Performance Characteristics of Eight Estradiol Immunoassays

Similar documents
Comparison of Five Automated Serum and Whole Blood Folate Assays

Immunoassay of Estradiol: Unanticipated Suppression by Unconjugated Estriol

Intermethod Differences in Results for Total PSA, Free PSA, and Percentage of Free PSA

Chapter 2. High-sensitivity C-reactive protein methods examined

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

Determination of hemoglobin is one of the most commonly

Analytic Bias of Thyroid Function Tests. Analysis of a College of American Pathologists Fresh Frozen Serum Pool by 3900 Clinical Laboratories

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

ACCESS hstni SCIENTIFIC LITERATURE

Evaluation of the LIAISON 25 OH Vitamin D TOTAL Assay on the new LIAISON XL analyser

In Vitro Diagnostic Glucose Test System

Fertility Diagnostics

Establishment of Reference Intervals for Thyroid- Stimulating Hormone and Free Thyroxine in Amniotic Fluid Using the Bayer ADVIA Centaur

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

Endocrinology of the Female Reproductive Axis

Table. Analytical characteristics of commercial and research cardiac troponin I and T assays declared by the manufacturer.

Shyamali Pal. Page 315 ejifcc2017vol28no4pp R B Diagnostic Private Limited

The Virtues and Pitfalls of Implementing a New Test

Harmonisation of TFTs

Cortisol Assays. The Good, The Bad and The Indifferent. David Ducroq. Cardiff and Vale ulhb WEQAS. Unit 6, Parc Tŷ Glas. Llanishen.

Multicenter Comparison of Four Automated Immunoassay Analyzers for Prostate Specific Antigen

Analytic Performance of the iq200 Automated Urine Microscopy Analyzer and Comparison With Manual Counts Using Fuchs-Rosenthal Cell Chambers

Reproductive FSH. Analyte Information

N. Shirazian, MD. Endocrinologist

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Standardization and Certification of the ADVIA Centaur Vitamin D Total Assay

CASE STUDY 3: PSA-STANDARDISATION & VARIABILITY COMPARATIVE STUDY IN HOSPITALS

Ultra-Sensitive Estradiol lumelisa Catalog No. GWB-AEB745, legacy id (96 Tests)

Method Comparison Report Semi-Annual 1/5/2018

Toward Standardization of Insulin Immunoassays

ImmunoCAP Systems. Optimized for allergy testing - Precision - Accuracy - Consistency

lbt lab tests t Conrolled Ovarian Hyperstimulation Dr Soheila Ansaripour

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats

in vitro fertilization

Hormone Balance - Female Report SAMPLE. result graph based on Luteal Phase. result graph based on Luteal Phase

Lab Tests Made Simple

Serology and International units

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey

Evaluation of the UniCel TM DxI 800 Immunoassay Analyzer in Measuring Five Tumor Markers

Traditional approaches to Reference Intervals

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

MARTIN FLEISHER, PH.D., FACB MEMORIAL SLOAN-KETTERING CANCER CENTER NEW YORK, N.Y.

This supplementary material has been provided by the authors to give readers

Superovulation of Beef Heifers with Follicle Stimulating Hormone or Human Menopausal Gonadotropin: Acute Effects on Hormone Secretion

Photo (compulsory) Columbia University Medical Center, New York, NY 2. University of Southern California, Keck School of Medicine, Los Angeles, CA 3

Guide to Fulfillment of Measurement Uncertainty

Total Cost of Ownership (TCO): An evidence-based approach to compare laboratory equipment

Measurement of total thyroxine concentration in serum from dogs and cats by use of various methods Objective Sample Population Procedure Results

SCIEX Vitamin D 200M Assay for the Topaz System

Fish follicle-stimulating hormone, FSH ELISA Kit

METHOD VALIDATION: WHY, HOW AND WHEN?

Reproductive. Estradiol Analyte Information

Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

The Adiponectin Turbidimetric Immunoassay Reagent Kit

8.2 Principles of Quantitative Hematologic Determinations (1)

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800

Commentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee

Total Thyroxine ELISA (T4)

1. During the follicular phase of the ovarian cycle, the hypothalamus releases GnRH.

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer

Superovulation of Beef Heifers with Follicle Stimulating Hormone or Human Menopausal Gonadotropin: Acute Effects on Hormone Secretion

Rat Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA kit

Commentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee

Commercial insulin immunoassays fail to detect commonly prescribed insulin analogues

New York State Soluble Tumor Markers Proficiency Test

Analytical Variability Among Methods for the Measurement of 25-Hydroxyvitamin D. Still Adding to the Noise

All stocks and calibration levels were prepared in water: methanol (50:50) v/v to cover range of all steroid concentrations (refer Table 1).

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls

Elevated Urine Zinc Concentration Reduces the Detection of Methamphetamine, Cocaine, THC and Opiates in Urine by EMIT

International Journal of Pharma and Bio Sciences ENDOCRINE MARKERS AND DECLINE IN REPRODUCTIVE POTENTIAL OF WOMEN ABSTRACT

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

BNP Fragment EIA (Cat.No. BI-20852W) For the Determination of BNP Fragment in Human Samples

POTASSIUM/K. 3.6 mmol/l

For In Vitro Diagnostic Use

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

CMV DNA Quantification Using an Automated Platform for Nucleic Acid Extraction and Real- time PCR Assay Set-up

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

Reproductive. Estrone sulfate Analyte Information

Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer

Bias in clinical chemistry. Elvar Theodorsson EFLM and Linköping University

Evaluation of Cheongmeak DCS TM Reagents for Chemistry Analyzers

For in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date

Bio-Rad Laboratories. Cardiac Assessment Controls. Value Assigned for Clinical Laboratory and Point of Care Test Systems

Hormonal Control of Human Reproduction

Verification and validation of diagnostic laboratory tests in clinical virology

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

The effect of serum matrix and gender on cortisol measurement by commonly used immunoassays

HCG (human chorionic gonadotropin); Novarel Pregnyl (chorionic gonadotropin); Ovidrel (choriogonadotropin alfa)

Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration

Introduction. Etienne Cavalier*, Olivier Rousselle, Nunzio Ferrante, Agnes Carlisi, Caroline Le Goff and Jean-Claude Souberbielle.

MICROWELL ELISA LUTEINIZING HORMONE (LH) ENZYMEIMMUNOASSAY TEST KIT LH ELISA. Cat # 4225Z

Persistent Low Concentration of Human Chorionic Gonadotropin in a Nonpregnant Woman

Establishing Reference Intervals for Sex Hormones on the Analytical Platforms Advia Centaur and Immulite 2000XP

Cortisol (Sheep) ELISA Kit

Annual Review Contents. Page 1. Introduction, scheme-wide trends, future developments Insulin C-Peptide 8-10

ab Estradiol E2 Human ELISA Kit

Transcription:

Clinical Chemistry / EVALUATION OF EIGHT ESTRADIOL IMMUNOASSAYS Performance Characteristics of Eight Estradiol Immunoassays David T. Yang, MD, 1 William E. Owen, MT(ASCP), 2 Carol S. Ramsay, 3 Hui Xie, MS, 3 and William L. Roberts, MD, PhD, 1,2 Key Words: Gas chromatography; Mass spectrometry; Accuracy; Imprecision DOI: 1.139/5N2R4HT4GMAGPBY Abstract Measurement of estradiol is useful in assisted reproduction, evaluation of infertility, menopause, and male feminization. The analytic performance of 8 estradiol immunoassays was evaluated. The imprecision and accuracy of the Access, ADVIA Centaur, ARCHITECT i2, AutoDELFIA, Elecsys 21, IMMULITE 2, and Vitros ECi estradiol assays (see text for proprietary information) were evaluated by using an isotope dilution gas chromatography mass spectrometry (ID-GC-MS) reference method. The coefficient of variation (CV) ranged from 6.9% on the Elecsys 21 to 42.6% on the ADVIA Centaur at an estradiol concentration of 18 pg/ml (66 pmol/l), with the ARCHITECT i2 assay in development and the Vitros ECi having a CV below 1% at this estradiol concentration. Agreement between the automated assays and ID-GC-MS was variable, with slopes ranging from.87 to 1.2. The Access, ARCHITECT i2 in development, and the IMMULITE 2 were the most accurate, with slopes of.99,.98, and 1.3, respectively. These findings indicate that the ARCHITECT i2 estradiol assay in development had the best precision and accuracy of the assays evaluated for measurement of serum estradiol concentrations. Estradiol, a steroid hormone, is the most potent of the naturally occurring estrogens, and in the nonpregnant female, it is derived almost exclusively from the ovaries. Estradiol is responsible for the development and maintenance of the female sex organs and secondary sex characteristics. It also has important roles in regulation of the menstrual cycle and in maintenance of pregnancy. Estradiol measurements frequently are used during evaluation of female infertility and menopause and feminization in males and in assisted reproduction to monitor follicular development. 1 Physiologic levels are low, often less than 5 pg/ml (184 pmol/l), in men, children, and postmenopausal women and can range widely, from 5 to 41 pg/ml (18-1,55 pmol/l), in menstruating women. 2 The wide reference intervals and the clinical applications of this assay dictate the need for high sensitivity, excellent accuracy and precision through a wide analytical range, and a prompt turnaround time. Concern regarding these requirements, accuracy in particular, led to efforts toward standardization of serum estradiol measurements through a reference serum panel with estradiol concentrations determined by isotope dilution gas chromatography mass spectrometry (ID-GC-MS). 3,4 Recently, estradiol assays have become available on a number of fully automated immunoassay analyzers. While these systems provide rapid turnaround times and high throughput, there presently are only limited reports regarding the analytic performance of these assays, and most do not assess accuracy with ID-GC-MS as the reference method. 5-8 The aim of our study was to assess the imprecision and accuracy of 8 automated estradiol assays with ID-GC-MS as the reference method. 332 Am J Clin Pathol 24;122:332-337 332 DOI: 1.139/5N2R4HT4GMAGPBY

Clinical Chemistry / ORIGINAL ARTICLE Materials and Methods The following automated methods and instruments were evaluated: the Access analyzer (Beckman Coulter, Brea, CA), the ADVIA Centaur analyzer (Bayer Diagnostics, Tarrytown, NY), the ARCHITECT i2 analyzer using current reagents and reagents for a method that is in development (Abbott Diagnostics, Abbott Park, IL), the AutoDELFIA automatic immunoassay system (Wallac Oy, Turku, Finland), the Elecsys 21 analyzer (Roche Diagnostics, Indianapolis, IN), the IMMULITE 2 analyzer (Diagnostics Products, Los Angeles, CA), and the Vitros ECi analyzer (Ortho-Clinical Diagnostics, Raritan, NJ). All of these immunoassay methods use a competitive format. All methods use rabbit antiestradiol antibodies except for the Vitros ECi assay, which uses both rabbit and sheep antiestradiol antibodies. The ARCHITECT i2 in development assay requires 15 µl of sample compared with 1 µl for the current assay and has a comparable resistance to interference by hemolysis, icterus, and lipemia. The Elecsys 21 assay adds mesterolone to release estradiol from the sample. Assay imprecision was assessed by assaying aliquots of 5 serum pools with estradiol concentrations of 18, 31, 93, 387, and 846 pg/ml (66, 114, 341, 1,42, and 3,15 pmol/l, respectively) in 6 replicates each on 2 different days for a total of 12 measurements. The estradiol concentration of these samples was determined by ID-GC-MS as previously described. 9 Functional sensitivity, defined as the estradiol concentration that would produce a coefficient of variation (CV) of 1%, was determined from a power curve fit of the CV plotted against the estradiol concentration determined by ID-GC-MS. A method comparison study between each of the 8 estradiol assays and the ID-GC-MS reference method was performed using 61 samples with estradiol concentrations ranging from to 1,11 pg/ml (-4,41 pmol/l). EP Evaluator Release 5 software (David G. Rhoads Associates, Kennett Square, PA) was used for simple precision calculations and Deming regression analysis. For Deming regression analysis, the same SD was used for each method. Results A summary of data from the imprecision study is shown in Table 1. CVs ranged from 1.2% to 42.6% for estradiol concentrations from 18 to 846 pg/ml (66-3,15 pmol/l). At 18 pg/ml (66 pmol/l), the range of CV observed for each method was broad, ranging from 6.9% on the Elecsys 21 to 42.6% on the ADVIA Centaur. At 93 pg/ml, CVs became more uniform and ranged from 2.5% on the ARCHITECT i2 in development to 9.4% on the IMMULITE 2. The functional sensitivities for a cutoff CV of 1% were determined from the data and are listed in Table 2. The ARCHI- TECT i2 in development, Elecsys 21, and Vitros ECi are capable of determining estradiol concentrations as low as 18 pg/ml (66 pmol/l) while maintaining CVs of less than 1%. The remaining instruments had functional sensitivities ranging from 35 to 8 pg/ml (128-294 pmol/l). It is noteworthy that the mean estradiol concentrations obtained by some immunoassay methods for the sample with the lowest estradiol concentration were substantially higher than the ID- GC-MS reference method. Comparison of results obtained for samples collected from individual subjects using the 8 automated estradiol immunoassays and the reference ID-GC-MS method by Deming regression analysis is shown Figure 1. There were varying degrees of assay accuracy apparent, with slopes ranging from.87 to 1.2. The Access, ARCHI- TECT i2 in development, and IMMULITE 2 assays had the best accuracy with slopes of.99,.98, and 1.3 respectively. The AutoDELFIA and Elecsys 21 assays showed the poorest accuracy with slopes of.87 and 1.2, respectively. Difference plots for the same data are shown Figure 2. The dispersion of data around the regression line, as reflected in the standard error of the estimate (S y/x ) and the correlation coefficient (r) and represented graphically in the difference plot, showed that the ARCHITECT i2 assay in development had the best agreement with the reference method with an S y/x of 23 pg/ml (84 pmol/l), r =.997, followed by the Elecsys 21 method with an S y/x of 28 pg/ml (13 pmol/l) and r =.997. The ADVIA Centaur had the largest amount of dispersion with an S y/x of 66 pg/ml (242 pmol/l) and r =.98. Discussion Our protocol for assessing imprecision was abbreviated owing to limited quantities of serum assayed by the ID-GC- MS reference method. Therefore, the true overall imprecision of the estradiol immunoassays during multiple days of analysis and multiple reagent lots is likely to be substantially higher than our data would indicate. Nevertheless, our results reflect the relative imprecision of each method. Of the 8 estradiol immunoassays evaluated, the ARCHITECT i2 assay in development, Elecsys 21, and Vitros ECi methods were the most precise at low estradiol concentrations, being able to quantify estradiol concentrations as low as 18 pg/ml (66 pmol/l) while maintaining CVs of less than 1%. The Access, ARCHITECT i2 assay in development, and IMMULITE 2 had the best average results compared with the reference method, with Am J Clin Pathol 24;122:332-337 333 333 DOI: 1.139/5N2R4HT4GMAGPBY 333

Yang et al / EVALUATION OF EIGHT ESTRADIOL IMMUNOASSAYS Table 1 Summary of Imprecision Data * Estradiol Concentration (pg/ml) Method Serum Pool ID-GC-MS Method Mean Coefficient of Variation (%) Access 1 18 38.7 14.7 2 31 52.5 8. 3 93 98.2 9.3 4 387 422. 3.3 5 846 824.9 3.9 ADVIA Centaur 1 18 4.5 42.6 2 31 57.4 19.4 3 93 16.2 6.5 4 387 443.7 6.8 5 846 1,65.5 7.6 ARCHITECT i2 1 18 58.6 12. 2 31 6.3 15.7 3 93 119.8 4.8 4 387 458.3 2. 5 846 >1, Not determined ARCHITECT i2 in development 1 18 25.1 9.9 2 31 3.2 5.1 3 93 75.9 2.5 4 387 256.4 1.2 5 846 893.3 2.5 AutoDELFIA 1 18 22.9 31.4 2 31 32.7 2.3 3 93 9.9 5.8 4 387 361.2 3.7 5 846 718.9 4.3 Elecsys 21 1 18 28.1 6.9 2 31 35.5 8.9 3 93 91.6 4.9 4 387 52. 3.6 5 846 1,23.1 3.6 IMMULITE 2 1 18 28.4 14.7 2 31 2.9 21. 3 93 67.9 9.4 4 387 368.6 4.7 5 846 91.6 4.4 Vitros ECi 1 18 34.9 8.9 2 31 38.1 6.6 3 93 73.7 6.1 4 387 338. 2.5 5 846 673.3 3.1 ID-GC-MS, isotope dilution gas chromatography mass spectrometry. * Values are given in conventional units; for conversion to Système International units (pmol/l), multiply by 3.67. For proprietary information, see the text. Table 2 Estimated Functional Sensitivity * Assay Functional Sensitivity (pg/ml) Access 37 ADVIA Centaur 116 (61) ARCHITECT i2 35 ARCHITECT i2 in development <18 AutoDELFIA 74 (57) Elecsys 21 <18 IMMULITE 2 8 Vitros ECi <18 * The functional sensitivity was defined as the estradiol concentration that would produce a coefficient of variation (CV) of 1%. It was estimated from a power curve fit of the data in Table 1 as shown in Figure 1. For proprietary information, see the text. Inspection of the power curves in Figure 1 revealed that the concentration of estradiol corresponding to a CV of 1% was overestimated for the ADVIA Centaur and AutoDELFIA methods. The imprecision data for the 18, 31, and 93 pg/ml samples were refitted with a power curve, and the result shown in parentheses provides a better estimate of the true functional sensitivity of these assays. slopes very close to 1.. The remainder of the instruments showed varying degrees of bias and possibly could benefit from improved calibration. This is especially true for the Elecsys 21, which demonstrated excellent precision but had a positive proportional bias. Ovarian stimulation protocols require serial monitoring of estradiol levels to direct dosage adjustments of gonadotropins for optimization of follicular growth and prevention of ovarian overstimulation. 1,11 Most of these decision points are between estradiol concentrations of 2 and 2, pg/ml (734-7,34 pmol/l), 1 and all 8 instruments have adequate precision within this range with CVs of less than 1%. Before initiating an ovarian stimulation protocol, differentiating populations with poor stimulation 334 Am J Clin Pathol 24;122:332-337 334 DOI: 1.139/5N2R4HT4GMAGPBY

Clinical Chemistry / ORIGINAL ARTICLE A B C D Access (pg/ml) ADVIA Centaur (pg/ml) ARCHITECT i2 (pg/ml) ARCHITECT i2 in Development (pg/ml) E F G H AutoDELFIA (pg/ml) Elecsys 21 (pg/ml) IMMULITE 2 (pg/ml) Vitros ECi (pg/ml) Figure 1 Comparison of estradiol immunoassay and isotope dilution gas chromatography mass spectrometry (ID-GC-MS) results using Deming regression. The solid lines are from Deming regression analysis, and the dashed lines are x = y. Data are given for the assay compared with the ID-GC-MS reference method. A, Access: slope,.99; intercept, 21 pg/ml; standard error of the estimate (S y/x ), 38 pg/ml; r =.992. B, ADVIA Centaur: slope, 1.1; intercept, 9 pg/ml; S y/x, 66 pg/ml; r =.98. C, ARCHITECT i2: slope, 1.12; intercept, 41 pg/ml; S y/x, 4 pg/ml; r =.993. D, ARCHITECT i2 in development: slope,.98; intercept, 1 pg/ml; S y/x, 23 pg/ml; r =.997. E, AutoDELFIA: slope,.87; intercept, 17 pg/ml; S y/x, 32 pg/ml; r =.993. F, Elecsys 21: slope, 1.2; intercept, 1 pg/ml; S y/x, 28 pg/ml; r =.997. G, IMMULITE 2: slope, 1.3; intercept, 7 pg/ml; S y/x, 36 pg/ml; r =.993. H, Vitros ECi: slope,.92; intercept, 9 pg/ml; S y/x, 32 pg/ml; r =.993. GC-MS, gas chromatography mass spectrometry. Intercept and S y/x values are given in conventional units; for conversion to Système International units (pmol/l), multiply by 3.67. For proprietary information, see the text. characteristics from those with good stimulation characteristics is a prognostic necessity to select the appropriate stimulation protocol. This can be done by measuring basal levels of estradiol and follicle stimulating hormone on day 3 of the menstrual cycle, and the determination point is an estradiol concentration of 6 pg/ml (22 pmol/l). 12 All of the assays except the IMMULITE 2 demonstrated what we consider to be adequate precision at low estradiol concentrations for this task (CV <1%). While the Access, ARCHITECT i2 assay in development, and IMMULITE 2 assays demonstrated good concordance with the reference method, additional calibration standardization of the ADVIA Centaur, AutoDELFIA, Elecsys 21, and Vitros ECi assays might improve their accuracy and could lead to better harmonization of estradiol results. Additional studies to examine interfering substances, particularly endogenous and exogenous estrogens, need to be performed because the presence of some potential crossreacting substances could contribute to the bias and scatter observed for some of the methods. The clinical applications of estradiol assays can be diverse. This study shows that the ARCHITECT i2 assay in development, Elecsys 21, and Vitros ECi assays have adequate low-end precision for the measurement of estradiol levels in men and postmenopausal women, but the accuracy of the Elecsys 21 method needs some improvement. The rapid analysis time and low technical demands of automated immunoassays have made these instruments popular in the field of assisted reproduction. Within the range of estradiol concentrations typically encountered in this application, the Access, ARCHITECT i2 in development, and Vitros ECi assays seem to have adequate precision and accuracy. The Am J Clin Pathol 24;122:332-337 335 335 DOI: 1.139/5N2R4HT4GMAGPBY 335

Yang et al / EVALUATION OF EIGHT ESTRADIOL IMMUNOASSAYS A B C D Access 2 1 1 2 ADVIA Centaur 2 1 1 2 ARCHITECT i2 2 1 1 2 ARCHITECT i2 in Development 2 1 1 2 E F G H AutoDELFIA 2 1 1 2 Elecsys 21 2 1 1 2 IMMULITE 2 2 1 1 2 Vitros ECi 2 1 1 2 Figure 2 Comparison of estradiol immunoassays with isotope dilution gas chromatography mass spectrometry by difference plots. The solid lines are the ± 2 SD limits of agreement, and the dashed lines are the mean bias. All values are given in conventional units (pg/ml); for conversion to Système International units (pmol/l), multiply by 3.67. A, Access: mean bias, +2; limits of agreement, 55 and +95. B, ADVIA Centaur: mean bias, +34; limits of agreement, 14 and +172. C, ARCHITECT i2: mean bias, +7; limits of agreement, 32 and +173. D, ARCHITECT i2 in development: mean bias, 5; limits of agreement, 54 and +44. E, AutoDELFIA: mean bias, 17; limits of agreement, 122 and +89. F, Elecsys 21: mean bias, 48; limits of agreement, 79 and +176. G, IMMULITE 2: mean bias, ; limits of agreement, 72 and +72. H, Vitros ECi: mean bias, 29; limits of agreement, 113 and +54. GC-MS, gas chromatography mass spectrometry. For proprietary information, see the text. changes made to the reagents for the ARCHITECT i2 in development have resulted in an improvement in assay performance compared with the current ARCHITECT i2 reagents. The IMMULITE 2, while accurate, lacked adequate precision at low estradiol concentrations for this application, and, conversely, the Elecsys 21, while precise, lacked the requisite accuracy for this application. From the 1 Department of Pathology, University of Utah Health Sciences Center, Salt Lake City; 2 ARUP Institute for Clinical and Experimental Pathology, Salt Lake City; and 3 Abbott Diagnostics Division, Abbott Park, IL. Supported by the Diagnostics Division of Abbott Laboratories and the ARUP Institute for Clinical and Experimental Pathology. Address reprint requests to Dr Roberts: ARUP Laboratories, 5 Chipeta Way, Salt Lake City, UT 8418. References 1. Gronowski AM, Landau-Levine M. Reproductive endocrine function. In: Burtis CA, Ashwood ER, eds. Tietz Textbook of Clinical Chemistry. 3rd ed. Philadelphia, PA: Saunders; 1999:161-1641. 2. Painter PC, Cope JY, Smith JL. Reference information for the clinical laboratory. In: Burtis CA, Ashwood ER, eds. Tietz Textbook of Clinical Chemistry. 3rd ed. Philadelphia, PA: Saunders; 1999:1788-1846. 3. Thienpont L. Meeting report: first and second estradiol international workshops [letter]. Clin Chem. 1996;42:1122-1124. 4. Thienpont LM, De Leenheer AP. Efforts by industry toward standardization of serum estradiol-17β measurements. Clin Chem. 1998;44:671-674. 5. Levesque A, Letellier M, Dillon P, et al. Analytical performance of Bayer Immuno I estradiol and progesterone assays. Clin Chem. 1997;43:161-169. 336 Am J Clin Pathol 24;122:332-337 336 DOI: 1.139/5N2R4HT4GMAGPBY

Clinical Chemistry / ORIGINAL ARTICLE 6. Dancoine F, Couplet G, Buvat J, et al. Analytical and clinical evaluation of the Immulite estradiol assay in serum from patients undergoing in vitro fertilization: estradiol increase in mature follicles. Clin Chem. 1997;43:1165-1171. 7. Rodriguez-Espinoza J, Otal-Entraigas C, Gascon-Roche N, et al. Analytical and clinical performance of an automated immunoassay system (Immulite) for estradiol in serum. Clin Chem Lab Med. 1998;36:969-974. 8. Reinsberg J, Jost E. Analytical performance of the fully automated AxSYM estradiol assay. Clin Chem Lab Med. 2;38:51-55. 9. Wu H, Ramsay C, Ozaeta P, et al. Serum estradiol quantified by isotope dilution gas chromatography/mass spectrometry. Clin Chem. 22;48:364-366. 1. Maclin VM. Ovarian stimulation and ovulation induction. In: Keel BA, May JV, De Jonge CJ, eds. Handbook of the Assisted Reproduction Laboratory. Boca Raton, FL: CRC Press; 2:139-178. 11. Navot D, Bergh PA, Laufer N. Ovarian hyperstimulation syndrome in novel reproductive technologies: prevention and treatment. Fertil Steril. 1992;58:249-261. 12. Evers JL, Slaats P, Land JA, et al. Elevated levels of basal estradiol-17β predict poor response in patients with normal basal levels of follicle-stimulating hormone undergoing in vitro fertilization. Fertil Steril. 1998;69:11-114. Am J Clin Pathol 24;122:332-337 337 337 DOI: 1.139/5N2R4HT4GMAGPBY 337