criteria of 3%, 2 mm showed that 95.7 percent of the points had a gamma index < 1. 8

Figures 7 8: (left) slice of the derived T2 maps of the irradiated phantom. High dose regions correspond to darker areas; (right) 1D profile comparison between calculated (TPS) and measured (RTsafe) dose distributions at the location depicted by the red line; error bars correspond to ±1 mm spatial uncertainty; 1D gamma index calculations are also given using passing criteria 5%/2 mm 9

4. Discussion The results show the feasibility of highly efficient radiosurgery delivery. The absolute dose delivery accuracy of within three percent, and within one percent when calculated on the phantom itself, suggest a high degree of dosimetric precision with Elekta Versa HD, Agility and Monaco. This agreement validates the very low MLC leakage and precise control systems in place in the Elekta digital linear accelerator, as well as the dosimetric accuracy of the Monaco Monte Carlo TPS. Regarding localization accuracy, the effect of rotations is still the primary concern for a highly accurate stereotactic delivery across the range of the brain. When analyzing dose profiles across a wide range of regions of interest, some highly conformal profiles were found in conjunction with regions of slight misalignment. Some disagreement is to be expected for any 1D analysis. For example, they may consider misinformation from neighboring slices where high dose gradients may exist. Nevertheless, the localization results were promising given that CBCT of a gel phantom provides a limited amount of image detail to fuse with a planning CT. From a patient standpoint, these results indicate a vast increase in treatment efficiency by using a single isocenter VMAT technique. This translates to reduced time spent in a mask, thus minimizing uncertainty due to patient motion and couch shifts between isocenters, further increasing treatment accuracy. In addition, treatment time is of the utmost concern given the increased presence of radiosurgery in the modern radiation therapy clinic. Patients with multiple metastases would specifically benefit from HDRS, although very small lesions in the periphery of the head could benefit from separate isocenters to minimize rotational uncertainties, unless a target expansion could be clinically feasible, to account for those uncertainties. Targets located near bony interfaces or near the sinuses, for example, may also specifically benefit from the increased dose calculation accuracy offered by the Monte Carlo algorithm. This is a topic for future investigation. 10

...the results at this site are very promising as far as accurate, dynamic and efficient dose delivery is concerned. 5. Conclusions Radiosurgery of multiple metastases in a single isocenter is clearly more efficient than multiple isocenter dose delivery but is subject to increased areas for concern. While it remains to be seen if end-to-end delivery verification is universally accurate across multiple radiation therapy departments, (something to be verified in a multiple-institution study already underway), the results at this site are very promising as far as accurate, dynamic and efficient dose delivery is concerned. Disclaimer This publication is based on the experience and application of a medical expert, and is intended as an illustration of an innovative use of Elekta solutions. It is not intended to promote or exclude any particular treatment approach to the management of a condition. Any such approach should be determined by a qualified medical practitioner. It is important to note that radiation treatments, while usually beneficial, may cause side effects that vary depending on the clinical site being treated along with other medical circumstances. The most frequent side effects are typically temporary and may include, but are not limited to, skin redness and irritation, hair loss, respiratory, digestive, urinary or reproductive system irritation, rib, bone, joint or soft tissue (muscle) pain, fatigue, nausea and vomiting. In some patients, these side effects may be severe. Treatment sessions may also vary in frequency, complexity and duration. Finally, radiation treatments are not appropriate for all cancers, and their use along with the potential benefits and risks should be discussed before treatment. 11

Elekta Offices Elekta AB Box 7593 SE 103 93 Stockholm, Sweden T +46 8 587 254 00 F +46 8 587 255 00 Europe, Middle East, Africa T +46 8 587 254 00 F +46 8 587 255 00 North America T +1 770 300 9725 F +1 770 448 6338 Latin America, South America T +55 11 5054 4550 F +55 11 5054 4568 Asia Pacific T +852 2891 2208 F +852 2575 7133 Japan T +81 3 6722 3800 F +81 3 6436 4231 China T +86 10 5669 2800 F +86 10 5669 2900 elekta.com /elekta @elekta /company/ elekta Art. No. LPCOX180906 2018 Elekta AB (publ.) All mentioned trademarks and registered trademarks are the property of the Elekta Group. All rights reserved. No part of this document may be reproduced in any form without written permission from the copyright holder.