SimpleLine II Surgical / Prosthesis Manual

Similar documents
SuperLine & Implantium Surgical / Prosthesis Manual

SuperLine & Implantium Surgical / Prosthesis Manual

PRODUCT CATALOG. Introduction. Prosthetic Procedure 1 Solid / Dual / SCA Abutment 13. Prosthetic Procedure 3 Screw Abutment 23.

Abutment level impression Ⅰ. Combi abutment/dual abutment Multi units

SlimLine Product Catalog & Manual 1. Unit:mm, Scale 1: 1.5 / mm. SlimLine. Product Catalog / Manual

NR Line Product Catalog 1. NR Line Product/Manual Catalog

Legacy Prosthetic Catalog

NR Line Product Catalog

Hex-Lock Abutment System. Restorative Manual

Treatment Options for Restoring Edentulous Jaws using One- and Two-Piece Implants from Implant Direct Int l

SS Implant System 2013 PROSTHETIC PROCEDURE

4766 Research Dr. San Antonio, TX insightdentalsystems.com

NR Line Product/Manual Catalog

IMPRESSION PROCEDURES TRAINING MANUAL

US Implant System 2013 PROSTHETIC PROCEDURE

INSTRUCTIONS FOR THE DENTIST. synocta impression procedure, Screw-retained and Snap-on. Screw-retained (Open tray) Snap-on (Closed tray)

For the oral rehabilitation

SwissPlant Prosthetic Catalog. SwissPlant

cement-retained single crowns using cementable abutments

UDELL DENTAL LABORATORY Instructions for Use PREAT Precision Attachments

synocta Meso abutment for cement-retained restorations PROSTHETICS Step-by-step instructions

C O N T E N T S. Introduction of DASK. Crestal Approach. Lateral Approach. Dentium - For Dentists By Dentists

DASK-0912 [Rev.3] Distributed by. DASK_rev3_1127: 레이아웃 오? 4:53 페이지 1. Specifications are subject to change without prior notice.

Implant Phase Flow Chart

TS Implant System 2013 PROSTHETIC PROCEDURE

Dentium Implant System. New Product

- Easy solution for critical problems which may occur the prosthetics process - 5 tools in 1 kit (Screw Remover / Abutment Hex Remover / Screw Tap

IMPRESSION-TAKING, BITE REGISTRATION, AND TEMPORARY RESTORATION ON CAMLOG IMPLANTS. a perfect fit

Smarter Thinking. Simpler Design. Prima Plus. Surgical Manual

OT EQUATOR CASTABLE Single Attachment for Overdentures

Blue Sky Bio Multi unit Abutments BIO Multi-Units are available for all our Bone Level Implants

SlimLine CATALOG & MANUAL

prosthetic technique manual

NARROW DIAMETER implant

Straight Octagon Abutment for Tissue Level Implant

The LOCATOR concept. Simplicity and versatility for prosthesis fixation

Introduction of Dentium

User Manual Complete Prosthetic Kit (CPK)

CONTENTS. Introduction to DASK. Crestal Approach. Lateral Approach

PROSTHETIC COMPONENTS

a perfect fit INSTRUCTION MANUAL CAMLOG TORQUE WRENCH WITH CONTINUOUS TORQUE ADJUSTMENT Handling Sterilization Care

INTRODUCTION. Why Choose Cortex Prosthetics?

Element-Z Screw-Retained Hybrid

PROSTHETIC PROCEDURE for TS & GS Implant System

Smarter Thinking. Simpler Design. TiLobe Prosthetic Manual for PrimaConnex & Genesis implants

Height* FC width Prep depth RC width 4.0mm+ 4.3mm N/A 6.3mm

INDEX. Pag. 12. Pag. 05. Pag. 06. Pag. 13. Pag. 06. Pag. 14. Pag. 15. Pag. 15. Pag. 15. Pag. 15. Pag. 08. Pag. 16. Pag. 09. Pag.

UFII : Unique Fixture, 2 nd Generation. UFII Design Characteristics

The REAL Mini Implant

PRODUCT CATALOG. Since 1985» Simple. Predictable. Profitable.

Service Set Training Presentation English. André Breisinger BU Prosthetics

Part 2: Internal Tri-Channel Connection. Nobel Biocare Compatible Components. Tri-Channel Internal Connection Implants

ball abutment overdenture: chairside pick-up using existing denture

CHECK-UP Determination of Impression type & Method after implant placement

INTERNAL HEX Implant System

How something small, can make such a big difference!! Your best choice in overdenture

Ankylos. Acuris conometric concept. Manual and product catalog

HEXAGON Implant System

FIXED-HYBRID EMERGENCY! Why Redo When You Can Rescue?

Real World Implant Prosthetics: Fixed and Removable Samuel M. Strong, DDS

Solid Zirconia Full-Arch Implant Prosthesis (Protocol C All-CAD with Multi-Unit Abutments) BruxZir. FIRST Appointment. The BruxZir

- RESTORATIVE PRODUCT CATALOG

OD Secure chairside pick-up using existing denture

Prosthodontic Procedure

Very small abutment head easy and secure handling. Ankylos. The SmartFix concept. Prosthetic solution on angled implants

Restoring Straumann implants. with Locator abutments.

SwishPlant. SwishPlant System Catalog. 4.8mmD mmD. 4.8mmD mmD. 3.7mmD mmD. 6.5mmD

Product Catalog. Since 1985» Simple. Predictable. Profitable.

Step-by-Step Cemented Bridge Using Using CPK Abutments. Internal Hex. Implant System

NEW. The SmartFix Concept Prosthetic solution on tilted implants. SmartFix

DIGITAL WORKFLOWS IN THE PRIVATE PRACTICE TOURO DENTAL SEMINAR MARCH 9, 2018

Ankylos. Product Catalog

Prosthetic Guideline. Dental Implant System Tissue Level

verification jig fabrication

Multi-Unit Abutment Guide

Straumann synocta Prosthetic System Basic Information. Technical Information

CROWN AND BRIDGE RESTORATIONS WITH THE SYNOCTA PROSTHETIC SYSTEM

ß For cement-retained crowns and bridges. Note: synocta cementable abutments. ß Can be especially helpful with large multi-unit cases Note:

SIC invent. SIC P2F Abutment Dr. Galip Gurel From the gingiva shaper to a provisional crown to the final aesthetic restoration

Screw-retained restorations

LOCATOR ROOT ATTACHMENT SYSTEM

Multi-Unit Abutment System SIC Safe on Four. Optimum use of available bone by angled placement of implants

. Simplification. High Performance. Safety and Longevity. The fundamental goal is to restore the quality of life of patients.

Trabecular Metal and Tapered Screw-Vent Implant Systems. Restorative Manual

UDELL DENTAL LABORATORY Instructions for Use

Universal Plunger Loc Attachment

Guidance for. implant removal. Straumann Dental Implant System

screw-retained crown using the Laser-Lok Easy Ti temp abutment

Implant Restorations: A Step-By-Step Guide

Prosthetic Flow Diagrams for SS System

Preci Clix Instructions for Ball Retained Overdenture Attachments

surgical and prosthetic options & impression technique overview

MINI IMPLANT SYSTEM English Instructions for Use MINI IMPLANTS

PART 6. Implant Prosthodontics

Legacy TM. System Catalog. Legacy TM 1 Legacy TM 2 Legacy TM 3 Legacy TM 4

A Clinician s Guide to Restoring Conus

Introduction of Dentium Implant. Help kit

InterActive System Catalog

EXTERNAL HEX INTERNAL HEX NTERNAL. ENDOPORE Product Catalog

Transcription:

SimpleLine II Surgical / Prosthesis Manual

SimpleLine II

SURGICAL MANUAL Surgical Drill Sequence 04 Fixture Installation 05 Fixture Connection 06 Surgical Kit Maintenance 08 Warnings 09

04 SimpleLine ll Surgical / Prosthesis Manual Surgical Drill Sequence 81012 14 16 First guide drill Second guide drill Final drill Tap drill 3.4 3.8 4.3 4.8 3.4 3.8 4.3 4.8 Ø2.2 Ø2.6 Ø2.85 Ø3.3 Ø3.85 Ø4.4 Ø3.4 Ø3.8 Ø4.3 Ø4.8 Platform: Ø4.8 Body: Ø3.4 Platform: Ø4.8 Body: Ø3.8 Platform: Ø4.8 /Ø6.5 Body: Ø4.3 Platform: Ø6.5 Body: Ø4.8 D1~3 Bone

SimpleLine ll Surgical / Prosthesis Manual 05 Fixture Installation D1 Bone Platform: Ø6.5 / Body: Ø4.8 Guide & Final Drill: 800~1,200rpm / 30~45N. cm Tap Drill: 20~60rpm / 30~45N. cm Final drill Ø4.3 Final drill Ø4.8 Tap drill Ø4.8 Ø6.5 Full depth Ø4.8 D2, 3 Bone Platform: Ø6.5 / Body: Ø4.8 Final drill Ø4.3 Final drill Ø4.8 Tap drill Ø4.8 Ø6.5 5mm 10mm Half depth Ø4.8 D4 Bone Platform: Ø6.5 / Body: Ø4.8 Final drill Ø4.3 Final drill Ø4.8 Ø6.5 Ø4.8

06 SimpleLine ll Surgical / Prosthesis Manual Fixture Connection Caution_ When opening the fixture package, hold it upright to avoid falling out of the fixture. Securely engage the adapter with the fixture. By hand-piece 20rpm / 35N cm By ratchet Directions when Using the Hand-piece / Ratchet Adapter Hand-piece adapter Ratchet adapter Connection Hand-piece adapter Ratchet adapter The ratchet adapter and the fixture interal octa must be firmly connected together.

SimpleLine ll Surgical / Prosthesis Manual 07 Removal of the Fixture Mount Removal Mechanism of the Fixture Mount By handpiece Run the hand-piece in reverse mode. Rotate the mount screw in reverse. Fix the mount holder firmly to the mount. Fix the mount holder firmly to the mount. By ratchet Separate the fixture from the mount. Use the ratchet to rotate the mount in reverse. Fix the mount holder firmly to the mount. * In case the power of motor is insufficient to rotate the mount in reverse

08 SimpleLine ll Surgical / Prosthesis Manual Surgical Kit Maintenance Sterilization and Instrument Care Procedures Please follow legal regulations, as well as hygienic guidelines to prevent contamination and infection. Please remember that you are responsible for the maintenance and sterilization of your medical / dental products/devices. It is important to use and follow proper cleaning, disinfection and sterilization procedures. It is also important to follow the manufacture s recommendation on the usage of the drills.. Please keep a log as to the number of times the drills are used. Drills are used per implant placed not per patient. Bone density determines the longevity of the drills. Replace white and red o-rings on the adapters and the hex drivers, if worn and dried out. Drills should be considered for replacement after about 40 uses based on bone density. 01 All instruments, immediately after use, must be pre-soaked for a few minutes in a germicidal bath to loosen and prevent debris from attaching to instruments. Do not soak over-night. 02 Scrub with a soft brush to remove any debris. 03 For internal irrigation drills use a reamer or small gauge needle to cleanout drill internally. 04 If using an ultrasonic cleaner, wrap drills in a 2 x 2 gauze to prevent the drills from rubbing against each other. 05 Rinse thoroughly under warm water. 06 Clean all instrument trays with a germicidal cleaner prior to replacing instruments in kit. 07 Dry completely and place back into kit. 08 Always check for damage or corrosion after rinsing and drying. 09 Seal the tray in a sterilization pouch. 10 Sterilize using a steam autoclave at 121 C / 250F for 30 minutes or refer to manufacture s recommendations. 11 Store in a dry area at room temperature. Maintenance Period for Surgical Drills All surgical drills should be replaced after approximately 40 uses based on bone density Guide Drill (First, Second) (800~1200rpm, 30~45N cm with Irrigation) Final Drill (800~1200rpm, 30~45N cm with Irrigation) Tap Drill (20~30rpm, 30~45N cm with Irrigation)

SimpleLine ll Surgical / Prosthesis Manual 09 Warnings Warnings Dental implant surgery and restoration involve complex dental procedures. Appropriate and adequate training in proper technique is strongly recommended prior to use. Improper medical examination and / or treatment plan can result in implant failure and / or loss of supportive bone. Improper initial stability and / or excessive occlusal forces during healing period may lead to osseointergration failure. Excessive insertion torque may lead to a mechanical failure or a implant biologic failure due to bone compression and necrosis. When forces or loads are greater than its design, implant or abutment fracture could occur. Therefore clinicians should make careful decisions in regards to clinical treatment planning to minimize the risk of fracture. Appropriate implant quantity, occlusal interface and a nightguard are essential. Potential excessive loading conditions may include the following: 01 Inadequate number of implants are placed 02 Implant width and / or length are inappropriate for a treatment site 03 Prosthesis which has excessive cantilever length due to inadequate biomechanical design 04 Continuous occlusal force may be generated by incomplete connection between implant and abutment and / or abutment screw loosening 05 Direct casting abutment angles are greater than 30 from the vertical axis of the implant. Angled abutment is excessively milled. 06 Occlusal interferences causing excessive lateral forces 07 Patient parafunctions such as bruxism 08 Inadequate dental laboratory casting procedures 09 Improper prosthesis fit 10 Trauma from patient habits or accidents 11 Excessive marginal bone loss caused by inadequate bone width and / or advanced periimplantitis

PROSTHESIS MANUAL Types of Abutment 11 Prosthetic Procedure 1 12 Abutment Level- Solid / Dual Abutment 13 Abutment Level- SCA Abutment 16 Prosthetic Procedure 2 19 Fixture Level [Pick-up Type]- Dual Abutment 20 Fixture Level [Transfer Type]- Dual Abutment 23 Fixture Level- SCA Abutment 26 Fixture Level- Dual Milling Abutment 28 Fixture Level- Angled Abutment 30 Fixture Level- Direct Casting Abutment 32 Prosthetic Procedure 3 34 Abutment Level- Screw Abutment 35 Cementation Repair Method (SCRP) 38 Prosthetic Procedure 4 40 Positioner 41 Ball Attachment 43 Magnetic Attachment 45

SimpleLine ll Surgical / Prosthesis Manual 11 Types of Abutment One-Piece Two-Pieces Octa Octa Non-Octa Solid Abutment Dual Abutment SCA Abutment Octa 15 25 Octa Non-Octa Screw Abutment Dual Milling Abutment Angled Abutment Direct-Casting Abutment Ti Plastic Cylinder Abutment Level Octa Non-Octa Metal-Casting Abutment Octa Non-Octa Octa Non-Octa Fixture Level Temporary Abutment Dual Abutment Solid Abutment SCA Abutment Abutment level Positioner Ball Overdenture Magnetic Dual Abutment SCA Abutment Dual Milling Abutment Angled Abutment (15 / 25 ) Direct-Casting Abutment Metal-Casting Abutment Temporary Abutment Fixture level Screw Abutment Positioner Ball Magnetic Screw retained (Abutment level) For denture use

12 SimpleLine ll Surgical / Prosthesis Manual Prosthetic Procedure 1 Impression Technique and Restoration Selection Solid / Dual / SCA Abutment Abutment Level Impression Closed Tray Technique Solid Abutment Ø4.8 / Ø6.5 Dual Abutment Octa Ø4.8 / Ø6.5 SCA Abutment Octa / Non-octa Ø4.8 / Ø6.5 Impression Coping Transfer (Snap on) Ø4.8 / Ø6.5 Impression Coping Transfer (Snap on) Ø5.5 / Ø6.5 Analog Ø4.8 / Ø6.5 Analog Ø5.5 / Ø6.5 Burn-out Cylinder Single Non-Rotational Ø4.8 / Ø6.5 Bridge Rotational Ø4.8 / Ø6.5 Single Non-Rotational Ø5.5 / Ø6.5 Bridge Rotational Ø5.5 / Ø6.5 Modification Cemented Restoration

SimpleLine ll Surgical / Prosthesis Manual 13 Abutment Level- Solid / Dual Abutment [Multiple Units] Clinical Procedure Healing Abutment Solid Abutment or Dual Abutment Abutment Level Impression Provisional Restoration with or without Comfort Cap Comfort Cap Chairside Fixture installation. Choose solid abutment or dual abutment. Tighten the abutment with 25~30N cm and retighten it after 15 minutes. Affix the impression coping on the abutment. Apply the impression material. Take the impression. Image of the set final impression with impression coping). Place comfort cap over the abutment.

14 SimpleLine ll Surgical / Prosthesis Manual Abutment Level- Solid / Dual Abutment [Multiple Units] Laboratory Procedure Gold Crown Porcelain Crown Lab Analog Connection Cylinder Crown Wax-up Final Restoration Cementation Type Lab Side Insert analogs into the embedded impression coping. Make sure the analogs are securely locked into the impression coping (line up the flat side of analog to the flat side of the coping). Soft tissue modeling. Create the master model. Snap on the burn-out cylinders securely onto the analogs. Cut the cylinder after measuring proper height based on the proximity of the opposing teeth. Prepare for wax-up by affixing the plastic bar. Completion of wax-up. Produce the metal framework.

SimpleLine ll Surgical / Prosthesis Manual 15 Abutment Level- Solid / Dual Abutment [Multiple Units] Shave off the extended margin by using the rubber wheel. Metal framework and reamer. Use the reamer to eliminate the Lip created by the "snap-on" mechanism. Metal Framework after the removal of the "Lip". Metal framework. [Only Dual Abutment] Porcelain build-up. SCRP : Once an access hole has been created, it could be converted to a SCRP (Screw & Cement Retained Prosthesis). Final prosthesis. Create an access hole when the burn-out cylinder is used for the wax-up. Image of the extended margin around the metal framework. Shave off the extended margin by using the rubber wheel. Metal framework and reamer. Use the reamer to eliminate the Lip created by the "snap-on" mechanism. Metal framework after the removal of the "Lip". Metal framework. Final prosthesis.

16 SimpleLine ll Surgical / Prosthesis Manual Abutment Level- SCA Abutment [Multiple Units] Clinical Procedure Healing Abutment SCA Abutment Abutment Level Impression Provisional Restoration with or without Comfort Cap Comfort Cap Chairside Fixture installation. Choose SCA abutment (Octa / Non-octa). Tighten the abutment with 25~30N cm and retighten it after 15 minutes. Affix the impression coping on the abutment. Apply the impression material. Take the impression. Image of the set final impression with impression coping). Place comfort cap over the abutment.

SimpleLine ll Surgical / Prosthesis Manual 17 Abutment Level- SCA Abutment [Multiple Units] Laboratory Procedure Gold Crown Porcelain Crown Lab Analog Connection Cylinder Crown Wax-up Final Restoration Cementation Type Lab Side Insert analogs into the embedded impression coping. Make sure the analogs are securely locked into the impression coping (line up the flat side of analog to the flat side of the coping). Soft tissue modeling. Create the master model. Snap on the burn-out cylinders securely onto the analogs. Cut the cylinder after measuring proper height based on the proximity of the opposing teeth. Prepare for wax-up by affixing the plastic bar. Completion of wax-up. Produce the metal framework.

18 SimpleLine ll Surgical / Prosthesis Manual Abutment Level- SCA Abutment [Multiple Units] Shave off the extended margin by using the rubber wheel. Metal framework and reamer. Use the reamer to eliminate the Lip created by the "snap-on" mechanism. Metal Framework after the removal of the "Lip". Metal framework. Porcelain build-up. SCRP : Once an access hole has been created, it could be converted to a SCRP (Screw & Cemented Retained Prosthesis). Final prosthesis. Create an access hole when the burn-out cylinder is used for the wax-up. Image of the extended margin around the metal framework. Shave off of the extended margin by using the rubber wheel. Metal framework and reamer. Use the reamer to eliminate the Lip created by the "snap-on" mechanism. Metal framework after the removal of the "Lip". Metal framework. Final prosthesis.

SimpleLine ll Surgical / Prosthesis Manual 19 Prosthetic Procedure 2 Impression Technique and Restoration Selection Dual / SCA / Dual Milling / Angled / Direct-Casting / Metal-Casting / Temporary Abutment Fixture Level Impression Octa Non-octa Octa Non-octa Impression Coping / Pick-up Open Tray technique (Complicated case) Ø4.8 / Ø6.5 Impression Coping / Transfer Closed Tray technique (Simple case) Ø4.8 / Ø6.5 Analog Ø4.8 / Ø6.5 Dual Abutment Octa Ø4.8 / Ø6.5 SCA Abutment Octa / Non-octa Ø4.8 / Ø6.5 Dual Milling Abutment Octa Ø4.8 / Ø6.5 / Ø7.5 Angled Abutment Octa 15 / 25 Ø4.8 / Ø6.5 Direct-Casting Abutment Octa / Non-octa Ø4.8 / Ø6.5 Metal-Casting Abutment Octa / Non-octa Ø4.8 / Ø6.5 Temporary Abutment Octa / Non-octa Ti / Plastic Ø4.8 / Ø6.5 Modification Modification Cemented Restoration Screw-Retained Restoration

20 SimpleLine ll Surgical / Prosthesis Manual Fixture Level [Pick-up Type]- Dual Abutment [Multiple Units] Clinical Procedure Healing Abutment Impression Coping Pick-up Type Fixture Level Impression Open Tray Chairside Pick-up type impression coping (Octa). Apply adhesive on the open tray. (Individual tray) Apply the impression material. Take the impression. Remove the screw before removing the impression tray. Image of the set final impression with impression coping.

SimpleLine ll Surgical / Prosthesis Manual 21 Fixture Level [Pick-up Type]- Dual Abutment [Multiple Units] Laboratory Procedure Gold Crown Porcelain Crown Lab Analog Connection Assemble the Dual Abutment Burn-out Cylinder cutting for height adjustment Crown Wax-up Final Restoration Cementation Type Lab Side Connect analogs with the embedded impression coping. Soft tissue modeling. Create the master model. Assemble the dual abutment. If deemed necessary, abutment milling is possible. Fabricate the positioning jig Fabricate the cap with pattern resin. Completion of wax-up. Metal framework.

22 SimpleLine ll Surgical / Prosthesis Manual Fixture Level [Pick-up Type]- Dual Abutment [Multiple Units] Chairside Final prosthesis. Use positioning jig to transfer the abutment from the model to the intraoral and then tighten it with 25~30N cm. Re-tighten it after 15 minutes. Cement the final prosthesis and make occlusal adjustment. * In the process of seating the prosthesis, the components can be rebounded by gingival tissue. In that case, it is advised to apply occlusal load on the prosthesis for 10~15 minutes. SCRP- Lab Side Create an access hole for pick-up coping screw. Completion of Wax-up. Metal framework. SCRP- Chairside Final prosthesis. Use positioning jig to transfer the abutment from the model to the intraoral and tighten with to 25~30N cm. Re-tighten it after 15 minutes. Cement the final prosthesis and make occlusal adjustment. * In the process of seating the prosthesis, the components can be rebounded by gingival tissue. In that case, it is advised to apply occlusal load on the prosthesis for 10~15 minutes.

SimpleLine ll Surgical / Prosthesis Manual 23 Fixture Level [Transfer Type]- Dual Abutment [Multiple Units] Clinical Procedure Healing Abutment Impression Coping Transfer Type Fixture Level Impression Closed Tray Chairside Transfer type impression coping (Octa). Connect the impression coping for fixture level impression. Apply the impression material. Take the impression. Image of the set final impression with impression coping.

24 SimpleLine ll Surgical / Prosthesis Manual Fixture Level [Transfer Type]- Dual Abutment [Multiple Units] Laboratory Procedure Gold Crown Porcelain Crown Lab Analog Connection Assemble the Dual Abutment Burn-out Cylinder cutting for height adjustment Crown Wax-up Final Restoration Cementation Type Lab Side Insert analogs into the set impression. Make sure the analogs are securely seated in the impression coping (line up the flat side of analog to the flat side of the coping). Soft tissue modeling. Create the master model. Examine the soft tissue condition after the retrieval of the impression coping. Assemble the dual abutment. If deemed necessary, abutment milling is possible. Fabricate the positioning jig. Fabricate the cap with pattern resin.

SimpleLine ll Surgical / Prosthesis Manual 25 Fixture Level [Transfer Type]- Dual Abutment [Multiple Units] Chairside Completion of wax-up. Metal framework. Final prosthesis build-up on the framework with porcelain. Use the positioning jig to transfer the abutment from the model to the intraoral and tighten it with 25~30N cm. Re-tighten after 15 minutes. Cement the final prosthesis and make occlusal adjustment. Place wax into the opening of the abutment to protect the screw head prior to the composite sealing. SCRP- Lab Side Create an access hole for the pick-up coping screw. Completion of Wax-up. Metal framework. SCRP- Chairside Final prosthesis. Use positioning jig to transfer abutment from the model to the intraoral and tighten it with 25~30N cm. Re-tighten after 15 minutes. Cement the final prosthesis and make occlusal adjustment. Place wax into the opening of the abutment to protect the screw head prior to the composite sealing. * In the process of seating the prosthesis, the components can be rebounded by gingival tissue. In that case, it is advised to apply occlusal load on the prosthesis for 10~15 minutes.

26 SimpleLine ll Surgical / Prosthesis Manual Fixture Level- SCA Abutment [Multiple Units] Clinical Procedure Healing Abutment Fixture Level Impression Pick-up Type Open Tray Fixture Level Impression Transfer Type Closed Tray Laboratory Procedure Gold Crown Porcelain Crown Lab Analog Connection Assemble the SCA Abutment Burn-out Cylinder cutting for height adjustment Crown Wax-up Final Restoration Cementation Type Lab Side Connect analogs with the embedded impression coping. Soft tissue modeling. Create the master model.

SimpleLine ll Surgical / Prosthesis Manual 27 Fixture Level- SCA Abutment [Multiple Units] Assemble the SCA abutment. If deemed necessary, abutment milling is possible. Fabricate the positioning jig Fabricate the cap with pattern resin Chairside Completion of wax-up. Metal framework. Final prosthesis. Use positioning jig to transfer the abutment from the model to the intraoral and tighten it with 25~30N cm. Re-tighten it after 15 minutes. Cement the final prosthesis and make occlusal adjustment. Place wax into the opening of the abutment to protect the screw head prior to the composite sealing. SCRP- Lab Side * In the process of seating the prosthesis, the components can be rebounded by gingival tissue. In that case, it is advised to apply occlusal load on the prosthesis for 10~15 minutes. Create an access hole for pick-up coping screw Completion of wax-up. SCRP-Chairside Metal framework. Final prosthesis. Use positioning jig to transfer the abutment from the model to the intraoral and tighten it with 25~30N cm. Re-tighten it after 15 minutes. Cement the final prosthesis and make occlusal adjustment. Place wax into the opening of the abutment to protect the screw head prior to the composite sealing. * In the process of seating the prosthesis, the components can be rebounded by gingival tissue. In that case, it is advised to apply occlusal load on the prosthesis for 10~15 minutes.

28 SimpleLine ll Surgical / Prosthesis Manual Fixture Level- Dual Milling Abutment [Single Unit] Clinical Procedure Healing Abutment Fixture Level Impression Pick-up Type Open Tray Fixture Level Impression Transfer Type Closed Tray Laboratory Procedure Lab Analog Connection Dual Milling Abutment Connection Modification Crown Wax-up Final Restoration Cementation Type Lab Side Connect analogs with the set impression material. Soft tissue modeling. Create the master model.

SimpleLine ll Surgical / Prosthesis Manual 29 Fixture Level- Dual Milling Abutment [Single Unit] Assemble the dual milling abutment. Milled the abutment to an appropriate size. Fabricate the positioning jig Fabricate the cap with pattern resin. Completion of wax-up. Metal framework. Chairside Final prosthesis. Use positioning jig to transfer the abutment from the model to the intraoral and tighten it with 25~30N cm. Re-tighten it after 15 minutes. Cement the final prosthesis and make occlusal adjustment. Place wax into the opening of the abutment to protect the screw head prior to the composite sealing. * In the process of seating the prosthesis, the components can be rebounded by gingival tissue. In that case, it is advised to apply occlusal load on the prosthesis for 10~15 minutes.

30 SimpleLine ll Surgical / Prosthesis Manual Fixture Level- Angled Abutment [Single Unit] Clinical Procedure Healing Abutment Fixture Level Impression Pick-up Type Open Tray Fixture Level Impression Transfer Type Closed Tray Laboratory Procedure Lab Analog Connection Angled Abutment Connection Modification Crown Wax-up Final Restoration Cementation Type Lab Side Connect analogs with the set impression material. Soft tissue modeling. Unscrew and separate the impression from the model.

SimpleLine ll Surgical / Prosthesis Manual 31 Fixture Level- Angled Abutment [Single Unit] Create the master model. Assemble the angled abutment. Milled the abutment to an appropriate size and fabricate the positioning jig. Fabricate the cap with pattern resin. Completion of wax-up. Metal framework. Chairside Final prosthesis. Use positioning jig to transfer the abutment from the model to the intraoral and tighten it with 25~30N cm. Re-tighten it after 15 minutes. Cement the final prosthesis and make occlusal adjustment. Place wax into the opening of the abutment to protect the screw head prior to the composite sealing.

32 SimpleLine ll Surgical / Prosthesis Manual Fixture Level- Direct-Casting Abutment [Single Unit] Clinical Procedure Healing Abutment Fixture Level Impression Pick-up Type Open Tray Fixture Level Impression Transfer Type Closed Tray Laboratory Procedure Lab Analog Connection Direct-Casting Abutment Connection Modification Crown Wax-up Final Restoration Cementation Type Lab Side Connect analogs with the set impression material. Soft tissue modeling. Unscrew and separate the impression from the model.

SimpleLine ll Surgical / Prosthesis Manual 33 Fixture Level- Direct-Casting Abutment [Single Unit] Assemble the direct casting abutment. Completed customized abutment. Fabricate the positioning jig. Fabrication of pattern resin cap Completion of wax-up. Metal framework. Chairside Final prosthesis. Use positioning jig to transfer the abutment from the model to the intraral and tighten it with 25~30N cm. Re-tighten it after 15 minutes. Cement the final prosthesis and make occlusal adjustment. Place wax into the opening of the abutment to protect the screw head prior to the composite sealing. Fixture Level- Temporary Abutment [Multiple Units] <Using Ti Abutment> <Using Plastic Abutment> Ti-Temporary Abutment Plastic Temporary Abutment Consider the opposing teeth before seating the temporary abutment. Trim off the abutment as needed and complete the temporary abutment prosthesis with direct resin.

34 SimpleLine ll Surgical / Prosthesis Manual Prosthetic Procedure 3 Impression Technique and Restoration Selection Screw Abutment Abutment Level Impression Screw Abutment Ø4.8 / Ø6.5 Closed Tray Technique Impression Coping Transfer Hex / Non-hex Ø4.8 / Ø6.5 Impression Coping Pick-up Hex / Non-hex Ø4.8 / Ø6.5 Open Tray Technique Analog Ø4.8 / Ø6.5 Burn-out Cylinder Gold Cylinder Metal Cylinder Titanium Cylinder Ø4.8 / Ø6.5 Ø4.8 / Ø6.5 Ø4.8 / Ø6.5 Ø4.8 / Ø6.5 Bridge (Non-hex) Single (Hex) Bridge (Non-hex) Single (Hex) Bridge (Non-hex) Single (Hex) Bridge (Non-hex) Single (Hex) Polishing Protector Ø4.8 / Ø6.5 Temporary Restoration Ti-Retaining Screw Screw-Retained Restoration

SimpleLine ll Surgical / Prosthesis Manual 35 Abutment Level- Screw Abutment [Multiple Units] Clinical Procedure Healing Abutment Screw Abutment Impression Coping Transfer Type Abutment Level Impression Chairside Screw abutment and delivery holder. Select and seat an appropriate screw abutment with delivery holder. Tighten it with 25~30N cm. Re-tighten it after 15 minutes with screw abutment adapter. Screw abutment transfer copings (abutment level). Connect the impression coping for abutment level impression. Apply the impression material. Take the impression. Image of the set final impression with impression coping. Place comfort cap over the screw abutment.

36 SimpleLine ll Surgical / Prosthesis Manual Abutment Level- Screw Abutment [Multiple Units] Laboratory Procedure Lab Analog Connection Cylinder Connection Modification Crown Wax-up Final Restoration Screw Retained Lab Side Insert analogs into the set impression. Make sure the analogs are securely seated in the impression coping (line up the flat side of analog to the flat side of the coping). Soft tissue modeling. Create the master model. Remove the impression coping. Connect the screw abutment cylinder and tighten it with Ti-retaining screw. Trim cylinder after measuring proper height based on the proximity of the opposing teeth.. Connect the plastic bar in the middle of trimmed burn-out cylinders to help support the resin pattern. Resin pattern may experience shrinkages. Completion of wax-up.

SimpleLine ll Surgical / Prosthesis Manual 37 Abutment Level- Screw Abutment [Multiple Units] Gold framework. Use the reamer to remove the "Lip" in the interior of the metal framework. Completion of gold framework. Final prosthesis. Insert the final prosthesis and make necessary occlusal adjustments. Tighten it with Ti-retaining screw (10 N cm).

38 SimpleLine ll Surgical / Prosthesis Manual Cementation Repair Method (SCRP) [Screw & Cement Retained Prosthesis] In light of Implant Prosthesis: Screw type restoration simplifies prosthetic repair or insertion and removal of the prosthesis to any given situation. Cement type restoration tend to have a stable occlusion and may enhance the adaptability. However the weak point is, it cannot be removed after permanent cementation. A SCA abutment can be cemented or screw retained. Solid abutments are cement retained and no occlusal hole is necessary. Screw Loosening or Prosthesis Repair In case of the following: screw loosening or prosthesis repair In order to unscrew, create access hole on the occlusal surface with a bur. Unscrew, and remove the prosthesis from the patient's mouth. Both cemented prosthesis and abutments are removed. Finish the repair and seat it inside the patient's mouth. Tighten the prosthesis with 25~30N cm with a screw driver * In case of screw abutment connection, Ti-Retain screw has to be tightened with 10N cm. Place a small piece of cotton to cover the head of the screw. Fill the remaining access space with a resin. Final prosthesis.

SimpleLine ll Surgical / Prosthesis Manual 39 Cementation Repair Method (SCRP) [Screw & Cement Retained Prosthesis] Separation of Prosthesis with Abutment due to Cement Loss Remove the screw completely with 25~30N cm and remove prosthesis from the patient's mouth. Apply cement to the prosthesis. Place it back into the patient's mouth. * In case of screw abutment connection, Ti-Retain screw has to be tightened with) 10N cm. Unscrew and remove the excessive cement. Finish the repair and seat it inside the patient's mouth. Tighten the prosthesis with 25~30N cm with a screw driver. Augmenting Interproximal Volume to Repair Prosthesis Loosening Adding volume to the interproximal surface to repair loosening. Create access hole on the occlusal surface with a bur. Unscrew and remove the cemented prosthesis with abutment from the patient's mouth. Add resin to the prepared space on the contact surface. Screw back in the prosthesis and perform light curing. Aftermath, polish the contact surface. Position the prosthesis in the mouth and tighten the screw with 25~30N cm. Fill in the access hole.

40 SimpleLine ll Surgical / Prosthesis Manual Prosthetic Procedure 4 Impression Technique and Restoration Type Overdenture Procedure Positoner / Ball / Magnetic Attachment Positioner Ø4.8 Ball Abutment Ø4.8 / Ø6.5 Implant Keeper Ø4.8 / Ø6.5 Abutment Level Impression Positioner Impression Coping Ball Impression Coping Positioner Analog Ball Analog Positioner Block Out Spacer Socket Spacer Magnetic Assay Positioner Metal Socket / Plastic Socket Mini Denture Socket and O-ring Positioner Attachment for Overdenture Ball and Socket Attachment for Overdenture Magnetic Attachment for Overdenture

SimpleLine ll Surgical / Prosthesis Manual 41 Positioner Chairside Connect the Positioner Abutment onto the fixture. Affix the impression coping on the Positioner Abutment. Take impression for the production of the individual tray. Produce the individual tray for denture impression. Abutment and the impression coping together, apply the impression material. Take the final impression with the prepared individual tray. After the impression material is set, discard the individual tray. Image of the set final impression (with impression coping). Lab side Positioner Analog. Insert the Positioner Analog into the embedded impression coping. Create the master model. Block out procedure to achieve the space required for the metal socket. Fabrication of the denture with conventional method

42 SimpleLine ll Surgical / Prosthesis Manual Positioner Case 1 Chairside Secure spaces for the female sockets. Place the block out spacer on the Positioner Abutment in the patient s mouth. Connect the metal socket onto the Positioner Abutment. Apply a small amount of resin into the space created for the metal socket. Position the denture in the mouth and wait until the resin is completely set. Remove the white plastic socket (100gf) using the positioner tool and replace with a regular plastic socket of a desired retention force (300, 500 or 1000gf). Remove the denture after the resin is fully set. Image of the denture with the metal socket. Case 2 Remove the block out spacer from the patient s mouth. Chairside Polish and the overdenture is complete. Create holes for the placement of the metal sockets. Place the block out spacer on the Positioner Abutment in the intraoral. Connect the metal socket onto the Positioner Abutment. Examine for interference between the inner surface of the holes and the female sockets. Apply the resin into the holes and wait until it is completely set. Remove the white plastic socket (100gf) using the positioner tool and replace with a regular plastic socket of a desired retention force (300, 500 or 1000gf). Apply additional resin around the metal socket where there is a shortage of resin. Apply resin around the metal socket. Polish and the overdenture is complete.

SimpleLine ll Surgical / Prosthesis Manual 43 Ball Attachment Chairside Connect the Ball Abutment with the fixture. Affix the impression coping on the Ball Abutment. Take impression for the production of the individual tray. Produce the individual tray for denture impression. Apply the impression material. Take the final impression with the prepared individual tray. After the impression material is set, discard the individual tray. Image of the set final impression (with impression coping). Lab side Ball Analog. Insert the analogs into the embedded impression coping. Create the master model. Socket spacer. Fabrication of the denture with conventional method.

44 SimpleLine ll Surgical / Prosthesis Manual Ball Attachment Case 1 Chairside Secure spaces for the female sockets. Connect the female sockets to the Ball Abutment in the intraoral. Apply small amount of the resin into the secured area. Position the denture in the mouth and wait until the resin is completely set. Case 2 Female sockets are placed in the denture. Chairside Polish and the overdenture is complete. Create holes for the placement of the female sockets. Connect the female sockets to the Ball Abutment in the intraoral. Examine for interference between the inner surface of the holes and the female sockets. Apply the resin into the holes and wait until it is completely set. Place the female sockets. Apply resin around the female sockets. Polish and the overdenture is complete.

SimpleLine ll Surgical / Prosthesis Manual 45 Magnetic Attachment Chairside Remove the Healing Abutment. Connect implant keeper with the fixture and tighten it with 25~30 N cm. Implant keepers connected with the fixtures. Position the magnetic assay on the implant keeper. Case 1 Secure spaces for the magnetic assays. Examine for interference between inner divets of the denture and the magnets. Apply resin on the divets of the denture s inner surface. Position the denture into the mouth and wait until the resin is completely set. Position the denture into the mouth and wait for initial setting. Remove the denture and apply resin around the magnets. After the resin is completely set, remove excess. Polish and the overdenture is complete.

46 SimpleLine ll Surgical / Prosthesis Manual Magnetic Attachment Case 2 Create holes for the placement of the magnets. Examine for interference between the inner surface of the holes and the magnets..apply small amount of resin into the hole Position the denture in the mouth and wait until the resin is completely set. After initial setting, remove denture from the mouth. Add the resin around the magnets. Polish and the overdenture is complete.

Copyright April, 2013 DENTIUM

SimpleLine II Surgical / Prosthesis Manual Specifications are subject to change without prior notice. Some productsthat are to be launched in the market after necessary approvals are also listed in this catalog. HEAD OFFICE 3105 Trade Tower, 513 Yeongdong-daero, Gangnam-gu, Seoul, Korea 135-731 T +82-2-501-8560 F +82-2-567-9578 HOMEPAGE www.dentium.com SL2M-1304 [Rev.2]