Transcatheter aortic valve replacement with intracardiac echocardiography from the right internal jugular vein

Similar documents
TAVR: Echo Measurements Pre, Post And Intra Procedure

Boston, MA 2 Service Chief, Cardiovascular Imaging, Department of Radiology, Massachusetts General Hospital,

Adult Echocardiography Examination Content Outline

JOINT MEETING 2 Tricuspid club Chairpersons: G. Athanassopoulos, A. Avgeropoulou, M. Khoury, G. Stavridis

Streamlining a TAVR Procedure From screening to post TAVR care Hatim Al Lawati

New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology

Intracardiac EchoCardiography (ICE) Common Views

Zanobini et al. Journal of Cardiothoracic Surgery (2017) 12:114 DOI /s

Original Date: October 2009 TRANSESOPHAGEAL (TEE) ECHO Page 1 of 6

Pre-procedural CT angiography for Transcatheter Aortic Valve Implantation: What a Radiologist Needs to Know?

First Transfemoral Aortic Valve Implantation In Bulgaria - Crossing The Valve With The Device Is Not Always

National Imaging Associates, Inc. Clinical guidelines CARDIAC CATHETERIZATION -LEFT HEART CATHETERIZATION. Original Date: October 2015 Page 1 of 5

Imaging in TAVI. Jeroen J Bax Dept of Cardiology Leiden Univ Medical Center The Netherlands Davos, feb 2013

Edwards Sapien. Medtronic CoreValve. Inoperable FDA approved High risk: in trials. FDA approved

Transcatheter closure of patent foramen ovale using the internal jugular venous approach

PROSTHETIC VALVE BOARD REVIEW

Echo Assessment Pre-TAVI

Minimalist Transcatheter Aortic Valve Replacement (MA-TAVR)

Dr Winnie Sze-Wun Chan. Cardiac Team Deputy Team Head Department of Radiology and Imaging Queen Elizabeth Hospital Hong Kong

Title of image and video article Sub-Acute Leaflet Thrombosis: A Reversible Cause of Aortic Stenosis

*Core lab for numerous trials, for which I receive no direct compensation from sponsors.

Pediatric Echocardiography Examination Content Outline

Severity of AS Degree of AV calcification (? Bicuspid AV), annulus size, & aortic root

Interesting Cases - A Case Report: Renal Cell Carcinoma With Tumor Mass In IVC And Heart. O Wenker, L Chaloupka, R Joswiak, D Thakar, C Wood, G Walsh

Echocardiography as a diagnostic and management tool in medical emergencies

22/06/2017. Oxford City. Transcatheter aortic valve replacement 2017 guidelines. 1. First time I have heard about it. 2.

CIPG Transcatheter Aortic Valve Replacement- When Is Less, More?

Copyright 2017 American College of Emergency Physicians. All rights reserved.

Aortic valve implantation using the femoral and apical access: a single center experience.

Indication, Timing, Assessment and Update on TAVI

Intracardiac echocardiography: an ideal guiding tool for device closure of interatrial communications

Mitral Valve Disease, When to Intervene

INTEGRATING ECHOCARDIOGRAPHY WITH CATHETER INTERVENTIONS FOR CONGENITAL HEART DISEASE. Krishna Kumar SevenHills Hospital, Mumbai, India

The Role of Imaging in Transcatheter Aortic Valve Implantation

Aortic Stenosis: Interventional Choice for a 70-year old- SAVR, TAVR or BAV? Interventional Choice for a 90-year old- SAVR, TAVR or BAV?

PART II ECHOCARDIOGRAPHY LABORATORY OPERATIONS ADULT TRANSTHORACIC ECHOCARDIOGRAPHY TESTING

Chapter 24: Diagnostic workup and evaluation: eligibility, risk assessment, FDA guidelines Ashwin Nathan, MD, Saif Anwaruddin, MD, FACC Penn Medicine

Persistent left superior vena cava with absent right superior vena cava detected during emergent coronary artery bypass grafting surgery

Jian Fang 1, Shaobo Xie 2, Lunchao Ma 2, Chao Yang 2. Original Article

Percutaneous occluder device closure through femoral vein guidance by transthoracic echocardiography in adult atrial septal defect patients

TGA atrial vs arterial switch what do we need to look for and how to react

Igor Palacios, MD Director of Interventional Cardiology Massachusetts General Hospital Professor of Medicine Harvard Medical School

Atrial Septal Defects

A case of mislabeled textbooks: misnomer of the traditional "bicaval" view.

Structural Heart Disease: Setting the Stage for Success

For Personal Use. Copyright HMP 2013

Breakout Session: Transesophageal Echocardiography

The earliest application of intracardiac imaging

2019 Qualified Clinical Data Registry (QCDR) Performance Measures

TAVR Cases. Disclosures 2/17/2018. February 17, :15 3:30 PM 15 min

Case Report Sinus Venosus Atrial Septal Defect as a Cause of Palpitations and Dyspnea in an Adult: A Diagnostic Imaging Challenge

Menachem M. Weiner Assistant Professor of Anesthesiology Icahn School of Medicine at Mount Sinai

Case Report. Percutaneous Mitral Repair with MitraClip as an Adjunct Therapy of Heart Failure. Introduction. Case Report. Keywords

IMAGES. in PAEDIATRIC CARDIOLOGY

Echocardiographic Guidance During Placement of the Buttoned Double-Disk Device for Atrial Septa1 Defect Closure

Atrial Septal Defect Closure. Stephen Brecker Director, Cardiac Catheterisation Labs

Echocardiography Conference

Preprocedural evaluation for TAVR

A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision

Outline. EuroScore II. Society of Thoracic Surgeons Score. EuroScore II

Transcatheter aortic valve replacement in high risk patients with different anaesthetic techniques

University of Florida Department of Surgery. CardioThoracic Surgery VA Learning Objectives

Successful Transfemoral Edwards Sapien Aortic. Valve Implantation in a Patient with Previous. Mitral Valve Replacement

Transcatheter Aortic Valve Implantation Management of risks and complications

Percutaneous VSD closure

Transcatheter Closures for Fistula Tract and Paravalvular Leak after Mitral Valve Replacement and Tricuspid Annuloplasty

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics

Clinicians and Facilities: RESOURCES WHEN CARING FOR WOMEN WITH ADULT CONGENITAL HEART DISEASE OR OTHER FORMS OF CARDIOVASCULAR DISEASE!!

Ch.15 Cardiovascular System Pgs {15-12} {15-13}

ISUOG Basic Training. Obtaining & Interpreting Heart Views Correctly Alfred Abuhamad, USA. Basic training. Editable text here

Mitral Regurgitation

Syncope Due to Intracavitary Left Ventricular Obstruction Secondary to Giant Esophageal Hiatus Hernia

J Somerville and V Grech. The chest x-ray in congenital heart disease 2. Images Paediatr Cardiol Jan-Mar; 12(1): 1 8.

Multimodality Imaging of Anomalous Left Coronary Artery from the Pulmonary

Portico (St. Jude Medical Inc, St.

Appendix II: ECHOCARDIOGRAPHY ANALYSIS

Image Assistance in TAVI Why CT? Won-Jang Kim, MD, PhD Clinical Assistant Professor of Medicine, Heart Institute, Asan Medical Center, Seoul, Korea

Revealing new insights. irotate electronic rotation and xplane adjustable biplane imaging. Ultrasound cardiology. irotate and xplane

Diversion of the inferior vena cava following repair of atrial septal defect causing hypoxemia

Successful percutaneous treatment of late-onset femoral pseudoaneurysm after transcatheter, aortic valve implantation procedure

Index. interventional.theclinics.com. Note: Page numbers of article titles are in boldface type.

Transcatheter Aortic Valve Implantation Procedure (TAVI)

Biventricular Pacing - Hemodynamic Benefit for Patients with Congestive Heart Failure

COMPREHENSIVE EVALUATION OF FETAL HEART R. GOWDAMARAJAN MD

2 Brigham and Women s Hospital, Boston, MA.

HISTORY. Question: What category of heart disease is suggested by this history? CHIEF COMPLAINT: Heart murmur present since early infancy.

Optimal Imaging Technique Prior to TAVI -Echocardiography-

Percutaneous atrial septal defect closure with the Occlutech Figulla Flex ASD Occluder.

Cigna - Prior Authorization Procedure List Cardiology

ATRIAL SEPTAL CLOSURE AND LEFT ATRIAL APPENDAGE OCCLUSION: INDICATIONS AND GUIDANCE ECHOCARDIOGRAPHY IN INTERVENTIONAL CARDIOLOGY

(Ann Thorac Surg 2008;85:845 53)

An anterior aortoventriculoplasty, known as the Konno-

Certificate in Clinician Performed Ultrasound (CCPU) Syllabus. Rapid Cardiac Echo (RCE)

25 different brand names >44 different models Sizes mm

Bioprosthetic Mitral Valve Dysfunction: Innovation and Evolution of a New Therapeutic Technique

Pacing in patients with congenital heart disease: part 1

Left ventricular guidewire pacing for transcatheter aortic valve. implantation

Echocardiographic Cardiovascular Risk Stratification: Beyond Ejection Fraction

From the University of California Sulpizio Cardiovascular Center, San Diego, California.

Catheter Based Valve Interventions. Matthew Caldwell, MD February 5, 2019 Puerto Vallarta, Mexico

Transcription:

Case Report Transcatheter aortic valve replacement with intracardiac echocardiography from the right internal jugular vein Hiroto Yagasaki 1, Yoshiaki Goto 1, Yoshio Mori 2, Toshiyuki Noda 1 1 Department of Cardiology, 2 Department of Cardiovascular Surgery, Gifu Prefectural General Correspondence to: Hiroto Yagasaki, MD.. Email: y.pilottio@gmail.com. Abstract: Transesophageal echocardiography (TEE) is a useful imaging modality to guide transcatheter aortic valve replacement (TAVR). The limitations of TEE include the need for general anesthesia (GA) and endotracheal intubation. In Europe, TAVR under local anesthesia (LA) is usually performed without TEE. An intracardiac echocardiography (ICE) catheter may be used as an alternative to TEE, but the catheter is usually inserted from the femoral vein (FV-ICE). There are some reports of TAVR with FV-ICE; however, there are no reports of ICE inserted from the right internal jugular vein (JV-ICE). In the first case of its kind, we successfully performed TAVR under LA with JV-ICE. JV-ICE allows for the evaluation of perioperative hemodynamics, measurement of the aortic valve complex, and assessment of major complications during the procedure without interference from the operator or fluoroscopes; these represent a distinct advantage over TEE, transthoracic echocardiography, and FV-ICE. Moreover, there is no need for the echocardiologist to use maximal barrier precautions; the catheter can be operated in the standing position, like TEE; and operability is higher than that with FV-ICE. TAVR with JV-ICE is a promising alternative to TAVR under GA with TEE and TAVR under LA with FV-ICE. Keywords: Aortic valve stenosis; valvular disease; transcatheter aortic valve implantation; intracardiac echocardiography (ICE) Submitted Apr 05, 2018. Accepted for publication Jun 11, 2018. doi: 10.21037/cdt.2018.07.06 View this article at: http://dx.doi.org/10.21037/cdt.2018.07.06 Introduction Transcatheter aortic valve replacement (TAVR) is performed for patients who are considered moderate- or high-risk candidates for surgical aortic valve replacement (1,2). Transesophageal echocardiography (TEE) is a useful imaging modality to guide TAVR; however, TEE requires general anesthesia (GA) and endotracheal intubation. In Europe, TAVR is generally performed under local anesthesia (LA) without TEE (3-5). TAVR performed under LA has been shown to be comparable with TAVR performed under GA with respect to patient safety but offers the advantage of shorter hospital stay, ICU days, and procedure time (6,7). TAVR under GA is usually performed under the guidance of only fluoroscopy, transthoracic echocardiography (TTE), or intracardiac echocardiography (ICE). TEE and TTE probe may interfere with the fluoroscopic image, which makes it difficult to simultaneously evaluate the aortic valve and left ventricle with fluoroscopy. The ICE catheter does not interfere with the fluoroscopic image, so it can display any step of the procedure. The ICE catheter is usually inserted into the femoral vein (FV) and advanced to the right atrium. There are some reports of TAVR performed under the guidance of an ICE catheter inserted from the FV (FV-ICE) (8-11); however, there are no reports of TAVR performed under the guidance of an ICE catheter inserted from the right internal jugular vein (JV-ICE). During JV-ICE, the echocardiologist can operate without interference from the operator and the sheath of the femoral artery; moreover, the procedure does not require the echocardiologist to be clean. TAVR under

526 Yagasaki et al. TAVR with JV-ICE Figure 1 The arrow multilumen access catheter. The lumen a and b together have a diameter of 9 Fr, and the lumen c independently have a diameter of 12 Ga. Figure 2 Long-axis ascending aorta view: there was no injury of the ascending aorta. LA with JV-ICE may be an alternative to TAVR with TEE, TTE, and FV-ICE. We report the first case of TAVR with JV-ICE. Case presentation The patient was an 84-year-old woman with a history of arteriosclerosis obliterans, chronic gastritis, hiatal hernia, and constipation. She was asymptomatic at presentation. Her height was 147 cm, and her weight was 41 kg. Her STS (the Society of Thoracic Surgery) score was 3.472%; Clinical Frailty Score was 4. TTE revealed a left ventricular (LV) ejection fraction of 68%, an aortic valve peak velocity of 4.3 m/s, a mean gradient of 47 mmhg, and an aortic valve area of 0.65 cm 2. Electrocardiogram (ECG)-gated computed tomography revealed that the aortic valve area was 392 mm 2, the circumference of the aortic valve was 72.4 mm, and the major and minor diameters of the aortic valve were 24.1 and 19.2 mm, respectively. The height of the valsalva sinus was 15 mm, and the diameter of the valsalva sinus was 29 mm. Because the femoral artery and aorta were deemed adequate for a femoral artery approach, we planned to implant an Evolut R 29-mm valve (Medtronic, Minneapolis, Minnesota, US) via the transfemoral approach. TEE probe insertion was not suitable because of the hiatal hernia; therefore, the patient underwent LA and laryngeal mask airway intubation, and TAVR was performed under ICE guidance. There were two puncture sites of the patient, one was a 6-Fr 10-cm sheath (Radifocus Introducer II H; Terumo, Tokyo, Japan) which was inserted from the right internal JV as a temporary pacemaker, and the other was a 14-Fr 10-cm sheath (Arrow Multi-Lumen Access Catheter; Teleflex, Pennsylvania, US) which was inserted from the cranial side for the ICE catheter insertion and inotrope drip (Figure 1,S1,S2). ICE was performed with an 8-Fr AcuNav catheter (Siemens, Erlangen, Germany) with imaging on a Vivid Q ultrasound laptop (GE Healthcare, Buckinghamshire, UK), advanced from the right internal JV to the superior vena cava. On advancing the catheter by about 5 cm with counterclockwise rotation, the ascending aorta was obtained (Figure 2,S3). The ascending aorta was evaluated prior to the procedure. The atrial septum was obtained on advancing the catheter by a few cm (Figure 3,S4). There was no atrial septal defect or patent foramen ovale. With clockwise rotation, a long-axis view of the right atrium, tricuspid valve, and right ventricle was obtained. Preoperative tricuspid regurgitation and estimated right ventricular systolic pressure (RVSP) were measured (Figure 4A,B,S5). Pushing forward for the right ventricle with anterior flexion and counterclockwise rotation, a longaxis view of the left ventricle was obtained (Figure 5A,S6). Before the procedure, this view showed LV contraction and pericardial effusion, and during the procedure, it showed the position of the wire (Figure 5B). When the catheter was returned in the clockwise direction and pulled to the right atrium, a long-axis view of the aortic valve was obtained

Cardiovascular Diagnosis and Therapy, Vol 8, No 4 August 2018 527 Figure 3 Atrial septum view: this view may be less useful during transcatheter aortic valve replacement (TAVR) but may be useful during interventions for other structural heart diseases that require the Brockenbrough method. (Figure 6A). In this view, the preprocedural aortic valve flow velocity was measured (Figure 6B). The diameter of the aortic valve complex was measured (Figure 6C). After implantation of the valve, mild-to-moderate paravalvular aortic regurgitation from the edge of the non-coronary cusp was observed (Figure 6D,S7). However, due to the appearance of an atrioventricular block, we did not perform post-dilatation. Finally, we evaluated the changes in pericardial effusion, LV wall motion, RVSP, mitral valve regurgitation, postoperative aortic valve flow velocity, and ascending aortic aorta. The ICE catheter was removed, and the patient was transferred to the intensive care unit with two in situ sheaths and a temporary pacemaker. The operation time was 55 minutes, and the anesthesia time was 124 minutes. The volume of the contrast agent used A B Figure 4 Long-axis tricuspid view. (A) There was mild tricuspid regurgitation; (B) the tricuspid regurgitation pressure gradient was 18 mmhg; 4 mmhg of the central venous pressure was added, and the estimated right systolic ventricular pressure was 22 mmhg. A B Figure 5 Long-axis left ventricular view. (A) There was no left ventricular wall motion synergy or signs of pericardial effusion; (B) we could confirm the position of the wire during the procedure.

528 Yagasaki et al. TAVR with JV-ICE A B long-axis tricuspid view, long-axis LV view, and long-axis aortic valve view. As imaging modality of echocardiography, there are four procedures; TTE, TEE, FV-ICE, and JV-ICE. JV-ICE vs. TEE C D It was easy to evaluate the ascending aorta with JV-ICE, which is typically difficult to evaluate with TEE. The accurate evaluation of RVSP and aortic valve flow velocity is also difficult with TEE; however, it is relatively easy with JV-ICE because tricuspid regurgitation and aortic valve flow velocity can be obtained. Although we did not compare with TEE in this case, we could measure the aortic valve complex with JV-ICE in a view that was akin to that obtained with TEE. Therefore, JV-ICE allows for the evaluation of perioperative hemodynamics, measurement of the aortic valve complex, and assessment of major complications during the procedure. Figure 6 Long-axis aortic valve view. (A) A view akin to the 135 view during transesophageal echocardiography (TEE) was obtained; (B) the aortic valve flow velocity as measured by jugular vein-intracardiac echocardiography (JV-ICE) was 4.75 m/s. It is typically difficult to measure the flow velocity with TEE; however, JV-ICE was able to measure it for straight; (C) the basal ring was 23.3 mm, the valsalva sinus was 31.4 mm, the sinotubular junction diameter was 24.0 mm, and the ascending aorta diameter was 27.1 mm; (D) there was mild-to-moderate paravalvular aortic regurgitation from the edge of the non-coronary cusp. was 60 ml, and the fluoroscopy time was 16.1 minutes. The heart rhythm improved from pacing rhythm to sinus rhythm on the first postoperative day; therefore, the temporary pacemaker was removed on the second postoperative day. Subsequently, she developed a urinary tract infection and underwent antibiotic therapy. She was finally discharged on the 15 th postoperative day. Discussion This is the first documented case of TAVR performed under LA with JV-ICE. Using JV-ICE, we were able to obtain the following five views akin to those obtained with TEE: long-axis ascending aorta view, atrial septum view, JV-ICE vs. TTE and FV-ICE For JV-ICE, there is no need for the echocardiologist to be clean, and the operator can perform the procedure in the standing position akin to that with TEE; lastly, JV-ICE offers greater operability than that with FV-ICE. All these points indicate the superiority of JV-ICE over FV-ICE. JV-ICE vs. TTE, TEE, and FV-ICE JV-ICE can obtain images without interference from the operator or the fluoroscope, which represents a distinct advantage over TEE, TTE, and FV-ICE. There are four limitations of JV-ICE. First, during insertion of the ICE catheter from the right internal JV, it is necessary to insert three sheaths for (one each for the pacemaker, central venous catheter for inotrope, and ICE). To minimize the number of punctures, we used a 14-Fr sheath with a central venous catheter combined with ICE lumen in this case; however, this led to an increase in the thickness of the sheath. Second, the ICE catheter may interfere with the pacemaker, although this can also occur with FV-ICE. Third, there is a possibility of injury to the heart and vessels, and the risk of arrhythmia is higher than that with TEE and TTE. Fourth, the resolution afforded by JV-ICE is lower than that with TEE; therefore, there is a possibility of underestimating the complications and changes, particularly paravalvular aortic valve regurgitation.

Cardiovascular Diagnosis and Therapy, Vol 8, No 4 August 2018 529 In the case of LA, there were only options to perform with FV-ICE, TTE or without echocardiography. However, JV-ICE offers the above advantages over these procedures, and TAVR with JV-ICE may potentially minimize the complications of TAVR under LA and improve patient prognosis. Conclusions The use of JV-ICE allows the operator and echocardiologist to perform without interference from each other, and it is possible to depict the image without overlapping with the fluoroscopic apparatus. JV-ICE is a promising alternative to TEE, TTE, or FV-ICE for TAVR performed under LA. Acknowledgements None. Footnote Conflicts of Interest: The authors have no conflicts of interest to declare. Informed Consent: Written informed consent was obtained from the patient for publication of this manuscript and any accompanying images. References 1. Baumgartner H, Falk V, Bax JJ, et al. 2017 ESC/EACTS Guidelines for the management of valvular heart disease. Eur Heart J 2017;38:2739-91. 2. Nishimura RA, Otto CM, Bonow RO, et al. 2014 AHA/ ACC guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol 2014;63:e57-185. 3. Bufton KA, Augoustides JG, Cobey FC. Anesthesia for transfemoral aortic valve replacement in North America and Europe. J Cardiothorac Vasc Anesth 2013;27:46-9. 4. Di Mario C, Eltchaninoff H, Moat N, et al. The 2011-12 pilot European Sentinel Registry of Transcatheter Aortic Valve Implantation: in-hospital results in 4,571 patients. EuroIntervention 2013;8:1362-71. 5. Auffret V, Lefevre T, Van Belle E, et al. Temporal Trends in Transcatheter Aortic Valve Replacement in France: FRANCE 2 to FRANCE TAVI. J Am Coll Cardiol 2017;70:42-55. 6. Yamamoto M, Meguro K, Mouillet G, et al. Effect of local anesthetic management with conscious sedation in patients undergoing transcatheter aortic valve implantation. Am J Cardiol 2013;111:94-9. 7. Ehret C, Rossaint R, Foldenauer AC, et al. Is local anaesthesia a favourable approach for transcatheter aortic valve implantation? A systematic review and metaanalysis comparing local and general anaesthesia. BMJ Open 22017;7:e016321. 8. Bartel T, Bonaros N, Müller L, et al. Intracardiac echocardiography: a new guiding tool for transcatheter aortic valve replacement. J Am Soc Echocardiogr 2011;24:966-75. 9. Kadakia MB, Silvestry FE, Herrmann HC. Intracardiac echocardiography-guided transcatheter aortic valve replacement. Catheter Cardiovasc Interv 2015;85:497-501. 10. Bartel T, Müller S, Biviano A, et al. Why is intracardiac echocardiography helpful? Benefits, costs, and how to learn. Eur Heart J 2014;35:69-76. 11. Bartel T, Edris A, Velik-Salchner C, et al. Intracardiac echocardiography for guidance of transcatheter aortic valve implantation under monitored sedation: a solution to a dilemma? Eur Heart J Cardiovasc Imaging 2016;17:1-8. Cite this article as: Yagasaki H, Goto Y, Mori Y, Noda T. Transcatheter aortic valve replacement with intracardiac echocardiography from the right internal jugular vein.. doi: 10.21037/ cdt.2018.07.06

Supplementary Video 3. Video of atrial septum view Figure S1 Image of the right jugular venous access with sheath, temporary pacemaker wire and ICE catheter of another patient. ICE, intracardiac echocardiography. Figure S4 Video of atrial septum view (Figure 3) (14). Available online: http://www.asvide.com/article/view/26388 Video 1. Video demonstrating access and placement of the sheath, temporary pacemaker wire and ICE catheter of another patient Video 4. Video of long-axis tricuspid view Figure S2 Video demonstrating access and placement of the sheath, temporary pacemaker wire and ICE catheter of another patient (12). ICE, intracardiac echocardiography. Available online: http://www.asvide.com/article/view/26386 Figure S5 Video of long-axis tricuspid view (Figure 4) (15). Available online: http://www.asvide.com/article/view/26389 Video 5. Video of long-axis left ventricular view Video 2. Video of long-axis ascending aorta view Figure S6 Video of long-axis left ventricular view (Figure 5) (16). Available online: http://www.asvide.com/article/view/26390 Figure S3 Video of long-axis ascending aorta view (Figure 2) (13). Available online: http://www.asvide.com/article/view/26387

References Video 6. Video of long-axis aortic valve view Figure S7 Video of long-axis aortic valve view (Figure 6) (17). Available online: http://www.asvide.com/article/view/26391 12. Yagasaki H, Goto Y, Mori Y, et al. Video demonstrating access and placement of the sheath, temporary pacemaker wire and ICE catheter of another patient. Asvide 2018;5:680. Available http://www.asvide.com/article/ view/26386 13. Yagasaki H, Goto Y, Mori Y, et al. Video of long-axis ascending aorta view. Asvide 2018;5:681. Available online: http://www.asvide.com/article/view/26387 14. Yagasaki H, Goto Y, Mori Y, et al. Video of atrial septum view. Asvide 2018;5:682. Available online: http://www. asvide.com/article/view/26388 15. Yagasaki H, Goto Y, Mori Y, et al. Video of long-axis tricuspid view. Asvide 2018;5:683. Available online: http:// www.asvide.com/article/view/26389 16. Yagasaki H, Goto Y, Mori Y, et al. Video of long-axis left ventricular view. Asvide 2018;5;684. Available online: http://www.asvide.com/article/view/26390 17. Yagasaki H, Goto Y, Mori Y, et al. Video of long-axis aortic valve view. Asvide 2018;5:685. Available online: http:// www.asvide.com/article/view/26391